Millimetre-scale magnetocardiography of living rats with thoracotomy
Open accessThe DOI matches its Crossref record; 67 of the 68 references it cites were found in a registry or an index.
Abstract
Abstract Magnetocardiography is a contactless imaging modality for electric current propagation in the cardiovascular system. Although conventional sensors provide sufficiently high sensitivity, their spatial resolution is limited to a centimetre-scale, which is inadequate for revealing the intra-cardiac electrodynamics such as rotational waves associated with ventricular arrhythmias. Here, we demonstrate invasive magnetocardiography of living rats at a millimetre-scale using a quantum sensor based on nitrogen-vacancy centres in diamond. The acquired magnetic images indicate that the cardiac signal source is well explained by vertically distributed current dipoles, pointing from the right atrium base via the Purkinje fibre bundle to the left ventricular apex. We also find that this observation is consistent with and complementary to an alternative picture of electric current density distribution calculated with a stream function method. Our technique will enable the study of the origin and progression of various cardiac arrhythmias, including flutter, fibrillation, and tachycardia.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures
![Fig. 1 Magnetocardiography based on a solid-state quantum sensor. a Schematic of the rat magnetocardiography (MCG) setup. A living rat ' s heart remains approximately one millimetre below a diamond chip containing an ensemble of nitrogen-vacancy (NV) centres. The rat is scanned automatically along the XYaxes for magnetic field mapping and manually along the Z-axis for height adjustment. An electrocardiography (ECG) signal is monitored through ECG profilers concurrently with the MCG. The NV centres are excited by a 2.0 W green laser light. This excitation entails spin-state-dependent fluorescence collected by an aspheric condenser lens. b NVcentre energy level diagram. The mS = ±1 ground states are split by a bias magnetic field and mixed by microwaves resonant with the NV transition frequencies. Each of the ground states are further split by hyperfine interactions with the host 14 N nuclear spin. c Enlarged view of the heart and diamond. Electric currents flowing through the heart generate a circulating field (blue arrows). The NV centres (red arrows) along the [111] orientation are sensitive to the Z-component of the magnetic field. d Magnetometry principle. The time-varying cardiac magnetic field (blue), which shifts the NV transition frequency, is converted to a change in the lock-in-demodulated fluorescence signal (red). Five peaks are observed in the lock-in optically detected magnetic resonance (ODMR) spectrum because three hyperfine transition frequencies are excited with three-tone microwaves. e Magnetic field sensitivity across the rat ' s heart signal frequency band of DC ~200 Hz. The black dashed line indicates 140 pT Hz -1/2 .](https://figures.cochat.ai/paper-figures/e2080dab110a42918fe784cbd9e371c2/millimetre-scale-magnetocardiography-of-living-rats-with-thoracotomy-figure-1.png)



Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Since the best reported diamond magnetometers’ sensitivity is between 12 and 50 pT/Hz1/2, they are suitable for experimental recording of magnetic fields from isolated nerve or muscle neuronal action potentials isolated nerve or muscle tissues (294) or even for invasive, close-proximity, high-resolution MCG of living rats (208), but do not have sufficient sensitivity to compete with SQUIDs or OPMs”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [208]
