Variability and Standardization of Quantitative Imaging: Monoparametric to Multiparametric Quantification, Radiomics, and Artificial Intelligence
Open accessThe DOI matches its Crossref record; 197 of the 197 references it cites were found in a registry or an index.
Abstract
Radiological images have been assessed qualitatively in most clinical settings by the expert eyes of radiologists and other clinicians. On the other hand, quantification of radiological images has the potential to detect early disease that may be difficult to detect with human eyes, complement or replace biopsy, and provide clear differentiation of disease stage. Further, objective assessment by quantification is a prerequisite of personalized/precision medicine. This review article aims to summarize and discuss how the variability of quantitative values derived from radiological images are induced by a number of factors and how these variabilities are mitigated and standardization of the quantitative values are achieved. We discuss the variabilities of specific biomarkers derived from magnetic resonance imaging and computed tomography, and focus on diffusion-weighted imaging, relaxometry, lung density evaluation, and computer-aided computed tomography volumetry. We also review the sources of variability and current efforts of standardization of the rapidly evolving techniques, which include radiomics and artificial intelligence.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 31,709 · cited 1×
“9 to 1).(20) A moderate ICC can be considered sufficient when a measurement is used for group-level comparisons for research purposes.”
Background(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Reduced Lung-Cancer Mortality with Low-Dose Computed Tomographic Screening
The National Lung Screening Trial Research Team (2011) · New England Journal of Medicine · doi:10.1056/nejmoa1102873 · Cited by 11,064
Radiomics: Images Are More than Pictures, They Are Data
Gillies, R. J., Kinahan, P. E., & Hricak, H. (2016) · Radiology · doi:10.1148/radiol.2015151169 · Cited by 8,421 · cited 1×
“Radiomics is based on the high-throughput computer extraction of potentially innumerable numbers of quantitative imaging metrics, or “radiomic features,”which will be collectively used for the prediction of diagnosis and prognosis and gene expression profiling.(198) These radiomic features can be combined with other p…”
BackgroundBiomarkers and surrogate endpoints: Preferred definitions and conceptual framework
(2001) · Clinical Pharmacology & Therapeutics · doi:10.1067/mcp.2001.113989 · Cited by 6,762
Effects of Diffusion on Free Precession in Nuclear Magnetic Resonance Experiments
Carr, H. Y., & Purcell, E. M. (1954) · Physical Review · doi:10.1103/physrev.94.630 · Cited by 6,466 · cited 1×
“A version of MET2, termed the Carr-Purcell-Meiboom-Gill sequence, accelerates MET2 by incorporating a 180-degree phase increment to refocusing pulses and is now considered to be the criterion standard for T2 measurement.(107,108) Another approach for T2 measurement is driven equilibrium single-pulse observation of T2…”
MethodModified Spin-Echo Method for Measuring Nuclear Relaxation Times
Meiboom, S., & Gill, D. (1958) · Review of Scientific Instruments · doi:10.1063/1.1716296 · Cited by 6,022 · cited 1×
“A version of MET2, termed the Carr-Purcell-Meiboom-Gill sequence, accelerates MET2 by incorporating a 180-degree phase increment to refocusing pulses and is now considered to be the criterion standard for T2 measurement.(107,108) Another approach for T2 measurement is driven equilibrium single-pulse observation of T2…”
MethodRadiomics: the bridge between medical imaging and personalized medicine
Lambin, P., Leijenaar, R. T. H., Deist, T. M., Peerlings, J., de Jong, E. E. C., van Timmeren, J., Sanduleanu, S., Larue, R. T. H. M., Even, A. J. G., Jochems, A., et al. (2017) · Nature Reviews Clinical Oncology · doi:10.1038/nrclinonc.2017.141 · Cited by 5,688
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220 · cited 3×
“However, this approach only evaluates the linear association between 2measurementswithout consideration of bias, and it does not give an indication of repeatability or agreement.(18) Further, a large betweensubject variation makes the correlation coefficient higher.”
BackgroundMethodInfluentialTofts, P. S., Brix, G., Buckley, D. L., Evelhoch, J. L., Henderson, E., Knopp, M. V., Larsson, H. B. W., Lee, T.-Y., Mayr, N. A., Parker, G. J. M., et al. (1999) · Journal of Magnetic Resonance Imaging · doi:10.1002/(sici)1522-2586(199909)10:3<223::aid-jmri2>3.0.co;2-s · Cited by 3,143 · cited 1×
“Mitigation of variation across centers can be achieved by (1) setting scanning and analysis procedures to be as identical as possible so that any systemic errors are replicated across participating centers(26) or (2) aiming for high accuracy at each center.(27) Although standardizing the scanning protocol is the simpl…”
BackgroundGuidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed
Kottner, J., Audigé, L., Brorson, S., Donner, A., Gajewski, B. J., Hróbjartsson, A., Roberts, C., Shoukri, M., & Streiner, D. L. (2011) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2010.03.002 · Cited by 2,174
Diffusion-Weighted Magnetic Resonance Imaging as a Cancer Biomarker: Consensus and Recommendations
Padhani, A. R., Liu, G., Mu-Koh, D., Chenevert, T. L., Thoeny, H. C., Takahara, T., Dzik-Jurasz, A., Ross, B. D., Van Cauteren, M., Collins, D., et al. (2009) · Neoplasia · doi:10.1593/neo.81328 · Cited by 1,989
Lakhani, P., & Sundaram, B. (2017) · Radiology · doi:10.1148/radiol.2017162326 · Cited by 1,775
Harmonization of cortical thickness measurements across scanners and sites
Fortin, J.-P., Cullen, N., Sheline, Y. I., Taylor, W. D., Aselcioglu, I., Cook, P. A., Adams, P., Cooper, C., Fava, M., McGrath, P. J., et al. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2017.11.024 · Cited by 1,580 · cited 1×
“If only postprocessed data (eg, fractional anisotropy map and cortical thickness) are available, regression analysis or more sophisticated statistical approaches can be performed for harmonization.(43,44)”
MethodMagnetic resonance fingerprinting
Ma, D., Gulani, V., Seiberlich, N., Liu, K., Sunshine, J. L., Duerk, J. L., & Griswold, M. A. (2013) · Nature · doi:10.1038/nature11971 · Cited by 1,577 · cited 1×
“In MRF, radiofrequency pulses and repetition times are simultaneously varied in a pseudorandom fashion to create signal evolutions that characterize the various relaxation processes unique for each type of tissue (so-called fingerprint).(131) The acquired signal evolutions are”
BackgroundHarmonization of multi-site diffusion tensor imaging data
Fortin, J.-P., Parker, D., Tunç, B., Watanabe, T., Elliott, M. A., Ruparel, K., Roalf, D. R., Satterthwaite, T. D., Gur, R. C., Gur, R. E., et al. (2017) · NeuroImage · doi:10.1016/j.neuroimage.2017.08.047 · Cited by 1,261 · cited 1×
“If only postprocessed data (eg, fractional anisotropy map and cortical thickness) are available, regression analysis or more sophisticated statistical approaches can be performed for harmonization.(43,44)”
MethodRepeatability and Reproducibility of Radiomic Features: A Systematic Review
Traverso, A., Wee, L., Dekker, A., & Gillies, R. (2018) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2018.05.053 · Cited by 903
Hsia, C. C. W., Hyde, D. M., Ochs, M., & Weibel, E. R. (2010) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200809-1522st · Cited by 902
Park, S. H., & Han, K. (2018) · Radiology · doi:10.1148/radiol.2017171920 · Cited by 894
Time Saving in Measurement of NMR and EPR Relaxation Times
Look, D. C., & Locker, D. R. (1970) · Review of Scientific Instruments · doi:10.1063/1.1684482 · Cited by 819
Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo
Rooney, W. D., Johnson, G., Li, X., Cohen, E. R., Kim, S., Ugurbil, K., & Springer, C. S. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21122 · Cited by 652 · cited 2×
“T1 values significantly increase with the field strength.(60) The criterion standard method for measuring T1 relaxation time is the inversion recovery technique, in which only one echo is acquired at a time and full spin relaxation is awaited (approximately 5 T1 periods) before the next spin inversion.”
BackgroundMeasuring Computed Tomography Scanner Variability of Radiomics Features
Mackin, D., Fave, X., Zhang, L., Fried, D., Yang, J., Taylor, B., Rodriguez-Rivera, E., Dodge, C., Jones, A. K., & Court, L. (2015) · Investigative Radiology · doi:10.1097/rli.0000000000000180 · Cited by 651
Statistics Notes: Measurement error and correlation coefficients
Bland, J. M., & Altman, D. G. (1996) · BMJ · doi:10.1136/bmj.313.7048.41 · Cited by 639
Berenguer, R., Pastor-Juan, M. D. R., Canales-Vázquez, J., Castro-García, M., Villas, M. V., Mansilla Legorburo, F., & Sabater, S. (2018) · Radiology · doi:10.1148/radiol.2018172361 · Cited by 622 · cited 2×
“In general, a lack of reproducibility in radiomic features is a limitation for radiomics to be widely used in clinical practice.(199,203,204) The stability of radiomic features is sensitive, at various degrees, to a number of processing factors, including image acquisition parameters, reconstruction algorithms, digita…”
BackgroundRobust Radiomics Feature Quantification Using Semiautomatic Volumetric Segmentation
Parmar, C., Rios Velazquez, E., Leijenaar, R., Jermoumi, M., Carvalho, S., Mak, R. H., Mitra, S., Shankar, B. U., Kikinis, R., Haibe-Kains, B., et al. (2014) · PLoS ONE · doi:10.1371/journal.pone.0102107 · Cited by 611 · cited 1×
“Semiautomated or fully automated methods improved feature reproducibility.(212) However, the articles included in this systematic review were mostly about CTand PET.”
