Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring
The DOI matches its Crossref record; 38 of the 38 references it cites were found in a registry or an index.
Abstract
The present study investigated the relationship between blood and subcutaneous interstitial fluid (ISF) glucose by employing an amperometric glucose sensor specifically developed for 3-day continuous glucose monitoring. The apparent sensor sensitivity and ISF glucose equilibration delay were estimated on separate days during hyperglycemic clamps in four dogs in which insulin was either suppressed with somatostatin, allowed to change, or increased with an exogenous infusion. A 2-h sensor “settling-in” period was allowed before the clamps. During insulin deficiency, the sensor sensitivity and ISF glucose delay were 0.23 ± 0.03 nA per mg/dl and 4.4 ± 0.8 min. Sensitivity was not affected by increases in endogenous (0.30 ± 0.04 vs. 0.28 ± 0.04 nA per mg/dl) or exogenous insulin (0.18 ± 0.01 vs. 0.16 ± 0.01 nA per mg/dl) nor was the delay (3.3 ± 1.2 vs. 5.7 ± 1.1 and 9.2 ± 2.6 vs. 12.3 ± 1.7 min; P > 0.05 for all). Sensor glucose accurately predicted plasma glucose without correcting for delays <10 min ( r > 0.9 for all), whereas for longer delays a digital corrective filter was used ( r = 0.91 with filter). We conclude that the relationship between blood and ISF glucose is not affected by insulin and that delays in ISF glucose equilibration can be corrected with digital filters.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Digital Signal Processing: Principles, Algorithms, and Applications
Proakis, J. G., & Manolakis, D. G. (1992) · Cited by 4,601
ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY
Clark, L. C., & Lyons, C. (1962) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1962.tb13623.x · Cited by 3,945
A microdialysis method allowing characterization of intercellular water space in humans
Lonnroth, P., Jansson, P. A., & Smith, U. (1987) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1987.253.2.e228 · Cited by 633
Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring
Bindra, D. S., Zhang, Y., Wilson, G. S., Sternberg, R., Thevenot, D. R., Moatti, D., & Reach, G. (1991) · Analytical Chemistry · doi:10.1021/ac00017a008 · Cited by 309 · cited 1×
“The principle is based on the generation of hydrogen peroxide from glucose and oxygen via the enzyme glucose oxidase (2, 5, 9, 18, 32).”
BackgroundMaggs, D. G., Jacob, R., Rife, F., Lange, R., Leone, P., During, M. J., Tamborlane, W. V., & Sherwin, R. S. (1995) · Journal of Clinical Investigation · doi:10.1172/jci118043 · Cited by 183
Measurement of glucose in diabetic subjects using noninvasive transdermal extraction
Tamada, J. A., Bohannon, N. J. V., & Potts, R. O. (1995) · Nature Medicine · doi:10.1038/nm1195-1198 · Cited by 182
Bolincier, J., Ungerstedt, U., & Arner, P. (1992) · Diabetologia · doi:10.1007/bf00401374 · Cited by 142
Steil, G. M., Ader, M., Moore, D. M., Rebrin, K., & Bergman, R. N. (1996) · Journal of Clinical Investigation · doi:10.1172/jci118572 · Cited by 138
Self-citationGlucose measurement in patients with diabetes mellitus with dermal interstitial fluid
Bantle, J. P., & Thomas, W. (1997) · Journal of Laboratory and Clinical Medicine · doi:10.1016/s0022-2143(97)90044-5 · Cited by 128
Schmidtke, D. W., Freeland, A. C., Heller, A., & Bonnecaze, R. T. (1998) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.95.1.294 · Cited by 118
Cited by
Type 1 diabetes: new perspectives on disease pathogenesis and treatment
Atkinson, M. A., & Eisenbarth, G. S. (2001) · The Lancet · doi:10.1016/s0140-6736(01)05415-0 · Cited by 1,491
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660
“Otherwise the ratio is defined by the complicated interplay of glucose physiology and insulin kinetics (Aussedat et al., 2000; Rebrin et al., 1999).”
BackgroundArtificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Diabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
“fall in IG would precede fall in BG [196], [212].”
Continuous glucose monitoring and closed‐loop systems
Hovorka, R. (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01672.x · Cited by 443
Wearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
The MiniMed Continuous Glucose Monitoring System
Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050214078 · Cited by 393
Self-citationA Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
Cengiz, E., & Tamborlane, W. V. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0002 · Cited by 390
“22 A two-compartment model, described by Steil, Rebrin, and co-workers, 23,24 provides an insight to the glucose dynamics between the plasma and IF compartments.”
Assessing Glycemia in Diabetes Using Self-monitoring Blood Glucose and Hemoglobin A 1c
Saudek, C. D., Derr, R. L., & Kalyani, R. R. (2006) · JAMA · doi:10.1001/jama.295.14.1688 · Cited by 373
Closed-loop insulin delivery—the path to physiological glucose control
Steil, G. (2004) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2003.08.011 · Cited by 313
Self-citationNon-Invasive Glucose Measurement Technologies: An Update from 1999 to the Dawn of the New Millennium
Khalil, O. S. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.660 · Cited by 255
Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 251
“7 min (2), which is in agreement with previously published estimates of time lag between BG and IG (8).”
Chico, A., Vidal-Ríos, P., Subirà, M., & Novials, A. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.1153 · Cited by 235
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R. L., Bugler, J. R., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
“6 min, consistent with previously published studies using multicompartment diffusion models (13,20,21).”
BackgroundA Pilot Study of the Continuous Glucose Monitoring System
Kaufman, F. R., Gibson, L. C., Halvorson, M., Carpenter, S., Fisher, L. K., & Pitukcheewanont, P. (2001) · Diabetes Care · doi:10.2337/diacare.24.12.2030 · Cited by 224
“surements of plasma glucose concentrations (5) and home glucose values (6).”
BackgroundAbi-Saab, W. M., Maggs, D. G., Jones, T., Jacob, R., Srihari, V., Thompson, J., Kerr, D., Leone, P., Krystal, J. H., Spencer, D. D., et al. (2002) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1097/00004647-200203000-00004 · Cited by 219
“These sensors, which measure glucose levels in the interstitial fluid of subcutaneous tissues, appear to do so with a lag time of up to 20 minutes with respect to plasma levels in humans (reviewed in Rebrin et al., 1999).”
BackgroundCan Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 213
“Digital filters that can correct for this error have been proposed by us, 44 and others.”
InfluentialSelf-citationContinuous glucose monitoring: an Endocrine Society Clinical Practice Guideline
Klonoff, D. C., Buckingham, B., Christiansen, J. S., Montori, V. M., Tamborlane, W. V., Vigersky, R. A., & Wolpert, H. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-2756 · Cited by 211
Continuous monitoring of circadian glycemic patterns in patients receiving prednisolone for COPD
Burt, M. G., Roberts, G. W., Aguilar-Loza, N. R., Frith, P., & Stranks, S. N. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-2729 · Cited by 203
“CGMS measure interstitial glucose concentration, which closely approximates plasma glucose after a short time lag, every 5 min for up to 72 h (10).”
BackgroundContinuous Subcutaneous Glucose Monitoring in Children With Type 1 Diabetes
Chase, H. P., Kim, L. M., Owen, S. L., MacKenzie, T. A., Klingensmith, G. J., Murtfeldt, R., & Garg, S. K. (2001) · Pediatrics · doi:10.1542/peds.107.2.222 · Cited by 194
Continuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
Bequette, B. W. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400222 · Cited by 192
Bode, B., Gross, K., Rikalo, N., Schwartz, S., Wahl, T., Page, C., Gross, T., & Mastrototaro, J. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504773731285 · Cited by 190
“13 The effectiveness of the real-time hypo-and hyperglycemia alerts produced by the Guardian were evaluated by calculating sensitivity and specificity and producing ROC curves.”