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“In fact, when the ECG and cardiac enzyme patterns cannot yet be diagnostic, the higher sensitivity of MCG in detecting myocardial ischemia at an early stage is only one of the advantages arising by the more comprehensive biomagnetic electrophysiological assessment, which has the potential to non-invasively detect electrogenic phenomena at cellular and even subcellular levels (86, 208).”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [208]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Vos, T., Lim, S. S., Abbafati, C., Abbas, K. M., Abbasi, M., Abbasifard, M., Abbasi-Kangevari, M., Abbastabar, H., Abd-Allah, F., Abdelalim, A., et al. (2020) · The Lancet · doi:10.1016/s0140-6736(20)30925-9 · Cited by 19,358
Degen, C. L., Reinhard, F., & Cappellaro, P. (2017) · Reviews of Modern Physics · doi:10.1103/revmodphys.89.035002 · Cited by 4,045
High-sensitivity diamond magnetometer with nanoscale resolution
Taylor, J. M., Cappellaro, P., Childress, L., Jiang, L., Budker, D., Hemmer, P. R., Yacoby, A., Walsworth, R., & Lukin, M. D. (2008) · Nature Physics · doi:10.1038/nphys1075 · Cited by 1,959
Geddes, L. A., & Baker, L. E. (1967) · Medical & Biological Engineering · doi:10.1007/bf02474537 · Cited by 1,610
The surgical treatment of atrial fibrillation
Cox, J. L., Schuessler, R. B., D’Agostino, H. J., Stone, C. M., Chang, B.-C., Cain, M. E., Corr, P. B., & Boineau, J. P. (1991) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/s0022-5223(19)36684-x · Cited by 1,464
Spatial and temporal organization during cardiac fibrillation
Gray, R. A., Pertsov, A. M., & Jalife, J. (1998) · Nature · doi:10.1038/32164 · Cited by 958
Artificial intelligence-enhanced electrocardiography in cardiovascular disease management
Siontis, K. C., Noseworthy, P. A., Attia, Z. I., & Friedman, P. A. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-00503-2 · Cited by 914
Gepstein, L., Hayam, G., & Ben-Haim, S. A. (1997) · Circulation · doi:10.1161/01.cir.95.6.1611 · Cited by 759
Temperature Dependence of the Nitrogen-Vacancy Magnetic Resonance in Diamond
Acosta, V. M., Bauch, E., Ledbetter, M. P., Waxman, A., Bouchard, L.-S., & Budker, D. (2010) · Physical Review Letters · doi:10.1103/physrevlett.104.070801 · Cited by 735
Optical magnetic imaging of living cells
Le Sage, D., Arai, K., Glenn, D. R., DeVience, S. J., Pham, L. M., Rahn-Lee, L., Lukin, M. D., Yacoby, A., Komeili, A., & Walsworth, R. L. (2013) · Nature · doi:10.1038/nature12072 · Cited by 730
Self-citationOptical magnetic detection of single-neuron action potentials using quantum defects in diamond
Barry, J. F., Turner, M. J., Schloss, J. M., Glenn, D. R., Song, Y., Lukin, M. D., Park, H., & Walsworth, R. L. (2016) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1601513113 · Cited by 637
Nuclear magnetic resonance detection and spectroscopy of single proteins using quantum logic
Lovchinsky, I., Sushkov, A. O., Urbach, E., de Leon, N. P., Choi, S., De Greve, K., Evans, R., Gertner, R., Bersin, E., Müller, C., et al. (2016) · Science · doi:10.1126/science.aad8022 · Cited by 536
Review on solving the forward problem in EEG source analysis
Hallez, H., Vanrumste, B., Grech, R., Muscat, J., De Clercq, W., Vergult, A., D'Asseler, Y., Camilleri, K. P., Fabri, S. G., Van Huffel, S., et al. (2007) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-4-46 · Cited by 527
Dréau, A., Lesik, M., Rondin, L., Spinicelli, P., Arcizet, O., Roch, J.-F., & Jacques, V. (2011) · Physical Review B · doi:10.1103/physrevb.84.195204 · Cited by 523
Rotors and the Dynamics of Cardiac Fibrillation
Pandit, S. V., & Jalife, J. (2013) · Circulation Research · doi:10.1161/circresaha.111.300158 · Cited by 404
Subpicotesla Diamond Magnetometry
Wolf, T., Neumann, P., Nakamura, K., Sumiya, H., Ohshima, T., Isoya, J., & Wrachtrup, J. (2015) · Physical Review X · doi:10.1103/physrevx.5.041001 · Cited by 378
