Accuracy of Heart Rate Watches: Implications for Weight Management
Open accessAbstract
BACKGROUND: Wrist-worn monitors claim to provide accurate measures of heart rate and energy expenditure. People wishing to lose weight use these devices to monitor energy balance, however the accuracy of these devices to measure such parameters has not been established. AIM: To determine the accuracy of four wrist-worn devices (Apple Watch, Fitbit Charge HR, Samsung Gear S and Mio Alpha) to measure heart rate and energy expenditure at rest and during exercise. METHODS: Twenty-two healthy volunteers (50% female; aged 24 ± 5.6 years) completed ~1-hr protocols involving supine and seated rest, walking and running on a treadmill and cycling on an ergometer. Data from the devices collected during the protocol were compared with reference methods: electrocardiography (heart rate) and indirect calorimetry (energy expenditure). RESULTS: None of the devices performed significantly better overall, however heart rate was consistently more accurate than energy expenditure across all four devices. Correlations between the devices and reference methods were moderate to strong for heart rate (0.67-0.95 [0.35 to 0.98]) and weak to strong for energy expenditure (0.16-0.86 [-0.25 to 0.95]). All devices underestimated both outcomes compared to reference methods. The percentage error for heart rate was small across the devices (range: 1-9%) but greater for energy expenditure (9-43%). Similarly, limits of agreement were considerably narrower for heart rate (ranging from -27.3 to 13.1 bpm) than energy expenditure (ranging from -266.7 to 65.7 kcals) across devices. CONCLUSION: These devices accurately measure heart rate. However, estimates of energy expenditure are poor and would have implications for people using these devices for weight loss.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures


![Fig 3 Bland-Altman plots for device [(A) Apple Watch; N = 22, (B) Fitbit Charge HR; N = 22, (C) Samsung Gear S; N = 22, (D) Mio ALPHA; N = 22] and electrocardiography (reference) average heart rate (bpm). The solid line represents the mean difference (bpm) between the two measures and the dashed lines are the 95% limits of agreement (bpm). Notes: bpm = beats per minute, LoA = limits of agreement, MD = mean difference.](/paper-figures/ec7329ed55e24deba7fd6b3688025697/accuracy-of-heart-rate-watches-implications-for-weight-management-figure-3.jpg)
![Fig 4 Bland-Altman plots for device [(A) Apple Watch; N = 22, (B) Fitbit Charge HR; N = 22, (C) Samsung Gear S; N = 19, (D) Mio ALPHA; N = 22] and METAMAX (reference) total energy expenditure (kcal). The solid line represents the mean difference (kcal) between the two measures and the dashed lines are the 95% limits of agreement (kcal). Notes: kcal = kilocalories, LoA = limits of agreement, MD = mean difference.](/paper-figures/0ff18f7ebafc4b36bad2d5daa198824a/accuracy-of-heart-rate-watches-implications-for-weight-management-figure-4.jpg)
![Fig 5 Bland-Altman plots for device [(A) Apple Watch; N = 21, (B) Fitbit Charge HR; N = 21, (C) Samsung Gear S; N = 20] and direct observation (reference) steps. The solid line represents the mean difference (steps) between the two measures and the dashed lines are the 95% limits of agreement (steps). Notes: LoA = limits of agreement, MD = mean difference.](/paper-figures/cfa0e8fbaba64507b721381b5ea24223/accuracy-of-heart-rate-watches-implications-for-weight-management-figure-5.jpg)
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Similar to previous studies [13,28,30-32], we found reasonable accuracy in HR and EE estimations for these 2 devices under certain exercise conditions.”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [32]
Experimental — accuracy not yet measuredIn the citing paper
“Of those, all five found the Apple Watch to be more accurate than the other commercial devices [2, 9–12].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [12]
In the citing paper
“The remaining two studies showed mixed results.”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [12]
In the citing paper
“Wallen et al. noted lower heart rate accuracy in participants with a Fitzpatrick scale score > 4 with the Apple Watch, but not in the other studied smartwatches [12].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [12]
In the citing paper
“Although some studies have attempted to address wearable HR accuracy using the gold-standard reference method of ECG, which directly measures HR via electrical impulses from the heart and is used for clinical accuracy38–44”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [43]
In the citing paper
“with the Apple Watch having slightly greater accuracy than Fitbit devices (Apple Watch: −1.3 bpm, Fitbit: −9.3 bpm42–44)”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [43]
In the citing paper
“lower mean difference standard deviation43, and higher agreement with ECG than Fitbit devices39,44.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [43]
In the citing paper
“Evidence is starting to accumulate on the importance of skin tone when it comes to accuracy and generalizability of wrist-worn wearable studies with darker skin tone (and tattoos) being found to be associated with reduced wrist-worn HR accuracy as they absorb more green light67, yet this is not found in all studies43,50.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [43]
In the citing paper
“Inaccuracy of activity trackers may also lead individuals to overestimate their level of physical activity, limiting their effectiveness for lifestyle interventions [142,143].”
