Biosensors for real-time in vivo measurements
Abstract
The current status of sensors capable of continuous measurement of analytes in biological media is reviewed. This review containing 173 references deals with devices whose use in single cells, tissue slices, animal models and humans has been demonstrated. In addition to sensors specific for glucose, lactate, glutamate, pyruvate, choline and acetylcholine, insights obtained from monitoring nitric oxide, Na(+), K(+), Ca(2+), and dopamine are presented. Performance criteria for sensor performance are described as is the subject of biosensor calibration. Biocompatibility issues are dealt with in some detail as is the status of continuous blood glucose monitoring in humans.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“As described earlier, the use of mediators to shuttle electrons between the enzyme and the electrode can also minimize the impact of interfering species by lowering the working potential required to oxidize hydrogen peroxide.56,58,62”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [58]
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY
Clark, L. C., & Lyons, C. (1962) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1962.tb13623.x · Cited by 3,945
HEPATOCELLULAR CARCINOMA AND HEPATITIS B VIRUS
Beasley, R. P., Lin, C., Hwang, L., & Chien, C. (1981) · The Lancet · doi:10.1016/s0140-6736(81)90585-7 · Cited by 2,856
Electrochemical biosensors: recommended definitions and classification
Thévenot, D. R., Tóth, K., Durst, R. A., & Wilson, G. S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00115-4 · Cited by 1,858
Self-citationEvaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Neural activity triggers neuronal oxidative metabolism followed by astrocytic glycolysis
Kasischke, K. A., Vishwasrao, H. D., Fisher, P. J., Zipfel, W. R., & Webb, W. W. (2004) · Science · doi:10.1126/science.1096485 · Cited by 742
Surveying for Surfaces that Resist the Adsorption of Proteins
Chapman, R., Ostuni, E., Takayama, S., Holmlin, R. E., Yan, L., & Whitesides, G. M. (2000) · Journal of the American Chemical Society · doi:10.1021/ja000774f · Cited by 600
Methods for reducing biosensor membrane biofouling
Wisniewski, N., & Reichert, M. (2000) · Colloids and Surfaces B: Biointerfaces · doi:10.1016/s0927-7765(99)00148-4 · Cited by 543
The Encyclopedia of Electrochemistry
(1965) · Students Quarterly Journal · doi:10.1049/sqj.1965.0065 · Cited by 501
Daunert, S., Barrett, G., Feliciano, J. S., Shetty, R. S., Shrestha, S., & Smith‐Spencer, W. (2000) · Chemical Reviews · doi:10.1021/cr990115p · Cited by 446
Enzyme-Based Biosensors for in Vivo Measurements
Wilson, G. S., & Hu, Y. (2000) · Chemical Reviews · doi:10.1021/cr990003y · Cited by 434
Self-citationPsychoanalytical Electrochemistry: Dopamine and Behavior
Venton, B. J., & Wightman, R. M. (2003) · Analytical Chemistry · doi:10.1021/ac031421c · Cited by 423
The MiniMed Continuous Glucose Monitoring System
Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050214078 · Cited by 394
Development and Application of Cell-Based Biosensors
Pancrazio, J. J., Whelan, J., Borkholder, D. A., Ma, W., & Stenger, D. A. (1999) · Annals of Biomedical Engineering · doi:10.1114/1.225 · Cited by 391
Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390
Rebrin, K., Steil, G. M., van Antwerp, W. P., & Mastrototaro, J. J. (1999) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1999.277.3.e561 · Cited by 367
Subsecond Adsorption and Desorption of Dopamine at Carbon-Fiber Microelectrodes
Bath, B. D., Michael, D. J., Venton, B. J., Joseph, J. D., Runnels, P. L., & Wightman, R. M. (2000) · Analytical Chemistry · doi:10.1021/ac000849y · Cited by 363
Characterization of implantable biosensor membrane biofouling
Wisniewski, N. A., Moussy, F. G., & Reichert, W. M. (2000) · Fresenius Journal of Analytical Chemistry · doi:10.1007/s002160051556 · Cited by 351
Antioxidant defense of the brain: a role for astrocytes
Wilson, J. X. (1997) · Canadian Journal of Physiology and Pharmacology · doi:10.1139/y97-146 · Cited by 349
Hu, Y., & Wilson, G. S. (1997) · Journal of Neurochemistry · doi:10.1046/j.1471-4159.1997.69041484.x · Cited by 337
Self-citationChapter 4 Mechanisms of inflammation and infection with implanted devices
Anderson, J. M. (1993) · Cardiovascular Pathology · doi:10.1016/1054-8807(93)90045-4 · Cited by 335
Closed-loop insulin delivery-the path to physiological glucose control.
