Attenuation of pathological tremors by functional electrical stimulation I: Method
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
In this study we explored the possibility of suppressing pathological tremors using closed-loop functional electrical stimulation (FES) to activate the tremorogenic muscles out-of-phase. A displacement signal monitored with a transducer was filtered so as to be "tuned" to the tremor frequency at the wrist or elbow. The filtered signal was used to amplitude-modulate the electrical stimulation. The design process was based on measurements of the open-loop frequency response characteristics of the forearm and hand to stimulation of the elbow and wrist flexors and extensors in a number of subjects. These data allowed us to identify closed-loop configurations, which attenuated 2-5 Hz tremors substantially, while only minimally attenuating functional movements in the 0-1 Hz range. There was a fairly delicate balance between efficacy and the risk of instability. However, designs were identified that offered enough tremor suppression and adequate immunity to muscle/load variations for the technique to be considered seriously for clinical application.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Although tremor in one patient did not decreased, the device achieved an average suppression of 58.1 ± 20.5% (in a range between 91 and 10%) in the rest of the patients.”
From A Review on Wearable Technologies for Tremor Suppression (2021) · cites this paper as [63]
Citations
Cites
The biomechanics of human movement
(1991) · Applied Ergonomics · doi:10.1016/0003-6870(91)90404-6 · Cited by 739
BIOMECHANICS OF HUMAN MOVEMENT
Mclntyre, D. R. (1980) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-198024000-00020 · Cited by 384
Experimental correction of footdrop by electrical stimulation of the peroneal nerve
Waters, R. L., McNeal, D. R., & Perry, J. (1975) · Journal of Bone and Joint Surgery · doi:10.2106/00004623-197557080-00002 · Cited by 152 · cited 1×
“, (20)) may provide viable alternatives in the future but at this stage are only commercially available for stimulation of the common peroneal nerve.”
BackgroundSignal-handling characteristics of load-moving skeletal muscle
Ld, P. (1966) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1966.210.5.1178 · Cited by 152 · cited 1×
“A priori, this seemed quite possible in that the dynamic response characteristics of muscle are known to be load-dependent (11): large loads could cause changes in the combined frequency characteristic, substantially reducing overall gain at the tremor frequency and destabilizing the closed loop.”
BackgroundMovement Disorders: A Comprehensive Survey
(1990) · Annals of Internal Medicine · doi:10.7326/0003-4819-112-4-314_1 · Cited by 138
Marsolais, E. B., & Kobetic, R. (1986) · PubMed · Cited by 137
Closed-Loop Control of Force During Electrical Stimulation of Muscle
Crago, P. E., Mortimer, J. T., & Peckham, P. H. (1980) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1980.326738 · Cited by 135 · cited 1×
“Closed-loop FES has been explored before to linearise the responses to externally generated movement command signals (1,15,16,19).”
BackgroundAN INVESTIGATION INTO THE VALUE OF TREATING INTENTION TREMOR BY WEIGHTING THE AFFECTED LIMB
HEWER, R. L., COOPER, R., & MORGAN, M. H. (1972) · Brain · doi:10.1093/brain/95.3.579 · Cited by 112 · cited 1×
“The only available alternative to control such tremors has been inertial loading of the affected limb (3).”
BackgroundControl of a skeletal joint by electrical stimulation of antagonists
Vodovnik, L., Crochetiere, W. J., & RESWICK, J. B. (1967) · Medical & Biological Engineering & Computing · doi:10.1007/bf02474498 · Cited by 90 · cited 2×
“Closed-loop FES has been explored before to linearise the responses to externally generated movement command signals (1,15,16,19).”
BackgroundAttenuation of pathological tremors by functional electrical stimulation II: Clinical evaluation
Javidan, M., Elek, J., & Prochazka, A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368522 · Cited by 80
Self-citation
Research works printed but not linked: 5
- Javidan, M.; Elek, J.; Prochazka, A. Tremor reduction by functional electrical stimulation. Neurol. 40:369; 1990.
- Morrice, B.L.; Lee, R.G.; Becker, W.J.; White, D.G.; Hoffer, J.A. Viscous loading improves manual tracking accuracy in humans with cerebellar incoordination. Soc. Neurosci. Abstracts 13:1700; 1987.
- Mortimer, J.T. Motor prostheses. In: Brooks, V.B., ed. American handbook of physiology II, Section I: Motor control. Baltimore, MD: Williams and Wilkins; 1981: pp. 155–187.
- Stein, R.B.; Prochazka, A.; Popovic, D.; Edamura, M.; Llewellyn, M.; Davis, L.A. Technology transfer and development for walking using functional electrical stimulation. In: Popovic, D., ed., Advances in external control of human extremities X. Belgrade: Yugoslav Committee for Electronics and Automation; 1990: pp. 161–175.
