Relevance and utility of the in-vivo and ex-vivo optical properties of the skin reported in the literature: a review [Invited]
Open accessAbstract
Imaging non-invasively into the human body is currently limited by cost (MRI and CT scan), image resolution (ultrasound), exposure to ionising radiation (CT scan and X-ray), and the requirement for exogenous contrast agents (CT scan and PET scan). Optical imaging has the potential to overcome all these issues but is currently limited by imaging depth due to the scattering and absorption properties of human tissue. Skin is the first barrier encountered by light when imaging non-invasively, and therefore a clear understanding of the way that light interacts with skin is required for progress on optical medical imaging to be made. Here we present a thorough review of the optical properties of human skin measured in-vivo and compare these to the previously collated ex-vivo measurements. Both in-vivo and ex-vivo published data show high inter- and intra-publication variability making definitive answers regarding optical properties at given wavelengths challenging. Overall, variability is highest for ex-vivo absorption measurements with differences of up to 77-fold compared with 9.6-fold for the in-vivo absorption case. The impact of this variation on optical penetration depth and transport mean free path is presented and potential causes of these inconsistencies are discussed. We propose a set of experimental controls and reporting requirements for future measurements. We conclude that a robust in-vivo dataset, measured across a broad spectrum of wavelengths, is required for the development of future technologies that significantly increase the depth of optical imaging.
AI snapshot
Written from the paper’s full text.
TL;DR
This review compiles published in-vivo and ex-vivo optical properties of human skin and finds large inter- and intra-publication variability, especially for ex-vivo absorption. The authors argue that reliable in-vivo data across a broad wavelength range are needed for developing deeper optical imaging.
Key findings
- Variability is highest for ex-vivo absorption measurements, with differences of up to 77-fold, compared with up to 9.6-fold for in-vivo absorption.
- Average in-vivo reduced scattering values beyond 480 nm are approximately 1.7-fold less than ex-vivo values; in the 800–1300 nm range this difference decreases to 1.4-fold.
- Minimum absorption for both ex-vivo and in-vivo data occurs between 800–1300 nm; in this range ex-vivo absorption coefficients are approximately 1.6-fold greater than in-vivo coefficients, compared with a 2.5-fold average difference across the whole spectrum.
- At 1400 nm, the transport mean free path for in-vivo data is maximal at 1.35 mm; light is predicted to travel 1.38 mm before becoming diffuse using in-vivo data versus 0.6 mm using ex-vivo data. Transmission through in-vivo skin peaks at 1400 nm, with ∼93% transmitted beyond the epidermis and ∼24% beyond the dermis.
Methods
- Literature review extracting reported skin optical properties and comparing ex-vivo with in-vivo data; mean absorption and reduced scattering data were collected from nine ex-vivo publications and twelve in-vivo publications after an extensive literature search.
- Data were pooled and averaged regardless of method, separated only by ex-vivo or in-vivo status and location where stated, to determine trends with wavelength.
- Transport mean free path and light transmission were calculated from the pooled coefficients; transmission examples used an epidermis of 0.1 mm and a dermis of 1.83 mm taken from Meglinski et al.
Limitations
- The authors state that both in-vivo and ex-vivo published data show high inter- and intra-publication variability, making definitive optical properties at given wavelengths challenging; in-vivo data have not been published beyond 1600 nm, limiting comparisons at longer wavelengths.
- The paper states that polarisation effects were generally ignored among the reviewed papers, and many publications do not report measurement site or skin pigmentation; standard deviations for individual data are often not published.
- The text makes evident that pooling averages regardless of method may obscure method-specific differences, and ex-vivo sample treatment—time to measurement, storage, compression, exsanguination, hydration—likely contributes to inconsistent coefficients.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 130 cited sources checked · 0 supported · 0 flagged
- Findingown data
Both in-vivo and ex-vivo published data show high inter- and intra-publication variability, making definitive answers regarding optical properties at given wavelengths challenging.
“Both in-vivo and ex-vivo published data show high inter- and intra-publication variability making definitive answers regarding optical properties at given wavelengths challenging.”
- Also stated in: § Introduction · plain
“There is a wide variation in the literature for the optical properties of these different layers and the skin as a whole.”
- Also stated in: § Introduction · plain
“However, despite picking data based on similarities and confidence in the data published, vast variability was still evident.”
Quote in the paperSubstantial claim - Also stated in: § Introduction · plain
- Findingcitation
Measured reduced scattering coefficients have up to 10-fold variability using DRS at 700 nm, ranging from 0.184 mm-1 to 2.19 mm-1.
“Similarly, measured reduced scattering coefficients have up to 10-fold variability using DRS at this wavelength, ranging from 0.184 mm-1 to 2.19 mm-1 [55,57,60,61].”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- cites [57] Doornbos, 1999, Phys. Med. Biol. source
Not checked in the free run.
- cites [60] Hung, 2015, Biomed. Opt. Express source
Not checked in the free run.
- cites [61] Jonasson, 2018, J. Biomed. Opt. source
Not checked in the free run.
- Findingnone given
Published ex-vivo and in-vivo absorption and scattering data are different.
“Published ex-vivo and in-vivo absorption and scattering data are different, however absorption increases beyond 1300 nm for both.”
Quote in the paperSubstantial claim - Findingnone given
When comparing DRS and SFD, reduced scattering coefficients tend to be larger when measured using DRS than SFD, being 1.2 - 1.6-fold greater in the 500 nm - 1000 nm spectrum.
“the reduced scattering coefficients tend to be larger when measured using DRS than SFD being from 1.2 - 1.6-fold greater in the 500 nm - 100 nm spectrum.”
Quote in the paperNot a substantial claim - Findingown data
At a wavelength of 1400 nm, the transport mean free path (TMFP) for in-vivo data is maximal at 1.35 mm.
“Using this calculation, the TMFP through in-vivo and ex-vivo skin using the published data was compared (Fig. 9) and shows that at a wavelength of 1400 nm the TMFP for in-vivo data is maximal (1.35 mm).”
Quote in the paperSubstantial claim
- Findingown data
In general, the ex-vivo and in-vivo absorption and reduced scattering show similar trends.
“In general, the ex-vivo and in-vivo absorption and reduced scattering show similar trends.”
Quote in the paperNot a substantial claim - Findingown data
In the 800 nm -1300 nm spectral range this difference in reduced scattering coefficients decreases to 1.4-fold.
“In the 800 nm -1300 nm spectral range this difference in reduced scattering coefficients decreases to 1.4-fold.”
Quote in the paperNot a substantial claim - Findingreasoning
Maximal transmission through ex-vivo samples is at 1700 nm.
“Maximal transmission through ex-vivo samples is at 1700nm.”
Quote in the paperSubstantial claim - Findingnone given
SFD absorption measurements tended to be between 1.3 - 1.8-fold greater than DRS in the 500 nm - 700 nm spectrum and at 1000 nm.
“SFD absorption measurements tended to be between 1.3 - 1.8-fold greater than DRS in the 500 nm - 700 nm spectrum and at 1000 nm.”
Quote in the paperSubstantial claim - Findingown data
In general, predicted transmission through in-vivo skin is greater than through ex-vivo skin across the measured spectrum.
“Figure 10 shows that in general, transmission through in-vivo skin is predicted to be greater than ex-vivo across the spectrum measured.”
Quote in the paperSubstantial claim - Findingown data
For more than ∼1% of light to be transmitted beyond the dermis, ∼78% of light must be transmitted beyond the epidermis.
“In general, the published data suggests that to achieve more than ∼5% transmission beyond the dermis more than ∼85% of light into the epidermis must be transmitted beyond it and ∼78% of light must be transmitted beyond the epidermis for more than ∼1% of light to be transmitted beyond the dermis.”
Quote in the paperSubstantial claim - Findingown data
For wavelengths greater than 400 nm, all the data reviewed suggest that light will keep traveling in a well-defined direction well into the dermis layer.
“From Fig. 9 it can been seen that at wavelengths greater than 400 nm all the data reviewed here suggest that light will keep traveling in a well-defined direction well into the dermis layer.”
Quote in the paperSubstantial claim - Findingreasoning
The lowest predicted light transmission is in the visible spectrum.
“The lowest amount of light transmitted is predicted to be in the visible spectrum and this is probably due to melanin absorption of visible light in the epidermis.”
Quote in the paperNot a substantial claim - Findingown data
On average, the difference in light transmission beyond the in-vivo and ex-vivo epidermis across the spectrum is approximately 11%.
“On average the difference in light transmission beyond the in-vivo and ex-vivo epidermis across the spectrum is approximately 11%, however this difference increases to 77% when comparing ex-vivo and in-vivo transmission through the epidermis and beyond the dermis.”
Quote in the paperNot a substantial claim - Findingnone given
For ex-vivo measurements, reduced scattering steadily decreases between 1500 nm and 1800 nm.
“Beyond 1450 nm in-vivo reduced scattering increases, however, for ex-vivo measurements it steadily decreases between 1500 nm and 1800nm.”
Quote in the paperSubstantial claim - Findingown data
Minimum absorption for both ex-vivo and in-vivo data is between 800 - 1300 nm.
“Minimum absorption for both ex-vivo and in-vivo data is between 800 - 1300 nm.”
Quote in the paperSubstantial claim - Findingown data
The difference in light transmission through the epidermis and beyond the dermis between ex-vivo and in-vivo is 77%.
“On average the difference in light transmission beyond the in-vivo and ex-vivo epidermis across the spectrum is approximately 11%, however this difference increases to 77% when comparing ex-vivo and in-vivo transmission through the epidermis and beyond the dermis.”
Quote in the paperSubstantial claim - Findingnone given
Ex-vivo, between 1600 nm and 1800 nm, troughs in reduced scattering and absorption are coincidental with peaks in TMFP and percentage of light transmission.
“Ex-vivo, between 1600 nm – 1800nm, troughs in reduced scattering and absorption are coincidental with peaks in TMFP and percentage of light transmission suggesting that this wavelength range might have the potential for deeper imaging than at shorter wavelengths.”
Quote in the paperSubstantial claim - Findingnone given
Beyond 1450 nm, in-vivo reduced scattering increases.
“Beyond 1450 nm in-vivo reduced scattering increases, however, for ex-vivo measurements it steadily decreases between 1500 nm and 1800nm.”
Quote in the paperSubstantial claim - Findingreasoning
Absorption coefficients are much greater ex-vivo than in-vivo.
“However, absorption coefficients are much greater ex-vivo than in-vivo and this is probably due to sample preparation techniques and rehydration leading to unrealistic water content and greater absorption by water for ex-vivo samples.”
Quote in the paperSubstantial claim - Findingown data
Minimal transmission in the visible region is ∼60% using in-vivo epidermis data compared with ∼42% using ex-vivo epidermis values.
“However, minimal transmission in this region is ∼60% using the in-vivo epidermis data compared with ∼42% when using the ex-vivo epidermis values.”
Quote in the paperNot a substantial claim - Findingreasoning
At these rates of transmission, virtually no light is transmitted beyond the dermis.
“At these rates of transmission virtually no light is transmitted beyond the dermis.”
Quote in the paperSubstantial claim - Findingnone given
At 800 nm there was little difference between DRS and SFD.
“However, at 800 nm there was little difference between the two methods.”
Quote in the paperNot a substantial claim - Findingown data
In the 800-1300 nm spectral range the ex-vivo absorption coefficients are approximately 1.6-fold greater than in-vivo absorption coefficients (compared with a 2.5-fold average difference across the whole spectrum).
“In this spectral range the ex-vivo absorption coefficients are approximately 1.6-fold greater than in-vivo absorption coefficients (compared with a 2.5-fold average difference across the whole spectrum).”
Quote in the paperNot a substantial claim - Findingcitation
There are vast discrepancies in the optical properties of skin for measurements taken in-vivo, particularly with regards to scattering.
“As shown for ex-vivo data, we report vast discrepancies in the optical properties of skin for measurements taken in-vivo, particularly with regards to scattering [8].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [8] Mignon, 2018, Biomed. Opt. Express source
Not checked in the free run.
- Findingnone given
Absorption increases beyond 1300 nm for both in-vivo and ex-vivo data.
“Published ex-vivo and in-vivo absorption and scattering data are different, however absorption increases beyond 1300 nm for both.”
Quote in the paperSubstantial claim - Findingown data
To achieve more than ∼5% transmission beyond the dermis, more than ∼85% of light into the epidermis must be transmitted beyond it.
“In general, the published data suggests that to achieve more than ∼5% transmission beyond the dermis more than ∼85% of light into the epidermis must be transmitted beyond it and ∼78% of light must be transmitted beyond the epidermis for more than ∼1% of light to be transmitted beyond the dermis.”
Quote in the paperSubstantial claim - Findingnone given
Transmission through in-vivo skin peaks at 1400 nm.
“Transmission through in-vivo skin peaks at 1400 nm, with ∼93% of the light into the epidermis being transmitted beyond it and ∼24% of this light being transmitted beyond the dermis.”
