A wearable chemical–electrophysiological hybrid biosensing system for real-time health and fitness monitoring
Open accessAbstract
Flexible, wearable sensing devices can yield important information about the underlying physiology of a human subject for applications in real-time health and fitness monitoring. Despite significant progress in the fabrication of flexible biosensors that naturally comply with the epidermis, most designs measure only a small number of physical or electrophysiological parameters, and neglect the rich chemical information available from biomarkers. Here, we introduce a skin-worn wearable hybrid sensing system that offers simultaneous real-time monitoring of a biochemical (lactate) and an electrophysiological signal (electrocardiogram), for more comprehensive fitness monitoring than from physical or electrophysiological sensors alone. The two sensing modalities, comprising a three-electrode amperometric lactate biosensor and a bipolar electrocardiogram sensor, are co-fabricated on a flexible substrate and mounted on the skin. Human experiments reveal that physiochemistry and electrophysiology can be measured simultaneously with negligible cross-talk, enabling a new class of hybrid sensing devices.
AI snapshot
Written from the paper’s full text.
TL;DR
A screen-printed, skin-worn hybrid patch simultaneously measures sweat lactate and electrocardiogram (ECG) with negligible cross-talk. In three healthy male subjects during cycling, ECG-derived heart rate correlated strongly with a commercial monitor, and lactate response tracked exertion.
Key findings
- The Chem–Phys patch combined a three-electrode amperometric lactate biosensor and bipolar ECG electrodes on a 50-μm polyester sheet; the full patch was 7 × 2 cm², with 1.5 × 1.5 cm² ECG electrodes separated by 4 cm and lactate electrodes of 3 × 2.5 mm active area.
- In vitro, the lactate biosensor responded linearly from 0–28 mM with a sensitivity of 96 nA/mM; human sweat lactate typically ranges from 0–25 mM.
- In three healthy male subjects during 15–30 min of intense cycling, ECG and lactate were measured simultaneously with negligible co-interference; heart rate extracted from the patch matched a commercial wristband monitor with Pearson’s r = 0.975.
- An enzyme-free control amperometric sensor produced negligible current under the same conditions, supporting selectivity of the LOx-based lactate sensor.
Methods
- The hybrid patch was fabricated by screen-printing Ag/AgCl, Prussian blue, and insulator inks on flexible polyester; the lactate working electrode was functionalized with lactate oxidase and coated with polyvinyl chloride, and a printed hydrophobic Ecoflex layer separated the amperometric and ECG electrodes.
- The patch interfaced with a custom 4-layer PCB containing a potentiostat, ECG analog front-end, and Bluetooth low-energy radio; in vitro lactate characterization used 0.1 M phosphate-buffered solution (pH 7.0) at −0.1 V versus Ag/AgCl for 60 s.
- Human testing involved 3 healthy male volunteers with no prior heart conditions, diabetes, or chronic skeletomuscular pain; the patch was placed on the fourth intercostal space between V1 and V2, and subjects cycled on a stationary bicycle with increasing resistance every 3 min followed by a 3-min cool-down.
Limitations
- The paper does not include a dedicated limitations section; it describes the device as “an important first step” and states that all relevant data are available from the authors on request.
- Evident from the text: the human validation was very small (3 healthy male volunteers) and short-term (15–30 min exercise bouts), with no female, clinical, or broader-population testing.
- Evident from the text: the study did not assess long-term wear, repeated use, or multi-day stability, and lactate was measured in sweat rather than blood.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 10 cited sources checked · 0 supported · 0 flagged
- Findingreasoning
The hydrophobic barrier between the lactate and ECG sensors assisted in minimizing potential cross-talk between the two sensing modalities.
“These on-body studies illustrate that the hybrid patch could monitor sweat lactate and electrocardiogram in a continuous and simultaneous manner, and that the hydrophobic barrier between the sensors assisted in minimizing potential cross-talk between the two sensing modalities.”
- Also stated in: § Hybrid patch design rationale and fabrication · plain
“This effectively increases the impedance between the electrocardiogram and amperometric electrodes via sweat, thus minimizing the cross-talk between the two sensors.”
Quote in the paperSubstantial claim - Also stated in: § Hybrid patch design rationale and fabrication · plain
- Findingown data
Heart rate extracted from the Chem-Phys patch ECG data matched a commercial wristband heart rate monitor with a Pearson's correlation coefficient of r=0.975.
“Extracted heart rate data matched the wrist-worn device with a Pearson's correlation coefficient of r=0.975.”
Quote in the paperSubstantial claim- Rests on an assumption: The commercial wristband heart rate monitor (BASIS) provides an accurate heart rate reference.
- Findingown data
The hydrophobic barrier had minimal effect on the supply of oxygen to the enzyme electrode required for biocatalytic detection of lactate.
“The data also demonstrates that such a barrier had minimal effect on the supply of oxygen to the enzyme electrode required for biocatalytic detection of lactate.”
Quote in the paperSubstantial claim - Findingown data
On-body ECG signals recorded using the fabricated electrodes have similar morphologies to those recorded using commercial 3M Red Dot ECG electrodes at the same location with the same AFE circuitry.
“As illustrated in Fig. 2b, on-body electrocardiogram signals recorded for the same subject using commercial and fabricated electrodes at the same location have similar morphologies when acquired using the same AFE circuitry.”
Quote in the paperSubstantial claim- Rests on an assumption: The commercially available 3M Red Dot ECG electrodes provide a valid reference for ECG morphology.
- Findingown data
Physiochemistry and electrophysiology can be measured simultaneously with negligible cross-talk.
“Human experiments reveal that physiochemistry and electrophysiology can be measured simultaneously with negligible cross-talk, enabling a new class of hybrid sensing devices.”
- Also stated in: § Epidermal evaluation of the Chem–Phys patch · hedged
“These on-body studies illustrate that the hybrid patch could monitor sweat lactate and electrocardiogram in a continuous and simultaneous manner, and that the hydrophobic barrier between the sensors assisted in minimizing potential cross-talk between the two sensing modalities.”
Quote in the paperSubstantial claim - Also stated in: § Epidermal evaluation of the Chem–Phys patch · hedged
- Findingown data
The Chem-Phys hybrid patch enables simultaneous real-time monitoring of lactate and electrocardiogram.
“Here, we introduce a skin-worn wearable hybrid sensing system that offers simultaneous real-time monitoring of a biochemical (lactate) and an electrophysiological signal (electrocardiogram), for more comprehensive fitness monitoring than from physical or electrophysiological sensors alone.”
- Also stated in: § Epidermal evaluation of the Chem–Phys patch · hedged
“These on-body studies illustrate that the hybrid patch could monitor sweat lactate and electrocardiogram in a continuous and simultaneous manner, and that the hydrophobic barrier between the sensors assisted in minimizing potential cross-talk between the two sensing modalities.”
- Also stated in: § Discussion · plain
“On-body epidermal testing in a realistic fitness environment revealed that electrocardiogram sensing is in-line with existing wearable devices, and is not adversely affected by simultaneous measurement of lactate via constant-potential amperometry.”
Quote in the paperSubstantial claim - Also stated in: § Epidermal evaluation of the Chem–Phys patch · hedged
- Interpretationreasoning
The Chem-Phys patch provides more comprehensive monitoring/view of a person's health status than current wearable fitness monitors or physical/electrophysiological sensors alone.
“Here, we introduce a skin-worn wearable hybrid sensing system that offers simultaneous real-time monitoring of a biochemical (lactate) and an electrophysiological signal (electrocardiogram), for more comprehensive fitness monitoring than from physical or electrophysiological sensors alone.”
