A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat
Open accessAbstract
A soft, skin-mounted microfluidic device captures microliter volumes of sweat and quantitatively measures biochemical markers by colorimetric analysis.
AI snapshot
Written from the paper’s full text.
TL;DR
A soft, skin-mounted microfluidic device captures sweat through microchannels and reservoirs and colorimetrically quantifies sweat rate, total loss, pH, chloride, lactate, and glucose, with optional NFC smartphone readout. Human testing during controlled cycling and a 104-km outdoor bicycle race demonstrated feasibility, though the colorimetric chemistry limits real-time and time-dependent marker tracking.
Key findings
- The device retained ~50 μL of sweat, corresponding to 1 to 6 hours of exercise depending on sweat loss and mounting location; sweat rates were 12 to 120 μL/hour per cm².
- In 9 subjects during controlled cycling at 38°C and 50% relative humidity, lower-back sweat rates were typically ~2.3-fold greater than volar forearm rates; large-harvesting-area devices agreed with absorbent pads, and colorimetric chloride, lactate, and pH results agreed with laboratory analysis.
- In 12 cyclists during the 104-km El Tour de Tucson, devices collected sweat without detachment, leakage, or reported discomfort; older subjects (50–69 years) had greater sweating rates than younger subjects (10–29 years), and males had greater rates than females.
- Colorimetric assays covered pH 5.0–7.0, lactate 1.5–100 mM, and glucose with an LOD of ~200 μM; image analysis could resolve 0.5 pH units and 0.2, 0.3, and 0.1 mM changes in chloride, lactate, and glucose, respectively, corresponding to a 1% change in the R channel.
Methods
- Soft lithography fabricated multilayer PDMS microfluidic devices with a skin adhesive layer, colorimetric reagent reservoirs, serpentine sweat-volume channel, and optional NFC electronics; in vitro testing used an artificial sweat-pore system with a syringe pump.
- Indoor human study: 9 healthy subjects exercised on cycle ergometers under controlled 38°C and 50% relative humidity, with devices on the lower back and volar forearm and two harvesting areas; results were compared with absorbent pads and laboratory chemical analysis.
- Outdoor human study: 12 healthy volunteers wore devices on the lower back and volar forearm during a 104-km bicycle race; sweat images were captured by smartphone and digital single-lens reflex cameras, and Pearson/Spearman correlation analyses were performed using SAS and JMP.
Limitations
- Stated: the colorimetric schemes do not afford real-time tracking of changes in analyte concentration, and the current channel/reservoir designs cannot easily provide time-dependent marker concentration information.
- Stated: limitations are primarily in the range of chemical reagents available for accurate colorimetric analysis at relevant concentrations; further field studies are needed to demonstrate accuracy in realistic, demanding scenarios.
- Stated and evident: the small sample size only partially demonstrated correlations across individual subjects, and glucose in healthy subjects fell below the LOD for both image and laboratory analysis; small harvesting areas yielded somewhat higher inferred sweat rates, perhaps due to altered perspiration behavior caused by the device’s physical presence.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 22 cited sources checked · 0 supported · 0 flagged
- Findingown data
Conformal contact of the adhesive layer inhibited lateral flow of sweat from outside the defined openings, ensuring fluid from the harvesting area dominated the sweat sample.
“The conformal contact of the adhesive layer inhibited lateral flow of sweat from regions located outside the defined openings, ensuring fluid issuing from the harvesting area dominated the sweat sample (fig. S3).”
Quote in the paperSubstantial claim - Findingreasoning
Quantitative colorimetric assay reagents in the reservoirs enabled assessment of pH and concentrations of glucose, lactate, and chloride through enzymatic or chromogenic reactions.
“Quantitative colorimetric assay reagents in the reservoirs enabled assessment of pH and the concentration of selected essential markers, including glucose, lactate, and chloride, through either enzymatic or chromogenic reactions.”
Quote in the paperSubstantial claim - Findingcitation
The acrylic adhesive exhibited about 5 times greater adhesion force than typical medical adhesives such as Tegaderm (1.02 N).
“This adhesive exhibited about 5 times greater adhesion force than typical medical adhesives such as Tegaderm (1.02 N) (33).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [33] Klode J Schöttler L Stoffels I Körber A Schadendorf D Dissemond J Investigation of adhesion of modern wound dressings: a comparative analysis of 56 different wound dressings Journal of the European Academy of Dermatology and Venereology 25 933 939 2011 21062365 source
Not checked in the free run.
- Findingown data
A medical-grade acrylic adhesive film ensured stable, strong, seamless adhesion (~5.7 N) of the device to skin without irritation, even at regions of substantial hair coverage or in the presence of sweat.
“A medical-grade acrylic adhesive film ensured stable, strong and seamless adhesion (~5.7 N) of the device to the skin without irritation in a manner that offered compatibility with skin, even at regions of substantial hair coverage or in the presence of sweat (fig. S2).”
Quote in the paperSubstantial claim - Findingcitation
The maximum normal and shear stresses at the device/phantom skin interface were far below the threshold for somatosensory perception of forces (20 kPa) during ~30% stretch of the skin.
“The maximum normal and shear stresses at the device/phantom skin interface were far below the threshold for somatosensory perception of forces (20 kPa) during ~30% stretch of the skin (fig. S4) (37).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [37] Lee CH Ma Y Jang KI Banks A Pan T Feng X Kim JS Kang D Raj MS McGrane BL Morey B Wang X Ghaffari R Huang Y Rogers JA Soft core/shell packages for stretchable electronics Adv Funct Mater 25 3698 3704 2015 source
Not checked in the free run.
- Findingown data
FEA results show deformation of a representative device under various mechanical distortions of an underlying phantom skin sample.
“The finite element analysis (FEA) results of strain/stress distributions and corresponding optical images in Fig. 1E show deformation of a representative device under various mechanical distortions of an underlying phantom skin sample as a support (a PDMS substrate exhibiting similar mechanical properties to skin).”
Quote in the paperSubstantial claim - Findingown data
Integrated electronics allow wireless interfaces to external computing and digital analysis systems using common platforms such as the smartphone.
“Integrated electronics allow wireless interfaces to external computing and digital analysis systems using common platforms such as the smartphone.”
Quote in the paperNot a substantial claim - Findingcitation
FEA results demonstrated that the maximum strain in the copper layer was below the elastic limit (0.3%) under all loading conditions.
“FEA results demonstrated that the maximum strain in the copper layer was below the elastic limit (0.3%) under all loading conditions (fig. S5) (39).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [39] Kim J Banks A Cheng HY Xie ZQ Xu S Jang KI Lee JW Liu ZJ Gutruf P Huang X Wei PH Liu F Li K Dalal M Ghaffari R Feng X Huang YG Gupta S Paik U Rogers JA Epidermal electronics with advanced capabilities in near-field communication Small 11 906 912 2015 25367846 source
Not checked in the free run.
- Findingown data
The overall designs allowed stretchable electronic systems to operate under physical deformation without significantly altering the mechanical properties of the soft microfluidic structures or the overall device.
“The overall designs allowed these stretchable electronic systems to operate under physical deformation without significantly altering the mechanical properties of the soft microfluidic structures or the overall device.”
Quote in the paperSubstantial claim - Findingcitation
Modeling yielded predictions for percentage changes in the volume of the serpentine channel associated with externally applied pressures between 100 and 400 Pa comparable to those that might be associated with a gentle touch by a fingertip.
“Modeling yielded predictions for percentage changes in the volume of the serpentine channel associated with externally applied pressures between 100 and 400 Pa (Fig. 2C) comparable to those that might be associated with a gentle touch by a fingertip (41).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [41] Pang C Lee GY Kim TI Kim SM Kim HN Ahn SH Suh KY A flexible and highly sensitive strain-gauge sensor using reversible interlocking of nanofibres Nature Mater 11 795 801 2012 22842511 source
Not checked in the free run.
- Findingreasoning
Owing to the thin layer (~25 μm) coated on the channel wall and the hydrophilic properties of the pHEMA hydrogel matrix, the hydrodynamics of flow within the channel were not influenced during conditions of momentary flow.
“Owing to the thin layer (~25 μm) coated on the channel wall and the hydrophilic properties of the pHEMA hydrogel matrix, the hydrodynamics of flow within the channel were not influenced during conditions of momentary flow (fig. S11, A to D).”
Quote in the paperSubstantial claim - Findingown data
As sweat entered the channel, anhydrous cobalt (II) chloride chelated with water to form hexahydrate cobalt chloride, generating a change in color from deep blue (λmax = 657) to pale purple (λmax = 511).
“As sweat entered the channel, the anhydrous cobalt (II) chloride chelated with water to form hexahydrate cobalt chloride (CoCl2•6H2O), generating a change in color from deep blue (λmax = 657) to pale purple (λmax = 511) (Fig. 3B).”
Quote in the paperSubstantial claim - Findingown data
In the orbicular serpentine channel, cobalt (II) chloride contained in a pHEMA hydrogel matrix served as a colorimetric indicator.
“In the orbicular serpentine channel, cobalt (II) chloride (that is, CoCl2) contained in a coating of a polyhydroxyethylmethacrylate (pHEMA) hydrogel matrix served as a colorimetric indicator.”
Quote in the paperSubstantial claim - Findingcitation
The color change in the detection reservoir correlated with the concentration of lactate throughout the relevant range expected in sweat (1.5 to 100 mM).
“The color change in the detection reservoir correlated with the concentration of lactate throughout the relevant range expected in sweat (1.5 to 100 mM) (Fig. 3C) (13, 46, 47).”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [13] Sonner Z Wilder E Heikenfeld J Kasting G Beyette F Swaile D Sherman F Joyce J Hagen J Kelley-Loughnane N Naik R The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications Biomicrofluidics 9 031301 2015 26045728 PMC4433483 source
Not checked in the free run.
- cites [46] Alvear-Ordenes I García-López D De Paz JA González-Gallego J Sweat lactate, ammonia, and urea in rugby players Int J Sports Med 26 632 637 2005 16158367 source
Not checked in the free run.
- cites [47] Lamont LS Sweat lactate secretion during exercise in relation to women’s aerobic capacity Journal of Applied Physiology 62 194 198 1987 3558180 source
Not checked in the free run.