Steil, G., Panteleon, A., & Rebrin, K. (2004) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2003.08.011 · Cited by 325
WEARABLE ARTIFICIAL ENDOCRINE PANCREAS WITH NEEDLE-TYPE GLUCOSE SENSOR
Shichiri, M., Yamasaki, Y., Kawamori, R., Hakui, N., & Abe, H. (1982) · The Lancet · doi:10.1016/s0140-6736(82)92788-x · Cited by 323
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312
Surface modification of polymers: chemical, biological and surface analytical challenges
Ratner, B. D. (1995) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(95)99218-a · Cited by 304
Burmeister, J. J., & Gerhardt, G. A. (2001) · Analytical Chemistry · doi:10.1021/ac0010429 · Cited by 288
Noninvasive Glucose Monitoring
Tamada, J. A., Garg, S. K., Jovanovič, L., Pitzer, K. R., Fermi, S. J., & Potts, R. O. (1999) · JAMA · doi:10.1001/jama.282.19.1839 · Cited by 281
Aptamer-based biosensor arrays for detection and quantification of biological macromolecules
McCauley, T. G., Hamaguchi, N., & Stanton, M. J. (2003) · Analytical Biochemistry · doi:10.1016/s0003-2697(03)00297-5 · Cited by 268
o-Phenylenediamine-Modified Carbon Fiber Electrodes for the Detection of Nitric Oxide
Friedemann, M. N., Robinson, S. W., & Gerhardt, G. A. (1996) · Analytical Chemistry · doi:10.1021/ac960093w · Cited by 266
Hu, Y., Mitchell, K. M., Albahadily, F., Michaelis, E. K., & Wilson, G. S. (1994) · Brain Research · doi:10.1016/0006-8993(94)90870-2 · Cited by 252
Self-citationPhysiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 252
Optical biosensors : present and future
Ligler, F. S., & Taitt, C. A. R. (2002) · Cited by 241
Engineering the tissue which encapsulates subcutaneous implants. I. Diffusion properties
Sharkawy, A. A., Klitzman, B. M., Truskey, G. A., & Reichert, W. M. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19971205)37:3 · Cited by 239
Kulagina, N. V., Shankar, L., & Michael, A. C. (1999) · Analytical Chemistry · doi:10.1021/ac990636c · Cited by 230
Two-dimensional enzyme electrode sensor for glucose
Gough, D. A., Lucisano, J. Y., & Tse, P. H. S. (1985) · Analytical Chemistry · doi:10.1021/ac00289a042 · Cited by 227
Effect of chemical treatments on tendon cellularity and mechanical properties
Cartmell, J. S., & Dunn, M. G. (2000) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(200001)49:1 · Cited by 226
Enzyme-induced biodegradation of polycarbonate-polyurethanes: Dependence on hard-segment chemistry
Tang, Y. W., Labow, R. S., & Santerre, J. P. (2001) · Journal of Biomedical Materials Research · doi:10.1002/1097-4636(20011215)57:4 · Cited by 198
Hickey, T., Kreutzer, D. L., Burgess, D. J., & Moussy, F. G. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10016 · Cited by 187
Aussedat, B., Dupire-Angel, M., Gifford, R., Klein, J. C., Wilson, G. S., & Reach, G. (2000) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.2000.278.4.e716 · Cited by 186
Self-citationRapid Changes in Local Extracellular Rat Brain Glucose Observed with an In Vivo Glucose Sensor
Hu, Y., & Wilson, G. S. (1997) · Journal of Neurochemistry · doi:10.1046/j.1471-4159.1997.68041745.x · Cited by 186
Self-citationFeldman, B., Brazg, R., Schwartz, S. L., & Weinstein, R. L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526978 · Cited by 184
Lowry, J. P., McAteer, K., Atrash, S. S. E., Duff, A., & O'Neill, R. D. (1994) · Analytical Chemistry · doi:10.1021/ac00082a025 · Cited by 183
Emerging biomedical sensing technologies and their applications
Cote, G. L., Lec, R. M., & Pishko, M. V. (2003) · IEEE Sensors Journal · doi:10.1109/jsen.2003.814656 · Cited by 183
An ultrastructural analysis of tissue surrounding a microdialysis probe
Clapp-Lilly, K. L., Roberts, R. C., Duffy, L. K., Irons, K. P., Hu, Y. S., & Drew, K. L. (1999) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(99)00064-3 · Cited by 181
In vivo leukocyte cytokine mRNA responses to biomaterials are dependent on surface chemistry
