Battery-free, skin-interfaced microfluidic/electronic systems for simultaneous electrochemical, colorimetric, and volumetric analysis of sweat
Open accessThe DOI matches its Crossref record; 51 of the 51 references it cites were found in a registry or an index.
Abstract
Battery-free, wireless microfluidic/electronic system for multiparameter sweat analysis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures


![Fig. 3 Lactate sensor characterization. (A) Exploded-view schematic illustration of the layer makeup of the biofuel cell–based lactate sensor. LOx, lactate oxidase; TTF, tetrathiafulvalene. (B) Image of the lactate sensor. (C) Real-time sensor response to increasing lactate concentration in phosphate buffer (pH 7.0) at 25°C and (D) the corresponding calibration (n = 3). (E) Plot illustrating a reversible response for the lactate sensor for four consecutive cycles of varying lactate concentration. Inset shows calibration plot that compares the sensor signal in (E) for the four cycles. V, voltage in millivolts; C, concentration in millimolar. (F) Real-time data acquired with increasing lactate concentration in artificial sweat under common physiological sweat conditions [temperature, ~30°C (pH 5.5)]. (G) Calibration plot obtained from lactate sensors at ~30°C in artificial sweat with different pH (n = 3). Photo credit: Philipp Gutruf, Northwestern University.](https://figures.cochat.ai/paper-figures/c95b352e68f64987841a96ff03a86af5/battery-free-skin-interfaced-microfluidic-electronic-systems-for-simultaneous-figure-3.jpg)
![Fig. 4 Glucose sensor characterization. (A) Exploded-view schematic illustration of the layer makeup of the biofuel cell–based glucose sensor. GOx, glucose oxidase; PtC, platinized carbon. (B) Image of the glucose sensor. (C) Real-time sensor response to increasing glucose concentration in phosphate buffer (pH 7.0) at 25°C and (D) the corresponding calibration (n = 3). (E) Plot illustrating the reversible response of the glucose sensor for four consecutive cycles of varying glucose concentration. V, voltage in millivolts; C, concentration in micromolar. Inset shows calibration plot comparing the sensor signal plotted in (E) for the four cycles. (F) Real-time data acquired with increasing glucose concentration in artificial sweat under common physiological sweat conditions [temperature, ~30°C (pH 5.5)]. (G) Calibration plot obtained from glucose sensors at ~30°C in artificial sweat with different pH (n = 3). Photo credit: Philipp Gutruf, Northwestern University.](https://figures.cochat.ai/paper-figures/4f4d71a41f0e492b9813431a7f23e27c/battery-free-skin-interfaced-microfluidic-electronic-systems-for-simultaneous-figure-4.jpg)



Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In recent years, with the popularity of smartphone cameras, the quantitative measurement of color is highly attractive for in-home monitoring of optical sensors.[237]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [237]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Preliminary or small-scale correlation studies have shown promising applications for other analytes such as metabolites and certain neuroimmune biomarkers present in sweat (3, 27–29, 36), including a recent work that shows how exercise sweat glucose of three healthy subjects roughly tracks their blood glucose, making a preliminary case for semiquantitative glucose prediction (29).”
From Regional and correlative sweat analysis using high-throughput microfluidic sensing patches toward decoding sweat (2019) · cites this paper as [29]
In the citing paper
“Sweat can be collected by capillary wicking into microfluidic pores fabricated into a synthetic polymer membrane that is directly in contact with the skin surface90–93.”