MethodIntrinsic dependencies of CT radiomic features on voxel size and number of gray levels
Shafiq‐ul‐Hassan, M., Zhang, G. G., Latifi, K., Ullah, G., Hunt, D. C., Balagurunathan, Y., Abdalah, M. A., Schabath, M. B., Goldgof, D. G., Mackin, D., et al. (2017) · Medical Physics · doi:10.1002/mp.12123 · Cited by 585
The principle of reciprocity in signal strength calculations?A mathematical guide
Hoult, D. I. (2000) · Concepts in Magnetic Resonance · doi:10.1002/1099-0534(2000)12:4<173::aid-cmr1>3.0.co;2-q · Cited by 550 · cited 1×
“The reciprocity principle can still be used when different coils are used by performing an additional acquisition in which the transmit coil is used for receiving.(55) However, the reciprocity principle becomes less accurate at a field strength of 3 T or higher.”
How to Read Articles That Use Machine Learning: Users’ Guides to the Medical Literature
Liu, Y., Chen, P.-H. C., Krause, J., & Peng, L. (2019) · JAMA · doi:10.1001/jama.2019.16489 · Cited by 516 · cited 1×
“the development and evaluation of an AI algorithm in the medical field.(219,227,228) In brief, desired steps to develop a reliable AI algorithm include the following: (1) using reliable reference standards, (2) using a training dataset that matches the intended use, (3) tuning hyperparameters on a dataset independent…”
MethodReproducibility of radiomics for deciphering tumor phenotype with imaging
Zhao, B., Tan, Y., Tsai, W.-Y., Qi, J., Xie, C., Lu, L., & Schwartz, L. H. (2016) · Scientific Reports · doi:10.1038/srep23428 · Cited by 509
Rapid magnetic resonance quantification on the brain: Optimization for clinical usage
Warntjes, J. B. M., Leinhard, O. D., West, J., & Lundberg, P. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21635 · Cited by 500 · cited 1×
“B1 + field measurements can be simultaneously performed based on the same acquisition data.(116) The performance of the MDME sequence was examined on 3 scanners from 3 different vendors.”
BackgroundPulmonary Arterial Enlargement and Acute Exacerbations of COPD
Wells, J. M., Washko, G. R., Han, M. K., Abbas, N., Nath, H., Mamary, A. J., Regan, E., Bailey, W. C., Martinez, F. J., Westfall, E., et al. (2012) · New England Journal of Medicine · doi:10.1056/nejmoa1203830 · Cited by 499
Deoni, S. C. L., Peters, T. M., & Rutt, B. K. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20314 · Cited by 478
Deep learning for undersampled MRI reconstruction
Hyun, C. M., Kim, H. P., Lee, S. M., Lee, S., & Seo, J. K. (2018) · Physics in Medicine & Biology · doi:10.1088/1361-6560/aac71a · Cited by 428
Jiang, Y., Ma, D., Seiberlich, N., Gulani, V., & Griswold, M. A. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25559 · Cited by 422
Airflow Limitation and Airway Dimensions in Chronic Obstructive Pulmonary Disease
Hasegawa, M., Nasuhara, Y., Onodera, Y., Makita, H., Nagai, K., Fuke, S., Ito, Y., Betsuyaku, T., & Nishimura, M. (2006) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200601-037oc · Cited by 413 · cited 1×
“One approach determines the percentage of low attenuation area in the lung, which reflects the destruction of the lung parenchyma, and the other determines the percentage of wall area in the bronchi, which reflects bronchial narrowing and wall thickening.(183) In addition, 3D airway luminal volumetry has been introduc…”
MethodRaunig, D. L., McShane, L. M., Pennello, G., Gatsonis, C., Carson, P. L., Voyvodic, J. T., Wahl, R. L., Kurland, B. F., Schwarz, A. J., Gönen, M., et al. (2015) · Statistical Methods in Medical Research · doi:10.1177/0962280214537344 · Cited by 394 · cited 5×
“If the relationship between Y and X is nonlinear, the ability of a measurement to detect change in the measurand is inconstant.(15) If there is lack of a standard reference, another imaging measurement”
BackgroundInfluentialSaghir, Z., Dirksen, A., Ashraf, H., Bach, K. S., Brodersen, J., Clementsen, P. F., Døssing, M., Hansen, H., Kofoed, K. F., Larsen, K. R., et al. (2012) · Thorax · doi:10.1136/thoraxjnl-2011-200736 · Cited by 390
Horeweg, N., Scholten, E. T., de Jong, P. A., van der Aalst, C. M., Weenink, C., Lammers, J.-W. J., Nackaerts, K., Vliegenthart, R., ten Haaf, K., Yousaf-Khan, U. A., et al. (2014) · The Lancet Oncology · doi:10.1016/s1470-2045(14)70387-0 · Cited by 384
Infante, M., Cavuto, S., Lutman, F. R., Passera, E., Chiarenza, M., Chiesa, G., Brambilla, G., Angeli, E., Aranzulla, G., Chiti, A., et al. (2015) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201408-1475oc · Cited by 365
Automated Abdominal Segmentation of CT Scans for Body Composition Analysis Using Deep Learning
Weston, A. D., Korfiatis, P., Kline, T. L., Philbrick, K. A., Kostandy, P., Sakinis, T., Sugimoto, M., Takahashi, N., & Erickson, B. J. (2019) · Radiology · doi:10.1148/radiol.2018181432 · Cited by 358
Becker, A. S., Marcon, M., Ghafoor, S., Wurnig, M. C., Frauenfelder, T., & Boss, A. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000358 · Cited by 352
Cheng, H. M., & Wright, G. A. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20791 · Cited by 351 · cited 1×
“The accuracy of flip angle depends on the B1 + inhomogeneity at a given spatial location, which can bemeasured by B1 + mapping.(51) Notably, every vendor uses their own radiofrequency pulse shapes that lead to variability in flip angle, complicating comparison across scanners from different vendors.”
BackgroundVerma, N., Cowperthwaite, M. C., Burnett, M. G., & Markey, M. K. (2013) · Neuro-Oncology · doi:10.1093/neuonc/nos307 · Cited by 351 · cited 1×
“Control of these variabilities may enable ADC to replace biopsy, such as for the differential diagnosis between tumor recurrence and necrosis.(77)”
BackgroundNodule management protocol of the NELSON randomised lung cancer screening trial
Xu, D. M., Gietema, H., de Koning, H., Vernhout, R., Nackaerts, K., Prokop, M., Weenink, C., Lammers, J.-W., Groen, H., Oudkerk, M., et al. (2006) · Lung Cancer · doi:10.1016/j.lungcan.2006.08.006 · Cited by 347
False Discovery Rates in PET and CT Studies with Texture Features: A Systematic Review
Chalkidou, A., O’Doherty, M. J., & Marsden, P. K. (2015) · PLOS ONE · doi:10.1371/journal.pone.0124165 · Cited by 344 · cited 1×
“It is largely unknown how the repeatability and reproducibility of each feature are propagated to the final result of radiomics, and therefore, validation of a radiomics algorithm against another independent dataset is considered to be crucial.(215) There are several multi-institutional efforts to standardize and incr…”
Nevill, A. M., & Atkinson, G. (1997) · British Journal of Sports Medicine · doi:10.1136/bjsm.31.4.314 · Cited by 317
Sasaki, M., Yamada, K., Watanabe, Y., Matsui, M., Ida, M., Fujiwara, S., & Shibata, E. (2008) · Radiology · doi:10.1148/radiol.2492071681 · Cited by 314 · cited 1×
“When interscanner CV is evaluated, one should be careful if the scanners are from the same vendor or from different vendors as significant intervendor bias in ADC measurement of the brain has been reported with a %bias up to 7%.(93)”
MethodKessler, L. G., Barnhart, H. X., Buckler, A. J., Choudhury, K. R., Kondratovich, M. V., Toledano, A., Guimaraes, A. R., Filice, R., Zhang, Z., Sullivan, D. C., et al. (2015) · Statistical Methods in Medical Research · doi:10.1177/0962280214537333 · Cited by 293 · cited 7×
“Overall, the terminologies related to variability used in this article conform to those suggested by the QIBATerminology Working Group in 2015.(9)”
BackgroundMethodInfluentialOn the accuracy of T1 mapping: Searching for common ground
Stikov, N., Boudreau, M., Levesque, I. R., Tardif, C. L., Barral, J. K., & Pike, G. B. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25135 · Cited by 285 · cited 1×
“The acquired B1 + map can be used for the measurement of tissue parameters such as T1 and magnetization transfer to correct the achieved flip angle for the intended one.(52,53) The B1 + field is smooth compared with anatomical structures, even at high field strengths, so B1 + maps are often acquired at low resolution…”
MethodTahmassebi, A., Wengert, G. J., Helbich, T. H., Bago-Horvath, Z., Alaei, S., Bartsch, R., Dubsky, P., Baltzer, P., Clauser, P., Kapetas, P., et al. (2019) · Investigative Radiology · doi:10.1097/rli.0000000000000518 · Cited by 274 · cited 1×
“Another possible application of multiparametric MRI in clinical practice is feeding the data into machine learning for diagnosis,(156) tumor grading,(157) or the prediction of treatment response.(158) Multiparametric quantitative MRI can also be used for extracting new measures that reflect subvoxel microstructural in…”
MethodOptimizing the precision in T1 relaxation estimation using limited flip angles
Wang, H. Z., Riederer, S. J., & Lee, J. N. (1987) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910050502 · Cited by 271
Peerlings, J., Woodruff, H. C., Winfield, J. M., Ibrahim, A., Van Beers, B. E., Heerschap, A., Jackson, A., Wildberger, J. E., Mottaghy, F. M., DeSouza, N. M., et al. (2019) · Scientific Reports · doi:10.1038/s41598-019-41344-5 · Cited by 266 · cited 1×
“85 for all tumor entities investigated (ie, ovarian cancer, lung cancer, and colorectal liver metastasis).(213) For magnetic field strength and vendor differences, 245 and 209 features, respectively, were stable.”
