A novel flexible cuff-like microelectrode for dual purpose, acute and chronic electrical interfacing with the mouse cervical vagus nerve

Open access
Caravaca, A S, Tsaava, T, Goldman, L, Silverman, H, Riggott, G, Chavan, S S, Bouton, C, Tracey, K J +4 more·2017·Journal of Neural Engineering

Abstract

Objective . Neural reflexes regulate immune responses and homeostasis. Advances in bioelectronic medicine indicate that electrical stimulation of the vagus nerve can be used to treat inflammatory disease, yet the understanding of neural signals that regulate inflammation is incomplete. Current interfaces with the vagus nerve do not permit effective chronic stimulation or recording in mouse models, which is vital to studying the molecular and neurophysiological mechanisms that control inflammation homeostasis in health and disease. We developed an implantable, dual purpose, multi-channel, flexible ‘microelectrode’ array, for recording and stimulation of the mouse vagus nerve. Approach . The array was microfabricated on an 8 µ m layer of highly biocompatible parylene configured with 16 sites. The microelectrode was evaluated by studying the recording and stimulation performance. Mice were chronically implanted with devices for up to 12 weeks. Main results . Using the microelectrode in vivo , high fidelity signals were recorded during physiological challenges (e.g potassium chloride and interleukin-1 β ), and electrical stimulation of the vagus nerve produced the expected significant reduction of blood levels of tumor necrosis factor (TNF) in endotoxemia. Inflammatory cell infiltration at the microelectrode 12 weeks of implantation was limited according to radial distribution analysis of inflammatory cells. Significance . This novel device provides an important step towards a viable chronic interface for cervical vagus nerve stimulation and recording in mice.

AI snapshot

Written from the paper’s full text.

TL;DR

The paper reports a flexible 16-channel parylene-C cuff-like microelectrode for acute and chronic recording and stimulation of the mouse cervical vagus nerve. It recorded high-fidelity compound action potentials during KCl and IL-1β challenges, reduced TNFα in endotoxemia with electrical stimulation, and showed limited inflammatory cell infiltration after 12 weeks.

Key findings

  • The microelectrode had 16 sites on an 8 μm parylene sheet, with site areas of 6400, 2500, or 225 μm²; 1 kHz impedance in physiological saline was 100±12, 411±21, and 910±75 kΩ, respectively (n=320 sites).
  • During KCl challenge, mean CAPs were 2.36±0.50 per channel; 2500 μm² sites gave 477±38 μV signal and SNR 8.63±1.36, while 6400 μm² sites gave 179.5±24.5 μV, SNR 5.98±0.81, and 3.25±0.45 features per channel. The 225 μm² sites had impedance too high for meaningful CAP isolation.
  • After intraperitoneal IL-1β, distinctive CAP shapes and interspike intervals appeared in bursts 5 minutes after injection, found only in a subset of 14 adjacently arranged recording channels spaced 250 μm apart. Delays of 6 ms and 12 ms across 250 μm corresponded to conduction velocities of 42 m/s and 21 m/s.
  • Vagus nerve stimulation (1 mA, 250 μs biphasic pulse, 50 μs interphase delay, 10 Hz, 60 s) significantly reduced serum TNFα versus sham-operated controls in endotoxemia (n=30; p<0.05). At 12 weeks, no significant differences in CD3 or I-Ab staining were found between non-implanted and implanted nerve areas by radial distribution analysis extending 150 μm from the nerve center.

Methods

  • The device was microfabricated as a parylene-C (4 μm)–WTi (300 nm)–parylene-C (4 μm) sandwich, with 16 electrode sites in tetrode-like or diagonal arrangements, a ribbon cable, and an Omnetics connector; impedance was measured at 1 kHz in physiological saline.
  • Male BALB/c and C57Bl/6J mice (8–12 weeks) were implanted on the left cervical vagus nerve. In vivo recordings under isoflurane were sampled at 30 kHz, high-pass filtered at 200 Hz, and CAPs sorted with principal component analysis using a 4σ threshold; challenges included topical KCl, intraperitoneal saline, and intraperitoneal IL-1β.
  • For stimulation, mice received left cervical vagus nerve stimulation, then 5 mg/kg endotoxin intraperitoneally after 3 h, were euthanized 90 minutes post-injection, and serum TNFα was measured by ELISA. Immunohistochemistry used anti-CD3 and anti-I-Ab antibodies with radial distribution analysis; statistics used unpaired or paired two-tailed Student’s t-tests with p<0.05 considered significant.

