Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Open accessAbstract
Wearable sensor technologies are essential to the realization of personalized medicine through continuously monitoring an individual's state of health. Sampling human sweat, which is rich in physiological information, could enable non-invasive monitoring. Previously reported sweat-based and other non-invasive biosensors either can only monitor a single analyte at a time or lack on-site signal processing circuitry and sensor calibration mechanisms for accurate analysis of the physiological state. Given the complexity of sweat secretion, simultaneous and multiplexed screening of target biomarkers is critical and requires full system integration to ensure the accuracy of measurements. Here we present a mechanically flexible and fully integrated (that is, no external analysis is needed) sensor array for multiplexed in situ perspiration analysis, which simultaneously and selectively measures sweat metabolites (such as glucose and lactate) and electrolytes (such as sodium and potassium ions), as well as the skin temperature (to calibrate the response of the sensors). Our work bridges the technological gap between signal transduction, conditioning (amplification and filtering), processing and wireless transmission in wearable biosensors by merging plastic-based sensors that interface with the skin with silicon integrated circuits consolidated on a flexible circuit board for complex signal processing. This application could not have been realized using either of these technologies alone owing to their respective inherent limitations. The wearable system is used to measure the detailed sweat profile of human subjects engaged in prolonged indoor and outdoor physical activities, and to make a real-time assessment of the physiological state of the subjects. This platform enables a wide range of personalized diagnostic and physiological monitoring applications.
Overview
AI snapshot
Written from the paper’s full text.
TL;DR
A flexible wearable sensing array with full on-board integration was built to measure sweat glucose, lactate, sodium, potassium, and skin temperature at the same time, with signal conditioning, processing, and Bluetooth transmission. Human exercise trials showed it could track sweat profiles in real time and detect dehydration-related electrolyte changes.
Key findings
- The platform merged plastic-based sensors with silicon integrated circuits on a flexible printed circuit board, enabling multiplexed sweat analysis without external analysis; five sensors measured metabolites, electrolytes, and skin temperature.
- In vitro sensitivities were 2.35 nA μM−1 for glucose over 0–200 μM, 220 nA mM−1 for lactate over 0–30 mM, 64.2 mV per decade for Na+ over 10–160 mM, 61.3 mV per decade for K+ over 1–32 mM, and about 0.18% per degree Celsius for temperature over 20–40 °C.
- Sensors showed negligible interference from non-target species, and temperature compensation corrected enzymatic sensor overestimation; long-term tests indicated stable sensitivities for at least four weeks, with Na+/K+ at about 62.5 and 59.5 mV per decade in ambient conditions and glucose/lactate within 5% of original values when stored at 4 °C.
- On-body trials with 26 healthy subjects (4 females, 22 males, aged 20–40) tracked sweat during cycling and outdoor running; dehydration trials without water intake showed sweat [Na+] increases after about 80 min and roughly 2.5% body-weight loss, while euhydration with 150 ml water every 5 min kept [Na+] and [K+] stable after initial changes.
Methods
- Design: a five-sensor array on PET with 3-mm electrodes, including enzymatic amperometric glucose/lactate sensors, ion-selective Na+/K+ sensors, and a resistive temperature sensor, integrated with an FPCB containing analogue conditioning, an ATmega328P microcontroller, and Bluetooth.
- Data: in vitro characterization in analyte solutions and artificial sweat; on-body testing in constant-load cycling (14 volunteers: 4 females, 10 males), graded cycling (7 males), and outdoor running (12 males), plus ex situ sweat comparisons.
- Analysis: chronoamperometry for glucose/lactate, open-circuit potentiometry for Na+/K+, resistance for temperature, with on-board calibration/compensation and wireless data relay to a mobile app.
Limitations
- Stated: glucose and lactate sensor responses degraded after about two hours of use because of reduced enzyme activity, though the sensor array can be replaced via the sensor–FPCB interface.
- Stated: wearing the plastic-substrate device raised local temperature, which may slightly bias actual skin-temperature readings, but on-board calibration was reported to prevent impact on electrolyte and metabolite measurements; in vitro temperature-compensation tests also had noise from uneven solution heating.
- Evident: the human sample was small and heavily male-skewed (26 total; 4 females, 22 males; graded and outdoor running trials included only males), and the paper does not report long-term continuous use beyond the exercise sessions for the enzymatic sensors.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
36 Own result · 18 No basis given · 13 No full text of the sources · 10 Own method · 10 Own interpretation · 8 Partly supported · 2 Supported · 1 Supported so far · 1 Couldn’t confirm · 1 Not checked yet · accuracy not yet measured
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“skin temperature is clinically informative of a variety of diseases and skin injuries such as pressure ulcers23,24.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [23] Webb, R. C. et al. Ultrathin conformal devices for precise and continuous thermal characterization of human skin. Nature Mater. 12, 938–944 (2013) sourceExperimental — accuracy not yet measured
The Abstract says that skin thermometry can provide clinically relevant information about cardiovascular health, malignancy and other aspects of physiology. It does not link skin temperature to pressure ulcers or establish its clinical value for assessing skin injuries; its mention of wound healing concerns local heating as a possible intervention. (LLM, conf=0.86)
- Not checked — no full text of the sourcecites [24] Sprigle, S., Linden, M., McKenna, D., Davis, K. & Riordan, B. Clinical skin temperature measurement to predict incipient pressure ulcers. Adv. Skin Wound Care 14, 133–137 (2001) source
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The decreases in concentration of lactate and glucose in sweat are expected, owing to the dilution effect caused by an increase in sweat rate, which is visually observed as exercise continues13.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) sourceExperimental — accuracy not yet measured
The Introduction (page not available in our copy) explains that increasing sweat flow can dilute biomarkers whose transport into sweat is slower than the flow. It does not establish that lactate and glucose concentrations both decrease as exercise continues; instead, it says high-exertion tests can show a spike in sweat lactate associated with sweat rate. (LLM, conf=0.83)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The relatively stable behaviour of [K+] is explained by its passive ion partitioning mechanism13.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) sourceExperimental — accuracy not yet measured
The section headed “Passive ion partitioning: Potassium (mM)” reports that sweat K+ concentration is independent of sweat rate, supporting the description of relatively stable behavior. But it says the precise partitioning mechanism has not been fully confirmed, so it does not establish passive partitioning as the explanation. No page number is available in our copy. (LLM, conf=0.86)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Assuming a single-centred flow model13, the best-case sampling interval can be calculated to be less than 1 min, based on the sweat rate (~3–4 mg min−1 cm−2)39 and the pad size (1.5 cm × 2 cm × 50 μm).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
What the sources say
- Supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) source
The citation is attached to the single-centred flow model, not the sweat rate or pad dimensions. Section C, “Expectations for biomarker sampling intervals,” defines that model as the best-case limit and gives the refill-time calculation used by the citing authors. (LLM, conf=0.94)
- Partly supportedcites [39] Vimieiro-Gomes, A. C. Comparison of sweat rate during graded exercise and the local rate induced by pilocarpine. Braz. J. Med. Biol. Res. 38, 1133–1139 (2005) sourceExperimental — accuracy not yet measured
The Abstract (page unavailable) reports local exercise-induced sweat rates of 2.1 ± 1.4 and 2.7 ± 1.2 mg cm−2 min−1 in nine young men. That supports using a sweat rate on the order of 3 for an estimate, but not the cited value of approximately 3–4 as a reported rate. The citation is attached to the sweat-rate input, not to the flow model, pad dimensions, or calculated interval. (LLM, conf=0.84)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- The authors state this as an assumption: Sweat flow can be modeled as a single-centred flow.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“However, lactate concentration becomes relatively stable after 1,100 s, indicating the stabilization of physiological responses to continuous, sub-maximal constant-exercise power output22.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [22] Derbyshire, P. J., Barr, H., Davis, F. & Higson, S. P. Lactate in human sweat: a critical review of research to the present day. J. Physiol. Sci. 62, 429–440 (2012) sourceExperimental — accuracy not yet measured