Brodbeck, W. G., Voskerician, G., Ziats, N. P., Nakayama, Y., Matsuda, T., & Anderson, J. M. (2003) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.10425 · Cited by 181
Response Times of Carbon Fiber Microelectrodes to Dynamic Changes in Catecholamine Concentration
Venton, B. J., Troyer, K. P., & Wightman, R. M. (2001) · Analytical Chemistry · doi:10.1021/ac010819a · Cited by 174
Bungay, P. M., Newton‐Vinson, P., Isele, W., Garris, P. A., & Justice, J. B. (2003) · Journal of Neurochemistry · doi:10.1046/j.1471-4159.2003.01904.x · Cited by 173
Quinn, C. P., Pathak, C. P., Heller, A., & Hubbell, J. A. (1995) · Biomaterials · doi:10.1016/0142-9612(95)98856-9 · Cited by 172
Stability of Oxidases Immobilized in Silica Gels
Chen, Q., Kenausis, G. L., & Heller, A. (1998) · Journal of the American Chemical Society · doi:10.1021/ja971750k · Cited by 168
Smith, P. J., Hammar, K. M., Porterfield, D. M., Sanger, R. H., & Trimarchi, J. R. (1999) · Microscopy Research and Technique · doi:10.1002/(sici)1097-0029(19990915)46:6 · Cited by 157
Avshalumov, M. V., Chen, B. T., Marshall, S. P., Peña, D. M., & Rice, M. E. (2003) · Journal of Neuroscience · doi:10.1523/jneurosci.23-07-02744.2003 · Cited by 156
Acetylcholine and Choline Amperometric Enzyme Sensors Characterized in Vitro and in Vivo
Mitchell, K. M. (2004) · Analytical Chemistry · doi:10.1021/ac034757v · Cited by 153
Fluorescent nano-PEBBLE sensors designed for intracellular glucose imaging
Xu, H., Aylott, J. W., & Kopelman, R. (2002) · The Analyst · doi:10.1039/b202782h · Cited by 143
A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.
Updike, S. J., Shults, M. C., Gilligan, B. J., & Rhodes, R. K. (2000) · Diabetes Care · doi:10.2337/diacare.23.2.208 · Cited by 135
von Recum, A. F. (1986) · Medical Entomology and Zoology · Cited by 134
Biosensors: new approaches in drug discovery
Keusgen, M. (2002) · Die Naturwissenschaften · doi:10.1007/s00114-002-0358-3 · Cited by 134
Optical Imaging Fiber-Based Single Live Cell Arrays: A High-Density Cell Assay Platform
Biran, I., & Walt, D. R. (2002) · Analytical Chemistry · doi:10.1021/ac020009e · Cited by 132
Performance of subcutaneously implanted glucose sensors for continuous monitoring
Gerritsen, M. (1999) · The Netherlands Journal of Medicine · doi:10.1016/s0300-2977(99)00006-6 · Cited by 122
Chen, X., Matsumoto, N., Hu, Y., & Wilson, G. S. (2002) · Analytical Chemistry · doi:10.1021/ac015628m · Cited by 120
Self-citationKennedy, R. T., Thompson, J. E., & Vickroy, T. W. (2002) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(01)00506-4 · Cited by 119
Eklund, S. E., Taylor, D. E., Kozlov, E., Prokop, A., & Cliffel, D. E. (2003) · Analytical Chemistry · doi:10.1021/ac034641z · Cited by 119
A model for biocompatibility and its evaluation
Williams, D. F. (1989) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(89)90138-6 · Cited by 119
Pellerin, L., & Magistretti, P. J. (2004) · Science · doi:10.1126/science.1100428 · Cited by 117
Dempsey, E., Diamond, D., Smyth, M. R., Urban, G. A., Jobst, G., Moser, I., Verpoorte, E. M. J., Manz, A., Widmer, H. M., Rabenstein, K., et al. (1997) · Analytica Chimica Acta · doi:10.1016/s0003-2670(97)90075-1 · Cited by 112
Critical guidelines for validation of the selectivity of in-vivo chemical microsensors
Phillips, P. E. M., & Wightman, R. M. (2003) · TrAC Trends in Analytical Chemistry · doi:10.1016/s0165-9936(03)00907-5 · Cited by 110
Sharkawy, A. A., Klitzman, B. M., Truskey, G. A., & Reichert, W. M. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19980615)40:4 · Cited by 110
Metabolic Oscillations in β-Cells
Kennedy, R. T., Kauri, L. M., Dahlgren, G. M., & Jung, S. (2002) · Diabetes · doi:10.2337/diabetes.51.2007.s152 · Cited by 110
Lowry, J. P., Miele, M., O'Neill, R. D., Boutelle, M. G., & Fillenz, M. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(97)00171-4 · Cited by 109