- Also stated in: § Abstract · plain
“By combining a lactate biosensor and an electrocardiogram sensor, the new Chem–Phys hybrid wearable patch represents a powerful platform capable of simultaneously tracking both physicochemical and electrophysiological attributes, thus providing a more comprehensive view of a person's health status than current wearable fitness monitors.”
Quote in the paperSubstantial claim - Also stated in: § Abstract · plain
- Interpretationcitation
The lactate biosensor data for each subject resembles the expected sweat-lactate profile for increasing intensity workouts.
“The lactate biosensor data for each subject resembles the expected sweat-lactate profile for increasing intensity workouts37.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [37] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013). source
Not checked in the free run.
- Premisecitation
Sweat rate increase causes a dilution factor that decreases lactate concentration.
“As the cycling continued, the sweat rate for each subject increased, leading to the well-documented phenomenon of dilution factor that causes decrease in the lactate concentration42.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [42] Buono, M. J., Lee, N. V. L. & Miller, P. W. The relationship between exercise intensity and the sweat lactate excretion rate. J. Physiol. Sci. 60, 103–107 (2010). source
Not checked in the free run.
- Study design and conductnone given
The potentiostat circuit was verified together with the lactate biosensor through an in-vitro amperometric experiment.
“The potentiostat circuit was verified together with the lactate biosensor through an in-vitro amperometric experiment.”
Quote in the paperSubstantial claim - Premisecitation
The working electrode of the lactate biosensor uses prussian blue ink because prussian blue has high selectivity towards hydrogen peroxide, a byproduct of the enzymatic oxidation of lactate.
“The working electrodes were printed using prussian blue ink due to the high selectivity of prussian blue towards hydrogen peroxide, a byproduct of the enzymatic oxidation of lactate3851 (Fig. 1e).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3851]
Not checked in the free run.
- Study design and conductown data
Electrode size, separation and placement parameters were determined through experiments on subjects with different chest sizes.
“Electrode size, separation and placement parameters were determined through a series of experiments involving the placement of Ag/AgCl-based electrocardiogram electrodes of various sizes (1 × 1, 1.5 × 1.5, 1.5 × 1 and 2 × 2 cm2) and separation distances (1–6 cm) on subjects with different chest sizes, and observing the resulting electrocardiogram waveforms.”
Quote in the paperSubstantial claim - Study design and conductown data
The dimensions of the electrodes and inter-electrode distances were optimized based on human trials to acquire a clean ECG signal and lactate response with minimal interference between the two sensors.
“The dimensions of the electrodes and the inter-electrode distances have been optimized based on the human trials to acquire a clean electrocardiogram signal and lactate response with minimal interference between the two sensors.”
Quote in the paperSubstantial claim
- May combine two claims. 10
- Classified as probably not a substantial claim. 5
May combine two claims.
Findingown dataThe lactate biosensor responds linearly to lactate concentrations in the physiological range of 0-28 mM with a sensitivity of 96 nA/mM.
“Figure 2a shows the amperometric response of the lactate biosensor to increasing lactate concentrations in the physiological range of 0–28 mM. It is evidenced from this figure that the biosensor responds linearly to the lactate concentrations in this range with a sensitivity of 96 nA/mM.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAn enzyme-free (LOx-free) amperometric sensor produces a negligible current response under the same experimental conditions, confirming the lactate biosensor's high selectivity for lactate.
“As shown in Fig. 4e, the control biosensor leads to a negligible current response without the presence of LOx, confirming the high selectivity of the lactate biosensor.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataDuring intense cycling, sweat-lactate concentration increases with increasing cycling resistance and correlates with physical exertion and heart rate, after a physiologic time delay.
“As the resistance increases, the sweat-lactate concentration too increases, as illustrated in Fig. 4a–c, showing a correlation between physical exertion, heart rate and, after a physiologic time delay, lactate generation.”
Quote in the paperSubstantial claim- Rests on an assumption: The increasing cycling resistance protocol invokes anaerobic metabolism at similar time scales, augmenting lactate excretion in perspiration.
May combine two claims.
PremisecitationThe chest region has a high sweat rate during physical exertion and can serve as an appropriate location to measure lactate levels in human perspiration.
“The chest region is not only convenient for measurement of electrocardiogram, but also has a high-sweat rate during physical excursion4950, and can thus serve as an appropriate location to also measure lactate levels in human perspiration.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [4950]
Not checked in the free run.
May combine two claims.
PremisecitationLactate concentration in human sweat depends on a person's metabolism and level of exertion and typically ranges from 0 to 25 mM.
“Lactate concentration in human sweat depends on a person's metabolism and level of exertion, and typically ranges from 0 to 25 mM (ref. 52).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [52] Green, J. M., Pritchett, R. C., Crews, T. R., McLester, J. R. & Tucker, D. C. Sweat lactate response between males with high and low aerobic fitness. Eur. J. Appl. Physiol. 91, 1–6 (2004). source
Not checked in the free run.
May combine two claims.
PremisecitationChemical information can be conventionally acquired via clinical labs or point-of-care devices, but such approaches do not support continuous, real-time measurements.
“Chemical information can be conventionally acquired via clinical labs or point-of-care devices323334; unfortunately, such approaches do not support continuous, real-time measurements, therefore limiting their utility to applications where stationary, infrequent tests are sufficient.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites 323334
Not checked in the free run.
May combine two claims.
PremisecitationGao et al. demonstrated a wearable patch that can simultaneously track levels of metabolites and electrolytes in human sweat, but electrophysiology sensors were not included.
“Recently, Gao et al.41 demonstrated a wearable patch that can simultaneously track levels of metabolites and electrolytes in human sweat. However, electrophysiology sensors were not included, and such multimodal sensor fusion is crucial to obtain a more comprehensive knowledge about a wearer's well-being.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [41] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
May combine two claims.
PremisecitationSweat lactate can be used to track an individual's performance and exertion level and is an important biomarker for tissue oxygenation and pressure ischaemia.
“When used in physical-exertion monitoring, electrocardiogram measurements can help monitor heart health and function, while sweat lactate can be used to track an individual's performance and exertion level, and is also an important biomarker for tissue oxygenation and pressure ischaemia424344454647.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites 424344454647
Not checked in the free run.
May combine two claims.
PremisecitationRecent epidermal electronic devices can measure chemicals such as electrolytes and metabolites continuously on the skin or through non-invasive monitoring of other body fluids, but these devices measure only a single parameter at once and are not integrated with other sensing modalities.
“While recent work, including our own, has demonstrated that chemicals such as electrolytes and metabolites can be measured continuously using epidermal electronics on the skin35363738, or through non-invasive monitoring of other body fluids383940, these devices measure only a single parameter at once, and are not integrated with other sensing modalities.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites 35363738
Not checked in the free run.
- cites 383940
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataInformed signed consent was obtained from each individual following a rigorous prescreening procedure.
“A total of 3 healthy male volunteers (recruited in response to follow-up from flyers) with no prior medical history of heart conditions, diabetes or chronic skeletomuscular pain were recruited for participation in the study, and informed, signed consent was obtained from each individual following a rigorous prescreening procedure.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe study was deemed by the IRB as posing no greater than minimal risk to the prescreened subjects.
“The study was deemed by the IRB as posing ‘no greater than minimal risk' to the prescreened subjects who were recruited for the investigation.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductnone givenThe printed circuit board was assembled and tested in-vitro to validate both functionality and performance.