Quote in the paperSubstantial claim - Findingnone given
Water absorption is dominant at 1460 nm for both in-vivo and ex-vivo data.
“For both in-vivo and ex-vivo data melanin absorption is dominant up to 600 nm and water absorption is dominant at 1460 nm showing the importance of avoiding chromophore absorption peaks for diagnostic and imaging applications.”
Quote in the paperSubstantial claim - Findingreasoning
A corresponding peak in transmission is not observed for ex-vivo skin.
“A corresponding peak in transmission is not observed for ex-vivo skin, in fact between 1300 nm and 1500 nm transmission decreased from ∼85% to ∼7 2% transmission trough the epidermis and ∼4.5% to ∼0.2% transmission beyond the dermis.”
Quote in the paperSubstantial claim - Findingnone given
For both in-vivo and ex-vivo data, melanin absorption is dominant up to 600 nm.
“For both in-vivo and ex-vivo data melanin absorption is dominant up to 600 nm and water absorption is dominant at 1460 nm showing the importance of avoiding chromophore absorption peaks for diagnostic and imaging applications.”
Quote in the paperSubstantial claim - Findingnone given
Polarisation effects were generally ignored among the papers reviewed here.
“Polarisation effects were generally ignored amongst the papers reviewed here and so we are unable to draw conclusions.”
Quote in the paperSubstantial claim - Premisenone given
Four optical windows in the NIR have been described: I (650-950 nm), II (1100-1350 nm), III (1600-1870 nm) and IV (2100-2300 nm).
“Four optical windows in the NIR have been described: I (650 nm - 950 nm), II (1100 nm -1350 nm), III (1600 nm - 1870nm) and IV (2100 nm - 2300 nm).”
Quote in the paperSubstantial claim - Premisecitation
For commercial OCT imaging of the skin, light at a wavelength of 1300 nm is commonly used.
“For commercial optical coherence tomography (OCT) imaging of the skin, light at a wavelength of 1300 nm is commonly used.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [19] Olsen, 2018, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
OCT is the only method currently used to image deeply (up to 1 mm in the skin) among OCT, reflectance confocal microscopy and multiphoton microscopy.
“Optical coherence tomography (OCT), reflectance confocal microscopy and multiphoton microscopy have become advanced enough for clinical imaging, diagnosis and monitoring of skin, however, OCT is the only method currently used to image deeply (up to 1 mm in the skin) [4].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [4] Lentsch, 2022, J. Invest. Dermatol. source
Not checked in the free run.
- Premisecitation
As wavelength increases from the visible towards the near infra-red, absorption by water, collagen and lipids becomes more prominent despite not being significant in the visible part of the spectrum.
“As the wavelength of light increases from the visible towards the near infra-red, absorption by water, collagen and lipids become more prominent despite not being significant in the visible part of the spectrum (Fig. 3).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [18] Wilson, 2015, J. Biomed. Opt. source
Not checked in the free run.
- Premisenone given
In the NIR region, water has the greatest effect on absorption.
“In the NIR region water has the greatest effect on absorption.”
Quote in the paperNot a substantial claim - Premisecitation
Light scattering and absorption differ at various locations on the body due to the heterogeneous nature of skin composition.
“Light scattering and absorption also differ at various locations due to the heterogeneous nature of skin composition [5–7].”
Quote in the paperSubstantial claim0 of 3 sources checked - Premisecitation
At 1300 nm, water absorption is low enough to acquire 500 μm image depth penetration in the dermis.
“At this wavelength, water absorption is low enough to acquire 500 µm image depth penetration in the dermis [19].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [19] Olsen, 2018, J. Biomed. Opt. source
Not checked in the free run.
- Interpretationreasoning
The golden window where absorption and scattering are low and imaging depth is optimal is in the NIR spectral region.
“The golden window where absorption and scattering are low and, importantly, imaging depth is optimal is in the NIR spectral region.”
Quote in the paperSubstantial claim - Premisecitation
For rat brain tissue imaging, optical window III was shown to provide greatest imaging depth due to reduced scattering and absorption in this wavelength range.
“The relationship between transmission and tissue thickness for rat brain tissue imaging was determined and optical window III was shown to provide greatest imaging depth due to reduced scattering and absorption in this wavelength range [30].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [30] Shi, 2016, J. Biophotonics source
Not checked in the free run.
- Premisecitation
Using ex-vivo rat tissue, Golovynskyi et al. showed that NIR III has the potential to be optimal for imaging skin due to reduced water absorption in this region.
“However, using ex-vivo rat tissue Golovynskyi et al., showed that NIR III has the potential to be optimal for imaging skin due to reduced water absorption in this region [29].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [29] Golovynskyi, 2018, J. Biophotonics source
Not checked in the free run.
- Premisenone given
The major chromophores that contribute to absorption and scattering in the skin are haemoglobin (oxygenated and deoxygenated), melanin, water and lipid.
“The major chromophores that contribute to absorption and scattering in the skin are haemoglobin (oxygenated and deoxygenated), melanin, water and lipid.”
Quote in the paperSubstantial claim - Premisecitation
In the epidermis, scattering decreased from ∼11 mm-1 at ∼360 nm to 2.3 mm-1 at 1600 nm.
“In the epidermis, scattering decreased from ∼11 mm-1 at ∼360 nm to 2.3 mm-1 at 1600 nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
The reduced scattering coefficients at 633 nm ranged from 2.73 mm−1 to 3.21 mm−1 in the epidermal/dermal layers tested from 5 samples compared with 1.26 mm−1 in the dermis.
“Similarly, the reduced scattering coefficients at 633 nm ranged from 2.73 mm−1 to 3.21 mm−1 in the epidermal/dermal layers tested from 5 samples compared with 1.26 mm−1 in the dermis.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Simpson, 1998, Phys. Med. Biol. source
Not checked in the free run.
- Premisecitation
Reduced scattering coefficients decreased with increasing wavelength in all skin layers.
“Reduced scattering coefficients decreased with the increasing wavelength in all skin layers.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Collagen, predominantly found in the dermal skin layer, has absorption peaks at 1050 nm, 1200 nm, 1500 nm, and 1725 nm.
“Collagen, predominantly found in the dermal skin layer, has absorption peaks at 1050 nm, 1200 nm, 1500 nm, and 1725nm [45].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [45] Nachabé, 2011, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Under the same conditions, scattering coefficients varied from 2.13 mm-1 to 6.91 mm-1 amongst the donors.
“Under the same conditions, scattering coefficients varied from 2.13 mm-1 - 6.91 mm-1 amongst the donors.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
- Premisecitation
Scattering tended to decrease across the spectrum measured from 3.4 mm-1 to 0.6 mm-1.
“Scattering tended to decrease across the spectrum measured from 3.4 mm-1 to 0.6 mm-1.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
- Premisecitation
Monem et al. showed comparative results for single and double integrating sphere optical property measurements of phantoms.
“Monem et al., showed comparative results for single and double integrating sphere optical property measurements of phantoms [48].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [48] Monem, 2015, Appl. Phys. B source
Not checked in the free run.
- Premisecitation
Scattering in the epidermis was shown to be higher than in the remaining layers across the spectrum tested.
“Scattering in the epidermis was shown to be higher than in the remaining layers across the spectrum tested.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
- Premisenone given
In the visible spectrum, melanin and haemoglobin have most effect on absorption.
“In the visible spectrum, melanin and haemoglobin have most effect on absorption.”
Quote in the paperNot a substantial claim - Premisecitation
During Prahl's ex-vivo measurement, the thickness of the sample changed due to dehydration, affecting the contribution of water to the coefficients being measured.
“During measurement, the thickness of the sample changed due to dehydration, therefore affecting the contribution of water to the coefficients being measured.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] Prahl, 1998
Not checked in the free run.
- Premisecitation
At 500 nm under varying pressure, even amongst samples from a single donor, the optical properties measured were variable, ranging from 14% difference between 3 absorbance measurements taken for an individual to 48%, and up to 95% difference between these measurements for scattering.
“However, at 500 nm under varying pressure, even amongst samples from a single donor, the optical properties measured were variable, ranging from 14% difference between 3 absorbance measurements taken for an individual to 48%, and up to 95% difference between these measurements for scattering.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
- Premisecitation
Dermal and subcutaneous tissue scattering coefficients were similar across the spectrum measured, ranging from ∼4.5 mm-1 and ∼3.5 mm-1 at ∼500 nm to 1.7 mm-1 and ∼1.5 mm-1 at 1600 nm respectively.
“Dermal and subcutaneous tissue scattering coefficients were similar across the spectrum measured, ranging from ∼4.5 mm-1 and ∼3.5 mm-1 at ∼500 nm to 1.7 mm-1 and ∼1.5 mm-1 at 1600 nm respectively.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
The calculated reduced scattering coefficients tend to become less similar with increasing sample thickness beyond approximately 2.4 mm.
“The calculated reduced scattering coefficients tend to become less similar with increasing sample thickness beyond approximately 2.4 mm.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [48] Monem, 2015, Appl. Phys. B source
Not checked in the free run.
- Premisecitation
Troy et al. observed variation in optical properties amongst the population.
“Variation in optical properties amongst the population was observed.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [54] Troy, 2001, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
An increase of reduced scattering coefficient at 1450 nm in the region of strong water absorption was noted.
“An increase of reduced scattering coefficient at 1450 nm in the region of strong water absorption was noted.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
- Interpretationreasoning
Although the epidermis in these samples contains small amounts of melanin, the increased scattering compared to the dermis and subcutaneous tissue layers can be explained by the difference in refractive index of melanin compared to its surroundings.
“Although the epidermis in these samples contains small amounts of melanin, the increased scattering compared to the dermis and subcutaneous tissue layers can be explained by the difference in refractive index of melanin compared to its surroundings.”
Quote in the paperSubstantial claim - Premisecitation
As age increased the reduced scattering coefficient decreased.
“As age increased the reduced scattering coefficient decreased”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [61] Jonasson, 2018, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Beyond 980 nm, light has the potential to penetrate the subcutaneous tissue travelling a distance of ∼1.65 mm.
“Beyond 980 nm, light has the potential to penetrate the subcutaneous tissue travelling a distance of ∼1.65 mm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [53] Shimojo, 2020, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Jonasson et al. found that reduced scattering coefficient decreased in the spectrum of 475 nm - 850 nm from 3.16 mm−1 to 1.13 mm−1.
“The reduced scattering coefficient decreased in the spectrum of 475 nm - 850 nm from 3.16 mm−1 to 1.13 mm−1.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [61] Jonasson, 2018, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Scattering was greatest in the forehead and least in the arm at 630, 660, and 700 nm (Doornbos et al.).
“However, scattering was greatest in the forehead and least in the arm at each of these wavelengths.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [57] Doornbos, 1999, Phys. Med. Biol. source
Not checked in the free run.
- Premisecitation
MC simulations showed that the interrogation depth at 500 nm was at least 46% less than at 900 nm.
“MC simulations were used to determine the penetration depth of photons in the skin and showed that the interrogation depth at 500 nm was at least 46% less than at 900 nm.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- Premisecitation
Minimal absorption was observed at 1100 nm for each skin layer.
“Minimal absorption was observed at 1100 nm for each skin layer.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Tseng et al. showed that there was a 2-fold decrease in absorption as the wavelength increased from 500 nm to 600 nm.
“Taking measurements from 18 subjects in the wavelength range 500 nm-1000 nm, Tseng et al., showed that there was a 2-fold decrease in absorption as the wavelength increased from 500 nm to 600 nm [55].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- Premisecitation
Skin location and colour have important effects on measured absorption and scattering in the skin.
“These authors showed that skin location and colour have important effects on measured absorption and scattering in the skin.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- Premisecitation
A small peak in absorption was observed at ∼970 nm for all skin-types corresponding with water absorption.
“A small peak in absorption was observed at ∼970 nm for all skin-types corresponding with water absorption.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- Premisecitation
In the visible region, both melanin in the epidermis and haemoglobin in the blood supply to the dermis contributed to the high absorption coefficients in some subjects.
“In the visible region both melanin in the epidermis and haemoglobin in the blood supply to the dermis contributed to the high absorption coefficients in some subjects.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- Premisecitation
Concentrations of tissue chromophores including melanin, haemoglobin, water and lipid can be determined using DRS measurements.
“The concentrations of tissue chromophores including melanin, haemoglobin water and lipid can also be determined using DRS measurements [56].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [56] Hsu, 2015, Biomed. Opt. Express source
Not checked in the free run.
- Premisecitation
At 800 nm, the absorption coefficients described by Tseng et al. were 5-fold greater than those measured by Hsu et al.
“At 800 nm the absorption coefficients described by Tseng et al., were 5-fold greater than those measured by Hsu et al. [55,56].”
Quote in the paperNot a substantial claim0 of 2 sources checked - Premisecitation
Peaks in absorbance were observed at 970 nm and 1150 nm - 1200 nm corresponding with water absorption.
