Timing of Changes in Interstitial and Venous Blood Glucose Measured With a Continuous Subcutaneous Glucose Sensor
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
The objective of this study was to use a subcutaneous continuous glucose sensor to determine time differences in the dynamics of blood glucose and interstitial glucose. A total of 14 patients with type 1 diabetes each had two sensors (Medtronic/MiniMed CGMS) placed subcutaneously in the abdomen, acquiring data every 5 min. Blood glucose was sampled every 5 min for 8 h, and two liquid meals were given. A smoothing algorithm was applied to the blood glucose and interstitial glucose curves. The first derivatives of the glucose traces defined and quantified the timing of rises, peaks, falls, and nadirs. Altogether, 24 datasets were used for the analysis of time differences between interstitial and blood glucose and between sensors in each patient. Time differences between blood and interstitial glucose ranged from 4 to 10 min, with the interstitial glucose lagging behind blood glucose in 81% of cases (95% CIs 72.5 and 89.5%). The mean (±SD) difference between the two sensors in each patient was 6.7 ± 5.1 min, representing random variation in sensor response. In conclusion, there is a time lag of interstitial glucose behind blood glucose, regardless of whether glycemia is rising or falling, but intersensor variability is considerable in this sensor system. Comparisons of interstitial and blood glucose kinetics must take statistical account of variability between sensors.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The MiniMed Continuous Glucose Monitoring System
Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050214078 · Cited by 393
Rebrin, K., Steil, G. M., van Antwerp, W. P., & Mastrototaro, J. J. (1999) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1999.277.3.e561 · Cited by 366
Noninvasive Glucose Monitoring: Comprehensive Clinical Results
Tamada, J. A., Garg, S., Jovanovic, L., Pitzer, K. R., Fermi, S., Potts, R. O., & and the Cygnus Research Team (1999) · JAMA · doi:10.1001/jama.282.19.1839 · Cited by 282
Kashiwagi, A., Verso, M. A., Andrews, J., Vasquez, B., Reaven, G., & Foley, J. E. (1983) · Journal of Clinical Investigation · doi:10.1172/jci111080 · Cited by 250
Glucose monitoring by reverse iontophoresis
Potts, R. O., A. Tamada, J., & J. Tierney, M. (2002) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.210 · Cited by 243
Zierler, K. (1999) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1999.276.3.e409 · Cited by 217
McGowan, K., Thomas, W., & Moran, A. (2002) · Diabetes Care · doi:10.2337/diacare.25.9.1499 · Cited by 148
Performance of subcutaneously implanted glucose sensors for continuous monitoring
Gerritsen, M., Jansen, J. A., & Lutterman, J. A. (1999) · The Netherlands Journal of Medicine · doi:10.1016/s0300-2977(99)00006-6 · Cited by 140
Thomé-Duret, V., Reach, G., Gangnerau, M. N., Lemonnier, F., Klein, J. C., Zhang, Y., Hu, Y., & Wilson, G. S. (1996) · Analytical Chemistry · doi:10.1021/ac960069i · Cited by 118
Characterization by microdialysis of intracellular glucose level in subcutaneous tissue in humans
Jansson, P. A., Fowelin, J., Smith, U., & Lonnroth, P. (1988) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1988.255.2.e218 · Cited by 118
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Wearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315
Bandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav3294 · Cited by 761
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Glucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584
Diabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
“of the direction of BG change [196], [207], [208].”
Phillip, M., Nimri, R., Bergenstal, R. M., Barnard-Kelly, K., Danne, T., Hovorka, R., Kovatchev, B. P., Messer, L. H., Parkin, C. G., Ambler-Osborn, L., et al. (2023) · Endocrine Reviews · doi:10.1210/endrev/bnac022 · Cited by 443
Skin-Attachable, Stretchable Electrochemical Sweat Sensor for Glucose and pH Detection
Oh, S. Y., Hong, S. Y., Jeong, Y. R., Yun, J., Park, H., Jin, S. W., Lee, G., Oh, J. H., Lee, H., Lee, S.-S., et al. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b03342 · Cited by 440
A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
Cengiz, E., & Tamborlane, W. V. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0002 · Cited by 390
Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy
Yadav, J., Rani, A., Singh, V., & Murari, B. M. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.01.005 · Cited by 374
Lipani, L., Dupont, B. G. R., Doungmene, F., Marken, F., Tyrrell, R. M., Guy, R. H., & Ilie, A. (2018) · Nature Nanotechnology · doi:10.1038/s41565-018-0112-4 · Cited by 340
Comprehensive Review on Wearable Sweat-Glucose Sensors for Continuous Glucose Monitoring
Zafar, H., Channa, A., Jeoti, V., & Stojanović, G. M. (2022) · Sensors · doi:10.3390/s22020638 · Cited by 329
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312
Noninvasive Monitoring of Blood Glucose with Raman Spectroscopy
Pandey, R., Paidi, S. K., Valdez, T. A., Zhang, C., Spegazzini, N., Dasari, R. R., & Barman, I. (2017) · Accounts of Chemical Research · doi:10.1021/acs.accounts.6b00472 · Cited by 290
Technologies for Continuous Glucose Monitoring: Current Problems and Future Promises
Vaddiraju, S., Burgess, D. J., Tomazos, I., Jain, F. C., & Papadimitrakopoulos, F. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400632 · Cited by 285
Torimoto, K., Okada, Y., Mori, H., & Tanaka, Y. (2013) · Cardiovascular Diabetology · doi:10.1186/1475-2840-12-1 · Cited by 284
Non-Invasive Glucose Measurement Technologies: An Update from 1999 to the Dawn of the New Millennium
Khalil, O. S. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.660 · Cited by 255
Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Kovatchev, B., Anderson, S., Heinemann, L., & Clarke, W. (2008) · Diabetes Care · doi:10.2337/dc07-2401 · Cited by 236
Xie, X., Doloff, J. C., Yesilyurt, V., Sadraei, A., McGarrigle, J. J., Omami, M., Veiseh, O., Farah, S., Isa, D., Ghani, S., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0273-3 · Cited by 235
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Ludvigsson, J., & Hanas, R. (2003) · Pediatrics · doi:10.1542/peds.111.5.933 · Cited by 228
Flexible inorganic bioelectronics
Chen, Y., Zhang, Y., Liang, Z., Cao, Y., Han, Z., & Feng, X. (2020) · npj Flexible Electronics · doi:10.1038/s41528-020-0065-1 · Cited by 227
Zhu, J., Zhou, X., Kim, H., Qu, M., Jiang, X., Lee, K., Ren, L., Wu, Q., Wang, C., Zhu, X., et al. (2020) · Small · doi:10.1002/smll.201905910 · Cited by 214
Time Lag of Glucose From Intravascular to Interstitial Compartment in Humans
Basu, A., Dube, S., Slama, M., Errazuriz, I., Amezcua, J. C., Kudva, Y. C., Peyser, T., Carter, R. E., Cobelli, C., & Basu, R. (2013) · Diabetes · doi:10.2337/db13-1132 · Cited by 205
“The time lag between plasma and ISF glucose appears to differ depending on whether plasma glucose values are rising or falling (9-11) or the type of CGM instrument and sensor algorithm used (17; 18).”
