Analysis of Al2O3—parylene C bilayer coatings and impact of microelectrode topography on long term stability of implantable neural arrays
The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Abstract Objective . Performance of many dielectric coatings for neural electrodes degrades over time, contributing to loss of neural signals and evoked percepts. Studies using planar test substrates have found that a novel bilayer coating of atomic-layer deposited (ALD) Al 2 O 3 and parylene C is a promising candidate for neural electrode applications, exhibiting superior stability to parylene C alone. However, initial results from bilayer encapsulation testing on non-planar devices have been less positive. Our aim was to evaluate ALD Al 2 O 3 -parylene C coatings using novel test paradigms, to rigorously evaluate dielectric coatings for neural electrode applications by incorporating neural electrode topography into test structure design. Approach . Five test devices incorporated three distinct topographical features common to neural electrodes, derived from the utah electrode array (UEA). Devices with bilayer (52 nm Al 2 O 3 + 6 µ m parylene C) were evaluated against parylene C controls ( N ⩾ 6 per device type). Devices were aged in phosphate buffered saline at 67 °C for up to 311 d, and monitored through: (1) leakage current to evaluate encapsulation lifetimes (>1 nA during 5VDC bias indicated failure), and (2) wideband (1–10 5 Hz) impedance. Main results . Mean-times-to-failure (MTTFs) ranged from 12 to 506 d for bilayer-coated devices, versus 10 to >2310 d for controls. Statistical testing (log-rank test, α = 0.05) of failure rates gave mixed results but favored the control condition. After failure, impedance loss for bilayer devices continued for months and manifested across the entire spectrum, whereas the effect was self-limiting after several days, and restricted to frequencies <100 Hz for controls. These results correlated well with observations of UEAs encapsulated with bilayer and control films. Significance . We observed encapsulation failure modes and behaviors comparable to neural electrode performance which were undetected in studies with planar test devices. We found the impact of parylene C defects to be exacerbated by ALD Al 2 O 3 , and conclude that inferior bilayer performance arises from degradation of ALD Al 2 O 3 when directly exposed to saline. This is an important consideration, given that neural electrodes with bilayer coatings are expected to have ALD Al 2 O 3 exposed at dielectric boundaries that delineate electrode sites. Process improvements and use of different inorganic coatings to decrease dissolution in physiological fluids may improve performance. Testing frameworks which take neural electrode complexities into account will be well suited to reliably evaluate such encapsulation schemes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896
Chronic recording capability of the Utah Intracortical Electrode Array in cat sensory cortex
Rousche, P. J., & Normann, R. A. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00031-4 · Cited by 646
Ludwig, K. A., Uram, J. D., Yang, J., Martin, D. C., & Kipke, D. R. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/1/007 · Cited by 638
Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066014 · Cited by 600
Wodlinger, B., Downey, J. E., Tyler-Kabara, E. C., Schwartz, A. B., Boninger, M. L., & Collinger, J. L. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016011 · Cited by 536
Simeral, J. D., Kim, S.-P., Black, M. J., Donoghue, J. P., & Hochberg, L. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/2/025027 · Cited by 496
Suner, S., Fellows, M. R., Vargas-Irwin, C., Nakata, G. K., & Donoghue, J. P. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.857687 · Cited by 415
Progress towards biocompatible intracortical microelectrodes for neural interfacing applications
Jorfi, M., Skousen, J. L., Weder, C., & Capadona, J. R. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/011001 · Cited by 384 · cited 1×
“However, an encapsulation scheme has yet to be recognized as capable of providing microelectrode performance stability for 10+ years [21].”
BackgroundSelective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 359
Davis, T. S., Wark, H. A. C., Hutchinson, D. T., Warren, D. J., O’Neill, K., Scheinblum, T., Clark, G. A., Normann, R. A., & Greger, B. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/3/036001 · Cited by 314
Prasad, A., Xue, Q.-S., Sankar, V., Nishida, T., Shaw, G., Streit, W. J., & Sanchez, J. C. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/5/056015 · Cited by 298 · cited 1×
“Materials used for this application include SiO2 or SiNx [10, 11], polyimide [12, 13], silicone [14], and parylene C [15–17].”