BackgroundLarue, R. T. H. M., van Timmeren, J. E., de Jong, E. E. C., Feliciani, G., Leijenaar, R. T. H., Schreurs, W. M. J., Sosef, M. N., Raat, F. H. P. J., van der Zande, F. H. R., Das, M., et al. (2017) · Acta Oncologica · doi:10.1080/0284186x.2017.1351624 · Cited by 262 · cited 1×
“However, the selection of interpolationmethods (eg, nearest neighbor, trilinear and tricubic interpolation) has been revealed to affect the reproducibility of radiomic features.(209) Also, how the software treats the boundary of the volume of interest also affects radiomic features; when the voxels outside the ROI are…”
BackgroundBonekamp, D., Nagae, L. M., Degaonkar, M., Matson, M., Abdalla, W. M. A., Barker, P. B., Mori, S., & Horská, A. (2007) · NeuroImage · doi:10.1016/j.neuroimage.2006.09.020 · Cited by 260
The effect of measuring error on the results of therapeutic trials in advanced cancer
Moertel, C. G., & Hanley, J. A. (1976) · Cancer · doi:10.1002/1097-0142(197607)38:1<388::aid-cncr2820380156>3.0.co;2-a · Cited by 257
Radiation Dose Reduction in Chest CT: A Review
Kubo, T., Lin, P.-J. P., Stiller, W., Takahashi, M., Kauczor, H.-U., Ohno, Y., & Hatabu, H. (2008) · American Journal of Roentgenology · doi:10.2214/ajr.07.2556 · Cited by 256
Self-citationHamilton, J. I., Jiang, Y., Chen, Y., Ma, D., Lo, W.-C., Griswold, M., & Seiberlich, N. (2017) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26216 · Cited by 246
Tracking tumor biology with radiomics: a systematic review utilizing a radiomics quality score
Sanduleanu, S., Woodruff, H. C., de Jong, E. E. C., van Timmeren, J. E., Jochems, A., Dubois, L., & Lambin, P. (2018) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2018.03.033 · Cited by 242
Deep Learning Based Noise Reduction for Brain MR Imaging: Tests on Phantoms and Healthy Volunteers
Kidoh, M., Shinoda, K., Kitajima, M., Isogawa, K., Nambu, M., Uetani, H., Morita, K., Nakaura, T., Tateishi, M., Yamashita, Y., et al. (2020) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.mp.2019-0018 · Cited by 242
Hagiwara, A., Warntjes, M., Hori, M., Andica, C., Nakazawa, M., Kumamaru, K. K., Abe, O., & Aoki, S. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000365 · Cited by 240
Self-citationRobustness and Reproducibility of Radiomics in Magnetic Resonance Imaging: A Phantom Study
Baeßler, B., Weiss, K., & Pinto dos Santos, D. (2019) · Investigative Radiology · doi:10.1097/rli.0000000000000530 · Cited by 229
DeepHarmony: A deep learning approach to contrast harmonization across scanner changes
Dewey, B. E., Zhao, C., Reinhold, J. C., Carass, A., Fitzgerald, K. C., Sotirchos, E. S., Saidha, S., Oh, J., Pham, D. L., Calabresi, P. A., et al. (2019) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2019.05.041 · Cited by 228
T2 quantitation of articular cartilage at 1.5 T
Maier, C. F., Tan, S. G., Hariharan, H., & Potter, H. G. (2003) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10263 · Cited by 217 · cited 1×
“T2 measurements are also affected by magnetization transfer effects between the water and macromolecular protons, resulting in diminished signal in the free water and inaccurate T2 measurement.(111,112) Similar to the discrepancy among T1 measurement methods, T2 measurement methods are known to show disagreement with…”
BackgroundVoxel size and gray level normalization of CT radiomic features in lung cancer
Shafiq-ul-Hassan, M., Latifi, K., Zhang, G., Ullah, G., Gillies, R., & Moros, E. (2018) · Scientific Reports · doi:10.1038/s41598-018-28895-9 · Cited by 215
Quantitative Assessment of Airway Remodeling Using High-Resolution CT*
Nakano, Y., Müller, N. L., King, G. G., Niimi, A., Kalloger, S. E., Mishima, M., & Paré, P. D. (2002) · Chest · doi:10.1378/chest.122.6_suppl.271s · Cited by 214
Multi-centre reproducibility of diffusion MRI parameters for clinical sequences in the brain
Grech‐Sollars, M., Hales, P. W., Miyazaki, K., Raschke, F., Rodriguez, D., Wilson, M., Gill, S. K., Banks, T., Saunders, D. E., Clayden, J. D., et al. (2015) · NMR in Biomedicine · doi:10.1002/nbm.3269 · Cited by 213
MR Fingerprinting for Rapid Quantitative Abdominal Imaging
Chen, Y., Jiang, Y., Pahwa, S., Ma, D., Lu, L., Twieg, M. D., Wright, K. L., Seiberlich, N., Griswold, M. A., & Gulani, V. (2016) · Radiology · doi:10.1148/radiol.2016152037 · Cited by 211
Lung Cancers Diagnosed at Annual CT Screening: Volume Doubling Times
Henschke, C. I., Yankelevitz, D. F., Yip, R., Reeves, A. P., Farooqi, A., Xu, D., Smith, J. P., Libby, D. M., Pasmantier, M. W., & Miettinen, O. S. (2012) · Radiology · doi:10.1148/radiol.12102489 · Cited by 208
Yamashita, A., Yahata, N., Itahashi, T., Lisi, G., Yamada, T., Ichikawa, N., Takamura, M., Yoshihara, Y., Kunimatsu, A., Okada, N., et al. (2019) · PLOS Biology · doi:10.1371/journal.pbio.3000042 · Cited by 201
Braithwaite, A. C., Dale, B. M., Boll, D. T., & Merkle, E. M. (2009) · Radiology · doi:10.1148/radiol.2502080849 · Cited by 195
Matsuoka, S., Washko, G. R., Yamashiro, T., Estepar, R. S. J., Diaz, A., Silverman, E. K., Hoffman, E., Fessler, H. E., Criner, G. J., Marchetti, N., et al. (2010) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200908-1189oc · Cited by 194
Hansell, D. M., Goldin, J. G., King, T. E., Lynch, D. A., Richeldi, L., & Wells, A. U. (2015) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(15)00096-x · Cited by 189
Mårtensson, G., Ferreira, D., Granberg, T., Cavallin, L., Oppedal, K., Padovani, A., Rektorova, I., Bonanni, L., Pardini, M., Kramberger, M. G., et al. (2020) · Medical Image Analysis · doi:10.1016/j.media.2020.101714 · Cited by 176 · cited 1×
“To develop an AI algorithm that is robust to variability in acquisition parameters, machine settings, and clinical conditions, the algorithm should be trained with a heterogeneous dataset.(229) The standardization/ harmonization of the input images could be one approach to making an algorithm that is generalizable to…”
MethodBuckler, A. J., Bresolin, L., Dunnick, N. R., Sullivan, D. C., & For the Group (2011) · Radiology · doi:10.1148/radiol.10100799 · Cited by 175 · cited 1×
“Q uantitative imaging, defined as the extraction of quantifiable features from radiological images,(1) has been increasingly performed for the measurement of normal biological and pathological processes, patient risk stratification, evaluation of treatment response and outcome, and drug development.(2) Such features o…”
BackgroundSimultaneous three-dimensional myocardial T1 and T2 mapping in one breath hold with 3D-QALAS
Kvernby, S., Warntjes, M. J. B., Haraldsson, H., Carlhäll, C., Engvall, J. E., & Ebbers, T. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-014-0102-0 · Cited by 175
Quantitative imaging biomarkers: A review of statistical methods for computer algorithm comparisons
Obuchowski, N. A., Reeves, A. P., Huang, E. P., Wang, X.-F., Buckler, A. J., Kim, H. J. (., Barnhart, H. X., Jackson, E. F., Giger, M. L., Pennello, G., et al. (2015) · Statistical Methods in Medical Research · doi:10.1177/0962280214537390 · Cited by 172 · cited 1×
“considered as the true value, these are more appropriately referred to as reference standards; they are described as well-accepted or commonly used methods for measuring a biomarker but also have associated bias and/or random error.(38) For example, histology and pathology are known to be affected by fixation and stai…”
MethodQuantitative magnetic resonance imaging phantoms: A review and the need for a system phantom
Keenan, K. E., Ainslie, M., Barker, A. J., Boss, M. A., Cecil, K. M., Charles, C., Chenevert, T. L., Clarke, L., Evelhoch, J. L., Finn, P., et al. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26982 · Cited by 170 · cited 4×
“The apparent diffusion coefficient (ADC) of the pure water changes approximately 3% per °C at room temperature.(50) The phantom should be stored in theMRI room and reach a temperature close to that in the magnet bore so that the temperature of the phantom does not fluctuate during the scan.”