- Findingcitation
Creatinine was detected in sweat using a mixture of enzymes (creatininase, creatinase, and peroxidase) and a corresponding responsive dye (4-aminophenazone).
“Similarly, creatinine, a vital marker of hydration status and renal function, was detected in sweat using a mixture of enzymes (creatininase, creatinase, and peroxidase) and a corresponding responsive dye (4-aminophenazone) (Fig. 3E) (53).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [53] Crocker H Shephard M White G Evaluation of an enzymatic method for determining creatinine in plasma Journal of clinical pathology 41 576 581 1988 3384988 PMC1141516 source
Not checked in the free run.
- Findingown data
The color changes occurred on timescales of <1 min.
“The color changes occurred on timescales of <1 min (movie S4).”
Quote in the paperNot a substantial claim - Findingnone given
The cellulose matrices in each reservoir could be filled with as little as 5 to 10 μl of sweat sample.
“The cellulose matrices in each reservoir could be filled with as little as 5 to 10 μl of sweat sample.”
Quote in the paperSubstantial claim - Findingown data
The formulation of enzyme and dyes in the detection cocktail solution ensured a dynamic range compatible with human sweat.
“The formulation of enzyme and dyes in the detection cocktail solution ensured a dynamic range compatible with human sweat.”
Quote in the paperSubstantial claim - Findingreasoning
The application digitized RGB color information on the screen, enabling the user to read the concentration of the marker.
“The application digitized RGB color information on the screen, enabling the user to read the concentration of the marker.”
Quote in the paperSubstantial claim - Findingown data
The proximity of a smartphone to the device initiated image capture and analysis software automatically using NFC.
“Figure 4A shows frames from a video clip (movie S1) in which the proximity of a smartphone to the device initiated image capture and analysis software automatically using NFC.”
Quote in the paperNot a substantial claim - Findingown data
Although PDMS is known to have some permeability to water and certain small molecules, the colorimetric responses in these devices are unaffected for most practical applications due to the relevant operational timescale and the analyte chemistries used.
“Although PDMS is known to have some permeability to water and certain small molecules, the colorimetric responses in these devices are unaffected for most practical applications due to the relevant operational timescale and the analyte chemistries used (Fig. S14).”
Quote in the paperSubstantial claim - Findingown data
A universal pH indicator that includes dyes such as bromothymol blue, methyl red, and phenolphthalein yielded colorimetric responses over a medically relevant range (pH 5.0 to 7.0).
“A universal pH indicator that includes dyes such as bromothymol blue, methyl red, and phenolphthalein yielded colorimetric responses over a medically relevant range (pH 5.0 to 7.0) (Fig. 3F).”
Quote in the paperSubstantial claim- Rests on an assumption: Sweat pH is an index of hydration state.
- Findingown data
After image correction, the digital color data (in %RGB format) were converted into analyte concentrations using calibration curves.
“After image correction, the digital color data (in %RGB format) were converted into analyte concentrations using calibration curves (Fig. 4D).”
Quote in the paperSubstantial claim - Findingown data
Devices with large harvesting areas showed agreement with absorbing pads for sweat rate.
“Rates determined using devices with large harvesting areas showed agreement with those obtained using absorbing pads (Fig. 5E).”
Quote in the paperSubstantial claim - Findingown data
Concentrations of chloride, glucose, lactate, and pH obtained by the colorimetric readouts demonstrated excellent agreement with conventional laboratory analysis of sweat collected from absorbing pads.
“Concentrations of the markers chloride, glucose, lactate, and pH obtained by the colorimetric readouts demonstrated excellent agreement with conventional laboratory analysis of sweat collected from absorbing pads as shown in Fig. 5F.”
Quote in the paperSubstantial claim - Findingown data
Unsealed microfluidic structures described here can hold captured sweat for ~75 hours upon removal from the skin with negligible deterioration of colorimetric analysis.
“we observed that the microfluidics structures described here can hold captured sweat for ~125 hours upon removal from the skin and sealing of the open channels (~75 hours without sealing) with negligible deterioration of colorimetric analysis.”
Quote in the paperSubstantial claim - Findingown data
Glucose concentration in sweat from healthy subjects fell below the limit of detection for both image and laboratory analysis.
“The glucose concentration in sweat from healthy subjects fell below the LOD for both image and laboratory analysis (P < 0.05 refers to differences in background noise only).”
Quote in the paperNot a substantial claim - Findingown data
Male subjects exhibited greater rates of sweating than female subjects.
“Older subjects (ages 50 to 69 years) had greater rates of sweating in comparison to that of younger subjects (ages 10 to 29 years), and male subjects exhibited greater rates of sweating than females (Fig. 6E).”
Quote in the paperNot a substantial claim - Findingcitation
Devices with small harvesting areas yielded somewhat higher inferred sweat rates than those with larger harvesting areas.
“Devices with small harvesting areas yielded somewhat higher inferred rates than those with larger harvesting areas, perhaps due to alternations in perspiration behavior caused by the physical presence of the device (Fig. 5E) (55).”
Quote in the paperNot a substantial claim - Findingown data
Sealed microfluidic structures described here can hold captured sweat for ~125 hours upon removal from the skin with negligible deterioration of colorimetric analysis.
“we observed that the microfluidics structures described here can hold captured sweat for ~125 hours upon removal from the skin and sealing of the open channels (~75 hours without sealing) with negligible deterioration of colorimetric analysis.”
Quote in the paperSubstantial claim - Findingown data
Participants in the outdoor cycling race reported no sense of discomfort or limitation in body or arm movement during cycling.
“Participants reported no sense of discomfort or limitation in body or arm movement during the cycling.”
Quote in the paperSubstantial claim - Interpretationown data
The thin geometry (25 μm) and low modulus (~17 kPa) of the adhesive layer provided stress release during skin deformation, facilitating comfort and long-term wearability.
“The thin geometry (25 μm) and low modulus (~17 kPa) of this layer provided stress release during deformation of the skin (fig. S2), facilitating comfort and long-term wearability.”
Quote in the paperSubstantial claim - Interpretationcitation
The soft mechanics and thin geometry enabled soft, nonirritating intimate contact with the skin through principles similar to those established for epidermal electronics.
“The soft mechanics and thin geometry enabled soft, nonirritating intimate contact with the skin through principles similar to those established for epidermal electronics (2, 36).”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [2] Rogers JA Electronics for the human body JAMA 313 561 562 2015 25668256 source
Not checked in the free run.
- cites [36] Kim DH Ghaffari R Lu N Rogers JA Flexible and Stretchable Electronics for Biointegrated Devices Annual Review of Biomedical Engineering 14 113 128 2012 22524391 source
Not checked in the free run.
- Interpretationown data
The serpentine channel layouts provided a convenient means to increase the total channel volume for a given device size.
“In addition, the serpentine channel layouts provided a convenient means to increase the total channel volume for a given device size (fig. S9).”
Quote in the paperSubstantial claim - Interpretationreasoning
The position of the leading edge that defines this color change, along with the dimensional characteristics and geometry of the channel, yields quantitative information on the sweat rate and volume.
“The position of the leading edge that defines this color change, along with the dimensional characteristics and geometry of the channel, yields quantitative information on the sweat rate and volume.”
Quote in the paperSubstantial claim - Interpretationcitation
For image capture once every 5 min, a spatial resolution of ~0.35 to 3.5 mm can easily resolve changes in the positions of the fluid fronts, providing 12 data points within a ~60-min timeframe.
“For image capture once every 5 min, a spatial resolution of ~0.35 to 3.5 mm can easily resolve changes in the positions of the fluid fronts, providing 12 data points within a ~60-min timeframe (13).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] Sonner Z Wilder E Heikenfeld J Kasting G Beyette F Swaile D Sherman F Joyce J Hagen J Kelley-Loughnane N Naik R The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications Biomicrofluidics 9 031301 2015 26045728 PMC4433483 source
Not checked in the free run.
- Premisecitation
The concentration of lactate in sweat is an indicator of exercise intolerance and tissue hypoxia.
“The concentration of lactate in sweat is an indicator of exercise intolerance and tissue hypoxia (44, 45).”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [44] Polliack A Taylor R Bader D Sweat analysis following pressure ischaemia in a group of debilitated subjects JRRD 34 303 308 1997 9239623
Not checked in the free run.
- cites [45] Biagi S Ghimenti S Onor M Bramanti E Simultaneous determination of lactate and pyruvate in human sweat using reversed-phase high-performance liquid chromatography: a noninvasive approach Biomed Chromatogr 26 1408 1415 2012 22311625 source
Not checked in the free run.
- Interpretationcitation
The reported devices' range of sensitivity could diagnose hyperglycemia, with a limit of detection of ~200 μM.
“the range of sensitivity in the reported devices could diagnose hyperglycemia, for example, limit of detection (LOD) = ~200 μM (fig. S13) (50).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [50] Boysen TC Yanagawa S Sato F Sato K A modified anaerobic method of sweat collection Journal of Applied Physiology 56 1302 1307 1984 6327585 source
Not checked in the free run.
- Rests on an assumption: Glucose concentration in sweat is typically one order of magnitude lower than in plasma.
- Study design and conductown data
The crosses allowed rotations/translations of the images to facilitate accurate analysis of sweat rate and volume on the serpentine channel.
“The crosses further allowed rotations/translations of the images to facilitate accurate analysis of sweat rate and volume on the serpentine channel (fig. S7).”
Quote in the paperSubstantial claim - Study design and conductown data
Reference color markers (true white and black) allowed white balancing to eliminate the dependence of the analysis on lighting conditions of practical relevance (daylight, shadow, and various light sources).
“Reference color markers (true white and black) allowed white balancing to eliminate the dependence of the analysis on lighting conditions of practical relevance (daylight, shadow, and various light sources) (fig. S6).”
Quote in the paperSubstantial claim - Interpretationown data
The y intercept in Fig. 5E corresponds to the limit of sweat measurement with the absorbing pad due to water evaporation (0.349 g/cm2s at 38°C and 50% relative humidity) during sample collection.