Efficacy and Reliability of the Continuous Glucose Monitoring System
Gross, T. M., & Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050214087 · Cited by 108
Lowry, J. P., O'Neill, R. D., Boutelle, M. G., & Fillenz, M. (1998) · Journal of Neurochemistry · doi:10.1046/j.1471-4159.1998.70010391.x · Cited by 105
Determination of Plasma Glucose During Rapid Glucose Excursions with a Subcutaneous Glucose Sensor
Steil, G. M., Rebrin, K., Mastrototaro, J. J., Bernaba, B., & Saad, M. F. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503763816436 · Cited by 104
Poscia, A., Mascini, M., Moscone, D., Luzzana, M., Caramenti, G., Cremonesi, P., Valgimigli, F., Bongiovanni, C., & Varalli, M. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00216-6 · Cited by 101
Taylor, C., Kenausis, G., Katakis, I., & Heller, A. (1995) · Journal of Electroanalytical Chemistry · doi:10.1016/0022-0728(95)04080-8 · Cited by 100
Zeng, G., Gao, L., Birklé, S., & Yu, R. K. (2000) · PubMed · Cited by 98
Mediation of in vivo glucose sensor inflammatory response via nitric oxide release
Gifford, R., Batchelor, M. M., Lee, Y., Gokulrangan, G., Meyerhoff, M. E., & Wilson, G. S. (2005) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.30359 · Cited by 96
Self-citationGlucose microbiosensor based on alumina sol–gel matrix/electropolymerized composite membrane
Chen, X., Hu, Y., & Wilson, G. S. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00093-3 · Cited by 95
Self-citationPoitout, V., Moatti-Sirat, D., Reach, G., Zhang, Y., Wilson, G. S., Lemonnier, F., & Klein, J. C. (1993) · Diabetologia · doi:10.1007/bf00404077 · Cited by 95
Self-citationMao, L., Osborne, P. G., Yamamoto, K., & Kato, T. (2002) · Analytical Chemistry · doi:10.1021/ac011261+ · Cited by 93
Zhang, Y., & Wilson, G. S. (1993) · Analytica Chimica Acta · doi:10.1016/0003-2670(93)85009-9 · Cited by 92
Self-citationDirect electrochemical measurement of nitric oxide in vascular endothelium
Brovkovych, V. M., Stolarczyk, E. I., Oman, J., Tomboulian, P., & Maliñski, T. (1999) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/s0731-7085(98)00090-9 · Cited by 92
[Measuring EEG complexity for studying the state of mental load].
Liu, J., He, T., Zheng, C., & Huang, Y. (1997) · PubMed · Cited by 91
Fabrication and Characterization of Sputtered-Carbon Microelectrode Arrays
Sreenivas, G., Ang, S. S., Fritsch, I., Brown, W. D., Gerhardt, G. A., & Woodward, D. J. (1996) · Analytical Chemistry · doi:10.1021/ac9508816 · Cited by 89
Influence of inflammatory cells and serum on the performance of implantable glucose sensors
Gerritsen, M., Jansen, J. A., Kros, A., Vriezema, D. M., Sommerdijk, N. A. J. M., Nolte, R. J. M., Lutterman, J. A., van Hövell, S. W. F. M., & van der Gaag, A. (2000) · Journal of Biomedical Materials Research · doi:10.1002/1097-4636(200101)54:1 · Cited by 89
Revzin, A., Sirkar, K., Simonian, A. L., & Pishko, M. V. (2002) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(01)00982-0 · Cited by 88
Heller, J., & Heller, A. (1998) · Journal of the American Chemical Society · doi:10.1021/ja973911q · Cited by 88
Invasive consequences of using micro-electrodes and microdialysis probes in the brain
Khan, A. S., & Michael, A. C. (2003) · TrAC Trends in Analytical Chemistry · doi:10.1016/s0165-9936(03)00908-7 · Cited by 86
Optimization of the Geometry and Porosity of Microelectrode Arrays for Sensor Design
Sandison, M. E., Anicet, N., Glidle, A., & Cooper, J. M. (2002) · Analytical Chemistry · doi:10.1021/ac025649w · Cited by 86
Neurochemistry and electroanalytical probes
Troyer, K. P., Heien, M. L., Venton, B. J., & Wightman, R. M. (2002) · Current Opinion in Chemical Biology · doi:10.1016/s1367-5931(02)00374-5 · Cited by 83
Subcutaneous glucose monitoring by means of electrochemical sensors: fiction or reality?