From Wearables and the Medical Revolution (2018) · cites this paper as [142]
In the citing paper
“Moreover, although the Apple Watch has become very popular, it has only been used in one validation study.”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [18]
In the citing paper
“The Apple Watch was found to underestimate both heart rate and EE, with a tendency to underestimate calories expended by more than 100 calories [18].”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [18]
In the citing paper
“Fitbit Charge HR has been found to underestimate heart rate and EE [18]”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [18]
In the citing paper
“Apple Watch has been previously found to underestimate calories by more than 100 calories [18], whereas our study found the device to overestimate calories expended.”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [18]
In the citing paper
“Most studies establishing validity only assess the overall physical activity intensity for an entire bout of physical activity [18].”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [18]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Health benefits of physical activity: the evidence
Warburton, D. E. R. (2006) · Canadian Medical Association Journal · doi:10.1503/cmaj.051351 · Cited by 8,007
HASKELL, W. L., LEE, I.-M., PATE, R. R., POWELL, K. E., BLAIR, S. N., FRANKLIN, B. A., MACERA, C. A., HEATH, G. W., THOMPSON, P. D., & BAUMAN, A. (2007) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e3180616b27 · Cited by 4,616
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339
DONNELLY, J. E., BLAIR, S. N., JAKICIC, J. M., MANORE, M. M., RANKIN, J. W., & SMITH, B. K. (2009) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e3181949333 · Cited by 2,658
Wearable Devices as Facilitators, Not Drivers, of Health Behavior Change
Patel, M. S., Asch, D. A., & Volpp, K. G. (2015) · JAMA · doi:10.1001/jama.2014.14781 · Cited by 808
Accuracy of Smartphone Applications and Wearable Devices for Tracking Physical Activity Data
Case, M. A., Burwick, H. A., Volpp, K. G., & Patel, M. S. (2015) · JAMA · doi:10.1001/jama.2014.17841 · Cited by 694
Ferguson, T., Rowlands, A. V., Olds, T., & Maher, C. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0201-9 · Cited by 468
Validity of Consumer-Based Physical Activity Monitors
LEE, J.-M., KIM, Y., & WELK, G. J. (2014) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000287 · Cited by 401
Prediction of energy expenditure from heart rate monitoring during submaximal exercise
Keytel, L. R., Goedecke, J. H., Noakes, T. D., Hiilloskorpi, H., Laukkanen, R. M. T., van der Merwe, L., & Lambert, E. V. (2005) · Journal of Sports Sciences · doi:10.1080/02640410470001730089 · Cited by 336
Methods of Prescribing Relative Exercise Intensity: Physiological and Practical Considerations
Mann, T., Lamberts, R. P., & Lambert, M. I. (2013) · Sports Medicine · doi:10.1007/s40279-013-0045-x · Cited by 333
Cited by
Sorted by citations
Items 101–200 of 336
Sprint, G., Cook, D. J., Weeks, D. L., Dahmen, J. B., & Fleur, A. L. (2017) · Sensors · doi:10.3390/s17102219 · Cited by 32
Wearable PZT Piezoelectric Sensor Device for Accurate Arterial Pressure Pulse Waveform Measurement
Li, M.-Y., Aoyama, J., Inayoshi, K., & Zhang, H. (2025) · Advanced Electronic Materials · doi:10.1002/aelm.202400852 · Cited by 31
Accuracy of a Wrist-Worn Heart Rate Sensing Device during Elective Pediatric Surgical Procedures
