On the Analysis of Fingertip Photoplethysmogram Signals
Open accessAbstract
Photoplethysmography (PPG) is used to estimate the skin blood flow using infrared light. Researchers from different domains of science have become increasingly interested in PPG because of its advantages as non-invasive, inexpensive, and convenient diagnostic tool. Traditionally, it measures the oxygen saturation, blood pressure, cardiac output, and for assessing autonomic functions. Moreover, PPG is a promising technique for early screening of various atherosclerotic pathologies and could be helpful for regular GP-assessment but a full understanding of the diagnostic value of the different features is still lacking. Recent studies emphasise the potential information embedded in the PPG waveform signal and it deserves further attention for its possible applications beyond pulse oximetry and heart-rate calculation. Therefore, this overview discusses different types of artifact added to PPG signal, characteristic features of PPG waveform, and existing indexes to evaluate for diagnoses. Keywords: Photoplethysmography, acceleration plethysmogram, second derivative plethysmogram, digital volume pulse, ageing, artery, autonomic function, blood pressure, cardiovascular, heart rate, pulse wave analysis, vascular disease
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures












![Fig. (13) Typical PPG waveforms show the parameters changes with age [62].](/paper-figures/8ee5a57901c347579a7a5160e4be1526/on-the-analysis-of-fingertip-photoplethysmogram-signals-figure-13.jpg)


![Fig. (16) APG waveforms and types of photoplethysmogram [73]. There are different types of APG waveforms. The first APG waveform A (far left) refers to good circulation, whereas the amplitude of b wave is lower than c wave. The last APG waveform G (far right) refers to distinctively bad circulation, whereas the amplitude of c wave is lower than b wave.](/paper-figures/54ec3ef6e18e4116ad4a30b26f4dd445/on-the-analysis-of-fingertip-photoplethysmogram-signals-figure-16.jpg)
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“| c/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“The shape of a clean PPG signal is closely related to the blood pulse wave of the human body25”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [25]
In the citing paper
“In this case, the pulsatile component of the PPG can be of the order of 1 mV in amplitude [55] although lower amplitudes can be caused by technical factors, such as inappropriate gain control, or physiological factors, such as vasoconstriction [9].”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [9]
In the citing paper
“In particular, obesity can lead to changes in skin thickness, blood flow, and oxygen saturation. This affects the optical properties of skin in addition to the distance light has to travel to reach a target vessel or vessels [48,59,60].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“While, to the authors’ knowledge, publications that experimentally and explicitly demonstrate the effect of obesity or BMI on the PPG waveform are limited, we will explore works which suggest that there would be a substantial effect [56,59,60,61,62,63,64,65,66,67,68,69,70].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“The ratio b/a has been found to correlate positively to age, while c/a, d/a and e/a have all been correlated negatively to age [29,59].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“| e/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“| d/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [59]
In the citing paper