BackgroundMethodInfluentialCetin Karayumak, S., Bouix, S., Ning, L., James, A., Crow, T., Shenton, M., Kubicki, M., & Rathi, Y. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2018.08.073 · Cited by 164 · cited 1×
“Harmonization of raw data is particularly important for diffusionweighted imaging (DWI) data to be analyzed by multicompartment models or tractography, and model-free methods for harmonization of raw DWI signals have also been suggested.(45,46) Harmonization by deep learning has also been proposed, but the algorithm s…”
MethodPinker, K., Moy, L., Sutton, E. J., Mann, R. M., Weber, M., Thakur, S. B., Jochelson, M. S., Bago-Horvath, Z., Morris, E. A., Baltzer, P. A., et al. (2018) · Investigative Radiology · doi:10.1097/rli.0000000000000465 · Cited by 163 · cited 1×
“One possible approach for implementing multiparametric MRI into clinical practice is setting a cutoff value for each parameter.(146,153) Pinker et al(154) evaluated the diagnostic accuracy of contrast-enhancedMRI, DWI, and MR spectroscopy in combinations of 2 or 3 and used a single measure by setting a threshold for e…”
MethodOh, J., Han, E. T., Pelletier, D., & Nelson, S. J. (2005) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2005.10.016 · Cited by 162 · cited 1×
“An alternative approach to T2 measurement is to separate the imaging section of the sequence from T2-weighting using the T2 preparation pulse (T2-prep), which enables acquisition of multiple echo with fast imaging technique.(110) As with T1 measurement, all these methods are affected by B1 + and B1 − inhomogeneities.”
MethodMR fingerprinting with simultaneous B1 estimation
Buonincontri, G., & Sawiak, S. J. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26009 · Cited by 160
Basics of iterative reconstruction methods in computed tomography: A vendor-independent overview
Stiller, W. (2018) · European Journal of Radiology · doi:10.1016/j.ejrad.2018.10.025 · Cited by 160 · cited 1×
“In addition, newly developed iterative reconstruction (IR) methods have been introduced and applied in routine clinical practice.(171,172) In fact, dose reduction strategies have been realized by employing a variety of techniques for data acquisition, such as tube current reduction, tube voltage reduction, increased h…”
MethodBiomarkers of progression of chronic obstructive pulmonary disease (COPD).
Shaw, J. G., Vaughan, A., Dent, A. G., O’Hare, P. E., Goh, F., Bowman, R., Fong, K. M., & Yang, I. A. (2014) · PubMed · doi:10.3978/j.issn.2072-1439.2014.11.33 · Cited by 157
Chenevert, T. L., Galbán, C. J., Ivancevic, M. K., Rohrer, S. E., Londy, F. J., Kwee, T. C., Meyer, C. R., Johnson, T. D., Rehemtulla, A., & Ross, B. D. (2011) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22363 · Cited by 157
Replicability of diffusion tensor imaging measurements of fractional anisotropy and trace in brain
Pfefferbaum, A., Adalsteinsson, E., & Sullivan, E. V. (2003) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10377 · Cited by 156
Jiang, Y., Ma, D., Keenan, K. E., Stupic, K. F., Gulani, V., & Griswold, M. A. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26509 · Cited by 156
Nyflot, M. J., Yang, F., Byrd, D. W., Bowen, S. R., Sandison, G. A., & Kinahan, P. E. (2015) · Journal of medical imaging · doi:10.1117/1.jmi.2.4.041002 · Cited by 150 · cited 1×
“In general, a lack of reproducibility in radiomic features is a limitation for radiomics to be widely used in clinical practice.(199,203,204) The stability of radiomic features is sensitive, at various degrees, to a number of processing factors, including image acquisition parameters, reconstruction algorithms, digita…”
BackgroundIntrarater and interrater reliability of the MS functional composite outcome measure
Cohen, J. A., Fischer, J. S., Bolibrush, D. M., Jak, A. J., Kniker, J. E., Mertz, L., Skaramagas, T., & Cutter, G. (2000) · Neurology · doi:10.1212/wnl.54.4.802 · Cited by 149 · cited 1×
“Intraclass correlation coefficient considers both the withinsubject variance originating from measurement error and variance originating from the difference between subjects.(19) The ICC is the fraction of the total variance that is attributed to the subjects and is calculated as:”
BackgroundPinker, K., Bogner, W., Baltzer, P., Gruber, S., Bickel, H., Brueck, B., Trattnig, S., Weber, M., Dubsky, P. C., Bagó-Horváth, Z., et al. (2014) · Investigative Radiology · doi:10.1097/rli.0000000000000029 · Cited by 144
Recommendations towards standards for quantitative MRI (qMRI) and outstanding needs
Keenan, K. E., Biller, J. R., Delfino, J. G., Boss, M. A., Does, M. D., Evelhoch, J. L., Griswold, M. A., Gunter, J. L., Hinks, R. S., Hoffman, S. W., et al. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26598 · Cited by 137 · cited 1×
“Magnetic resonance imaging can extract a variety of quantitative tissue properties that include not only length and volume but also relaxation properties (T1, T2, and T2*), diffusion, perfusion, phase, fat fraction, temperature, tissue chemical properties (eg, spectroscopy and chemical exchange saturation transfer), a…”
BackgroundCohen, O., Huang, S., McMahon, M. T., Rosen, M. S., & Farrar, C. T. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.27221 · Cited by 135
Physical and numerical phantoms for the validation of brain microstructural MRI: A cookbook
Fieremans, E., & Lee, H. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2018.06.046 · Cited by 135 · cited 1×
“However, simulations often do not match measurements in vivo due to the negligence of biological effects different than those being simulated.(67)”
BackgroundNew and Current Clinical Imaging Techniques to Study Chronic Obstructive Pulmonary Disease
Coxson, H. O., Mayo, J. R., Lam, S. C., Santyr, G. E., Párraga, G., & Sin, D. D. (2009) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200901-0159pp · Cited by 129 · cited 1×
“For both clinical and academic purposes, several commercially available and proprietary software and visual scoring systems have been adopted for the CT-based assessment of pulmonary emphysema.(168,177) Two major approaches have been reported for the quantitative assessment of COPD.”
MethodMagnetic resonance fingerprinting: a technical review
Mehta, B. B., Coppo, S., McGivney, D., Hamilton, J., Chen, Y., Jiang, Y., Ma, D., Seiberlich, N., Gulani, V., & Griswold, M. A. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.27403 · Cited by 128 · cited 1×
“Two major approaches are quantitative synthetic MRI(102) and MR fingerprinting (MRF).(115) Quantitative synthetic MRI is commonly performed by a 2D multidynamic multiecho (MDME) sequence, which is a turbo spin-echo sequence typically performed with 4 delay times and 2 echo times in the brain so that the scan time is c…”
BackgroundKalpathy-Cramer, J., Mamomov, A., Zhao, B., Lu, L., Cherezov, D., Napel, S., Echegaray, S., Rubin, D., McNitt-Gray, M., Lo, P., et al. (2016) · Tomography · doi:10.18383/j.tom.2016.00235 · Cited by 127 · cited 1×
“In general, a lack of reproducibility in radiomic features is a limitation for radiomics to be widely used in clinical practice.(199,203,204) The stability of radiomic features is sensitive, at various degrees, to a number of processing factors, including image acquisition parameters, reconstruction algorithms, digita…”
BackgroundFat fraction mapping using magnetic resonance imaging: insight into pathophysiology
Bray, T., Chouhan, M. D., Punwani, S., Bainbridge, A. P., & Hall-Craggs, M. A. (2017) · British Journal of Radiology · doi:10.1259/bjr.20170344 · Cited by 125 · cited 1×
“Proton density fat fraction is vulnerable to B0 field nonuniformity because differentiation between the phase shifts, due to B0 nonuniformity and those due to chemical shift utilized for extracting fat signal, is difficult.(59)”
BackgroundHagiwara, A., Hori, M., Cohen‐Adad, J., Nakazawa, M., Suzuki, Y., Kasahara, A., Horita, M., Haruyama, T., Andica, C., Maekawa, T., et al. (2018) · Investigative Radiology · doi:10.1097/rli.0000000000000510 · Cited by 122 · cited 1×
“The performance of the MDME sequence was examined on 3 scanners from 3 different vendors.(117) The highest intrascanner wCVs for T1, T2, and”
MethodSelf-citationLiao, C., Bilgiç, B., Manhard, M. K., Zhao, B., Cao, X., Zhong, J., Wald, L. L., & Setsompop, K. (2017) · NeuroImage · doi:10.1016/j.neuroimage.2017.08.030 · Cited by 122 · cited 1×
“High-resolution 3DMRFwith the resolution of 1 mm isovoxelwas also proposed with a scan time less than 8 minutes for full brain coverage.(147) Kato et al(148) investigated FISP-based MRF with B1 + correction and T1 and T2 measurements on the ISMRM/NIST phantom scanned for 100 days and showed high repeatability with a C…”
ResultReproducibility and Repeatability of MR Fingerprinting Relaxometry in the Human Brain
Koerzdoerfer, G., Kirsch, R., Liu, K., Pfeuffer, J., Hensel, B., Jiang, Y., Ma, D., Gratz, M., Bär, P., Bogner, W., et al. (2019) · Radiology · doi:10.1148/radiol.2019182360 · Cited by 119
Measurement and correction of transmitter and receiver induced nonuniformities in vivo
Wang, J., Qiu, M., Yang, Q., Smith, M. B. H., & Constable, R. T. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20354 · Cited by 118
Modeling the Presence of Myelin and Edema in the Brain Based on Multi-Parametric Quantitative MRI
Warntjes, M. J. B., Engström, M., Tisell, A., & Lundberg, P. (2016) · Frontiers in Neurology · doi:10.3389/fneur.2016.00016 · Cited by 114
Multiparametric estimation of brain hemodynamics with MR fingerprinting ASL
Su, P., Mao, D., Liu, P., Li, Y., Pinho, M. C., Welch, B. G., & Lu, H. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26587 · Cited by 112
