On the Analysis of Fingertip Photoplethysmogram Signals
Open accessAbstract
Photoplethysmography (PPG) is used to estimate the skin blood flow using infrared light. Researchers from different domains of science have become increasingly interested in PPG because of its advantages as non-invasive, inexpensive, and convenient diagnostic tool. Traditionally, it measures the oxygen saturation, blood pressure, cardiac output, and for assessing autonomic functions. Moreover, PPG is a promising technique for early screening of various atherosclerotic pathologies and could be helpful for regular GP-assessment but a full understanding of the diagnostic value of the different features is still lacking. Recent studies emphasise the potential information embedded in the PPG waveform signal and it deserves further attention for its possible applications beyond pulse oximetry and heart-rate calculation. Therefore, this overview discusses different types of artifact added to PPG signal, characteristic features of PPG waveform, and existing indexes to evaluate for diagnoses. Keywords: Photoplethysmography, acceleration plethysmogram, second derivative plethysmogram, digital volume pulse, ageing, artery, autonomic function, blood pressure, cardiovascular, heart rate, pulse wave analysis, vascular disease
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures












![Fig. (13) Typical PPG waveforms show the parameters changes with age [62].](/paper-figures/8ee5a57901c347579a7a5160e4be1526/on-the-analysis-of-fingertip-photoplethysmogram-signals-figure-13.jpg)


![Fig. (16) APG waveforms and types of photoplethysmogram [73]. There are different types of APG waveforms. The first APG waveform A (far left) refers to good circulation, whereas the amplitude of b wave is lower than c wave. The last APG waveform G (far right) refers to distinctively bad circulation, whereas the amplitude of c wave is lower than b wave.](/paper-figures/54ec3ef6e18e4116ad4a30b26f4dd445/on-the-analysis-of-fingertip-photoplethysmogram-signals-figure-16.jpg)
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“| c/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“The shape of a clean PPG signal is closely related to the blood pulse wave of the human body25”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [25]
In the citing paper
“In this case, the pulsatile component of the PPG can be of the order of 1 mV in amplitude [55] although lower amplitudes can be caused by technical factors, such as inappropriate gain control, or physiological factors, such as vasoconstriction [9].”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [9]
In the citing paper
“In particular, obesity can lead to changes in skin thickness, blood flow, and oxygen saturation. This affects the optical properties of skin in addition to the distance light has to travel to reach a target vessel or vessels [48,59,60].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“While, to the authors’ knowledge, publications that experimentally and explicitly demonstrate the effect of obesity or BMI on the PPG waveform are limited, we will explore works which suggest that there would be a substantial effect [56,59,60,61,62,63,64,65,66,67,68,69,70].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“The ratio b/a has been found to correlate positively to age, while c/a, d/a and e/a have all been correlated negatively to age [29,59].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“| e/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“| d/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“Furthermore, PPG can be used to index heart rate variability (HRV), blood pressure, oxygen saturation, and cardiac output24, but as of yet have not be widely rolled out in consumer-use devices and therefore data collection other than HR will not be discussed further.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [24]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
Assessment of fat-free mass using bioelectrical impedance measurements of the human body