MethodLudwig, K. A., Langhals, N. B., Joseph, M. D., Richardson-Burns, S. M., Hendricks, J. L., & Kipke, D. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/1/014001 · Cited by 283
Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants
Prasad, A., Xue, Q.-S., Dieme, R., Sankar, V., Mayrand, R. C., Nishida, T., Streit, W. J., & Sanchez, J. C. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00002 · Cited by 178 · cited 1×
“Post-surgical reductions of electrode impedance have been repeatedly observed and attributed to dielectric degradation [2, 17, 22, 23], and encapsulation degradation is frequently cited as the mechanism for loss of recorded neural signals and reduced signal-to-noise ratio [11, 24–27].”
BackgroundWinslow, B. D., Christensen, M. B., Yang, W.-K., Solzbacher, F., & Tresco, P. A. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.05.050 · Cited by 171
Self-citationBarrese, J. C., Aceros, J., & Donoghue, J. P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/026003 · Cited by 161
Wark, H. A. C., Sharma, R., Mathews, K. S., Fernandez, E., Yoo, J., Christensen, B., Tresco, P., Rieth, L., Solzbacher, F., Normann, R. A., et al. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/4/045003 · Cited by 148
Self-citationThe insulation performance of reactive parylene films in implantable electronic devices
Seymour, J. P., Elkasabi, Y. M., Chen, H.-Y., Lahann, J., & Kipke, D. R. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2009.07.061 · Cited by 147
Biostability of micro‐photodiode arrays for subretinal implantation
Hämmerle, H., Kobuch, K., Kohler, K., Nisch, W., Sachs, H., & Stelzle, M. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00185-5 · Cited by 146 · cited 1×
“Materials used for this application include SiO2 or SiNx [10, 11], polyimide [12, 13], silicone [14], and parylene C [15–17].”
MethodCorrosion of silicon integrated circuits and lifetime predictions in implantable electronic devices
Vanhoestenberghe, A., & Donaldson, N. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/031002 · Cited by 140 · cited 1×
“Materials with low WVTR are desirable for their stable dielectric properties, as they do not hydrate as readily as those with higher WVTR, and more effectively prevent water nucleation around device surface contaminants [35].”
BackgroundXie, X., Rieth, L., Williams, L., Negi, S., Bhandari, R., Caldwell, R., Sharma, R., Tathireddy, P., & Solzbacher, F. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/2/026016 · Cited by 137
Self-citationDowney, J. E., Weiss, J. M., Muelling, K., Venkatraman, A., Valois, J.-S., Hebert, M., Bagnell, J. A., Schwartz, A. B., & Collinger, J. L. (2016) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-016-0134-9 · Cited by 111
A wafer-scale etching technique for high aspect ratio implantable MEMS structures
Bhandari, R., Negi, S., Rieth, L., & Solzbacher, F. (2010) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2010.06.011 · Cited by 71
Self-citationExcimer-laser deinsulation of Parylene-C coated Utah electrode array tips
Yoo, J.-M., Sharma, A., Tathireddy, P., Rieth, L. W., Solzbacher, F., & Song, J.-I. (2012) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2012.03.073 · Cited by 15
Self-citation
Cited by
Self‐Powered Implantable Medical Devices: Photovoltaic Energy Harvesting Review
Zhao, J., Ghannam, R., Htet, K. O., Liu, Y., Law, M., Roy, V. A. L., Michel, B., Imran, M. A., & Heidari, H. (2020) · Advanced Healthcare Materials · doi:10.1002/adhm.202000779 · Cited by 218
Mariello, M., Kim, K., Wu, K., Lacour, S. P., & Leterrier, Y. (2022) · Advanced Materials · doi:10.1002/adma.202201129 · Cited by 195
Recent advances in neural electrode–tissue interfaces
Woeppel, K., Yang, Q., & Cui, X. T. (2017) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2017.09.003 · Cited by 99
“…the insulation by first depositing a 52 nm adhesive layer of Al 2 O 3 via atomic layer deposition before the Parylene C coating [102], however, results seem to vary between positive results for planar electrodes [103] and potentially detrimental results for bed of needle style electrodes [104].”
Recent advances in materials and flexible electronics for peripheral nerve interfaces
Bettinger, C. J. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0007-6 · Cited by 77
An ASIC for Recording and Stimulation in Stacked Microchannel Neural Interfaces
Lancashire, H. T., Jiang, D., Demosthenous, A., & Donaldson, N. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2891284 · Cited by 15
Metadata
- Authors
- Caldwell, R., Mandal, H., Sharma, R., Solzbacher, F., Tathireddy, P., & Rieth, L.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2017-05-31
- Volume
- 14
- Issue
- 4
- Pages
- 046011
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2599624695
- Cited by
- 36
- References
- 23
- FWCI
- 1.88
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.