Current Advances of Artificial Pancreas Systems: A Comprehensive Review of the Clinical Evidence
Open accessThe DOI matches its Crossref record; 42 of the 42 references it cites were found in a registry or an index.
Abstract
Since Banting and Best isolated insulin in the 1920s, dramatic progress has been made in the treatment of type 1 diabetes mellitus (T1DM). However, dose titration and timely injection to maintain optimal glycemic control are often challenging for T1DM patients and their families because they require frequent blood glucose checks. In recent years, technological advances in insulin pumps and continuous glucose monitoring systems have created paradigm shifts in T1DM care that are being extended to develop artificial pancreas systems (APSs). Numerous studies that demonstrate the superiority of glycemic control offered by APSs over those offered by conventional treatment are still being published, and rapid commercialization and use in actual practice have already begun. Given this rapid development, keeping up with the latest knowledge in an organized way is confusing for both patients and medical staff. Herein, we explore the history, clinical evidence, and current state of APSs, focusing on various development groups and the commercialization status. We also discuss APS development in groups outside the usual T1DM patients and the administration of adjunct agents, such as amylin analogues, in APSs.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
DiMeglio, L. A., Evans-Molina, C., & Oram, R. A. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31320-5 · Cited by 1,561
Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Raghinaru, D., Lum, J. W., Buckingham, B. A., Kudva, Y. C., Laffel, L. M., Levy, C. J., Pinsker, J. E., Wadwa, R. P., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1907863 · Cited by 1,030 · cited 4×
“, infusion set error) rather than automated insulin delivery function [37,45,76,78].”
BackgroundResultInfluentialEffectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Peoples, T., Perkins, B. A., Welsh, J. B., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1002853 · Cited by 894 · cited 2×
“However, SAP still could not improve severe hypoglycemia [15,58] and so the group developed an SAP with an LGS function that automatically stops insulin infusion in the event of Fig.”
BackgroundMethodFeig, D. S., Donovan, L. E., Corcoy, R., Murphy, K. E., Amiel, S. A., Hunt, K. F., Asztalos, E., Barrett, J. F. R., Sanchez, J. J., de Leiva, A., et al. (2017) · The Lancet · doi:10.1016/s0140-6736(17)32400-5 · Cited by 803
SEARCH for Diabetes in Youth Study Group (2006) · Pediatrics · doi:10.1542/peds.2006-0690 · Cited by 726
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652 · cited 1×
“2013 [16] Medtronic Medtronic Enlite Minimed 530G Minimed 530G RCT, parallel 247 16–70 Home 3 months LGS SAP SAP Nocturnal hypoglycemic events: 1.”
BackgroundSafety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes
Bergenstal, R. M., Garg, S., Weinzimer, S. A., Buckingham, B. A., Bode, B. W., Tamborlane, W. V., & Kaufman, F. R. (2016) · JAMA · doi:10.1001/jama.2016.11708 · Cited by 644 · cited 6×
“For single-hormone HCL studies, although each study presented a different form, TBR <54 or <50 mg/dL were significantly improved in HCL compared to CSII (Florence system, TBR <50 mg/dL, 0.3% vs. 1.0%, P <0.01; Minimed 670G, TBR <54 mg/dL, 0.2% vs. 0.9%, P <0.01) [24,77].”
BackgroundMethodInfluentialExercise-stimulated glucose uptake — regulation and implications for glycaemic control
Sylow, L., Kleinert, M., Richter, E. A., & Jensen, T. E. (2017) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2016.162 · Cited by 498 · cited 1×
“However, because of the time gap between the onset of insulin action and the rise in blood glucose, single-hormone APS users remain at high risk of developing hypoglycemia after meals, especially if they exercise [102].”
BackgroundYeh, H.-C., Brown, T. T., Maruthur, N., Ranasinghe, P., Berger, Z., Suh, Y. D., Wilson, L. M., Haberl, E. B., Brick, J., Bass, E. B., et al. (2012) · Annals of Internal Medicine · doi:10.7326/0003-4819-157-5-201209040-00508 · Cited by 493
Cappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477
A Randomized Trial of Closed-Loop Control in Children with Type 1 Diabetes
Breton, M. D., Kanapka, L. G., Beck, R. W., Ekhlaspour, L., Forlenza, G. P., Cengiz, E., Schoelwer, M., Ruedy, K. J., Jost, E., Carria, L., et al. (2020) · New England Journal of Medicine · doi:10.1056/nejmoa2004736 · Cited by 455 · cited 1×
“In 2020, the study was expanded to 101 children aged 6 to 13 years, and their TIR improved compared with the SAP group for 16 weeks (67% vs. 55%, P 0.001) [83], which allowed the Tandem Control-IQ to gain approval for use in children aged 6 years and older.”