BackgroundMethodKovatchev, B. P., Clarke, W. L., Breton, M., Brayman, K., & McCall, A. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.849 · Cited by 201
“The accomplish this QUANTIFYING GLUCOSE VARIABILITY time delays that may be inherent with interstitial glucose measurement—recent publications 3,25 6 For the purposes of this manuscript we assume that CGMs provide a sufficiently accurate representation recorded in a frequently sampled time series.”
Mazze, R. S., Strock, E., Wesley, D., Borgman, S., Morgan, B., Bergenstal, R., & Cuddihy, R. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0293 · Cited by 182
“21,22,30,31 To further reduce the inherent differences be-tween measuring glucose in blood and interstitial fluid, we employed a smoothing algo-rithm to compute values that comprise the AGP frequency curves.”
Ribet, F., Stemme, G., & Roxhed, N. (2018) · Biomedical Microdevices · doi:10.1007/s10544-018-0349-6 · Cited by 181
“…lag time between blood glucose and interstitial fluid glucose concentrations, which is of the order of 4 to 12 min (Aussedat et al. 2000; Boyne et al. 2003), and the CGMS delay itself impedes real-time monitoring, and hence the possibility to take medical decisions, such as insulin…”
Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals
Gough, D. A., Kumosa, L. S., Routh, T. L., Lin, J. T., & Lucisano, J. Y. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3001148 · Cited by 179
Freckmann, G., Hagenlocher, S., Baumstark, A., Jendrike, N., Gillen, R. C., Rössner, K., & Haug, C. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100513 · Cited by 174
Analysis, Modeling, and Simulation of the Accuracy of Continuous Glucose Sensors
Breton, M., & Kovatchev, B. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200517 · Cited by 173
Time Delay of CGM Sensors: Relevance, Causes, and Countermeasures
Schmelzeisen-Redeker, G., Schoemaker, M., Kirchsteiger, H., Freckmann, G., Heinemann, L., & del Re, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815590154 · Cited by 161
Rebrin, K., Sheppard, N. F., & Steil, G. M. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400507 · Cited by 152
Optimal prandial timing of bolus insulin in diabetes management: a review
Slattery, D., Amiel, S. A., & Choudhary, P. (2018) · Diabetic Medicine · doi:10.1111/dme.13525 · Cited by 132
Wearable Skin Sensors and Their Challenges: A Review of Transdermal, Optical, and Mechanical Sensors
Ahmad Tarar, A., Mohammad, U., & K. Srivastava, S. (2020) · Biosensors · doi:10.3390/bios10060056 · Cited by 119
Clinical Use of Continuous Glucose Monitoring in Adults with Type 1 Diabetes
Slattery, D., & Choudhary, P. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0051 · Cited by 87
A Century of Diabetes Technology: Signals, Models, and Artificial Pancreas Control
Kovatchev, B. (2019) · Trends in Endocrinology & Metabolism · doi:10.1016/j.tem.2019.04.008 · Cited by 80
Continuous Glucose Sensors: Continuing Questions about Clinical Accuracy
Clarke, W. L., & Kovatchev, B. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100510 · Cited by 71
Burnett, D. R., Huyett, L. M., Zisser, H. C., Doyle, F. J., & Mensh, B. D. (2014) · Diabetes · doi:10.2337/db13-1649 · Cited by 52
Wolf, A. K., Qin, Y., Major, T. C., & Meyerhoff, M. E. (2015) · Chinese Chemical Letters · doi:10.1016/j.cclet.2015.03.002 · Cited by 21
Metadata
- Authors
- Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D.
- Venue
- Diabetes
- Publisher
- American Diabetes Association
- Published
- 2003-11-01
- Volume
- 52
- Issue
- 11
- Pages
- 2790-2794
- ISSN
- 0012-1797 (ISSN-L)
- 1939-327X
- OpenAlex
- W2142534938
- Cited by
- 404
- References
- 16
- FWCI
- 7.74
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.