Factors affecting two-point discrimination in argus II patients
Open accessAbstract
Two of the main obstacles to the development of epiretinal prosthesis technology are electrodes that require current amplitudes above safety limits to reliably elicit percepts, and a failure to consistently elicit pattern vision. Here, we explored the causes of high current amplitude thresholds and poor spatial resolution within the Argus II epiretinal implant. We measured current amplitude thresholds and two-point discrimination (the ability to determine whether one or two electrodes had been stimulated) in 3 blind participants implanted with Argus II devices. Our data and simulations show that axonal stimulation, lift and retinal damage all play a role in reducing performance in the Argus 2, by either limiting sensitivity and/or reducing spatial resolution. Understanding the relative role of these various factors will be critical for developing and surgically implanting devices that can successfully subserve pattern vision.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Several factors might limit the spatial resolution of prosthetic vision, including retinal damage, electrode-retina distance, and inadvertent stimulation of nerve fiber bundles.
“Most recently, Yücel et al (2022) identified several factors that might limit the spatial resolution of prosthetic vision, which included retinal damage, electrode-retina distance, and the inadvertent stimulation of nerve fiber bundles.”
From Axonal stimulation affects the linear summation of single-point perception in three Argus II users (2024) · cites this paper as Yücel et al 2022
Phosphenes tended to appear larger and rounder as stimulus amplitude increased.
“We found that phosphenes tended to appear larger and rounder as stimulus amplitude increased (table 4), which is consistent with previous considerations about the current spread in the retina (de Balthasar et al 2008, Granley and Beyeler 2021, Yücel et al 2022).”
From Axonal stimulation affects the linear summation of single-point perception in three Argus II users (2024) · cites this paper as Yücel et al 2022
The probability of perceiving two distinct phosphenes increases with inter-electrode distance.
“Yücel et al (2022) previously demonstrated that the probability of perceiving two distinct phosphenes increases with inter-electrode distance.”
From Axonal stimulation affects the linear summation of single-point perception in three Argus II users (2024) · cites this paper as Yücel et al (2022)
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Wong, W. L., Su, X., Li, X., Cheung, C. M. G., Klein, R., Cheng, C., & Wong, T. Y. (2014) · The Lancet Global Health · doi:10.1016/s2214-109x(13)70145-1 · Cited by 5,403 · see citing works
The psychometric function: I. Fitting, sampling, and goodness of fit
Wichmann, F. A., & Hill, N. J. (2001) · Perception & Psychophysics · doi:10.3758/bf03194544 · Cited by 2,359 · see citing works
Age-related macular degeneration
Mitchell, P., Liew, G., Gopinath, B., & Wong, T. Y. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31550-2 · Cited by 1,572 · see citing works
Applied Linear Statistical Models
John, R. C. S. (1986) · Journal of Quality Technology · doi:10.1080/00224065.1986.11978989 · Cited by 1,320 · see citing works
Interim results from the international trial of Second Sight's visual prosthesis
Humayun, M. S., Dorn, J. D., da Cruz, L., Dagnelie, G., Sahel, J.-A., Stanga, P. E., Cideciyan, A. V., Duncan, J. L., Eliott, D., Filley, E., et al. (2012) · Ophthalmology · doi:10.1016/j.ophtha.2011.09.028 · Cited by 700 · see citing works
Self-citationDendritic field size and morphology of midget and parasol ganglion cells of the human retina.
Dacey, D. M., & Petersen, M. R. (1992) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.89.20.9666 · Cited by 495 · see citing works
Discrimination at threshold: Labelled detectors in human vision
Watson, A. B., & Robson, J. G. (1981) · Vision Research · doi:10.1016/0042-6989(81)90014-6 · Cited by 423 · see citing works
Five-year safety and performance results from the argus II retinal prosthesis system clinical trial.
da Cruz, L., Dorn, J. D., Humayun, M. S., Dagnelie, G., Handa, J., Barale, P.-O., Sahel, J.-A., Stanga, P. E., Hafezi, F., Safran, A. B., et al. (2016) · Ophthalmology · doi:10.1016/j.ophtha.2016.06.049 · Cited by 359 · see citing works
Self-citationFirst-in-human trial of a novel suprachoroidal retinal prosthesis
Ayton, L. N., Blamey, P. J., Guymer, R. H., Luu, C. D., Nayagam, D. A. X., Sinclair, N. C., Shivdasani, M. N., Yeoh, J., McCombe, M. F., Briggs, R. J., et al. (2014) · PLoS ONE · doi:10.1371/journal.pone.0115239 · Cited by 333 · see citing works
Non-linear projection of the retinal image in a wide-angle schematic eye.