Self-citationMagnetic spin imaging under ambient conditions with sub-cellular resolution
Steinert, S., Ziem, F., Hall, L. T., Zappe, A., Schweikert, M., Götz, N., Aird, A., Balasubramanian, G., Hollenberg, L., & Wrachtrup, J. (2013) · Nature Communications · doi:10.1038/ncomms2588 · Cited by 331
Magnetic field imaging with nitrogen-vacancy ensembles
Pham, L. M., Le Sage, D., Stanwix, P. L., Yeung, T. K., Glenn, D., Trifonov, A., Cappellaro, P., Hemmer, P. R., Lukin, M. D., Park, H., et al. (2011) · New Journal of Physics · doi:10.1088/1367-2630/13/4/045021 · Cited by 306
High sensitivity magnetic imaging using an array of spins in diamond
Steinert, S., Dolde, F., Neumann, P., Aird, A., Naydenov, B., Balasubramanian, G., Jelezko, F., & Wrachtrup, J. (2010) · Review of Scientific Instruments · doi:10.1063/1.3385689 · Cited by 297
Broadband magnetometry and temperature sensing with a light-trapping diamond waveguide
Clevenson, H., Trusheim, M. E., Teale, C., Schröder, T., Braje, D., & Englund, D. (2015) · Nature Physics · doi:10.1038/nphys3291 · Cited by 296
Simultaneous Broadband Vector Magnetometry Using Solid-State Spins
Schloss, J. M., Barry, J. F., Turner, M. J., & Walsworth, R. L. (2018) · Physical Review Applied · doi:10.1103/physrevapplied.10.034044 · Cited by 270
Efficient Readout of a Single Spin State in Diamond via Spin-to-Charge Conversion
Shields, B. J., Unterreithmeier, Q. P., de Leon, N. P., Park, H., & Lukin, M. D. (2015) · Physical Review Letters · doi:10.1103/physrevlett.114.136402 · Cited by 240
Nonthrombogenic, stretchable, active multielectrode array for electroanatomical mapping
Lee, W., Kobayashi, S., Nagase, M., Jimbo, Y., Saito, I., Inoue, Y., Yambe, T., Sekino, M., Malliaras, G. G., Yokota, T., et al. (2018) · Science Advances · doi:10.1126/sciadv.aau2426 · Cited by 227
Self-citationTwo-photon Excited Fluorescence of Nitrogen-Vacancy Centers in Proton-Irradiated Type Ib Diamond
Wee, T.-L., Tzeng, Y.-K., Han, C.-C., Chang, H.-C., Fann, W., Hsu, J.-H., Chen, K.-M., & Yu, Y.-C. (2007) · The Journal of Physical Chemistry A · doi:10.1021/jp073938o · Cited by 196
A CMOS-integrated quantum sensor based on nitrogen–vacancy centres
Kim, D., Ibrahim, M. I., Foy, C., Trusheim, M. E., Han, R., & Englund, D. R. (2019) · Nature Electronics · doi:10.1038/s41928-019-0275-5 · Cited by 184
Alem, O., Sander, T. H., Mhaskar, R., LeBlanc, J., Eswaran, H., Steinhoff, U., Okada, Y., Kitching, J., Trahms, L., & Knappe, S. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/12/4797 · Cited by 182
Efficient photon detection from color centers in a diamond optical waveguide
Le Sage, D., Pham, L. M., Bar-Gill, N., Belthangady, C., Lukin, M. D., Yacoby, A., & Walsworth, R. L. (2012) · Physical Review B · doi:10.1103/physrevb.85.121202 · Cited by 180
Kuwahata, A., Kitaizumi, T., Saichi, K., Sato, T., Igarashi, R., Ohshima, T., MASUYAMA, Y., Iwasaki, T., Hatano, M., Jelezko, F., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-59064-6 · Cited by 167
Self-citationOptical control of excitation waves in cardiac tissue
Burton, R. A. B., Klimas, A., Ambrosi, C. M., Tomek, J., Corbett, A. D., Entcheva, E. G., & Bub, G. (2015) · Nature Photonics · doi:10.1038/nphoton.2015.196 · Cited by 156
Jensen, K., Leefer, N., Jarmola, A., Dumeige, Y., Acosta, V. M., Kehayias, P., Patton, B., & Budker, D. (2014) · Physical Review Letters · doi:10.1103/physrevlett.112.160802 · Cited by 147
Davis, H. C., Ramesh, P., Bhatnagar, A., Lee-Gosselin, A., Barry, J. F., Glenn, D. R., Walsworth, R. L., & Shapiro, M. G. (2018) · Nature Communications · doi:10.1038/s41467-017-02471-7 · Cited by 144
Spatial Resolution Requirements for Accurate Identification of Drivers of Atrial Fibrillation