“Furthermore, PPG can be used to index heart rate variability (HRV), blood pressure, oxygen saturation, and cardiac output24, but as of yet have not be widely rolled out in consumer-use devices and therefore data collection other than HR will not be discussed further.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [24]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
Assessment of fat-free mass using bioelectrical impedance measurements of the human body
Lukaski, H., Johnson, P., Bolonchuk, W., & Lykken, G. (1985) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/41.4.810 · Cited by 1,896
Investigation of Chronic Venous Insufficiency: A Consensus Statement
Nicolaides, A. N. (2000) · Circulation · doi:10.1161/01.cir.102.20.e126 · Cited by 945
Fifteen years experience with finger arterial pressure monitoring: assessment of the technology
Imholz, B. P. M. (1998) · Cardiovascular Research · doi:10.1016/s0008-6363(98)00067-4 · Cited by 841
Determination of age-related increases in large artery stiffness by digital pulse contour analysis
Millasseau, S., Kelly, R. P., Ritter, J., & Chowienczyk, P. (2002) · Clinical Science · doi:10.1042/cs1030371 · Cited by 624
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589
Photoelectric measurement of blood pressure, volume and flow in the finger
Peňáz, J. (1973) · Medical Entomology and Zoology · Cited by 528
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 460
Research works printed but not linked: 3
- [59]Takazawa KTN Fujita M Matsuoka O Saiki T Aikawa M Tamura S Ibukiyama C Assessment of vasocative agents and vascular aging by the second derivative of photoplethysmogram waveform Hypertension 1998 32 365 70 9719069
- [80]Elgendi M Jonkman M De Boer F Fred A Filipe J Gamboa h Heart Rate Variability and Acceleration Plethysmogram measured at rest Biomedical Engineering Systems and Technologies: 2011 Springer 266 77
- [81]Tokutaka H MY Gonda E Yamamoto M Construction of a General Physical Condition Judgment System Using Acceleration Plethysmogram Pulse-Wave Analysis 2009 Berlin/Heidelberg Springer 5629 307 15
Cited by
Sorted by citations
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945
Multifunctional materials for implantable and wearable photonic healthcare devices
Lee, G., Moon, H., Kim, H., Lee, G. H., Kwon, W., Yoo, S., Myung, D. J., Yun, S. H., Bao, Z., & Hahn, S. K. (2020) · Nature Reviews Materials · doi:10.1038/s41578-019-0167-3 · Cited by 683
A wearable cardiac ultrasound imager
Hu, H., Huang, H., Li, M., Gao, X., Yin, L., Qi, R., Wu, R. S., Chen, X., Ma, Y., Shi, K., et al. (2023) · Nature · doi:10.1038/s41586-022-05498-z · Cited by 644
The use of photoplethysmography for assessing hypertension
Elgendi, M., Fletcher, R., Liang, Y., Howard, N., Lovell, N. H., Abbott, D., Lim, K., & Ward, R. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0136-7 · Cited by 625
Self-citationCuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621
Self‐Powered Pulse Sensor for Antidiastole of Cardiovascular Disease
Ouyang, H., Tian, J., Sun, G., Zou, Y., Liu, Z., Li, H., Zhao, L., Shi, B., Fan, Y., Fan, Y., et al. (2017) · Advanced Materials · doi:10.1002/adma.201703456 · Cited by 470
A Wireless Textile Based Sensor System for Self-Powered Personalized Health Care
Meng, K., Zhao, S., Zhou, Y., Wu, Y., Zhang, S., He, Q., Wang, X., Zhou, Z., Fan, W., Tan, X., et al. (2019) · Matter · doi:10.1016/j.matt.2019.12.025 · Cited by 454
Cuff-less high-accuracy calibration-free blood pressure estimation using pulse transit time
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2015) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2015.7168806 · Cited by 351
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348
“The ratio b/a has been found to correlate positively to age, while c/a, d/a and e/a have all been correlated negatively to age [29,59].”