Lukaski, H., Johnson, P., Bolonchuk, W., & Lykken, G. (1985) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/41.4.810 · Cited by 1,896
Investigation of Chronic Venous Insufficiency: A Consensus Statement
Nicolaides, A. N. (2000) · Circulation · doi:10.1161/01.cir.102.20.e126 · Cited by 945
Fifteen years experience with finger arterial pressure monitoring: assessment of the technology
Imholz, B. P. M. (1998) · Cardiovascular Research · doi:10.1016/s0008-6363(98)00067-4 · Cited by 841
Determination of age-related increases in large artery stiffness by digital pulse contour analysis
Millasseau, S., Kelly, R. P., Ritter, J., & Chowienczyk, P. (2002) · Clinical Science · doi:10.1042/cs1030371 · Cited by 624
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589
Photoelectric measurement of blood pressure, volume and flow in the finger
Peňáz, J. (1973) · Medical Entomology and Zoology · Cited by 528
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 460
Research works printed but not linked: 3
- [59]Takazawa KTN Fujita M Matsuoka O Saiki T Aikawa M Tamura S Ibukiyama C Assessment of vasocative agents and vascular aging by the second derivative of photoplethysmogram waveform Hypertension 1998 32 365 70 9719069
- [80]Elgendi M Jonkman M De Boer F Fred A Filipe J Gamboa h Heart Rate Variability and Acceleration Plethysmogram measured at rest Biomedical Engineering Systems and Technologies: 2011 Springer 266 77
- [81]Tokutaka H MY Gonda E Yamamoto M Construction of a General Physical Condition Judgment System Using Acceleration Plethysmogram Pulse-Wave Analysis 2009 Berlin/Heidelberg Springer 5629 307 15
Cited by
Sorted by citations
Items 201–300 of 1,001
Trovato, V., Teblum, E., Kostikov, Y., Pedrana, A., Re, V., Nessim, G. D., & Rosace, G. (2019) · Materials Chemistry and Physics · doi:10.1016/j.matchemphys.2019.122218 · Cited by 36
Tang, Q., Chen, Z., Allen, J., Alian, A. A., Menon, C., Ward, R. K., & Elgendi, M. (2020) · Frontiers in Medicine · doi:10.3389/fmed.2020.597774 · Cited by 36
Self-citationA Continuous Cuffless Blood Pressure Estimation Using Tree-Based Pipeline Optimization Tool
Fati, S. M., Muneer, A., Akbar, N. A., & Taib, S. M. (2021) · Symmetry · doi:10.3390/sym13040686 · Cited by 38
Liu, B., Zhang, Z., Di, X., Wang, X., Xie, L., Xie, W., & Zhang, J. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2021.733264 · Cited by 35
The mountaineer's method for peak detection in photoplethysmographic signals
Prada, E. J. A. (2019) · Revista Facultad de Ingeniería Universidad de Antioquia · doi:10.17533/udea.redin.n90a06 · Cited by 35
Mohebbian, M. R., Dinh, A., Wahid, K. A., & Alam, M. S. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101682 · Cited by 35
Huthart, S., Elgendi, M., Zheng, D., Stansby, G. P., & Allen, J. (2020) · Frontiers in Digital Health · doi:10.3389/fdgth.2020.619692 · Cited by 35
Self-citationEstimation of phase distortions of the photoplethysmographic signal in digital IIR filtering
Lapitan, D. G., Рогаткин, Д. А., Molchanova, E. A., & Tarasov, A. P. (2024) · Scientific Reports · doi:10.1038/s41598-024-57297-3 · Cited by 34
A review: Blood pressure monitoring based on PPG and circadian rhythm
Chen, G., Zou, L., & Ji, Z. (2024) · APL Bioengineering · doi:10.1063/5.0206980 · Cited by 34
Rajala, S., Ahmaniemi, T., Lindholm, H. J., & Taipalus, T. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8036809 · Cited by 34
Bilgaiyan, A., Elsamnah, F., Ishidai, H., Shim, C.-H., Misran, M. A. B., Adachi, C., & Hattori, R. (2020) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.0c00082 · Cited by 35
Mindfulness-Based Student Training Leads to a Reduction in Physiological Evaluated Stress
Voss, A., Bogdanski, M., Langohr, B., Albrecht, R., & Sandbothe, M. (2020) · Frontiers in Psychology · doi:10.3389/fpsyg.2020.00645 · Cited by 37
Age-related changes in pulse risetime measured by multi-site photoplethysmography
Allen, J., O’Sullivan, J. J., Stansby, G. P., & Murray, A. (2020) · Physiological Measurement · doi:10.1088/1361-6579/ab9b67 · Cited by 34
Reddy, G. N. K., Manikandan, M. S., & Murty, N. V. L. N. (2022) · IEEE Access · doi:10.1109/access.2022.3148256 · Cited by 34