Pelizzo, G., Guddo, A., Puglisi, A., Silvestri, A. D., Comparato, C., Valenza, M., Bordonaro, E., & Calcaterra, V. (2018) · Children · doi:10.3390/children5030038 · Cited by 31
Role of Paced Breathing for Treatment of Hypertension
Cernes, R., & Zimlichman, R. (2017) · Current Hypertension Reports · doi:10.1007/s11906-017-0742-1 · Cited by 33
Accuracy of Consumer Monitors for Estimating Energy Expenditure and Activity Type
WOODMAN, J. A., CROUTER, S. E., BASSETT, D. R., FITZHUGH, E. C., & BOYER, W. R. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001090 · Cited by 29
Weatherall, J., Paprocki, Y., Meyer, T. M., Kudel, I., & Witt, E. A. (2018) · JMIR mhealth and uhealth · doi:10.2196/mhealth.8122 · Cited by 29
Mesfin, G., Saleme, E. B., Ademoye, O. A., Kani‐Zabihi, E., Santos, C. A. S., & Ghinea, G. (2020) · IEEE Transactions on Multimedia · doi:10.1109/tmm.2020.2992948 · Cited by 28
Deoni, S., D’Sa, V. A., Volpe, A. R., Beauchemin, J., Croff, J. M., Elliott, A., Pini, N., Lucchini, M., & Fifer, W. P. (2022) · Developmental Cognitive Neuroscience · doi:10.1016/j.dcn.2022.101059 · Cited by 29
Dorn, D., Gorzelitz, J., Gangnon, R. E., Bell, D. R., Koltyn, K. F., & Cadmus‐Bertram, L. (2019) · JMIR mhealth and uhealth · doi:10.2196/13547 · Cited by 28
CaRiSMA 1.0: Cardiac Risk Self-Monitoring Assessment
Agostinelli, A., Morettini, M., Sbrollini, A., Maranesi, E., Migliorelli, L., Nardo, F. D., Fioretti, S., & Burattini, L. (2017) · The Open Sports Sciences Journal · doi:10.2174/1875399x01710010179 · Cited by 27
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · JMIR mhealth and uhealth · doi:10.2196/14534 · Cited by 27
Humans interacting with multi-robot systems: a natural affect-based approach
Villani, V., Capelli, B., Secchi, C., Fantuzzi, C., & Sabattini, L. (2019) · Autonomous Robots · doi:10.1007/s10514-019-09889-6 · Cited by 26
de Arriba-Pérez, F., Santos-Gago, J. M., Rodríguez, M. C., & Merino, M. R. (2019) · Journal of Ambient Intelligence and Humanized Computing · doi:10.1007/s12652-019-01188-3 · Cited by 26
Commercial Smart Watches and Heart Rate Monitors: A Concurrent Validity Analysis
Montalvo, S., Martinez, A., Arias, S., Lozano, A., Gonzalez, M., Dietze-Hermosa, M., Boyea, B. L., & Dorgo, S. (2023) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000004482 · Cited by 28
Dry electrode made from carbon nanotubes for continuous recording of bio-signals
Jung, J., Shin, S., & Kim, Y. T. (2018) · Microelectronic Engineering · doi:10.1016/j.mee.2018.11.003 · Cited by 25
Huber, D. L., Thomas, D. G., Danduran, M., Meier, T. B., McCrea, M., & Nelson, L. D. (2019) · Journal of Athletic Training · doi:10.4085/1062-6050-93-18 · Cited by 23
O’Driscoll, R., Turicchi, J., Hopkins, M., Gibbons, C., Larsen, S. C., Palmeira, A. L., Heitmann, B. L., Horgan, G. W., Finlayson, G., & Stubbs, R. J. (2019) · Health and Technology · doi:10.1007/s12553-019-00392-7 · Cited by 22
Mahloko, L., & Adebesin, F. (2020) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-45002-1_9 · Cited by 23
A meta-analysis of Fitbit devices: same company, different models, different validity evidence
Leung, W. C. W., Case, L., Sung, M., & Jung, J. (2021) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2021.2006350 · Cited by 23
Bădău, D., & Bădău, A. (2018) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph15122614 · Cited by 23
Validity of the Apple Watch ® for monitoring push counts in people using manual wheelchairs