BackgroundOptimal Signal Quality Index for Photoplethysmogram Signals
Elgendi, M. (2016) · Bioengineering · doi:10.3390/bioengineering3040021 · Cited by 338
Self-citationBota, P. J., Wang, C., Fred, A. L. N., & Silva, H. (2019) · IEEE Access · doi:10.1109/access.2019.2944001 · Cited by 348
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101870 · Cited by 300
Thin, soft, wearable system for continuous wireless monitoring of artery blood pressure
Li, J., Jia, H., Zhou, J., Huang, X., Xu, L., Jia, S., Gao, Z., Yao, K., Li, D., Zhang, B., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-40763-3 · Cited by 321
Flexible graphene photodetectors for wearable fitness monitoring
Polat, E. O., Mercier, G., Nikitskiy, I., Puma, E., Galán, T., Gupta, S., Marques, M. M., Piqueras, J. J., Bouwens, M., Durduran, T., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw7846 · Cited by 301
Elgendi, M., Norton, I., Brearley, M. B., Abbott, D., & Schuurmans, D. (2013) · PLoS ONE · doi:10.1371/journal.pone.0076585 · Cited by 282
Self-citationWearable-Based Affect Recognition—A Review
Schmidt, P., Reiss, A., Dürichen, R., & Laerhoven, K. V. (2019) · Sensors · doi:10.3390/s19194079 · Cited by 291
Chowdhury, M. H., Shuzan, M. N. I., Chowdhury, M. E. H., Mahbub, Z. B., Uddin, M. M., Khandakar, A., & Reaz, M. B. I. (2020) · Sensors · doi:10.3390/s20113127 · Cited by 276
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263
How to use digital devices to detect and manage arrhythmias: an EHRA practical guide
Svennberg, E., Tjong, F., Goette, A., Akoum, N., Di Biase, L., Bordachar, P., Boriani, G., Burri, H., Conte, G., Deharo, J. C., et al. (2022) · Europace · doi:10.1093/europace/euac038 · Cited by 261
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250
An optimal filter for short photoplethysmogram signals
Liang, Y., Elgendi, M., Chen, Z., & Ward, R. (2018) · Scientific Data · doi:10.1038/sdata.2018.76 · Cited by 242
Self-citationGuidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
Nelson, B. W., Low, C. A., Jacobson, N., Areán, P., Torous, J., & Allen, N. B. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0297-4 · Cited by 241
HeartPy: A novel heart rate algorithm for the analysis of noisy signals
van Gent, P., Farah, H., van Nes, N., & van Arem, B. (2019) · Transportation Research Part F: Traffic Psychology and Behaviour · doi:10.1016/j.trf.2019.09.015 · Cited by 234
Blood pressure estimation from appropriate and inappropriate PPG signals using A whole-based method
Mousavi, S. S., Firouzmand, M., Charmi, M., Hemmati, M. R., Moghadam, M., & Ghorbani, Y. (2018) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2018.08.022 · Cited by 234
Heart rate variability with photoplethysmography in 8 million individuals: a cross-sectional study
Natarajan, A., Pantelopoulos, A., Emir-Farinas, H., & Natarajan, P. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(20)30246-6 · Cited by 229
Local Crack‐Programmed Gold Nanowire Electronic Skin Tattoos for In‐Plane Multisensor Integration
Gong, S. P., Yap, L. W., Zhu, B., Zhai, Q., Liu, Y., Lyu, Q., Wang, K., Yang, M., Ling, Y., Lai, D. T. H., et al. (2019) · Advanced Materials · doi:10.1002/adma.201903789 · Cited by 224
EM-Wave Biosensors: A Review of RF, Microwave, mm-Wave and Optical Sensing
Mehrotra, P., Chatterjee, B., & Sen, S. (2019) · Sensors · doi:10.3390/s19051013 · Cited by 222
Real-world heart rate norms in the Health eHeart study
Avram, R., Tison, G. H., Aschbacher, K., Kuhar, P., Vittinghoff, E., Butzner, M., Runge, R., Wu, N., Pletcher, M. J., Marcus, G. M., et al. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0134-9 · Cited by 217