- Wood, R.C. Automatic linear control system analysis and design software. Los Angeles, CA: West Educational Publishing Co.; 1988.
Cited by
Adaptive cancelling of physiological tremor for improved precision in microsurgery
Riviere, C. N., Rader, R. S., & Thakor, N. V. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.686791 · Cited by 260
Prochazka, A., Gauthier, M., Wieler, M., & Kenwell, Z. (1997) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(97)90426-3 · Cited by 250
Positive Force Feedback Control of Muscles
Prochazka, A., Gillard, D., & Bennett, D. J. (1997) · Journal of Neurophysiology · doi:10.1152/jn.1997.77.6.3226 · Cited by 171
Implications of Positive Feedback in the Control of Movement
Prochazka, A., Gillard, D., & Bennett, D. J. (1997) · Journal of Neurophysiology · doi:10.1152/jn.1997.77.6.3237 · Cited by 165
Electrical stimulation for the suppression of pathological tremor
Popović Maneski, L., Jorgovanović, N., Ilić, V., Došen, S., Keller, T., Popović, M. B., & Popović, D. B. (2011) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-011-0803-6 · Cited by 129
“The first FES system that was designed to suppress tremor was proposed 20 years ago [6, 15 ].”
MethodOnline tremor suppression using electromyography and low-level electrical stimulation
Dosen, S., Muceli, S., Dideriksen, J. L., Romero, J. P., Rocon, E., Pons, J., & Farina, D. (2015) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2014.2328296 · Cited by 120
“As in the majority of previous tremor suppression studies based on FES [15], [16], [19] the stimulation was delivered out of phase with the estimated tremorogenic activity.”
BackgroundResultModeling and canceling tremor in human-machine interfaces
Riviere, C. N., & Thakor, N. V. (1996) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.499755 · Cited by 114
Non-invasive characterization of motor unit behaviour in pathological tremor
Holobar, A., Glaser, V., Gallego, J. Á., Dideriksen, J. L., & Farina, D. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/5/056011 · Cited by 96
“Detecting motor unit activities during tremor contractions non-invasively would allow novel physiological insights into tremor pathology as well as the possibility of developing EMG-driven technologies for tremor suppression, e.g. EMGdriven functional electrical stimulation (Prochazka et al 1992).”
BackgroundZhang, D., Poignet, P., Widjaja, F., & Tech Ang, W. (2011) · Control Engineering Practice · doi:10.1016/j.conengprac.2010.08.009 · Cited by 88
Tremor suppression through impedance control
Pledgie, S., Barner, K. E., Agrawal, S. K., & Rahman, T. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.830949 · Cited by 85
A neuroprosthesis for tremor management through the control of muscle co-contraction
Gallego, J. Á., Rocon, E., Belda-Lois, J. M., & Pons, J. L. (2013) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-10-36 · Cited by 81
“Notice that transcutaneous neurostimulation for tremor management had been preliminarily evaluated in [32].”
BackgroundAttenuation of pathological tremors by functional electrical stimulation II: Clinical evaluation
Javidan, M., Elek, J., & Prochazka, A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368522 · Cited by 80
“I N T R O D U C T I O N In the prev ious p a p e r (7) we descr ibed a new m e t h o d for select ively a t t enua t i ng t r e m o r by using c losed- loop func t iona l e lectr ical s t imu la t ion (FES) to ac t iva te the t r emorogen ic muscles ou t -o f -phase .”
BackgroundMethodSelf-citationMedical Devices for Tremor Suppression: Current Status and Future Directions
Mo, J., & Priefer, R. (2021) · Biosensors · doi:10.3390/bios11040099 · Cited by 78
“The recognized limitations of these early works included the potentially unreliable placement of surface electrodes, which could lead to insufficient tremor suppression [74,75].”
BackgroundTuned vibration absorber for suppression of rest tremor in Parkinson’s disease
Hashemi, S. M., Golnaraghi, M. F., & Patla, A. E. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02351012 · Cited by 68
Electrical stimulation of afferent pathways for the suppression of pathological tremor
Dideriksen, J. L., Laine, C. M., Dosen, S., Muceli, S., Rocon, E., Pons, J. L., Benito-Leon, J., & Farina, D. (2017) · Frontiers in Neuroscience · doi:10.3389/fnins.2017.00178 · Cited by 67
“The most common strategy (out-of-phase) applies electrical stimulation to the muscles so that they generate forces opposite to those arising from the tremorogenic bursts of activity (Javidan et al., 1992; Prochazka et al., 1992; Gillard et al., 1999; Popović Maneski et al., 2011; Dosen et al., 2015).”