“The printed circuit board was assembled and tested in-vitro to validate both functionality and performance.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe ECG AFE was characterized using a CONTEC MS400 Multi-parameter Patient Simulator.
“The electrocardiogram AFE was characterized using a CONTEC MS400 Multi-parameter Patient Simulator (electrocardiogram simulator).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenAll on-body experiments were performed in strict compliance with IRB guidelines and were approved by the Human Research Protections Program at University of California, San Diego (Project number 130,003).
“All on-body experiments were performed in strict compliance with the guidelines of Institutional Review Boards (IRB) and were approved by the Human Research Protections Program at University of California, San Diego (Project name: Epidermal Electrochemical Sensors and Biosensors. Project number: 130,003).”
Quote in the paperNot a substantial claim
The paper relies on this without support.
The increasing cycling resistance protocol invokes anaerobic metabolism at similar time scales, augmenting lactate excretion in perspiration.
“This ensured that the anaerobic metabolism was invoked at similar time scales, hence augmenting the excretion of lactate in the perspiration in a controlled manner.”
The paper relies on this without support.
The epidermis and muscle tissues over the chest locations do not experience complex 3D strains and remain fairly stable even during intense physical activities.
“In addition, the epidermis and muscle tissues over these locations do not experience complex 3D strains and remain fairly stable even during intense physical activities, making measurements here especially convenient.”
The paper relies on this without support.
The performance of amperometric lactate electrodes is not compromised by reducing their dimension.
“Since the performance of amperometric lactate electrodes is not compromised by reducing their dimension, they were fabricated between the two electrocardiogram electrodes, as shown in Fig. 1d.”
The paper relies on this without support.
The commercial wristband heart rate monitor (BASIS) provides an accurate heart rate reference.
“To validate electrocardiogram data over long time series, even under the presence of experimentally-induced motion, heart rate as extracted from the electrocardiogram data is benchmarked against a commercial wristband heat rate monitor (BASIS) for subjects 1 and 3.”
The paper relies on this without support.
The commercially available 3M Red Dot ECG electrodes provide a valid reference for ECG morphology.
“The ability of the printed electrocardiogram electrodes to record electrocardiogram signals was validated by comparing recordings from the fabricated electrodes with commercially-available 3M Red Dot electrocardiogram electrodes.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Polyurethane-based ion-selective membranes were integrated with textile-based potentiometric sensors to create stretchable diagnostic devices.
“In another study, polyurethane‐based ion‐selective membranes were integrated with textile‐based potentiometric sensors to create stretchable diagnostic devices.35”
From Wearables in Medicine (2018) · cites this paper as [35]
The Methods section (“Printing and functionalization of Chem–Phys hybrid patch”; page unavailable) describes an amperometric lactate sensor and ECG electrodes printed on a flexible polyester substrate, not polyurethane-based ion-selective membranes integrated with textile potentiometric sensors. The Results section (“Hybrid patch design rationale and fabrication”; page unavailable) calls the device flexible, but does not describe it as the stretchable textile device claimed. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredWearable sweat sensors address the limitation of traditional sweat testing by enabling in situ sweat measurements with real-time feedback, creating potential for preventive care, timely diagnosis, and treatment.
“By taking advantage of flexible and hybrid electronics (13–17), wearable sweat sensors address this limitation by enabling in situ sweat measurements with real-time feedback, creating potential for preventive care, timely diagnosis, and treatment (18–34).”
From Regional and correlative sweat analysis using high-throughput microfluidic sensing patches toward decoding sweat (2019) · cites this paper as [19]
The Abstract says conventional clinical-lab and point-of-care testing cannot provide continuous, real-time measurements, and the Results (Epidermal evaluation; page unavailable) demonstrate continuous on-body sweat-lactate monitoring. The paper discusses monitoring and diagnostic opportunities, but does not specifically establish the claimed potential for preventive care, timely diagnosis, or treatment. (LLM, conf=0.83)
Experimental — accuracy not yet measuredNa+ and K+ ion selective electrodes showed a limit of detection of ≈10−5 m.
“Na+ and K+ ion selective electrodes showed a limit of detection of ≈10−5 m.35”
From Wearables in Medicine (2018) · cites this paper as [35]
The Results section on in-vitro characterization reports a detection range and sensitivity for a lactate biosensor, not detection limits for Na+ or K+ ion-selective electrodes. The Methods section on in-vitro studies likewise describes lactate and ECG sensors. Searches for Na+ and K+ electrodes and a 10−5 M detection limit found no passage addressing the claimed value; page numbers were unavailable. (LLM, conf=0.72)
Experimental — accuracy not yet measuredReal-time monitoring showed that the ECG compared to current wearable devices was not affected by concurrent lactate detection.
“Real-time monitoring showed that the ECG compared to current wearable devices was not affected by concurrent lactate detection.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [48]
Lactate levels measured by the biosensor closely approximated the expected sweat-lactate profile for increasing intensity workouts.
“In addition, lactate levels measured by the biosensor closely approximated the expected sweat-lactate profile for increasing intensity workouts.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [48]
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836 · see citing works
Multifunctional wearable devices for diagnosis and therapy of movement disorders
Son, D., Lee, J., Qiao, S., Ghaffari, R., Kim, J., Lee, J. E., Song, C., Kim, S. J., Lee, D. J., Jun, S. W., et al. (2014) · Nature Nanotechnology · doi:10.1038/nnano.2014.38 · Cited by 1,462 · see citing works
Non-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247 · see citing works
Self-citationResistance Exercise in Individuals With and Without Cardiovascular Disease
Pollock, M. L., Franklin, B. A., Balady, G. J., CHAITMAN, B. R., Fleg, J. L., Fletcher, B. J., Limacher, M. C., Piña, I. L., Stein, R. A., Williams, M. A., et al. (2000) · Circulation · doi:10.1161/01.cir.101.7.828 · Cited by 1,054 · see citing works
Electrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862 · see citing works
Self-citationFlexible and Stretchable Electronics for Biointegrated Devices
Kim, D.-H., Ghaffari, R., Lu, N., & Rogers, J. A. (2012) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071811-150018 · Cited by 749 · see citing works
Tattoo-Based Noninvasive Glucose Monitoring: A Proof-of-Concept Study
Bandodkar, A. J., Jia, W., Yardımcı, C., Wang, X., Ramirez, J., & Wang, J. (2015) · Analytical Chemistry · doi:10.1021/ac504300n · Cited by 700 · see citing works
Self-citationWearable Electrochemical Sensors and Biosensors: A Review
Windmiller, J. R., & Wang, J. (2012) · Electroanalysis · doi:10.1002/elan.201200349 · Cited by 699 · see citing works
Self-citationWearable salivary uric acid mouthguard biosensor with integrated wireless electronics
Kim, J., Imani, S., de Araujo, W. R., Warchall, J., Valdés-Ramírez, G., Paixão, T. R. L. C., Mercier, P. P., & Wang, J. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.07.039 · Cited by 678 · see citing works
Self-citationInteraction of physiological mechanisms during exercise.