Karinharju, K. S., Boughey, A. M., Tweedy, S. M., Clanchy, K., Trost, S. G., & Gomersall, S. R. (2019) · Journal of Spinal Cord Medicine · doi:10.1080/10790268.2019.1576444 · Cited by 23
Self-citationMatsuura, H., Mukaino, M., Otaka, Y., Kagaya, H., Aoshima, Y., Suzuki, T., Inukai, A., Hattori, E., Ogasawara, T., & Saitoh, E. (2019) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-019-0140-x · Cited by 23
Improving Heart Rate Estimation on Consumer Grade Wrist-Worn Device Using Post-Calibration Approach
Choksatchawathi, T., Ponglertnapakorn, P., Ditthapron, A., Leelaarporn, P., Wisutthisen, T., Piriyajitakonkij, M., & Wilaiprasitporn, T. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2979191 · Cited by 25
The Validity and Reliability of the Mi Band Wearable Device for Measuring Steps and Heart Rate
Paradiso, C., Colino, F., & Liu, S. (2020) · International journal of exercise science · doi:10.70252/njhq9420 · Cited by 23
Validation of Electronic Activity Monitor Devices During Treadmill Walking
Tam, K. M., & Cheung, S. Y. (2018) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2017.0263 · Cited by 22
Taylor, J. L., Keating, S. E., LEVERITT, M. D., Holland, D. J., Gomersall, S. R., & Coombes, J. S. (2017) · Contemporary Clinical Trials Communications · doi:10.1016/j.conctc.2017.10.002 · Cited by 22
Self-citationMeasuring Moderate-Intensity Exercise with the Apple Watch: Validation Study
Abt, G., Bray, J., & Benson, A. C. (2018) · JMIR Cardio · doi:10.2196/cardio.8574 · Cited by 21
Sun, J., & Liu, Y. (2020) · JMIR mhealth and uhealth · doi:10.2196/17699 · Cited by 21
HRS White Paper on Clinical Utilization of Digital Health Technology
Wan, E. Y., Ghanbari, H., AKOUM, N. W., Attia, Z. I., Asirvatham, S. J., Chung, E. H., Dagher, L., Al‐Khatib, S. M., Mendenhall, G. S., McManus, D. D., et al. (2021) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2021.07.001 · Cited by 21
Evaluation of Commercial-Off-The-Shelf Wrist Wearables to Estimate Stress on Students
de Arriba-Pérez, F., Santos-Gago, J. M., Rodríguez, M. C., & Iglesias, M. J. F. (2018) · Journal of Visualized Experiments · doi:10.3791/57590 · Cited by 21
Digital phenotyping to improve prediction of suicidal urges in treatment: Study protocol
Brown, L. A., Taylor, D. J., Bryan, C., Wiley, J. F., Pruiksma, K., Khazem, L., Baker, J. C., Young, J., & O'Leary, K. (2022) · Aggression and Violent Behavior · doi:10.1016/j.avb.2022.101733 · Cited by 20
Lakkireddy, D. (2020) · Journal of Atrial Fibrillation · doi:10.4022/jafib.20200446 · Cited by 20
Hunter, A., Leckie, T., Coe, O., Hardy, B., Fitzpatrick, D., Gonçalves, A., Standing, M.-K., Koulouglioti, C., Richardson, A., & Hodgson, L. E. (2022) · JMIR Rehabilitation and Assistive Technologies · doi:10.2196/25494 · Cited by 23
Kasparian, A. M., & Badawy, S. M. (2022) · mHealth · doi:10.21037/mhealth-21-28 · Cited by 19
Hendrikx, J., Ruijs, L. S., Cox, L. G. E., Lemmens, P. M. C., Schuijers, E. G., & Goris, A. H. C. (2017) · JMIR mhealth and uhealth · doi:10.2196/mhealth.6893 · Cited by 19
Dementyev, A., Hernandez, J., Choi, I., Follmer, S., & Paradiso, J. A. (2018) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3264912 · Cited by 20
Rodriguez-Pulido, F., Castillo-Sánchez, G., Hamrioui, S., Martin, L. D., Vazquez-Beltrán, P., Díez, I. D. L. T., & Franco, M. (2020) · Journal of Medical Systems · doi:10.1007/s10916-020-1543-7 · Cited by 21
Heart rate as a predictor of cardiovascular risk