“Peaks in absorbance were observed at 970 nm and 1150 nm - 1200 nm corresponding with water absorption.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [60] Hung, 2015, Biomed. Opt. Express source
Not checked in the free run.
- Premisecitation
At 405-532 nm, light penetrates to a maximum depth of ∼0.3 mm reaching the upper part of dermis.
“At 405 - 532 nm, the light penetrates to a maximum depth of ∼0.3 mm reaching the upper part of dermis [53].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [53] Shimojo, 2020, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
Skin location has an important effect on absorption and scattering coefficients (Doornbos et al.).
“Like Tseng et al., skin location was shown to have an important effect on absorption and scattering coefficients by Doornbos et al. [55,57].”
Quote in the paperNot a substantial claim0 of 2 sources checked - Interpretationreasoning
The huge variation in values reported in the literature makes producing reliable values for light transmission and TMFP at different wavelengths challenging.
“However, the huge variation in values reported in the literature makes producing reliable values for light transmission and TMFP at different wavelengths challenging.”
Quote in the paperSubstantial claim - Interpretationreasoning
Using in-vivo values predicts that light travels further into the skin than using ex-vivo values over the 400-1600 nm wavelength range.
“The implications of using the values measured in-vivo is that light will be predicted to travel further into the skin compared to the values measured ex-vivo in the 400 nm – 1600 nm wavelength range.”
Quote in the paperSubstantial claim - Premisecitation
Baseline scattering varied amongst measurement location (Phan et al.).
“Baseline scattering varied amongst measurement location.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [65] Phan, 2021, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
A significant peak was observed at 640 nm for the in-vivo data published by Hsu et al.
“However, a significant peak was observed at 640 nm for the in-vivo data published by Hsu et al., potentially representing scattering due to collagen bundles in the skin [56].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [56] Hsu, 2015, Biomed. Opt. Express source
Not checked in the free run.
- Interpretationreasoning
The decrease in ex-vivo transmission may be attributed to water absorption of light in ex-vivo skin due to rehydration during sample preparation and unrealistic water content.
“This decrease in transmission may be attributed to water absorption of light in ex-vivo skin, due to rehydration during sample preparation and unrealistic water content.”
Quote in the paperSubstantial claim - Interpretationreasoning
The inconsistency in published values has the potential to affect the design of deep imaging systems using longer wavelengths.
“This inconsistency in the published values has the potential to affect the design of deep imaging systems using longer wavelengths and needs to be investigated further.”
Quote in the paperSubstantial claim - Interpretationreasoning
Avoiding chromophore absorption peaks is important for diagnostic and imaging applications.
“For both in-vivo and ex-vivo data melanin absorption is dominant up to 600 nm and water absorption is dominant at 1460 nm showing the importance of avoiding chromophore absorption peaks for diagnostic and imaging applications.”
Quote in the paperSubstantial claim - Premisecitation
Changes in the polarisation state of light can be used to provide information about tissues, including the superficial layers of the skin, by characterising the depolarisation of light as it propagates.
“In addition, changes in the polarisation state of light can be used to provide information about tissues, including the superficial layers of the skin, by characterising the depolarisation of light as it propagates [88,89].”
Quote in the paperSubstantial claim0 of 2 sources checked - Interpretationreasoning
Comparisons cannot be drawn beyond 1600 nm between ex-vivo and in-vivo data.
“Therefore, comparisons cannot be drawn beyond 1600 nm between these data.”
Quote in the paperSubstantial claim - Premisecitation
Saager et al. showed clear differences between measured absorption and reduced scattering coefficients when subjects were grouped based upon skin pigmentation.
“These authors showed that there were clear differences between both measured absorption and reduced scattering coefficients when subjects were grouped based upon their skin pigmentation.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [64] Saager, 2015, J. Biomed. Opt. source
Not checked in the free run.
- Interpretationreasoning
The 1600-1800 nm range might have the potential for deeper imaging than at shorter wavelengths.
“Ex-vivo, between 1600 nm – 1800nm, troughs in reduced scattering and absorption are coincidental with peaks in TMFP and percentage of light transmission suggesting that this wavelength range might have the potential for deeper imaging than at shorter wavelengths.”
Quote in the paperSubstantial claim - Premisecitation
Phan et al. found that the least variation in scattering between subjects was at 851 nm.
“They found that the least variation in scattering between subjects was at 851 nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [65] Phan, 2021, J. Biomed. Opt. source
Not checked in the free run.
- Interpretationreasoning
The greater ex-vivo absorption coefficients are probably due to sample preparation techniques and rehydration leading to unrealistic water content and greater absorption by water for ex-vivo samples.
“However, absorption coefficients are much greater ex-vivo than in-vivo and this is probably due to sample preparation techniques and rehydration leading to unrealistic water content and greater absorption by water for ex-vivo samples.”
Quote in the paperSubstantial claim - Interpretationreasoning
The decrease in inter-subject variation with increasing wavelength may be attributed to melanin absorption being lowest at 851 nm, which affects absorption measurements and reduced scattering measurements due to deeper photon penetration and scattering by skin components at greater depths.
“This may be attributed to melanin absorption being lowest at 851 nm (the longest wavelength measured by these authors), which not only affects absorption measurements at this wavelength but also reduced scattering measurements due to the ability of photons to penetrate deeper, and therefore to be scattered by skin components at greater depths.”
Quote in the paperNot a substantial claim - Study design and conductnone given
For published in-vivo data, optical coefficients were only measured up to 1600 nm.
“Since the optical coefficients were only measured up to 1600 nm for the published in-vivo data conclusions on transmission at even longer wavelengths cannot be drawn however, it would be useful to determine the potential of longer wavelengths for increased transmission.”
Quote in the paperSubstantial claim - Premisecitation
Both reduced scattering and absorption tended to decrease with increasing wavelength (Saager et al.).
“Both reduced scattering and absorption tended to decrease with increasing wavelength.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [64] Saager, 2015, J. Biomed. Opt. source
Not checked in the free run.
- Interpretationreasoning
The low transmission in the visible spectrum is probably due to melanin absorption of visible light in the epidermis.
“The lowest amount of light transmitted is predicted to be in the visible spectrum and this is probably due to melanin absorption of visible light in the epidermis.”
Quote in the paperNot a substantial claim - Premisecitation
What happens first, loss of ballistic photons due to scattering or depolarisation of light as it propagates, is dependent upon the size (essentially the g factor), shape and concentration of the scatterer.
“In reality, what happens first, loss of ballistic photons due to scattering or depolarisation of light as it propagates, is dependent upon the size (essentially the g factor), shape and concentration of the scatterer [92].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] Morgan, 1997, Appl. Opt. source
Not checked in the free run.
- Study design and conductreasoning
We are unable to draw conclusions about polarisation effects from the reviewed papers.
“Polarisation effects were generally ignored amongst the papers reviewed here and so we are unable to draw conclusions.”
Quote in the paperSubstantial claim - Premisecitation
OCT currently images to depths of approximately 1 mm in the skin.
“Despite the deeply scattering nature of the human tissues, light has been used for multiple biological applications including cancer diagnosis and studying cellular changes like apoptosis and it can provide structural information deriving from tissue boundaries, cells and organelles with OCT currently imaging to depths of approximately 1 mm in the skin [4].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [4] Lentsch, 2022, J. Invest. Dermatol. source
Not checked in the free run.
- May combine two claims. 65
- The quote only partly states this claim. 1
- Classified as probably not a substantial claim. 23
May combine two claims.
Findingnone givenOverall, variability is highest for ex-vivo absorption measurements, with differences of up to 77-fold compared with 9.6-fold for the in-vivo absorption case.
“Overall, variability is highest for ex-vivo absorption measurements with differences of up to 77-fold compared with 9.6-fold for the in-vivo absorption case.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenSeveral authors using integrating sphere techniques to determine ex-vivo skin optical properties present data that all differ markedly, with 2-10-fold differences between absorption coefficients across a range of wavelengths and 2-3-fold variation in scattering coefficients.
“Several authors have used integrating sphere techniques to determine the optical properties of the skin ex-vivo [15,21,49–54]. However, the data that they present all differ markedly, with 2–10-fold differences between absorption coefficients across a range of wavelengths and 2–3-fold variation in scattering coefficients.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationNumerous authors have measured optical properties of skin on the forearm using DRS and absorption coefficients vary 9-fold amongst publications from 0.01 mm-1 to 0.09 mm-1 at 700 nm.
“Numerous authors have measured the optical properties of skin on the forearm using DRS and the absorption coefficients vary 9-fold amongst the publications from 0.01 mm-1 - 0.09 mm-1 at 700 nm [37,46,55,57,60].”
Quote in the paperNot a substantial claim0 of 5 sources checked- cites [37] Torricelli, 2001, Phys. Med. Biol. source
Not checked in the free run.
- cites [46] Zonios, 2006, Opt. Express source
Not checked in the free run.
- cites [55] Tseng, 2009, Opt. Express source
Not checked in the free run.
- cites [57] Doornbos, 1999, Phys. Med. Biol. source
Not checked in the free run.
- cites [60] Hung, 2015, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
Findingown dataThe wavelength range for ex-vivo data was greater than for in-vivo data, ranging from 400 nm - 2200 nm and 390 nm - 1600 nm respectively.
“The wavelength range for ex-vivo data was greater than for in-vivo data, ranging from 400 nm - 2200 nm and 390 nm - 1600 nm respectively.”
Quote in the paperSubstantial claimThe quote only partly states this claim.
Findingown dataBy averaging the ex-vivo published data, the 640 nm peak became more apparent.
“By averaging the ex-vivo published data, this peak became more apparent.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor both data sets, a large scattering peak is observed at 640 nm; this is the most significant peak amongst the in-vivo data.
“For both data sets, a large scattering peak is observed at 640 nm. This is the most significant peak amongst the in-vivo data.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingreasoningBetween 1300 nm and 1500 nm, ex-vivo transmission decreased from ∼85% to ∼72% through the epidermis and from ∼4.5% to ∼0.2% beyond the dermis.
“A corresponding peak in transmission is not observed for ex-vivo skin, in fact between 1300 nm and 1500 nm transmission decreased from ∼85% to ∼7 2% transmission trough the epidermis and ∼4.5% to ∼0.2% transmission beyond the dermis.”
Quote in the paperSubstantial claimMay combine two claims.
FindingreasoningPublished data available for skin in the in-vivo situation are variable and incomplete.
“It is clear that the published data available for skin in the in-vivo situation are variable and incomplete and a comprehensive dataset covering a broad set of wavelength measurements is required for optimising skin treatments and imaging.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenThis 640 nm peak does not tend to be seen in the published data when taken individually; reduced scattering coefficients generally reduce linearly with increasing wavelength.
“This does not tend to be seen in the published data when taken individually; reduced scattering coefficients generally reduce linearly with increasing wavelength.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingreasoningAt 1700 nm, TMFP is at its greatest and scattering and absorption troughs are coincident.
“At this point TMFP is at its greatest and scattering and absorption troughs are coincident suggesting that this wavelength range might have the potential to provide deeper imaging and better resolution than at shorter wavelengths.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenAt 1400 nm, ∼93% of light into the epidermis is transmitted beyond it and ∼24% of this light is transmitted beyond the dermis.
“Transmission through in-vivo skin peaks at 1400 nm, with ∼93% of the light into the epidermis being transmitted beyond it and ∼24% of this light being transmitted beyond the dermis.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataBoth in-vivo and ex-vivo absorption and reduced scattering data display multiple peaks, some being true resonant peaks whilst other represent the noise amongst the data.
“Both in-vivo and ex-vivo absorption and reduced scattering data display multiple peaks, some being true resonant peaks whilst other represent the noise amongst the data.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingreasoningAt 1400 nm, using in-vivo data, light is expected to travel 1.38 mm into the skin before becoming diffuse, compared with 0.6 mm using ex-vivo data.
“However, if one uses the in-vivo data recorded at 1400 nm, light would be expected to travel 1.38 mm in to the skin before it became diffuse compared to 0.6 mm for the same wavelength using the ex-vivo data.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationNIR II and III have been exploited for deep-tissue high-resolution optical imaging in other tissues.
“Similar has been shown to be true for other tissues, and NIR II and III have also been exploited for deep-tissue high-resolution optical imaging [27,31–36].”
Quote in the paperNot a substantial claim0 of 7 sources checked- cites [27] Sordillo, 2014, J. Biomed. Opt. source
Not checked in the free run.
- cites [31] Smith, 2009, Nat. Nanotechnol. source
Not checked in the free run.
- cites [32] Poelman, 2020, J. Appl. Phys. source
Not checked in the free run.
- cites [33] Hemmer, 2016, Nanoscale Horiz. source
Not checked in the free run.
- cites [34] Shaw, 2022, Front. Chem. source
Not checked in the free run.