Wasserman, K., Kessel, A. L. V., & Burton, G. G. (1967) · Journal of Applied Physiology · doi:10.1152/jappl.1967.22.1.71 · Cited by 527 · see citing works
Chemical Sensors and Biosensors
Eggins, B. R. (2002) · Cited by 496 · see citing works
Acute physiological effects of exhaustive whole‐body vibration exercise in man
Rittweger, J., Beller, & Felsenberg (2000) · Clinical Physiology · doi:10.1046/j.1365-2281.2000.00238.x · Cited by 469 · see citing works
Sensor for Measuring Strain in Textile
Mattmann, C., Clemens, F. J., & Tröster, G. (2008) · Sensors · doi:10.3390/s8063719 · Cited by 465 · see citing works
Variations in Regional Sweat Composition in Normal Human Males
Patterson, M. J., Galloway, S. D. R., & Nimmo, M. A. (2000) · Experimental Physiology · doi:10.1111/j.1469-445x.2000.02058.x · Cited by 438 · see citing works
Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry
Tudos, A. J., Besselink, G. A. J., & Schasfoort, R. B. M. (2001) · Lab on a Chip · doi:10.1039/b106958f · Cited by 426 · see citing works
Smart Vest: Wearable multi-parameter remote physiological monitoring system
Pandian, P. S., Mohanavelu, K., Safeer, K. P., Kotresh, T. M., Shakunthala, D. T., Gopal, P., & Padaki, V. C. (2007) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.05.014 · Cited by 420 · see citing works
Chemical Sensors and Biosensors for Medical and Biological Applications
Spichiger‐Keller, U. E. (1998) · Wiley eBooks · doi:10.1002/9783527612284 · Cited by 419 · see citing works
Yarasheski, K. E., Zachwieja, J. J., & Bier, D. M. (1993) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1993.265.2.e210 · Cited by 406 · see citing works
Wireless sensor network based wearable smart shirt for ubiquitous health and activity monitoring
Lee, Y., & Chung, W. (2009) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2009.04.040 · Cited by 383 · see citing works
Tattoo-based potentiometric ion-selective sensors for epidermal pH monitoring
Bandodkar, A. J., Hung, V. W. S., Jia, W., Valdés-Ramírez, G., Windmiller, J. R., Martinez, A. G., Ramírez, J., Chan, G., Kerman, K., & Wang, J. (2013) · The Analyst · doi:10.1039/c2an36422k · Cited by 379 · see citing works
Self-citationPoint-of-care biosensor systems for cancer diagnostics/prognostics
Soper, S. A., Brown, K. C., Ellington, A. D., Frazier, B., Garcia‐Manero, G., Gau, V., Gutman, S. I., Hayes, D. F., Korte, B. J., Landers, J. L., et al. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.01.006 · Cited by 377 · see citing works
Self-citationNon-invasive mouthguard biosensor for continuous salivary monitoring of metabolites
Kim, J., Valdés-Ramírez, G., Bandodkar, A. J., Jia, W., Martinez, A. G., Ramírez, J., Mercier, P., & Wang, J. (2014) · The Analyst · doi:10.1039/c3an02359a · Cited by 367 · see citing works
Self-citationEffects of heat stress on physiological responses and exercise performance in elite cyclists
Tatterson, A., Hahn, A. G., Martin, D. T., & Febbraio, M. A. (2000) · Journal of science and medicine in sport · doi:10.1016/s1440-2440(00)80080-8 · Cited by 314 · see citing works
Amperometric biosensors for clinical and therapeutic drug monitoring: a review
Wang, J. (1999) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/s0731-7085(98)00056-9 · Cited by 294 · see citing works
Self-citationWearable, Redundant Fabric-Based Sensor Arrays for Reconstruction of Body Segment Posture
Lorussi, F., Rocchia, W., Scilingo, E. P., Tognetti, A., & Rossi, D. D. (2004) · IEEE Sensors Journal · doi:10.1109/jsen.2004.837498 · Cited by 261 · see citing works
Strain Sensing Fabric for Hand Posture and Gesture Monitoring
Lorussi, F., Scilingo, E. P., Tesconi, M., Tognetti, A., & DeRossi, D. (2005) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2005.854510 · Cited by 207 · see citing works
Male and female upper body sweat distribution during running measured with technical absorbents
Havenith, G., Fogarty, A. L., Bartlett, R., Smith, C. J. W., & Ventenat, V. (2007) · European Journal of Applied Physiology · doi:10.1007/s00421-007-0636-z · Cited by 195 · see citing works
Lee, S. M., Byeon, H. J., Lee, J. H., Baek, D. H., Lee, K. H., Hong, J. S., & Lee, S.-H. (2014) · Scientific Reports · doi:10.1038/srep06074 · Cited by 181 · see citing works
Kostis, J. B., Moreyra, A. E., Amendo, M. T., Pietro, J. D., Cosgrove, N. M., & Kuo, P. T. (1982) · Circulation · doi:10.1161/01.cir.65.1.141 · Cited by 173 · see citing works
Textile-Based Capacitive Sensors for Respiration Monitoring
Merritt, C. R., Nagle, H. T., & Grant, E. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2008.2010356 · Cited by 154 · see citing works
Textile Electrodes for EEG Recording — A Pilot Study
Löfhede, J., Seoane, F., & Thordstein, M. (2012) · Sensors · doi:10.3390/s121216907 · Cited by 145 · see citing works
The relationship between exercise intensity and the sweat lactate excretion rate
Buono, M. J., Lee, N. V. L., & Miller, P. W. (2010) · The Journal of Physiological Sciences · doi:10.1007/s12576-009-0073-3 · Cited by 128 · see citing works
Non-contact Low Power EEG/ECG Electrode for High Density Wearable Biopotential Sensor Networks
Yu, M., Deiss, S. R., & Cauwenberghs, G. (2009) · International Workshop on Wearable and Implantable Body Sensor Networks (BSN) · doi:10.1109/bsn.2009.52 · Cited by 128 · see citing works
Wireless Batteryless Implantable Blood Pressure Monitoring Microsystem for Small Laboratory Animals
Cong, P., Ko, W. H., & Young, D. J. (2010) · IEEE Sensors Journal · doi:10.1109/jsen.2009.2030982 · Cited by 127 · see citing works
Photonic textiles for pulse oximetry
Rothmaier, M., Selm, B., Spichtig, S., Haensse, D., & Wolf, M. (2008) · Optics Express · doi:10.1364/oe.16.012973 · Cited by 112 · see citing works
Biagi, S., Ghimenti, S., Onor, M., & Bramanti, E. (2012) · Biomedical Chromatography · doi:10.1002/bmc.2713 · Cited by 110 · see citing works
A mobile EEG system with dry electrodes
Gargiulo, G. D., Bifulco, P., Calvo, R. A., Cesarelli, M., Jin, C. T., & André, V. S. (2008) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2008.4696927 · Cited by 98 · see citing works
Textile Technology for the Vital Signs Monitoring in Telemedicine and Extreme Environments
Rienzo, M. D., Meriggi, P., Rizzo, F., Castiglioni, P., Lombardi, C., Ferratini, M., & Parati, G. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2010.2048921 · Cited by 88 · see citing works
Point-of-care temperature and respiration monitoring sensors for smart fabric applications
Jung, S., Ji, T., & Varadan, V. K. (2006) · Smart Materials and Structures · doi:10.1088/0964-1726/15/6/042 · Cited by 73 · see citing works
Fabrication of a wearable fabric tactile sensor produced by artificial hollow fiber
Hasegawa, Y., Shikida, M., Ogura, D., Suzuki, Y., & Sato, K. (2008) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/18/8/085014 · Cited by 72 · see citing works
Revisited: The Conception of Lability of Metal Complexes
van Leeuwen, H. P. (2001) · Electroanalysis · doi:10.1002/1521-4109(200106)13:10 · Cited by 72 · see citing works
Sweat lactate response between males with high and low aerobic fitness
Green, J. M., Pritchett, R. C., Crews, T. R., McLester, J. R., & Tucker, D. C. (2004) · European Journal of Applied Physiology · doi:10.1007/s00421-003-0968-2 · Cited by 70 · see citing works
Wearable Respiratory Rate Monitoring using Piezo-resistive Fabric Sensor
Jeong, J. W., Jang, Y., Lee, I., Shin, S.-S., & Kim, S. (2009) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-03904-1_78 · Cited by 59 · see citing works
Sweat lactate in exercising children and adolescents of varying physical maturity
Falk, B., Bar‐Or, O., MacDOUGALL, J. D., McGillis, L. H., Calvert, R., & Meyer, F. (1991) · Journal of Applied Physiology · doi:10.1152/jappl.1991.71.5.1735 · Cited by 58 · see citing works
Electrocardiographic Electrode Misplacement, Misconnection, and Artifact
Harrigan, R. A., Chan, T. C., & Brady, W. J. (2012) · Journal of Emergency Medicine · doi:10.1016/j.jemermed.2012.02.024 · Cited by 50 · see citing works
Muehlsteff, J., Aubert, X. L., & Morren, G. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4649501 · Cited by 48 · see citing works
Sweat analysis following pressure ischaemia in a group of debilitated subjects.