The Lung Image Database Consortium (LIDC)
Reeves, A. P., Biancardi, A., Apanasovich, T. V., Meyer, C. R., MacMahon, H., van Beek, E. J. R., Kazerooni, E. A., Yankelevitz, D. F., McNitt‐Gray, M. F., McLennan, G., et al. (2007) · Academic Radiology · doi:10.1016/j.acra.2007.09.005 · Cited by 108 · cited 1×
“Specialist markings (eg, setting a boundary of a region for volumetry) are also sometimes considered as the true value, but they often have a variable degree of interreader variation and should be considered as a reference standard.(39)”
BackgroundImaging biomarker analysis of advanced multiparametric MRI for glioma grading
Vamvakas, A., Williams, S., Theodorou, K., Kapsalaki, E. Z., Fountas, K. N., Kappas, C., Vassiou, K. G., & Tsougos, I. (2019) · Physica Medica · doi:10.1016/j.ejmp.2019.03.014 · Cited by 103
Christen, T., Pannetier, N. A., Ni, W. W., Qiu, D., Moseley, M. E., Schuff, N., & Zaharchuk, G. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2013.11.052 · Cited by 97 · cited 3×
“Magnetic resonance fingerprinting can measure any property that can be simulated by the Bloch equation, for example, and recent works have also incorporated the measurements of B1 ,(132,133) T2*,(134) magnetization transfer,(135) amide,(136) spectroscopy, perfusion,(137) andmicrovascular characteristics into the MRF.(…”
BackgroundInfluentialMagnetization transfer in magnetic resonance fingerprinting
Hilbert, T., Xia, D., Block, K. T., Yu, Z., Lattanzi, R., Sodickson, D. K., Kober, T., & Cloos, M. A. (2019) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28096 · Cited by 97
Whittall, K. P., MacKay, A. L., & Li, D. K. B. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199906)41:6<1255::aid-mrm23>3.0.co;2-i · Cited by 96 · cited 1×
“By changing the echo time, T2 relaxation time can be measured by spin echo technique with as few as 2 measurements— assuming monoexponential decay; however, the measurement suffers from partial volume and is susceptible to noise.(106) Further, the acquisition time is very long because full T1 relaxation is required du…”
BackgroundQuantitative Imaging Test Approval and Biomarker Qualification: Interrelated but Distinct Activities
Buckler, A. J., Bresolin, L., Dunnick, N. R., Sullivan, D. C., & For the Group (2011) · Radiology · doi:10.1148/radiol.10100800 · Cited by 94 · cited 1×
“For example, if a QIB is qualified by the Food andDrug Administration for drug development, it could help deliver a new therapy to the public through either a traditional or accelerated approval pathway.(10) In addition to the clinical relevance and sensitivity to the disease process, good reproducibility is the key e…”
BackgroundHagiwara, A., Otsuka, Y., Hori, M., Tachibana, Y., Yokoyama, K., Fujita, S., Andica, C., Kamagata, K., Irie, R., Koshino, S., et al. (2019) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5927 · Cited by 93 · cited 1×
“Although the image quality of synthetic FLAIR is generally perceived to be inferior to FLAIR acquired by conventional methods, improvement of synthetic FLAIR quality by deep learning has also been suggested.(124) Hence, relaxometry data derived from MDME may become an adjunct to contrast-weighted images in clinical se…”
BackgroundSelf-citationBelli, G., Busoni, S., Ciccarone, A., Coniglio, A. D., Esposito, M., Giannelli, M., Mazzoni, L. N., Nocetti, L., Sghedoni, R., Tarducci, R., et al. (2015) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24956 · Cited by 92
Kumamaru, K. K., Fujimoto, S., Otsuka, Y., Kawasaki, T., Kawaguchi, Y. O., Kato, E., Takamura, K., Aoshima, C., Kamo, Y., Kogure, Y., et al. (2019) · European Heart Journal - Cardiovascular Imaging · doi:10.1093/ehjci/jez160 · Cited by 88 · cited 1×
“For example, it can predict the functional flow reserve from cardiac CT data by point estimatioin.(226) The major advantages of AI approaches over manual approaches in terms of decreasing variability in QIB are as follows: (1) no variance is caused by fatigue as in human analysts, and (2) AI returns consistent results…”
BackgroundSelf-citationMultiparametric Magnetic Resonance Imaging in Prostate Cancer Management
Scheenen, T. W. J., Rosenkrantz, A. B., Haider, M. A., & Fütterer, J. J. (2015) · Investigative Radiology · doi:10.1097/rli.0000000000000163 · Cited by 88
Jambor, I., Merisaari, H., Taimen, P., Boström, P. J., Minn, H. R. I., Pesola, M., & Aronen, H. J. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25323 · Cited by 87
A Phantom for diffusion‐weighted imaging of acute stroke
Laubach, H., Jakob, P. M., Lövblad, K., Baird, A. E., Bovo, M. P., Edelman, R. R., & Warach, S. J. (1998) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1880080627 · Cited by 86
Fujita, S., Hagiwara, A., Hori, M., Warntjes, M. J. B., Kamagata, K., Fukunaga, I., Andica, C., Maekawa, T., Irie, R., Takemura, M. Y., et al. (2019) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2019.08.031 · Cited by 86 · cited 1×
“The 3-dimensional (3D) version of the MDME, namely, 3D-QALAS (3D-quantification using an interleaved LL acquisition sequence with T2-prep pulse), was recently developed for the heart(127) and has also been applied to the brain.(128,129) SyntheticMR angiography constructed by deep learning is also feasible based on the…”
MethodSelf-citationPohl, K. M., Sullivan, E. V., Rohlfing, T., Chu, W., Kwon, D., Nichols, B. N., Zhang, Y., Brown, S. A., Tapert, S. F., Cummins, K., et al. (2016) · NeuroImage · doi:10.1016/j.neuroimage.2016.01.061 · Cited by 85
Correction of systematic errors in quantitative proton density mapping
Volz, S., Nöth, U., & Deichmann, R. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23206 · Cited by 85 · cited 1×
“However, the reciprocity principle becomes less accurate at a field strength of 3 T or higher.(56,57) B1 − nonuniformity affects the spatial distribution of image intensity and thereby any quantitativeMRI, especially the measurement of proton density and absolute metabolite concentration.”
BackgroundBadve, C., Yu, A., Rogers, M., Ma, D., Liu, Y., Schluchter, M., Sunshine, J., Griswold, M., & Gulani, V. (2015) · Tomography · doi:10.18383/j.tom.2015.00166 · Cited by 82
Hoffman, E. A., Lynch, D. A., Barr, R. G., van Beek, E. J. R., Párraga, G., & IWPFI Investigators (2015) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25010 · Cited by 78 · cited 1×
“For both clinical and academic purposes, several commercially available and proprietary software and visual scoring systems have been adopted for the CT-based assessment of pulmonary emphysema.(168,177) Two major approaches have been reported for the quantitative assessment of COPD.”
MethodFedorov, A., Vangel, M. G., Tempany, C. M., & Fennessy, F. M. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000382 · Cited by 76
History of Research in Medical Image Perception
Kundel, H. L. (2006) · Journal of the American College of Radiology · doi:10.1016/j.jacr.2006.02.023 · Cited by 76 · cited 1×
“This is partly because the extraction of most QIBs from radiological images is not yet fully automated, and it requires a radiologist or other experienced practitioner to engage in the analysis process, which introduces an inevitable variability arising from human perception.(12) Further, the variability of a QIB is a…”
BackgroundDesign of a breast phantom for quantitative MRI
Keenan, K. E., Wilmes, L. J., Aliu, S. O., Newitt, D. C., Jones, E. F., Boss, M. A., Stupic, K. F., Russek, S. E., & Hylton, N. M. (2016) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25214 · Cited by 74 · cited 1×
“For example, the breast phantoms were developed partly because the previous phantoms were not physically compatible with a breast coil.(64) The properties of the phantom may vary over time”
BackgroundHagiwara, A., Hori, M., Yokoyama, K., Takemura, M. Y., Andica, C., Kumamaru, K. K., Nakazawa, M., Takano, N., Kawasaki, H., Sato, S., et al. (2016) · American Journal of Neuroradiology · doi:10.3174/ajnr.a4977 · Cited by 73
Self-citationHagiwara, A., Kamagata, K., Shimoji, K., Yokoyama, K., Andica, C., Hori, M., Fujita, S., Maekawa, T., Irie, R., Akashi, T., et al. (2019) · American Journal of Neuroradiology · doi:10.3174/ajnr.a6209 · Cited by 72
Self-citationHeijmen, L., ter Voert, E. E. G. W., Nagtegaal, I. D., Span, P. N., Bussink, J., Punt, C. J. A., de Wilt, J. H. W., Sweep, F. C. G. J., Heerschap, A., & van Laarhoven, H. W. M. (2012) · European Radiology · doi:10.1007/s00330-012-2654-4 · Cited by 72
Perkuhn, M., Stavrinou, P., Thiele, F., Shakirin, G., Mohan, M., Garmpis, D., Kabbasch, C., & Borggrefe, J. (2018) · Investigative Radiology · doi:10.1097/rli.0000000000000484 · Cited by 71 · cited 1×
“Automated techniques using deep learning trained with a large dataset is expected to reduce the variability in tasks such as tumor segmentation.(35)”
MethodEchegaray, S., Gevaert, O., Shah, R. P., Kamaya, A., Louie, J. D., Kothary, N., & Napel, S. A. (2015) · Journal of medical imaging · doi:10.1117/1.jmi.2.4.041011 · Cited by 70
Lee, S. M., Choi, Y. H., You, S. K., Lee, W. K., Kim, W. H., Kim, H. J., Lee, S. Y., & Cheon, H. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000435 · Cited by 70
Fujita, S., Hagiwara, A., Otsuka, Y., Hori, M., Takei, N., Hwang, K., Irie, R., Andica, C., Kamagata, K., Akashi, T., et al. (2020) · Investigative Radiology · doi:10.1097/rli.0000000000000628 · Cited by 69 · cited 1×
“SyntheticMR angiography constructed by deep learning is also feasible based on the 3D-QALAS data of high resolution.(130) Another promising approach of simultaneous relaxometry is MRF.”