MethodInfluentialKarges, B., Schwandt, A., Heidtmann, B., Kordonouri, O., Binder, E., Schierloh, U., Boettcher, C., Kapellen, T., Rosenbauer, J., & Holl, R. W. (2017) · JAMA · doi:10.1001/jama.2017.13994 · Cited by 452 · cited 1×
“In a large retrospective study, CSII was reported to be more effective for glycemic control and hypoglycemia than MDI, but many randomized controlled trials (RCTs) did not show improvement [11-13].”
BackgroundHome Use of an Artificial Beta Cell in Type 1 Diabetes
Thabit, H., Tauschmann, M., Allen, J. M., Leelarathna, L., Hartnell, S., Wilinska, M. E., Acerini, C. L., Dellweg, S., Benesch, C., Heinemann, L., et al. (2015) · New England Journal of Medicine · doi:10.1056/nejmoa1509351 · Cited by 450
Tauschmann, M., Thabit, H., Bally, L., Allen, J. M., Hartnell, S., Wilinska, M. E., Ruan, Y., Sibayan, J., Kollman, C., Cheng, P., et al. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31947-0 · Cited by 440 · cited 2×
“, infusion set error) rather than automated insulin delivery function [37,45,76,78].”
BackgroundLy, T. T., Nicholas, J. A., Retterath, A., Lim, E. M., Davis, E. A., & Jones, T. W. (2013) · JAMA · doi:10.1001/jama.2013.277818 · Cited by 411 · cited 2×
“2013 [59] Medtronic Medtronic Enlite Minimed 530G Minimed 530G RCT, parallel 95 4–50 Home 6 months LGS SAP CSII Severe to moderate hypoglycemic events: 9.”
BackgroundA Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes
El-Khatib, F. H., Russell, S. J., Nathan, D. M., Sutherlin, R. G., & Damiano, E. R. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3000619 · Cited by 403
Bergenstal, R. M., Nimri, R., Beck, R. W., Criego, A., Laffel, L., Schatz, D., Battelino, T., Danne, T., Weinzimer, S. A., Sibayan, J., et al. (2021) · The Lancet · doi:10.1016/s0140-6736(20)32514-9 · Cited by 344
Closed-Loop Insulin Delivery during Pregnancy in Women with Type 1 Diabetes
Stewart, Z. A., Wilinska, M. E., Hartnell, S., Temple, R. C., Rayman, G., Stanley, K. P., Simmons, D., Law, G. R., Scott, E. M., Hovorka, R., et al. (2016) · New England Journal of Medicine · doi:10.1056/nejmoa1602494 · Cited by 279
Collyns, O. J., Meier, R. A., Betts, Z. L., Chan, D. S. H., Frampton, C., Frewen, C. M., Hewapathirana, N. M., Jones, S. D., Roy, A., Grosman, B., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2250 · Cited by 276 · cited 1×
“, infusion set error) rather than automated insulin delivery function [37,45,76,78].”
BackgroundInsulin-pump therapy for type 1 diabetes mellitus
Pickup, J. C. (2012) · New England Journal of Medicine · doi:10.1056/nejmct1113948 · Cited by 270
Hopkins, D., Lawrence, I., Mansell, P., Thompson, G., Amiel, S., Campbell, M., & Heller, S. (2012) · Diabetes Care · doi:10.2337/dc11-1579 · Cited by 265 · cited 1×
“However, the reality is that the process is neither easy nor comfortable [9].”
BackgroundOne Year Real-World Use of the Control-IQ Advanced Hybrid Closed-Loop Technology
Breton, M. D., & Kovatchev, B. P. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0097 · Cited by 250 · cited 1×
“In the case of Control-IQ, there was a 1-year retrospective study of 9,451 child and adult patients who mostly used Basal IQ (PLGS SAP) in the past and changed to use Control-IQ (HCL) [98].”