- cites [35] Kenry, 2018, Adv. Mater. source
Not checked in the free run.
- cites [36] Zhao, 2018, J. Mater. Chem. B source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationMore than 200 papers describing the optical properties of tissues both in-vivo and ex-vivo from multiple animal sources but not focussed on the skin have been published annually since 2011.
“There has been a recent surge in publications describing the optical properties of tissues both in-vivo and ex-vivo from multiple animal sources but not focussed on the skin, with more than 200 papers being published annually since 2011 [1,2].”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationTechnologies currently available for imaging the layers of the skin are limited in depth, and therefore for skin diseases like skin cancer, biopsies are still required for reliable diagnosis.
“Technologies currently available for imaging the layers of the skin, are limited in depth and therefore for skin diseases, like skin cancer, biopsies are still required for reliable diagnosis [3].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3] Li, 2021, JID Innov. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationHuman skin is made up of approximately 65% water and 9% lipid.
“In general terms, human skin is made up of approximately 65% water and 9% lipid [37].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [37] Torricelli, 2001, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
PremisecitationBetween 600 nm and 700 nm, absorption decreases and scattering becomes dominated by collagen and elastin bundles in the dermis.
“Between 600 nm and 700 nm absorption decreases and scattering becomes dominated by collagen and elastin bundles in the dermis [16].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16] Gajinov, 2010, Serbian J. Dermatol. Venereol. source
Not checked in the free run.
May combine two claims.
PremisecitationFor visible and near infra-red wavelengths, scattering dominates over absorption in skin, being 100-1000 times stronger (and reduced scattering being 10-100 times stronger).
“For visible and near infra-red (NIR) wavelengths, scattering dominates over absorption in skin, being 100 - 1000 times stronger (and reduced scattering being 10 - 100 times stronger) [10].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [10] Cheong, 1990, IEEE J. Quantum Electron. source
Not checked in the free run.
May combine two claims.
PremisecitationIn skin, chromophores such as melanin in the epidermis and haemoglobin in the dermis are the major absorbers of light, and the absorption spectra depend on the amount of these chromophores present.
“In the skin, chromophores such as melanin in the epidermis and haemoglobin in the dermis are the major absorbers of light and the absorption spectra will depend on the amount of these chromophores present [16].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [16] Gajinov, 2010, Serbian J. Dermatol. Venereol. source
Not checked in the free run.
May combine two claims.
PremisecitationAt visible wavelengths, light is absorbed by melanin and haemoglobin, whereas structures such as cell components and melanosomes of less than 300 μm give rise to scatter.
“At visible wavelengths, light is absorbed by melanin and haemoglobin; whereas structures, such as cell components and melanosomes of less than 300 µm, give rise to scatter [10].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Cheong, 1990, IEEE J. Quantum Electron. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationAge and obesity affect skin composition, with a doubling of lipid content reported in abdominal subcutaneous fat content with high levels of obesity.
“However, factors such as age and obesity affect the composition with a doubling of the lipid content being reported in the abdominal subcutaneous fat content with high levels of obesity [38].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [38] Rodriguez, 2022, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe skin is generally considered to be stacked layers of epidermis, dermis, and subcutaneous tissue, each with varying thickness and different optical characteristics.
“The skin is generally considered to be stacked layers of epidermis, dermis, and subcutaneous tissue each with varying thickness and different optical characteristics associated within each layer [8] (Fig. 1).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [8] Mignon, 2018, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
PremisecitationNIR I has been well characterised for imaging human tissue, and NIR II, III and IV have been shown to offer increased penetration and optical transparency over NIR I in a range of ex-vivo and in-vivo tissue including normal and malignant breast and prostate tissue, human aorta, pig and mouse brain and chicken and rat tissue.
“NIR I has been well characterised for imaging human tissue [12,18,20–25], and NIR II, III and IV have been shown to offer increased penetration and optical transparency over NIR I in a range of ex-vivo and in-vivo tissue including normal and malignant breast and prostate tissue, human aorta, pig and mouse brain and chicken and rat tissue [18,26–28].”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [18] Wilson, 2015, J. Biomed. Opt. source
Not checked in the free run.
- cites [26] Sordillo, 2015, J. Biomed. Opt. source
Not checked in the free run.
- cites [27] Sordillo, 2014, J. Biomed. Opt. source
Not checked in the free run.
- cites [28] Hong, 2014, Nat. Photonics source
Not checked in the free run.
May combine two claims.
PremisecitationFor skin, the anisotropy factor g tends to be between 0.8 and 0.95, i.e. scattering occurs mostly in the forward direction.
“For skin, g tends to be between 0.8-0.95, i.e. scattering occurs mostly in the forward direction.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Cheong, 1990, IEEE J. Quantum Electron. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningIn general, the scattering coefficient is much higher at shorter wavelengths and changes relatively smoothly with wavelength.
“In general, this leads to a much higher scattering coefficient at the shorter wavelengths and, overall, a scattering coefficient that changes relatively smoothly with wavelength.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone givenIn the Rayleigh regime the scattering intensity is proportional to λ-4, whereas in the Mie regime the relationship is typically λ-1.
“In the Rayleigh regime the scattering intensity is proportional to λ-4, whereas in the Mie regime the relationship is more complex but is typically λ-1.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone givenAt NIR wavelengths, scattering of photons in skin tends to decrease and these wavelengths are able to penetrate deepest into the skin.
“At NIR wavelengths, scattering of photons in skin tends to decrease and these wavelengths are able to penetrate deepest into the skin.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationMelanosomes affect scattering in the 600 nm to 700 nm range, and due to their high refractive index, their scattering coefficient is at least one order of magnitude higher than their absorption coefficient at 650 nm, even though melanin is a strong absorber.
“Melanosomes also effect scattering in the 600 nm to 700 nm range and, due to their high refractive index in comparison to their surroundings, their scattering coefficient is at least one order of magnitude higher than their absorption coefficient at 650 nm, even though melanin is a strong absorber [17].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17] Song, 2017, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
PremisecitationLipid absorption peaks at 900 nm, with minor absorption peaks at 1040 nm, 1210 nm, 1400 nm, 1730 nm and 1760 nm.
“Lipid absorption, which greatly affect the subcutaneous skin layer, peaks at 900 nm, but it also has minor absorption peaks at 1040 nm, 1210 nm, 1400 nm, 1730nm and 1760nm [18,43].”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationTwo absorption peaks were observed at 1460 nm and 1950 nm with absorption being 2.2 mm-1 at 2050 nm.
“Two absorption peaks were observed at 1460 nm and 1950nm with absorption being 2.2 mm-1 at 2050nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [54] Troy, 2001, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationAbsorption in the epidermis was greater than in the dermis and subcutaneous tissue and decreased with wavelength except for a peak at 1450 nm corresponding with water absorption.
“Similarly, these authors described absorption in the epidermis being greater than in the dermis and subcutaneous tissue and that this decreased with wavelength except for a peak at 1450 nm corresponding with water absorption.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenIn the visible spectrum, water absorption is negligible.
“In the visible spectrum water absorption is negligible.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPrahl suggested that the optical depth was quite sensitive to (bloodless) tissue coagulation, which changes in the first 50-75 seconds of exposure to light.
“Prahl suggested that the optical depth was quite sensitive to (bloodless) tissue coagulation, which changes in the first 50 - 75 seconds of exposure to light [15].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] Prahl, 1998
Not checked in the free run.
May combine two claims.
PremisecitationBetween 1000 nm and 2200 nm, scattering remained between 0.8 mm-1 and 1.3 mm-1 and tended to decrease with increasing wavelength.
“Between 1000 nm and 2200 nm scattering remained between 0.8 mm-1 and 1.3 mm-1 and tended to decrease with increasing wavelength.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [54] Troy, 2001, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationAlthough haemoglobin is a strong absorber, its total effect on skin absorption is modest because blood makes up only a few percent of total skin volume and melanin is confined to the thin epidermal layer.
“For example, although melanin absorbs strongly at visible wavelengths, it is only located in the very thin epidermal layer of the skin. Blood only makes up a few percent of the total skin volume, and is different depending upon the layer in question, therefore, although haemoglobin is a strong absorber, its total effect is modest [41].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [41] Campolmi, 2016, Med. Princ. Pract. source
Not checked in the free run.
May combine two claims.
PremisecitationOxyhaemoglobin has absorption peaks at 418 nm, 542 nm and 577 nm, while deoxyhaemoglobin has peak absorption at 430 nm and 555 nm.
“Oxyhaemoglobin has peaks at 418 nm, 542 nm and 577 nm, while deoxyhaemoglobin has peak absorption at 430 nm and 555 nm [41,42].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [41] Campolmi, 2016, Med. Princ. Pract. source
Not checked in the free run.
- cites [42]
Not checked in the free run.
May combine two claims.
PremisecitationWater has multiple absorption peaks with increasing wavelength, with minor peaks at 740 nm and 835 nm and major peaks at 970 nm, 1180 nm, 1430 nm, 1650 nm, 1930 nm and 1975 nm.
“Water has multiple peaks with increasing wavelength, with minor peaks at 740 nm and 835 nm and major peaks at 970 nm, 1180 nm, 1430 nm, 1650 nm, 1930nm and 1975nm [25,44].”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationOverall, compression was shown to increase absorption and scattering coefficients by up to ∼75%, however it didn't follow a monotonic trend.
“Overall, compression was shown to increase absorption and scattering coefficients by up to ∼75%, however it didn’t follow a monotonic trend.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
May combine two claims.
PremisecitationThe optical properties for the epidermal and dermal skin section were 2.5-18.5-fold greater than those of the subcutaneous layer; e.g., at 633 nm the mean absorption coefficient for the epidermal/dermal layer ranged from 0.033 mm−1 to 0.241 mm−1 compared to 0.013 mm−1 for the subcutaneous layer.
“The optical properties for the epidermal and dermal skin section were 2.5 - 18.5-fold greater than those of the subcutaneous layer e.g., at 633 nm the mean absorption coefficient for the epidermal/dermal layer ranged from 0.033 mm−1 to 0.241 mm−1 compared to 0.013 mm−1 for the subcutaneous layer.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Simpson, 1998, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
InterpretationcitationTroy et al.'s data corroborates Simpson et al., whose data showed that absorption may increase beyond 1000 nm.
“This corroborates Simpson et al., whose data showed that absorption may increase beyond 1000 nm [50].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Simpson, 1998, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
InterpretationcitationSingle and double integrating sphere methods are potentially limited by sample thickness, and this should be a consideration when choosing a method for measuring optical properties of samples.
“This potentially suggests that single and double integrating sphere methods are limited by sample thickness and this should therefore be a consideration when choosing a method for measuring the optical properties of samples.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationWith no applied pressure, the absorption coefficient ranged from 0.34 mm-1 to 0.59 mm-1, with a 2-fold difference in measurements within samples from one donor.
“With no applied pressure the absorption coefficient ranged from 0.34 mm-1 - 0.59 mm-1, with a 2-fold difference in measurements within samples from one donor.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
May combine two claims.
PremisecitationAt 1000 nm, absorption was recorded as an average of ∼0.1 mm-1, tending to increase slowly across the measured spectrum of 1000 nm-2200 nm.
“At 1000 nm absorption was recorded as an average of ∼ 0.1 mm-1, tending to increase slowly across the measured spectrum of 1000 nm - 2200 nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [54] Troy, 2001, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFrom 1700 nm, absorption starts to increase again with increasing wavelength.
“From 1700nm absorption starts to increase again with increasing wavelength.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationBeyond the NIR, water also has an absorption peak at 3400 nm.
“Beyond the NIR, water also has an absorption peak at 3400 nm [44].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [44] Hale, 1973, Appl. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationError in Simpson et al.'s data was large and attributed to differences in reflectance from imaging back to front and vice versa.
“Error in the data was large and attributed to differences in reflectance from imaging back to front and vice versa.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Simpson, 1998, Phys. Med. Biol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationA linear increase in anisotropy corresponded with tissue dehydration and thinning, which was proposed to reduce the distance between scatterers and increase their effective size.
“A linear increase in anisotropy was shown to correspond with tissue dehydration and thinning which was proposed to reduce the distance between scatterers and increase their effective size.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] Prahl, 1998
Not checked in the free run.
May combine two claims.
PremisecitationChan et al. showed that absorption decreased rapidly from 1.4 mm-1 at 400 nm to 0.2 mm-1 at 1250 nm and peaked at 1450 nm to 2.1 mm-1, corresponding with maximal water absorption, with a sharp decrease thereafter.
“Absorption was shown to decrease rapidly from 1.4 mm-1 at 400 nm to 0.2 mm-1 at 1250 nm and to peak at 1450 nm to 2.1 mm-1, corresponding with maximal water absorption, with a sharp decrease in absorption thereafter.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] Chan, 1997, IEEE J. Select. Topics Quantum Electron. source
Not checked in the free run.
May combine two claims.