Self-citationThennadil, S. N., Rennert, J. L., Wenzel, B. J., Hazen, K. H., Ruchti, T. L., & Block, M. B. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/15209150152607132 · Cited by 166
Time Lag of Glucose From Intravascular to Interstitial Compartment in Type 1 Diabetes
Basu, A., Dube, S., Veettil, S., Slama, M., Kudva, Y. C., Peyser, T., Carter, R. E., Cobelli, C., & Basu, R. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814554797 · Cited by 164
Experience with the Continuous Glucose Monitoring System® in a Medical Intensive Care Unit
Goldberg, P. A., Siegel, M. D., Russell, R. R., Sherwin, R. S., Halickman, J. I., Cooper, D. A., Dziura, J. D., & Inzucchi, S. E. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504774198034 · Cited by 158
“18 The accuracy of the CGMS may improve slightly when calibrations are performed by an experienced nurse.”
Rebrin, K., Sheppard, N. F., & Steil, G. M. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400507 · Cited by 152
Self-citationModeling the glucose sensor error
Facchinetti, A., Del Favero, S., Sparacino, G., Castle, J. R., Ward, W. K., & Cobelli, C. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2284023 · Cited by 130
“able BG references (one every 15 min for about 40 h) and IG profile is estimated by feeding a first-order linear time-invariant dynamic model [32] with interpolated BG data (time constant of”
MethodNorton, L. W., Koschwanez, H. E., Wisniewski, N. A., Klitzman, B., & Reichert, W. M. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31088 · Cited by 128
Chen, R., Yogev, Y., Ben-Haroush, A., Jovanovic, L., Hod, M., & Phillip, M. (2003) · The Journal of Maternal-Fetal & Neonatal Medicine · doi:10.1080/jmf.14.4.256.260 · Cited by 101
Care of diabetes in children and adolescents: controversies, changes, and consensus
Cameron, F. J., & Wherrett, D. K. (2015) · The Lancet · doi:10.1016/s0140-6736(15)60971-0 · Cited by 99
Realistic Expectations and Practical Use of Continuous Glucose Monitoring for the Endocrinologist
Hirsch, I. B. (2009) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2008-2625 · Cited by 93
Wiesli, P., Schmid, C., Kerwer, O., Nigg-Koch, C., Klaghofer, R., Seifert, B., Spinas, G. A., & Schwegler, K. (2005) · Diabetes Care · doi:10.2337/diacare.28.8.1910 · Cited by 88
Facchinetti, A., Del Favero, S., Sparacino, G., & Cobelli, C. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1226-y · Cited by 80
“1, the BG signal is transformed into the IG signal according to the BG-to-IG kinetics process, which can be modeled using the linear time-invariant twocompartment model of [29].”
MethodA Century of Diabetes Technology: Signals, Models, and Artificial Pancreas Control
Kovatchev, B. (2019) · Trends in Endocrinology & Metabolism · doi:10.1016/j.tem.2019.04.008 · Cited by 80
Evolution of Devices in Diabetes Management
Penfornis, A., Personeni, E., & Borot, S. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0058 · Cited by 49
“42 Oxidation of glucose by the enzyme-coated sensor removes an electron from each glucose molecule in the ISF, resulting in an electrical current, which the sensor converts into an interstitial glucose reading.”
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
“The most widely accepted model for BG-to-IG dynamics is the simple two-compartment diffusion model [45]:”
BackgroundAssessment of transcapillary glucose exchange in human skeletal muscle and adipose tissue
Regittnig, W., Ellmerer, M., Fauler, G., Sendlhofer, G., Trajanoski, Z., Leis, H.-J., Schaupp, L., Wach, P., & Pieber, T. R. (2003) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.00351.2002 · Cited by 41
Metadata
- Authors
- Rebrin, K., Steil, G. M., van Antwerp, W. P., & Mastrototaro, J. J.
- Venue
- American Journal of Physiology-Endocrinology and Metabolism
- Publisher
- American Physiological Society
- Published
- 1999-09-01
- Volume
- 277
- Issue
- 3
- Pages
- E561-E571
- ISSN
- 0193-1849 (ISSN-L)
- 1522-1555
- OpenAlex
- W2095833129
- Cited by
- 366
- References
- 38
- FWCI
- 6.81
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.