MethodForlenza, G. P., Li, Z., Buckingham, B. A., Pinsker, J. E., Cengiz, E., Wadwa, R. P., Ekhlaspour, L., Church, M. M., Weinzimer, S. A., Jost, E., et al. (2018) · Diabetes Care · doi:10.2337/dc18-0771 · Cited by 246 · cited 2×
“2018 [63] Tandem Dexcom G5 t:slim X2 Basal IQ RCT, crossover 103 6–72 Home 6 weeks PLGS SAP SAP TIR 70–180: 65% vs.”
BackgroundMethodHovorka, R., Kumareswaran, K., Harris, J., Allen, J. M., Elleri, D., Xing, D., Kollman, C., Nodale, M., Murphy, H. R., Dunger, D. B., et al. (2011) · BMJ · doi:10.1136/bmj.d1855 · Cited by 244 · cited 1×
“org ments in overnight time in range (TIR) 70 to 144 mg/dL and TBR <70 mg/dL compared with CSII [21,22].”
MethodClosed-loop insulin delivery for glycemic control in noncritical care
Bally, L., Thabit, H., Hartnell, S., Andereggen, E., Ruan, Y., Wilinska, M. E., Evans, M. L., Wertli, M. M., Coll, A. P., Stettler, C., et al. (2018) · New England Journal of Medicine · doi:10.1056/nejmoa1805233 · Cited by 232
Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes
Castle, J. R., Engle, J. M., Youssef, J. E., Massoud, R. G., Yuen, K. C. J., Kagan, R., & Ward, W. K. (2010) · Diabetes Care · doi:10.2337/dc09-2254 · Cited by 232
Benhamou, P.-Y., Franc, S., Reznik, Y., Thivolet, C., Schaepelynck, P., Renard, E., Guerci, B., Chaillous, L., Lukas-Croisier, C., Jeandidier, N., et al. (2019) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30003-2 · Cited by 217
One year clinical experience of the first commercial hybrid closed-loop system
Lal, R. A., Basina, M., Maahs, D. M., Hood, K., Buckingham, B., & Wilson, D. M. (2019) · Diabetes Care · doi:10.2337/dc19-0855 · Cited by 215
Real-world use of open source artificial Pancreas systems
Lewis, D., Leibrand, S., & Community, #. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816665635 · Cited by 180
Stewart, Z. A., Wilinska, M. E., Hartnell, S., O’Neil, L. K., Rayman, G., Scott, E. M., Barnard, K., Farrington, C., Hovorka, R., & Murphy, H. R. (2018) · Diabetes Care · doi:10.2337/dc17-2534 · Cited by 168 · cited 1×
“In 2018, a similar study design was used to assess similar subjects all-day, and TIR did not differ, but TBR 63 mg/dL was improved (1.6% vs. 2.7%, P =0.04) [122].”
MethodInfluentialHaidar, A., Legault, L., Dallaire, M., Alkhateeb, A., Coriati, A., Messier, V., Cheng, P., Millette, M., Boulet, B., & Rabasa-Lhoret, R. (2013) · Canadian Medical Association Journal · doi:10.1503/cmaj.121265 · Cited by 167
McAuley, S. A., Lee, M. H., Paldus, B., Vogrin, S., de Bock, M. I., Abraham, M. B., Bach, L. A., Burt, M. G., Cohen, N. D., Colman, P. G., et al. (2020) · Diabetes Care · doi:10.2337/dc20-1447 · Cited by 158 · cited 3×
“0 events); however, most events were not related to the control algorithm [38,47,76].”
BackgroundMethodCastle, J. R., El Youssef, J., Wilson, L. M., Reddy, R., Resalat, N., Branigan, D., Ramsey, K., Leitschuh, J., Rajhbeharrysingh, U., Senf, B., et al. (2018) · Diabetes Care · doi:10.2337/dc18-0228 · Cited by 152
Kovatchev, B., Cheng, P., Anderson, S. M., Pinsker, J. E., Boscari, F., Buckingham, B. A., Doyle, F. J., Hood, K. K., Brown, S. A., Breton, M. D., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0333 · Cited by 141
Taborsky, G. J., & Mundinger, T. O. (2012) · Endocrinology · doi:10.1210/en.2011-2040 · Cited by 140 · cited 1×
“Patients with T1DM often have impairments in counter-regulatory response to hypoglycemia as well as insulin secretion [101].”