A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China
Liang, Y., Chen, Z., Liu, G., & Elgendi, M. (2018) · Scientific Data · doi:10.1038/sdata.2018.20 · Cited by 215
Self-citationKirigami‐Inspired Pressure Sensors for Wearable Dynamic Cardiovascular Monitoring
Meng, K., Xiao, X., Liu, Z., Shen, S., Tat, T., Wang, Z., Lu, C., Ding, W., He, X., Yang, J., et al. (2022) · Advanced Materials · doi:10.1002/adma.202202478 · Cited by 210
Diagnostic Features and Potential Applications of PPG Signal in Healthcare: A Systematic Review
Almarshad, M. A., Islam, M. S., Al-Ahmadi, S. A., & BaHammam, A. S. (2022) · Healthcare · doi:10.3390/healthcare10030547 · Cited by 215
Tanveer, M. S., & Hasan, M. K. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.02.028 · Cited by 200
Cuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180
Hu, Z., Lou, S., Xing, Y., Wang, X., Cao, D., & Lv, C. (2022) · IEEE Transactions on Intelligent Vehicles · doi:10.1109/tiv.2022.3195635 · Cited by 189
Hartmann, V., Liu, H., Chen, F., Qiu, Q., Hughes, S., & Zheng, D. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00198 · Cited by 173
Photoplethysmography and Deep Learning: Enhancing Hypertension Risk Stratification
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Biosensors · doi:10.3390/bios8040101 · Cited by 176
Self-citationKaisti, M., Panula, T., Leppänen, J., Punkkinen, R., Tadi, M. J., Vasankari, T., Jaakkola, S., Kiviniemi, T., Airaksinen, J. K. E., Kostiainen, P., et al. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0117-x · Cited by 172
Camera Measurement of Physiological Vital Signs
McDuff, D. (2022) · ACM Computing Surveys · doi:10.1145/3558518 · Cited by 176
Li, K., Cardoso, C., Moctezuma-Ramirez, A., Elgalad, A., & Perin, E. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20247146 · Cited by 160
A novel feature ranking algorithm for biometric recognition with PPG signals
Kavsaoğlu, A. R., Polat, K., & Bozkurt, M. R. (2014) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2014.03.005 · Cited by 160
Stressing the accuracy: Wrist‐worn wearable sensor validation over different conditions
Menghini, L., Gianfranchi, E., Cellini, N., Patron, E., Tagliabue, M., & Sarlo, M. (2019) · Psychophysiology · doi:10.1111/psyp.13441 · Cited by 157
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102301 · Cited by 156
Arza, A., Garzón-Rey, J. M., Lázaro, J., Gil, E., Lopez-Anton, R., de la Camara, C., Laguna, P., Bailon, R., & Aguiló, J. (2019) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-018-1879-z · Cited by 148
Martinez-Ríos, E. A., Montesinos, L., Alfaro-Ponce, M., & Pecchia, L. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102813 · Cited by 147
Blood Pressure Estimation Using Photoplethysmogram Signal and Its Morphological Features
Hasanzadeh, N., Ahmadi, M. M., & Mohammadzade, H. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2961411 · Cited by 144
Noninvasive Blood Glucose Monitoring Systems Using Near-Infrared Technology—A Review
Hina, A., & Saadeh, W. (2022) · Sensors · doi:10.3390/s22134855 · Cited by 156
Charlton, P. H., Bonnici, T., Tarassenko, L., Alastruey, J., Clifton, D. A., Beale, R. J., & Watkinson, P. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa670e · Cited by 149
Cuffless blood pressure estimation from PPG signals and its derivatives using deep learning models
El-Hajj, C., & Kyriacou, P. A. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102984 · Cited by 140
von Wowern, E., Östling, G., Nilsson, P. M., & Olofsson, P. (2015) · PLOS ONE · doi:10.1371/journal.pone.0135659 · Cited by 134
“The e wave represents the dictrotic notch and corresponds to the aortic valve closure [11].”