BackgroundInfluentialGallego, J. A., Dideriksen, J. L., Holobar, A., Ibanez, J., Glaser, V., Romero, J. P., Benito-Leon, J., Pons, J. L., Rocon, E., & Farina, D. (2015) · Journal of Neuroscience · doi:10.1523/jneurosci.0106-15.2015 · Cited by 67
“However, any eventual effect cannot be easily dissociated from that of the low-pass filter characteristics of the muscle and limb mechanics (Milner-Brown et al., 1973; Prochazka et al., 1992; Baldissera et al., 1998).”
BackgroundPeripheral electrical stimulation to reduce pathological tremor: a review
Pascual-Valdunciel, A., Hoo, G. W., Avrillon, S., Barroso, F. O., Goldman, J. G., Hernandez-Pavon, J. C., & Pons, J. L. (2021) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-021-00811-9 · Cited by 66
A Review on Wearable Technologies for Tremor Suppression
Lora-Millan, J. S., Delgado-Oleas, G., Benito-León, J., & Rocon, E. (2021) · Frontiers in Neurology · doi:10.3389/fneur.2021.700600 · Cited by 62
Gillard, D. M., Cameron, T., Prochazka, A., & Gauthier, M. J. A. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.788474 · Cited by 61
Functional electrical stimulation in rehabilitation engineering: a survey
Zhang, D., Guan, T. H., Widjaja, F., & Ang, W. T. (2007) · International Convention on Rehabilitation Engineering & Assistive Technology · doi:10.1145/1328491.1328546 · Cited by 62
“A special case related to upper limb is the pathological tremor suppression via FES [25], [ 79 ].”
BackgroundOptimal digital filtering for tremor suppression
Gonzalez, J. G., Heredia, E. A., Rahman, T., Barner, K. E., & Arce, G. R. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.841338 · Cited by 56
Pathological Tremor and Voluntary Motion Modeling and Online Estimation for Active Compensation
Bo, A. P. L., Poignet, P., & Geny, C. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2010.2091283 · Cited by 53
“Likewise, these online methods are a fundamental part in active devices designed to directly attenuate tremor, such as orthosis [6] or solutions based on electrical stimulation [ 7 ].”
MethodAdaptive Suppression of Severe Pathological Tremor by Torque Estimation Method
Taheri, B., Case, D., & Richer, E. (2015) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2014.2317948 · Cited by 53
“Functional electrical stimulation has been used to stimulate the tremorogenic muscles and attenuate tremor amplitude [6]–[8].”
MethodA Multimodal Human–Robot Interface to Drive a Neuroprosthesis for Tremor Management
Gallego, J. Á., Ibáñez, J., Dideriksen, J. L., Serrano, J. I., del Castillo, M. D., Farina, D., & Rocón, E. (2012) · IEEE Transactions on Systems Man and Cybernetics Part C (Applications and Reviews) · doi:10.1109/tsmcc.2012.2200101 · Cited by 53
“In addition, functional electrical stimulation (FES) for tremor suppression has already been preliminarily validated in [7].”
BackgroundEMG-Based Characterization of Pathological Tremor Using the Iterated Hilbert Transform
Dideriksen, J. L., Gianfelici, F., Maneski, L. Z. P., & Farina, D. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2163069 · Cited by 42
“This may be an issue when using FES to suppress tremor in specific muscles.”
BackgroundCharacterization of Parkinsonian Hand Tremor and Validation of a High-Order Tremor Estimator
Zhou, Y., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2018) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2018.2859793 · Cited by 42
“that applies functional electrical stimulation [16]–[23].”
BackgroundAdaptive sliding bandlimited multiple fourier linear combiner for estimation of pathological tremor
Wang, S., Gao, Y., Zhao, J., & Cai, H. (2014) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2013.10.004 · Cited by 42
Vib-bracelet: a passive absorber for attenuating forearm tremor
Buki, E., Katz, R., Zacksenhouse, M., & Schlesinger, I. (2017) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-017-1742-7 · Cited by 42
“Alternatively, active tremor suppression can be achieved by out-of-phase closed-loop FES [12, 16].”
BackgroundOn the Use of Fixed‐Intensity Functional Electrical Stimulation for Attenuating Essential Tremor
Bó, A. P. L., Azevedo‐Coste, C., Geny, C., Poignet, P., & Fattal, C. (2014) · Artificial Organs · doi:10.1111/aor.12261 · Cited by 40
“Also, increase in tremor amplitude may occur due to errors in tremor phase estimation or in controller parameters, which may be aggravated if manual tuning of parameters before each use is required (8).”