Premisenone givenIn the NIR spectrum (750-1400 nm), oxyhaemoglobin absorption has maximal absorption at 900 nm and deoxyhaemoglobin absorption peaks at 960 nm.
“In the NIR spectrum (750 nm – 1400 nm) oxyhaemoglobin absorption has maximal absorption at 900 nm and deoxyhaemoglobin absorption peaks at 960 nm.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone givenMelanin absorption peaks at 335 nm but reduces steadily with increasing wavelength.
“Melanin absorption peaks at 335 nm but this reduces steadily with increasing wavelength.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationIn agreement with Doornbos et al., the forehead was the location of greatest scattering amongst the subjects and the arm the least.
“In agreement with Doornbos et al., the forehead was the location of greatest scattering amongst the subjects and the arm the least [57].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [57] Doornbos, 1999, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
PremisecitationPeak absorption and scattering coefficients for all skin layers was at 405 nm, steadily reducing with increasing wavelength, with the greatest absorption and scattering coefficients being in the epidermis.
“Peak absorption and scattering coefficients for all skin layers was at 405 nm steadily reducing with increasing wavelength with the greatest absorption and scattering coefficients being in the epidermis.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [53] Shimojo, 2020, J. Biomed. Opt. source
Not checked in the free run.
Classified as probably not a substantial claim.
InterpretationreasoningBecause of the change of penetration depth associated with wavelength, different wavelengths could be used to determine the optical properties of different skin layers.
“Because of the change of penetration depth associated with wavelength, it was suggested that different wavelengths could be used to determine the optical properties of different skin layers.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningThe greater scattering in the forehead may be due to 1 cm - 2 cm tissue depth being probed by photons, the skin being thinnest at the forehead and photons being scattered by the skull beneath.
“This may have been due to 1 cm - 2 cm tissue depth being probed by photons, the skin being thinnest at the forehead and photons being scattered by the skull beneath.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationAbsorption in the epidermis, dermis and subcutaneous layer ranged from ∼1.5 mm-1, ∼1 mm-1 and ∼1.8 mm-1 at ∼400 nm to 0.8 mm-1, ∼0.8 mm-1 and ∼0.4 mm-1 at 1600 nm respectively; on average a reduction of ∼2-fold for the absorption coefficient at 1600 nm compared to 400 nm.
“Absorption in the epidermis, dermis and subcutaneous layer ranged from ∼1.5 mm-1, ∼1 mm-1 and ∼1.8 mm-1 at ∼400 nm to 0.8 mm-1, ∼0.8 mm-1 and ∼0.4 mm-1 at 1600 nm respectively; on average a reduction of ∼2-fold for the absorption coefficient at 1600 nm compared to 400 nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationAt the wavelength spectrum of 405 nm-1064 nm, absorption ranged from 3.32 mm-1 to 0.13 mm-1 and reduced scattering ranged from 9.95 mm-1 to 2.85 mm-1 in the epidermis (standard deviations for both coefficients tended to reduce with increasing wavelength).
“At the wavelength spectrum of 405 nm - 1064 nm, absorption ranged from 3.32 mm-1 - 0.13 mm-1 and reduced scattering ranged from 9.95 mm-1 - 2.85 mm-1 in the epidermis (standard deviations for both coefficients tended to reduce with increasing wavelength).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [53] Shimojo, 2020, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationSalomatina et al. suggested that scattering properties were dependent upon location and that collagen-elastin networks in fat from the scalp were thicker and denser than those in the subcutaneous adipose tissue of skin taken from the back.
“Salomatina et al., suggested that scattering properties were dependent upon location and that collagen-elastin networks in fat from the scalp were thicker and denser than those in the subcutaneous adipose tissue of skin taken from the back [52].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [52] Salomatina, 2006, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationIn Doornbos et al., at 630 nm absorption was greatest in the forehead compared to the sole of the foot and the arm; at 660 nm it was greatest in the sole of the foot; at 700 nm the arm had greatest absorption.
“Wavelength measurements were limited to 630 nm, 660 nm and 700 nm and at 630 nm absorption was greatest in the forehead compared to the sole of the foot and the arm. At 660 nm it was greatest in the sole of the foot, and at 700 nm the arm had greatest absorption.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [57] Doornbos, 1999, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
PremisecitationTorricelli et al. found that scattering decreased progressively with increasing wavelength, while absorption showed features of water, haemoglobin, and lipid, with average spectral peaks at ∼600 nm, 760 nm, and 950 nm of approximately 0.017 mm-1, 0.013 mm-1, and 0.027 mm-1 respectively.
“Again, scattering decreased progressively with increasing wavelength, while the absorption showed the features of absorption by water, haemoglobin, and lipid, with average spectral peaks at ∼600 nm, 760 nm and 950 nm of approximately 0.017 mm-1, 0.013 mm-1 and 0.027 mm-1 respectively.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [37] Torricelli, 2001, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
PremisecitationKono et al. found that average absorption coefficients amongst all data showed spectral peaks at 600 nm, 1000 nm, 1200 nm and 1450 nm corresponding with absorbers in the skin, with main differences between locations being observed in the visible spectrum.
“Across the subjects the average absorption coefficients amongst all data showed spectral peaks at 600 nm, 1000 nm, 1200 nm and 1450 nm corresponding with absorbers in the skin, with main differences between locations being observed in the visible spectrum.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [63] Kono, 2019, Int. J. Thermophys. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationKono et al. found scattering ranged from 12 mm-1 down to 3.7 mm-1 between 450 nm and 1400 nm.
“Scattering ranged from 12 mm-1 down to 3.7 mm-1 between 450 nm and 1400 nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [63] Kono, 2019, Int. J. Thermophys. source
Not checked in the free run.
May combine two claims.
PremisecitationThe amount of absorption associated with each skin constituent depended upon skin location and the most absorbing locations varied amongst subjects.
“The amount of absorption associated with each of these skin constituents depended upon skin location and the most absorbing locations varied amongst subjects.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [37] Torricelli, 2001, Phys. Med. Biol. source
Not checked in the free run.
May combine two claims.
PremisecitationHung et al. data showed a downward trend in optical scattering across the spectrum from 2.45 mm-1 to 1.15 mm-1; standard deviations for measurements above 1000 nm are large suggesting no significant differences between scattering coefficients beyond 1000 nm.
“Their data showed a downward trend in optical scattering across the spectrum from 2.45 mm-1 to 1.15 mm-1; however, the standard deviations for measurements taken above 1000 nm are large suggesting no significant differences between the scattering coefficients beyond 1000 nm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [60] Hung, 2015, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
PremisecitationJonasson et al. found a significant difference between men and women; measured reduced scattering coefficients were greater in men than women across the spectrum measured.
“there was a significant difference between men and women; measured reduced scattering coefficients were greater in men than women across the spectrum measured.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [61] Jonasson, 2018, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningThe decrease in reduced scattering with age may be due to decreasing collagen concentrations in skin with age.
“potentially due to decreasing collagen concentrations in skin with age.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationBetween 1200 nm and ∼1275 nm the absorbance declined rapidly and then increased sharply beyond this wavelength.
“Between 1200 nm and ∼1275 nm the absorbance declined rapidly and then increased sharply beyond this wavelength.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [60] Hung, 2015, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
PremisecitationTseng and Hsu's scattering and absorption profiles followed a similar pattern to other authors, with absorption greatest between 500 nm and 600 nm, declining with increasing wavelength except for a small peak at ∼970 nm.
“However, their scattering and absorption profiles followed a similar pattern to the other authors’ with absorption being greatest between 500 nm and 600 nm beyond which it declines steadily with increasing wavelength except for a small peak observed at ∼970 nm.”
Quote in the paperNot a substantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationThis resulted in increased scattering coefficients compared to the back where connective tissue was thin.
“This resulted in increased scattering coefficients compared to the back where connective tissue was thin.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationA new SFD model reduced the median relative error by 10% for μs' and 64% for μa compared to previously used models.
“This model reduced the median relative error by 10% for µs’ and 64% for µa compared to previously used models.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [69] Post, 2023, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe 1700 nm wavelength range might have the potential to provide deeper imaging and better resolution than at shorter wavelengths.
“At this point TMFP is at its greatest and scattering and absorption troughs are coincident suggesting that this wavelength range might have the potential to provide deeper imaging and better resolution than at shorter wavelengths.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationLike Saager et al., Phan et al. showed that skin pigmentation affected measured coefficients and that inter-subject variation for both absorption and scattering decreased amongst all locations measured with increasing wavelength.
“Like Saager et al., these authors showed that skin pigmentation affected the measured coefficients and that inter-subject variation for both absorption and scattering was found to decrease amongst all locations measured with increasing wavelength [64,65].”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationPhan et al. found not enough measurements across the spectrum to observe absorption peaks associated with skin components, although the general trend was for a reduction in absorption with increasing wavelength.
“There were not enough measurements made across the spectrum to observe absorption peaks associated with the components of the skin, although the general trend was for a reduction in absorption with increasing wavelength.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [65] Phan, 2021, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationThe 640 nm peak in Hsu et al. data potentially represents scattering due to collagen bundles in the skin.
“potentially representing scattering due to collagen bundles in the skin [56].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [56] Hsu, 2015, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
PremisecitationAguénounon et al. improved SFD imaging by using Single Snapshot imaging of Optical Properties (SSOP), reducing image acquisitions to one and applying machine learning to improve accuracy, reduce edge artifacts, improve resolution and increase imaging speed.
“Improvements to SFD imaging have been made by Aguénounon et al., to enable its use in surgical settings, by using Single Snapshot imaging of Optical Properties (SSOP). This reduces the number of image acquisitions to one and applies machine learning approaches to improve accuracy, reduce edge artifacts, improve resolution and increase imaging speed.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationAt 900 nm both absorbance and reduced scattering in the skin of all subjects was similar, however at shorter wavelengths variability between measurements was greater.
“At 900 nm both the absorbance and reduced scattering in the skin of all subjects was similar, however at shorter wavelengths variability between measurements was greater.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [64] Saager, 2015, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationSaager et al. data followed the same general trends as other authors, but optical properties were only measured up to 900 nm so water peaks beyond this were not observed.
“Although their data followed the same general trends as that of other authors, the optical properties were only measured up to 900 nm so peaks associated with the water beyond this were not observed.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [64] Saager, 2015, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe epidermis and dermis thicknesses used were taken from Meglinski et al. as an example of average skin layer depths.
“The thickness of the epidermis and dermis used were taken from Meglinski et al., and used as an example of average skin layer depths [58].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [58] Meglinski, 2002, Physiol. Meas. source
Not checked in the free run.
May combine two claims.
PremisecitationSFD imaging is limited due to slow and sequential acquisitions of light patterns and edge artifacts and is therefore difficult to apply to clinical applications that require real-time feedback such as surgery.
“SFD imaging, discussed here, is limited due to slow and sequential acquisitions of light patterns and edge artifacts and is therefore difficult to apply to clinical applications that require real-time feedback such as surgery [68].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [68] Aguénounon, 2020, Biomed. Opt. Express source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationMelanin has been shown to depolarise light.
“For example, melanin has been shown to depolarise light [93].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [93] Baumann, 2012, Biomed. Opt. Express source
Not checked in the free run.
May combine two claims.
PremisecitationThe degree of depolarisation has allowed differentiation between cancerous and benign lesions, showing importance for detecting skin cancers.
“This has been shown to be of importance for detecting skin cancers where the degree of depolarisation has allowed differentiation between cancerous and benign lesions [90,91].”
Quote in the paperSubstantial claim0 of 2 sources checked
The paper relies on this without support.
Skin is the first barrier encountered by light when imaging non-invasively.
“Skin is the first barrier encountered by light when imaging non-invasively”
“If we want to image inside the human body using light, skin is the first barrier to light transmission.”
The paper relies on this without support.
Photons encounter approximately 10 scattering events per millimetre when travelling through the skin.
“Photons encounter approximately 10 scattering events per millimetre when travelling through the skin”
The paper relies on this without support.
Skin thickness varies with location on the body; the epidermis is thickest on the soles of the feet (~660 µm) and thinnest in the eyelids (~130 µm), and the dermis is thickest on the back (up to 4 mm) and thinnest on the eyelids (~215 µm).
“For example, the epidermis is thickest on the soles of the feet (∼660 µm), and thinnest in the eyelids (∼130 µm). The dermis, however, is thickest on the back (up to 4 mm) and thinnest on the eyelids (∼ 215 µm).”
The paper relies on this without support.
The optical properties of the skin must be accurately determined to advance imaging depth and methodologies.
“To advance imaging depth and methodologies the optical properties of the skin must be accurately determined.”
The paper relies on this without support.
Scattering in skin is the main loss mechanism of photons, dominating over absorption.
“Scattering in skin is the main loss mechanism of photons dominating over absorption.”
The paper relies on this without support.
At wavelengths in the visible spectrum, absorption by the epidermis is dominated by melanin, which decreases as wavelength increases.