The review’s “Sweat lactate variation with exercise intensity” section (page unavailable) reports that sweat lactate decayed and then reached a plateau during incremental cycling. It does not give 1,100 s as the time of stabilization or interpret the plateau as stabilization of physiological responses during continuous, sub-maximal constant-power exercise. (LLM, conf=0.84)
hedgeddirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“Owing to Faraday and capacitive currents38, the responses of both sensors showed drift initially but stabilized within 1 min of the data recording.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
What the sources say
- Could not verifycites [38] Grieshaber, D., MacKenzie, R., Vörös, J. & Reimhult, E. Electrochemical biosensors—sensor principles and architectures. Sensors 8, 1400–1458 (2008) sourceExperimental — accuracy not yet measured
The Electrochemical Detection Techniques section (page unavailable) describes time-dependent current in chronoamperometry, but attributes current changes to the diffusion layer. Searches for Faradaic and capacitive currents, sensor drift, and one-minute stabilization found no passage connecting those currents to initial drift in glucose and lactate sensor responses. (LLM, conf=0.73)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The independent and selective operation of individual sensors is preserved during multiplexed measurements by employing highly specific surface chemistries and by electrically decoupling the operating points of each sensor's interface.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
In the paper
“By using poly(3,4-ethylenedioxythiophene) polystyrene sulfonate (PEDOT:PSS) as an ion-to-electron transducer in the ISEs and carbon nanotubes in the PVB reference membrane25, robust potentiometric sensors (with voltage output) can be obtained for long-term continuous measurements with negligible voltage drift.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [25] Zhu, J., Qin, Y. & Zhang, Y. Preparation of all solid-state potentiometric ion sensors with polymer-CNT composites. Electrochem. Commun. 11, 1684–1687 (2009) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“These enzymatic sensors autonomously generate current signals proportional to the abundance of the corresponding metabolites between the working electrode and the Ag/AgCl electrode.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The measurement of Na+ and K+ levels is facilitated through the use of ion-selective electrodes (ISEs), coupled with a polyvinyl butyral (PVB)-coated reference electrode to maintain a stable potential in solutions with different ionic strengths (Extended Data Fig. 2b–d).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Probably not a substantial claim. 22
- It may state two separate claims as one. 30
- The quote only partly states this claim. 1
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Wearable sensor technologies are essential to the realization of personalized medicine through continuously monitoring an individual's state of health1–12.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [1] Kim, D. H. et al. Epidermal electronics. Science 333, 838–843 (2011) sourceExperimental — accuracy not yet measured
The paper’s Abstract describes skin-integrated sensors for physiological-status monitoring, and its ECG, EMG and EEG demonstrations show that they can measure an individual’s signals. It does not establish that wearable sensors are essential to personalized medicine or demonstrate continuous health monitoring; it notes that long-term use requires further work (Abstract; page unavailable). (LLM, conf=0.84)
- Not checked — no full text of the sourcecites [2] Lipomi, D. J. et al. Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes. Nature Nanotechnol. 6, 788–792 (2011) source
- Not checked — no full text of the sourcecites [3] Takei, K. et al. Nanowire active-matrix circuitry for low-voltage macroscale artificial skin. Nature Mater. 9, 821–826 (2010) source
- Not checked — no full text of the sourcecites [4] Wu, W. et al. Piezoelectricity of single-atomic-layer MoS2 for energy conversion and piezotronics. Nature 514, 470–474 (2014) source
- Partly supportedcites [5] McAlpine, M. C., Ahmad, H., Wang, D. & Heath, J. R. Highly ordered nanowire arrays on plastic substrates for ultrasensitive flexible chemical sensors. Nature Mater. 6, 379–384 (2007) sourceExperimental — accuracy not yet measured
The Abstract (p. 1) says flexible chemical and biological sensors could enable wearable monitoring systems. It does not establish that wearable sensors are essential to personalized medicine or that they continuously monitor an individual's health; those claims are stronger and more specific than the paper's proposed application. (LLM, conf=0.86)
- Not checked — no full text of the sourcecites [6] Kaltenbrunner, M. et al. An ultra-lightweight design for imperceptible plastic electronics. Nature 499, 458–463 (2013) source
- Partly supportedcites [7] Xu, S. et al. Soft microfluidic assemblies of sensors, circuits, and radios for the skin. Science 344, 70–74 (2014) sourceExperimental — accuracy not yet measured
The Abstract says skin-mounted sensors can provide clinical-quality health monitoring for continuous use and that continuous physiological monitoring is needed to treat many medical conditions. It does not say wearable sensors are essential to the realization of personalized medicine; that part overstates what the paper supports. (LLM, conf=0.85)
- Not checked — no full text of the sourcecites [8] Wang, C. et al. User-interactive electronic skin for instantaneous pressure visualization. Nature Mater. 12, 899–904 (2013) source
- Not checked — no full text of the sourcecites [9] Liang, J., Li, L., Niu, X., Yu, Z. & Pei, Q. Elastomeric polymer light-emitting devices and displays. Nature Photon. 7, 817–824 (2013) source
- Not checked — no full text of the sourcecites [10] Kim, D. H., Ghaffari, R. & Rogers, J. A. Flexible and stretchable electronics for biointegrated devices. Annu. Rev. Biomed. Eng. 14, 113–128 (2012) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence.
- Partly supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“Previously reported sweat-based and other non-invasive biosensors either can only monitor a single analyte at a time or lack on-site signal processing circuitry and sensor calibration mechanisms for accurate analysis of the physiological state14–18.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Supportedcites [14] Schazmann, B. et al. A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration. Anal. Methods 2, 342–348 (2010) source
The Abstract describes this device as a sodium-selective sweat sensor, and the Conclusions discuss a sensor array for other ions as a future possibility. The paper therefore supports its use as an example of a previously reported single-analyte sweat sensor; the citation need not establish that this device also lacked on-site processing or calibration. (LLM, conf=0.90)
- Not checked — no full text of the sourcecites [15] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013) source
- Could not verifycites [16] Coyle, S. et al. BIOTEX—biosensing textiles for personalised healthcare management. IEEE Trans. Inf. Technol. Biomed. 14, 364–370 (2010) sourceExperimental — accuracy not yet measured
The Introduction describes limitations of earlier sweat-collection methods, but does not say that previously reported biosensors were restricted to one analyte or lacked on-site processing and calibration. It instead describes BIOTEX’s multiparametric patch; the sensor and control-electronics sections discuss calibration before trials and signal processing, without establishing the claimed either-or limitation of prior devices. (LLM, conf=0.76)
- Not checked — no full text of the sourcecites [17] Bandodkar, A. J. et al. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 54, 603–609 (2014) source
- Not checked — no full text of the sourcecites [18] Rose, D. P. et al. Adhesive RFID sensor patch for monitoring of sweat electrolytes. IEEE Trans. Biomed. Eng. 62, 1457–1465 (2015) source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“sweat lactate can potentially serve as a sensitive marker of pressure ischaemia22;”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Supportedcites [22] Derbyshire, P. J., Barr, H., Davis, F. & Higson, S. P. Lactate in human sweat: a critical review of research to the present day. J. Physiol. Sci. 62, 429–440 (2012) source
The “Potential applications for sweat lactate measurements” section (page not provided) calls sweat lactate “potentially a very useful early warning indicator of pressure ischemia.” This supports the claim’s tentative wording. (LLM, conf=0.93)
hedgedcorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Sampling human sweat, which is rich in physiological information13, could enable non-invasive monitoring.”