“Notably, the y intercept in Fig. 5E corresponds to the limit of sweat measurement with the absorbing pad due to water evaporation (0.349 g/cm2s of water evaporation at 38 °C, 50% relative humidity) during sample collection, highlighting one of its limitations.”
Quote in the paperNot a substantial claim
- May combine two claims. 29
- Classified as probably not a substantial claim. 14
- The quote only partly states this claim. 1
May combine two claims.
FindingcitationThe soft epidermal microfluidic device adheres and conforms to skin and captures and routes sweat through microchannels and reservoirs using capillarity and natural perspiration pressure (~70 kPa) for volumetric assessment and chemical analysis in situ.
“The soft, epidermal microfluidic device technology introduced here adheres and conforms to the skin in a manner that captures and routes sweat through a network of microchannels and reservoirs—using a combination of capillarity and action of the natural pressure (~70 kPa) associated with perspiration—for volumetric assessment and chemical analysis in situ (13).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] Sonner Z Wilder E Heikenfeld J Kasting G Beyette F Swaile D Sherman F Joyce J Hagen J Kelley-Loughnane N Naik R The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications Biomicrofluidics 9 031301 2015 26045728 PMC4433483 source
Not checked in the free run.
- Rests on an assumption: The pressure that drives fluid flow through the device arises from the action of the sweat glands themselves, assisted by capillary effects in the microchannels and the materials embedded within them.
May combine two claims.
FindingcitationThe specific device designs can retain ~50 μL of sweat, corresponding to an effective working time of 1 to 6 hours depending on sweat loss rate and mounting location (12 to 120 μL/hour per cm2).
“The specific designs described can retain ~50 μL of sweat corresponding to an effective working time of 1 to 6 hours of exercise, depending on the rate of sweat loss and the mounting location on the body (12 to 120 μL/hour per cm2) (32).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] Sato K Sato F Individual variations in structure and function of human eccrine sweat gland Am J Physiol Regul Integr Comp Physiol 245 R203 R208 1983 6881378 source
Not checked in the free run.
Classified as probably not a substantial claim.
Findingnone givenThe colorimetric schemes embedded in the current devices did not afford real-time tracking of changes in analyte concentration.
“Hence, the colorimetric schemes embedded in the current devices did not afford real-time tracking of changes in analyte concentration.”
Quote in the paperSubstantial claimMay combine two claims.
FindingreasoningA water-responsive chromogenic reagent in the serpentine channel allowed determination of the extent of filling with sweat, which could be converted to overall sweat rate and volume.
“A water-responsive chromogenic reagent in the serpentine channel allowed determination of the extent of filling with sweat, which could be converted to overall sweat rate and volume.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationThe elastomeric microfluidic devices exhibited an effective modulus of ~0.16 MPa, comparable to human skin and previously reported epidermal devices.
“These elastomeric microfluidic devices exhibited an effective modulus of ~0.16 MPa, comparable to human skin and previously reported epidermal devices (37, 38).”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [37] Lee CH Ma Y Jang KI Banks A Pan T Feng X Kim JS Kang D Raj MS McGrane BL Morey B Wang X Ghaffari R Huang Y Rogers JA Soft core/shell packages for stretchable electronics Adv Funct Mater 25 3698 3704 2015 source
Not checked in the free run.
- cites [38] Liang X Boppart S Biomechanical properties of in vivo human skin from dynamic optical coherence elastography IEEE Trans Biomed Eng 57 953 959 2010 19822464 PMC3699319 source
Not checked in the free run.
May combine two claims.
Findingown dataThe water vapor loss showed little dependence on the length of the outlet channel, whereas backpressure was linearly proportional to this length according to calculations for a model system.
“The water vapor loss showed little dependence on the length of the outlet channel, whereas backpressure was linearly proportional to this length according to calculations for a model system (fig. S8 and Fig. 2B).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe technology capitalized on NFC schemes to launch image capture and analysis software on an external device and/or to read temperature from an integrated sensor.
“Our technology capitalized on NFC schemes to launch image capture and analysis software on such an external device and/or to read temperature from an integrated sensor (movies S1 and S2).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataCalculations indicated that vapor loss with 100-μm-wide channels was ~3.2-fold lower than with 800-μm-wide channels, while 25-μm-wide channels differed by only ~1.1-fold from 100-μm-wide channels.
“Our calculations further indicated that the vapor loss with 100-μm-wide channels was ~3.2-fold lower than with 800-μm-wide channels, while 25-μm-wide channels differed by only ~1.1-fold from 100-μm-wide channels (Fig. 2B).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataA device (0.07 cm2 harvesting surface area) laminated on the perforated membrane captured dyed water pumped at 5.5 μl/hour, demonstrating the first step in quantitative analysis of liquid uptake.
“A device (0.07 cm2 of harvesting surface area) laminated on the perforated membrane captured dyed water pumped at 5.5 μl/hour (Fig. 2G), demonstrating the first step in quantitative analysis of liquid uptake (movie S3).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe experiments revealed negligible loss of water vapor and no fluid leakage under these conditions.
“The experiments revealed negligible loss of water vapor and no fluid leakage under these conditions.”
Quote in the paperSubstantial claimMay combine two claims.
FindingreasoningThe reservoirs fill within ~0.3 to 3.2 hours at these sweating rates, with times that scale linearly with reservoir volume.
“The reservoirs fill within ~0.3 to 3.2 hours at these sweating rates, with times that scale linearly with reservoir volume.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor widths under 100 μm, backpressure notably increased and bending deformation of the device was obstructed, with negligible effects on vapor loss.
“For widths under 100 μm, the backpressure notably increased and bending deformation of the device was obstructed, with negligible effects on vapor loss.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataHuman studies presented subsequently showed volumetric sweat harvesting rates of ~1.2 to 12 μl/hour, corresponding to linear filling rates of ~0.07 to 0.7 mm/min along the serpentine channels.
“For a device layout with harvesting area (~10 mm2), the human studies presented subsequently showed volumetric sweat harvesting rates of ~1.2 to 12 μl/hour, corresponding to linear filling rates of ~0.07 to 0.7 mm/min along the serpentine channels.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe experimentally determined harvested liquid volume in the channel is consistent with the input volume introduced by the syringe pump and with linear hydrodynamic flow in the microfluidic channel.
“As shown in movie S3, the experimentally determined harvested liquid volume in the channel is consistent with the input volume introduced by the syringe pump and with linear hydrodynamic flow in the microfluidic channel.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe volume change increased with the aspect ratio (AR; width to height); for AR 5 the volume change was about fivefold greater than for AR 3.3.
“The volume change increased with the aspect ratio (AR; width to height). For example, the volume change for AR 5 (width, 1.5 mm; height, 300 μm) was about fivefold greater than that for AR 3.3 (width, 1.0 mm; height, 300 μm).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataSweat could continue to travel slowly through the channel by spontaneous internal flow (0.68 μl/hour), with the possibility of a ~2% reading error.
“The sweat could, however, continue to travel slowly through the channel by spontaneous internal flow (0.68 μl/hour), with the possibility of a ~2% reading error (fig. S11, E and F).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThis artifact does not occur in channels without the hydrogel; its effect could be eliminated, for practical purposes, by patterning the hydrogel into short segments.
“This artifact does not occur in channels without the hydrogel; its effect could be eliminated, for practical purposes, by patterning the hydrogel into short segments (fig. S12).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe colorimetric sensing approach allowed simple, rapid quantitative assessment of instantaneous rate and total volume of sweat loss, pH, and concentration of chloride, lactate, and glucose in sweat.
“The colorimetric sensing approach allowed simple, rapid quantitative assessment of instantaneous rate and total volume of sweat loss, the pH, as well as the concentration of chloride, lactate, and glucose in the sweat (Fig. 3A).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataGlucose concentration could also be analyzed by an enzymatic reaction.
“Glucose concentration could also be analyzed by an enzymatic reaction (Fig. 3D).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataRecording color changes and converting them into quantitative information were accomplished by digital image capture and analysis.
“Recording color changes and converting them into quantitative information were accomplished by digital image capture and analysis.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationThe previously reported ultrathin NFC electronics integrated on the top of the microfluidic device enabled wireless communication to external devices, with stable operation and a soft, biocompatible set of mechanical properties, even under a 30% strain condition.
“The previously reported ultrathin NFC electronics (39) integrated on the top of the microfluidic device enabled wireless communication to external devices, with stable operation and a soft, biocompatible set of mechanical properties, even under a 30% strain condition (39).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [39] Kim J Banks A Cheng HY Xie ZQ Xu S Jang KI Lee JW Liu ZJ Gutruf P Huang X Wei PH Liu F Li K Dalal M Ghaffari R Feng X Huang YG Gupta S Paik U Rogers JA Epidermal electronics with advanced capabilities in near-field communication Small 11 906 912 2015 25367846 source
Not checked in the free run.
May combine two claims.
Findingown dataIn the presence of chloride ions, iron ions (Fe2+) bind with TPTZ, whereas Hg2+ participates as HgCl2, thereby inducing a change in color from transparent to blue.
“In the presence of chloride ions, iron ions (Fe2+) bind with TPTZ, whereas Hg2+ participates as HgCl2, thereby inducing a change in color from transparent to blue as shown in Fig. 3G.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe NFC electronics facilitated image capture, and built-in sensors provided wireless, digital data on skin temperature.
“The NFC electronics facilitated image capture, and built-in sensors provided wireless, digital data on skin temperature (fig. S15 and movie S2).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenThe calibration curve in Fig. 4C captures the pseudo-linearities associated with the serpentine shape of the channel.
“The calibration curve in Fig. 4C captures the pseudo-linearities associated with the serpentine shape of the channel.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe device could reliably measure changes of 0.5 pH units and 0.2, 0.3, and 0.1 mM of chloride, lactate, and glucose concentrations, respectively, corresponding to a 1% change in the R channel of the RGB images.
“We could reliably measure changes of 0.5 pH units and 0.2, 0.3, and 0.1 mM of chloride, lactate, and glucose concentrations, respectively, corresponding to a 1% change in the R channel of the RGB images.”
Quote in the paperSubstantial claimThe quote only partly states this claim.