Song, L., Xu, E., Yu, Y., Jie, J., Xia, Y., Chen, S., Jiang, Y., Xu, G., Li, D., & Jie, J. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200555 · Cited by 34
Continuous Blood Pressure Monitoring using Wrist-worn Bio-impedance Sensors with Wet Electrodes
Ibrahim, B., & Jafari, R. (2018) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2018.8584783 · Cited by 34
Béres, S., Holczer, L., & Hejjel, L. (2019) · Measurement Science Review · doi:10.2478/msr-2019-0030 · Cited by 33
Gruwez, H., Ezzat, D., Puyvelde, T. V., Dhont, S., Meekers, E. M., Bruckers, L. M., Wouters, F., Kellens, M., Herendael, H. V., Rivero‐Ayerza, M., et al. (2024) · EP Europace · doi:10.1093/europace/euae065 · Cited by 35
PPG embedded system for blood pressure monitoring
Agrò, D., Canicattì, R., Tomasino, A., Giordano, A., Adamo, G., Parisi, A., Pernice, R., Stivala, S. S., Giaconia, G. C., Busacca, A., et al. (2014) · AEIT International Annual Conference · doi:10.1109/aeit.2014.7002012 · Cited by 33
Ushakov, N. A., Александрович, М. А., Kulik, D., & Борисович, Л. Л. (2021) · Photonics · doi:10.3390/photonics8050142 · Cited by 33
McKnight, J. C., Mulder, E., Ruesch, A., Kainerstorfer, J. M., Wu, J., Hakimi, N., Balfour, S., Bronkhorst, M., Horschig, J. M., Pernett, F., et al. (2021) · Philosophical Transactions of the Royal Society B Biological Sciences · doi:10.1098/rstb.2020.0349 · Cited by 33
A Survey of Photoplethysmography and Imaging Photoplethysmography Quality Assessment Methods
Desquins, T., Bousefsaf, F., Pruski, A., & Maaoui, C. (2022) · Applied Sciences · doi:10.3390/app12199582 · Cited by 35
PPGMotion: Model-based detection of motion artifacts in photoplethysmography signals
Maity, A. K., Veeraraghavan, A., & Sabharwal, A. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.103632 · Cited by 32
Evaluating Photoplethysmogram as a Real-Time Cognitive Load Assessment during Game Playing
Zhang, X., Lyu, Y., Hu, X., Hu, Z., Shi, Y., & Yin, H. (2018) · International Journal of Human-Computer Interaction · doi:10.1080/10447318.2018.1461763 · Cited by 32
Addressing gain-bandwidth trade-off by a monolithically integrated photovoltaic transistor
Li, Y., Chen, G., Zhao, S., Liu, C., & Zhao, N. (2022) · Science Advances · doi:10.1126/sciadv.abq0187 · Cited by 32
Computerized approach for cardiovascular risk level detection using photoplethysmography signals
Ramachandran, D., Vanathi, P. T., & Rajaguru, H. (2019) · Measurement · doi:10.1016/j.measurement.2019.107048 · Cited by 31
Speckleplethysmographic (SPG) Estimation of Heart Rate Variability During an Orthostatic Challenge
Dunn, C. E., Monroe, D. C., Crouzet, C., Hicks, J. W., & Choi, B. (2019) · Scientific Reports · doi:10.1038/s41598-019-50526-0 · Cited by 31
Robust Reconstruction of Electrocardiogram Using Photoplethysmography: A Subject-Based Model
Tang, Q., Chen, Z., Guo, Y., Liang, Y., Ward, R. K., Menon, C., & Elgendi, M. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.859763 · Cited by 31
Self-citationAdvances in Cuffless Continuous Blood Pressure Monitoring Technology Based on PPG Signals
Qin, C., Wang, X., Xu, G., & Ma, X. (2022) · BioMed Research International · doi:10.1155/2022/8094351 · Cited by 31
Haugg, F., Elgendi, M., & Menon, C. (2023) · Bioengineering · doi:10.3390/bioengineering10020243 · Cited by 31
Self-citationAttivissimo, F., Palma, L. D., Nisio, A. D., Scarpetta, M., & Lanzolla, A. M. L. (2023) · Sensors · doi:10.3390/s23042321 · Cited by 31
A Pulse Rate Detection Method for Mouse Application Based on Multi-PPG Sensors
Lin, S.-T., Chen, W., & Lin, Y. (2017) · Sensors · doi:10.3390/s17071628 · Cited by 31
Feng, J., Huang, Z., Zhou, C., & Ye, X. (2018) · Australasian Physical & Engineering Sciences in Medicine · doi:10.1007/s13246-018-0637-8 · Cited by 31
Automated detection of the onset and systolic peak in the pulse wave using Hilbert transform
Ferro, B. T. R., Aguilera, A. R., & Prieto, R. R. F. D. L. V. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.04.009 · Cited by 31
Don’t add or miss a beat: a guide to cleaner heart rate variability recordings.