InfluentialA digital biomarker of diabetes from smartphone-based vascular signals
Avram, R., Olgin, J. E., Kuhar, P., Hughes, J. W., Marcus, G. M., Pletcher, M. J., Aschbacher, K., & Tison, G. H. (2020) · Nature Medicine · doi:10.1038/s41591-020-1010-5 · Cited by 136
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138
Hypertension Assessment via ECG and PPG Signals: An Evaluation Using MIMIC Database
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Diagnostics · doi:10.3390/diagnostics8030065 · Cited by 132
Self-citationAssessing mental stress from the photoplethysmogram: a numerical study
Charlton, P. H., Celka, P., Farukh, B., Chowienczyk, P., & Alastruey, J. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aabe6a · Cited by 131
Wearable Emotion Recognition System based on GSR and PPG Signals
Udovičić, G., Đerek, J., Russo, M., & Sikora, M. (2017) · International Workshop on Multimedia for Personal Health and Health Care (MMHealth) · doi:10.1145/3132635.3132641 · Cited by 129
A Critical Review of Proactive Detection of Driver Stress Levels Based on Multimodal Measurements
RASTGOO, M. N., Nakisa, B., Rakotonirainy, A., Chandran, V., & Tjondronegoro, D. (2018) · ACM Computing Surveys · doi:10.1145/3186585 · Cited by 131
Varaki, E. S., Gargiulo, G. D., Penkala, S., & Breen, P. P. (2018) · BioMedical Engineering OnLine · doi:10.1186/s12938-018-0494-4 · Cited by 126
Body Sensor Network for Mobile Health Monitoring, a Diagnosis and Anticipating System
Wannenburg, J., & Malekian, R. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2464773 · Cited by 122
Ou‐Yang, W., Liu, L., Xie, M., Zhou, S., Hu, X., Wu, H., Tian, Z., Chen, X., Zhu, Y., & Li, J. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.109151 · Cited by 117
PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement
Chandrasekhar, A., Yavarimanesh, M., Natarajan, K., Hahn, J., & Mukkamala, R. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.2976989 · Cited by 123
Toward Generating More Diagnostic Features from Photoplethysmogram Waveforms
Elgendi, M., Liang, Y., & Ward, R. K. (2018) · Diseases · doi:10.3390/diseases6010020 · Cited by 107
Self-citationDu, Y., & Stephanus, A. (2018) · Sensors · doi:10.3390/s18072322 · Cited by 107
Wang, W., Mohseni, P., Kilgore, K. L., & Najafizadeh, L. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3128383 · Cited by 109
Ibrahim, B., & Jafari, R. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2946661 · Cited by 108
Harfiya, L. N., Chang, C.-C., & Li, Y. (2021) · Sensors · doi:10.3390/s21092952 · Cited by 110
Detection of a and b waves in the acceleration photoplethysmogram
Elgendi, M., Norton, I., Brearley, M. B., Abbott, D., & Schuurmans, D. (2014) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-13-139 · Cited by 100
Self-citationA Noninvasive Glucose Monitoring SoC Based on Single Wavelength Photoplethysmography
Hina, A., & Saadeh, W. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.2979514 · Cited by 99
Hypertension Assessment Using Photoplethysmography: A Risk Stratification Approach
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Journal of Clinical Medicine · doi:10.3390/jcm8010012 · Cited by 99
Self-citationMeasurement of respiratory rate using wearable devices and applications to COVID-19 detection
Natarajan, A., Su, H.-W., Heneghan, C., Blunt, L., O’Connor, C., & Niehaus, L. (2021) · npj Digital Medicine · doi:10.1038/s41746-021-00493-6 · Cited by 99
Cuff-less PPG based continuous blood pressure monitoring — A smartphone based approach
Gaurav, A., Maheedhar, M., Tiwari, V. N., & Narayanan, R. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7590775 · Cited by 100
