A Flat Interface Nerve Electrode With Integrated Multiplexer
Open accessThe DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
One of the goals of peripheral nerve cuff electrode development is the design of an electrode capable of selectively activating a specific population of axons in a common nerve trunk. Several designs such as the round spiral electrode or the flat interface nerve electrode (FINE) have shown such ability. However, multiple contact electrodes require many leads, making the implantation difficult and potentially damaging to the nerve. Taking advantage of the flat geometry of the FINE, multiplexers were embedded within the cuff electrode to reduce the number of leads needed to control 32 channels. The circuit was implemented on a polyimide film using off-the-shelf electronic components. The electronic module was surface-mounted directly onto the electrode's flat substrate. Two circuit designs were designed, built, and tested: 1) a single supply design with only two wires but limited to cathodic-first pulse and 2) a dual-supply design requiring three lead wires but an arbitrary stimulation waveform. The electrode design includes 32 contacts in a 1 mm x 8 mm opening. The contact size is 300 microm x 400 microm with access resistance less than 1 k ohm. This electrode is not intended for long-term use, but developed as a feasibility study for future development using low-water-absorption materials such as liquid crystal polymer and an application specific integrated circuit.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · cited 1×
“epimysial [3], intramuscular [4], epi-neural [5]–[7], and intrafascicular electrodes [8]–[11].”
BackgroundFunctionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398 · cited 3×
“A flat interface nerve electrode (FINE) was developed to overcome this problem [16].”
BackgroundSelf-citationSelective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 359 · cited 2×
“epimysial [3], intramuscular [4], epi-neural [5]–[7], and intrafascicular electrodes [8]–[11].”
BackgroundSelective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · cited 3×
“Multicontact spiral cuff electrodes have also been shown to be selective in stimulating individual fascicles in a common nerve [6], [7].”
BackgroundMethodAn implantable multielectrode array with on-chip signal processing
Najafi, K., & Wise, K. D. (1986) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1986.1052646 · Cited by 310 · cited 1×
“Integrated circuits have been used to reduce the number of interconnecting leads for both central and peripheral nerve interfaces [24], [25].”
MethodSmith, B., Tang, Z., Johnson, M. W., Pourmehdi, S., Gazdik, M. M., Buckett, J. R., & Peckham, P. H. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.664202 · Cited by 270 · cited 1×
“lating waveform for the single-supply design is limited to a cathodic-first pulse followed by low-amplitude anodic pulse, this type of stimulating waveform has been used clinically for over a decade with no complications [27], [28].”
Electrodeposited iridium oxide for neural stimulation and recording electrodes
Meyer, R. D., Cogan, S. F., Nguyen, T. H., & Rauh, R. D. (2001) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/7333.918271 · Cited by 255 · cited 1×
“However, electrodeposition of iridium oxide would be required for nerve stimulation to increase the charge injection limit and prevent metal dissolution [31].”
BackgroundLago, N., Yoshida, K., Koch, K. P., & Navarro, X. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.886617 · Cited by 227 · cited 2×
“epimysial [3], intramuscular [4], epi-neural [5]–[7], and intrafascicular electrodes [8]–[11].”
BackgroundA nerve cuff technique for selective excitation of peripheral nerve trunk regions
Sweeney, J. D., Ksienski, D. A., & Mortimer, J. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55681 · Cited by 196 · cited 2×
“Multicontact spiral cuff electrodes have also been shown to be selective in stimulating individual fascicles in a common nerve [6], [7].”
BackgroundMethodSchiefer, M. A., Triolo, R. J., & Tyler, D. J. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2008.918425 · Cited by 148 · cited 5×
“flat area of the electrode is a suitable location for housing an electronic module, especially for a large nerve such as a femoral nerve that is 12 mm wide and 4 mm thick [17].”
BackgroundInfluentialRecruitment properties of monopolar and bipolar epimysial electrodes
Grandjean, P. A., & Mortimer, J. T. (1986) · Annals of Biomedical Engineering · doi:10.1007/bf02364648 · Cited by 134
Crago, P. E., Peckham, P. H., Mortimer, J. T., & Van Der Meulen, J. P. (1974) · Annals of Biomedical Engineering · doi:10.1007/bf02368496 · Cited by 133 · cited 1×
“Several electrode designs have been utilized to interface the stimulator with the biological tissue, such as surface [2],”
MethodSelectivity of multiple-contact nerve cuff electrodes: a simulation analysis
Choi, A. Q., Cavanaugh, J. K., & Durand, D. M. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.909637 · Cited by 121 · cited 2×
“Both computer modeling and animal studies suggest that the selectivity could be further improved by increasing the number of contacts around the nerve [18], [20]; however, that will also require more interconnecting leads, making the implantation more difficult.”
BackgroundSelf-citationInversion of the current-distance relationship by transient depolarization
Grill, W. M., & Mortimer, J. T. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.553708 · Cited by 109 · cited 1×
“over, arbitrary stimulating waveforms can be used to further enhance the spatial selectiveness of the electrode [29], without modifying the electrode.”
MethodA modeling study of nerve fascicle stimulation
Veltink, P. H., van Veen, B. K., Struijk, J. J., Holsheimer, J., & Boom, H. B. K. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32100 · Cited by 95
Subfascicle stimulation selectivity with the flat interface nerve electrode
Leventhal, D. K., & Durand, D. M. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1569266 · Cited by 81 · cited 2×
“Since only a few contacts in the electrode are used during selective stimulation [17], [20], the”
BackgroundSelf-citationMultielectrode intrafascicular and extraneural stimulation
Veltink, P. H., van Alsté, J. A., & Boom, H. B. K. (1989) · Medical and Biological Engineering and Computing · doi:10.1007/bf02442165 · Cited by 62 · cited 1×
“The electrode’s circular cross section, however, makes it difficult to activate selectively central axon populations [13]–[15].”
BackgroundChronic Measurement of the Stimulation Selectivity of the Flat Interface Nerve Electrode
Leventhal, D. K., & Durand, D. M. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.827535 · Cited by 55
Self-citationSelective motor unit recruitment via intrafascicular multielectrode stimulation
McDonnall, D., Clark, G. A., & Normann, R. A. (2004) · Canadian Journal of Physiology and Pharmacology · doi:10.1139/y04-047 · Cited by 53 · cited 1×
“While, intrafascicular electrodes, such as the Utah Slanted Electrode Array (USEA) [8], [9] and the longitudinal intrafascicular electrode (LIFE) [10], [11], have shown excellent selectivity, their reliability for longterm implant has yet to be validated.”
BackgroundChronic histological effects of the flat interface nerve electrode
Leventhal, D. K., Cohen, M., & Durand, D. M. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/2/004 · Cited by 43
Self-citation
Cited by
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133
Corrosion of silicon integrated circuits and lifetime predictions in implantable electronic devices
Vanhoestenberghe, A., & Donaldson, N. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/031002 · Cited by 140
Motor-commands decoding using peripheral nerve signals: a review
Hong, K.-S., Aziz, N., & Ghafoor, U. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aab383 · Cited by 63
“To solving the problem of circuit wiring, the electronic module was surfacemounted directly onto the electrode’s flat sub strate [158].”
MethodWu, Y., & Guo, L. (2020) · Neural Interface Engineering · doi:10.1007/978-3-030-41854-0_5 · Cited by 3
Metadata
- Authors
- Lertmanorat, Z., Montague, F. W., & Durand, D. M.
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2009-04
- Volume
- 17
- Issue
- 2
- Pages
- 176-182
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- OpenAlex
- W2133215897
- Cited by
- 18
- References
- 20
- FWCI
- 0.26
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.