Seravalle, G., Trevano, F. Q., & Grassi, G. M. (2021) · Minerva Medica · doi:10.23736/s0026-4806.20.06695-1 · Cited by 20
Herkert, C., Kraal, J. J., van Loon, E. M. A., van Hooff, M., & Kemps, H. M. C. (2019) · JMIR mHealth and uHealth · doi:10.2196/15045 · Cited by 19
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21
Evaluation of physical health status beyond daily step count using a wearable activity sensor
Xu, Z., Zahradka, N., Ip, S., Koneshloo, A., Roemmich, R. T., Sehgal, S., Highland, K. B., & Searson, P. C. (2022) · npj Digital Medicine · doi:10.1038/s41746-022-00696-5 · Cited by 19
Støve, M. P., & Hansen, E. C. K. (2022) · Journal of Sports Sciences · doi:10.1080/02640414.2023.2180160 · Cited by 19
Hänsel, K., Poguntke, R., Haddadi, H., Alomainy, A., & Schmidt, A. (2018) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3173574.3173719 · Cited by 19
Validity of the Polar Vantage M watch when measuring heart rate at different exercise intensities
Shumate, T., Link, M., Furness, J. W., Kemp‐Smith, K., Simas, V., & Climstein, M. (2021) · PeerJ · doi:10.7717/peerj.10893 · Cited by 19
Sensing Physiological Change and Mental Stress in Older Adults From Hot Weather
Cheong, S., Isaza, C. A. B., & Ortiz, L. (2020) · IEEE Access · doi:10.1109/access.2020.2982153 · Cited by 18
Koyama, S., Fujimoto, A., Yasuda, Y., & Satou, Y. (2020) · Sensors · doi:10.3390/s20247076 · Cited by 18
Accuracy of wrist-worn wearable devices for determining exercise intensity
Ho, W.-T., Yang, Y.-J., & Li, T.-C. (2022) · Digital Health · doi:10.1177/20552076221124393 · Cited by 19
A Validation Study to Confirm the Accuracy of Wearable Devices Based on Health Data Analysis
Hrabovska, N., Kajáti, E., & Zolotová, I. (2023) · Electronics · doi:10.3390/electronics12112536 · Cited by 21
Evaluation of Earbud and Wristwatch Heart Rate Monitors during Aerobic and Resistance Training
Bunn, J. A., Wells, E. K., Manor, J., & Webster, M. J. (2019) · International journal of exercise science · doi:10.70252/loox1045 · Cited by 18
Lee, E. S., Lee, J. S., Joo, M. C., Kim, J. H., & Noh, S. E. (2017) · Annals of Rehabilitation Medicine · doi:10.5535/arm.2017.41.1.129 · Cited by 17
van der Stam, J. A., Mestrom, E. H. J., Scheerhoorn, J., Jacobs, F. E. N. B., Nienhuijs, S. W., Boer, A., van Riel, N. A. W., de Morree, H. M., Bonomi, A., Scharnhorst, V., et al. (2023) · JMIR Perioperative Medicine · doi:10.2196/40474 · Cited by 16
Ibrahim, N. S., Rampal, S., Lee, W. L., Pek, E. W., & Suhaimi, A. (2023) · Cardiovascular Engineering and Technology · doi:10.1007/s13239-023-00693-z · Cited by 16
Guidoux, R., Duclos, M., Fleury, G., Lacomme, P., Lamaudière, N., Saboul, D., Ren, L., & Rousset, S. (2017) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2017.04.005 · Cited by 17
Lee, J.-E., Lee, D. H., Oh, T. J., Kim, K. M., Choi, S. H., Lim, S. K., Park, Y. J., Park, D. J., Jang, H. C., & Moon, J. H. (2018) · JMIR Research Protocols · doi:10.2196/resprot.8119 · Cited by 19
Diabetes Technology and Exercise
Riddell, M. C., Pooni, R., Fontana, F. Y., & Scott, S. N. (2019) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2019.10.011 · Cited by 18
Congruent Accuracy of Wrist-worn Activity Trackers during Controlled and Free-living Conditions
Dondzila, C. J., Lewis, C., Lopez, J. R., & Parker, T. (2018) · International Journal of Exercise Science · doi:10.70252/ubbt6886 · Cited by 16
“Furthermore, there is a dearth of research on the Mio devices for any physical activity metric, other than heart rate (4, 20, 24, 28).”