- Also stated in: § Abstract · plain
“Measurements of human sweat could enable such insight, because it contains physiologically and metabolically rich information that can be retrieved non-invasively13.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) source
The Introduction describes sweat as a source of biomarkers and blood solutes accessible non-invasively. Section V explains that many blood solutes enter sweat quickly enough for continuous physiological monitoring, supporting the claim that sweat sampling could provide useful physiological information. (LLM, conf=0.92)
plainconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Also stated in: § Abstract · plain
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The two-electrode system is a common strategy for low-current electrochemical sensing34,35.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Design of electrochemical sensors
What the sources say
- Not checked — no full text of the sourcecites [34] Kudo, H. et al. A flexible and wearable glucose sensor based on functional polymers with Soft-MEMS techniques. Biosens. Bioelectron. 22, 558–562 (2006) source
- Partly supportedcites [35] Bandodkar, A. J. et al. Tattoo-based noninvasive glucose monitoring: a proof-of-concept study. Anal. Chem. 87, 394–398 (2015) sourceExperimental — accuracy not yet measured
The Results and Discussion section describes this tattoo glucose sensor using an Ag/AgCl electrode as a combined counter/reference electrode alongside a working electrode (page unavailable). It does not say that this two-electrode configuration is a common or standard strategy for low-current electrochemical sensing; its discussion of lower current density concerns glucose extraction by reverse iontophoresis, not the sensing configuration. (LLM, conf=0.85)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“These trends are probably caused by increased blood serum [Na+] and [K+] with dehydration and increased neural stimulation, a conclusion in agreement with previous ex situ sweat analyses33.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [33] Morgan, R. M., Patterson, M. J. & Nimmo, M. A. Acute effects of dehydration on sweat composition in men during prolonged exercise in the heat. Acta Physiol. Scand. 182, 37–43 (2004) sourceExperimental — accuracy not yet measured
The Results report higher serum Na+ and K+ during dehydration, and the Discussion proposes greater extracellular Na+ and sympathetic nervous activity as possible explanations for higher sweat Na+. The Discussion’s concluding mechanism emphasizes Na+ and Cl−, aldosterone, and sympathetic activity; it does not identify increased serum K+ as a specific cause. The Experimental procedures describe sweat collected from a forearm patch and analyzed after collection, consistent with ex situ analysis. Page numbers are unavailable in our copy. (LLM, conf=0.86)
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Partly supported
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Here we present a mechanically flexible and fully integrated (that is, no external analysis is needed) sensor array for multiplexed in situ perspiration analysis, which simultaneously and selectively measures sweat metabolites (such as glucose and lactate) and electrolytes (such as sodium and potassium ions), as well as the skin temperature (to calibrate the response of the sensors).”
- Also stated in: § Abstract · plain
“Here we present a wearable flexible integrated sensing array (FISA) for simultaneous and selective screening of a panel of biomarkers in sweat (Fig. 1a).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Also stated in: § Abstract · plain
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The wearable system is used to measure the detailed sweat profile of human subjects engaged in prolonged indoor and outdoor physical activities, and to make a real-time assessment of the physiological state of the subjects.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The use of Prussian blue dye as a mediator minimizes the reduction potentials to approximately 0 V (versus Ag/AgCl) (Extended Data Fig. 2a), and thus eliminates the need for an external power source to activate the sensors.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The pad could absorb about 10 μl of sweat, which was sufficient to provide stable sensor readings.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
hedgeddirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A claim is checked only against the cited paper’s full text, never its abstract, and none is available for its sources.
In the paper
“Sweat [Na+] increases and [K+] decreases in the beginning of perspiration, in line with the previous ex situ studies from the collected sweat samples28,29.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [28] Cage, G. W., Wolfe, S. M., Thompson, R. H. & Gordon, R. S. Jr. Effects of water intake on composition of thermal sweat in normal human volunteers. J. Appl. Physiol. 29, 687–690 (1970) source
- Not checked — no full text of the sourcecites [29] Patterson, M. J., Galloway, S. D. R. & Nimmo, M. A. Variations in regional sweat composition in normal human males. Exp. Physiol. 85, 869–875 (2000) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Importantly, the entire system is mechanically flexible, thus delivering a practical wearable sensor technology that can be used for prolonged indoor and outdoor physical activities.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“This conditioning process was important to minimize the potential drift further.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Preparation of Na+ and K+ selective sensors
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The quote only partly states this claim.
In the paper
“No apparent difference is observed for glucose concentration at different power output settings, a finding consistent with the response of blood glucose to graded, short-term exercise30.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [30] Messonnier, L. A. et al. Lactate kinetics at the lactate threshold in trained and untrained men. J. Appl. Physiol. 114, 1593–1602 (2013) source
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Extended Data Fig. 6a–d show that Na+ and K+ sensors had a relative standard deviation of ~1% in sensitivity while glucose and lactate sensors had a relative standard deviation of ~5% in sensitivity.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Although temperature has a minimal effect on the potentiometric sensors, it greatly influences the performance of the enzymatic sensors.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“After the power is raised, the sweat rate visibly increases, followed by a sharp increase in skin temperature and sweat [Na+] as well as a slight increase in [K+] (in three of the seven subjects, [K+] remained stable).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirectn = 3- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Parylene is used as an insulating layer to ensure reliable sensor reading by preventing electrical contact of the metal lines with skin and sweat.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Both [Na+] and [K+] stabilize as the cycling continues.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Figure 2h illustrates that with the increase of temperature, the uncompensated sensor readouts can lead to substantial overestimation of the actual concentration of the given glucose and lactate solutions; however, the temperature compensation allows for accurate and consistent readings.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Sweat analyte levels on the wrist follow similar trends but with concentrations different from those obtained at the forehead (Extended Data Fig. 9).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“During the exercise, the heart rate, oxygen consumption (VO2), and pulmonary minute ventilation were measured using external monitoring instruments, and were found to increase proportionally with increasing power output as shown in Fig. 3c.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“On the other hand, a substantial increase in sweat [Na+] and a smaller increase in sweat [K+] (no clear increase in [K+] was observed in two out of six subjects) were observed in dehydration trials (without water intake) after 80 min when subjects had lost a large amount of water (~2.5% of body weight) (Fig. 4d and e).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirectn = 2- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“With continued perspiration, the skin temperature rises at about 400 s because of muscle heat conductance to skin and then remains stable, while the concentration of both lactate and glucose in sweat decrease gradually.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“A linear relationship between current and analyte concentrations with sensitivities of 2.35 nA μM−1 for glucose sensors and 220 nA mM−1 for lactate sensors was observed.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The sensor–FPCB interface allows for convenient replacement of the fresh sensor arrays for subsequent use.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“With the cessation of exercise, these physiological responses decrease and then remain stable.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The sensors could be used without pre-conditioning (with a small drift of ~2–3 mV h−1).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Preparation of Na+ and K+ selective sensors
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The current sensing signal paths were capable of measuring current levels as low as 1 nA, which was much lower than the minimum signal in our measurements (tens of nanoamperes).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Signal conditioning, processing and wireless transmission circuit design
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“An increase in temperature was observed when the ‘smart headband’ or ‘smart wristband’ was worn owing to the use of the plastic substrate on skin.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This platform enables a wide range of personalized diagnostic and physiological monitoring applications.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence.
In the paper
“For example, excessive loss of sodium and potassium in sweat could result in hyponatremia, hypokalemia, muscle cramps or dehydration20;”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [20] Speedy, D. B., Noakes, T. D. & Schneider, C. Exercise-associated hyponatremia: a review. Emerg. Med. 13, 17–27 (2001) source
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may state two separate claims as one.
In the paper
“For these applications, the sample collection and analysis are performed separately, failing to provide a real-time profile of sweat content secretion, while requiring extensive laboratory analysis using bulky instrumentation19.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [19] Boysen, T. C., Yanagawa, S., Sato, F. & Sato, K. A modified anaerobic method of sweat collection. J. Appl. Physiol. 56, 1302–1307 (1984) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Given the complexity of sweat secretion, simultaneous and multiplexed screening of target biomarkers is critical and requires full system integration to ensure the accuracy of measurements.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may state two separate claims as one.
- Rests on an assumption: Sweat secretion involves multivariate mechanisms and is complex.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Additionally, skin temperature measurements are needed to compensate for and eliminate the influence of temperature variation in the readings of the chemical sensors through a built-in signal processor.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: fails — The paper may give no reason why it holds.
Probably not a substantial claim. The paper may give no reason why it holds.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This approach decouples the stringent mechanical requirements at the sensor level and electrical requirements at the signal conditioning, processing and transmission levels, and at the same time exploits the strengths of the underlying technologies.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The convection and non-uniform distributions of solution temperature could result in noticeable noise in the signal measurements.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This platform could be exploited or reconfigured for in situ analyses of other biomarkers within sweat and other human fluid samples to facilitate personalized and real-time physiological and clinical investigations.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“During on-body tests, the newly generated sweat would refill the pad and ‘rinse away’ the old sweat.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
hedgeddirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“On the other hand, the silicon integrated-circuit technology does not provide sufficiently large active areas nor the intimate skin contact required to achieve stable and sensitive on-body measurements.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim.