Rebrin, K., Fischer, U., von Dorsche, H. H., von Woetke, T., Abel, P., & Brunstein, E. (1992) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(92)90033-h · Cited by 81
Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00304-9 · Cited by 76
Self-citationKerner, W., Kiwit, M., Linke, B., Keck, F. S., Zier, H., & Pfeiffer, E. F. (1993) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(93)80032-k · Cited by 75
Simultaneous detection of diverse analytes with an aptazyme ligase array
Hesselberth, J. R., Robertson, M. P., Knudsen, S. M., & Ellington, A. D. (2003) · Analytical Biochemistry · doi:10.1016/s0003-2697(02)00441-4 · Cited by 75
Ceramic-based multisite microelectrode array for rapid choline measures in brain tissue
Burmeister, J. J., Palmer, M. R., & Gerhardt, G. A. (2003) · Analytica Chimica Acta · doi:10.1016/s0003-2670(03)00067-9 · Cited by 74
In vitro and in vivo mineralization of Nafion membrane used for implantable glucose sensors
Mercado, R. C., & Moussy, F. (1998) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(97)00112-7 · Cited by 73
Nitric Oxide-Releasing Sol−Gel Particle/Polyurethane Glucose Biosensors
Shin, J. H., Marxer, S. M., & Schoenfisch, M. H. (2004) · Analytical Chemistry · doi:10.1021/ac049776z · Cited by 71
Direct, Real-Time Sensing of Free Radical Production by Activated Human Glioblastoma Cells
Manning, P., McNeil, C. J., Cooper, J. M., & Hillhouse, E. W. (1998) · Free Radical Biology and Medicine · doi:10.1016/s0891-5849(97)00455-3 · Cited by 70
Fundamental Studies of Glucose Oxidase Deposition on a Pt Electrode
Matsumoto, N., Chen, X., & Wilson, G. S. (2001) · Analytical Chemistry · doi:10.1021/ac015536x · Cited by 67
Self-citationOperation of a Miniature Redox Hydrogel-Based Pyruvate Sensor in Undiluted Deoxygenated Calf Serum
Gajovic, N., Binyamin, G. N., Warsinke, A., Scheller, F. W., & Heller, A. (2000) · Analytical Chemistry · doi:10.1021/ac991021i · Cited by 66
Schmidtke, D. W., & Heller, A. (1998) · Analytical Chemistry · doi:10.1021/ac970932u · Cited by 65
In Vitro and In Vivo Degradation of Glucose Oxidase Enzyme Used for an Implantable Glucose Biosensor
Valdes, T. I., & Moussy, F. G. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194233 · Cited by 64
A Pyruvate Oxidase Electrode Based on an Electrochemically Deposited Redox Polymer
Gajovic, N., Habermüller, K., Warsinke, A., Schuhmann, W. W., & Scheller, F. W. (1999) · Electroanalysis · doi:10.1002/(sici)1521-4109(199912)11:18 · Cited by 63
Fabre, B., Burlet, S., Cespuglio, R., & Bidan, G. (1997) · Journal of Electroanalytical Chemistry · doi:10.1016/s0022-0728(96)04985-6 · Cited by 63
Small-Volume On-Line Sensor for Continuous Measurement of γ-Aminobutyric Acid
Niwa, O., Kurita, R., Horiuchi, T., & Torimitsu, K. (1998) · Analytical Chemistry · doi:10.1021/ac970740z · Cited by 63
Galeska, I., Hickey, T., Moussy, F. G., Kreutzer, D. L., & Papadimitrakopoulos, F. (2001) · Biomacromolecules · doi:10.1021/bm010112y · Cited by 62
Calibration of a subcutaneous amperometric glucose sensor
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V. H., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00306-2 · Cited by 60
Self-citationShram, N. F., Netchiporouk, L. I., & Cespuglio, R. (2002) · European Journal of Neuroscience · doi:10.1046/j.1460-9568.2002.02081.x · Cited by 60
Lowry, J., & Fillenz, M. (2001) · Bioelectrochemistry · doi:10.1016/s1567-5394(01)00109-8 · Cited by 59
Microdialysis of Glucose in Subcutaneous Adipose Tissue up to 3 Weeks in Healthy Volunteers
Wientjes, K. J. C., Vonk, P., Klei, Y. V.-V., Schoonen, A. J. M., & Kossen, N. W. F. (1998) · Diabetes Care · doi:10.2337/diacare.21.9.1481 · Cited by 58
Enzymatic glucose sensors. Improved long-term performance in vitro and in vivo.