BackgroundNeurophysiology and neural engineering: a review
Procházka, A. (2017) · Journal of Neurophysiology · doi:10.1152/jn.00149.2017 · Cited by 40
A soft wearable robot for tremor assessment and suppression
Gallego, J. Á., Rocón, E., Ibáñez, J., Dideriksen, J. L., Koutsou, A. D., Paradiso, R., Popović, M. B., Lois, J. M. B., Gianfelici, F., Farina, D., et al. (2011) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2011.5979639 · Cited by 39
“Notice that attenuation of writs tremors based on muscle stimulation has been previously evaluated in [7], demonstrating the feasibility of the approach.”
MethodRepetitive Control of Electrical Stimulation for Tremor Suppression
Copur, E. H., Freeman, C. T., Chu, B., & Laila, D. S. (2019) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2017.2771327 · Cited by 38
“Classical feedback systems also introduce large high-frequency stimulation transients that cause significant discomfort [19].”
BackgroundMethodSensory control of normal movement and of movement aided by neural prostheses
Procházka, A. (2015) · Journal of Anatomy · doi:10.1111/joa.12311 · Cited by 43
“The only closed-loop FES devices to have been tested in participants going about their daily lives were a foot-drop stimulator (Prochazka & Wiles, 1983) and a tremor-suppressing system (Javidan et al. 1992; Prochazka et al. 1992).”
BackgroundNon-invasive electrical stimulation of peripheral nerves for the management of tremor
Valdunciel, A. P., Rajagopal, A., Pons, J. L., & Delp, S. L. (2022) · Journal of the Neurological Sciences · doi:10.1016/j.jns.2022.120195 · Cited by 38
A Model of the Surface Electromyogram in Pathological Tremor
Dideriksen, J. L., Enoka, R. M., & Farina, D. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2118756 · Cited by 35
“Current methods for the treatment of tremor include medication, deep brain stimulation [5], functional electrical stimulation (FES) [6], [7], and orthotic exoskeletons [8].”
MethodDideriksen, J. L., Negro, F., & Farina, D. (2015) · Journal of Neurophysiology · doi:10.1152/jn.00247.2015 · Cited by 36
Muceli, S., Poppendieck, W., Hoffmann, K.-P., Dosen, S., Benito-León, J., Barroso, F. O., Pons, J. L., & Farina, D. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab047a · Cited by 33
“Electrical stimulation has been also used in neuroprostheses for tremor suppression (Prochazka et al 1992, Gallego et al 2013).”
A Wearable System for Attenuating Essential Tremor Based on Peripheral Nerve Stimulation
Kim, J., Wichmann, T., Inan, O. T., & Deweerth, S. P. (2020) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2020.2985058 · Cited by 33
“Previous attempts at modulating tremor characteristics were conducted with a range of (empirically determined) stimulation parameters [16]–[24], [28].”
BackgroundMethodInfluentialZhang, Z., Chu, B., Liu, Y., Li, Z., & Owens, D. H. (2022) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2022.3150301 · Cited by 37
“As an alternative, functional electrical stimulation (FES) is shown promising due to its minor side effects, low cost, and portability [11]–[15].”
BackgroundMethodReciprocal EMG Controlled FES for Pathological Tremor Suppression of Forearm
Zhang, D., & Ang, W. T. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353416 · Cited by 33
“In recent years, assistive technology has rapidly emerged to be a promising method for pathological tremor suppression; for example, orthotics [6] and functional electrical stimulation (FES) [ 10 ].”
MethodShahtalebi, S., Atashzar, S. F., Samotus, O., Patel, R. V., Jog, M. S., & Mohammadi, A. (2020) · Scientific Reports · doi:10.1038/s41598-020-58912-9 · Cited by 32
“PHT therapies such as oral medication, therapeutic lesions, Gamma-knife radiosurgery, and Deep Brain Stimulation (DBS) 18 focus on alleviating tremor severity and improving arm functionality 19–21 .”
BackgroundTremor Suppression of Elbow Joint via Functional Electrical Stimulation: A Simulation Study
Zhang, D., & Ang, W. (2006) · 2006 IEEE International Conference on Automation Science and Engineering · doi:10.1109/coase.2006.326877 · Cited by 32
“The assistive technique functions in two forms: (i) to suppress the tremor at the man-machine interface [3], [5], [6], and (ii) to attenuate the tremor actively and directly at the limbs [2], [4].”