In the paper
“In this case, because the subject had a lower sweat rate at the wrist29, the sensors were activated at a later time.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [29] Patterson, M. J., Galloway, S. D. R. & Nimmo, M. A. Variations in regional sweat composition in normal human males. Exp. Physiol. 85, 869–875 (2000) source
plaindirectSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“Monitoring hydration status is of the utmost importance to athletes because fluid deficit impairs endurance performance and increases carbohydrate reliance32.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [32] Barr, S. I. Effects of dehydration on exercise performance. Can. J. Appl. Physiol. 24, 164–172 (1999) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“We believe that this wearable platform may enable new fundamental physiology studies through further on-body evaluation.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“To ensure the fidelity of sensor readings, the data collection of each channel took place when a sufficient sweat sample was present, as shown by stabilization of sensor readings within the physiologically relevant range (see Methods).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“The power output was calibrated and monitored through the ergometer.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“All subjects gave written, informed consent before participation in the study.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“The on-body evaluation of the FISA was performed in compliance with the protocol that was approved by the institutional review board at the University of California, Berkeley (CPHS 2014-08-6636).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“Although this may result in a small error in measuring the actual skin temperature, it should be noted that this does not have an impact on the measurement of the electrolytes and metabolites, owing to the on-board temperature calibration.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“The sensor arrays were calibrated, and the subjects’ foreheads and wrists were cleaned with alcohol swabs and gauze before sensors were worn on-body.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper relies on this without support.
Commercially available wearable sensors only track physical activities and vital signs and fail to provide molecular-level health insight.
“At present, commercially available wearable sensors are only capable of tracking an individual's physical activities and vital signs (such as heart rate), and fail to provide insight into the user's health state at molecular levels.”
The paper relies on this without support.
Sweat secretion involves multivariate mechanisms and is complex.
“Given the complexity of sweat secretion, simultaneous and multiplexed screening of target biomarkers is critical and requires full system integration to ensure the accuracy of measurements.”
“Given the multivariate mechanisms that are involved in sweat secretion, an attractive strategy would be to devise a fully integrated multiplexed sensing system to extract the complex information available from sweat.”
The paper relies on this without support.
Amperometric sensors with a larger area produce a larger current signal.
“It is known that amperometric sensors with larger area provide larger current signal.”
The paper relies on this without support.
A thinner Prussian blue layer provides better sensitivity, which is essential for measuring low glucose levels in sweat.
“A thinner Prussian blue layer can provide better sensitivity, which is essential for low-glucose-level measurements in sweat.”
The paper relies on this without support.
A thicker Prussian blue layer yields a broader linear response range, which is important for measuring lactate in sweat.
“A thicker Prussian blue layer can provide a wider linear response range, which is crucial for lactate measurement in sweat.”
The authors state this as an assumption.
Sweat flow can be modeled as a single-centred flow.
“Assuming a single-centred flow model13, the best-case sampling interval can be calculated to be less than 1 min, based on the sweat rate (~3–4 mg min−1 cm−2)39 and the pad size (1.5 cm × 2 cm × 50 μm).”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The sensor remains stable throughout the measurement period with a sensitivity of 56.2 mV/decade, similar to sensitivities achieved by lithography-based Na+ sensors and close to the Nernstian ideal (23).”
From Regional and correlative sweat analysis using high-throughput microfluidic sensing patches toward decoding sweat (2019) · cites this paper as [23]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Recently, a fully integrated wearable sensor array for multiplexed in-situ perspiration analysis was reported, where the sodium, potassium, lactate, and glucose content of sweat can be measured simultaneously by electrochemical methods.”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [83]
The cited source says what the claim attributes to it.
In the citing paper
“Conventional commercially available integrated-circuit components (more than ten chips) can be applied on a FPCB to serve as data processing and transmission.[83]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [83]
The cited source says what the claim attributes to it.
In the citing paper
“Sweat [Na+] and [K+] were deemed stable throughout running in euhydration trials (with water intake of 150 ml per 5 min) after the initial [Na+] increase and [K+] decrease.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
The cited source says what the claim attributes to it.
In the citing paper
“Although such systems can support various types of biophysical measurements of physiological health (8–13), additional advances are needed to meet the challenging requirements of the NICU, where comprehensive, continuous sensing with wireless functionality, clinical-grade measurement fidelity, and mechanical form factors that eliminate risk of harm to exceptionally fragile neonatal skin are essent”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [10]
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Recently, tactile pressure sensors have been developed in the form of active‐matrix arrays consisting of pressure‐sensitive graphene transistors with air–dielectric layers.33”
From Wearables in Medicine (2018) · cites this paper as [33]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“These tactile sensors allowed detecting a broad pressure range from 250 Pa to 3 MPa.”
From Wearables in Medicine (2018) · cites this paper as [33]
Experimental — accuracy not yet measuredIn the citing paper
“Javey group, Gao group and Wang group also focused on electrochemical biosensing of sweat using various wearable sweat patches for sweat rate [30], L-Dopa [138,139], and C-reactive protein [140], cortisol [141,142], ketone [143], glucose [144], and therapeutic drugs [145].”
From The emergence of AI-based wearable sensors for digital health technology: a review (2023) · cites this paper as [30]
In the citing paper
“Wearable sensors are helping clinical and healthcare services with the possible shift from centralized clinical diagnostics to more individualized and personalized home care [29,30] (Figure 1a).”
From The emergence of AI-based wearable sensors for digital health technology: a review (2023) · cites this paper as [30]
In the citing paper
“In addition, since sweat is not sampled but sensed directly on the skin, the skin topography, skin temperature[ 10 ] and pH[ 11 ] may interfere considerably with the measurement results.”
From MicroSweat: A wearable microfluidic patch for noninvasive and reliable sweat collection enables human stress monitoring (2023) · cites this paper as [10]
In the citing paper
“In this work, the Berkeley team demonstrated the simultaneous multiplexed detection of sweat metabolites (glucose and lactate) and electrolytes (sodium and potassium ions), along with skin temperature by integrating a multi-sensing array.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [35]
In the citing paper
“The reported system proved advantageous for monitoring fitness parameters during exercise, but its utility would be limited in continuous monitoring applications due to its reliance on physical exertion for sweat generation.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [35]
In the citing paper
“Results of long-term stability studies of these electrochemical sensors indicate that the sensitivities of the biosensors are consistent over 2 wk with sensitivity variations of <5% (16).”
From Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose monitoring using a fully integrated wearable platform (2017) · cites this paper as [16]
In the citing paper
“We recently demonstrated a fully integrated wearable sensing system for real-time monitoring of multiple analytes in human perspiration during physical exercise which allows accurate measurement of sweat analytes through signal processing and calibration (16).”
From Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose monitoring using a fully integrated wearable platform (2017) · cites this paper as [16]
In the citing paper
“The system allowed for realtime monitoring of body temperature, electrolytes (sodium and potassium ions), and metabolites (glucose and lactate) during prolonged indoor and outdoor physical activities.”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“The multisensor array exhibited good mechanical resiliency with good RSD (1% for electrolytes and 5% for metabolites).”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“Javey and co-workers reported an integrated system consisting of a flexible multiplexed sensor array, a flexible printed circuitry board and a customized app for perspiration/sweat analysis (Figure 8b).[199]”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“Simultaneous and continuous monitoring of sweat profile during exercise can be conducted using wearable electrochemical sensors.[66,199]”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“As illustrated in Figure 10d, by wearing the device consisting of a multiplexed sensor array and a flexible printed circuitry board in the form of a wristband or headband, variations of skin temperature, electrolytes, and sweat metabolites in perspiration can be monitored wirelessly using a customized app on a smart phone.[199]”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“An increase in sweat [Na+] and a smaller increase in sweat [K+] were observed in dehydration trials (without water intake) after 80 min when subjects had lost a large amount of water (~2.5% of body weight).”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“Gao et al. developed a flexible sweat sensor for the real-time detection of Na+, lactate, K+, glucose, and skin temperature.44”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“As previously mentioned, Gao et al.44 fabricated an epidermal sensor on a PET substrate for the concurrent and continuous detection of sodium, lactate, potassium, glucose, and skin temperature.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“Ex situ measurements of [Na+] and [K+] from collected sweat samples demonstrated similar phenomena.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“Recently, Gao et al.41 demonstrated a wearable patch that can simultaneously track levels of metabolites and electrolytes in human sweat. However, electrophysiology sensors were not included, and such multimodal sensor fusion is crucial to obtain a more comprehensive knowledge about a wearer's well-being.”