Polliack, A. A., Taylor, R. P., & Bader, D. L. (1997) · PubMed · Cited by 41 · see citing works
Flexible thick-film glucose biosensor: Influence of mechanical bending on the performance
Chuang, M., Yang, Y.-L., Tseng, T.-F., Chou, T., Lou, S.-L., & Wang, J. (2009) · Talanta · doi:10.1016/j.talanta.2009.11.029 · Cited by 37 · see citing works
Self-citationSecretion of eccrine sweat glands during exercise.
Pilardeau, P., Vaysse, J., Garnier, M., Joublin, M., & Valeri, L. (1979) · British Journal of Sports Medicine · doi:10.1136/bjsm.13.3.118 · Cited by 32 · see citing works
Chiarugi, F., Karatzanis, I., Zacharioudakis, G., Meriggi, P., Rizzo, F., Stratakis, M., Louloudakis, S., Biniaris, C. G., Valentini, M., Rienzo, M. D., et al. (2008) · Computers in cardiology · doi:10.1109/cic.2008.4749188 · Cited by 21 · see citing works
Effect of different work-loads on sweat production and composition in man.
Pilardeau, P., Lavie, F., Vaysse, J., Garnier, M., Harichaux, P., Jn, M., & Chalumeau, M. T. (1988) · PubMed · Cited by 16 · see citing works
Thermal analysis of ultrathin, compliant sensors for characterization of the human skin
Bian, Z., Song, J., Webb, R. C., Bonifas, A. P., Rogers, J. A., & Huang, Y. Y. (2013) · RSC Advances · doi:10.1039/c3ra45277h · Cited by 14 · see citing works
Integration of textile-based sensors and Shimmer for breathing rate and volume measurement
Rovira-Borràs, C., Coyle, S. M., Diamont, D., Stroiescu, F., Daly, K., & Corcoran, B. (2011) · International Conference on Pervasive Computing Technologies for Healthcare · doi:10.4108/icst.pervasivehealth.2011.246104 · Cited by 14 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373 · see citing works
Self-citationYang, J. C., Mun, J., Kwon, S. Y., Park, S., Bao, Z., & Park, S. (2019) · Advanced Materials · doi:10.1002/adma.201904765 · Cited by 1,641 · see citing works
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434 · see citing works
Self-citationBio-Integrated Wearable Systems: A Comprehensive Review
Ray, T. R., Choi, J., Bandodkar, A. J., Krishnan, S., Gutruf, P., Tian, L., Ghaffari, R., & Rogers, J. A. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00573 · Cited by 1,350 · see citing works
Self-citationWearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315 · see citing works
Advanced Carbon for Flexible and Wearable Electronics
Wang, C., Xia, K., Wang, H., Liang, X., Yin, Z., & Zhang, Y. (2018) · Advanced Materials · doi:10.1002/adma.201801072 · Cited by 1,304 · see citing works
A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat
Yang, Y., Song, Y., Bo, X., Min, J., Pak, O. S., Zhu, L., Wang, M., Tu, J., Kogan, A., Zhang, H., et al. (2020) · Nature Biotechnology · doi:10.1038/s41587-019-0321-x · Cited by 1,270 · see citing works
Lee, H., Song, C., Hong, Y. S., Kim, M., Cho, H. R., Kang, T., Shin, K., Choi, S. H., Hyeon, T., & Kim, D.-H. (2017) · Science Advances · doi:10.1126/sciadv.1601314 · Cited by 1,118 · see citing works
Disposable Sensors in Diagnostics, Food, and Environmental Monitoring
Dincer, C., Bruch, R. C., Costa‐Rama, E., Fernández-Abedul, M. T., Merkoçi, A., Manz, A., Urban, G. A., & Güder, F. (2019) · Advanced Materials · doi:10.1002/adma.201806739 · Cited by 972 · see citing works
Maduraiveeran, G., Sasidharan, M., & Ganesan, V. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2017.12.031 · Cited by 919 · see citing works
Emaminejad, S., Gao, W., Wu, E., Davies, Z. A., Yin Yin Nyein, H., Challa, S., Ryan, S. P., Fahad, H. M., Chen, K., Shahpar, Z., et al. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1701740114 · Cited by 839 · see citing works
Enzyme‐Based Glucose Sensor: From Invasive to Wearable Device
Lee, H., Hong, Y. J., Baik, S., Hyeon, T., & Kim, D. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201701150 · Cited by 778 · see citing works
Bandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav3294 · Cited by 761 · see citing works
Self-citationRecent advances in wearable tactile sensors: Materials, sensing mechanisms, and device performance
Yang, T., Xie, D., Li, Z., & Zhu, H. (2017) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2017.02.001 · Cited by 755 · see citing works
Skin-Interfaced Wearable Sweat Sensors for Precision Medicine
Min, J., Tu, J., Xu, C., Lukas, H., Shin, S., Yang, Y., Solomon, S. A., Mukasa, D., & Gao, W. F. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00823 · Cited by 728 · see citing works
Self‐Powered Real‐Time Arterial Pulse Monitoring Using Ultrathin Epidermal Piezoelectric Sensors
Park, D. Y., Joe, D. J., Kim, D. H., Park, H., Han, J. H., Jeong, C. K., Park, H., Park, J. G., Joung, B., & Lee, K. J. (2017) · Advanced Materials · doi:10.1002/adma.201702308 · Cited by 695 · see citing works
A wearable biosensing system with in-sensor adaptive machine learning for hand gesture recognition
Moin, A., Zhou, A., Rahimi, A., Menon, A., Benatti, S., Alexandrov, G., Tamakloe, S., Ting, J., Yamamoto, N. A. D., Khan, Y., et al. (2020) · Nature Electronics · doi:10.1038/s41928-020-00510-8 · Cited by 685 · see citing works
An epidermal patch for the simultaneous monitoring of haemodynamic and metabolic biomarkers
Sempionatto, J. R., Lin, M., Yin, L., De la paz, E., Pei, K., Sonsa-ard, T., de Loyola Silva, A. N., Khorshed, A. A., Zhang, F., Tostado, N., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00685-1 · Cited by 658 · see citing works
Self-citationTehrani, F., Teymourian, H., Wuerstle, B. L., Kavner, J., Patel, R., Furmidge, A., Aghavali, R., Hosseini-Toudeshki, H., Brown, C., Zhang, F., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00887-1 · Cited by 651 · see citing works
Self-citationFunctional Fibers and Fabrics for Soft Robotics, Wearables, and Human–Robot Interface
Xiong, J., Chen, J., & Lee, P. S. (2020) · Advanced Materials · doi:10.1002/adma.202002640 · Cited by 640 · see citing works
Flexible Hybrid Electronics for Digital Healthcare
Ma, Y., Zhang, Y., Cai, S., Han, Z., Liu, X., Wang, F., Cao, Y., Wang, Z., Li, H., Chen, Y., et al. (2020) · Advanced Materials · doi:10.1002/adma.201902062 · Cited by 620 · see citing works