Shaffer, F., & Combatalade, D. C. (2013) · Biofeedback · doi:10.5298/1081-5937-41.3.04 · Cited by 30
Enhancing accuracy of arrhythmia classification by combining logical and machine learning techniques
Kalidas, V., & Tamil, L. S. (2015) · Computing in cardiology · doi:10.1109/cic.2015.7411015 · Cited by 30
Micro-patterned graphene-based sensing skins for human physiological monitoring
Wang, L., Loh, K. J., Chiang, W., & Manna, K. (2018) · Nanotechnology · doi:10.1088/1361-6528/aaa709 · Cited by 30
Using imaging Photoplethysmography (iPPG) Signal for Blood Pressure Estimation
Goudarzi, R. H., Mousavi, S. S., & Charmi, M. (2020) · Iranian Conference on Machine Vision and Image Processing · doi:10.1109/mvip49855.2020.9116902 · Cited by 30
Cuffless Blood Pressure Monitors: Principles, Standards and Approval for Medical Use
TAMURA, T. (2020) · IEICE Transactions on Communications · doi:10.1587/transcom.2020hmi0002 · Cited by 31
Camera-Based Blood Pressure Estimation via Windkessel Model and Waveform Features
Wu, B.-J., Wu, B., & Hsu, C.-P. (2022) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2022.3224534 · Cited by 30
Dutt, D. N., & Shruthi, S. (2015) · International Conference on Communications and Signal Processing (ICCSP) · doi:10.1109/iccsp.2015.7322766 · Cited by 30
Schäck, T., Harb, Y. S., Muma, M., & Zoubir, A. M. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8036773 · Cited by 30
Progresses towards a processing pipeline in photoplethysmogram (PPG) based on SiPMs
Vinciguerra, V., Ambra, E., Maddiona, L., Oliveri, S., Romeo, M. F., Mazzillo, M. C., Rundo, F., & Fallica, G. (2017) · European Conference on Circuit Theory and Design · doi:10.1109/ecctd.2017.8093327 · Cited by 29
Yang, C., Abodurexiti, A., & Maimaitiyiming, X. (2020) · Macromolecular Materials and Engineering · doi:10.1002/mame.202000287 · Cited by 29
Study on PPG Biometric Recognition Based on Multifeature Extraction and Naive Bayes Classifier
Yang, J., Huang, Y., Zhang, R., Huang, F., Meng, Q., & Feng, S. (2021) · Scientific Programming · doi:10.1155/2021/5597624 · Cited by 29
Beat-Based PPG-ABP Cleaning Technique for Blood Pressure Estimation
Salah, M., Omer, O. A., Hassan, L., Ragab, M., Hassan, A. M., & Abdelreheem, A. A. (2022) · IEEE Access · doi:10.1109/access.2022.3175436 · Cited by 29
Comparison and Noise Suppression of the Transmitted and Reflected Photoplethysmography Signals
Li, S., Liu, L., Wu, J., Tang, B., & Li, D. (2018) · BioMed Research International · doi:10.1155/2018/4523593 · Cited by 29
Rabow, S., Hjorth, U., Schönbeck, S., & Olofsson, P. (2018) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-018-2029-1 · Cited by 29
Yamashita, Y., Wang, L., Nakamura, A., Nanba, F., Saito, S., Toda, T., Nakagawa, J., & Ashida, H. (2020) · Archives of Biochemistry and Biophysics · doi:10.1016/j.abb.2020.108408 · Cited by 29
Vahlsing, T., Delbeck, S., Leonhardt, S., & Heise, H. M. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818798347 · Cited by 28
“Various physiological and pharmacological factors exist, which influence the systolic amplitude owing, for example, to poor finger perfusion, dysrhythmia, and others.(26) To round up the discussion of physiological parameters that are detectable in general by SW-NIR measurements, it is evident that hemoglobin chromoph…”
BackgroundCardiac arrhythmia classification using multi-modal signal analysis
Kalidas, V., & Tamil, L. S. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/8/1253 · Cited by 28