From A wearable chemical–electrophysiological hybrid biosensing system for real-time health and fitness monitoring (2016) · cites this paper as [41]
In the citing paper
“Several research groups have recently developed mechanically flexible and fully integrated wearable sweat sensors (in wristband, temporary tattoo, eyeglasses, mouthguard) for analysis of biofluid metabolites and electrolytes (e.g., glucose, lactate, sodium, potassium) and pH sensing [156-160].”
From Wearables and the Medical Revolution (2018) · cites this paper as [156]
In the citing paper
“The wristband device also measures skin temperature for calibration, and that data is also useful in analyzing health as described earlier [126,156].”
From Wearables and the Medical Revolution (2018) · cites this paper as [156]
In the citing paper
“These devices have potential applications for measuring heat stress, dehydration and metabolism in athletes, soldiers, firefighters, and industrial laborers who exercise or work in hot environments.”
From A Critical Review of Consumer Wearables, Mobile Applications, and Equipment for Providing Biofeedback, Monitoring Stress, and Sleep in Physically Active Populations (2018) · cites this paper as Gao et al., 2016
Citations
Cites
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836 · cited 1×
“Cloud servers Transimpedance amplifier (1) Transimpedance amplifier (3)”
BackgroundSkin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes
Lipomi, D. J., Vosgueritchian, M., Tee, B. C.-K., Hellstrom, S. L., Lee, J. A., Fox, C. H., & Bao, Z. (2011) · Nature Nanotechnology · doi:10.1038/nnano.2011.184 · Cited by 3,188
An ultra-lightweight design for imperceptible plastic electronics
Kaltenbrunner, M., Sekitani, T., Reeder, J., Yokota, T., Kuribara, K., Tokuhara, T., Drack, M., Schwödiauer, R., Graz, I., Bauer-Gogonea, S., et al. (2013) · Nature · doi:10.1038/nature12314 · Cited by 2,485 · cited 2×
“Differential amplifier (6) Voltage buffer (8)”
BackgroundA review of wearable sensors and systems with application in rehabilitation
Patel, S., Park, H., Bonato, P., Chan, L., & Rodgers, M. (2012) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-9-21 · Cited by 2,396 · cited 1×
“Bluetooth transceiver (11) Iglucose Ilactate Rtemperature Vsodium Vref Vpotassium Ag/AgCl Gluc ose”
BackgroundElectrochemical Biosensors - Sensor Principles and Architectures
Grieshaber, D., MacKenzie, R., Vörös, J., & Reimhult, E. (2008) · Sensors · doi:10.3390/s80314000 · Cited by 2,329
Piezoelectricity of single-atomic-layer MoS2 for energy conversion and piezotronics
Wu, W., Wang, L., Li, Y., Zhang, F., Lin, L., Niu, S., Chenet, D., Zhang, X., Hao, Y., Heinz, T. F., et al. (2014) · Nature · doi:10.1038/nature13792 · Cited by 2,287 · cited 1×
“Low-pass filter (2) Low-pass filter (4) Voltage divider”
MethodNanowire active-matrix circuitry for low-voltage macroscale artificial skin
Takei, K., Takahashi, T., Ho, J. C., Ko, H., Gillies, A. G., Leu, P. W., Fearing, R. S., & Javey, A. (2010) · Nature Materials · doi:10.1038/nmat2835 · Cited by 1,274 · cited 1×
“Cloud servers Transimpedance amplifier (1) Transimpedance amplifier (3)”
BackgroundSelf-citationNon-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247
Ultrathin conformal devices for precise and continuous thermal characterization of human skin
Webb, R. C., Bonifas, A. P., Behnaz, A., Zhang, Y., Yu, K. J., Cheng, H., Shi, M., Bian, Z., Liu, Z., Kim, Y.-S., et al. (2013) · Nature Materials · doi:10.1038/nmat3755 · Cited by 1,229 · cited 1×
“…dehydration 20 ; sweat glucose is reported to be metabolically related to blood glucose 21 ; sweat lactate can potentially serve as a sensitive marker of pressure ischaemia 22 ; and skin temperature is clinically informative of a variety of diseases and skin injuries such as pressure ulcers 23,24 .”
BackgroundSoft Microfluidic Assemblies of Sensors, Circuits, and Radios for the Skin
Xu, S., Zhang, Y., Jia, L., Mathewson, K. E., Jang, K.-I., Kim, J., Fu, H., Huang, X., Chava, P., Wang, R., et al. (2014) · Science · doi:10.1126/science.1250169 · Cited by 1,199 · cited 1×
“Low-pass filter (7) Low-pass filter (9)”
Cited by
Sorted by citations
Items 1,601–1,700 of 5,047
Carbon nanotube electronics for IoT sensors
Cardenas, J. A., Andrews, J. B., Noyce, S. G., & Franklin, A. D. (2019) · Nano Futures · doi:10.1088/2399-1984/ab5f20 · Cited by 58
Tang, Y., Li, X., Lv, H., Wang, W., Zhi, C., & Li, H. (2020) · InfoMat · doi:10.1002/inf2.12102 · Cited by 55
“Compared with sensors, the power supplies are usually much bulkier, of which flexible miniaturization is more important.(10,14) Simultaneously, the other performance indexes of power supplies such as energy density, power density, mechanical stability, and service life need to be considered as well.”
BackgroundMethodInfluentialExchange Bias in a La0.67Sr0.33MnO3/NiO Heterointerface Integrated on a Flexible Mica Substrate
Huang, J., Wang, H., Wang, H., et al. (2020) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.0c12935 · Cited by 55
Study of the Dielectric Properties of Artificial Sweat Mixtures at Microwave Frequencies
Eldamak, A. R., Thorson, S., & Fear, E. C. (2020) · Biosensors · doi:10.3390/bios10060062 · Cited by 58
All-Solid-State Ion-Selective Electrodes: A Tutorial for Correct Practice
Criscuolo, F., Hanitra, I. N., Taurino, I., Carrara, S., & Micheli, G. D. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3099209 · Cited by 57
Realizing High‐Energy and Stable Wire‐Type Batteries with Flexible Lithium–Metal Composite Yarns
Gao, Y., Hu, H., Chang, J., et al. (2021) · Advanced Energy Materials · doi:10.1002/aenm.202101809 · Cited by 54
Integrated Aptasensor Array for Sweat Drug Analysis
Zhang, X., Tang, Y., Wu, H., Wang, Y., Niu, L., & Li, F. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.2c00736 · Cited by 54
Sinha, A. K., Goh, G. L., Yeong, W. Y., & Cai, Y. (2022) · Advanced Materials Interfaces · doi:10.1002/admi.202200621 · Cited by 55
Chen, Y., Ling, Y., & Yin, R. (2022) · Sensors · doi:10.3390/s22249716 · Cited by 59
Power‐Free Contact Lens for Glucose Sensing
Li, Z., Yun, J., Li, X., Kim, M., Li, J., Lee, D., Wu, A., & Lee, S. W. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202304647 · Cited by 57
Flexible organic electrochemical transistors for bioelectronics
Zhao, Z., Tian, Z., & Yan, F. (2023) · Cell Reports Physical Science · doi:10.1016/j.xcrp.2023.101673 · Cited by 58
Liu, L., Luo, T., Kuang, X., et al. (2024) · ACS Sensors · doi:10.1021/acssensors.4c00164 · Cited by 54
Permeable Bioelectronics toward Biointegrated Systems
Lee, S., Liang, X., Kim, J. S., Yokota, T., Fukuda, K., & Someya, T. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00823 · Cited by 59
Zhuang, Q., Yao, K., Song, X., Zhang, Q., Zhang, C., Wang, H., Yang, R., Zhao, G., Li, S., Shu, H., et al. (2025) · Science Advances · doi:10.1126/sciadv.adu3146 · Cited by 65
Continuous biochemical profiling of the gastrointestinal tract using an integrated smart capsule
Min, J., Ahn, H., Lukas, H., Ma, X., Bhansali, R., Sunwoo, S., Wang, C., Xu, Y., Yao, D. R., Kim, G., et al. (2025) · Nature Electronics · doi:10.1038/s41928-025-01407-0 · Cited by 59
“For stabilization, the electrode was preconditioned in a solution containing 1 M KCl for 3 h before measurements 46 .”