Drasdo, N., & Fowler, C. W. (1974) · British Journal of Ophthalmology · doi:10.1136/bjo.58.8.709 · Cited by 306 · see citing works
Ahuja, A. K., Dorn, J. D., Caspi, A., McMahon, M. J., Dagnelie, G., daCruz, L., Stanga, P., Humayun, M. S., Greenberg, R. J., & Argus II Study Group (2011) · British Journal of Ophthalmology · doi:10.1136/bjo.2010.179622 · Cited by 289 · see citing works
Self-citationLong-Term Results from an Epiretinal Prosthesis to Restore Sight to the Blind
Ho, A. C., Humayun, M. S., Dorn, J. D., da Cruz, L., Dagnelie, G., Handa, J. T., Barale, P. O., Sahel, J., Stanga, P. E., Hafezi, F., et al. (2015) · Ophthalmology · doi:10.1016/j.ophtha.2015.04.032 · Cited by 262 · see citing works
Self-citationAn update on retinal prostheses
Ayton, L. N., Barnes, N., Dagnelie, G., Fujikado, T., Goetz, G., Hornig, R., Jones, B. W., Muqit, M. M. K., Rathbun, D. L., Stingl, K., et al. (2020) · Clinical Neurophysiology · doi:10.1016/j.clinph.2019.11.029 · Cited by 223 · see citing works
Self-citationAxonal Sodium-Channel Bands Shape the Response to Electric Stimulation in Retinal Ganglion Cells
Fried, S. I., Lasker, A., Desai, N. J., Eddington, D. K., & Rizzo, J. F. (2009) · Journal of Neurophysiology · doi:10.1152/jn.91081.2008 · Cited by 206 · see citing works
Factors affecting perceptual thresholds in epiretinal prostheses
de Balthasar, C., Patel, S., Roy, A., Freda, R., Greenwald, S., Horsager, A., Mahadevappa, M., Yanai, D., McMahon, M. J., Humayun, M. S., et al. (2008) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.07-0696 · Cited by 204 · see citing works
Self-citationThe functional performance of the Argus II retinal prosthesis
Stronks, H. C., & Dagnelie, G. (2014) · Expert Review of Medical Devices · doi:10.1586/17434440.2014.862494 · Cited by 189 · see citing works
Self-citationDesign and validation of a foldable and photovoltaic wide-field epiretinal prosthesis
Ferlauto, L., Airaghi Leccardi, M. J. I., Chenais, N. A. L., Gilliéron, S. C. A., Vagni, P., Bevilacqua, M., Wolfensberger, T. J., Sivula, K., & Ghezzi, D. (2018) · Nature Communications · doi:10.1038/s41467-018-03386-7 · Cited by 183 · see citing works
Persistent remodeling and neurodegeneration in late-stage retinal degeneration
Pfeiffer, R. L., Marc, R. E., & Jones, B. W. (2019) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2019.07.004 · Cited by 178 · see citing works
Nanduri, D., Fine, I., Horsager, A., Boynton, G. M., Humayun, M. S., Greenberg, R. J., & Weiland, J. D. (2012) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.11-8401 · Cited by 163 · see citing works
Jensen, R. J., Ziv, O. R., & Rizzo, J. F. (2005) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.04-1018 · Cited by 159 · see citing works
Jansonius, N. M., Nevalainen, J., Selig, B., Zangwill, L. M., Sample, P. A., Budde, W. M., Jonas, J. B., Lagrèze, W. A., Airaksinen, P. J., Vonthein, R., et al. (2009) · Vision Research · doi:10.1016/j.visres.2009.04.029 · Cited by 155 · see citing works
A model of ganglion axon pathways accounts for percepts elicited by retinal implants
Beyeler, M., Nanduri, D., Weiland, J. D., Rokem, A., Boynton, G. M., & Fine, I. (2019) · Scientific Reports · doi:10.1038/s41598-019-45416-4 · Cited by 143 · see citing works
Self-citationFactors Affecting Perceptual Threshold in Argus II Retinal Prosthesis Subjects
Ahuja, A. K., Yeoh, J., Dorn, J. D., Caspi, A., Wuyyuru, V., McMahon, M. J., Humayun, M. S., Greenberg, R. J., daCruz, L., & Argus II Study Group (2013) · Translational Vision Science & Technology · doi:10.1167/tvst.2.4.1 · Cited by 123 · see citing works
Retinitis pigmentosa and allied conditions today: a paradigm of translational research
Ayuso, C., & Millán, J. M. (2010) · Genome Medicine · doi:10.1186/gm155 · Cited by 121 · see citing works