Delbeck, S., Vahlsing, T., Leonhardt, S. K., Steiner, G. R., & Heise, H. M. (2018) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1395-x · Cited by 95
Liang, Y., Abbott, D., Howard, N., Lim, K. I., Ward, R. K., & Elgendi, M. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8030337 · Cited by 94
Self-citationTrueHeart: Continuous Authentication on Wrist-worn Wearables Using PPG-based Biometrics
Zhao, T., Wang, Y., Liu, J., Chen, Y. J., Cheng, J., & Yu, J. (2020) · IEEE Conference on Computer Communications (INFOCOM) · doi:10.1109/infocom41043.2020.9155526 · Cited by 92
Hsu, Y.-C., Li, Y., Chang, C.-C., & Harfiya, L. N. (2020) · Sensors · doi:10.3390/s20195668 · Cited by 92
Data quality evaluation in wearable monitoring
Böttcher, S., Vieluf, S., Bruno, E., Joseph, B., Epitashvili, N., Biondi, A., Zabler, N., Glasstetter, M., Dümpelmann, M., Van Laerhoven, K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-25949-x · Cited by 91
An Algorithm for Real-Time Pulse Waveform Segmentation and Artifact Detection in Photoplethysmograms
Fischer, C., Dömer, B., Wibmer, T., & Penzel, T. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2518202 · Cited by 95
Song, Z., Li, W., Bao, Y., Wang, W., Liu, Z., Han, F., Han, D., & Niu, L. (2018) · Advanced Electronic Materials · doi:10.1002/aelm.201800252 · Cited by 90
Chan, G., Cooper, R. E., Hosanee, M., Welykholowa, K., Kyriacou, P. A., Zheng, D., Allen, J., Abbott, D., Lovell, N. H., Fletcher, R., et al. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8111827 · Cited by 89
Self-citationSensors Capabilities, Performance, and Use of Consumer Sleep Technology
de Zambotti, M., Cellini, N., Menghini, L., Sarlo, M., & Baker, F. C. (2020) · Sleep Medicine Clinics · doi:10.1016/j.jsmc.2019.11.003 · Cited by 95
Development and Validation of a Novel Cuff-Less Blood Pressure Monitoring Device
Watanabe, N., Bando, Y., Kawachi, T., Yamakita, H., Futatsuyama, K., Honda, Y., Yasui, H., Nishimura, K., Kamihara, T., Okumura, T., et al. (2017) · JACC Basic to Translational Science · doi:10.1016/j.jacbts.2017.07.015 · Cited by 88
Rong, M., & Li, K. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102772 · Cited by 85
Estimating blood pressure trends and the nocturnal dip from photoplethysmography
Radha, M., de Groot, K., Rajani, N. B., Wong, C. C. P., Kobold, N., Vos, V., Fonseca, P. M., Mastellos, N., Wark, P. A., Velthoven, N., et al. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab030e · Cited by 90
Pankaj, Kumar, A., Komaragiri, R., & Kumar, M. (2021) · Archives of Computational Methods in Engineering · doi:10.1007/s11831-021-09597-4 · Cited by 84
Shrivastava, A., Chakkaravarthy, M., & Shah, M. A. (2023) · Healthcare Analytics · doi:10.1016/j.health.2023.100219 · Cited by 84
A Wearable Device and System for Movement and Biometric Data Acquisition for Sports Applications
Kos, M., & Kramberger, I. (2017) · IEEE Access · doi:10.1109/access.2017.2675538 · Cited by 83
Finnegan, E., Davidson, S. M., Harford, M., Watkinson, P., Tarassenko, L., & Villarroel, M. (2023) · Scientific Reports · doi:10.1038/s41598-022-27170-2 · Cited by 87
A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability
Sarhaddi, F., Kazemi, K., Azimi, I., Cao, R., Niela‐Vilén, H., Axelin, A., Liljeberg, P., & Rahmani, A. M. (2022) · PLoS ONE · doi:10.1371/journal.pone.0268361 · Cited by 93
Maqsood, S., Xu, S., Springer, M., & Mohawesh, R. (2021) · IEEE Access · doi:10.1109/access.2021.3117969 · Cited by 84
Kotzen, K., Charlton, P. H., Salabi, S., Amar, L., Landesberg, A., & Behar, J. A. (2022) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2022.3225363 · Cited by 80