BackgroundMethodRepetitive control of functional electrical stimulation for induced tremor suppression
Freeman, C. T., Sampson, P., Burridge, J. H., & Hughes, A.-M. (2015) · Mechatronics · doi:10.1016/j.mechatronics.2015.10.008 · Cited by 31
Gillard, D. M., Yakovenko, S., Cameron, T., & Procházka, A. (2000) · Journal of Biomechanics · doi:10.1016/s0021-9290(00)00127-5 · Cited by 31
Active tremor compensation in handheld instrument for microsurgery
Ang, W. T., Khosla, P. K., & Riviere, C. N. (2004) · Cited by 36
Sensory electrical stimulation for suppression of postural tremor in patients with essential tremor
Heo, J.-H., Kim, J.-W., Kwon, Y., Lee, S.-K., Eom, G.-M., Kwon, D.-Y., Lee, C.-N., Park, K.-W., & Manto, M. (2015) · Bio-Medical Materials and Engineering · doi:10.3233/bme-151372 · Cited by 30
FES-controlled co-contraction strategies for pathological tremor compensation
Bo, A. P. L., Poignet, P., Zhang, D., & Ang, W. T. (2009) · 2009 IEEE/RSJ International Conference on Intelligent Robots and Systems · doi:10.1109/iros.2009.5354397 · Cited by 30
“In addition, it may be also used to simulate different tremor compensation strategies, including strategies based on motion (such as [3]) and impedance control.”
BackgroundMethodAdaptive band-pass filter (ABPF) for tremor extraction from inertial sensor data
Popović, L. Z., Šekara, T. B., & Popović, M. B. (2010) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2010.03.018 · Cited by 29
“At present, tremulous movement is most commonly recorded with tiny accelerometers or gyroscopes [2] attached to the skin of the body part of interest, or less often, with an externally positioned device for displacement measurement [3,4].”
BackgroundHao, M.-Z., Xu, S.-Q., Hu, Z.-X., Xu, F.-L., Niu, C.-X. M., Xiao, Q., & Lan, N. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0286-2 · Cited by 29
Simulated Tremor Propagation in the Upper Limb: From Muscle Activity to Joint Displacement
Corie, T., & Charles, S. K. (2019) · Journal of Biomechanical Engineering · doi:10.1115/1.4043442 · Cited by 30
Biomechanical loading as an alternative treatment for tremor: a review of two approaches
Rocon, E., Gallego, J. Á., Belda-Lois, J. M., Benito-León, J., & Pons, J. L. (2012) · Tremor and Other Hyperkinetic Movements · doi:10.5334/tohm.107 · Cited by 26
Tremor attenuation using FES-based joint stiffness control
Bó, A. P. L., & Poignet, P. (2010) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.2010.5509560 · Cited by 26
“In his pioneer work, [4], [5] proposed a closed-loop FES-based tremor compensation system in which a pair of antagonist muscles were stimulated in order to contract out of phase with respect to tremor.”
BackgroundEssential tremor amplitude modulation by median nerve stimulation
Reis, C., Arruda, B. S., Pogosyan, A. H., Brown, P., & Cagnan, H. (2021) · Scientific Reports · doi:10.1038/s41598-021-96660-6 · Cited by 26
Rationale and Evidence for Peripheral Nerve Stimulation for Treating Essential Tremor
Shukla, A. W. (2022) · Tremor and Other Hyperkinetic Movements · doi:10.5334/tohm.685 · Cited by 25
“FES was observed to result in nearly 70% tremor improvement [36, 51].”
BackgroundActive vibration control in human forearm model using paired piezoelectric sensor and actuator
Hosseini, S. M., Kalhori, H., & Al-Jumaily, A. A. (2020) · Journal of Vibration and Control · doi:10.1177/1077546320957533 · Cited by 24
“Email: Hamed.kalhori@uts.edu.au et al., 2000; Prochazka et al., 1992), vibration therapy (Feys et al., 2006; Haas et al., 2006), limb weights (Hewer et al., 1972; Meshack and Norman, 2002), bright light therapy (Nagaoka et al., 2012) and transcranial magnetic stimulation (Filipović et al., 2010)…”
BackgroundEvaluation of a controlled-energy-dissipation orthosis for tremor suppression
Arnold, A. S., Rosen, M. J., & Aisen, M. L. (1993) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(05)80001-x · Cited by 24
Marchis, C. D., Schmid, M., & Conforto, S. (2012) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2011.12.017 · Cited by 21
Mechanical vibration does not systematically reduce the tremor in essential tremor patients