BackgroundMultinational Home Use of Closed-Loop Control Is Safe and Effective
Anderson, S. M., Raghinaru, D., Pinsker, J. E., Boscari, F., Renard, E., Buckingham, B. A., Nimri, R., Doyle, F. J., Brown, S. A., Keith-Hynes, P., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2468 · Cited by 120 · cited 1×
“After performing mid-to long-term single arm studies in 2016 to 2017 [80,81], the TypeZero group conducted the largest HCL study to date in 2019 of 168 adolescents and adults for 6 months using a Tandem pump called t:slim X2, and the re-sults showed TIR (71% vs. 59%, P 0.001) and TBR (1.58% vs.…”
MethodInfluentialJacobs, P. G., El Youssef, J., Reddy, R., Resalat, N., Branigan, D., Condon, J., Preiser, N., Ramsey, K., Jones, M., Edwards, C., et al. (2016) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.12707 · Cited by 116 · cited 1×
“In the same year, the Oregon and IRCM groups showed better hypoglycemia control with dual-hormone HCL than with SAP and single-hormone HCL, respectively, in short-term studies of exercising patients [106,107].”
BackgroundFirst use of model predictive control in outpatient wearable artificial pancreas
Del Favero, S., Bruttomesso, D., Di Palma, F., Lanzola, G., Visentin, R., Filippi, A., Scotton, R., Toffanin, C., Messori, M., Scarpellini, S., et al. (2014) · Diabetes Care · doi:10.2337/dc13-1631 · Cited by 103 · cited 1×
“Previously, HCL studies from another group had used the Omnipod pump (Insulet) [23,89], but the Insulet group began research using its own control algorithm in the late 2010s.”
MethodChanging the landscape for type 1 diabetes: the first step to prevention
Dayan, C. M., Korah, M., Tatovic, D., Bundy, B. N., & Herold, K. C. (2019) · The Lancet · doi:10.1016/s0140-6736(19)32127-0 · Cited by 96
Dassau, E., Renard, E., Place, J., Farret, A., Pelletier, M., Lee, J., Huyett, L. M., Chakrabarty, A., Doyle, F. J., & Zisser, H. C. (2017) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.12999 · Cited by 94 · cited 1×
“Furthermore, to overcome the limitations of insulin-based APSs, other peptide hormones such as amylin analogues and glucagon-like peptide 1 (GLP-1) receptor (GLP-1R) agonists have been investigated, and intraperitoneal (IP) insulin delivery systems have been in development [40-42].”
BackgroundDel Favero, S., Boscari, F., Messori, M., Rabbone, I., Bonfanti, R., Sabbion, A., Iafusco, D., Schiaffini, R., Visentin, R., Calore, R., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2815 · Cited by 94
Sherr, J. L., Buckingham, B. A., Forlenza, G. P., Galderisi, A., Ekhlaspour, L., Wadwa, R. P., Carria, L., Hsu, L., Berget, C., Peyser, T. A., et al. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0286 · Cited by 92
Cited by
Zeng, B., Gao, L., Yang, Q., Jia, H., & Sun, F. (2023) · Diabetes Care · doi:10.2337/dc23-0504 · Cited by 100
Moon, S. J., Kim, K., Lee, W. J., Lee, M. Y., Vigersky, R., & Park, C. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14852 · Cited by 69
Self-citationPemberton, J. S., Wilmot, E. G., Barnard‐Kelly, K., Leelarathna, L., Oliver, N., Randell, T., Taplin, C. E., Choudhary, P., & Adolfsson, P. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14962 · Cited by 44
Kim, J. Y., Kim, S., & Kim, J. H. (2025) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2024.0804 · Cited by 13
Metadata
- Authors
- Moon, S. J., Jung, I., & Park, C.-Y.
- Venue
- Diabetes & Metabolism Journal
- Publisher
- Korean Diabetes Association
- Published
- 2021-11-30
- Volume
- 45
- Issue
- 6
- Pages
- 813-839
- ISSN
- 2233-6079 (ISSN-L)
- 2233-6087
- OpenAlex
- W3215260405
- Cited by
- 98
- References
- 42
- FWCI
- 9.35
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.