Photoplethysmogram-based Cognitive Load Assessment Using Multi-Feature Fusion Model
Zhang, X., Lyu, Y., Qu, T., Qiu, P., Luo, X., Zhang, J., Fan, S., & Shi, Y. (2019) · ACM Transactions on Applied Perception · doi:10.1145/3340962 · Cited by 28
A computer vision approach for classifying isometric grip force exertion levels
Asadi, H., Zhou, G., Lee, J. J., Aggarwal, V., & Yu, D. (2020) · Ergonomics · doi:10.1080/00140139.2020.1745898 · Cited by 28
Bradke, B. S., & Everman, B. (2020) · Biosensors · doi:10.3390/bios10040034 · Cited by 28
dSVRI: A PPG-Based Novel Feature for Early Diagnosis of Type-II Diabetes Mellitus
Gupta, S., Singh, A. K., Sharma, A., & Tripathy, R. K. (2022) · IEEE Sensors Letters · doi:10.1109/lsens.2022.3203609 · Cited by 28
Photoplethysmography temporal marker-based machine learning classifier for anesthesia drug detection
Khalid, S. G., Ali, S. M., Liu, H., Qurashi, A. G., & Ali, U. (2022) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02658-1 · Cited by 28
Stress Detection Using PPG Signal and Combined Deep CNN-MLP Network
Hasanpoor, Y., Motaman, K., Tarvirdizadeh, B., Alipour, K., & Ghamari, M. (2022) · National and International Iranian Conference on Biomedical Engineering (ICBME) · doi:10.1109/icbme57741.2022.10052957 · Cited by 28
Postoperative Pain Assessment Model Based on Pulse Contour Characteristics Analysis
Seok, H. S., Choi, B., Noh, G., & Shin, H. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2018.2890482 · Cited by 28
The Use of Photoplethysmography in the Assessment of Mental Health: Scoping Review
Lyzwinski, L. N., Elgendi, M., & Menon, C. (2023) · JMIR Mental Health · doi:10.2196/40163 · Cited by 29
Self-citationSmart phone based blood pressure indicator
Visvanathan, A., Banerjee, R., Choudhury, A. D., Sinha, A., & Kundu, S. (2014) · International Workshop on Pervasive Wireless Healthcare · doi:10.1145/2633651.2633657 · Cited by 27
ASCII-character-encoding based PPG compression for tele-monitoring system
Mukhopadhyay, S. K., Ahmad, M. O., & Swamy, M. N. S. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.09.021 · Cited by 27
Heart Rate Monitoring Using PPG With Smartphone Camera
Ayesha, A. H., Qiao, D., & Zulkernine, F. H. (2021) · IEEE International Conference on Bioinformatics and Biomedicine (BIBM) · doi:10.1109/bibm52615.2021.9669735 · Cited by 27
Bousefsaf, F., Desquins, T., Djeldjli, D., Ouzar, Y., Maaoui, C., & Pruski, A. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.103895 · Cited by 27
The Effect of Light Conditions on Photoplethysmographic Image Acquisition Using a Commercial Camera
Liu, H., Wang, Y., & Wang, L. (2014) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2014.2360200 · Cited by 27
Ghamari, M., Soltanpur, C., Cabrera, S. D., Romero, R., Martínek, R., & Nazeran, H. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591842 · Cited by 27
VITAL-ECG: A portable wearable hospital
Randazzo, V., Pasero, E. G. A., & Navaretti, S. (2018) · IEEE Sensors Applications Symposium (SAS) · doi:10.1109/sas.2018.8336776 · Cited by 27
Integrated Soft Optoelectronics for Wearable Health Monitoring
Biswas, S., Shao, Y., Hachisu, T., Nguyen-Dang, T., & Visell, Y. (2020) · Advanced Materials Technologies · doi:10.1002/admt.202000347 · Cited by 27
Madhura, M. (2014) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2014/7385.4778 · Cited by 26
A Survey on Denoising Techniques of PPG Signal