Lora-Millán, J. S., López-Blanco, R., Gallego, J. Á., Méndez-Guerrero, A., González de la Aleja, J., & Rocon, E. (2019) · Scientific Reports · doi:10.1038/s41598-019-52988-8 · Cited by 20
Pigg, A. C., Thompson‐Westra, J., Mente, K., Maurer, C. W., Haubenberger, D., Hallett, M., & Charles, S. K. (2020) · Clinical Neurophysiology · doi:10.1016/j.clinph.2020.08.010 · Cited by 20
“…al., 2004, Kotovsky and Rosen, 1998, Loureiro et al., 2005) or through electrical (Dosen et al., 2015, Freeman et al., 2015, Maneski et al., 2011, Prochazka et al., 1992) or vibratory stimulation (Feys et al., 2006, King et al., 2009) offers a potential alternative to medication and surgery, but…”
BackgroundDesigning a hand rest tremor dynamic vibration absorber using H 2 optimization method
Rahnavard, M., Hashemi, M., Farahmand, F., & Dizaji, A. F. (2014) · Journal of Mechanical Science and Technology · doi:10.1007/s12206-014-0104-8 · Cited by 19
Dideriksen, J. L., Gallego, J. Á., Holobar, A., Rocón, E., Pons, J. L., & Farina, D. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/4/046019 · Cited by 20
RBF-Based Technique for Statistical Demodulation of Pathological Tremor
Gianfelici, F. (2013) · IEEE Transactions on Neural Networks and Learning Systems · doi:10.1109/tnnls.2013.2263288 · Cited by 19
Modeling and experimental study of a hand tremor suppression system
Abbasi, M., & Afsharfard, A. (2018) · Mechanism and Machine Theory · doi:10.1016/j.mechmachtheory.2018.04.009 · Cited by 18
Frequency domain-based models of skeletal muscle
Baratta, R. V., Solomonow, M., & Zhou, B. (1998) · Journal of Electromyography and Kinesiology · doi:10.1016/s1050-6411(97)00024-2 · Cited by 18
Tremor Suppression Using Functional Electrical Stimulation
Habibollahi, Z., Zhou, Y., Jenkins, M. E., Garland, S. J., Friedman, E., Naish, M. D., & Trejos, A. L. (2024) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2024.3453222 · Cited by 17
“FES contracts the flexor and extensor muscles simultaneously (co-contraction method) [11], [13], or in an alternating way to counteract tremor (out-of-phase) [8], [9], [10], [12], [14].”
BackgroundLambrecht, S., Gallego, J. A., Rocon, E., & Pons, J. L. (2014) · Frontiers in Neuroscience · doi:10.3389/fnins.2014.00221 · Cited by 16
Zhang, Z., Chu, B., Liu, Y., Ren, H., Li, Z., & Owens, D. H. (2021) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2021.3111107 · Cited by 18
“The results of FMI-MP-RC algorithm, GB-MPRC algorithm, frequency modified inverse single periodic RC (FMI-SP-RC) algorithm [15] and the conventional PI high-pass filter (PI-HF) based control algorithm [17] will be compared.”
MethodBioinformatic Approaches Used in Modelling Human Tremor
Mario, M., Giuliana, G., Thomas, L., Dario, F., Lana, P., Silvia, C., Tommaso, D., Juan-Manuel, B.-L., Jose-Luis, P., & Eduardo, R. (2009) · Current Bioinformatics · doi:10.2174/157489309788184747 · Cited by 17
Non-invasive Central and Peripheral Stimulation: New Hope for Essential Tremor?
Chalah, M. A., Lefaucheur, J., & Ayache, S. S. (2015) · Frontiers in Neuroscience · doi:10.3389/fnins.2015.00440 · Cited by 17
“Old” and emerging Therapies of Human Tremor
Grimaldi, G., & Manto, M. U. (2010) · Clinical Medicine Insights Therapeutics · doi:10.4137/cmt.s2999 · Cited by 18
Zhou, Y., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2016) · 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7592069 · Cited by 16
Widjaja, F., Shee, C. Y., Au, W. L., Poignet, P., & Ang, W. T. (2011) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2011.5979865 · Cited by 16
“[2] have shown that it is possible to use feedback-controlled Functional Electrical Stimulation (FES) to attenuate pathological tremor.”
Kim, J., Wichmann, T., Inan, O. T., & DeWeerth, S. P. (2022) · Journal of Personalized Medicine · doi:10.3390/jpm12010076 · Cited by 16
“Tremor can also be suppressed by functional electrical stimulation (FES), which aims to stimulate arm muscles of the upper limbs [30–36].”
BackgroundWIDJAJA, F., SHEE, C. Y., ANG, W. T., AU, W. L., & POIGNET, P. (2011) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519411004344 · Cited by 14
“Prochazka et al. [Prochazka et al., 1992] have shown that it is possible to use feedback-controlled functional electrical stimulation (FES) to attenuate pathological tremor.”