Self-citationByun, J., Oh, E., Lee, B., Kim, S., Lee, S., & Hong, Y. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201701912 · Cited by 54
Vairo, D., Bruzzese, L., Marlinge, M., Fuster, L., Adjriou, N., Kipson, N., Brunet, P., Cautela, J., Jammes, Y., Mottola, G., et al. (2017) · Scientific Reports · doi:10.1038/s41598-017-12211-y · Cited by 55
Furniturewalla, A., Chan, M., Sui, J., Ahuja, K., & Javanmard, M. (2018) · Microsystems & Nanoengineering · doi:10.1038/s41378-018-0019-0 · Cited by 56
Fabric-based self-powered noncontact smart gloves for gesture recognition
Wu, H., Guo, H., Su, Z., Shi, M., Chen, X., Cheng, X., Han, M., & Zhang, H. (2018) · Journal of Materials Chemistry A · doi:10.1039/c8ta08276f · Cited by 52
Chun, S., Son, W., Lee, G., et al. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b20143 · Cited by 54
Carbon Nanomaterial-Based Biosensors: A Review of Design and Applications
Lin, Z., Wu, G., Zhao, L., & Lai, K. W. C. (2019) · IEEE Nanotechnology Magazine · doi:10.1109/mnano.2019.2927774 · Cited by 56
“[112]”
Gelatin-hydrogel based organic synaptic transistor
Lai, D., Li, E., Yan, Y., Liu, Y., Zhong, J., Lv, D., Ke, Y., Chen, H., & Guo, T. (2019) · Organic Electronics · doi:10.1016/j.orgel.2019.105409 · Cited by 55
Self-Powered, Self-Healed, and Shape-Adaptive Ultraviolet Photodetectors
Tsai, M.-S., Shen, T., Wu, H.-M., Liao, Y., Liao, Y.-K., Lee, W., Kuo, H., Lai, Y., & Chen, Y. (2020) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.9b21446 · Cited by 53
Wang, R., Du, Z., Xia, Z., Liu, J., Li, P., Wu, Z., Yue, Y., Xiang, Y., Meng, J., Liu, D., et al. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202107682 · Cited by 54
A Low-Cost Multi-Parameter Water Quality Monitoring System
Alam, A. U., Clyne, D., & Deen, M. J. (2021) · Sensors · doi:10.3390/s21113775 · Cited by 64
“The commercial Ag/AgCl pastes are composed of polymer resins such as polyvinyl butyral (PVB), which facilitates maintaining a stable potential over a long period of time in samples with different ionic strengths [17].”
BackgroundFazio, R. D., Vittorio, M. D., & Visconti, P. (2021) · Electronics · doi:10.3390/electronics10141660 · Cited by 57
“sampling [52] (Reprinted with permission from ref.”
BackgroundInfluentialIsland Effect in Stretchable Inorganic Electronics
Li, K., Shuai, Y., Cheng, X., Luan, H., Liu, S., Yang, C., Xue, Z., Huang, Y. Y., & Zhang, Y. (2022) · Small · doi:10.1002/smll.202107879 · Cited by 57
Chu, S. S., Nguyen, H. A., Zhang, J., Tabassum, S., & Cao, H. (2022) · Sensors · doi:10.3390/s22145200 · Cited by 56
Heat-shedding with photonic structures: radiative cooling and its potential
Heo, S., Lee, G. J., & Song, Y. M. (2022) · Journal of Materials Chemistry C · doi:10.1039/d2tc00318j · Cited by 52
Kim, Y.-J., Lee, J., Hong, H., Park, S., & Ryu, W. (2022) · Advanced Engineering Materials · doi:10.1002/adem.202200873 · Cited by 55
Khan, A., Haque, M. N., Kabiraz, D. C., Yeasin, A., Rashid, H. A., Sarker, A. C., & Hossain, G. (2022) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2022.114093 · Cited by 57
A 3D-printing approach toward flexible piezoelectronics with function diversity
Yuan, X., Mai, Z., Li, Z., et al. (2023) · Materials Today · doi:10.1016/j.mattod.2023.08.023 · Cited by 53
Li, Y., Guo, Y., Chen, H., Xiao, X., Long, F., Zhong, H., Wang, K., Guo, Z., Zhuang, Z., & Liu, Z. (2024) · ACS Photonics · doi:10.1021/acsphotonics.3c01490 · Cited by 53
Active electronic skin: an interface towards ambient haptic feedback on physical surfaces
Guo, Y. C., Wang, Y., Tong, Q., Shan, B., He, L., Zhang, Y., & Wang, D. (2024) · npj Flexible Electronics · doi:10.1038/s41528-024-00311-5 · Cited by 65
“…propel-ling the rapid development of fl exible tactile sensing and its wide applications in fi elds such as robotics 190,191 , healthcare 192,193 , and human-computer interaction 194,195 Piezoresistive sensors: Piezoresistive sensors measure interactive information based on the resistive effect.”
Flexible epidermal wearable sensor for Athlete's sweat biomarkers monitoring
Cai, J., Cao, M., Bai, J., Sun, M., Ma, C., Emran, M. Y., Kotb, A., Bo, X., & Zhou, M. (2024) · Talanta · doi:10.1016/j.talanta.2024.126986 · Cited by 59
Real-time personal healthcare data analysis using edge computing for multimodal wearable sensors
Matsumura, G., Honda, S., Kikuchi, T., Mizuno, Y., Hara, H., Kondo, Y., Nakamura, H., Watanabe, S., Hayakawa, K., Nakajima, K., et al. (2024) · Device · doi:10.1016/j.device.2024.100597 · Cited by 60
Piezoelectric composites for gas sensing: evolution of sensing and transduction designs
Li, W., Xie, G., Huo, X., Que, L., Tai, H., Jiang, Y., & Su, Y. (2025) · Journal of Materials Chemistry C · doi:10.1039/d5tc01383f · Cited by 56
Skin-like drift-free biosensors with stretchable diode-connected organic field-effect transistors
Zhao, C., Park, J., Maulà, D., Yuan, Y., Zhong, D., Wang, W., Liu, Q., Xu, C., Zheng, Y. Q., Mow, R. K., et al. (2025) · Nature Electronics · doi:10.1038/s41928-025-01465-4 · Cited by 60
“We used the well-developed glucose oxidase (GOx)-enzyme-based glucose sensing 50,51 .”
Hooton, K., Han, W., & Li, L. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b01930 · Cited by 58
Detection of tuberculosis in patients with the use of portable SPR device
Trzaskowski, M., Napiórkowska, A., AUGUSTYNOWICZ-KOPEĆ, E. M., & Ciach, T. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2017.12.183 · Cited by 52
Flexible Crossbar‐Structured Phase Change Memory Array via Mo‐Based Interfacial Physical Lift‐Off
Kim, D. H., Lee, H. E., You, B. K., Cho, S. B., Mishra, R., Kang, I., & Lee, K. J. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201806338 · Cited by 52
Liang, G., Ruan, Z., Liu, Z., Li, H., Wang, Z., Tang, Z., Mo, F., Yang, Q., Ma, L., Wang, D., et al. (2019) · Advanced Electronic Materials · doi:10.1002/aelm.201900553 · Cited by 53
Flexible Ultralow-Power Sensor Interfaces for E-Skin
Jiang, C. H., Cheng, X., & Nathan, A. (2019) · Proceedings of the IEEE · doi:10.1109/jproc.2019.2936105 · Cited by 55
“In addition, e-skin relies on batteries and warrants eco-friendly electronic devices, so minimizing power consumption and fabrication costs are important [7]–[11].”