Findingown dataIn nine human subjects during intermediate-level cycle ergometer exercise under 38°C and 50% relative humidity, sweat rates measured on the lower back were typically about 2.3-fold greater than on the volar forearm.
“Although the rates exhibited great variation among individual subjects, those measured on the lower back were typically ~2.3-fold greater than on the volar forearm, consistent with expectations from studies using conventional techniques (54).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [54] Smith CJ Havenith G Body mapping of sweating patterns in male athletes in mild exercise-induced hyperthermia Eur J Appl Physiol 111 1391 1404 2011 21153660 source
Not checked in the free run.
May combine two claims.
Findingown dataDevices on 12 volunteer riders in the El Tour de Tucson competitive long-distance outdoor bicycling race performed as anticipated, successfully collecting sweat with regional colorimetric change without patch detachment, even with substantial changes in temperature and humidity.
“In all cases, the devices performed as anticipated, successfully collecting sweat with regional colorimetric change without patch detachment, even with substantial changes in temperature and humidity.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataOlder subjects (ages 50 to 69 years) had greater rates of sweating than younger subjects (ages 10 to 29 years).
“Older subjects (ages 50 to 69 years) had greater rates of sweating in comparison to that of younger subjects (ages 10 to 29 years), and male subjects exhibited greater rates of sweating than females (Fig. 6E).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenThe reported embodiments can detect sweat volume and rate, as well as glucose, creatinine, lactate, chloride, and pH.
“The reported embodiments can detect sweat volume and rate, as well as several key markers including glucose, creatinine, lactate, chloride, and pH.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe devices accurately captured sweat volume and rate information continuously, without the need for removal.
“Furthermore, the devices accurately captured the volume and rate information continuously, without the need for removal.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationSpecially formulated PDMS variants offer optical transparency, ease of patterning into microfluidic systems, biocompatibility, and favorable mechanics (low modulus ~145 kPa; high elasticity up to ~200% strain at break).
“Specially formulated variants of PDMS offer physical characteristics that are attractive for this application, including optical transparency, ease of patterning into microfluidic systems, biocompatibility, and favorable mechanics (low modulus, ~145 kPa; high elasticity, up to ~200% strain at break) (35).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [35] Lee JN Park C Whitesides GM Solvent compatibility of poly (dimethylsiloxane)-based microfluidic devices Anal Chem 75 6544 6554 2003 14640726 source
Not checked in the free run.
Classified as probably not a substantial claim.
Study design and conductown dataReference marks on the top of the device platform included a white dot and black crosses for color balancing to allow accurate color extraction under arbitrary lighting conditions.
“Reference marks on the top of the device platform (Fig. 1B) included a white dot and black crosses for color balancing to allow accurate color extraction under arbitrary lighting condition (fig. S6).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe crosses helped determine the position and orientation from the images.
“The crosses also helped determine the position and orientation from the images (fig. S7).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenDecreasing the cross-sectional area of the channel increases the filling rate proportionally.
“Decreasing the cross-sectional area of the channel increases the filling rate proportionally.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone givenThree factors determine the resolution in determination of sweat rate: the rate of fluid flow into the reservoirs and serpentine channel, the harvesting area, and the time and spatial resolution of the camera system and image analysis software.
“Regarding device design, three factors determine the resolution in determination of sweat rate: (i) the rate of fluid flow into the reservoirs and the serpentine channel, (ii) the harvesting area, and (iii) the time and spatial resolution of the camera system and image analysis software.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe concentration of chloride ions serves as a marker of cystic fibrosis and altered electrolyte levels correspond to a sodium ion imbalance.
“The concentration of chloride ions serves as a marker of cystic fibrosis and altered electrolyte levels correspond to a sodium ion imbalance (17).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17] Jadoon S Karim S Akram MR Kalsoom Khan A Zia MA Siddiqi AR Murtaza G Recent developments in sweat analysis and its applications 2015 2015 7 PMC4369929 25838824 source
Not checked in the free run.
Classified as probably not a substantial claim.
Study design and conductown dataA white dot in the middle of the device and four black crosses distributed near the center established values for 100% and 0% in %RGB coordinates, respectively.
“In particular, a white dot in the middle of the device and four black crosses distributed near the center established values for 100% and 0% in %RGB coordinates, respectively (Fig. 4B).”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe angle of the filling front (the leading edge of the color change) in the serpentine channel defined the volume of sweat collected, thereby allowing calculation of total sweat loss and, with the time interval, total sweat rate.
“The angle of the filling front (the leading edge of the color change) in the serpentine channel defined the volume of sweat collected, thereby allowing calculation of total sweat loss and, with the time interval, total sweat rate.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenCompared to previously described technologies for sweat analysis using porous materials and fabrics or hydrogels as fluidic interfaces, our systems are unique in their use of fully integrated, soft microfluidics consisting of a network of functionalized channels and reservoirs for sweat capture, routing, and storage with spatially separated regions for analysis.
“Compared to previously described technologies for sweat analysis consisting of porous materials and fabrics or hydrogels as fluidic interfaces, our systems are unique in their use of fully integrated, soft microfluidics consisting of a network of functionalized channels and reservoirs for sweat capture, routing, and storage with spatially separated regions for analysis.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenThe limitations of the current devices are primarily in the range of chemical reagents available for accurate colorimetric analysis of markers at relevant ranges of concentration.
“The limitations of the current devices are primarily in the range of chemical reagents that are available for accurate colorimetric analysis of markers at relevant ranges of concentration.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe correlation between variations in marker analysis across individual subjects was only partially demonstrated because of the small sample size.
“The correlation between variations in marker analysis across individual subjects was only partially demonstrated because of the small sample size.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Interpretationnone givenThe microfluidic channel/reservoir designs reported here cannot be easily designed to provide information on time-dependent changes in marker concentration detection.
“However, the microfluidic channel/reservoir designs reported here cannot be easily designed to provide information on time-dependent changes in marker concentration detection.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe experimental conditions, including temperature, humidity, time course of application of device, and weight of absorbing Webril pads for sweat collection were all controlled and documented.
“The experimental conditions, including temperature, humidity, time course of application of device, and weight of absorbing Webril pads for sweat collection were all controlled and documented.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
The pressure that drives fluid flow through the device arises from the action of the sweat glands themselves, assisted by capillary effects in the microchannels and the materials embedded within them.
“The pressure that drives fluid flow arises from the action of the sweat glands themselves, assisted by capillary effects in the microchannels and the materials embedded within them.”
The paper relies on this without support.
Sweat pH is an index of hydration state.
“In sweat, pH is often considered an index of hydration state.”
The paper relies on this without support.
Glucose concentration in sweat is typically one order of magnitude lower than in plasma.
“We note that glucose concentration in sweat is typically one order of magnitude lower than in plasma;”
The paper relies on this without support.
The conventional absorbing pad and laboratory analysis methods provide accurate reference measurements of sweat rate and marker concentrations.
“We compared the performance of the device in situ to conventional procedures that use absorbing pads applied onto the skin with subsequent weighing and laboratory-based analyses, such as spectrophotometry.”
“Concentrations of the markers chloride, glucose, lactate, and pH obtained by the colorimetric readouts demonstrated excellent agreement with conventional laboratory analysis of sweat collected from absorbing pads as shown in Fig. 5F.”
The paper relies on this without support.
Sweat marker concentrations are indicative of health and disease status.
“The epidermal microfluidic devices introduced here represent versatile platforms for evaluating athletic performance and monitoring health and disease status.”
The paper relies on this without support.
The physical presence of the device does not significantly alter local sweat rate or composition when large harvesting areas are used.
“Devices with small harvesting areas yielded somewhat higher inferred rates than those with larger harvesting areas, perhaps due to alternations in perspiration behavior caused by the physical presence of the device (Fig. 5E) (55).”
The paper relies on this without support.
The constituent materials of the device are biocompatible and safe for prolonged skin contact.
“The soft mechanical properties, biocompatible constituent materials, digitally analyzable colorimetric responses, and overall careful optimization of structural, evaporative and fluidic properties are integral to the effectiveness of these devices and differentiate them from other sweat analysis technologies.”
The paper relies on this without support.
The tested populations (nine indoor subjects and 12 outdoor cyclists) are representative of the broader populations for whom the device is intended.
“The epidermal microfluidic devices introduced here represent versatile platforms for evaluating athletic performance and monitoring health and disease status.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Koh et al.'s closed microfluidic system proved efficacious when tested on humans during cycling.
“The device proved to be efficacious when tested on humans during cycling.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [55]
Sweat pads/patches developed at academic institutions have potential applications for measuring heat stress, dehydration and metabolism in athletes, soldiers, firefighters, and industrial laborers who exercise or work in hot environments.
“These devices have potential applications for measuring heat stress, dehydration and metabolism in athletes, soldiers, firefighters, and industrial laborers who exercise or work in hot environments.”
From A Critical Review of Consumer Wearables, Mobile Applications, and Equipment for Providing Biofeedback, Monitoring Stress, and Sleep in Physically Active Populations (2018) · cites this paper as Koh et al., 2016
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Stated sample volume exceeds each reservoir's geometric capacity
MinorA circular chamber 4 mm in diameter and 300 μm deep has a geometric volume of about 3.8 μL, before accounting for the paper inside it. That is less than the stated 5 to 10 μL of sweat needed to fill each reservoir's cellulose matrix.
“The microfluidic system consisted of a bottom polydimethylsiloxane (PDMS) layer (thickness, 500 μm) embossed with appropriate relief geometry (uniform depth, 300 μm) and filled with reagents for colorimetric analysis (Fig. 1, A and B). A top-capping layer of PDMS served as a seal (thickness, 200 μm). Our particular layout included four circular chambers (diameter, 4 mm)”
“The cellulose matrices in each reservoir could be filled with as little as 5 to 10 μl of sweat sample.”