InfluentialWearable Peripheral Electrical Stimulation Devices for the Reduction of Essential Tremor: A Review
Karamesinis, A. D., Sillitoe, R. V., & Kouzani, A. Z. (2021) · IEEE Access · doi:10.1109/access.2021.3084819 · Cited by 14
“amounts of hardware are required, reducing the likelihood of having a comfortable wearable device [65], [66].”
BackgroundInfluentialStandring, D. J., Pigg, A. C., Thompson‐Westra, J., Mente, K., Maurer, C. W., Haubenberger, D., Hallett, M., & Charles, S. K. (2022) · Clinical Neurophysiology · doi:10.1016/j.clinph.2022.07.001 · Cited by 15
A two-step model to optimise transcutaneous electrical stimulation of the human upper arm
Goffredo, M., Schmid, M., Conforto, S., Bilotti, F., Palma, C., Vegni, L., & D’Alessio, T. (2014) · COMPEL The International Journal for Computation and Mathematics in Electrical and Electronic Engineering · doi:10.1108/compel-04-2013-0118 · Cited by 14
“Recently, FES effectiveness in reducing pathological tremor has been successfully demonstrated (Prochazka et al., 1992; Rocon et al., 2010; Zhang and Ang, 2007).”
BackgroundTremor Suppression With Mechanical Vibration Stimulation
Liu, W., Kai, T., & Kiguchi, K. (2020) · IEEE Access · doi:10.1109/access.2020.3045023 · Cited by 13
“As the first study of tremor suppression using a periodic TVR with mechanical vibration stimulation, the effectiveness of the proposed method was confirmed with the healthy individuals similar to the other attempt [17][22].”
MethodPuttaraksa, G., Muceli, S., Barsakcioglu, D. Y., Holobar, A., Clarke, A. K., Charles, S. K., Pons, J. L., & Farina, D. (2022) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2022.3158606 · Cited by 13
SUPPRESSION OF ACTION TREMOR BY SENSORY ELECTRICAL STIMULATION IN PATIENTS WITH ESSENTIAL TREMOR
HEO, J.-H., KWON, Y., JEON, H.-M., KWON, D.-Y., LEE, C.-N., PARK, K.-W., MANTO, M., KIM, J.-W., & EOM, G.-M. (2016) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519416400261 · Cited by 12
Estimation of instantaneous tremor parameters for FES-based tremor suppression
Gallego, J. Á., Rocón, E., & Pons, J. L. (2010) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.2010.5509663 · Cited by 15
“In [5], the authors designed a controller that stimulated the tremorogenic muscles out of phase, attenuating the tremor without significantly affecting concomitant voluntary movement.”
BackgroundMultimodal Tremor Suppression of the Wrist Using FES and Electric Motors–A Simulation Study
Habibollahi, Z., Zhou, Y., Jenkins, M. E., Garland, S. J., Naish, M. D., & Trejos, A. L. (2023) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2023.3316607 · Cited by 11
“Wearable tremor suppression devices (WTSDs) have been developed over the past few decades to reduce and manage tremor using mechanical loading [3], [4] or external electrical stimulation of the target limb [5], [6], [7].”
BackgroundMethodPrediction of pathological tremor using adaptive multiple oscillators linear combiner
Xiao, F., Gao, Y., Wang, S., & Zhao, J. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.01.006 · Cited by 11
Pan, B., Qu, X., Ai, Y., Fu, Y., & Liu, C. (2016) · Computer Assisted Surgery · doi:10.1080/24699322.2016.1240316 · Cited by 10
“weakness for real-time control.[12] Real-time tremor”
System identification for FES-based tremor suppression
Çopur, E. H., Freeman, C. T., Chu, B., & Laila, D. S. (2015) · European Journal of Control · doi:10.1016/j.ejcon.2015.12.003 · Cited by 10
“In [33, 34], FES was applied to suppress tremor at the wrist or elbow using filters implemented in a closed2 loop feedback arrangement.”
BackgroundMethodInfluentialRobust higher order repetitive control applied to human tremor suppression
Verstappen, R. J. L. M., Freeman, C. T., Rogers, E., Sampson, T., & Burridge, J. H. (2012) · The proceedings of the IEEE Conference on Control Applications/The proceedings of the ... IEEE Conference on Control Applications · doi:10.1109/isic.2012.6398283 · Cited by 10
“In most cases, medication is the preferred treatment option by physicians but it is reported that almost 50% of the patients do not respond well [5].”