“At wavelengths in the visible spectrum, absorption by the epidermis is dominated by melanin which decreases as wavelength increases.”
The paper relies on this without support.
Absorption in the dermis and subcutaneous tissue is dominated by the presence of blood and haemoglobin.
“Absorption in the dermis and subcutaneous tissue is dominated by the presence of blood and haemoglobin.”
The paper relies on this without support.
As the wavelength of light increases into the NIR, absorption by all skin layers becomes dominated by water and lipid content.
“As the wavelength of light increases into the NIR absorption by all skin layers becomes dominated by water and lipid content.”
The paper relies on this without support.
An understanding of the optical absorption and scattering properties of tissues provides useful information regarding likely penetration depths.
“An understanding of the optical absorption and scattering properties of tissues provides useful information regarding likely penetration depths.”
The paper relies on this without support.
These models rely on accurate optical properties from the literature.
“However, these models rely on accurate optical properties from the literature.”
The paper relies on this without support.
TMFP is related to the reduced scattering coefficient.
“Because the TMFP is related to the reduced scattering coefficient, peaks in the scattering graph (Fig. 6) correspond with troughs in the TMFP graphs and reasons that scattering may be lower in the in-vivo situation were discussed in section 6.”
The paper relies on this without support.
Variability in in-vivo optical properties is intrinsic.
“Variability is intrinsic, and a major contributor to the intrinsic variability in the in-vivo measurements is skin pigmentation.”
The paper relies on this without support.
A major contributor to the intrinsic variability in in-vivo measurements is skin pigmentation.
“Variability is intrinsic, and a major contributor to the intrinsic variability in the in-vivo measurements is skin pigmentation.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In vivo published data show high inter- and intrapublication variability, with differences among published absorption data being up to 9.6-fold and 2-fold for reported scattering data.
“We have previously shown that in vivo published data show high inter- and intrapublication variability with differences among the published absorption data being up to 9.6- and 2-fold for reported scattering data.36”
From Effect of skin color on optical properties and the implications for medical optical technologies: a review (2024) · cites this paper as [36]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Two transport mean free paths reported at 1400 nm
MinorThe in-vivo transport mean free path at 1400 nm is reported as both 1.35 mm and 1.38 mm. Both values are given for the same wavelength and in-vivo data.
“at a wavelength of 1400 nm the TMFP for in-vivo data is maximal (1.35 mm).”
“if one uses the in-vivo data recorded at 1400 nm, light would be expected to travel 1.38 mm in to the skin before it became diffuse”
Epidermal absorption is not always greater than subcutaneous absorption
MinorThe text says epidermal absorption is greater than subcutaneous absorption, but its reported values at approximately 400 nm are 1.5 mm-1 and 1.8 mm-1, respectively.
“these authors described absorption in the epidermis being greater than in the dermis and subcutaneous tissue”
“Absorption in the epidermis, dermis and subcutaneous layer ranged from ∼1.5 mm-1, ∼1 mm-1 and ∼1.8 mm-1 at ∼400 nm”
Cites
Optical Constants of Water in the 200-nm to 200-μm Wavelength Region
Hale, G. M., & Querry, M. R. (1973) · Applied Optics · doi:10.1364/ao.12.000555 · Cited by 4,479 · see citing works
Optical properties of biological tissues: A review
Jacques, S. L. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/11/r37 · Cited by 4,008 · see citing works
Second window for in vivo imaging
Smith, A. M., Mancini, M. C., & Nie, S. (2009) · Nature Nanotechnology · doi:10.1038/nnano.2009.326 · Cited by 2,848 · see citing works
A review of the optical properties of biological tissues
Cheong, W. F., Prahl, S. A., & Welch, A. J. (1990) · IEEE Journal of Quantum Electronics · doi:10.1109/3.64354 · Cited by 2,530 · see citing works
Anderson, R. R., & Parrish, J. A. (1981) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12479191 · Cited by 2,456 · see citing works
Bashkatov, A. N., Genina, E. A., Kochubey, V. I., & Tuchin, V. V. (2005) · Journal of Physics D: Applied Physics · doi:10.1088/0022-3727/38/15/004 · Cited by 1,720 · see citing works
Farrell, T. J., Patterson, M. S., & Wilson, B. C. (1992) · Medical Physics · doi:10.1118/1.596777 · Cited by 1,380 · see citing works
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · see citing works
Through-skull fluorescence imaging of the brain in a new near-infrared window
Hong, G., Diao, S., Chang, J., Antaris, A. L., Chen, C., Zhang, B., Zhao, S., Atochin, D. N., Huang, P. L., Andreasson, K. I., et al. (2014) · Nature Photonics · doi:10.1038/nphoton.2014.166 · Cited by 1,052 · see citing works
Ash, C., Dubec, M., Donne, K., & Bashford, T. (2017) · Lasers in Medical Science · doi:10.1007/s10103-017-2317-4 · Cited by 1,040 · see citing works
Recent Advances of Optical Imaging in the Second Near‐Infrared Window
Kenry, K., Duan, Y., & Liu, B. (2018) · Advanced Materials · doi:10.1002/adma.201802394 · Cited by 813 · see citing works
OPTICAL PROPERTIES OF SKIN, SUBCUTANEOUS, AND MUSCLE TISSUES: A REVIEW
Bashkatov, A. N., Genina, E. A., & Tuchin, V. V. (2011) · Journal of Innovative Optical Health Sciences · doi:10.1142/s1793545811001319 · Cited by 765 · see citing works
Sandby-Møller, J., Poulsen, T., & Wulf, H. C. (2003) · Acta Dermato-Venereologica · doi:10.1080/00015550310015419 · Cited by 737 · see citing works
Hemmer, E., Benayas, A., Légaré, F., & Vetrone, F. (2016) · Nanoscale Horizons · doi:10.1039/c5nh00073d · Cited by 732 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · see citing works
Quantitation and mapping of tissue optical properties using modulated imaging
Cuccia, D. J., Bevilacqua, F. P., Durkin, A. J., Ayers, F. R., & Tromberg, B. J. (2009) · Journal of Biomedical Optics · doi:10.1117/1.3088140 · Cited by 659 · see citing works
Simpson, C. R., Kohl, M., Essenpreis, M., & Cope, M. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/9/003 · Cited by 651 · see citing works
Shcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625 · see citing works
Salomatina, E. V., Jiang, B., Novak, J., & Yaroslavsky, A. N. (2006) · Journal of Biomedical Optics · doi:10.1117/1.2398928 · Cited by 599 · see citing works
Imaging skin pathology with polarized light
Jacques, S. L., Ramella-Roman, J. C., & Lee, K. (2002) · Journal of Biomedical Optics · doi:10.1117/1.1484498 · Cited by 582 · see citing works
Optical properties of human skin in the near infrared wavelength range of 1000 to 2200 nm
Troy, T. L., & Thennadil, S. N. (2001) · Journal of Biomedical Optics · doi:10.1117/1.1344191 · Cited by 420 · see citing works
Deep optical imaging within complex scattering media
Yoon, S., Kim, M., Jang, M., Choi, Y., Choi, W., Kang, S., & Choi, W. (2020) · Nature Reviews Physics · doi:10.1038/s42254-019-0143-2 · Cited by 420 · see citing works
Wilson, R. H., Nadeau, K. P., Jaworski, F. B., Tromberg, B. J., & Durkin, A. J. (2015) · Journal of Biomedical Optics · doi:10.1117/1.jbo.20.3.030901 · Cited by 415 · see citing works
Meglinski, I. V., & Matcher, S. J. (2002) · Physiological Measurement · doi:10.1088/0967-3334/23/4/312 · Cited by 408 · see citing works
Deep optical imaging of tissue using the second and third near-infrared spectral windows
Sordillo, L. A., Pu, Y., Pratavieira, S., Budansky, Y., & Alfano, R. R. (2014) · Journal of Biomedical Optics · doi:10.1117/1.jbo.19.5.056004 · Cited by 407 · see citing works
Polarized light interaction with tissues
Tuchin, V. V. (2016) · Journal of Biomedical Optics · doi:10.1117/1.jbo.21.7.071114 · Cited by 384 · see citing works
Doornbos, R. M. P., Lang, R., Aalders, M. C., Cross, F. W., & Sterenborg, H. J. C. M. (1999) · Physics in Medicine and Biology · doi:10.1088/0031-9155/44/4/012 · Cited by 380 · see citing works
NIR-I-to-NIR-II fluorescent nanomaterials for biomedical imaging and cancer therapy
Zhao, J., Zhong, D., & Zhou, S. (2017) · Journal of Materials Chemistry B · doi:10.1039/c7tb02573d · Cited by 354 · see citing works
Transmission in near‐infrared optical windows for deep brain imaging
Shi, L., Sordillo, L. A., Rodríguez‐Contreras, A., & Alfano, R. (2016) · Journal of Biophotonics · doi:10.1002/jbio.201500192 · Cited by 346 · see citing works
Depth Penetration of Light into Skin as a Function of Wavelength from 200 to 1000 nm
Finlayson, L., Barnard, I. R. M., McMillan, L., Ibbotson, S. H., Brown, C. T. A., Eadie, E., & Wood, K. (2022) · Photochemistry and Photobiology · doi:10.1111/php.13550 · Cited by 302 · see citing works
van Veen, R. L. P., Sterenborg, H. J. C. M., Pifferi, A., Torricelli, A. M., Chikoidze, E., & Cubeddu, R. (2005) · Journal of Biomedical Optics · doi:10.1117/1.2085149 · Cited by 277 · see citing works
Oltulu, P., İnce, B., Kökbudak, N., Fındık, S., & Kılınç, F. (2018) · Turkish Journal of Plastic Surgery · doi:10.4103/tjps.tjps_2_17 · Cited by 276 · see citing works
Effects of compression on soft tissue optical properties
Chan, E. K., Sorg, B., Protsenko, D., O'Neil, M., Motamedi, M., & Welch, A. J. (1996) · IEEE Journal of Selected Topics in Quantum Electronics · doi:10.1109/2944.577320 · Cited by 256 · see citing works
Origins of Tissue Optical Properties in the UVA, Visible, and NIR Regions
Jacques, S. L. (1996) · Advances in Optical Imaging and Photon Migration · doi:10.1364/aoipm.1996.opc364 · Cited by 247 · see citing works
In vivo absorption and scattering spectroscopy of biological tissues
Taroni, P., Pifferi, A., Torricelli, A. M., Comelli, D., & Cubeddu, R. (2003) · Photochemical & Photobiological Sciences · doi:10.1039/b209651j · Cited by 220 · see citing works
Chromophore concentrations, absorption and scattering properties of human skin in-vivo
Tseng, S.-H., Bargo, P., Durkin, A., & Kollias, N. (2009) · Optics Express · doi:10.1364/oe.17.014599 · Cited by 217 · see citing works
Golovynskyi, S., Golovynska, I., Stepanova, L. I., Datsenko, O. I., Liu, L., Qu, J., & Ohulchanskyy, T. Y. (2018) · Journal of Biophotonics · doi:10.1002/jbio.201800141 · Cited by 209 · see citing works
Gender differences in skin: A review of the literature
Dao, H., & Kazin, R. A. (2007) · Gender Medicine · doi:10.1016/s1550-8579(07)80061-1 · Cited by 199 · see citing works
Zonios, G. I., & Dimou, A. D. (2006) · Optics Express · doi:10.1364/oe.14.008661 · Cited by 197 · see citing works
Persistent phosphors for the future: Fit for the right application
Poelman, D., der Heggen, D. V., Du, J., Cosaert, E., & Smet, P. F. (2020) · Journal of Applied Physics · doi:10.1063/5.0032972 · Cited by 187 · see citing works
Torricelli, A. M., Pifferi, A., Taroni, P., Giambattistelli, E., & Cubeddu, R. (2001) · Physics in Medicine and Biology · doi:10.1088/0031-9155/46/8/313 · Cited by 172 · see citing works
Advances in optical coherence tomography in dermatology—a review
Olsen, J., Holmes, J., & Jemec, G. B. E. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.4.040901 · Cited by 165 · see citing works
Baumann, B., Baumann, S. O., Konegger, T., Pircher, M., Götzinger, E., Schlanitz, F., Schütze, C., Sattmann, H., Litschauer, M., Schmidt-Erfurth, U., et al. (2012) · Biomedical Optics Express · doi:10.1364/boe.3.001670 · Cited by 163 · see citing works
Nachabe, R., Hendriks, B. H., Lucassen, G. W., van der Voort, M., Evers, D. J., Rutgers, E. J., Peeters, M.-J. V., Van der Hage, J. A., Oldenburg, H. S., Ruers, T. J., et al. (2011) · Journal of Biomedical Optics · doi:10.1117/1.3611010 · Cited by 161 · see citing works
Spatial frequency domain imaging in 2019: principles, applications, and perspectives
Gioux, S., Mazhar, A., & Cuccia, D. J. (2019) · Journal of Biomedical Optics · doi:10.1117/1.jbo.24.7.071613 · Cited by 161 · see citing works
Measurement of tissue optical properties in the context of tissue optical clearing
Bashkatov, A. N., Berezin, K. V., Dvoretski, K. N., Chernavina, M. L., Genina, E. A., Genin, V. D., Kochubey, V. I., Lazareva, E. N., Pravdin, A. B., Евгеньевна, Ш. М., et al. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.9.091416 · Cited by 146 · see citing works
Epidermal Thickness and Biomechanical Properties of Plantar Tissues in Diabetic Foot
Chao, C. Y. L., Zheng, Y.-P., & Cheing, G. L. Y. (2011) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2011.04.004 · Cited by 143 · see citing works
Optical properties of the subcutaneous adipose tissue in the spectral range 400–2500 nm
Bashkatov, A. N., Genina, E. A., Kochubey, V. I., & Tuchin, V. V. (2005) · Optics and Spectroscopy · doi:10.1134/1.2135863 · Cited by 142 · see citing works
Morgan, S. P., Khong, M. P., & Somekh, M. G. (1997) · Applied Optics · doi:10.1364/ao.36.001560 · Cited by 136 · see citing works
Self-citationShimojo, Y., Nishimura, T., Hazama, H., Ozawa, T., & Awazu, K. (2020) · Journal of Biomedical Optics · doi:10.1117/1.jbo.25.4.045002 · Cited by 132 · see citing works
Optical properties of porcine skin dermis between 900 nm and 1500 nm
Du, Y., Hu, X., Cariveau, M. J., Ma, X., Kalmus, G. W., & Lu, J. Q. (2000) · Physics in Medicine and Biology · doi:10.1088/0031-9155/46/1/312 · Cited by 114 · see citing works
Effect of temperature on the optical properties ofex vivohuman dermis and subdermis
Laufer, J. G., Simpson, R., Kohl‐Bareis, M., Essenpreis, M., & Cope, M. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/9/004 · Cited by 99 · see citing works
Koerber, D., Khan, S., Shamsheri, T., Kirubarajan, A., & Mehta, S. (2023) · Journal of Racial and Ethnic Health Disparities · doi:10.1007/s40615-022-01446-9 · Cited by 91 · see citing works
Saager, R. B., Balu, M., Crosignani, V., Sharif, A., Durkin, A. J., Kelly, K. M., & Tromberg, B. J. (2015) · Journal of Biomedical Optics · doi:10.1117/1.jbo.20.6.066005 · Cited by 76 · see citing works
Two-Photon Absorption: An Open Door to the NIR-II Biological Window?