BackgroundDautta, M., Alshetaiwi, M., Escobar, A. R., Torres, F., Bernardo, N., & Tseng, P. (2020) · Advanced Electronic Materials · doi:10.1002/aelm.201901311 · Cited by 55
Self-powered wearable electronics
Tan, P., Zou, Y., Fan, Y., & Li, Z. (2020) · Wearable Technologies · doi:10.1017/wtc.2020.3 · Cited by 58
“In 2016, Gao et al. developed a wearable flexible integrated sensing array for multiplexed in situ perspiration analysis (Gao et al., 2016) (Figure 5a).”
BackgroundHow is flexible electronics advancing neuroscience research?
Chen, Y., Rommelfanger, N. J., Mahdi, A. I., Wu, X., Keene, S. T., Obaid, A., Salleo, A., Wang, H., & Hong, G. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.120559 · Cited by 58
Trilayer MXene Fabric for Integrated Ultrasensitive Pressure Sensor and Wearable Heater
Deng, C., Zhao, S., Su, E., Li, Y., & Wu, F. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202100574 · Cited by 53
Conformal electrodes for on‐skin digitalization
Wang, C., He, K., Li, J., & Chen, X. (2021) · SmartMat · doi:10.1002/smm2.1068 · Cited by 54
“functionality, the preferred approach is to combine rigid chips with flexible/stretchable substrates/interconnections to achieve sensor digitization on the skin.(65) The system structure usually is designed with three strategies, for example, in‐plane interconnections,(66) vertical integration strategies,(67) and “isl…”
MethodZhu, Y., Wang, B., Deng, C., et al. (2021) · Nano Energy · doi:10.1016/j.nanoen.2021.105801 · Cited by 52
Özer, T., & Henry, C. S. (2021) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/abfc9f · Cited by 52
A review of inkjet printing technology for personalized-healthcare wearable devices
Du, X., Wankhede, S. P., Prasad, S., Shehri, A., Morse, J. D., & Lakal, N. (2022) · Journal of Materials Chemistry C · doi:10.1039/d2tc02511f · Cited by 56
Battery-free, wireless, and electricity-driven soft swimmer for water quality and virus monitoring
Li, D., Zhou, J., Zhao, Z., Huang, X., Li, H., Qu, Q., Zhou, C., Yao, K., Liu, Y., Wu, M., et al. (2024) · Science Advances · doi:10.1126/sciadv.adk6301 · Cited by 51
“An Ag/AgCl layer was fabricated on the middle circular electrode and coated with a polyvinyl butyral (PVB) reference cocktail to form the reference electrode ( 40 ).”
Sensors, society, and sustainability
Chaudhary, V., Gaur, P., & Rustagi, S. (2024) · Sustainable materials and technologies · doi:10.1016/j.susmat.2024.e00952 · Cited by 59
Nanotechnology-Based Wearable Electrochemical Biosensor for Disease Diagnosis
Yoon, J., Kwon, N., Lee, Y., Kim, S., Lee, T., & Choi, J. (2025) · ACS Sensors · doi:10.1021/acssensors.4c03371 · Cited by 56
Wearable Carbon Nanotube-Based Biosensors on Gloves for Lactate
Luo, X., Shi, W., Yu, H., Xie, Z., Li, K., & Cui, Y. (2018) · Sensors · doi:10.3390/s18103398 · Cited by 54
Mano, T., Nagamine, K., Ichimura, Y., Shiwaku, R., Furusawa, H., Matsui, H., Kumaki, D., & Tokito, S. (2018) · ChemElectroChem · doi:10.1002/celc.201801129 · Cited by 52
Lin, Z., Wu, Y., He, Q., Sun, C., Fan, E., Zhou, Z., Liu, M., Wei, W., & Yang, J. (2019) · Nanoscale · doi:10.1039/c9nr00083f · Cited by 51
General-purpose passive wireless point–of–care platform based on smartphone
Escobedo, P., Erenas, M. M., Martínez-Olmos, A., Carvajal, M. A., González-Chocano, S., Capitán-Vallvey, L. F., & Palma, A. J. (2019) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2019.111360 · Cited by 52
Precision pH Sensor Based on WO 3 Nanofiber-Polymer Composites and Differential Amplification
Choi, S., Savagatrup, S., Kim, Y., Lang, J. H., & Swager, T. M. (2019) · ACS Sensors · doi:10.1021/acssensors.9b01579 · Cited by 52
The Role and Challenges of Body Channel Communication in Wearable Flexible Electronics
Zhao, B., Mao, J., Zhao, J. P., Yang, H., & Lian, Y. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.2966285 · Cited by 53
“detect various physiological signals, such as tactile [6], ECG [7], [8], perspiration [9], [10], temperature [11], displacement [12],”
BackgroundMethodDou, J., Tang, L., Mou, L., Zhang, R., & Jiang, X. (2020) · Composites Science and Technology · doi:10.1016/j.compscitech.2020.108237 · Cited by 52
Hondred, J. A., Johnson, Z. T., & Claussen, J. C. (2020) · Journal of Materials Chemistry C · doi:10.1039/d0tc01423k · Cited by 53
Wang, X., Meng, X. M., Zhu, Y., Ling, H., Chen, Y., Li, Z., Hartel, M. C., Dokmeci, M. R., Zhang, S., & Khademhosseini, A. (2020) · IEEE Electron Device Letters · doi:10.1109/led.2020.3042310 · Cited by 52
Systematic Review on Human Skin-Compatible Wearable Photoplethysmography Sensors
Lee, I., Park, N., Lee, H., Hwang, C., Kim, J. H., & Park, S. (2021) · Applied Sciences · doi:10.3390/app11052313 · Cited by 52
“Wearable devices are also becoming widely used to gather information about an individual’s muscle activity, sleep quality, and other physiological activities in daily life [8,9].”
BackgroundWearable Thermocells Based on Gel Electrolytes for the Utilization of Body Heat
Yang, P., Liu, K., Chen, Q., Mo, X., Zhou, Y., Li, S., Feng, G., & Zhou, J. (2016) · Angewandte Chemie · doi:10.1002/ange.201606314 · Cited by 51
A self-powered brain multi-perception receptor for sensory-substitution application
Fu, Y., Zhang, M., Dai, Y., Zeng, H. J., Sun, C., Han, Y., Xing, L., Wang, S., Xue, X., Zhan, Y., et al. (2017) · Nano Energy · doi:10.1016/j.nanoen.2017.11.068 · Cited by 52
Mahanty, B., Ghosh, S. K., Jana, S., Mallick, Z., Sarkar, S., & Mandal, D. (2021) · Sustainable Energy & Fuels · doi:10.1039/d1se00444a · Cited by 53
Zhang, L., Huang, K., Wen, P., et al. (2021) · Energy storage materials · doi:10.1016/j.ensm.2021.07.041 · Cited by 51
Oral wearable sensors: Health management based on the oral cavity
Li, Y., Tang, H., Liu, Y., Qiao, Y., Xia, H.-Q., & Zhou, J. (2022) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2022.100135 · Cited by 56
Wearable Near-Field Communication Sensors for Healthcare: Materials, Fabrication and Application
Sun, X., Zhao, C., Li, H., Yu, H., Zhang, J., Qiu, H., Liang, J., Wu, J., Su, M., Shi, Y., et al. (2022) · Micromachines · doi:10.3390/mi13050784 · Cited by 58
A responsive hydrogel-based microneedle system for minimally invasive glucose monitoring
Wang, Y., Liu, H., Yang, X., Shi, Z., Li, J., Xue, L., Liu, S., & Lei, Y. (2022) · Smart Materials in Medicine · doi:10.1016/j.smaim.2022.07.006 · Cited by 53
Moving toward smart biomedical sensing
Naghdi, T., Ardalan, S., Adib, Z. A., Sharifi, A. R., & Golmohammadi, H. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.115009 · Cited by 52
Nanostructured wearable electrochemical and biosensor towards healthcare management: a review
Khaleque, M. A., Hossain, M. I., Ali, M. R., Bacchu, M. S., Aly, M. A. S., & Khan, M. Z. H. (2023) · RSC Advances · doi:10.1039/d3ra03440b · Cited by 55
Skin-interfacing wearable biosensors for smart health monitoring of infants and neonates
Zhou, L., Guess, M., Kim, K. R., & Yeo, W. (2024) · Communications Materials · doi:10.1038/s43246-024-00511-6 · Cited by 61