Jacoutot, P., Scaccabarozzi, A. D., Nodari, D., Panidi, J., Qiao, Z., Schiza, A., Nega, A. D., Dimitrakopoulou‐Strauss, A., Gregoriou, V. G., Heeney, M., et al. (2023) · Science Advances · doi:10.1126/sciadv.adh2694 · Cited by 80
InstaBP: Cuff-less Blood Pressure Monitoring on Smartphone using Single PPG Sensor
Dey, J., Gaurav, A., & Tiwari, V. N. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8513189 · Cited by 79
An Overview of the Sensors for Heart Rate Monitoring Used in Extramural Applications
Galli, A., Montree, R. J. H., Que, S., Peri, E., & Vullings, R. (2022) · Sensors · doi:10.3390/s22114035 · Cited by 87
Weber, B., Fischer, T., & Riedl, R. (2021) · Journal of Systems and Software · doi:10.1016/j.jss.2021.110946 · Cited by 82
Sharma, P., Choudhary, K., Gupta, K., Chawla, R., Gupta, D. K., & Sharma, A. (2019) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2019.101752 · Cited by 76
Evaluation of the Time Stability and Uniqueness in PPG-Based Biometric System
Hwang, D. Y., Taha, B., Lee, D. S., & Hatzinakos, D. (2020) · IEEE Transactions on Information Forensics and Security · doi:10.1109/tifs.2020.3006313 · Cited by 76
Sleep Deprivation Deteriorates Heart Rate Variability and Photoplethysmography
Bourdillon, N., Jeanneret, F., Nilchian, M., Albertoni, P., Ha, P., & Millet, G. P. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.642548 · Cited by 80
Natta, L., Mastronardi, V., Guido, F., Algieri, L., Puce, S., Pisano, F., Rizzi, F., Pulli, R., Qualtieri, A., & Vittorio, M. D. (2019) · Scientific Reports · doi:10.1038/s41598-019-44784-1 · Cited by 76
Golgouneh, A., & Tarvirdizadeh, B. (2019) · Neural Computing and Applications · doi:10.1007/s00521-019-04278-7 · Cited by 78
Analysis on Four Derivative Waveforms of Photoplethysmogram (PPG) for Fiducial Point Detection
Suboh, M. Z., Jaafar, R., Nayan, N. A., Harun, N. H., & Mohamad, M. S. F. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.920946 · Cited by 75
Djeldjli, D., Bousefsaf, F., Maaoui, C., Bereksi-Reguig, F., & Pruski, A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102242 · Cited by 74
Video-based real-time monitoring for heart rate and respiration rate
Alnaggar, M., Siam, A. I., Handosa, M., Medhat, T., & Rashad, M. Z. (2023) · Expert Systems with Applications · doi:10.1016/j.eswa.2023.120135 · Cited by 74
Metadata
- Authors
- Elgendi, M.
- Affiliations
- Charles Darwin University (AU)
- Corresponding author
- Elgendi, Mohamed
- ORCID
- Elgendi, Mohamed
- Venue
- Current Cardiology Reviews
- Publisher
- Bentham Science Publishers Ltd.
- Published
- 2012-06-01
- Volume
- 8
- Issue
- 1
- Pages
- 14-25
- ISSN
- 1573-403X (ISSN-L)
- PMC
- PMC3394104
- OpenAlex
- W1981918735
- Cited by
- 1,206
- References
- 88
- Citations by year
- 2026: 120 · 2025: 106 · 2024: 116 · 2023: 109 · 2022: 130 · 2021: 114 · 2020: 118 · 2019: 118 · 2018: 86 · 2017: 65 · 2016: 46 · 2015: 42 · 2014: 24 · 2013: 11 · 2012: 1
- FWCI
- 2.17
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Current Cardiology Reviews (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DR-NTU (Nanyang Technological University) (submittedVersion)
- Funders
- Univerzita Karlova v Praze
- Australian Government
- Charles Darwin University
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- photoplethysmography, digital volume pulse, aging, autonomic function, blood pressure, heart rate, pulse wave analysis, cardiovascular disease, vascular disease, atherosclerosis screening
- MeSH terms
- Fingers, Hemodynamics, Humans, Skin, Artifacts, Photoplethysmography, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review