Limitations

  • The paper does not include a formal limitations section. It states that 225 μm² sites had impedance too high for meaningful CAP isolation, that IL-1β-associated bursts were found only in a subset of 14 adjacently arranged channels, and that studies are ongoing to further characterize chronic recording and stimulation performance.
  • Evident from the text: chronic evaluation focuses on 12-week histological inflammatory response rather than long-term chronic recording or stimulation performance; the IL-1β recording findings come from a subset of channels and mice, and the paper does not report detailed chronic electrophysiology over the implantation period.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1088/1741-2552/aa7a42

Cites

38 of 38 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 16 cited sources checked · 0 supported · 0 flagged

  • Findingown data

    Using the microelectrode in vivo, high fidelity signals were recorded during physiological challenges such as potassium chloride and interleukin-1 beta.

    “Using the microelectrode in vivo, high fidelity signals were recorded during physiological challenges (e.g potassium chloride and interleukin-1 β)”
    Quote in the paperSubstantial claim
  • Findingown data

    Electrical stimulation of the vagus nerve through the microelectrode produced a significant reduction of blood TNF alpha levels in endotoxemia compared to sham-operated controls (n=30; p<0.05).

    “electrical stimulation of the vagus nerve produced the expected significant reduction of blood levels of tumor necrosis factor (TNF) in endotoxemia.”
    • Also stated in: § Electrical stimulation of the cervical vagus nerve using the microelectrode reduced TNFα in endotoxemia · plain
      “The TNFα levels were significantly lower in the electrically stimulated mice as compared to sham-operated controls (n=30; p<0.05)”
    Quote in the paperSubstantial claim
  • Findingown data

    Inflammatory cell infiltration at the microelectrode implantation at 12 weeks of implantation was limited in the radial distribution analysis of inflammatory cells.

    “Inflammatory cell infiltration at the microelectrode implantation at 12 weeks of implantation was limited in our radial distribution analysis of inflammatory cell.”
    Quote in the paperSubstantial claim
  • Findingown data

    No significant differences in either CD3 or I-Ab staining were found between non-implanted and implanted nerve areas as determined by normalised radial distribution analysis extending 150 um from the cross-sectional nerve center.

    “No significant differences in either CD3 or l-Ab staining were found between non-implanted and implanted nerve areas (Figure 6) as determined by normalised radial distribution analysis extending 150 um from the cross-sectional nerve center [18].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [18] Pavlov, 2003, The cholinergic anti-inflammatory pathway: a missing link in neuroimmunomodulation, Mol. Med. source

      Not checked in the free run.

  • Findingown data

    Electrode impedance at 1 kHz in physiological saline was 100±12, 411±21, and 910±75 kΩ for 6400, 2500 and 225 μm2 recording sites, respectively (n=320 sites).

    “Electrode impedance as measured at 1 kHz in physiological saline was 100±12, 411 ±21, and 910±75 kΩ for 6400, 2500 and 225 μm2 recording sites, respectively (n=320 sites).”
    Quote in the paperSubstantial claim
  • May combine two claims. 8
  • The quote only partly states this claim. 1
  • Classified as probably not a substantial claim. 2
  • May combine two claims.

    Findingown data

    No change in amplitude and frequency of the recorded signal was observed after saline injection compared to baseline, consistent across all animals.

    “No change in amplitude and frequency of the recorded signal was observed after saline injection as compared to baseline which was consistent across all animals.”
    Quote in the paperNot a substantial claim
  • May combine two claims. The quote only partly states this claim.