From End-to-end design of wearable sensors (2022) · cites this paper as [91]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Materials and Mechanics for Stretchable Electronics
Rogers, J. A., Someya, T., & Huang, Y. (2010) · Science · doi:10.1126/science.1182383 · Cited by 5,042
Self-citationThe rise of plastic bioelectronics
Someya, T., Bao, Z., & Malliaras, G. G. (2016) · Nature · doi:10.1038/nature21004 · Cited by 1,884
Lee, H., Choi, T. K., Lee, Y. B., Cho, H. R., Ghaffari, R., Wang, L., Choi, H. J., Chung, T. D., Lu, N., Hyeon, T., et al. (2016) · Nature Nanotechnology · doi:10.1038/nnano.2016.38 · Cited by 1,802
Self-citationWearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434
Self-citationA soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat
Koh, A., Kang, D., Xue, Y., Lee, S., Pielak, R. M., Kim, J., Hwang, T., Min, S., Banks, A., Bastien, P., et al. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf2593 · Cited by 1,308
Self-citationNon-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247
Self-citationElectrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862
Self-citationImani, S., Bandodkar, A. J., Mohan, A. M. V., Kumar, R., Yu, S., Wang, J., & Mercier, P. P. (2016) · Nature Communications · doi:10.1038/ncomms11650 · Cited by 861
Self-citationEmaminejad, S., Gao, W., Wu, E., Davies, Z. A., Yin Yin Nyein, H., Challa, S., Ryan, S. P., Fahad, H. M., Chen, K., Shahpar, Z., et al. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1701740114 · Cited by 839
Extremely Elastic Wearable Carbon Nanotube Fiber Strain Sensor for Monitoring of Human Motion
Ryu, S., Lee, P., Chou, J. B., Xu, R., Zhao, R., Hart, A. J., & Kim, S.-G. (2015) · ACS Nano · doi:10.1021/acsnano.5b00599 · Cited by 770
Printable elastic conductors by in situ formation of silver nanoparticles from silver flakes
Matsuhisa, N., Inoue, D., Zalar, P., Jin, H., Matsuba, Y., Itoh, A., Yokota, T., Hashizume, D., & Someya, T. (2017) · Nature Materials · doi:10.1038/nmat4904 · Cited by 750
Rational design of all-solid-state ion-selective electrodes and reference electrodes
Hu, J., Stein, A., & Bühlmann, P. (2016) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2015.11.004 · Cited by 586
Wearable non-invasive epidermal glucose sensors: A review
Kim, J., Campbell, A. S., & Wang, J. (2018) · Talanta · doi:10.1016/j.talanta.2017.08.077 · Cited by 586
Self-citationTee, B. C. ‐., Chortos, A., Dunn, R. R., Schwartz, G., Eason, E., & Bao, Z. (2014) · Advanced Functional Materials · doi:10.1002/adfm.201400712 · Cited by 538
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475
Self-citationSkin-Attachable, Stretchable Electrochemical Sweat Sensor for Glucose and pH Detection
Oh, S. Y., Hong, S. Y., Jeong, Y. R., Yun, J., Park, H., Jin, S. W., Lee, G., Oh, J. H., Lee, H., Lee, S.-S., et al. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b03342 · Cited by 440
Self-citationSkin-interfaced systems for sweat collection and analytics
Choi, J., Ghaffari, R., Baker, L. B., & Rogers, J. A. (2018) · Science Advances · doi:10.1126/sciadv.aar3921 · Cited by 433
Self-citationPrinted Carbon Nanotube Electronics and Sensor Systems
Chen, K., Gao, W., Emaminejad, S., Kiriya, D., Ota, H., Nyein, H. Y. Y., Takei, K., & Javey, A. (2016) · Advanced Materials · doi:10.1002/adma.201504958 · Cited by 425
Adhesive RFID Sensor Patch for Monitoring of Sweat Electrolytes
Rose, D. P., Ratterman, M. E., Griffin, D. K., Hou, L., Kelley-Loughnane, N., Naik, R. R., Hagen, J. A., Papautsky, I., & Heikenfeld, J. C. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2369991 · Cited by 416
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
Tattoo-based potentiometric ion-selective sensors for epidermal pH monitoring
Bandodkar, A. J., Hung, V. W. S., Jia, W., Valdés-Ramírez, G., Windmiller, J. R., Martinez, A. G., Ramírez, J., Chan, G., Kerman, K., & Wang, J. (2013) · The Analyst · doi:10.1039/c2an36422k · Cited by 379
Self-citationAbellán-Llobregat, A., Jeerapan, I., Bandodkar, A., Vidal, L., Canals, A., Wang, J., & Morallón, E. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2017.01.058 · Cited by 348
Self-citationStretchable biofuel cells as wearable textile-based self-powered sensors
Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J.-M., & Wang, J. (2016) · Journal of Materials Chemistry A · doi:10.1039/c6ta08358g · Cited by 317
Choi, J., Kang, D., Han, S., Kim, S. B., & Rogers, J. A. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201601355 · Cited by 315
Self-citationDiagnostic Sweat Testing: The Cystic Fibrosis Foundation Guidelines
LeGrys, V. A., Yankaskas, J. R., Quittell, L. M., Marshall, B. C., & Mogayzel, P. J. (2007) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2007.03.002 · Cited by 310
“Cut‐and‐Paste” Manufacture of Multiparametric Epidermal Sensor Systems
Yang, S., Chen, Y., Nicolini, L., Pasupathy, P., Sacks, J., Su, B., Yang, R., Sanchez, D., Chang, Y., Wang, P., et al. (2015) · Advanced Materials · doi:10.1002/adma.201502386 · Cited by 304
Advances in wearable chemical sensor design for monitoring biological fluids
Matzeu, G., Florea, L., & Diamond, D. (2015) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2015.01.077 · Cited by 300
Non‐invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016
Heikenfeld, J. (2016) · Electroanalysis · doi:10.1002/elan.201600018 · Cited by 294
A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration
Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D. (2010) · Analytical Methods · doi:10.1039/b9ay00184k · Cited by 286
Formulation and stability of a novel artificial human sweat under conditions of storage and use
Harvey, C. J., LeBouf, R. F., & Stefaniak, A. B. (2010) · Toxicology in Vitro · doi:10.1016/j.tiv.2010.06.016 · Cited by 274
All‐Printed Stretchable Electrochemical Devices
Bandodkar, A. J., Nuñez‐Flores, R., Jia, W., & Wang, J. (2015) · Advanced Materials · doi:10.1002/adma.201500768 · Cited by 199
Self-citationKim, J., Banks, A., Xie, Z., Heo, S. Y., Gutruf, P., Lee, J. W., Xu, S., Jang, K., Liu, F., Brown, G., et al. (2015) · Advanced Functional Materials · doi:10.1002/adfm.201501590 · Cited by 193
Self-citationMicroneedle-based self-powered glucose sensor
Valdés-Ramírez, G., Li, Y.-C., Kim, J., Jia, W., Bandodkar, A. J., Nuñez-Flores, R., Miller, P. R., Wu, S.-Y., Narayan, R., Windmiller, J. R., et al. (2014) · Electrochemistry Communications · doi:10.1016/j.elecom.2014.07.014 · Cited by 188
Self-citationAraki, H., Kim, J., Zhang, S., Banks, A., Crawford, K. E., Sheng, X., Gutruf, P., Shi, Y., Pielak, R. M., & Rogers, J. A. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201604465 · Cited by 183
Self-citationKim, S. B., Zhang, Y., Won, S. M., Bandodkar, A. J., Sekine, Y., Xue, Y., Koo, J., Harshman, S. W., Martin, J. A., Park, J. M., et al. (2018) · Small · doi:10.1002/smll.201703334 · Cited by 167
Self-citationChoi, J., Xue, Y., Xia, W. J., Ray, T. R., Reeder, J. T., Bandodkar, A. J., Kang, D., Xu, S., Huang, Y. Y., & Rogers, J. A. (2017) · Lab on a Chip · doi:10.1039/c7lc00525c · Cited by 150
Self-citationThe relevance of sweat testing for the diagnosis of cystic fibrosis in the genomic era.
Mishra, A., Greaves, R. F., & Massie, J. (2005) · Europe PMC (PubMed Central) · Cited by 128
Hyponatremia Associated With Exercise: Risk Factors and Pathogenesis
Montain, S. J., Sawka, M. N., & Wenger, C. B. (2001) · Exercise and Sport Sciences Reviews · doi:10.1097/00003677-200107000-00005 · Cited by 115
Biagi, S., Ghimenti, S., Onor, M., & Bramanti, E. (2012) · Biomedical Chromatography · doi:10.1002/bmc.2713 · Cited by 110
Godek, S. F., Bartolozzi, A. R., & Godek, J. J. (2005) · British Journal of Sports Medicine · doi:10.1136/bjsm.2004.011767 · Cited by 92
The Language of Glove: Wireless gesture decoder with low-power and stretchable hybrid electronics
O’Connor, T. F., Fach, M. E., Miller, R., Root, S. E., Mercier, P. P., & Lipomi, D. J. (2017) · PLoS ONE · doi:10.1371/journal.pone.0179766 · Cited by 92
Recent advances in RF-based passive device-free localisation for indoor applications
Palipana, S., Pietropaoli, B., & Pesch, D. (2017) · Ad Hoc Networks · doi:10.1016/j.adhoc.2017.06.007 · Cited by 76
An integrated laser blood flowmeter
Higurashi, E., Sawada, R., & Ito, T. (2003) · Journal of Lightwave Technology · doi:10.1109/jlt.2003.809551 · Cited by 74
Bandodkar, A. J., Jia, W., Ramírez, J., & Wang, J. (2015) · Advanced Healthcare Materials · doi:10.1002/adhm.201400808 · Cited by 67