Updike, S. J., Shults, M. C., Rhodes, R. K., GILLIGAN, B. J., LUEBOW, J. O., & von Heimburg, D. (1994) · PubMed · Cited by 57
Preparation and characterization of implantable sensors with nitric oxide release coatings
Frost, M. C., Batchelor, M. M., Lee, Y.-M., Zhang, H., Kang, Y., Oh, B. K., Wilson, G. S., Gifford, R., Rudich, S. M., & Meyerhoff, M. E. (2003) · Microchemical Journal · doi:10.1016/s0026-265x(03)00033-x · Cited by 55
Self-citationResolving dynamic changes in brain metabolism using biosensors and on-line microdialysis
Parkin, M. C., Hopwood, S. E., Boutelle, M. G., & Strong, A. J. (2003) · TrAC Trends in Analytical Chemistry · doi:10.1016/s0165-9936(03)00912-9 · Cited by 53
Updike, S. J., Shults, M. C., Rhodes, R. K., GILLIGAN, B. J., LUEBOW, J. O., & von Heimburg, D. (1994) · ASAIO Journal · doi:10.1097/00002480-199404000-00007 · Cited by 52
A Theoretical Description of Microdialysis with Mass Transport Coupled to Chemical Events
Yang, H., Peters, J. L., Allen, C., Chern, S.-S., Coalson, R. D., & Michael, A. C. (2000) · Analytical Chemistry · doi:10.1021/ac991186r · Cited by 51
Glucose sensor with improved haemocompatibilty
Yang, Y. (2000) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(00)00082-8 · Cited by 49
Kahlert, S., & Reiser, G. (2004) · Cell Calcium · doi:10.1016/j.ceca.2004.02.009 · Cited by 49
McAteer, K., & O'Neill, R. D. (1996) · The Analyst · doi:10.1039/an9962100773 · Cited by 47
Brown, F. O., & Lowry, J. (2003) · The Analyst · doi:10.1039/b300266g · Cited by 45
Sources of instability of ‘wired’ enzyme anodes in serum: urate and transition metal ions
Binyamin, G. N., Chen, T., & Heller, A. (2001) · Journal of Electroanalytical Chemistry · doi:10.1016/s0022-0728(00)00267-9 · Cited by 44
Varalli, M., Marelli, G., Maran, A., Bistoni, S., Luzzana, M., Cremonesi, P., Caramenti, G., Valgimigli, F., & Poscia, A. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00215-4 · Cited by 44
Modification of the sensitivity of glucose sensor implanted into subcutaneous tissue.
Thomé-Duret, V., Gangnerau, M.-N., Zhang, Y., Wilson, G. S., & Reach, G. (1996) · PubMed · Cited by 42
Self-citationWetterö, J., Bengtsson, T., & Tengvall, P. (2000) · Journal of Biomedical Materials Research · doi:10.1002/1097-4636(20000915)51:4 · Cited by 42
Healey, B. G., & Walt, D. R. (1997) · Analytical Chemistry · doi:10.1021/ac961058s · Cited by 42
Lutgers, H. L. (2000) · The Netherlands Journal of Medicine · doi:10.1016/s0300-2977(00)00022-x · Cited by 42
More Biocompatible Electrochemical Sensors Using Nitric Oxide Release Polymers
Mowery, K. A., Schoenfisch, M. H., Baliga, N., Wahr, J. A., & Meyerhoff, M. E. (1999) · Electroanalysis · doi:10.1002/(sici)1521-4109(199907)11:10/11 · Cited by 42
Prevention of Hypoglycemia Using Risk Assessment With a Continuous Glucose Monitoring System
Choleau, C., Dokládal, P., Klein, J. C., Ward, W. K., Wilson, G. S., & Reach, G. (2002) · Diabetes · doi:10.2337/diabetes.51.11.3263 · Cited by 42
Self-citationLowry, J. P., Ryan, M. R., & O'Neill, R. D. (1998) · Analytical Communications · doi:10.1039/a708747k · Cited by 41
Liu, Z., Niwa, O., Horiuchi, T., Kurita, R., & Torimitsu, K. (1999) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(99)00041-x · Cited by 41
Cui, J., Kulagina, N. V., & Michael, A. C. (2001) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(00)00343-5 · Cited by 40
Real-time multisite observation of glutamate release in rat hippocampal slices
Kasai, N., Jimbo, Y., Niwa, O., Matsue, T., & Torimitsu, K. (2001) · Neuroscience Letters · doi:10.1016/s0304-3940(01)01763-3 · Cited by 38
Development and comparison of biosensors for in-vivo applications
Georganopoulou, D. G., Carley, R., JONES, D. A., & Boutelle, M. G. (2000) · Faraday Discussions · doi:10.1039/b003256p · Cited by 38
Gramsbergen, J. B., Leegsma‐Vogt, G., Venema, K., Noraberg, J., & Korf, J. (2003) · Journal of Neurochemistry · doi:10.1046/j.1471-4159.2003.01673.x · Cited by 37
Yao, T., Yano, T., & Nishino, H. (2004) · Analytica Chimica Acta · doi:10.1016/j.aca.2003.12.062 · Cited by 36
Lowry, J., Demestre, M., & Fillenz, M. (1998) · Developmental Neuroscience · doi:10.1159/000017298 · Cited by 34
Nitric oxide and oxygen metabolism
Brown, G. C., McBride, A. G., Fox, E. J., McNaught, K. S. P., & Borutaitė, V. (1997) · Biochemical Society Transactions · doi:10.1042/bst0250901 · Cited by 34
Liposomes coated with chemically modified dextran interact with human endothelial cells
Cansell, M., Parisel, C., Jozefonvicz, J., & Letourneur, D. (1999) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199902)44:2 · Cited by 33
Analysis of antibody selection by phage display utilizing anti-phenobarbital antibodies
Malone, J. E., & Sullivan, M. A. (1996) · Journal of Molecular Recognition · doi:10.1002/(sici)1099-1352(199634/12)9:5/6 · Cited by 31
The Use of Optical Fiber Bundles Combined with Electrochemistry for Chemical Imaging
Szunerits, S., & Walt, D. R. (2003) · ChemPhysChem · doi:10.1002/cphc.200390029 · Cited by 30
Stability of immobilized enzymes as biosensors for continuous application in vitro and in vivo
Abel, P., von Woedtke, T., Schulz, B., Bergann, T., & Schwock, A. (1999) · Journal of Molecular Catalysis B Enzymatic · doi:10.1016/s1381-1177(99)00024-7 · Cited by 29
An independently addressable microbiosensor array: What are the limits of sensing element density?