Wearable chemical sensors for biomarker discovery in the omics era
Sempionatto, J. R., Lasalde-Ramírez, J. A., Mahato, K., Wang, J., & Gao, W. (2022) · Nature Reviews Chemistry · doi:10.1038/s41570-022-00439-w · Cited by 613 · see citing works
Self-citationNoninvasive Alcohol Monitoring Using a Wearable Tattoo-Based Iontophoretic-Biosensing System
Kim, J., Jeerapan, I., Imani, S., Cho, T. N., Bandodkar, A., Cinti, S., Mercier, P. P., & Wang, J. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00356 · Cited by 599 · see citing works
Self-citationYetisen, A. K., Martinez‐Hurtado, J. L., Ünal, B., Khademhosseini, A., & Butt, H. (2018) · Advanced Materials · doi:10.1002/adma.201706910 · Cited by 569 · see citing works
Integrated textile sensor patch for real-time and multiplex sweat analysis
He, W., Wang, C., Wang, H., Jian, M., Lu, W., Liang, X., Zhang, X., Yang, F., & Zhang, Y. (2019) · Science Advances · doi:10.1126/sciadv.aax0649 · Cited by 564 · see citing works
Wearable sensors for monitoring the physiological and biochemical profile of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0150-9 · Cited by 537 · see citing works
Digital health for aging populations
Chen, C., Ding, S., & Wang, J. (2023) · Nature Medicine · doi:10.1038/s41591-023-02391-8 · Cited by 525 · see citing works
Self-citationA Survey on Wearable Technology: History, State-of-the-Art and Current Challenges
Ometov, A., Shubina, V., Klus, L., Skibińska, J., Saafi, S., Pascacio, P., Flueratoru, L., Quezada-Gaibor, D., Chukhno, N., Chukhno, O., et al. (2021) · Computer Networks · doi:10.1016/j.comnet.2021.108074 · Cited by 523 · see citing works
Flexible Hybrid Sensors for Health Monitoring: Materials and Mechanisms to Render Wearability
Gao, Y., Yu, L., Yeo, J. C., & Lim, C. T. (2019) · Advanced Materials · doi:10.1002/adma.201902133 · Cited by 508 · see citing works
Shrivastava, S., Trung, T. Q., & Lee, N. (2020) · Chemical Society Reviews · doi:10.1039/c9cs00319c · Cited by 495 · see citing works
Toward a new generation of smart skins
Someya, T., & Amagai, M. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0079-1 · Cited by 485 · see citing works
Disruptive, Soft, Wearable Sensors
Ling, Y., An, T., Yap, L. W., Zhu, B., Gong, S. P., & Cheng, W. (2019) · Advanced Materials · doi:10.1002/adma.201904664 · Cited by 483 · see citing works
Advanced Materials for Printed Wearable Electrochemical Devices: A Review
Kim, J., Kumar, R., Bandodkar, A. J., & Wang, J. (2017) · Advanced Electronic Materials · doi:10.1002/aelm.201600260 · Cited by 452 · see citing works
Self-citationA MXene‐Based Wearable Biosensor System for High‐Performance In Vitro Perspiration Analysis
Lei, Y., Zhao, W., Zhang, Y., Jiang, Q. M., He, J., Baeumner, A. J., Wolfbeis, O. S., Wang, Z. L., Salama, K. N., & Alshareef, H. N. (2019) · Small · doi:10.1002/smll.201901190 · Cited by 443 · see citing works
Skin-Attachable, Stretchable Electrochemical Sweat Sensor for Glucose and pH Detection
Oh, S. Y., Hong, S. Y., Jeong, Y. R., Yun, J., Park, H., Jin, S. W., Lee, G., Oh, J. H., Lee, H.-C., Lee, S., et al. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b03342 · Cited by 442 · see citing works
Han, L., Yan, L., Wang, M., Wang, K., Fang, L., Zhou, J., Fang, J., Ren, F., & Lu, X. (2018) · Chemistry of Materials · doi:10.1021/acs.chemmater.8b01446 · Cited by 438 · see citing works
Hybrid multimodal wearable sensors for comprehensive health monitoring
Mahato, K., Saha, T., Ding, S., Sandhu, S. S., Chang, A., & Wang, J. (2024) · Nature Electronics · doi:10.1038/s41928-024-01247-4 · Cited by 435 · see citing works
Self-citationSkin-interfaced systems for sweat collection and analytics
Choi, J., Ghaffari, R., Baker, L. B., & Rogers, J. A. (2018) · Science Advances · doi:10.1126/sciadv.aar3921 · Cited by 433 · see citing works
Dong, K., Deng, J., Zi, Y., Wang, Y., Xu, C., Zou, H., Ding, W., Dai, Y., Gu, B., Sun, B., et al. (2017) · Advanced Materials · doi:10.1002/adma.201702648 · Cited by 433 · see citing works
Wu, H., Yang, G., Zhu, K., Liu, S., Guo, W., Jiang, Z., & Li, Z. (2021) · Advanced Science · doi:10.1002/advs.202001938 · Cited by 395 · see citing works
Liu, Y., Wang, H., Zhao, W., Zhang, M., Qin, H., & Xie, Y. (2018) · Sensors · doi:10.3390/s18020645 · Cited by 392 · see citing works
Reviews of wearable healthcare systems: Materials, devices and system integration
Lou, Z., Wang, L., Jiang, K., Wei, Z., & Shen, G. (2019) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2019.100523 · Cited by 389 · see citing works
Nyein, H. Y. Y., Bariya, M., Kivimäki, L., Uusitalo, S., Liaw, T. S., Jansson, E., Ahn, C. H., Hangasky, J. A., Zhao, J., Lin, Y., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw9906 · Cited by 388 · see citing works
Wearable Sensors for Biochemical Sweat Analysis
Bandodkar, A. J., Jeang, W. J., Ghaffari, R., & Rogers, J. A. (2019) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061318-114910 · Cited by 387 · see citing works
Self-citationNakata, S., Arie, T., Akita, S., & Takei, K. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00047 · Cited by 379 · see citing works
Wearable Biosensors: An Alternative and Practical Approach in Healthcare and Disease Monitoring
Sharma, A., Badea, M., Tiwari, S., & Marty, J. (2021) · Molecules · doi:10.3390/molecules26030748 · Cited by 373 · see citing works
Choi, J., Bandodkar, A. J., Reeder, J. T., Ray, T. R., Turnquist, A., Kim, S. B., Nyberg, N., Hourlier-Fargette, A., Model, J. B., Aranyosi, A. J., et al. (2019) · ACS Sensors · doi:10.1021/acssensors.8b01218 · Cited by 365 · see citing works
Self-citationLim, C., Hong, Y. J., Jung, J., Shin, Y., Sunwoo, S.-H., Baik, S., Park, O. K., Choi, S. H., Hyeon, T., Kim, J. H., et al. (2021) · Science Advances · doi:10.1126/sciadv.abd3716 · Cited by 365 · see citing works
From Diagnosis to Treatment: Recent Advances in Patient-Friendly Biosensors and Implantable Devices
Li, P., Lee, G., Kim, S. Y., Kwon, S. Y., Kim, H., & Park, S. (2021) · ACS Nano · doi:10.1021/acsnano.0c06688 · Cited by 362 · see citing works
Open-Source Potentiostat for Wireless Electrochemical Detection with Smartphones