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836 · see citing works
Self-citationSkin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes
Lipomi, D. J., Vosgueritchian, M., Tee, B. C.-K., Hellstrom, S. L., Lee, J. A., Fox, C. H., & Bao, Z. (2011) · Nature Nanotechnology · doi:10.1038/nnano.2011.184 · Cited by 3,188 · see citing works
Patterned Paper as a Platform for Inexpensive, Low‐Volume, Portable Bioassays
Martinez, A. W., Phillips, S. T., Butte, M. J., & Whitesides, G. M. (2007) · Angewandte Chemie International Edition · doi:10.1002/anie.200603817 · Cited by 2,882 · see citing works
Solvent Compatibility of Poly(dimethylsiloxane)-Based Microfluidic Devices
Lee, J. N., Park, C., & Whitesides, G. M. (2003) · Analytical Chemistry · doi:10.1021/ac0346712 · Cited by 2,643 · see citing works
An ultra-lightweight design for imperceptible plastic electronics
Kaltenbrunner, M., Sekitani, T., Reeder, J., Yokota, T., Kuribara, K., Tokuhara, T., Drack, M., Schwödiauer, R., Graz, I., Bauer-Gogonea, S., et al. (2013) · Nature · doi:10.1038/nature12314 · Cited by 2,485 · see citing works
Graphene Oxide: Intrinsic Peroxidase Catalytic Activity and Its Application to Glucose Detection
Song, Y., Qu, K., Zhao, C., Ren, J., & Qu, X. (2010) · Advanced Materials · doi:10.1002/adma.200903783 · Cited by 2,027 · see citing works
A flexible and highly sensitive strain-gauge sensor using reversible interlocking of nanofibres
Pang, C., Lee, G.-Y., Kim, T.-I., Kim, S. M., Kim, H. N., Ahn, S.-H., & Suh, K.-Y. (2012) · Nature Materials · doi:10.1038/nmat3380 · Cited by 1,695 · see citing works
Ultrasensitive mechanical crack-based sensor inspired by the spider sensory system
Kang, D., Pikhitsa, P. V., Choi, Y. W., Lee, C., Shin, S. S., Piao, L., Park, B., Suh, K.-Y., Kim, T.-I., & Choi, M. (2014) · Nature · doi:10.1038/nature14002 · Cited by 1,626 · see citing works
Self-citationMartinez, A. W., Phillips, S. T., Carrilho, E., Thomas, S. W., Sindi, H., & Whitesides, G. M. (2008) · Analytical Chemistry · doi:10.1021/ac800112r · Cited by 1,445 · see citing works
Wei, H., & Wang, E. (2008) · Analytical Chemistry · doi:10.1021/ac702203f · Cited by 1,413 · see citing works
Nanowire active-matrix circuitry for low-voltage macroscale artificial skin
Takei, K., Takahashi, T., Ho, J. C., Ko, H., Gillies, A. G., Leu, P. W., Fearing, R. S., & Javey, A. (2010) · Nature Materials · doi:10.1038/nmat2835 · Cited by 1,274 · see citing works
Ultrathin conformal devices for precise and continuous thermal characterization of human skin
Webb, R. C., Bonifas, A. P., Behnaz, A., Zhang, Y., Yu, K. J., Cheng, H., Shi, M., Bian, Z., Liu, Z., Kim, Y.-S., et al. (2013) · Nature Materials · doi:10.1038/nmat3755 · Cited by 1,229 · see citing works
Self-citationInjectable, Cellular-Scale Optoelectronics with Applications for Wireless Optogenetics
Kim, T.-I., McCall, J. G., Jung, Y. H., Huang, X., Siuda, E. R., Li, Y., Song, J., Song, Y. M., Pao, H. A., Kim, R.-H., et al. (2013) · Science · doi:10.1126/science.1232437 · Cited by 1,174 · see citing works
Self-citationViventi, J., Kim, D.-H., Vigeland, L., Frechette, E. S., Blanco, J. A., Kim, Y.-S., Avrin, A. E., Tiruvadi, V. R., Hwang, S.-W., Vanleer, A. C., et al. (2011) · Nature Neuroscience · doi:10.1038/nn.2973 · Cited by 1,136 · see citing works
Self-citationThin Plates and Shells: Theory, Analysis, and Applications
Ventsel, E. S., Krauthammer, T., & Carrera, E. (2002) · Applied Mechanics Reviews · doi:10.1115/1.1483356 · Cited by 1,018 · see citing works
Electrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862 · see citing works
Materials and Optimized Designs for Human‐Machine Interfaces Via Epidermal Electronics
Jeong, J., Yeo, W., Akhtar, A., Norton, J. J. S., Kwack, Y., Li, S., Jung, S., Su, Y., Lee, W., Xia, J., et al. (2013) · Advanced Materials · doi:10.1002/adma.201301921 · Cited by 827 · see citing works
Self-citationFlexible and Stretchable Electronics for Biointegrated Devices
Kim, D.-H., Ghaffari, R., Lu, N., & Rogers, J. A. (2012) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071811-150018 · Cited by 749 · cited 1× · see citing works
“The soft mechanics and thin geometry enabled soft, nonirritating intimate contact with the skin through principles similar to those established for epidermal electronics (2, 36).”
BackgroundSelf-citationSonner, Z., Wilder, E., Heikenfeld, J., Kasting, G., Beyette, F., Swaile, D., Sherman, F., Joyce, J., Hagen, J., Kelley-Loughnane, N., et al. (2015) · Biomicrofluidics · doi:10.1063/1.4921039 · Cited by 724 · see citing works
Point-of-care colorimetric detection with a smartphone
Shen, L., Hagen, J. A., & Papautsky, I. (2012) · Lab on a Chip · doi:10.1039/c2lc40741h · Cited by 610 · see citing works
Dagdeviren, C., Shi, Y., Joe, P., Ghaffari, R., Balooch, G., Usgaonkar, K., Gur, O., Tran, P. L., Crosby, J. R., Meyer, M., et al. (2015) · Nature Materials · doi:10.1038/nmat4289 · Cited by 513 · see citing works
Self-citationWireless Optofluidic Systems for Programmable In Vivo Pharmacology and Optogenetics
Jeong, J., McCall, J. G., Shin, G., Zhang, Y., Al‐Hasani, R., Kim, M. K., Li, S., Sim, J. Y., Jang, K., Shi, Y., et al. (2015) · Cell · doi:10.1016/j.cell.2015.06.058 · Cited by 493 · see citing works
Self-citationBandodkar, A. J., Molinnus, D., Mirza, O., Guinovart, T., Windmiller, J. R., Valdés-Ramírez, G., Andrade, F. J., Schöning, M. J., & Wang, J. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.11.039 · Cited by 490 · see citing works
Taylor, N. A., & Machado-Moreira, C. A. (2013) · Extreme Physiology & Medicine · doi:10.1186/2046-7648-2-4 · Cited by 490 · see citing works
Biomechanical Properties of <i>In Vivo</i> Human Skin From Dynamic Optical Coherence Elastography
Liang, X., & Boppart, S. A. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2033464 · Cited by 470 · see citing works
Adhesive RFID Sensor Patch for Monitoring of Sweat Electrolytes
Rose, D. P., Ratterman, M. E., Griffin, D. K., Hou, L., Kelley-Loughnane, N., Naik, R. R., Hagen, J. A., Papautsky, I., & Heikenfeld, J. C. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2369991 · Cited by 416 · see citing works
Body mapping of sweating patterns in male athletes in mild exercise-induced hyperthermia
Smith, C. J., & Havenith, G. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1744-8 · Cited by 375 · see citing works
Smartphone based health accessory for colorimetric detection of biomarkers in sweat and saliva
Oncescu, V., O’Dell, D., & Erickson, D. (2013) · Lab on a Chip · doi:10.1039/c3lc50431j · Cited by 362 · see citing works
A potentiometric tattoo sensor for monitoring ammonium in sweat
Guinovart, T., Bandodkar, A. J., Windmiller, J. R., Andrade, F. J., & Wang, J. (2013) · The Analyst · doi:10.1039/c3an01672b · Cited by 343 · see citing works
A short history of sweat gland biology
Wilke, K., Martin, A. E., Terstegen, L., & Biel, S. S. (2007) · International Journal of Cosmetic Science · doi:10.1111/j.1467-2494.2007.00387.x · Cited by 333 · see citing works
BIOTEX—Biosensing Textiles for Personalised Healthcare Management
Coyle, S., Lau, K.-T., Moyna, N., O'Gorman, D., Diamond, D., Di Francesco, F., Costanzo, D., Salvo, P., Trivella, M. G., De Rossi, D. E., et al. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2038484 · Cited by 330 · see citing works
Advances in wearable chemical sensor design for monitoring biological fluids
Matzeu, G., Florea, L., & Diamond, D. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.01.077 · Cited by 303 · see citing works
Non‐invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016
Heikenfeld, J. (2016) · Electroanalysis · doi:10.1002/elan.201600018 · Cited by 294 · see citing works
Stretchable, Wireless Sensors and Functional Substrates for Epidermal Characterization of Sweat
Huang, X. L., Liu, Y., Chen, K., Shin, W., Lu, C., Kong, G., Patnaik, D., Lee, S., Cortes, J. F., & Rogers, J. A. (2014) · Small · doi:10.1002/smll.201400483 · Cited by 282 · see citing works
Self-citationEpidermal Electronics with Advanced Capabilities in Near-Field Communication
Kim, J., Banks, A., Cheng, H., Xie, Z., Xu, S., Jang, K.-I., Lee, J. W., Liu, Z., Gutruf, P., Huang, X., et al. (2015) · Small · doi:10.1002/smll.201402495 · Cited by 274 · see citing works
Self-citationGao, L. H., Zhang, Y., Malyarchuk, V., Jia, L., Jang, K., Webb, R. C., Fu, H., Shi, Y., Zhou, G., Shi, L., et al. (2014) · Nature Communications · doi:10.1038/ncomms5938 · Cited by 271 · see citing works
Self-citationShirreffs, S. M., & Maughan, R. J. (1997) · Journal of Applied Physiology · doi:10.1152/jappl.1997.82.1.336 · Cited by 251 · see citing works
Stamp Collapse in Soft Lithography
Huang, Y. Y., Zhou, W., Hsia, K. J., Menard, E., Park, J., Rogers, J. A., & Alleyne, A. G. (2005) · Langmuir · doi:10.1021/la0502185 · Cited by 224 · see citing works