Yu, P., & Wilson, G. S. (2000) · Faraday Discussions · doi:10.1039/b001611j · Cited by 29
Self-citationIn vivo calibration of the subcutaneous amperometric glucose sensors using a non-enzyme electrode
Jeong, R.-A., Hwang, J. Y., Joo, S., Chung, T. D., Park, S., Kang, S. K., Lee, W., & Kim, H. C. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00219-7 · Cited by 28
Statistics for Critical Clinical Decision Making Based on Readings of Pairs of Implanted Sensors
Schmidtke, D. W., Pishko, M. V., Quinn, C. P., & Heller, A. (1996) · Analytical Chemistry · doi:10.1021/ac9602027 · Cited by 27
Novel Hyaluronic Acid Coating for Potential Use in Glucose Sensor Design
Praveen, S., Hanumantha, R., Belovich, J., & Davis, B. L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503765691893 · Cited by 26
Microfabrication technology of flexible membrane based sensors for in vivo applications
Buck, R. P., Coșofreţ, V. V., Lindner, E., Ufer, S., Mudaras, M. B., Johnson, T. A., Ash, R. B., & Neumun, M. R. (1995) · Electroanalysis · doi:10.1002/elan.1140070910 · Cited by 26
Localized Exocytosis Detected by Spatially Resolved Amperometry in Single Pancreatic β-Cells
Paras, C. D., Qian, W., Lakey, J. R. T., Tan, W., & Kennedy, R. T. (2000) · Cell Biochemistry and Biophysics · doi:10.1385/cbb:33:3:227 · Cited by 25
A Wire-Based Dual-Analyte Sensor for Glucose and Lactate: In Vitro and In Vivo Evaluation
Ward, W. K., House, J. L., Birck, J., Anderson, E. M., & Jansen, L. B. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504774198106 · Cited by 25
A Model for Transport of Glucose in Adipose Tissue to a Microdialysis Probe
Schoonen, A. J. M., & Wientjes, K. J. C. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250604 · Cited by 25
Rigby, G., Ahmed, S., Horseman, G., & Vadgama, P. M. (1999) · Analytica Chimica Acta · doi:10.1016/s0003-2670(98)00665-5 · Cited by 25
Simulations of the Frequency Response of Implantable Glucose Sensors
Jablecki, M., & Gough, D. A. (2000) · Analytical Chemistry · doi:10.1021/ac991018z · Cited by 24
Principles of Sensorial Radical Detection-A Minireview
Lisdat, F., & Scheller, F. W. (2000) · Analytical Letters · doi:10.1080/00032710008543032 · Cited by 24
Glucose uptake and flux through phosphofructokinase in wounded rat skeletal muscle
Förster, J., Morris, A. S., Shearer, J. D., Mastrofrancesco, B., Inman, K. C., Lawler, R. G., Bowen, W. D., & Caldwell, M. D. (1989) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1989.256.6.e788 · Cited by 24
Ultrafiltration of human serum to determine the size of species that poison voltammetric electrodes
Elbicki, J. M., & Weber, S. G. (1989) · Biosensors · doi:10.1016/0265-928x(89)80039-2 · Cited by 24
Meyerhoff, C., Mennel, F. J., Bischof, F., Sternberg, F., & Pfeiffer, E. F. (1994) · Hormone and Metabolic Research · doi:10.1055/s-2007-1001752 · Cited by 23
Yang, D., Tsai, T., Cheng, C.-H., Lee, C.-W., Chen, S.-H., & Cheng, F.-C. (2001) · Journal of Chromatography A · doi:10.1016/s0021-9673(00)01033-5 · Cited by 22
Oxygen tension at the subcutaneous implantation site of glucose sensors.