Temporal properties of visual perception on electrical stimulation of the retina
Pérez Fornos, A., Sommerhalder, J., da Cruz, L., Sahel, J. A., Mohand-Said, S., Hafezi, F., & Pelizzone, M. (2012) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.11-9344 · Cited by 113 · see citing works
AnIn VitroModel of a Retinal Prosthesis
Ahuja, A. K., Behrend, M. R., Kuroda, M., Humayun, M. S., & Weiland, J. D. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.919126 · Cited by 111 · see citing works
Predicting visual sensitivity in retinal prosthesis patients
Horsager, A., Greenwald, S. H., Weiland, J. D., Humayun, M. S., Greenberg, R. J., McMahon, M. J., Boynton, G. M., & Fine, I. (2009) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.08-2595 · Cited by 107 · see citing works
Self-citationBrightness as a Function of Current Amplitude in Human Retinal Electrical Stimulation
Greenwald, S. H., Horsager, A., Humayun, M. S., Greenberg, R. J., McMahon, M. J., & Fine, I. (2009) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.08-2897 · Cited by 95 · see citing works
Self-citationThe Argus II Retinal Prosthesis: 12-Month Outcomes from a Single-Study Center
Rizzo, S., Belting, C., Cinelli, L., Allegrini, L., Genovesi-Ebert, F., Barca, F., & Bartolo, E. D. (2014) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2014.02.039 · Cited by 83 · see citing works
Erickson-Davis, C., & Korzybska, H. (2021) · PLoS ONE · doi:10.1371/journal.pone.0229189 · Cited by 83 · see citing works
Factors Affecting Perceptual Thresholds in a Suprachoroidal Retinal Prosthesis
Shivdasani, M. N., Sinclair, N. C., Dimitrov, P. N., Varsamidis, M., Ayton, L. N., Luu, C. D., Perera, T., McDermott, H. J., & Blamey, P. J. (2014) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.14-14396 · Cited by 78 · see citing works
Ayton, L. N., Rizzo, J. F., Bailey, I. L., Colenbrander, A., Dagnelie, G., Geruschat, D. R., Hessburg, P. C., McCarthy, C. D., Petoe, M. A., Rubin, G. S., et al. (2020) · Translational Vision Science & Technology · doi:10.1167/tvst.9.8.25 · Cited by 75 · see citing works
Self-citationFine, I., & Boynton, G. M. (2015) · Philosophical Transactions of the Royal Society B: Biological Sciences · doi:10.1098/rstb.2014.0208 · Cited by 55 · see citing works
Self-citationLuo, Y. H.-L., Zhong, J. J., Clemo, M., & da Cruz, L. (2016) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2016.07.021 · Cited by 47 · see citing works
One-Year Safety and Performance Assessment of the Argus II Retinal Prosthesis: A Postapproval Study
Schaffrath, K., Schellhase, H., Walter, P., Augustin, A., Chizzolini, M., Kirchhof, B., Grisanti, S., Wiedemann, P., Szurman, P., Richard, G., et al. (2019) · JAMA Ophthalmology · doi:10.1001/jamaophthalmol.2019.1476 · Cited by 44 · see citing works
Retinotopic to spatiotopic mapping in blind patients implanted with the Argus II retinal prosthesis
Caspi, A., Roy, A., Dorn, J. D., & Greenberg, R. J. (2017) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.16-20398 · Cited by 32 · see citing works
Esler, T. B., Kerr, R. R., Tahayori, B., Grayden, D. B., Meffin, H., & Burkitt, A. N. (2018) · PLoS ONE · doi:10.1371/journal.pone.0193598 · Cited by 28 · see citing works
pulse2percept: A Python-based simulation framework for bionic visionPreprint
Beyeler, M., Boynton, G. M., Fine, I., & Rokem, A. (2017) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/148015 · Cited by 27 · see citing works
Self-citationGregori, N. Z., Callaway, N. F., Hoeppner, C. G., Yuan, A., Rachitskaya, A. V., Feuer, W. J., Ameri, H., Arévalo, J. F., Augustin, A. J., Birch, D. G., et al. (2018) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2018.06.012 · Cited by 27 · see citing works