Shaw, P. A., Forsyth, E., Haseeb, F., Yang, S., Bradley, M., & Klausen, M. (2022) · Frontiers in Chemistry · doi:10.3389/fchem.2022.921354 · Cited by 75 · see citing works
In vivo near-infrared spectral detection of pressure-induced changes in breast tissue
Jiang, S., Pogue, B. W., Paulsen, K. D., Kogel, C. A., & Poplack, S. P. (2003) · Optics Letters · doi:10.1364/ol.28.001212 · Cited by 74 · see citing works
Measurement of tissue optical properties in a wide spectral range: a review [Invited]
Martins, I. S., Silva, H., Lazareva, E. N., Chernomyrdin, N. V., Zaytsev, K. I., Oliveira, L. M. C., & Tuchin, V. V. (2022) · Biomedical Optics Express · doi:10.1364/boe.479320 · Cited by 73 · see citing works
Boyle, C. J., Plotczyk, M., Villalta, S. F., Patel, S. R., Hettiaratchy, S. P., Masouros, S. D., Masen, M. A., & Higgins, C. A. (2019) · Science Advances · doi:10.1126/sciadv.aay0244 · Cited by 70 · see citing works
Crooks, C. J., West, J., Morling, J. R., Simmonds, M., Juurlink, I., Briggs, S., Cruickshank, S., Hammond-Pears, S., Shaw, D., Card, T. R., et al. (2022) · European Respiratory Journal · doi:10.1183/13993003.03246-2021 · Cited by 68 · see citing works
Response To: Investigating sources of inaccuracy in wearable optical heart rate sensors
Colvonen, P. J. (2021) · npj Digital Medicine · doi:10.1038/s41746-021-00408-5 · Cited by 65 · see citing works
In Vivo Measurement of Optical Properties of Human Skin for 450–800 nm and 950–1600 nm Wavelengths
Kono, T., & Yamada, J. (2019) · International Journal of Thermophysics · doi:10.1007/s10765-019-2515-3 · Cited by 62 · see citing works
Gender variations in the optical properties of skin in murine animal models
Calabro, K., Curtis, A., Galarneau, J.-R., Krucker, T., & Bigio, I. J. (2011) · Journal of Biomedical Optics · doi:10.1117/1.3525565 · Cited by 58 · see citing works
Optical properties of animal tissues in the wavelength range from 350 to 2600 nm
Filatova, S. A., Shcherbakov, I. A., & Цветков, В. Б. (2017) · Journal of Biomedical Optics · doi:10.1117/1.jbo.22.3.035009 · Cited by 58 · see citing works
Seeing through the Skin: Photoacoustic Tomography of Skin Vasculature and Beyond
Li, D., Humayun, L. L., Vienneau, E. P., Vu, T., & Yao, J. (2021) · JID Innovations · doi:10.1016/j.xjidi.2021.100039 · Cited by 57 · see citing works
Honda, N., Ishii, K., Kajimoto, Y., Kuroiwa, T., & Awazu, K. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.7.075006 · Cited by 54 · see citing works
Aguénounon, E., Smith, J. T., Al‐Taher, M., Diana, M., Intes, X., & Gioux, S. (2020) · Biomedical Optics Express · doi:10.1364/boe.397681 · Cited by 54 · see citing works
Jonasson, H., Fredriksson, I., Bergstrand, S., Östgren, C. J., Larsson, M., & Strömberg, T. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.12.121608 · Cited by 52 · see citing works
Polarization speckle imaging as a potential technique for in vivo skin cancer detection
Tchvialeva, L., Dhadwal, G., Lui, H., Kalia, S., Zeng, H., McLean, D. I., & Lee, T. K. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.6.061211 · Cited by 49 · see citing works
Optical properties of neonatal skin measured in vivo as a function of age and skin pigmentation
Bosschaart, N., Mentink, R., van Leeuwen, A. G. J. M. T., Aalders, M. C., & Kok, J. H. (2011) · Journal of Biomedical Optics · doi:10.1117/1.3622629 · Cited by 49 · see citing works
Hsu, C.-K., Tzeng, S.-Y., Yang, C.-C., Lee, J. Y.-Y., Huang, L. L.-H., Chen, W.-R., Hughes, M., Chen, Y.-W., Liao, Y.-K., & Tseng, S.-H. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.000390 · Cited by 45 · see citing works
Song, W., Zhang, L., Ness, S., & Yi, J. (2017) · Biomedical Optics Express · doi:10.1364/boe.8.003966 · Cited by 45 · see citing works
Degree of optical polarization as a tool for detecting melanoma: proof of principle
Louie, D. C., Phillips, J., Tchvialeva, L., Kalia, S., Lui, H., Wang, W., & Lee, T. K. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.12.125004 · Cited by 44 · see citing works
Mignon, C., Tobin, D. J., Zeitouny, M., & Uzunbajakava, N. E. (2018) · Biomedical Optics Express · doi:10.1364/boe.9.000852 · Cited by 43 · see citing works
Optical properties of the human skin / Optičke osobine ljudske kože
Gajinov, Z., Matić, M., Prćić, S., & Đuran, V. (2010) · Serbian Journal of Dermatology and Venerology · doi:10.2478/v10249-011-0029-5 · Cited by 42 · see citing works
Temperature induced changes in the optical properties of skin in vivo
Iorizzo, T. W., Jermain, P. R., Salomatina, E. V., Muzikansky, A., & Yaroslavsky, A. N. (2021) · Scientific Reports · doi:10.1038/s41598-020-80254-9 · Cited by 41 · see citing works
Lentsch, G., Baugh, E. G., Lee, B., Aszterbaum, M., Zachary, C. B., Kelly, K. M., & Balu, M. (2022) · Journal of Investigative Dermatology · doi:10.1016/j.jid.2022.01.016 · Cited by 41 · see citing works
Phan, T., Rowland, R., Ponticorvo, A., Le, B. C., Wilson, R. H., Sharif, S. A., Kennedy, G. T., Bernal, N. P., & Durkin, A. J. (2021) · Journal of Biomedical Optics · doi:10.1117/1.jbo.26.2.026001 · Cited by 38 · see citing works
Skin optical properties in the obese and their relation to body mass index: a review
Rodriguez, A. J., Boonya-Ananta, M. T., Gonzalez, M., Le, V. N. D., Fine, J. N., Palacios, C. S., McShane, M. J., Coté, G. L., & Ramella‐Roman, J. C. (2022) · Journal of Biomedical Optics · doi:10.1117/1.jbo.27.3.030902 · Cited by 32 · see citing works
Recent Advances in Time-Resolved NIR Spectroscopy for Nondestructive Assessment of Fruit Quality
Torricelli, A. M., Contini, D., Mora, A. D. D., Martinenghi, E., Tamborini, D., Villa, F., Tosi, A., & Spinelli, L. (2015) · Chemical engineering transactions · doi:10.3303/cet1544008 · Cited by 26 · see citing works
Sordillo, L. A., Lindwasser, L. W., Budansky, Y., Leproux, P., & Alfano, R. R. (2015) · Journal of Biomedical Optics · doi:10.1117/1.jbo.20.3.030501 · Cited by 24 · see citing works
Kourbaj, G., Bielfeldt, S., Seise, M., & Wilhelm, K. (2020) · Skin Research and Technology · doi:10.1111/srt.12948 · Cited by 17 · see citing works
Zhang, X. U., Faber, D. J., van Leeuwen, T. G. J. M., & Sterenborg, H. J. C. M. (2020) · Journal of Physics Photonics · doi:10.1088/2515-7647/ab9071 · Cited by 16 · see citing works
Botelho, A. R., Silva, H., Martins, I. S., Carneiro, I. C., Carvalho, S. D., Henrique, R., Tuchin, V. V., & Oliveira, L. M. C. (2022) · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · doi:10.1016/j.saa.2022.122002 · Cited by 14 · see citing works
Skin, Light and their Interactions, an In-Depth Review for Modern Light-Based Skin Therapies
Albany Cosmetic And Laser Center, & Alhallak, K. (2021) · Clinical Dermatology & Therapy · doi:10.24966/cdt-8771/100081 · Cited by 12 · see citing works
Efficacy of Rhodamine Light in the Treatment of Superficial Vascular Lesions of the Face
Campolmi, P., Cannarozzo, G., Dragoni, F., Conti, R., & Moretti, S. (2016) · Medical Principles and Practice · doi:10.1159/000446987 · Cited by 11 · see citing works
Hung, C.-H., Chou, T.-C., Hsu, C.-K., & Tseng, S.-H. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.000443 · Cited by 8 · see citing works
Koerber, D. M., Khan, S., Shamsheri, T., et al. (2022) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(22)02981-3 · Cited by 8 · see citing works
Mok, A. T. Y., Shea, J., Wu, C., Xia, F., Tatarsky, R., Yapici, N., & Xu, C. (2021) · Biomedical Optics Express · doi:10.1364/boe.441844 · Cited by 7 · see citing works
Monem, S., Singh, A., Karsten, A. E., Amin, R., & Harith, M. A. (2015) · Applied Physics B · doi:10.1007/s00340-015-6226-6 · Cited by 4 · see citing works
Model for the diffuse reflectance in spatial frequency domain imaging
Post, A. L., Faber, D. J., & van Leeuwen, T. G. J. M. (2023) · Journal of Biomedical Optics · doi:10.1117/1.jbo.28.4.046002 · Cited by 4 · see citing works
Study on the influence of contact pressure on diffuse spectroscopy measurement of in vivo tissue
Li, C., Xia, H., Zhou, Y., Li, S., Liu, R., Chen, W., & Jiang, J. (2021) · Infrared Physics & Technology · doi:10.1016/j.infrared.2021.103669 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Setchfield, K., Gorman, A., Simpson, A. H. R. W., Somekh, M. G., & Wright, A. J. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.1.010901 · Cited by 72 · see citing works
Self-citationSeeing through the skin: Optical methods for visualizing transdermal drug delivery with microneedles
Limcharoen, B., Wanichwecharungruang, S. P., Banlunara, W., & Darvin, M. E. (2024) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2024.115478 · Cited by 21 · see citing works
Evaluating facial dermis aging in healthy Caucasian females with LC-OCT and deep learning
Assi, A., Fischman, S., Lopez, C., Pedrazzani, M., Grignon, G., Missodey, R., Korichi, R., Cauchard, J.-H., Ralambondrainy, S., & Bonnier, F. (2024) · Scientific Reports · doi:10.1038/s41598-024-74370-z · Cited by 13 · see citing works
Review of shortwave infrared imaging and spectroscopy in tissue [Invited]
Roblyer, D. M., Pilvar, A., Pham, T., Spink, S. S., Hayakawa, C. K., Venugopalan, V., & Pierce, M. C. (2025) · Biomedical Optics Express · doi:10.1364/boe.577163 · Cited by 12 · see citing works
Vision of Life’s Code: Molecular Probe Nanoarchitectonics for Deep RNA/DNA Illumination
Sutrisno, L., Sun, K., & Ariga, K. (2026) · ACS Nano · doi:10.1021/acsnano.5c22282 · Cited by 8 · see citing works
Kim, J., & Lee, O. (2024) · Skin Research and Technology · doi:10.1111/srt.13654 · Cited by 7 · see citing works
The Biophysics of Flash Radiotherapy: Tools for Measuring Tumor and Normal Tissues Microenvironment
Ali, I. G., & Naqa, I. E. (2025) · Antioxidants · doi:10.3390/antiox14080899 · Cited by 6 · see citing works