Asaduzzaman, M., Faruk, O., Samad, A. A., Kim, H., Reza, M. S., Lee, Y., & Park, J. Y. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202405651 · Cited by 52
Advanced functional chitosan-based nanocomposite materials for performance-demanding applications
Guo, Y., Qiao, D., Zhao, S., Zhang, B., & Xie, F. (2024) · Progress in Polymer Science · doi:10.1016/j.progpolymsci.2024.101872 · Cited by 67
Minimally and non-invasive glucose monitoring: the road toward commercialization
Min, S. Q., Geng, H., He, Y., Xu, T., Liu, Q., & Zhang, X. (2025) · Sensors & Diagnostics · doi:10.1039/d4sd00360h · Cited by 56
Wearable microfluidics: fabric-based digital droplet flowmetry for perspiration analysis
Yang, Y., Xing, S., Fang, Z., Li, R., Koo, H. S., & Pan, T. (2017) · Lab on a Chip · doi:10.1039/c6lc01522k · Cited by 50
Kim, J., Kim, N., Kwon, M., & Lee, J. (2017) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.7b05264 · Cited by 51
Sankhala, D., Muthukumar, S., & Prasad, S. (2018) · SLAS TECHNOLOGY · doi:10.1177/2472630318759257 · Cited by 51
Highly Stable Battery Pack via Insulated, Reinforced, Buckling‐Enabled Interconnect Array
Yin, L., Seo, J. K., Kurniawan, J. F., Kumar, R., Lv, J., Xie, L., Liu, X., Xu, S. H., Meng, Y. S., & Wang, J. (2018) · Small · doi:10.1002/smll.201800938 · Cited by 51
Ling, W., Hao, Y., Wang, H., Xu, H., & Huang, X. L. (2019) · Nanotechnology · doi:10.1088/1361-6528/ab30b6 · Cited by 51
Wearable capillary microfluidics for continuous perspiration sensing
Ma, B., Chi, J., Xu, C., Ni, Y., Zhao, C., & Liu, H. (2020) · Talanta · doi:10.1016/j.talanta.2020.120786 · Cited by 51
Dual-Mode Photonic Sensor Array for Detecting and Discriminating Hydrazine and Aliphatic Amines
Liu, T., Yang, L., Feng, W. Y., et al. (2020) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.0c00568 · Cited by 51
Wearable electrochemical sensors for noninvasive monitoring of health—a perspective
Liu, H., & Zhao, C. (2020) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2020.03.008 · Cited by 53
Dong, H., Zhang, L., Wu, T., Song, H., Luo, J., Huang, F., & Zuo, C. (2020) · Organic Electronics · doi:10.1016/j.orgel.2020.106044 · Cited by 51
Liu, X., Zhang, W., Lin, Z., Meng, Z., Shi, C., Xu, Z., Yang, L., & Liu, X. Y. (2021) · Small Methods · doi:10.1002/smtd.202000926 · Cited by 52
Karki, H. P., Jang, Y., Jung, J., & Oh, J. (2021) · Journal of Nanobiotechnology · doi:10.1186/s12951-021-00814-7 · Cited by 51
Park, T., Woo, H. K., Jung, B. K., Park, B., Bang, J., Kim, W., Jeon, S., Ahn, J., Lee, Y., Lee, Y. M., et al. (2021) · ACS Nano · doi:10.1021/acsnano.0c09835 · Cited by 52
Flexible energy storage devices for wearable bioelectronics
Ma, X., Jiang, Z., & Lin, Y. (2021) · Journal of Semiconductors · doi:10.1088/1674-4926/42/10/101602 · Cited by 52
Polymers and Plastics Modified Electrodes for Biosensors: A Review
Lanzalaco, S., & Molina, B. G. (2020) · Molecules · doi:10.3390/molecules25102446 · Cited by 51
Zhang, X., Xia, Y., Liu, Y., Mugo, S. M., & Zhang, Q. M. (2021) · Analytical Chemistry · doi:10.1021/acs.analchem.1c03884 · Cited by 50
Zhang, S., Suresh, L., Yang, J., Zhang, X., & Tan, S. C. (2022) · Advanced Intelligent Systems · doi:10.1002/aisy.202100194 · Cited by 56
Huang, Z., Xu, Y., Cheng, Y., Xue, M., Deng, M., Jaffrezic-Renault, N. J., & Guo, Z. (2022) · Sensors & Diagnostics · doi:10.1039/d2sd00037g · Cited by 53
Chen, S., Hou, K., Li, T., Wu, X., Wang, Z., Wei, L., & Leong, W. L. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200611 · Cited by 57
Flexible organic integrated electronics for self-powered multiplexed ocular monitoring
Lin, B., Wang, M., Zhao, C., Wang, S., Chen, K., Li, X., Long, Z., Zhao, C., Song, X., Yan, S., et al. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00211-6 · Cited by 59
Self-citationWang, H., Xiang, Z., Zhao, P., Wan, J., Miao, L., Guo, H., Xu, C., Zhao, W. G., Han, M., & Zhang, H. (2022) · ACS Nano · doi:10.1021/acsnano.2c05299 · Cited by 52
Paper-based sensors for rapid important biomarkers detection
Kuswandi, B., Hidayat, M. A., & Noviana, E. (2022) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2022.100246 · Cited by 50
Mugo, S. M., Lu, W., & Robertson, S. V. (2022) · Biosensors · doi:10.3390/bios12100854 · Cited by 52
Min, Z., Chang, B., Shao, C., Su, X., Wang, N., Li, Z., Wang, H., Zhao, Y., Fan, M., & Wang, J. (2022) · Applied Catalysis B: Environmental · doi:10.1016/j.apcatb.2022.122185 · Cited by 50
Metadata
- Authors
- Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., Lien, D.-H., Brooks, G. A., Davis, R. W., & Javey, A.
- Affiliations
- Lawrence Berkeley National Laboratory (US)
- Stanford Medicine (US)
- University of California, Berkeley (US)
- Corresponding author
- Javey, Ali
- ORCID
- Venue
- Nature
- Published
- 2016-01-27
- Volume
- 529
- Issue
- 7587
- Pages
- 509-514
- ISSN
- 0028-0836 (ISSN-L)
- 1476-4687
- PMC
- PMC4996079
- OpenAlex
- W2266175888
- Semantic Scholar
- 52709f85aaaaea3dbceb33a9666e3cb1dfb932a3
- Cited by
- 5,087
- References
- 39
- Citations by year
- 2026: 462 · 2025: 581 · 2024: 528 · 2023: 543 · 2022: 531 · 2021: 585 · 2020: 538 · 2019: 502 · 2018: 415 · 2017: 282 · 2016: 117
- FWCI
- 211.10
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Nature (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) (submittedVersion)
- CaltechAUTHORS (California Institute of Technology) (submittedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- eScholarship (California Digital Library) (submittedVersion)
- Rare & Special e-Zone (The Hong Kong University of Science and Technology) (submittedVersion)
- Funders
- National Science Foundation (DEAC02-05CH11231)
- U.S. Department of Energy (-AC02-05CH11231, 05CH11231, DE AC02 05CH11231, DE-AC02-)
- National Institutes of Health (DE-AC02–05CH11231, HG000205)
- Office of Science (DEAC02–05CH11231, AC02-05CH11231, DE-AC02)
- Japan Society for the Promotion of Science
- National Human Genome Research Institute (P01HG000205)
- Basic Energy Sciences (DE-AC02, DE-AC02-05CH11231., AC02–05CH11231)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Keywords
- wearable sensors, sweat analysis, sweat electrolytes, glucose, lactate, sodium, potassium, skin temperature, flexible electronics, silicon integrated circuits, multiplexed biosensing, wireless transmission, real-time physiological monitoring, non-invasive monitoring, wearable biosensors, physical activity monitoring, personalized medicine, drug abuse detection, athletic performance
- MeSH terms
- Adult, Bicycling, Body Water, Calibration, Electrolytes, Female, Glucose, Humans, Male, Monitoring, Physiologic, Running, Skin, Skin Temperature, Sweat, Reproducibility of Results, Lactic Acid, Young Adult, Precision Medicine, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-09-28, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-09-28 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review