Roney, C. H., Cantwell, C. D., Bayer, J. D., Qureshi, N. A., Lim, P. B., Tweedy, J. H., Kanagaratnam, P., Peters, N. S., Vigmond, E. J., & Ng, F. S. (2017) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.116.004899 · Cited by 138
Self-organization and the dynamical nature of ventricular fibrillation
Jalife, J., Gray, R. A., Morley, G. E., & Davidenko, J. M. (1998) · Chaos An Interdisciplinary Journal of Nonlinear Science · doi:10.1063/1.166289 · Cited by 136
Fujiwara, K., Oogane, M., Kanno, A., Imada, M., Jono, J., Terauchi, T., Okuno, T., Aritomi, Y., Morikawa, M., Tsuchida, M., et al. (2018) · Applied Physics Express · doi:10.7567/apex.11.023001 · Cited by 133
A new wearable multichannel magnetocardiogram system with a SERF atomic magnetometer array
Yang, Y., Xu, M., Liang, A., Yin, Y., Ma, X., Gao, Y., & Ning, X. (2021) · Scientific Reports · doi:10.1038/s41598-021-84971-7 · Cited by 125
SQUIDs in biomagnetism: a roadmap towards improved healthcare
Körber, R., Storm, J.-H., Seton, H. C., Mäkelä, J. P., Paetau, R., Parkkonen, L. T., Pfeiffer, C., Riaz, B., Schneiderman, J. F., Dong, H., et al. (2016) · Superconductor Science and Technology · doi:10.1088/0953-2048/29/11/113001 · Cited by 122
Catheter Ablation of Atrial Fibrillation: State of the Art and Future Perspectives
Rottner, L., Bellmann, B., Lin, T., Reißmann, B., Tönnis, T., Schleberger, R., Nies, M., Jungen, C., Dinshaw, L., Klatt, N., et al. (2020) · Cardiology and Therapy · doi:10.1007/s40119-019-00158-2 · Cited by 110
Diamond Magnetometry and Gradiometry Towards Subpicotesla dc Field Measurement
Zhang, C., Shagieva, F., Widmann, M., Kübler, M., Vorobyov, V., Kapitanova, P., Nenasheva, E., Corkill, R. M., Rhrle, O., Nakamura, K., et al. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.15.064075 · Cited by 95
Self-citationSQUID magnetocardiography: status and perspectives
Koch, H. (2001) · IEEE Transactions on Appiled Superconductivity · doi:10.1109/77.919284 · Cited by 91
Webb, J. L., Troise, L., Hansen, N. W., Olsson, C., Wojciechowski, A. M., Achard, J., Brinza, O., Staacke, R., Kieschnick, M., Meijer, J., et al. (2021) · Scientific Reports · doi:10.1038/s41598-021-81828-x · Cited by 84
Improved Current Density and Magnetization Reconstruction Through Vector Magnetic Field Measurements
Broadway, D. A., Lillie, S. E., Scholten, S. C., Rohner, D., Dontschuk, N., Maletinsky, P., Tetienne, J., & Hollenberg, L. C. L. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.14.024076 · Cited by 83
Jensen, K., Skarsfeldt, M. A., Stærkind, H., Arnbak, J., Balabas, M. V., Olesen, S.-P., Bentzen, B. H., & Polzik, E. S. (2018) · Scientific Reports · doi:10.1038/s41598-018-34535-z · Cited by 79
Gupta, A., Hacquebard, L. R. P., & Childress, L. (2016) · Journal of the Optical Society of America B · doi:10.1364/josab.33.000b28 · Cited by 77
N - V –Diamond Magnetic Microscopy Using a Double Quantum 4-Ramsey Protocol
Hart, C., Schloss, J. M., Turner, M. J., Scheidegger, P., Bauch, E., & Walsworth, R. L. (2021) · Physical Review Applied · doi:10.1103/physrevapplied.15.044020 · Cited by 67
The Future of Cardiovascular Imaging
Carli, M. F. D., Geva, T., & Davidoff, R. (2016) · Circulation · doi:10.1161/circulationaha.116.023511 · Cited by 65
Cardiac optical mapping – State-of-the-art and future challenges
O’Shea, C., Kabir, S. N., Holmes, A. P., Lei, M., Fabritz, L., Rajpoot, K., & Pavlović, D. (2020) · The International Journal of Biochemistry & Cell Biology · doi:10.1016/j.biocel.2020.105804 · Cited by 64
Laser spectroscopy of NV- and NV0 colour centres in synthetic diamond
Fraczek, E., Savitski, V. G., Dale, M. W., Breeze, B. G., Diggle, P., Markham, M., Bennett, A., Dhillon, H. K., Newton, M. E., & Kemp, A. J. (2017) · Optical Materials Express · doi:10.1364/ome.7.002571 · Cited by 60