Fischer, U., Hidde, A., Herrmann, S., von Woedtke, T., Rebrin, K., & Abel, P. (1989) · PubMed · Cited by 22
Direct measurement of glucose gradients and mass transport within islets of Langerhans
Kauri, L. M., Jung, S., & Kennedy, R. T. (2003) · Biochemical and Biophysical Research Communications · doi:10.1016/s0006-291x(03)00595-3 · Cited by 21
The Role of the Independent Variable to Glucose Sensor Calibration
Panteleon, A. E., Rebrin, K., & Steil, G. M. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503765691901 · Cited by 20
Implantable glucose sensor: progress and problems
Moussy, F. G. (2003) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2002.1037097 · Cited by 19
Biomaterials Community Examines Biosensor Biocompatibility
Moussy, F. G., & Reichert, W. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194341 · Cited by 18
Ultrasmall Enzyme Electrodes with Response Time Less than 100 Milliseconds
Meyerhoff, J. B., Ewing, M. A., & Ewing, A. G. (1999) · Electroanalysis · doi:10.1002/(sici)1521-4109(199905)11:5 · Cited by 17
Analysis of chemical dynamics in microenvironments
Wightman, R. M., Runnels, P. L., & Troyer, K. P. (1999) · Analytica Chimica Acta · doi:10.1016/s0003-2670(99)00605-4 · Cited by 16
Mediated Electron‐transfer between Redox‐enzymes and Electrode Supports
Katz, E. Y., Shipway, A. N., & Willner, I. (2002) · Encyclopedia of Electrochemistry · doi:10.1002/9783527610426.bard090017 · Cited by 12
Direct biologically based biosensing of dynamic physiological function
Christini, D. J., Walden, J., & Edelberg, J. M. (2001) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.2001.280.5.h2006 · Cited by 10
Chen, T., Schmidtke, D. W., & Heller, A. (2002) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2002.135 · Cited by 8
Genetic Engineering of Signaling Molecules
Feltus, A., & Daunert, S. (2002) · Elsevier eBooks · doi:10.1016/b978-044450974-1/50010-0 · Cited by 7
Germany - Microdialysis plus glucose sensor, theory and practical application
Owen, V. M. (1995) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(95)96875-y · Cited by 5
Expression of vascular endothelial growth factor in primary non-small cell lung carcinoma
Zhang, L., Guoli, Y., Yuquan, X., & Xu, W. (1999) · Chinese Journal of Cancer Research · doi:10.1007/s11670-999-0037-4 · Cited by 3
Mathur (1997) · doi:10.1002/(sici)1097-4636(199708)36:2
Kokot (1998) · doi:10.1002/(sici)1521-4109(199811)10:16
Flagged first, then by citations
4 of the 174 references couldn’t be matched to a known paper.
Research works printed but not linked: 4
- Clark, 1998, Electrochemistry in neuronal microenvironments, Electroanal. Chem., 20, 227
- Ishihara, 1997, Preventing the protein adsorption by the phospholipid polymers, Polym. Mater. Sci. Eng., 77, 574
- Reichert, 1999, vol. 28
- Sharkawy, 1998, Engineering the tissue which encapsulates subcutaneous implants. Part III. Effective tissue response times, J. Biomed. Mater. Res., 40, 598, 10.1002/(SICI)1097-4636(19980615)40:4<598::AID-JBM11>3.0.CO;2-C · DOI
Cited by
Sorted by citations
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Immobilization strategies to develop enzymatic biosensors
Sassolas, A., Blum, L. J., & Leca‐Bouvier, B. D. (2011) · Biotechnology Advances · doi:10.1016/j.biotechadv.2011.09.003 · Cited by 1,194
Electrochemical Non-enzymatic Glucose Sensors: A Perspective and an Evaluation
Toghill, K. E., & Compton, R. G. (2010) · International Journal of Electrochemical Science · doi:10.1016/s1452-3981(23)15359-4 · Cited by 930
Metadata
- Authors
- Wilson, G. S., & Gifford, R.
- Affiliations
- University of Kansas (US)
- Corresponding author
- Wilson, George S.
- ORCID
- Wilson, George S.
- Venue
- Biosensors and Bioelectronics
- Publisher
- Elsevier BV
- Published
- 2005-06
- Volume
- 20
- Issue
- 12
- Pages
- 2388-2403
- ISSN
- 0956-5663 (ISSN-L)
- OpenAlex
- W2090868198
- Cited by
- 660
- References
- 174
- Citations by year
- 2026: 17 · 2025: 21 · 2024: 14 · 2023: 16 · 2022: 17 · 2021: 13 · 2020: 23 · 2019: 27 · 2018: 36 · 2017: 25 · 2016: 37 · 2015: 39 · 2014: 40 · 2013: 50 · 2012: 52
- FWCI
- 2.85
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Biosensors and Bioelectronics (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- biosensors, in vivo monitoring, glucose sensor, lactate sensor, glutamate sensor, nitric oxide sensor, ion sensors, dopamine sensor, sensor calibration, biocompatibility, continuous glucose monitoring
- MeSH terms
- Blood Glucose, Equipment Design, Transducers, Blood Glucose Self-Monitoring, Biosensing Techniques, Monitoring, Ambulatory, Cell Culture Techniques, Prostheses and Implants
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review