Mishra, B., & Nirala, N. (2020) · International Conference for Innovation in Technology (INOCON) · doi:10.1109/inocon50539.2020.9298358 · Cited by 26
Bousefsaf, F., Djeldjli, D., Ouzar, Y., Maaoui, C., & Pruski, A. (2021) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2021.104860 · Cited by 26
Silva, R., Salvador, G. F. D., Bota, P. J., Fred, A. L. N., & Silva, H. (2022) · Neural Computing and Applications · doi:10.1007/s00521-022-07212-6 · Cited by 26
Enhancing the Robustness of Smartphone Photoplethysmography: A Signal Quality Index Approach
Liu, I., Ni, S., & Peng, K. (2020) · Sensors · doi:10.3390/s20071923 · Cited by 26
Sex-Related Differences in Photoplethysmography Signals Measured From Finger and Toe
Dehghanojamahalleh, S., & Kaya, M. (2019) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2019.2938506 · Cited by 25
Khalili, M., Lingawi, S., Hutton, J., Fordyce, C. B., Christenson, J. M., Shadgan, B., Grunau, B. E., & Kuo, C. (2024) · Scientific Reports · doi:10.1038/s41598-024-74117-w · Cited by 26
Physiological parameters measurements in a cardiac cycle via a combo PPG-ECG system
Oreggia, D., Guarino, S., Parisi, A., Pernice, R., Adamo, G., Mistretta, L., Buono, P. D., Fallica, G., Ferla, G., Cino, A. C., et al. (2015) · AEIT International Annual Conference · doi:10.1109/aeit.2015.7415214 · Cited by 25
Web of Objects Based Ambient Assisted Living Framework for Emergency Psychiatric State Prediction
Alam, M. G. R., Abedin, S. F., Ameen, M. A., & Hong, C. S. (2016) · Sensors · doi:10.3390/s16091431 · Cited by 25
Blood pressure prediction from speech recordings
Ankışhan, H. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101842 · Cited by 25
Veluswamy, S., Shah, P., Khaleel, M., Thuptimdang, W., Chalacheva, P., Sunwoo, J., Denton, C. C., Kato, R. M., Detterich, J. A., Wood, J. C., et al. (2020) · Blood · doi:10.1182/blood.2020005045 · Cited by 25
Aziz, M. H., Hasan, M. K., Mahmood, A., Love, R. R., & Ahamed, S. I. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3068658 · Cited by 25
Fischer, C., Glos, M., Penzel, T., & Fietze, I. (2017) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-017-0115-7 · Cited by 25
Shin, H., Sun, S., Lee, J., & Kim, H. C. (2021) · IEEE Access · doi:10.1109/access.2021.3078534 · Cited by 26
PBGAN: Learning PPG Representations From GAN for Time-Stable and Unique Verification System
Hwang, D. Y., Taha, B., & Hatzinakos, D. (2021) · IEEE Transactions on Information Forensics and Security · doi:10.1109/tifs.2021.3122817 · Cited by 25
A wearable-based aging clock associates with disease and behavior
Miller, A. C., Futoma, J., Abbaspourazad, S., Heinze‐Deml, C., Emrani, S., Shapiro, I., & Sapiro, G. (2025) · Nature Communications · doi:10.1038/s41467-025-64275-4 · Cited by 25
Nomoni, M., May, J. M., & Kyriacou, P. A. (2020) · Sensors · doi:10.3390/s20154246 · Cited by 24
In-Ear SpO₂ for Classification of Cognitive Workload
Davies, H. J., Williams, I., Hammour, G., Yarici, M. C., Stacey, M. J., Seemungal, B. M., & Mandic, D. P. (2022) · IEEE Transactions on Cognitive and Developmental Systems · doi:10.1109/tcds.2022.3196841 · Cited by 27
Aziz, S., Awais, M., Khan, M. U., Iqtidar, K., & Qamar, U. (2021) · Expert Systems · doi:10.1111/exsy.12664 · Cited by 24
Measurement of Pulsation Strain at the Fingertip Using a Plastic FBG Sensor
Koyama, S., Haseda, Y., Ishizawa, H., Okazaki, F., Bonefacino, J., & Tam, H. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3079432 · Cited by 24