BackgroundResultInfluentialAffective Change With Variations in Zumba Fitness Intensity as Measured by a Smartwatch
Lee, J., Park, J., Kim, Y., & Woo, M. (2021) · Perceptual and Motor Skills · doi:10.1177/00315125211022700 · Cited by 16
Guha, A., Xiang, X., Haddad, D. A., Buck, B. H., Gao, X., Dunleavy, M. P., Liu, E. Y., Patel, D., Fedorov, V. V., & Daoud, E. G. (2017) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.13248 · Cited by 16
Weeks, D. L., Sprint, G., Stilwill, V., Meisen-Vehrs, A. L., & Cook, D. J. (2018) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2017.0306 · Cited by 16
Tang, W.-H., Ho, W., & Chen, Y. J. (2018) · BioMedical Engineering OnLine · doi:10.1186/s12938-018-0574-5 · Cited by 15
Exercise training in paediatric congenital heart disease: fit for purpose?
Amir, N. H., Dorobantu, D. M., Wadey, C. A., Caputo, M., Stuart, G., Pieles, G. E., & Williams, C. A. (2021) · Archives of Disease in Childhood · doi:10.1136/archdischild-2020-321390 · Cited by 17
Instantaneous VO2 from a wearable device
Cook, A. J., Ng, B., Gargiulo, G. D., Hindmarsh, D., Pitney, M., Lehmann, T., & Hamilton, T. J. (2018) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2017.12.008 · Cited by 16
Assessing the ability of the Fitbit Charge 2 to accurately predict VO2max
Freeberg, K. A., Baughman, B. R., Vickey, T., Sullivan, J. A., & Sawyer, B. J. (2019) · mHealth · doi:10.21037/mhealth.2019.09.07 · Cited by 16
The accuracy of Apple Watch measurements: a living systematic review and meta-analysis
Lambe, R., Baldwin, M., O’Grady, B., Schumann, M., Caulfield, B., & Doherty, C. (2026) · npj Digital Medicine · doi:10.1038/s41746-025-02238-1 · Cited by 17
The Role of Uncertainty as a Facilitator to Reflection in Self-Tracking
Alqahtani, D. A., Jay, C., & Vigo, M. (2020) · Designing Interactive Systems Conference · doi:10.1145/3357236.3395448 · Cited by 15
Jagim, A. R., Koch-Gallup, N., Camic, C. L., KROENING, L., Nolte, C., SCHROEDER, C., Gran, L., & Erickson, J. L. (2021) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.20.11151-4 · Cited by 16
Comparing the Validity of a GPS Monitor and a Smartphone Application to Measure Physical Activity
Adamakis, M. (2017) · Journal of Mobile Technology in Medicine · doi:10.7309/jmtm.6.2.4 · Cited by 15
Accuracy of swimming wearable watches for estimating energy expenditure
Lee, M., Lee, H., & Park, S. (2018) · IJASS(International Journal of Applied Sports Sciences) · doi:10.24985/ijass.2018.30.1.80 · Cited by 15
Accuracy of Heart Rate Measurements by a Smartwatch in Low Intensity Activities
Spinsante, S., Porfiri, S., & Scalise, L. (2019) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2019.8802216 · Cited by 15
Khushhal, A. A., Abdelaal, A. A. M., & Elsayed, M. (2024) · Journal of Multidisciplinary Healthcare · doi:10.2147/jmdh.s449071 · Cited by 14
The Accuracy and Validity of iOS-Based Heart Rate Apps During Moderate to High Intensity Exercise
Bouts, A. M., Brackman, L., Martin, E., Subasic, A. M., & Potkanowicz, E. S. (2018) · International journal of exercise science · doi:10.70252/qwrt8315 · Cited by 14