Self-citationCorrie, S. R., Coffey, J. W., Islam, J., Markey, K. A., & Kendall, M. A. F. (2015) · The Analyst · doi:10.1039/c5an00464k · Cited by 220 · see citing works
Recent Developments in Sweat Analysis and Its Applications
Jadoon, S., Karim, S., Akram, M. R., Kalsoom Khan, A., Zia, M. A., Siddiqi, A. R., & Murtaza, G. (2015) · International Journal of Analytical Chemistry · doi:10.1155/2015/164974 · Cited by 216 · see citing works
Individual variations in structure and function of human eccrine sweat gland
Sato, K., & Sato, F. (1983) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1983.245.2.r203 · Cited by 215 · see citing works
Curto, V. F., Fay, C., Coyle, S. M., Byrne, R. J., O’Toole, C., Barry, C., Hughes, S. M., Moyna, N. M., Diamond, D., & Benito‐Lopez, F. (2012) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2012.06.048 · Cited by 211 · see citing works
Severity of compensatory sweating after thoracoscopic sympathectomy
Licht, P. B., & Pilegaard, H. K. (2004) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2004.02.087 · Cited by 166 · see citing works
A Wearable Sensor for Measuring Sweat Rate
Salvo, P., Francesco, F. D., Costanzo, D., Ferrari, C., Trivella, M. G., & Rossi, D. D. (2010) · IEEE Sensors Journal · doi:10.1109/jsen.2010.2046634 · Cited by 146 · see citing works
Optofluidics: Fundamentals, Devices, and Applications
Fainman, Y., Lee, L. P., Psaltis, D., & Yang, C. (2009) · Cited by 142 · see citing works
Soft Core/Shell Packages for Stretchable Electronics
Lee, C. H., Ma, Y., Jang, K., Banks, A., Pan, T., Feng, X., Kim, J. S., Kang, D., Raj, M. S., McGrane, B. L., et al. (2015) · Advanced Functional Materials · doi:10.1002/adfm.201501086 · Cited by 131 · see citing works
Self-citationBiagi, S., Ghimenti, S., Onor, M., & Bramanti, E. (2012) · Biomedical Chromatography · doi:10.1002/bmc.2713 · Cited by 110 · see citing works
Electronics for the Human Body
Rogers, J. A. (2015) · JAMA · doi:10.1001/jama.2014.17915 · Cited by 91 · see citing works
Self-citationA modified anaerobic method of sweat collection
Boysen, T. C., Yanagawa, S., Sato, F., & Sato, K. (1984) · Journal of Applied Physiology · doi:10.1152/jappl.1984.56.5.1302 · Cited by 91 · see citing works
Klode, J., Schöttler, L., Stoffels, I., Körber, A., Schadendorf, D., & Dissemond, J. (2011) · Journal of the European Academy of Dermatology and Venereology · doi:10.1111/j.1468-3083.2010.03886.x · Cited by 73 · see citing works
Sweat Lactate, Ammonia, and Urea in Rugby Players
Alvear-Ordenes, I., García-López, D., De Paz, J. A., & González-Gallego, J. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-830380 · Cited by 67 · see citing works
Artificial microfluidic skin for in vitro perspiration simulation and testing
Hou, L., Hagen, J. A., Wang, X., Papautsky, I., Naik, R. K., Kelley‐Loughnane, N., & Heikenfeld, J. C. (2013) · Lab on a Chip · doi:10.1039/c3lc41231h · Cited by 55 · see citing works
Evaluation of an enzymatic method for determining creatinine in plasma.
Crocker, H., Shephard, M. D., & White, G. H. (1988) · Journal of Clinical Pathology · doi:10.1136/jcp.41.5.576 · Cited by 49 · see citing works
Sweat lactate secretion during exercise in relation to women's aerobic capacity
Lamont, L. S. (1987) · Journal of Applied Physiology · doi:10.1152/jappl.1987.62.1.194 · Cited by 46 · see citing works
Sweat analysis following pressure ischaemia in a group of debilitated subjects.
Polliack, A. A., Taylor, R. P., & Bader, D. L. (1997) · PubMed · Cited by 41 · see citing works
Flagged first, then by citations
2 of the 58 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- White F. M., Fluid Mechanics (McGraw-Hill Higher Education, ed. 7, 2010).
Cited by
Sorted by citations
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373 · see citing works
Self-citationLab-on-Skin: A Review of Flexible and Stretchable Electronics for Wearable Health Monitoring
Liu, Y., Pharr, M., & Salvatore, G. A. (2017) · ACS Nano · doi:10.1021/acsnano.7b04898 · Cited by 1,725 · see citing works
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486 · see citing works
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434 · see citing works
Self-citationTechnology Roadmap for Flexible Sensors
Luo, Y., Abidian, M. R., Ahn, J., Akinwande, D., Andrews, A. M., Antonietti, M., Bao, Z., Magnus Berggren, A., Berkey, C. A., Bettinger, C. J., et al. (2023) · ACS Nano · doi:10.1021/acsnano.2c12606 · Cited by 1,419 · see citing works
Self-citationBio-Integrated Wearable Systems: A Comprehensive Review
Ray, T. R., Choi, J., Bandodkar, A. J., Krishnan, S., Gutruf, P., Tian, L., Ghaffari, R., & Rogers, J. A. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00573 · Cited by 1,350 · see citing works
Self-citationWearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315 · see citing works
End-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306 · see citing works
Flexible Electronics toward Wearable Sensing
Gao, W. F., Ota, H., Kiriya, D., Takei, K., & Javey, A. (2019) · Accounts of Chemical Research · doi:10.1021/acs.accounts.8b00500 · Cited by 1,281 · see citing works
A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat
Yang, Y., Song, Y., Bo, X., Min, J., Pak, O. S., Zhu, L., Wang, M., Tu, J., Kogan, A., Zhang, H., et al. (2020) · Nature Biotechnology · doi:10.1038/s41587-019-0321-x · Cited by 1,270 · see citing works
Rich, S. I., Wood, R. J., & Majidi, C. (2018) · Nature Electronics · doi:10.1038/s41928-018-0024-1 · Cited by 1,025 · see citing works
Lim, H., Kim, H. S., Qazi, R., Kwon, Y., Jeong, J., & Yeo, W. (2019) · Advanced Materials · doi:10.1002/adma.201901924 · Cited by 997 · see citing works
Disposable Sensors in Diagnostics, Food, and Environmental Monitoring
Dincer, C., Bruch, R. C., Costa‐Rama, E., Fernández-Abedul, M. T., Merkoçi, A., Manz, A., Urban, G. A., & Güder, F. (2019) · Advanced Materials · doi:10.1002/adma.201806739 · Cited by 972 · see citing works
Device integration of electrochemical biosensors
Wu, J., Liu, H., Chen, W., Ma, B., & Ju, H. (2023) · Nature Reviews Bioengineering · doi:10.1038/s44222-023-00032-w · Cited by 931 · see citing works
A wearable electrochemical biosensor for the monitoring of metabolites and nutrients
Wang, M., Yang, Y., Min, J., Song, Y., Tu, J., Mukasa, D., Ye, C., Xu, C., Heflin, N., McCune, J. S., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00916-z · Cited by 920 · see citing works
Wearable health devices-vital sign monitoring, systems and technologies
Dias, D., & Paulo Silva Cunha, J. (2018) · Sensors · doi:10.3390/s18082414 · Cited by 919 · see citing works
PDMS with designer functionalities—Properties, modifications strategies, and applications
Wolf, M., Salieb–Beugelaar, G. B., & Hunziker, P. (2018) · Progress in Polymer Science · doi:10.1016/j.progpolymsci.2018.06.001 · Cited by 855 · see citing works
Wound Healing: From Passive to Smart Dressings
Farahani, M., & Shafiee, A. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100477 · Cited by 822 · see citing works
Wearable and flexible electrochemical sensors for sweat analysis: a review
Gao, F., Liu, C., Zhang, L., Liu, T., Wang, Z., Song, Z., Cai, H., Fang, Z., Chen, J., Wang, J., et al. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00443-6 · Cited by 813 · see citing works
Light in diagnosis, therapy and surgery
Yun, S. H., & Kwok, S. J. J. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-016-0008 · Cited by 806 · see citing works
Advancing Biosensors with Machine Learning
Cui, F., Yue, Y., Zhang, Y., Zhang, Z., & Zhou, H. S. (2020) · ACS Sensors · doi:10.1021/acssensors.0c01424 · Cited by 786 · see citing works
Enzyme‐Based Glucose Sensor: From Invasive to Wearable Device
Lee, H., Hong, Y. J., Baik, S., Hyeon, T., & Kim, D. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201701150 · Cited by 778 · see citing works
Bandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav3294 · Cited by 761 · see citing works
Self-citationElectrical bioadhesive interface for bioelectronics
Deng, J., Yuk, H., Wu, J., Varela, C. E., Chen, X., Roche, E. T., Guo, C. F., & Zhao, X. (2020) · Nature Materials · doi:10.1038/s41563-020-00814-2 · Cited by 736 · see citing works
Skin-Interfaced Wearable Sweat Sensors for Precision Medicine
Min, J., Tu, J., Xu, C., Lukas, H., Shin, S., Yang, Y., Solomon, S. A., Mukasa, D., & Gao, W. F. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00823 · Cited by 728 · see citing works
Advances in healthcare wearable devices
Iqbal, S. M. A., Mahgoub, I., Du, E., Leavitt, M. A., & Asghar, W. (2021) · npj Flexible Electronics · doi:10.1038/s41528-021-00107-x · Cited by 714 · see citing works
Hydrogel interfaces for merging humans and machines
Yuk, H., Wu, J., & Zhao, X. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00483-4 · Cited by 683 · see citing works
Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays
Park, J., Kim, J., Kim, S.-Y., Cheong, W. H., Jang, J., Park, Y.-G., Na, K., Kim, Y.-T., Heo, J. H., Lee, C. Y., et al. (2018) · Science Advances · doi:10.1126/sciadv.aap9841 · Cited by 671 · see citing works
Self-citationYu, Y., Nassar, J., Xu, C., Min, J., Yang, Y., Dai, A., Doshi, R., Huang, A., Song, Y., Gehlhar, R., et al. (2020) · Science Robotics · doi:10.1126/scirobotics.aaz7946 · Cited by 655 · see citing works
Wireless battery-free wearable sweat sensor powered by human motion
Song, Y., Min, J., Yu, Y., Wang, H., Yang, Y., Zhang, H., & Gao, W. (2020) · Science Advances · doi:10.1126/sciadv.aay9842 · Cited by 652 · see citing works
Tehrani, F., Teymourian, H., Wuerstle, B. L., Kavner, J., Patel, R., Furmidge, A., Aghavali, R., Hosseini-Toudeshki, H., Brown, C., Zhang, F., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00887-1 · Cited by 651 · see citing works
Evolution of Wearable Devices with Real-Time Disease Monitoring for Personalized Healthcare
Guk, K., Han, G., Lim, J., Jeong, K., Kang, T., Lim, E.-K., & Jung, J. (2019) · Nanomaterials · doi:10.3390/nano9060813 · Cited by 624 · see citing works
Flexible Hybrid Electronics for Digital Healthcare
Ma, Y., Zhang, Y., Cai, S., Han, Z., Liu, X., Wang, F., Cao, Y., Wang, Z., Li, H., Chen, Y., et al. (2020) · Advanced Materials · doi:10.1002/adma.201902062 · Cited by 620 · see citing works
Wearable chemical sensors for biomarker discovery in the omics era
Sempionatto, J. R., Lasalde-Ramírez, J. A., Mahato, K., Wang, J., & Gao, W. (2022) · Nature Reviews Chemistry · doi:10.1038/s41570-022-00439-w · Cited by 613 · see citing works
Multifunctional Skin‐Inspired Flexible Sensor Systems for Wearable Electronics
Xu, K., Lu, Y., & Takei, K. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201800628 · Cited by 602 · see citing works
Wearable non-invasive epidermal glucose sensors: A review
Kim, J., Campbell, A. S., & Wang, J. (2017) · Talanta · doi:10.1016/j.talanta.2017.08.077 · Cited by 585 · see citing works
Self-citationIntegrated textile sensor patch for real-time and multiplex sweat analysis
He, W., Wang, C., Wang, H., Jian, M., Lu, W., Liang, X., Zhang, X., Yang, F., & Zhang, Y. (2019) · Science Advances · doi:10.1126/sciadv.aax0649 · Cited by 564 · see citing works
An electrically conductive silver–polyacrylamide–alginate hydrogel composite for soft electronics
Ohm, Y., Pan, C., Ford, M. J., Huang, X., Liao, J., & Majidi, C. (2021) · Nature Electronics · doi:10.1038/s41928-021-00545-5 · Cited by 551 · see citing works
Wearable sensors for monitoring the physiological and biochemical profile of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0150-9 · Cited by 537 · see citing works
Peake, J. M., Kerr, G., & Sullivan, J. P. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.00743 · Cited by 523 · see citing works
Enabling Technologies for Personalized and Precision Medicine
Ho, D., Quake, S. R., McCabe, E. R. B., Chng, W. J., Chow, E. K., Ding, X., Gelb, B. D., Ginsburg, G. S., Hassenstab, J., Ho, C.-M., et al. (2020) · Trends in Biotechnology · doi:10.1016/j.tibtech.2019.12.021 · Cited by 517 · see citing works
Flexible Hybrid Sensors for Health Monitoring: Materials and Mechanisms to Render Wearability
Gao, Y., Yu, L., Yeo, J. C., & Lim, C. T. (2019) · Advanced Materials · doi:10.1002/adma.201902133 · Cited by 508 · see citing works
Multifunctional conductive hydrogel-based flexible wearable sensors
Wang, L., Xu, T., & Zhang, X. (2021) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2020.116130 · Cited by 507 · see citing works
Self-citationShrivastava, S., Trung, T. Q., & Lee, N. (2020) · Chemical Society Reviews · doi:10.1039/c9cs00319c · Cited by 495 · see citing works
Toward a new generation of smart skins
Someya, T., & Amagai, M. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0079-1 · Cited by 485 · see citing works
Disruptive, Soft, Wearable Sensors
Ling, Y., An, T., Yap, L. W., Zhu, B., Gong, S. P., & Cheng, W. (2019) · Advanced Materials · doi:10.1002/adma.201904664 · Cited by 483 · see citing works
Shao, Y., Ying, Y. B., & Ping, J. F. (2020) · Chemical Society Reviews · doi:10.1039/c9cs00587k · Cited by 480 · see citing works
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475 · see citing works
Self-citationSkin-like biosensor system via electrochemical channels for noninvasive blood glucose monitoring
Chen, Y., Lu, S., Zhang, S., Li, Y., Qu, Z., Chen, Y., Lu, B., Wang, X., & Feng, X. (2017) · Science Advances · doi:10.1126/sciadv.1701629 · Cited by 470 · see citing works
Mimicking Human and Biological Skins for Multifunctional Skin Electronics
Lee, Y., Park, J., Choe, A., Cho, S., Kim, J., & Ko, H. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201904523 · Cited by 465 · see citing works
Self-citationSkin-inspired soft bioelectronic materials, devices and systems
Zhao, C., Park, J., Root, S. E., & Bao, Z. (2024) · Nature Reviews Bioengineering · doi:10.1038/s44222-024-00194-1 · Cited by 451 · see citing works
A Wearable Microfluidic Sensing Patch for Dynamic Sweat Secretion Analysis
Nyein, H. Y. Y., Tai, L., Ngo, Q. P., Chao, M., Zhang, G. B., Gao, W. F., Bariya, M., Bullock, J., Kim, H., Fahad, H. M., et al. (2018) · ACS Sensors · doi:10.1021/acssensors.7b00961 · Cited by 450 · see citing works
Self-assembled three dimensional network designs for soft electronics
Jang, K.-I., Li, K., Chung, H. U., Xu, S., Jung, H. N., Yang, Y., Kwak, J. W., Jung, H. H., Song, J., Yang, C., et al. (2017) · Nature Communications · doi:10.1038/ncomms15894 · Cited by 445 · see citing works
Self-citationWearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442 · see citing works
Skin-interfaced systems for sweat collection and analytics
Choi, J., Ghaffari, R., Baker, L. B., & Rogers, J. A. (2018) · Science Advances · doi:10.1126/sciadv.aar3921 · Cited by 433 · see citing works
Self-citationA wearable patch for continuous analysis of thermoregulatory sweat at rest
Nyein, H. Y. Y., Bariya, M., Tran, B., Ahn, C. H., Brown, B. J., Ji, W., Davis, N., & Javey, A. (2021) · Nature Communications · doi:10.1038/s41467-021-22109-z · Cited by 426 · see citing works
A universal interface for plug-and-play assembly of stretchable devices
Jiang, Y., Ji, S., Sun, J., Huang, J., Li, Y., Zou, G., Salim, T., Wang, C., Li, W., Jin, H., et al. (2023) · Nature · doi:10.1038/s41586-022-05579-z · Cited by 422 · see citing works
Materials-Driven Soft Wearable Bioelectronics for Connected Healthcare
Gong, S. P., Yan, L., Yin, J., Levin, A., & Cheng, W. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00502 · Cited by 424 · see citing works
Shirzaei Sani, E., Xu, C., Wang, C., Song, Y., Min, J., Tu, J., Solomon, S. A., Li, J., Banks, J. L., Armstrong, D. G., et al. (2023) · Science Advances · doi:10.1126/sciadv.adf7388 · Cited by 405 · see citing works
A wireless patch for the monitoring of C-reactive protein in sweat
Tu, J., Min, J., Song, Y., Xu, C., Li, J., Moore, J., Hanson, J., Hu, E., Parimon, T., Wang, T.-Y., et al. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01059-5 · Cited by 395 · see citing works
Xu, W., Jiao, L., Yan, H., Wu, Y., Chen, L., Gu, W., Du, D., Lin, Y., & Zhu, C. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.9b03004 · Cited by 395 · see citing works
Liu, Y., Wang, H., Zhao, W., Zhang, M., Qin, H., & Xie, Y. (2018) · Sensors · doi:10.3390/s18020645 · Cited by 392 · see citing works
Reviews of wearable healthcare systems: Materials, devices and system integration
Lou, Z., Wang, L., Jiang, K., Wei, Z., & Shen, G. (2019) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2019.100523 · Cited by 389 · see citing works
Self-citationAn Advanced Multifunctional Hydrogel‐Based Dressing for Wound Monitoring and Drug Delivery
Mirani, B., Pagan, E., Currie, B., Siddiqui, M. A., Hosseinzadeh, R., Mostafalu, P., Zhang, Y. S., Ghahary, A., & Akbari, M. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700718 · Cited by 388 · see citing works
Wearable Sensors for Biochemical Sweat Analysis
Bandodkar, A. J., Jeang, W. J., Ghaffari, R., & Rogers, J. A. (2019) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061318-114910 · Cited by 387 · see citing works
Self-citationFlexible Electronics Based on Micro/Nanostructured Paper
Zhang, Y., Zhang, L., Cui, K., Ge, S., Cheng, X. Y., Yan, M., Yu, J., & Liu, H. (2018) · Advanced Materials · doi:10.1002/adma.201801588 · Cited by 381 · see citing works
Wearable flexible sweat sensors for healthcare monitoring: a review
Chung, M., Fortunato, G., & Radacsi, N. (2019) · Journal of The Royal Society Interface · doi:10.1098/rsif.2019.0217 · Cited by 378 · see citing works
Wearable Biosensors: An Alternative and Practical Approach in Healthcare and Disease Monitoring
Sharma, A., Badea, M., Tiwari, S., & Marty, J. (2021) · Molecules · doi:10.3390/molecules26030748 · Cited by 373 · see citing works
Yin, L., Cao, M., Kim, K. N., Lin, M., Moon, J., Sempionatto, J. R., Yu, J., Liu, R., Wicker, C., Trifonov, A., et al. (2022) · Nature Electronics · doi:10.1038/s41928-022-00843-6 · Cited by 370 · see citing works