Yoshida, K.-I., & Yanagisawa, H. (2024) · Biomedical Optics Express · doi:10.1364/boe.527858 · Cited by 5 · see citing works
Papadoliopoulou, M., Koutsoumpos, S., Margaris, I., Matiatou, M., Giannios, P., Arkadopoulos, N. F., Moutzouris, K., & Michalopoulos, N. V. (2025) · Journal of Personalized Medicine · doi:10.3390/jpm15020071 · Cited by 4 · see citing works
Habibul, A., Mao, J., Xin, S., Zhang, J., Song, W., Song, X., & Tang, K. (2026) · Materials & Design · doi:10.1016/j.matdes.2026.115618 · Cited by 4 · see citing works
Côrrea, L., Meirelles, L., Carneiro, R. C. G., da Silva, F. F. A., Wisotzky, E. L., & Berrocal, E. (2025) · Lasers in Medical Science · doi:10.1007/s10103-025-04507-7 · Cited by 3 · see citing works
Tu, D., Lemberger, N. S., Wallmeier, K., Riseman, J., Kuzma, B. A., Wei, Y., Khoo, T. C., Rantou, E., Ghosh, P., Luke, M. C., et al. (2025) · Pharmaceutics · doi:10.3390/pharmaceutics17091193 · Cited by 3 · see citing works
Jones, Z. D., Reitzle, D., & Kienle, A. (2025) · Biomedical Optics Express · doi:10.1364/boe.573069 · Cited by 3 · see citing works
Enhancing tumor’s skin photothermal therapy using Gold nanoparticles : a Monte Carlo simulation
Zerakni, F., Dib, A. S. A., & Attili, A. (2024) · Lasers in Medical Science · doi:10.1007/s10103-024-04072-5 · Cited by 3 · see citing works
Hoffman, M. A. T., Keppler, M. A., Smith, A. L., Fasci, A. T., Macasadia, M. E., Tijerina, A. J., Hood, R. L., DeLisi, M. P., & Bixler, J. N. (2025) · Journal of Biomedical Optics · doi:10.1117/1.jbo.30.1.015001 · Cited by 3 · see citing works
da Silva, G. P., Doorduin, J., Szymański, W., da Silva, R. S., Elsinga, P. H., & Bonan, C. D. (2025) · Drug Discovery Today · doi:10.1016/j.drudis.2025.104477 · Cited by 3 · see citing works
Bachir, W. (2024) · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · doi:10.1016/j.saa.2024.125456 · Cited by 2 · see citing works
Age Prediction of Hematoma from Hyperspectral Images Using Convolutional Neural Networks
Keshavarz, A., Bieber, G., Wulff, D., Babian, C., & Lüdtke, S. (2026) · Journal of Imaging · doi:10.3390/jimaging12020078 · Cited by 2 · see citing works
Thermal comfort comparison of localized infrared radiant heaters from a spectral perspective
Qin, M., Zhan, H., Han, X., Qi, X., Su, H., Yuan, Y., & Tao, S. (2026) · Applied Thermal Engineering · doi:10.1016/j.applthermaleng.2026.131318 · Cited by 2 · see citing works
Рогаткин, Д. А., Tarasov, A. P., & Shtyflyuk, M. E. (2024) · Laser Medicine · doi:10.37895/2071-8004-2023-27-4-8-15 · Cited by 2 · see citing works
The effect of contact pressure on photoplethysmography
Gong, X., Yu, C., & Xuan, Y. (2024) · Indonesian Journal of Electrical Engineering and Computer Science · doi:10.11591/ijeecs.v33.i3.pp1589-1603 · Cited by 1 · see citing works
The coupling effect between skin strain and blood condition on its reflectance spectrum in-vivo
Huo, Z., Wang, S., Tan, K., Li, L., Li, C., & Wang, Z. (2024) · Optics & Laser Technology · doi:10.1016/j.optlastec.2024.110990 · Cited by 1 · see citing works
Wei, L., Yi, X., Cheng, W. P., Ma, Y., & Lin, Y. (2025) · Journal of Biophotonics · doi:10.1002/jbio.202500303 · Cited by 1 · see citing works
Zhao, Y., Wang, X., Wei, D., He, Y., Wang, Y., Xu, X., Zhang, M., Tian, W., Xu, Z., & Wang, J. (2026) · Results in Engineering · doi:10.1016/j.rineng.2026.109186 · Cited by 1 · see citing works
Evaluating the effect of skin color on photoplethysmography signals in a vascular finger phantom
Osorio-Sanchez, L., May, J. M., & Kyriacou, P. (2026) · Design and Quality for Biomedical Technologies · doi:10.1117/12.3079657 · Cited by 1 · see citing works
박수빈Park, S., Hyeon, J., & So, H. (2026) · Lasers in Medical Science · doi:10.1007/s10103-026-04892-7 · Cited by 1 · see citing works
Machine Learning Estimate the Optical Properties of Tissue
Zhang, Y., Deng, B., Wright, A. J., Somekh, M. G., Pounds, M. P., & Parkes, A. J. (2024) · Imaging and Applied Optics · doi:10.1364/aopt.2024.oth4f.2
Self-citationYu, H., Ko, D., & Jung, B. (2024) · Conference on Lasers and Electro-Optics Pacific Rim (CLEO-PR) · doi:10.1109/cleo-pr60912.2024.10676481
The interstitial channel system as a photobiomodulation regulatory system of the body
Letyagin, A. Y., Nebrat, V., & Bgatova, N. P. (2024) · International Multi-Conference on Engineering, Computer and Information Sciences (SIBIRCON) · doi:10.1109/sibircon63777.2024.10758515
Non-Invasive Laser Surgery With Deep Operating Depth Using Multibeam Interference
Liu, J. (2025) · Medical Devices Evidence and Research · doi:10.2147/mder.s498706
Deng, B., Zhang, Y., Ghahremani, M., Bentley, A., Parkes, A. J., Wright, A. J., Pound, M. P., & Somekh, M. G. (2025) · Optics Letters · doi:10.1364/ol.564068
Self-citationIn silico Evaluation Based on Tissue Optics for Picosecond Laser Treatment of Pigmented Lesions
Shimojo, Y., Tsuruta, D., Ozawa, T., & Kono, T. (2025) · Nippon Laser Igakkaishi · doi:10.2530/jslsm.jslsm-46_0011
Miyata, K., Yamagami, J., Takenaka, Y., Onuki, M., Sawayanagi, T., & Ishiguro, N. (2025) · Scientific Reports · doi:10.1038/s41598-025-28953-z
Preliminary Evaluation of a Remote Photoplethysmography Method for Heart Rate Estimation
González, A. J., Petracchi, F. M., & B., G. F. B. (2025) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-032-06401-1_131
Investigating the Dose Parameters of Low-Level Laser Therapy to Optimize Therapeutic Efficacy
Shurrab, K. (2025) · Polish Journal of Medical Physics And Engineering · doi:10.2478/pjmpe-2025-0035
Lee, W., Chang, M. H., Kim, S., Lee, B., Jung, C. H., & Kim, K. (2025) · IEEE Access · doi:10.1109/access.2025.3646742
Al-Ibadi, Z. (2026) · International Journal of Physics and Applications · doi:10.33545/26647575.2026.v8.i1b.247
Diagnosis and Treatment of Metastatic Photocarcinoma
Swarnkar, J., Saloki, A., & Saraf, S. (2026) · Springer eBooks · doi:10.1007/978-981-95-7358-5_14
Vis–NIR spectroscopy and deep learning for non-invasive in vivo detection of acute liver injury
Wang, X., Wang, Y., Ma, Q., Jin, A., Li, J., Lin, Y., Zhang, H., & He, Z. (2026) · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · doi:10.1016/j.saa.2026.127888
Farmer, J., Oian, C. A., & Khan, T. R. (2026) · Applied Sciences · doi:10.3390/app16094172
Wearable Dual-Mode Biosensing System for Dynamic Light Dosimetry in Tissues
Wei, J., Ma, L., Li, W., Xu, P., Wang, Y., Zhou, F., & Cai, F. (2026) · Biosensors · doi:10.3390/bios16050263
Wireless Charging Method for Implantable Electronic Devices Using Near Infrared Light
Fuada, S., Särestöniemi, M., Ma, G., & Katz, M. (2026) · IEEE Journal on Wireless Power Technologies · doi:10.1109/jwpt.2026.3700072
Fuada, S., Särestöniemi, M., & Katz, M. D. (2026) · Communications in computer and information science · doi:10.1007/978-3-032-28829-5_21
Zhang, Z., Huang, Z., Zhang, C., Qin, W., Zhong, C., Yin, S., Chen, Q., & Deng, X. (2026) · Colloids and Surfaces B Biointerfaces · doi:10.1016/j.colsurfb.2026.115996
A Modular Bio-Hybrid Skin Model for Optical Testing ApplicationsPreprint
Bajrami, D., Wei, K., Spano, F., Agah, N., Bonmarin, M., & Rossi, R. M. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.07.27.740463
Ni, D., Klämpfl, F., Lahoz, F. J., Schmidt, M., & Hohmann, M. (2026) · Advanced Materials Technologies · doi:10.1002/admt.71202
Lei, S., Meng, Y., Yu, C., SUN, N., & Liu, Y. (2026) · Journal of Cosmetic Dermatology · doi:10.1111/jocd.71163
Su, J., Lin, J.-M., Tsao, A., Lin, C.-K., & Lin, C. H. (2026) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/aea7ae
MacKenzie, L. E. (2026) · ChemRxiv · doi:10.26434/chemrxiv.15008974/v1
Bridging Visible and SWIR Spectroscopy via Dual-Spectrometer for Skin Tissue Characterization
Li, G., Li, W., Li, Y., & Cai, F. (2026) · Photonics · doi:10.3390/photonics13090880
Closing the Loop: High‐Precision 3D Photofabrication in Living Tissues
Zeynali, A. M. H., Malinauskas, M., & Chirico, G. G. (2026) · Advanced Science · doi:10.1002/advs.77853
Timimi, Z. A., & Tamimi, Z. J. M. (2026) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/aeada9
Osorio-Sanchez, L., May, J. M., & Kyriacou, P. A. (2026) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-032-39674-7_17
Solouma, N. H., & Hamdy, O. (2026) · Bioengineering · doi:10.3390/bioengineering13111246
MacKenzie, L. E. (2026) · ChemRxiv · doi:10.26434/chemrxiv.15008974/v3
- Authors
- Setchfield, K., Gorman, A., Simpson, A. H. R. W., Somekh, M. G., & Wright, A. J.
- Affiliations
- University of Nottingham (GB)
- University of Edinburgh (GB)
- University of Edinburgh
- Venue
- Biomedical Optics Express
- Publisher
- Optica Publishing Group
- Published
- 2023-06-21
- Volume
- 14
- Issue
- 7
- Pages
- 3555
- ISSN
- 2156-7085 (ISSN-L)
- PMC
- PMC10368038
- OpenAlex
- W4380090304
- Semantic Scholar
- 37f3b99ec68c7e4126ac96074eef0fa753853a4c
- Cited by
- 49
- References
- 90
- Citations by year
- 2026: 20 · 2025: 16 · 2024: 13
- FWCI
- 2.07
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Biomedical Optics Express (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Repository@Nottingham (University of Nottingham) (submittedVersion, cc-by)
- Repository@Nottingham (University of Nottingham) (submittedVersion)
- Funders
- Engineering and Physical Sciences Research Council (EP/T020997/1)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Physical Sciences › Engineering › Biomedical Engineering › Optical Coherence Tomography Applications
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Keywords
- skin optical properties, in vivo measurement, ex vivo measurement, optical absorption, light scattering, optical penetration depth, transport mean free path, measurement variability, reporting guidelines, non-invasive optical imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review