    Findingown data

    Good signal quality (179.5±24.5 μV) and SNR (5.98±0.81) with 3.25±0.45 features per channel were obtained from the 6400 μm2 sites.

    “Good signal quality (179.5±24.5 μV) and SNR (5.98±0.81) with 3.25±0.45 features per channel were obtained from the 64002 μm sites (Figure 2E).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Application of 50 μL of 4 mM KCl directly on the vagus nerve increased overall electrical activity and yielded a good number of CAPs across all recording channels (mean±SEM 2.36±0.50 per channel).

    “We observed an increase in overall electrical activity and obtained a good number of CAPs across all recording channels (Figure 2C, Mean±SEM CAPs = 2.36±0.50 per channel).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    In this study, recorded peak-to-peak signal amplitudes ranged between 100 and 2500 μV in response to pharmacological stimuli, and differed with the size of recording sites.

    “In this study, recorded peak-to-peak signal amplitudes ranged between 100 and 2500 μV in response to pharmacological stimuli, and differed with the size of recording sites.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Many current vagus nerve electrode implants cannot be successfully implanted chronically due to their size and can damage the nerve long-term.

    “Further, many of these implants can not be successfully implanted chronically due to their size and can damage the nerve long-term [9].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [9] Koopman, 2003, Vagus nerve stimulation inhibits cytokine production and attenuates disease severity in rheumatoid arthritis, Proc. Natl Acad. Sci. USA source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Previous studies of peripheral implants used for rodents have reported peak-to-peak signal amplitudes ranging from 100-900 μV in response to electrical stimuli.

    “Previous studies of peripheral implants used for rodents have reported peak-to-peak signal amplitudes ranging from 100–900 μV in response to electrical stimuli [22–27].”
    Quote in the paperNot a substantial claim0 of 6 sources checked
    • cites [22] Olofsson, 2015, Single-pulse and unidirectional electrical activation of the cervical vagus nerve reduces tumor necrosis factor in endotoxemia, Bioelectron. Med. source

      Not checked in the free run.

    • cites [23] Seymour, 2009, The insulation performance of reactive parylene films in implantable electronic devices, Biomaterials source

      Not checked in the free run.

    • cites [24] Yu, 2014, A cable-tie-type parylene cuff electrode for peripheral nerve interfaces source

      Not checked in the free run.

    • cites [25] Lee, 2016, Flexible epineural strip electrode for recording in fine nerves, IEEE Trans. Biomed. Eng. source

      Not checked in the free run.

    • cites [26] Badia, 2016, Spatial and functional selectivity of peripheral nerve signal recording with the transversal intrafascicular multichannel electrode (TIME), IEEE Trans. Neural Syst. Rehabil. Eng. source

      Not checked in the free run.

    • cites [27] Lee, 2016, Selective stimulation and neural recording on peripheral nerves using flexible split ring electrodes, Sensors Actuators B source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    The ultra-thin, inherently flexible, and customisable design of the microelectrode allows for improved interfacing with small diameter peripheral nerves in mice and avoids the increased tissue pressure associated with bulky silicone cuffs.

    “The ultra-thin, inherently flexible, and customisable design of the microelectrode developed in this study allows for improved interfacing with small diameter peripheral nerves in mice and avoids the increased tissue pressure associated with bulky silicone cuffs.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationcitation

    The current device design is fabricated with 16 channels, but it can easily be scaled up to ~1000 channels with the use of previously established process flows.

    “The current device design is fabricated with 16 channels, but it can easily be scaled up to ~1000 channels with the use of previously established process flows [29].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [29] Kagaya, 2002, Characterization of the anandamide induced depolarization of guinea-pig isolated vagus nerve, Br. J. Pharmacol. source

      Not checked in the free run.

  • Classified as probably not a substantial claim.

    Premisecitation

    Steinberg et al. recently demonstrated cytokine-specific neurograms in the sensory vagus nerve in mice.