Choi, J., Bandodkar, A. J., Reeder, J. T., Ray, T. R., Turnquist, A., Kim, S. B., Nyberg, N., Hourlier-Fargette, A., Model, J. B., Aranyosi, A. J., et al. (2019) · ACS Sensors · doi:10.1021/acssensors.8b01218 · Cited by 365 · see citing works
Self-citationFrom Diagnosis to Treatment: Recent Advances in Patient-Friendly Biosensors and Implantable Devices
Li, P., Lee, G., Kim, S. Y., Kwon, S. Y., Kim, H., & Park, S. (2021) · ACS Nano · doi:10.1021/acsnano.0c06688 · Cited by 362 · see citing works
The Era of Digital Health: A Review of Portable and Wearable Affinity Biosensors
Tu, J., Torrente-Rodríguez, R. M., Wang, M., & Gao, W. F. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201906713 · Cited by 359 · see citing works
Battery-free, wireless sensors for full-body pressure and temperature mapping
Han, S., Kim, J., Won, S. M., Ma, Y., Kang, D., Xie, Z., Lee, K., Chung, H. U., Banks, A. R., Min, S., et al. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aan4950 · Cited by 356 · see citing works
Self-citationDevice-assisted transdermal drug delivery
Lee, H., Song, C., Baik, S., Kim, D., Hyeon, T., & Kim, D.-H. (2017) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2017.08.009 · Cited by 347 · see citing works
Mishra, R. K., Hubble, L. J., Martín, A., Kumar, R., Barfidokht, A., Kim, J., Musameh, M. M., Kyratzis, I. L., & Wang, J. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00051 · Cited by 345 · see citing works
Self-citationWang, Y., Zhao, C., Wang, J., Luo, X., Xie, L., Zhan, S., Kim, J., Wang, X., Liu, X., & Ying, Y. B. (2021) · Science Advances · doi:10.1126/sciadv.abe4553 · Cited by 345 · see citing works
Self-citationXu, G., Lu, Y., Cheng, C., Li, X., Xu, J., Liu, Z., Liu, J., Liu, G., Shi, Z., Chen, Z., et al. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202100852 · Cited by 344 · see citing works
Comprehensive Review on Wearable Sweat-Glucose Sensors for Continuous Glucose Monitoring
Zafar, H., Channa, A., Jeoti, V., & Stojanović, G. M. (2022) · Sensors · doi:10.3390/s22020638 · Cited by 333 · see citing works
Yao, S., Ren, P., Song, R., Liu, Y., Huang, Q., Dong, J., O’Connor, B. T., & Zhu, Y. (2019) · Advanced Materials · doi:10.1002/adma.201902343 · Cited by 326 · see citing works
Skin bioelectronics towards long-term, continuous health monitoring
Wang, Y., Haick, H., Guo, S., Wang, C., Lee, S., Yokota, T., & Someya, T. (2022) · Chemical Society Reviews · doi:10.1039/d2cs00207h · Cited by 320 · see citing works
Self-citationAdvanced Materials for Health Monitoring with Skin‐Based Wearable Devices
Jin, H., Abu‐Raya, Y. S., & Haick, H. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700024 · Cited by 317 · see citing works
Choi, J., Kang, D., Han, S., Kim, S. B., & Rogers, J. A. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201601355 · Cited by 317 · see citing works
Self-citationApplied body-fluid analysis by wearable devices
Brasier, N., Wang, J., Gao, W. F., Sempionatto, J. R., Dincer, C., Ates, H. C., Güder, F., Olenik, S., Schauwecker, I., Schaffarczyk, D., et al. (2024) · Nature · doi:10.1038/s41586-024-08249-4 · Cited by 316 · see citing works
Self-citationRecent Progress in Inkjet‐Printed Thin‐Film Transistors
Chung, S., Cho, K., & Lee, T. (2019) · Advanced Science · doi:10.1002/advs.201801445 · Cited by 316 · see citing works
Eyeglasses based wireless electrolyte and metabolite sensor platform
Sempionatto, J. R., Nakagawa, T., Pavinatto, A., Mensah, S. T., Imani, S., Mercier, P., & Wang, J. (2017) · Lab on a Chip · doi:10.1039/c7lc00192d · Cited by 311 · see citing works
Functional Fiber Materials to Smart Fiber Devices
Chen, C., Feng, J., Li, J., Guo, Y., Shi, X., & Peng, H. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00192 · Cited by 312 · see citing works
Methylxanthine Drug Monitoring with Wearable Sweat Sensors
Tai, L., Gao, W., Chao, M., Bariya, M., Ngo, Q. P., Shahpar, Z., Nyein, H. Y. Y., Park, H., Sun, J., Jung, Y., et al. (2018) · Advanced Materials · doi:10.1002/adma.201707442 · Cited by 310 · see citing works
Reeder, J. T., Choi, J., Xue, Y., Gutruf, P., Hanson, J., Liu, M., Ray, T., Bandodkar, A. J., Avila, R., Xia, W., et al. (2019) · Science Advances · doi:10.1126/sciadv.aau6356 · Cited by 308 · see citing works
Self-citationStrategies for body-conformable electronics
Liu, S., Rao, Y., Jang, H., Tan, P., & Lu, N. (2022) · Matter · doi:10.1016/j.matt.2022.02.006 · Cited by 308 · see citing works
Mayer, M., & Baeumner, A. J. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00719 · Cited by 306 · see citing works
Bioinspired Adhesive Architectures: From Skin Patch to Integrated Bioelectronics
Baik, S., Lee, H. J., Kim, D. W., Kim, J. W., Lee, Y., & Pang, C. (2019) · Advanced Materials · doi:10.1002/adma.201803309 · Cited by 305 · see citing works
Self-citationJao, Y.-T., Yang, P., Chiu, C.-M., Lin, Y.-J., Chen, S.-W., Choi, D., & Lin, Z. (2018) · Nano Energy · doi:10.1016/j.nanoen.2018.05.071 · Cited by 296 · see citing works
Xiao, J., Liu, Y., Su, L., Zhao, D., Zhao, L., & Zhang, X. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b03110 · Cited by 294 · see citing works
Smart Sensor Systems for Wearable Electronic Devices
An, B. W., Shin, J. H., Kim, S.-Y., Kim, J., Ji, S., Park, J., Lee, Y., Jang, J., Park, Y., Cho, E., et al. (2017) · Polymers · doi:10.3390/polym9080303 · Cited by 293 · see citing works
Self-citationBioinspired Composite Microfibers for Skin Adhesion and Signal Amplification of Wearable Sensors
Drotlef, D., Amjadi, M., Yunusa, M., & Sitti, M. (2017) · Advanced Materials · doi:10.1002/adma.201701353 · Cited by 292 · see citing works
Lee, I., Probst, D., Klonoff, D. C., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 291 · see citing works
Wearable plasmonic paper–based microfluidics for continuous sweat analysis
Mogera, U., Guo, H., Namkoong, M., Rahman, M. S., Nguyen, N. T., & Tian, L. (2022) · Science Advances · doi:10.1126/sciadv.abn1736 · Cited by 288 · see citing works
Yamamoto, Y., Yamamoto, D., Takada, M., Naito, H., Arie, T., Akita, S., & Takei, K. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700495 · Cited by 283 · see citing works
Dong, K., Deng, J., Ding, W., Wang, A. C., Wang, P., Cheng, C., Wang, Y., Jin, L., Gu, B., Sun, B., et al. (2018) · Advanced Energy Materials · doi:10.1002/aenm.201801114 · Cited by 283 · see citing works
A Fully Integrated and Self-Powered Smartwatch for Continuous Sweat Glucose Monitoring
Zhao, J., Lin, Y., Wu, J., Nyein, H. Y. Y., Bariya, M., Tai, L., Chao, M., Ji, W., Zhang, G. B., Fan, Z., et al. (2019) · ACS Sensors · doi:10.1021/acssensors.9b00891 · Cited by 283 · see citing works
- Authors
- Koh, A., Kang, D., Xue, Y., Lee, S., Pielak, R. M., Kim, J., Hwang, T., Min, S., Banks, A., Bastien, P., Manco, M. C., Wang, L., Ammann, K. R., Jang, K.-I., Won, P., Han, S., Ghaffari, R., Paik, U., Slepian, M. J., Balooch, G., Huang, Y., & Rogers, J. A.
- Affiliations
- University of Illinois Urbana-Champaign (US)
- Northwestern University (US)
- L’Oréal Technology Incubator, San Francisco, CA 94107, USA.
- L'Oréal (France) (FR)
- L’Oréal Early Clinical, Clark, NJ 07066, USA.
- University of Arizona (US)
- MC10 (United States) (US)
- Hanyang University (KR)
- Corresponding author
- Koh, Ahyeon
- Kang, Daeshik
- Rogers, John A.
- ORCID
- Venue
- Science Translational Medicine
- Published
- 2016-11-23
- Volume
- 8
- Issue
- 366
- Pages
- 366ra165-366ra165
- ISSN
- 1946-6234 (ISSN-L)
- 1946-6242
- PMC
- PMC5429097
- OpenAlex
- W2552540631
- Semantic Scholar
- 84a9944cb082f216174ed23bb7d69071b54fd62b
- Cited by
- 1,308
- References
- 58
- Citations by year
- 2026: 86 · 2025: 129 · 2024: 149 · 2023: 153 · 2022: 168 · 2021: 160 · 2020: 150 · 2019: 162 · 2018: 103 · 2017: 47 · 2016: 1
- FWCI
- 50.08
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Science Translational Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- KAIST Institutional Repository (KAIST) (submittedVersion)
- Funders
- Global Research Laboratory Program (K20704000003TA050000310)
- National Research Foundation of Korea
- NIH (R01EB019337)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
- Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
- Keywords
- sweat analysis, colorimetric sensing, sweat collection, stretchable electronics, wireless communication, chloride determination, lactate measurement, sweat pH, glucose, sweat rate, health monitoring, cycling, sweat loss
- MeSH terms
- Smartphone, Wearable Electronic Devices, Adolescent, Adult, Aged, Child, Chlorides, Colorimetry, Equipment Design, Female, Glucose, Humans, Hydrogen-Ion Concentration, Image Processing, Computer-Assisted, Male, Middle Aged, Sweat, User-Computer Interface, Biosensing Techniques, Lactic Acid, Microfluidics, Young Adult, Lab-On-A-Chip Devices
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review