BackgroundInfluentialFiltering voluntary motion for pathological tremor compensation
Bo, A. P. L., Poignet, P., & Geny, C. (2009) · 2009 IEEE/RSJ International Conference on Intelligent Robots and Systems · doi:10.1109/iros.2009.5353972 · Cited by 9
Arruda, B. S., Reis, C., Sermon, J. J., Pogosyan, A. H., Brown, P., & Cagnan, H. (2021) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-021-00973-6 · Cited by 10
Zhang, Z., Huo, B., Liu, Y., Dong, A., & Yu, H. (2024) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2024.3450599 · Cited by 12
Grimaldi, G., Manto, M., & Jdaoudi, Y. (2013) · F1000Research · doi:10.12688/f1000research.2-282.v2 · Cited by 9
““"After (1) hearing an acoustic signal, the patient (2) prepared themselves mentally for the execution of movement and (3) performed the task.”
BackgroundMethodInfluentialUpper Limb Neuroprostheses: Recent Advances and Future Directions
Koutsou, A. D., Summa, S., Nasser, B., Gutiérrez-Martínez, J., & Thangaramanujam, M. (2013) · Biosystems & biorobotics · doi:10.1007/978-3-642-38556-8_11 · Cited by 9
Kim, J., Parnell, C., Wichmann, T., & DeWeerth, S. P. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7592136 · Cited by 8
“electrical stimulation [4], [5] or muscle vibration [6], [7]) for patients with treatment-resistant tremor.”
BackgroundOn the use of FES to attenuate tremor by modulating joint impedance
Bó, A. P. L., Coste, C., Poignet, P., Gény, C., & Fattal, C. (2011) · IEEE Conference on Decision and Control (CDC) · doi:10.1109/cdc.2011.6161352 · Cited by 8
“Also, no particular remarks on discomfort due to stimulation were made, which may indicate that using less varying stimulation may be more comfortable to the patient with respect to the method presented in [4].”
Impact of suppressed tremor: Is suppression of proximal joints sufficient?
Zhou, Y., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2018) · IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI ...) · doi:10.1109/bhi.2018.8333369 · Cited by 7
EFFECT OF GYROSCOPE PARAMETERS ON GYROSCOPIC TREMOR SUPPRESSION IN A SINGLE DEGREE OF FREEDOM
Allen, B. C., & Charles, S. K. (2019) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519419500246 · Cited by 8
“Alternative approaches aim at intervening peripherally and include wearable devices that suppress tremor through passive filtering [14-21], active cancelling with an exoskeleton [22, 23], and electrical stimulation [24-29].”
BackgroundWrist tremor suppression based on repetitive control with multi-muscle electrical stimulation
Zhang, Z., Chu, B., Liu, Y., Ren, H., & Owens, D. H. (2019) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2019.12.617 · Cited by 7
Marchis, C. D., Conforto, S., Severini, G., Schmid, M., & D’Alessio, T. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091851 · Cited by 6
Giuliana, G., Mario, M., & Yassin, J. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091899 · Cited by 6
“combination of Functional Electrical Stimulation (FES) [1][2] with a Brain-computer interface (BCI).”
BackgroundInfluentialGrimaldi, G., Fernandez, A., & Manto, M. U. (2013) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-10-100 · Cited by 5
“It has been proposed that surface muscular functional electrical stimulation (FES) represents an alternative approach to drugs or surgery to reduce neurological tremor [13,14].”
BackgroundLee, S. K., Kim, J. W., Kwon, Y., Lee, Y. J., Lee, J. H., Eom, G., Kwon, D., Lee, C. N., Seo, Y.-W., Kim, M. K., et al. (2013) · Journal of Biomedical Engineering Research · doi:10.9718/jber.2013.34.3.129 · Cited by 5
“FES (functional electrical stimulation)를 이용한 진전 억 제 방법을 제안하였다[12-14].”
Metadata
- Authors
- Prochazka, A., Elek, J., & Javidan, M.
- Affiliations
- University of Bonn (DE)
- University of Alberta (CA)
- Venue
- Annals of Biomedical Engineering
- Published
- 1992-03
- Volume
- 20
- Issue
- 2
- Pages
- 205-224
- ISSN
- 0090-6964 (ISSN-L)
- 1573-9686
- OpenAlex
- W2093029239
- Cited by
- 150
- References
- 16
- Citations by year
- 2026: 2 · 2025: 5 · 2024: 7 · 2023: 8 · 2022: 7 · 2021: 10 · 2020: 8 · 2019: 5 · 2018: 5 · 2017: 5 · 2016: 7 · 2015: 13 · 2014: 8 · 2013: 10 · 2012: 4
- FWCI
- 1.25
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Keywords
- pathological tremor, functional electrical stimulation, feedback control, tremor suppression, amplitude modulation, frequency response
- MeSH terms
- Electric Stimulation, Forearm, Humans, Isometric Contraction, Movement, Muscles, Reference Values, Tremor
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.