MethodSelf-citationPanda, A., O’Connor, G., Lo, W. C., Jiang, Y., Margevicius, S. P., SCHLUCHTER, M. D., Ponsky, L. E., & Gulani, V. (2019) · Investigative Radiology · doi:10.1097/rli.0000000000000569 · Cited by 68 · cited 1×
“One possible approach for implementing multiparametric MRI into clinical practice is setting a cutoff value for each parameter.(146,153) Pinker et al(154) evaluated the diagnostic accuracy of contrast-enhancedMRI, DWI, and MR spectroscopy in combinations of 2 or 3 and used a single measure by setting a threshold for e…”
MethodA simple correction for B1 field errors in magnetization transfer ratio measurements
Samson, R. S., Wheeler‐Kingshott, C. A. M. G., Symms, M. R., Tozer, D. J., & Tofts, P. S. (2005) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2005.10.025 · Cited by 67
Developing a quality control protocol for diffusion imaging on a clinical MRI system
Delakis, I., Moore, E. M., Leach, M. O., & Wilde, J. P. D. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/8/003 · Cited by 65
NCTN Assessment on Current Applications of Radiomics in Oncology
Nie, K., Al‐Hallaq, H., Li, X. A., Benedict, S., Sohn, J. W., Moran, J. M., Fan, Y., Huang, M., Knopp, M. V., Michalski, J. M., et al. (2019) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2019.01.087 · Cited by 64 · cited 2×
“The NRG Oncology investigators have provided recommendations and a guideline specifically for use in the National Clinical Trials Network.216 They suggest that the radiomics quality score217,218 may serve to evaluate the quality of radiomics studies.”
MethodRepeatability of echo-planar-based diffusion measurements of the human prostate at 3 T
Gibbs, P., Pickles, M. D., & Turnbull, L. W. (2007) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2007.03.030 · Cited by 64
Nawa, T., Nakagawa, T., Mizoue, T., Kusano, S., Chonan, T., Hayashihara, K., Suito, T., & Endo, K. (2012) · Lung Cancer · doi:10.1016/j.lungcan.2012.09.012 · Cited by 61
Intensity warping for multisite MRI harmonization
Wrobel, J., Martin, M. L., Bakshi, R., Calabresi, P. A., Elliot, M., Roalf, D. R., Gur, R. C., Gur, R. E., Henry, R. G., Nair, G., et al. (2020) · NeuroImage · doi:10.1016/j.neuroimage.2020.117242 · Cited by 60
Brain microstructure by multi-modal MRI: Is the whole greater than the sum of its parts?
Cercignani, M., & Bouyagoub, S. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2017.10.052 · Cited by 60 · cited 1×
“Multiparametric quantitative MRI can also be used for extracting new measures that reflect subvoxel microstructural information such as myelin and axon density, axon diameter, and membrane permeability.(152,159) Geometric distortion of images, image misregistration, and different interpolation techniques will introduc…”
MethodSorace, A. G., Wu, C., Barnes, S. L., Jarrett, A. M., Avery, S., Patt, D. A., Goodgame, B. W., Luci, J., Kang, H., Abramson, R. G., et al. (2018) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26011 · Cited by 59
A Practical Guide to Artificial Intelligence–Based Image Analysis in Radiology
Weikert, T., Cyriac, J., Yang, S., Nešić, I., Parmar, V., & Stieltjes, B. (2019) · Investigative Radiology · doi:10.1097/rli.0000000000000600 · Cited by 59 · cited 2×
“Artificial intelligence, including machine learning and deep learning, has been increasingly applied to medical imaging.(219) Promising results have been shown in various tasks related to radiological images, such as the detection of lesions,(220) segmentation (eg, labeling organs),(221) classification (eg, pneumonia…”
BackgroundMethodKobayashi, Y., & Terada, Y. (2018) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.tn.2018-0027 · Cited by 59 · cited 1×
“Larger variations of T2 were likely attributed to scanner imperfections related to certain system characteristics, such as different eddy current behaviors and the diffusion effect.(132)”
BackgroundHuang, E. P., Wang, X. F., Choudhury, K. R., McShane, L. M., Gönen, M., Ye, J., Buckler, A. J., Kinahan, P. E., Reeves, A. P., Jackson, E. F., et al. (2014) · Statistical Methods in Medical Research · doi:10.1177/0962280214537394 · Cited by 58 · cited 1×
“Ideally, this evaluation should involve summaries frommultiple studies to overcome any limitations arising from a small sample size (typically 10 to 20 subjects) of a study concerning technical performance and include a wider range of relevant clinical settings and patient populations.(28) Although a meta-analysis of…”
BackgroundWagner, F. E., Laun, F. B., Kuder, T. A., Mlynarska-Bujny, A., Maier, F. C., Faust, J. F., Demberg, K., Lindemann, L., Rivkin, B. B., Nagel, A. M., et al. (2017) · PLoS ONE · doi:10.1371/journal.pone.0179276 · Cited by 58 · cited 1×
“1 10−3 mm(2)/s, or alternatively, polyvinylpyrrolidone solutions with a range of ADC values.(99)”
BackgroundEstablishing intra‐ and inter‐vendor reproducibility of T1 relaxation time measurements with 3T MRI
Lee, Y., Callaghan, M. F., Acosta‐Cabronero, J., Lutti, A., & Nagy, Z. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.27421 · Cited by 58
Stocker, D., Manoliu, A., Becker, A. S., Barth, B. K., Nanz, D., Klarhöfer, M., & Donati, O. F. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000429 · Cited by 58
Jafar, M. M., Parsaï, A., & Miquel, M. E. (2016) · World Journal of Radiology · doi:10.4329/wjr.v8.i1.21 · Cited by 57
Prognostic Importance of Volumetric Measurements in Stage I Lung Adenocarcinoma
Yanagawa, M., Tanaka, Y., Leung, A. N. C., Morii, E., Kusumoto, M., WATANABE, S.-I. I., Watanabe, H., Inoue, M., Okumura, M., Gyobu, T., et al. (2014) · Radiology · doi:10.1148/radiol.14131903 · Cited by 55
Eshaghi, A., Riyahi-Alam, S., Saeedi, R., Roostaei, T., Nazeri, A., Aghsaei, A., Doosti, R., Ganjgahi, H., Bodini, B., Shakourirad, A., et al. (2015) · NeuroImage Clinical · doi:10.1016/j.nicl.2015.01.001 · Cited by 53
Essentials of Statistical Methods for Assessing Reliability and Agreement in Quantitative Imaging
Anvari, A., Halpern, E. F., & Samir, A. E. (2017) · Academic Radiology · doi:10.1016/j.acra.2017.09.010 · Cited by 53 · cited 4×
“The size of the region of interest (ROI) has been known to affect the repeatability and reproducibility of QIBs.29 AQIB is oftenmeasured as themeanvalue of amapwithin an ROI, where an increase in the size of the ROI leads to a smaller variance (ie, higher repeatability/reproducibility) of the measurement.”
BackgroundInfluentialExploring the sensitivity of magnetic resonance fingerprinting to motion
Yu, Z., Zhao, T., Assländer, J., Lattanzi, R., Sodickson, D. K., & Cloos, M. A. (2018) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2018.09.002 · Cited by 53 · cited 1×
“The effect of motion on the resulting image is also small as long as the errors are incoherent in such away that pattern-matching is still possible; however, MRF is known to be more vulnerable to through-planemotion than to in-plane motion.(140)”
BackgroundOhno, Y., Yaguchi, A., Okazaki, T., Aoyagi, K., Yamagata, H., Sugihara, N., Koyama, H., Yoshikawa, T., & Sugimura, K. (2016) · European Journal of Radiology · doi:10.1016/j.ejrad.2016.05.001 · Cited by 53
Self-citationRadiation dose reduction techniques for chest CT: Principles and clinical results
Ohno, Y., Koyama, H., Seki, S., Kishida, Y., & Yoshikawa, T. (2018) · European Journal of Radiology · doi:10.1016/j.ejrad.2018.12.017 · Cited by 53 · cited 1×
“In addition, newly developed iterative reconstruction (IR) methods have been introduced and applied in routine clinical practice.(171,172) In fact, dose reduction strategies have been realized by employing a variety of techniques for data acquisition, such as tube current reduction, tube voltage reduction, increased h…”
MethodSelf-citationReproducibility of Structural and Diffusion Tensor Imaging in the TACERN Multi-Center Study
Prohl, A. K., Scherrer, B., Tomas-Fernandez, X., Filip‐Dhima, R., Kapur, K., Velasco‐Annis, C., Clancy, S., Carmody, E., Dean, M., Valle, M., et al. (2019) · Frontiers in Integrative Neuroscience · doi:10.3389/fnint.2019.00024 · Cited by 52 · cited 1×
“Reproducibility is especially important in multicenter studies where reproduction of the same measurement is required across different centers and often with different kinds of scanners.(24) Multicenter studies of human subjects enable a comprehensive investigation of the disease.”