O’Shea, C., Holmes, A. P., Winter, J., Correia, J., Ou, X., Dong, R., He, S., Kirchhof, P. F., Fabritz, L., Rajpoot, K., et al. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00182 · Cited by 60
Mäkinen, A. J., Zetter, R., Iivanainen, J., Zevenhoven, K. C. J., Parkkonen, L. T., & Ilmoniemi, R. J. (2020) · Journal of Applied Physics · doi:10.1063/5.0016090 · Cited by 60
Room-temperature operation of a radiofrequency diamond magnetometer near the shot-noise limit
Shin, C. S., Avalos, C. E., Butler, M. C., Trease, D. R., Seltzer, S. J., Peter Mustonen, J., Kennedy, D. J., Acosta, V. M., Budker, D., Pines, A., et al. (2012) · Journal of Applied Physics · doi:10.1063/1.4771924 · Cited by 59
Simultaneous thermometry and magnetometry using a fiber-coupled quantum diamond sensor
Hatano, Y., Shin, J.-D., Nishitani, D., Iwatsuka, H., Masuyama, Y., Sugiyama, H., Ishii, M., Onoda, S., Ohshima, T., Arai, K., et al. (2021) · Applied Physics Letters · doi:10.1063/5.0031502 · Cited by 59
Self-citationRequirements for Coregistration Accuracy in On-Scalp MEG
Zetter, R., Iivanainen, J., Stenroos, M., & Parkkonen, L. (2018) · Brain Topography · doi:10.1007/s10548-018-0656-5 · Cited by 56
Tunneling magnetoresistance sensor with pT level 1/f magnetic noise
Deak, J. G., Zhou, Z., & Shen, W. (2017) · AIP Advances · doi:10.1063/1.4978465 · Cited by 52
Boukens, B. J., Hoogendijk, M. G., Verkerk, A. O., Linnenbank, A., van Dam, P., Remme, C.-A., Fiolet, J. W., Opthof, T., Christoffels, V. M., & Coronel, R. (2013) · Cardiovascular Research · doi:10.1093/cvr/cvs299 · Cited by 51
Investigations of sensitivity and resolution of ECG and MCG in a realistically shaped thorax model
Mäntynen, V., Konttila, T., & Stenroos, M. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/23/7141 · Cited by 45
Magnetic field modeling with surface currents. Part II. Implementation and usage of bfieldtools
Zetter, R., J. Mäkinen, A., Iivanainen, J., Zevenhoven, K. C. J., Ilmoniemi, R. J., & Parkkonen, L. (2020) · Journal of Applied Physics · doi:10.1063/5.0016087 · Cited by 45
Large diamonds grown at high pressure conditions
Kanda, H. (2000) · Brazilian Journal of Physics · doi:10.1590/s0103-97332000000300003 · Cited by 41
Surface structures of synthetic diamonds
Kanda, H., Akaishi, M., Setaka, N., Yamaoka, S., & Fukunaga, O. (1980) · Journal of Materials Science · doi:10.1007/bf00550541 · Cited by 38
Alday, E. A. P., Ni, H., Zhang, C., Colman, M. A., Gan, Z., & Zhang, H. (2016) · PLOS ONE · doi:10.1371/journal.pone.0160999 · Cited by 32
High resolution optical mapping of cardiac electrophysiology in pre-clinical models
O’Shea, C., Winter, J., Kabir, S. N., O’Reilly, M., Wells, S. P., Baines, O., Sommerfeld, L. C., Correia, J., Lei, M., Kirchhof, P. F., et al. (2022) · Scientific Data · doi:10.1038/s41597-022-01253-1 · Cited by 29
Kawakami, S., Takaki, H., Hashimoto, S., Kimura, Y., Nakashima, T., Aiba, T., Kusano, K. F., Kamakura, S., Yasuda, S., & Sugimachi, M. (2017) · Circulation Journal · doi:10.1253/circj.cj-16-0683 · Cited by 27
Acute regional left atrial ischemia causes acceleration of atrial drivers during atrial fibrillation
Yamazaki, M., Avula, U. M. R., Bandaru, K., Atreya, A. R., Boppana, V. S., Honjo, H., Kodama, I., Kamiya, K., & Kalifa, J. (2013) · Heart Rhythm · doi:10.1016/j.hrthm.2013.02.023 · Cited by 23
Self-citationUltralong Dephasing Times in Solid-State Spin Ensembles via Quantum Control
Bauch, E., Hart, C., Schloss, J. M., Turner, M. J., Barry, J. F., Kehayias, P., Singh, S., & Walsworth, R. L. (2018) · DSpace@MIT (Massachusetts Institute of Technology) · doi:10.48550/arxiv.1801.03793 · Cited by 23
Roudijk, R. W., Boonstra, M., Brummel, R., Kassenberg, W., Blom, L. J., Oostendorp, T. F., Riele, A. S. J. M. T., van der Heijden, J. F., Asselbergs, F. W., van Dam, P. M., et al. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2021.730736 · Cited by 19