Personalized Imputation on Wearable-Sensory Time Series via Knowledge Transfer
Wu, X., Mattingly, S. M., Mirjafari, S., Huang, C., & Chawla, N. V. (2020) · ACM International Conference on Information and Knowledge Management (CIKM) · doi:10.1145/3340531.3411879 · Cited by 14
Ilicic, J., & Brennan, S. M. (2022) · European Journal of Marketing · doi:10.1108/ejm-01-2021-0038 · Cited by 15
Algorithmic Social Injustices: Antecedents and Mitigations
Tanriverdi̇, H., & Akinyemi, J.-P. O. (2025) · MIS Quarterly · doi:10.25300/misq/2025/18314 · Cited by 15
Assessing Heart Rate Using Consumer Technology Association Standards
Reece, J. D., Bunn, J. A., Choi, M., & Navalta, J. W. (2021) · Technologies · doi:10.3390/technologies9030046 · Cited by 12
Buendía, R., Karpefors, M., Folkvaljon, F., HUNTER, R., Sillén, H., LUU, L., Docherty, K. F., & Cowie, M. (2024) · Journal of Cardiac Failure · doi:10.1016/j.cardfail.2024.01.016 · Cited by 19
Cai, Y., Wang, Z., Zhang, W., Kong, W., Jiang, J., Zhao, R., Wang, D., Feng, L., & Ni, G. (2022) · Sensors · doi:10.3390/s22134661 · Cited by 13
Helgerud, J., Haglo, H. P., & Hoff, J. (2022) · JMIR Cardio · doi:10.2196/38570 · Cited by 13
Lawanont, W., Mongkolnam, P., Nukoolkit, C., & Inoue, M. (2018) · Smart innovation, systems and technologies · doi:10.1007/978-3-319-92028-3_2 · Cited by 13
Støve, M. P., Holm, R. S., Kjaersgaard, A. S., Duncker, K., Jensen, M. R., & Larsen, B. T. (2020) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2020.1753836 · Cited by 12
Towards pulse rate parametrization during free-living activities using smart wristband
Rapalis, A., Petrėnas, A., Šimaitytė, M., Bailón, R., & Marozas, V. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac24a · Cited by 11
Chen, C.-C., Chen, Y., Tang, L., & Chieng, W.-H. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph19084810 · Cited by 17
Association between Estimated Cardiorespiratory Fitness and Abnormal Glucose Risk: A Cohort Study
Sloan, R., Kim, Y., Kenyon, J., Visentini-Scarzanella, M., Sawada, S., Sui, X.-M., Lee, I., Myers, J., & Lavie, C. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12072740 · Cited by 5
Islam, B., McElwain, N. L., Li, J., Davila, M. I., Hu, Y., Hu, K., Bodway, J. M., Dhekne, A., Choudhury, R. R., & Hasegawa‐Johnson, M. (2024) · Sensors · doi:10.3390/s24030901 · Cited by 11
Smits-Engelsman, B. C. M., & Bonney, E. (2019) · Research in Developmental Disabilities · doi:10.1016/j.ridd.2019.103461 · Cited by 11
Variable agreement between wearable heart rate monitors during exercise in cystic fibrosis
Gaynor, M., Sawyer, A., JENKINS, S. C., & Wood, J. (2019) · ERJ Open Research · doi:10.1183/23120541.00006-2019 · Cited by 10
Mestrom, E. H. J., Deneer, R., Bonomi, A., Margarito, J., Gelissen, J., Haakma, R., Korsten, H. H. M., Scharnhorst, V., & Bouwman, R. A. (2021) · JMIR Cardio · doi:10.2196/27765 · Cited by 10
Can accelerometry data improve estimates of heart rate variability from wrist pulse PPG sensors?