Morales-Fajardo, H., Rodríguez-Arce, J., Gutiérrez-Cedeño, A., Viñas, J. C., Lagos, J. J. R., Abarca-Castro, E. A., Ramírez, C. I. L., & Vilchis-González, A. H. (2022) · Sensors · doi:10.3390/s22103780 · Cited by 26
A 186μW Photoplethysmography-Based Noninvasive Glucose Sensing SoC
Hina, A., & Saadeh, W. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3180893 · Cited by 24
Moscato, S., Orlandi, S., Giannelli, A., Ostan, R., & Chiari, L. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871990 · Cited by 24
Azudin, K., Gan, K. B., Jaafar, R., & Jaáfar, M. H. (2023) · Sensors · doi:10.3390/s23146484 · Cited by 24
Multimodal Monitoring of Cardiovascular Responses to Postural Changes
Mol, A., Maier, A. B., Wezel, R. J. A. V., & Meskers, C. G. M. (2020) · Frontiers in Physiology · doi:10.3389/fphys.2020.00168 · Cited by 24
A touch-based multimodal and cryptographic bio-human–machine interface
Lin, S., Zhu, J., Yu, W., Wang, B., Sabet, K. A., Zhao, Y., Cheng, X., Hojaiji, H., Lin, H., Tan, J., et al. (2022) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2201937119 · Cited by 24
Acceleration plethysmogram based biometric identification
Jaafar, N. A. L., Sidek, K. A., & Azam, S. N. A. M. (2015) · International Conference on BioSignal Analysis, Processing and Systems (ICBAPS) · doi:10.1109/icbaps.2015.7292210 · Cited by 24
Kumar, S., Yadav, S., & Kumar, A. (2024) · Sensors & Diagnostics · doi:10.1039/d3sd00201b · Cited by 23
Palanisamy, S., & Rajaguru, H. (2023) · Bioengineering · doi:10.3390/bioengineering10060678 · Cited by 23
Liu, J., Hu, S., Xiao, Z., Hu, Q., Wang, D., & Yang, C. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105184 · Cited by 23
Ontiveros, R. C., Elgendi, M., & Menon, C. (2024) · Communications Medicine · doi:10.1038/s43856-024-00519-6 · Cited by 23
Self-citationPeláez‐Coca, M. D., Lozano, M. T., Hernando, A., Aiger, M., & Gil, E. (2018) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2018.2882142 · Cited by 23
Chalacheva, P., Kato, R. M., Shah, P., Veluswamy, S., Denton, C. C., Sunwoo, J., Thuptimdang, W., Wood, J. C., Detterich, J. A., Coates, T. D., et al. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00381 · Cited by 23
Metadata
- Authors
- Elgendi, M.
- Affiliations
- Charles Darwin University (AU)
- Corresponding author
- Elgendi, Mohamed
- ORCID
- Elgendi, Mohamed
- Venue
- Current Cardiology Reviews
- Publisher
- Bentham Science Publishers Ltd.
- Published
- 2012-06-01
- Volume
- 8
- Issue
- 1
- Pages
- 14-25
- ISSN
- 1573-403X (ISSN-L)
- PMC
- PMC3394104
- OpenAlex
- W1981918735
- Cited by
- 1,206
- References
- 88
- Citations by year
- 2026: 120 · 2025: 106 · 2024: 116 · 2023: 109 · 2022: 130 · 2021: 114 · 2020: 118 · 2019: 118 · 2018: 86 · 2017: 65 · 2016: 46 · 2015: 42 · 2014: 24 · 2013: 11 · 2012: 1
- FWCI
- 2.17
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Current Cardiology Reviews (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DR-NTU (Nanyang Technological University) (submittedVersion)
- Funders
- Univerzita Karlova v Praze
- Australian Government
- Charles Darwin University
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- photoplethysmography, digital volume pulse, aging, autonomic function, blood pressure, heart rate, pulse wave analysis, cardiovascular disease, vascular disease, atherosclerosis screening
- MeSH terms
- Fingers, Hemodynamics, Humans, Skin, Artifacts, Photoplethysmography, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review