Self-citationMultimodal epidermal devices for hydration monitoring
Krishnan, S. R., Shi, Y., Webb, R. C., Ma, Y., Bastien, P., Crawford, K. E., Wang, A., Feng, X., Manco, M. C., Kurniawan, J. F., et al. (2017) · Microsystems & Nanoengineering · doi:10.1038/micronano.2017.14 · Cited by 65
Self-citationColorimetric Determination of Chloride with Mercuric Chloranilate
Barney, J. E., & Bertolacini, R. J. (1957) · Analytical Chemistry · doi:10.1021/ac60128a023 · Cited by 54
Harker, M., Coulson, H., Fairweather, I., Taylor, D., & Daykin, C. A. (2006) · Metabolomics · doi:10.1007/s11306-006-0024-4 · Cited by 48
Wang, S., Zhang, G., Meng, H., & Li, L. (2012) · Skin Research and Technology · doi:10.1111/j.1600-0846.2012.00645.x · Cited by 37
Experimental low cost reflective type oximeter for wearable health systems
Rodrigues, E. M. G., Godina, R., Cabrita, C. M. P., & Catalào, J. P. S. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.09.013 · Cited by 33
Developing advanced traffic violation detection system with RFID technology for smart city
Wong, S. F., Mak, H. C., Ku, C.-H., et al. (2017) · doi:10.1109/ieem.2017.8289907 · Cited by 21
Cited by
Technology Roadmap for Flexible Sensors
Luo, Y., Abidian, M. R., Ahn, J.-H., Akinwande, D., Andrews, A. M., Antonietti, M., Bao, Z., Berggren, M., Berkey, C. A., Bettinger, C. J., et al. (2023) · ACS Nano · doi:10.1021/acsnano.2c12606 · Cited by 1,383
Self-citationEnd-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306
A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat
Yang, Y., Song, Y., Bo, X., Min, J., Pak, O. S., Zhu, L., Wang, M., Tu, J., Kogan, A., Zhang, H., et al. (2020) · Nature Biotechnology · doi:10.1038/s41587-019-0321-x · Cited by 1,270
Lim, H., Kim, H. S., Qazi, R., Kwon, Y., Jeong, J., & Yeo, W. (2020) · Advanced Materials · doi:10.1002/adma.201901924 · Cited by 1,002
Smart Textiles for Electricity Generation
Chen, G., Li, Y., Bick, M., & Chen, J. (2020) · Chemical Reviews · doi:10.1021/acs.chemrev.9b00821 · Cited by 931
Metadata
- Authors
- Bandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., Ghaffari, R., Su, C.-J., Leshock, J. P., Ray, T., Verrillo, A., Thomas, K., Krishnamurthi, V., Han, S., Kim, J., Krishnan, S., Hang, T., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- Japan Atomic Energy Agency (JP)
- Washington University in St. Louis (US)
- RMIT University (AU)
- Ajou University (KR)
- Kwangwoon University (KR)
- Shanghai Jiao Tong University (CN)
- Corresponding author
- Bandodkar, Amay J.
- Gutruf, Philipp
- Choi, Jungil
- Rogers, John A.
- ORCID
- Bandodkar, Amay J.
- Gutruf, Philipp
- Choi, Jungil
- Lee, KunHyuck
- Sekine, Yurina
- Reeder, Jonathan T.
- Jeang, William J.
- Aranyosi, Alexander J.
- Lee, Stephen P.
- Model, Jeffrey B.
- Ghaffari, Roozbeh
- Ray, Tyler
- Verrillo, Anthony
- Thomas, Kyle
- Krishnamurthi, Vaishnavi
- Han, Seungyong
- Kim, Jeonghyun
- Krishnan, Siddharth
- Hang, Tao
- Rogers, John A.
- Venue
- Science Advances
- Published
- 2019-01-04
- Volume
- 5
- Issue
- 1
- Pages
- eaav3294-eaav3294
- ISSN
- 2375-2548 (ISSN-L)
- PMC
- PMC6357758
- OpenAlex
- W2908711986
- Cited by
- 761
- References
- 51
- Citations by year
- 2026: 84 · 2025: 104 · 2024: 95 · 2023: 115 · 2022: 115 · 2021: 119 · 2020: 82 · 2019: 46
- FWCI
- 31.78
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by-nc
- Locations
- Science Advances (publishedVersion, cc-by-nc)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- UA Campus Repository (The University of Arizona) (acceptedVersion, cc-by-nc)
- RMIT Research Repository (RMIT University Library) (submittedVersion, cc-by-nc)
- Figshare (submittedVersion, cc-by-nc)
- Funders
- Air Force Research Laboratory (FA8650-14-D-6516)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
- Keywords
- wearable sweat sensors, battery-free electronics, microfluidic platform, colorimetric assay, wireless sensing, sweat glucose, sweat lactate, sweat pH, sweat chloride, sweat rate
- MeSH terms
- Wearable Electronic Devices, Bioelectric Energy Sources, Chlorides, Colorimetry, Glucose, Humans, Hydrogen-Ion Concentration, Male, Skin, Sweat, Biosensing Techniques, Lactic Acid, Microfluidics, Lab-On-A-Chip Devices, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.