Electrocardiography in Veterinary Medicine
Varshney, J. P. (2020) · Springer eBooks · doi:10.1007/978-981-15-3699-1 · Cited by 11
Yamazaki, M., Tomii, N., Tsuneyama, K., Takanari, H., Niwa, R., Honjo, H., Kodama, I., Arafune, T., Makita, N., Sakuma, I., et al. (2021) · Heart Rhythm · doi:10.1016/j.hrthm.2021.10.012 · Cited by 11
Self-citationECG Localization Method Based on Volume Conductor Model and Kalman Filtering
Nakano, Y., Rashed, E. A., Nakane, T., Laakso, I., & Hirata, A. (2021) · Sensors · doi:10.3390/s21134275 · Cited by 6
Cited by
Sensitive ac and dc magnetometry with nitrogen-vacancy-center ensembles in diamond
Barry, J. F., Steinecker, M., Alsid, S. T., Majumder, J. A., Pham, L., O’Keeffe, M. F., & Braje, D. (2024) · Physical Review Applied · doi:10.1103/physrevapplied.22.044069 · Cited by 140
Diamond Integrated Quantum Nanophotonics: Spins, Photons and Phonons
Shandilya, P. K., Flågan, S., Carvalho, N. C., Zohari, E., Kavatamane, V. K., Losby, J. E., & Barclay, P. E. (2022) · Journal of Lightwave Technology · doi:10.1109/jlt.2022.3210466 · Cited by 99
“NV based sensors are beginning to be employed in biomedical imaging [330], [337], [340] but”
BackgroundClinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Recent progress in hybrid diamond photonics for quantum information processing and sensing
Katsumi, R., Takada, K., Jelezko, F., & Yatsui, T. (2025) · Communications Engineering · doi:10.1038/s44172-025-00398-2 · Cited by 96
“---------------------------------------------------------------------- Reviewer #1”
Fiber-Coupled Diamond Magnetometry with an Unshielded Sensitivity of 30 pT / Hz
Graham, S. M., Rahman, A., Munn, L., Patel, R. L., Newman, A. J., Stephen, C. J., Colston, G., Nikitin, A. P., Edmonds, A. M., Twitchen, D. J., et al. (2023) · Physical Review Applied · doi:10.1103/physrevapplied.19.044042 · Cited by 71
“High-sensitivity NVC magnetometers with portable fiber-coupled sensor heads are more suitable than purely table-top setups for a wide range of applications, from magnetocardiography (MCG) [33, 47] to the detection of corrosion in steel [13, 15].”
BackgroundInfluential
Metadata
- Authors
- Arai, K., Kuwahata, A., Nishitani, D., Fujisaki, I., Matsuki, R., Nishio, Y., Xin, Z., Cao, X., Hatano, Y., Onoda, S., Shinei, C., Miyakawa, M., Taniguchi, T., Yamazaki, M., Teraji, T., Ohshima, T., Hatano, M., Sekino, M., & Iwasaki, T.
- Affiliations
- Tokyo Institute of Technology (JP)
- Tohoku University (JP)
- The University of Tokyo (JP)
- National Institutes for Quantum Science and Technology (JP)
- National Institute for Materials Science (JP)
- Nagano Municipal Hospital (JP)
- Corresponding author
- Hatano, Mutsuko
- Sekino, Masaki
- Iwasaki, Takayuki
- ORCID
- Venue
- Communications Physics
- Published
- 2022-08-23
- Volume
- 5
- Issue
- 1
- Pages
- 200(1)-200(10)
- ISSN
- 2399-3650 (ISSN-L)
- OpenAlex
- W4292697381
- Cited by
- 85
- References
- 68
- Citations by year
- 2026: 15 · 2025: 19 · 2024: 15 · 2023: 33 · 2022: 3
- FWCI
- 7.18
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Communications Physics (publishedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Institutional Repositories DataBase (IRDB) (submittedVersion)
- QST-Repository (National Institutes for Quantum and Radiological Science and Technology) (submittedVersion)
- Funders
- MEXT | JST | Precursory Research for Embryonic Science and Technology (JPMJPR20B1)
- Ministry of Education, Culture, Sports, Science and Technology (JPMXS0118068379, JPMXS0118067395)
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- magnetocardiography, nitrogen-vacancy centers, diamond quantum sensors, current dipoles, ventricular arrhythmias, cardiac arrhythmias, thoracotomy, rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.