BackgroundHagiwara, A., Hori, M., Yokoyama, K., Nakazawa, M., Ueda, R., Horita, M., Andica, C., Abe, O., & Aoki, S. (2017) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5312 · Cited by 49
Self-citationFujita, S., Hagiwara, A., Hori, M., Warntjes, M. J. B., Kamagata, K., Fukunaga, I., Goto, M., Haruyama, T., Takasu, K., Andica, C., et al. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26744 · Cited by 49 · cited 1×
“The 3-dimensional (3D) version of the MDME, namely, 3D-QALAS (3D-quantification using an interleaved LL acquisition sequence with T2-prep pulse), was recently developed for the heart(127) and has also been applied to the brain.(128,129) SyntheticMR angiography constructed by deep learning is also feasible based on the…”
MethodSelf-citationTofts, P. S., Steens, S. C. A., Cercignani, M., Admiraal-Behloul, F., Hofman, P. A. M., van Osch, M. J. P., Teeuwisse, W. M., Tozer, D. J., van Waesberghe, J. H. T. M., Yeung, R., et al. (2006) · Magnetic Resonance Materials in Physics, Biology and Medicine · doi:10.1007/s10334-006-0049-8 · Cited by 46 · cited 1×
“Body coil excitation is preferable for uniform transfmission.(26) Calibration of the transmitter output can be carried out periodically as part of routine maintenance.”
BackgroundBerre, A. L., Kamagata, K., Otsuka, Y., Andica, C., Hatano, T., Saccenti, L., Ogawa, T., Takeshige-Amano, H., Wada, A., Suzuki, M., et al. (2019) · Neuroradiology · doi:10.1007/s00234-019-02279-w · Cited by 46
Self-citationAndica, C., Hagiwara, A., Hori, M., Nakazawa, M., Goto, M., Koshino, S., Kamagata, K., Kumamaru, K. K., & Aoki, S. (2017) · Journal of Neuroradiology · doi:10.1016/j.neurad.2017.10.002 · Cited by 44 · cited 1×
“The effect of variability in in-plane resolution on the volumetry based on the MDME data was found to be little in healthy volunteers and MS patients, presumably because the segmentation algorithm considers tissue partial volumes in the interval of 0% to 100% rather than assigning a single tissue type to each voxel.(1…”
MethodSelf-citationPetrick, N., Kim, H. J. G., Clunie, D., Borradaile, K., Ford, R., Zeng, R., Gavrielides, M. A., McNitt‐Gray, M. F., Lu, Z., Fenimore, C., et al. (2013) · Academic Radiology · doi:10.1016/j.acra.2013.09.020 · Cited by 44 · cited 1×
“For example, bias for tumors with different sizes, shapes, and densities can be reported as a bias profile for CT volumetry.(17) Inconstant biases should be specified cautiously, especially when assessment of change in the QIB is the focus; this is because the different biases do not cancel out in calculating the chan…”
BackgroundJambor, I., Merisaari, H., Aronen, H. J., Järvinen, J., Saunavaara, J. P., Kauko, T., Borra, R., & Pesola, M. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24271 · Cited by 40
Multicentre imaging measurements for oncology and in the brain
Tofts, P. S., & Collins, D. J. (2011) · British Journal of Radiology · doi:10.1259/bjr/74316620 · Cited by 39 · cited 5×
“In addition to the clinical relevance and sensitivity to the disease process, good reproducibility is the key element of a qualified biomarker.(11) Although QIBs can be used similarly to laboratory assays, its clinical application has been hindered by its generally lower reproducibility.”
BackgroundInfluentialMulti-Site Harmonization of Diffusion MRI Data via Method of Moments
Huynh, K. M., Chen, G., Wu, Y., Shen, D. G., & Yap, P. (2019) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2019.2895020 · Cited by 38
Tunali, I., Stringfield, O., Güveniş, A., Wang, H., Liu, Y., Balagurunathan, Y., Lambin, P., Gillies, R. J., & Schabath, M. B. (2017) · Oncotarget · doi:10.18632/oncotarget.21629 · Cited by 36
MR fingerprinting using the quick echo splitting NMR imaging technique
Jiang, Y., Ma, D., Jerec̆ić, R., Duerk, J. L., Seiberlich, N., Gulani, V., & Griswold, M. A. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26173 · Cited by 36 · cited 1×
“Sequence design is flexible in MRF, and several sequence designs have been used, including balanced steady-state free precession,(131) fast imaging with steady-state precession (FISP),(142) RF-spoiled gradient echo, and quick echo splitting nuclear magnetic resonance.(143) The MRF has been primarily investigated for b…”
MethodPaldino, M. J., Barboriak, D. P., Desjardins, A., Friedman, H. S., & Vredenburgh, J. J. (2009) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21732 · Cited by 36
Statistical analysis of lung nodule volume measurements with CT in a large‐scale phantom study
Li, Q., Gavrielides, M. A., Sahiner, B., Myers, K. J., Zeng, R., & Petrick, N. (2015) · Medical Physics · doi:10.1118/1.4921734 · Cited by 35
Saccenti, L., Andica, C., Hagiwara, A., Yokoyama, K., Takemura, M. Y., Fujita, S., Maekawa, T., Kamagata, K., Berre, A. L., Hori, M., et al. (2019) · Neuroradiology · doi:10.1007/s00234-019-02241-w · Cited by 35 · cited 1×
“The effect of variability in in-plane resolution on the volumetry based on the MDME data was found to be little in healthy volunteers and MS patients, presumably because the segmentation algorithm considers tissue partial volumes in the interval of 0% to 100% rather than assigning a single tissue type to each voxel.(1…”
MethodSelf-citationAndica, C., Hori, M., Kamiya, K., Koshino, S., Hagiwara, A., Kamagata, K., Fukunaga, I., Hamasaki, N., Suzuki, M., Feiweier, T., et al. (2017) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.cr.2017-0111 · Cited by 35
Self-citationEffect of spiral undersampling patterns on FISP MRF parameter maps
Koerzdoerfer, G., Pfeuffer, J., Kluge, T., Gebhardt, M., Hensel, B., Meyer, C. H., & Nittka, M. (2019) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2019.01.011 · Cited by 34 · cited 1×
“The pattern matching can be performed even in the presence of undersampling artifacts; hence, the scan can be highly accelerated to reduce the scan time.(139) The effect of motion on the resulting image is also small as long as the errors are incoherent in such away that pattern-matching is still possible; however, MR…”
BackgroundAndica, C., Hagiwara, A., Kamagata, K., Yokoyama, K., Shimoji, K., Saito, A., Takenaka, Y., Nakazawa, M., Hori, M., Cohen‐Adad, J., et al. (2019) · Scientific Reports · doi:10.1038/s41598-019-44615-3 · Cited by 34
Self-citationMulshine, J. L., Gierada, D. S., Armato, S. G., Avila, R., Yankelevitz, D. F., Kazerooni, E. A., McNitt‐Gray, M. F., Buckler, A. J., & Sullivan, D. C. (2015) · Journal of the American College of Radiology · doi:10.1016/j.jacr.2014.12.003 · Cited by 33 · cited 1×
“Further, in the last decade, some societies have considered the application of quantitative CT-based indexes as QIBs for patient management, including therapeutic planning and treatment response assessment, and have worked on standardizing the CT protocols for quantitative assessment of CT-based indexes.(160,161) In a…”
Deckers, F., Foer, B. D., Mieghem, F. V., Botelberge, T., Weytjens, R., Padhani, A. R., & Pouillon, M. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24359 · Cited by 33
Hori, M., Hagiwara, A., Fukunaga, I., Ueda, R., Kamiya, K., Suzuki, Y., Liu, W., Murata, K., Takamura, T., Hamasaki, N., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-23527-8 · Cited by 32 · cited 1×
“Multiparametric quantitative MRI can also be used for extracting new measures that reflect subvoxel microstructural information such as myelin and axon density, axon diameter, and membrane permeability.(152,159) Geometric distortion of images, image misregistration, and different interpolation techniques will introduc…”
MethodSelf-citationRapid multicomponent relaxometry in steady state with correction of magnetization transfer effects
Liu, F., Block, W. F., Kijowski, R., & Samsonov, A. A. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25672 · Cited by 32 · cited 1×
“T2 measurements are also affected by magnetization transfer effects between the water and macromolecular protons, resulting in diminished signal in the free water and inaccurate T2 measurement.(111,112) Similar to the discrepancy among T1 measurement methods, T2 measurement methods are known to show disagreement with…”
BackgroundMishima, M., Oku, Y., Kawakami, K., Sakai, N., Fukui, M., Hirai, T., Chin, K., Ohi, M., Nishimura, K., Itoh, H., et al. (1997) · PubMed · Cited by 32
Assessing effects of scanner upgrades for clinical studies
Keenan, K. E., Gimbutas, Z., Dienstfrey, A., & Stupic, K. F. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26785 · Cited by 31 · cited 5×
“Hardware and scanner software upgrades may introduce more bias and/or less precision in the derived QIB.(36) For example, Keenan et al(36) showed that the variable flip angle (VFA) T1 measurements on upgraded systems (hardware and software) had an overestimation of approximately 18% compared with the measurements of t…”
ResultInfluentialThe Quantitative Imaging Network in Precision Medicine
Nordstrom, R. J. (2016) · Tomography · doi:10.18383/j.tom.2016.00190 · Cited by 31
Diffusion-Weighted MRI for Assessment of Early Cancer Treatment Response
Galbán, S., Brisset, J., Rehemtulla, A., Chenevert, T. L., Ross, B. D., & Galbán, C. J. (2010) · Current Pharmaceutical Biotechnology · doi:10.2174/138920110792246627 · Cited by 31 · cited 1×
“The presumption for using ADC in clinical practice for managing tumors is that treatment-associated change in the microenvironment precedes changes in the lesion size, thereby encouraging the use of ADC as a biomarker of treatment response.(78) Conformance to the specifications of the QIBA DWI Profile(13) by all relev…”
MethodMeasuring water content using T2 relaxation at 3T: Phantom validations and simulations
Meyers, S. M., Kolind, S., Laule, C., & MacKay, A. L. (2015) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2015.11.008 · Cited by 29 · cited 1×
“Calculation of PD is based on the estimation of the magnetization at equilibrium (M0), which represents the signal intensity in the absence of any relaxation.(103) T1 relaxation time characterizes the approach of the polarized spins to equilibrium in the direction of the external magnetic field, and it is affected by…”
BackgroundMulti‐parametric T2* magnetic resonance fingerprinting using variable echo times
Wyatt, C. R., Smith, T., Sammi, M. K., Rooney, W. D., & Guimarães, A. R. (2018) · NMR in Biomedicine · doi:10.1002/nbm.3951 · Cited by 26
Wallaert, L., Hagiwara, A., Andica, C., Hori, M., Yamashiro, K., Koshino, S., Maekawa, T., Kamagata, K., & Aoki, S. (2017) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.ci.2017-0110 · Cited by 26 · cited 1×
“Synthetic MRI enables the creation of clinically used contrast-weighted images including T1-weighted, T2-weighted, and fluid-attenuated inversion recovery (FLAIR) images based on the T1, T2, and PD maps.(123) Although the image quality of synthetic FLAIR is generally perceived to be inferior to FLAIR acquired by conve…”
MethodSelf-citationChoroid plexus cysts analyzed using diffusion-weighted imaging with short diffusion-time
Maekawa, T., Hori, M., Murata, K., Feiweier, T., Andica, C., Fukunaga, I., Koshino, S., Hagiwara, A., Kamiya, K., Kamagata, K., et al. (2018) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2018.12.010 · Cited by 25
Self-citationEffect of Region of Interest Size on the Repeatability of Quantitative Brain Imaging Biomarkers
Jafari‐Khouzani, K., Paynabar, K., Hajighasemi, F., & Rosen, B. R. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2860928 · Cited by 24 · cited 4×
“The size of the region of interest (ROI) has been known to affect the repeatability and reproducibility of QIBs.(29) AQIB is oftenmeasured as themeanvalue of amapwithin an ROI, where an increase in the size of the ROI leads to a smaller variance (ie, higher repeatability/reproducibility) of the measurement.”