Ainla, A., Mousavi, M. P. S., Tsaloglou, M., Redston, J., Bell, J. G., Fernández-Abedul, M. T., & Whitesides, G. M. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b00850 · Cited by 359 · see citing works
Vital-sign monitoring and spatial tracking of multiple people using a contactless radar-based sensor
Mercuri, M., Lorato, I. R., Liu, Y., Wieringa, F. P., Hoof, C. A. V., & Torfs, T. (2019) · Nature Electronics · doi:10.1038/s41928-019-0258-6 · Cited by 357 · see citing works
Weaving Sensing Fibers into Electrochemical Fabric for Real‐Time Health Monitoring
Wang, L., Wang, L., Zhang, Y., Pan, J., Li, S., Sun, X., Zhang, B., & Peng, H. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201804456 · Cited by 353 · see citing works
Wearable and flexible sensors for user-interactive health-monitoring devices
Ha, M., Lim, S., & Ko, H. (2018) · Journal of Materials Chemistry B · doi:10.1039/c8tb01063c · Cited by 352 · see citing works
Flexible Electrochemical Bioelectronics: The Rise of In Situ Bioanalysis
Yu, Y., Nyein, H. Y. Y., Gao, W. F., & Javey, A. (2019) · Advanced Materials · doi:10.1002/adma.201902083 · Cited by 347 · see citing works
Wang, Y., Zhao, C., Wang, J., Luo, X., Xie, L., Zhan, S., Kim, J., Wang, X., Liu, X., & Ying, Y. B. (2021) · Science Advances · doi:10.1126/sciadv.abe4553 · Cited by 345 · see citing works
Self-citationEnergy autonomous electronic skin
García Núñez, C., Manjakkal, L., & Dahiya, R. (2019) · npj Flexible Electronics · doi:10.1038/s41528-018-0045-x · Cited by 332 · see citing works
Wireless battery-free body sensor networks using near-field-enabled clothing
Lin, R., Kim, H.-J., Achavananthadith, S., Kurt, S. A., Tan, S. C. C., Yao, H., Tee, B. C. K., Lee, J. K. W., & Ho, J. S. (2020) · Nature Communications · doi:10.1038/s41467-020-14311-2 · Cited by 331 · see citing works
Zhao, Y., Zhai, Q., Dong, D., An, T., Gong, S. P., Shi, Q., & Cheng, W. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b00152 · Cited by 330 · see citing works
Stretchable wireless system for sweat pH monitoring
Dang, W., Manjakkal, L., Navaraj, W. T., Lorenzelli, L., Vinciguerra, V., & Dahiya, R. (2018) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2018.02.025 · Cited by 326 · see citing works
Recent Progress of Self‐Powered Sensing Systems for Wearable Electronics
Lou, Z., Li, L., Wang, L., & Shen, G. (2017) · Small · doi:10.1002/smll.201701791 · Cited by 323 · see citing works
Stable and Biocompatible Carbon Nanotube Ink Mediated by Silk Protein for Printed Electronics
Liang, X., Li, H., Dou, J., Wang, Q., He, W., Wang, C., Li, D., Lin, J., & Zhang, Y. (2020) · Advanced Materials · doi:10.1002/adma.202000165 · Cited by 321 · see citing works
Khan, Y., Garg, M., Gui, Q., Schadt, M. J., Gaikwad, A. M., Han, D., Yamamoto, N. A. D., Hart, P., Welte, R., Wilson, W., et al. (2016) · Advanced Functional Materials · doi:10.1002/adfm.201603763 · Cited by 320 · see citing works
Advanced Materials for Health Monitoring with Skin‐Based Wearable Devices
Jin, H., Abu‐Raya, Y. S., & Haick, H. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700024 · Cited by 317 · see citing works
Choi, J., Kang, D., Han, S., Kim, S. B., & Rogers, J. A. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201601355 · Cited by 317 · see citing works
Recent Progress in Inkjet‐Printed Thin‐Film Transistors
Chung, S., Cho, K., & Lee, T. (2019) · Advanced Science · doi:10.1002/advs.201801445 · Cited by 316 · see citing works
Applied body-fluid analysis by wearable devices
Brasier, N., Wang, J., Gao, W. F., Sempionatto, J. R., Dincer, C., Ates, H. C., Güder, F., Olenik, S., Schauwecker, I., Schaffarczyk, D., et al. (2024) · Nature · doi:10.1038/s41586-024-08249-4 · Cited by 316 · see citing works
Self-citationEyeglasses based wireless electrolyte and metabolite sensor platform
Sempionatto, J. R., Nakagawa, T., Pavinatto, A., Mensah, S. T., Imani, S., Mercier, P., & Wang, J. (2017) · Lab on a Chip · doi:10.1039/c7lc00192d · Cited by 311 · see citing works
Self-citationMethylxanthine Drug Monitoring with Wearable Sweat Sensors
Tai, L., Gao, W., Chao, M., Bariya, M., Ngo, Q. P., Shahpar, Z., Nyein, H. Y. Y., Park, H., Sun, J., Jung, Y., et al. (2018) · Advanced Materials · doi:10.1002/adma.201707442 · Cited by 310 · see citing works
Reeder, J. T., Choi, J., Xue, Y., Gutruf, P., Hanson, J., Liu, M., Ray, T., Bandodkar, A. J., Avila, R., Xia, W., et al. (2019) · Science Advances · doi:10.1126/sciadv.aau6356 · Cited by 308 · see citing works
Self-citationMayer, M., & Baeumner, A. J. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00719 · Cited by 306 · see citing works
Wearable and Implantable Sensors for Biomedical Applications
Koydemir, H. C., & Özcan, A. (2018) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061417-125956 · Cited by 296 · see citing works
Wearable Bioelectronics: Enzyme-Based Body-Worn Electronic Devices
Kim, J., Jeerapan, I., Sempionatto, J. R., Barfidokht, A., Mishra, R. K., Campbell, A. S., Hubble, L. J., & Wang, J. (2018) · Accounts of Chemical Research · doi:10.1021/acs.accounts.8b00451 · Cited by 290 · see citing works
Self-citationYamamoto, Y., Yamamoto, D., Takada, M., Naito, H., Arie, T., Akita, S., & Takei, K. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700495 · Cited by 283 · see citing works
A Fully Integrated and Self-Powered Smartwatch for Continuous Sweat Glucose Monitoring
Zhao, J., Lin, Y., Wu, J., Nyein, H. Y. Y., Bariya, M., Tai, L., Chao, M., Ji, W., Zhang, G. B., Fan, Z., et al. (2019) · ACS Sensors · doi:10.1021/acssensors.9b00891 · Cited by 283 · see citing works
Ning, C., Cheng, R., Jiang, Y., Sheng, F., Yi, J., Shen, S., Zhang, Y., Peng, X., Dong, K., & Wang, Z. L. (2022) · ACS Nano · doi:10.1021/acsnano.1c09792 · Cited by 283 · see citing works
Reshaping healthcare with wearable biosensors
Smith, A. A., Li, R., & Tse, Z. T. H. (2023) · Scientific Reports · doi:10.1038/s41598-022-26951-z · Cited by 283 · see citing works
Wearable electrochemical biosensors in North America
Min, J., Sempionatto, J. R., Teymourian, H., Wang, J., & Gao, W. F. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112750 · Cited by 280 · see citing works