    “Using bipolar electrodes, Steinberg et al. [13] recently demonstrated cytokine-specific neurograms in the sensory vagus nerve in mice.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [13] Steinberg, 2016, Cytokine-specific neurograms in the sensory vagus nerve, Bioelectron. Med. source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Information on signal propagation velocities may identify specific fibre types involved in a response, and indicate whether the activity is efferent or afferent.

    “Hence, information on signal propagation velocities may identify specific fibre types involved in a response, and indicate whether the activity is efferent or afferent.”
    Quote in the paperSubstantial claim
10 of 10
  • The paper relies on this without support.

    Optimal parameters for vagus nerve stimulation in inflammatory diseases have not been established.

    “However, the optimal parameters for vagus nerve stimulation in inflammatory diseases have not been established”
  • The paper relies on this without support.

    Technology for chronic implantation of electrodes in relevant mouse models has been lacking.

    “in part because technology for chronic implantation of electrodes in relevant mouse models has been lacking”
  • The paper relies on this without support.

    The neurophysiology of long-term electrical vagus nerve stimulation has not been mechanistically addressed.

    “For the same reason, the neurophysiology of long-term electrical vagus nerve stimulation has not been mechanistically addressed.”
  • The paper relies on this without support.

    Little is known about the electrical activity in the sensory, afferent arc of the inflammatory reflex.

    “In addition, little is known about the electrical activity in the sensory, afferent arc of the inflammatory reflex.”
  • The paper relies on this without support.

    The thin film parylene-C microfabrication technology has not yet been implemented for interfacing with peripheral nerves.

    “To our knowledge, this technology has not yet been implemented for interfacing with peripheral nerves.”
  • The paper relies on this without support.

    Many current electrode designs have low channel counts allowing the capture of only a certain number of compound action potentials (CAPs) to any given pharmacological challenge of the nerve.

    “Many current electrode designs have low channel counts allowing the capture of only a certain number of compound action potentials (CAPs) to any given pharmacological challenge of the nerve.”
6 of 6

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

1 of the 39 references couldn’t be matched to a known paper.

Research works printed but not linked: 1

  1. Sohal, 2016, Design and microfabrication considerations for reliable flexible intracortical implants, Front. Mech. Eng. China, 2

Cited by

Sorted by citations

Authors
Caravaca, A. S., Tsaava, T., Goldman, L., Silverman, H., Riggott, G., Chavan, S. S., Bouton, C., Tracey, K. J., Desimone, R., Boyden, E. S., Sohal, H. S., & Olofsson, P. S.
Affiliations
  • Karolinska University Hospital (SE)
  • Feinstein Institute for Medical Research (US)
  • Massachusetts Institute of Technology (US)
  • MIT Media Lab (US)
Corresponding author
  • Sohal, H S
  • Olofsson, P S
Venue
Journal of Neural Engineering
Published
2017-11-01
Volume
14
Issue
6
Pages
066005
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W2643626930
Cited by
57
References
39
Citations by year
2026: 2 · 2025: 2 · 2024: 3 · 2023: 2 · 2022: 6 · 2021: 13 · 2020: 8 · 2019: 14 · 2018: 7
FWCI
3.88
Type
journal article
Language
en
Open access
Locations
  • Journal of Neural Engineering (publishedVersion)
  • PubMed (publishedVersion)
  • Europe PMC (PubMed Central) (acceptedVersion)
  • PubMed Central (acceptedVersion)
Funders
  • New York Stem Cell Foundation Robertson Award
  • Knut and Alice Wallenberg’s Foundation (20140212)
  • Heart Lung Foundation (20150767)
  • NIH (1R01EY023173)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
  • Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
Keywords
cervical vagus nerve, flexible microelectrode arrays, neurostimulation, neural recording, chronic implantation, parylene, bioelectronic medicine, inflammation, endotoxemia, TNF-α, IL-1β, inflammatory cell infiltration, mouse model
MeSH terms
Action Potentials, Animals, Cervical Vertebrae, Electric Stimulation, Electrodes, Implanted, Male, Mice, Inbred BALB C, Mice, Inbred C57BL, Microelectrodes, Vagus Nerve, Mice, Vagus Nerve Stimulation

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review