BackgroundMethodInfluentialKato, Y., Ichikawa, K., Okudaira, K., Taoka, T., Kawaguchi, H., Murata, K., Maruyama, K., Koerzdoerfer, G., Pfeuffer, J., Nittka, M., et al. (2019) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.mp.2019-0016 · Cited by 23
Andica, C., Hagiwara, A., Hori, M., Haruyama, T., Fujita, S., Maekawa, T., Kamagata, K., Yoshida, M., Suzuki, M., Sugano, H., et al. (2019) · Neuroradiology · doi:10.1007/s00234-019-02250-9 · Cited by 23 · cited 1×
“Estimating the myelin volume fraction in each voxel by using a 4-compartment model based on the MDME data has also been suggested(118) and applied to diseases such as multiple sclerosis(33,119) and Sturge-Weber syndrome.(120) For applications in other organs, the sequence may have to be adjusted to each target tissue.”
MethodSelf-citationBayesian Generative Models for Knowledge Transfer in MRI Semantic Segmentation Problems
Kuzina, A., Egorov, E., & Burnaev, E. V. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00844 · Cited by 21 · cited 1×
“In relation to standardization of QIB, an AI algorithm that automates the process of QIB extraction has the capability to decrease variability, such as through an automated pipeline that can reduce ambiguity and variability in lesion segmentation.(225) Extracting a QIB using AI in a fully automated manner is also feas…”
MethodMaekawa, T., Hori, M., Murata, K., Feiweier, T., Fukunaga, I., Andica, C., Hagiwara, A., Kamagata, K., Koshino, S., Abe, O., et al. (2018) · Japanese Journal of Radiology · doi:10.1007/s11604-018-0737-0 · Cited by 20
Self-citationSagawa, H., Kataoka, M., Kanao, S., et al. (2018) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.mp.2018-0015 · Cited by 19 · cited 1×
“Repetitive sampling is also much easier for imaging than a biopsy, and imaging data can be dynamically obtained in some cases in the order of seconds to milliseconds.(6,7) In addition, QIBs may enable the detection of a subclinical presentation of disease that is too subtle to be detected by human eyes(8); this leads…”
BackgroundHuo, J., Alger, J. R., Kim, H., Brown, M. S., Okada, K., Pope, W. B., & Goldin, J. G. (2015) · Clinical Neuroradiology · doi:10.1007/s00062-015-0381-3 · Cited by 17
Koyama, H., Ohno, Y., Nishio, M., Matsumoto, S., Sugihara, N., Yoshikawa, T., Seki, S., & Sugimura, K. (2014) · European Radiology · doi:10.1007/s00330-014-3207-9 · Cited by 17
Self-citationMagnetic Resonance Fingerprinting Reconstruction Using Recurrent Neural Networks
Hoppe, E., Thamm, F., Koerzdoerfer, G., Syben, C., Schirrmacher, F., Nittka, M., Pfeuffer, J., Meyer, H., & Maier, A. (2019) · Studies in health technology and informatics · doi:10.3233/shti190816 · Cited by 15 · cited 1×
“The pattern-matching process may benefit from deep learning in terms of both accuracy and speed.(141) The pattern-matching process is a distinctive factor of MRF in view of standardization because resulting maps are dependent on the structure of the dictionary (Fig.”
BackgroundKoyama, H., Ohno, Y., Nishio, M., Takenaka, D., Yoshikawa, T., Matsumoto, S., Nishimura, Y., & Sugimura, K. (2012) · British Journal of Radiology · doi:10.1259/bjr/22602417 · Cited by 15
Self-citationQuantitative quality assurance in a multicenter HARDI clinical trial at 3 T
Zhou, X., Sakaie, K., Debbins, J. P., KIRSCH, J. E., Tatsuoka, C. M., Fox, R. J., & Lowe, M. J. (2016) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2016.08.022 · Cited by 14 · cited 1×
“However, if reproducibility across scanners is low, variability across scanners may reduce the statistical power of detecting differences between groups and annul the benefit of using data frommultiple centers.(25)Mitigation of variation across centers can be achieved by (1) setting scanning and analysis procedures to…”
BackgroundOhno, Y., Fujisawa, Y., Fujii, K., Sugihara, N., Kishida, Y., Seki, S., & Yoshikawa, T. (2019) · Japanese Journal of Radiology · doi:10.1007/s11604-019-00823-5 · Cited by 14
Self-citationJeong, D. K., Malalis, C., Arrington, J. A., Field, A. S., Choi, J. W., & Koçak, M. (2015) · PubMed · doi:10.3978/j.issn.2223-4292.2015.12.05 · Cited by 14
Koyama, H., Ohno, Y., Yamazaki, Y., Onishi, Y., Takenaka, D., Yoshikawa, T., Nishio, M., Matsumoto, S., Murase, K., Nishimura, Y., et al. (2011) · European Journal of Radiology · doi:10.1016/j.ejrad.2010.12.042 · Cited by 11
Self-citationJutras, J., Wachowicz, K., & Zanche, N. D. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26074 · Cited by 7
Cited by
Artificial intelligence and machine learning in cancer imaging
Koh, D.-M., Papanikolaou, N., Bick, U., Illing, R., Kahn, C. E., Kalpathi-Cramer, J., Matos, C., Martí-Bonmatí, L., Miles, A., Mun, S. K., et al. (2022) · Communications Medicine · doi:10.1038/s43856-022-00199-0 · Cited by 311
Nam, D., Chapiro, J., Paradis, V., Seraphin, T. P., & Kather, J. N. (2022) · JHEP Reports · doi:10.1016/j.jhepr.2022.100443 · Cited by 226
Lassau, N., Ammari, S., Chouzenoux, E., Gortais, H., Herent, P., Devilder, M., Soliman, S., Meyrignac, O., Talabard, M.-P., Lamarque, J.-P., et al. (2021) · Nature Communications · doi:10.1038/s41467-020-20657-4 · Cited by 190
Belue, M. J., Harmon, S. A., Lay, N. S., Daryanani, A., Phelps, T. E., Choyke, P. L., & Turkbey, B. (2023) · Journal of the American College of Radiology · doi:10.1016/j.jacr.2022.05.022 · Cited by 27
Bhandari, A., Scott, L., Weilbach, M., Marwah, R., & Lasocki, A. (2023) · Neuroradiology · doi:10.1007/s00234-023-03126-9 · Cited by 21
Metadata
- Authors
- Hagiwara, A., Fujita, S., Ohno, Y., & Aoki, S.
- Venue
- Investigative Radiology
- Published
- 2020-03-24
- Volume
- 55
- Issue
- 9
- Pages
- 601-616
- ISSN
- 0020-9996 (ISSN-L)
- 1536-0210
- OpenAlex
- W3013522077
- Cited by
- 183
- References
- 197
- FWCI
- 9.97
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.