Kos, M., Li, X., Far, I. K., Gordon, C. M., Misha Pavel, C., & Jimison, H. B. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8037141 · Cited by 11
Cheung, Y. K., Hsueh, P.-Y. S., Ensari, I., Willey, J. Z., & Diaz, K. M. (2018) · Sensors · doi:10.3390/s18093056 · Cited by 11
Yoon, E. S., Choo, J., Kim, J., & Jae, S. Y. (2019) · The Korean Journal of Sports Medicine · doi:10.5763/kjsm.2019.37.4.162 · Cited by 11
Apple watch accuracy in monitoring health metrics: a systematic review and meta-analysis
Choe, J.-P., & Kang, M. (2025) · Physiological Measurement · doi:10.1088/1361-6579/adca82 · Cited by 11
Ravanelli, N., Lefebvre, K., Brough, A., Paquette, S., & Lin, W. (2025) · Sensors · doi:10.3390/s25092926 · Cited by 9
Study-based evaluation of accuracy and usability of wearable devices in manual assembly
Tropschuh, B., Windecker, S., & Reinhart, G. (2022) · Production & Manufacturing Research · doi:10.1080/21693277.2022.2100505 · Cited by 11
Occupational physical activity in brewery and office workers
Thompson, J. F., Severson, R. L., & Rosecrance, J. C. (2018) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2018.1492136 · Cited by 9
DUCLOS, M., Lacomme, P., Lambert, C., Pereira, B., Ren, L., Fleury, G., OVIGNEUR, H., DESCHAMPS, T., FEARNBACH, N., Vanhelst, J., et al. (2022) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.21.12203-0 · Cited by 12
Hailey, K., Fortin, J. M., Pratt, P. G., Forbes, P. W., & McCabe, M. A. (2021) · Holistic Nursing Practice · doi:10.1097/hnp.0000000000000475 · Cited by 9
Use of ultra-low cost fitness trackers as clinical monitors in low resource emergency departments
Dagan, A. S., & Mechanic, O. J. (2020) · Clinical and Experimental Emergency Medicine · doi:10.15441/ceem.19.081 · Cited by 9
Oost, N. C. V., Masci, F., Malisse, A., Schyvens, A.-M., Peters, B., Dirix, H., Ross, V., Wets, G., Neven, A., Verbraecken, J. A., et al. (2025) · Sensors · doi:10.3390/s25206319 · Cited by 12
Gielen, J., Oost, N. C. V., Debard, G., Sels, R., Witte, N. A. J. D., Colman, T., Bonroy, B., & Aerts, J. (2026) · JMIR Formative Research · doi:10.2196/85186 · Cited by 8
Lazar, J. K., Feng, J. H., & Hochheiser, H. (2017) · Elsevier eBooks · doi:10.1016/b978-0-12-805390-4.00013-3 · Cited by 10
Metadata
- Authors
- Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S.
- Affiliations
- The University of Queensland (AU)
- Norwegian University of Science and Technology (NO)
- Corresponding author
- Coombes, Jeff S.
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2016-05-27
- Volume
- 11
- Issue
- 5
- Pages
- e0154420
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4883747
- OpenAlex
- W2395644623
- Cited by
- 336
- References
- 17
- Citations by year
- 2026: 12 · 2025: 20 · 2024: 18 · 2023: 28 · 2022: 30 · 2021: 30 · 2020: 41 · 2019: 54 · 2018: 62 · 2017: 35 · 2016: 6
- FWCI
- 42.08
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- FedUni ResearchOnline (Federation University Australia) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- NORA - Norwegian Open Research Archives (submittedVersion, cc-by)
- Funders
- Coca-Cola
- Topics
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Physical Activity and Health
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
- Keywords
- heart rate monitors, wrist-worn devices, energy expenditure, measurement accuracy, weight loss, wearable activity trackers, Apple Watch, Fitbit, physical activity, energy balance, indirect calorimetry, electrocardiogram
- MeSH terms
- Adult, Aged, Body Weight, Energy Metabolism, Female, Heart Rate, Humans, Male, Middle Aged, Monitoring, Physiologic, Wrist, Young Adult, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review