Self-citationRizzo, S., Cinelli, L., Finocchio, L., Tartaro, R., Santoro, F., & Gregori, N. Z. (2019) · JAMA Ophthalmology · doi:10.1001/jamaophthalmol.2018.6375 · Cited by 20 · see citing works
Spatially patterned bi-electrode epiretinal stimulation for axon avoidance at cellular resolution
Vilkhu, R. S., Madugula, S. S., Grosberg, L. E., Gogliettino, A. R., Hottowy, P., Dąbrowski, W., Sher, A., Litke, A. M., Mitra, S., & Chichilnisky, E. J. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac3450 · Cited by 18 · see citing works
Assessing the utility of visual acuity measures in visual prostheses
Caspi, A., & Zivotofsky, A. Z. (2015) · Vision Research · doi:10.1016/j.visres.2015.01.006 · Cited by 17 · see citing works
Arévalo, J. F., Rashaed, S. A., Alhamad, T. A., Kahtani, E. A., Al-Dhibi, H. A., for the KKESH Collaborative Retina Study Group, Mura, M., Kahtani, E. A., Nowilaty, S. R., Rashaed, S. A., et al. (2021) · International Journal of Retina and Vitreous · doi:10.1186/s40942-021-00324-6 · Cited by 17 · see citing works
Testing vision is not testing for vision
Peli, E. (2020) · Translational Vision Science & Technology · doi:10.1167/tvst.9.13.32 · Cited by 14 · see citing works
Lin, T., Wang, L., Lan, Y., Zhang, Y., Falabella, P., Zhu, D., Hinton, D. R., Rao, N. A., Birch, D. G., Spencer, R., et al. (2019) · Translational Vision Science & Technology · doi:10.1167/tvst.8.3.31 · Cited by 13 · see citing works
Naidu, A., Ghani, N., Yazdanie, M. S., et al. (2020) · BMC Ophthalmology · doi:10.1186/s12886-020-01631-6 · Cited by 12 · see citing works
Xu, L. T., Rachitskaya, A. V., DeBenedictis, M. J., Bena, J., Morrison, S. L., & Yuan, A. (2019) · European Journal of Ophthalmology · doi:10.1177/1120672119885799 · Cited by 9 · see citing works
Jansonius, N. M., & Schiefer, U. (2020) · Translational Vision Science & Technology · doi:10.1167/tvst.9.11.3 · Cited by 8 · see citing works
Inter-electrode Discriminability Correlates With Spatial Visual Performance In ArgusTM II Subjects
Lauritzen, T. Z., Nanduri, D., Weiland, J., Dorn, J. D., McClure, K., & Greenberg, R. J. (2011) · Investigative Ophthalmology & Visual Science · Cited by 7 · see citing works
EPIRETINAL FIBROSIS REMOVAL IN AN ARGUS II–IMPLANTED EYE
Patelli, F., Falleni, M., Colombo, L., Martinelli, C., Tosi, D., Bulfamante, G., & Rossetti, L. M. (2020) · Retina · doi:10.1097/iae.0000000000002780 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Axonal stimulation affects the linear summation of single-point perception in three Argus II users
Hou, Y., Nanduri, D., Granley, J., Weiland, J. D., & Beyeler, M. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad31c4 · Cited by 14 · see citing works
Self-citation
- Authors
- Yücel, E. I., Sadeghi, R., Kartha, A., Montezuma, S. R., Dagnelie, G., Rokem, A., Boynton, G. M., Fine, I., & Beyeler, M.
- Affiliations
- Johns Hopkins University (US)
- University of Minnesota (US)
- University of Washington (US)
- University of California, Santa Barbara (US)
- Corresponding author
- Fine, Ione
- ORCID
- Venue
- Frontiers in Neuroscience
- Publisher
- Frontiers Media SA
- Published
- 2022-08-24
- Volume
- 16
- Pages
- 901337-901337
- ISSN
- 1662-453X (ISSN-L)
- PMC
- PMC9448992
- OpenAlex
- W4292860335
- Cited by
- 13
- References
- 50
- Citations by year
- 2026: 5 · 2025: 3 · 2024: 3 · 2023: 1 · 2022: 1
- FWCI
- 0.88
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Frontiers in Neuroscience (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Funders
- National Eye Institute (R01-62-5032, R00 EY02932)
- Washington Research Foundation
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Keywords
- Argus II, epiretinal prosthesis, two-point discrimination, spatial resolution, retinal damage, pattern vision, visual prosthesis, blind patients
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review