Self-citationWearable-Based Affect Recognition—A Review
Schmidt, P., Reiss, A., Dürichen, R., & Laerhoven, K. V. (2019) · Sensors · doi:10.3390/s19194079 · Cited by 267 · see citing works
Smart plant-wearable biosensor for in-situ pesticide analysis
Zhao, F., He, J., Li, X., Bai, Y., Ying, Y. B., & Ping, J. F. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112636 · Cited by 263 · see citing works
Kumar, R., Shin, J., Yin, L., You, J., Meng, Y. S., & Wang, J. (2017) · Advanced Energy Materials · doi:10.1002/aenm.201602096 · Cited by 261 · see citing works
Self-citationTissue adhesive hydrogel bioelectronics
Li, S., Cong, Y., & Fu, J. (2021) · Journal of Materials Chemistry B · doi:10.1039/d1tb00523e · Cited by 260 · see citing works
A wearable pH sensor with high sensitivity based on a flexible charge-coupled device
Nakata, S., Shiomi, M., Fujita, Y., Arie, T., Akita, S., & Takei, K. (2018) · Nature Electronics · doi:10.1038/s41928-018-0162-5 · Cited by 256 · see citing works
Khan, S., Ali, S., & Bermak, A. (2019) · Sensors · doi:10.3390/s19051230 · Cited by 256 · see citing works
Electronic textiles: New age of wearable technology for healthcare and fitness solutions
Meena, J. S., Choi, S. B., Jung, S., & Kim, J. (2023) · Materials Today Bio · doi:10.1016/j.mtbio.2023.100565 · Cited by 253 · see citing works
Physical sensors for skin‐inspired electronics
Li, S., Zhang, Y., Wang, Y., Xia, K., Yin, Z., Wang, H., Zhang, M., Liang, X., Lü, H., Zhu, M., et al. (2019) · InfoMat · doi:10.1002/inf2.12060 · Cited by 245 · see citing works
Wearable and Mobile Sensors for Personalized Nutrition
Sempionatto, J. R., Montiel, V. R.-V., Vargas, E., Teymourian, H., & Wang, J. (2021) · ACS Sensors · doi:10.1021/acssensors.1c00553 · Cited by 237 · see citing works
Self-citationRecent advances and perspectives in sweat based wearable electrochemical sensors
Mohan, A. M. V., Rajendran, V., Mishra, R. K., & Mathiyarasu, J. (2020) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2020.116024 · Cited by 234 · see citing works
Self-citationKwon, Y.-T., Kim, Y.-S., Kwon, S., Mahmood, M., Lim, H.-R., Park, S.-W., Kang, S.-O., Choi, J. J., Herbert, R., Jang, Y. C., et al. (2020) · Nature Communications · doi:10.1038/s41467-020-17288-0 · Cited by 233 · see citing works
Pacifier Biosensor: Toward Noninvasive Saliva Biomarker Monitoring
García-Carmona, L., Martín, A., Sempionatto, J. R., Moreto, J. R., González, M. C., Wang, J., & Escarpa, A. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b03379 · Cited by 230 · see citing works
Self-citationHong, Y. J., Lee, H., Kim, J., Lee, M., Choi, H. J., Hyeon, T., & Kim, D. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201805754 · Cited by 229 · see citing works
Brown, M. S., Ashley, B. K., & Koh, A. (2018) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2018.00047 · Cited by 226 · see citing works
Diving into Sweat: Advances, Challenges, and Future Directions in Wearable Sweat Sensing
Childs, A., Mayol, B., Lasalde‐Ramírez, J. A., Song, Y., Sempionatto, J. R., & Gao, W. F. (2024) · ACS Nano · doi:10.1021/acsnano.4c10344 · Cited by 226 · see citing works
Flexible potentiometric pH sensors for wearable systems
Manjakkal, L., Dervin, S., & Dahiya, R. (2020) · RSC Advances · doi:10.1039/d0ra00016g · Cited by 224 · see citing works
Herbert, R., Kim, J.-H., Kim, Y., Lee, H., & Yeo, W.-H. (2018) · Materials · doi:10.3390/ma11020187 · Cited by 220 · see citing works
A highly integrated sensing paper for wearable electrochemical sweat analysis
Li, M., Wang, L., Liu, R., Li, J., Zhang, Q., Shi, G., Li, Y., Hou, C., & Wang, H. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112828 · Cited by 220 · see citing works
Recent Advancements in Development of Wearable Gas Sensors
Bag, A., & Lee, N. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202000883 · Cited by 219 · see citing works
Inorganic semiconducting materials for flexible and stretchable electronics
Yu, K. J., Yan, Z., Han, M., & Rogers, J. A. (2017) · npj Flexible Electronics · doi:10.1038/s41528-017-0003-z · Cited by 218 · see citing works
Bae, C. W., Toi, P. T., Kim, B. Y., Lee, W. I., Lee, H. B., Hanif, A., Lee, E. H., & Lee, N. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.9b00848 · Cited by 217 · see citing works
Advanced Wearable Microfluidic Sensors for Healthcare Monitoring
Li, S., Ma, Z., Cao, Z., Pan, L., & Shi, Y. (2019) · Small · doi:10.1002/smll.201903822 · Cited by 214 · see citing works
A fully inkjet-printed disposable glucose sensor on paper
Bihar, E., Wustoni, S., Pappa, A., Salama, K. N., Baran, D., & Inal, S. (2018) · npj Flexible Electronics · doi:10.1038/s41528-018-0044-y · Cited by 209 · see citing works
- Authors
- Imani, S., Bandodkar, A. J., Mohan, A. M. V., Kumar, R., Yu, S., Wang, J., & Mercier, P. P.
- Affiliations
- University of California San Diego (US)
- Corresponding author
- Wang, Joseph
- Mercier, Patrick P.
- Venue
- Nature Communications
- Published
- 2016-05-23
- Volume
- 7
- Issue
- 1
- Pages
- 11650-11650
- ISSN
- 2041-1723 (ISSN-L)
- PMC
- PMC4879240
- OpenAlex
- W2401823338
- Semantic Scholar
- ca05604884f0b3cf597f80f1d61b2fcdd0f4707f
- Cited by
- 861
- References
- 54
- Citations by year
- 2026: 49 · 2025: 73 · 2024: 79 · 2023: 77 · 2022: 102 · 2021: 115 · 2020: 100 · 2019: 110 · 2018: 92 · 2017: 54 · 2016: 9
- FWCI
- 42.52
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Nature Communications (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- Funders
- Arnold and Mabel Beckman Foundation
- National Institutes of Health
- Samsung
- National Institute of Biomedical Imaging and Bioengineering (R21EB019698)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- wearable biosensors, lactate biosensor, electrocardiogram, amperometric biosensor, flexible electronics, fitness tracking, health monitoring, biomarkers
- MeSH terms
- Electrocardiography, Exercise Test, Humans, Male, Monitoring, Physiologic, Sweat, Biosensing Techniques, Biomarkers, Lactic Acid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review