Contact Lens Sensors in Ocular Diagnostics
Abstract
Contact lenses as a minimally invasive platform for diagnostics and drug delivery have emerged in recent years. Contact lens sensors have been developed for analyzing the glucose composition of tears as a surrogate for blood glucose monitoring and for the diagnosis of glaucoma by measuring intraocular pressure. However, the eye offers a wider diagnostic potential as a sensing site and therefore contact lens sensors have the potential to improve the diagnosis and treatment of many diseases and conditions. With advances in polymer synthesis, electronics and micro/nanofabrication, contact lens sensors can be produced to quantify the concentrations of many biomolecules in ocular fluids. Non‐ or minimally invasive contact lens sensors can be used directly in a clinical or point‐of‐care setting to monitor a disease state continuously. This article reviews the state‐of‐the‐art in contact lens sensor fabrication, their detection, wireless powering, and readout mechanisms, and integration with mobile devices and smartphones. High‐volume manufacturing considerations of contact lenses are also covered and a case study of an intraocular pressure contact lens sensor is provided as an example of a successful product. This Review further analyzes the contact lens market and the FDA regulatory requirements for commercialization of contact lens sensors.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Blood and tear fluid have a compositional difference due to the blood–tear barrier; however, the concentrations of analytes in both fluids have a relationship as a result of plasma leakage.57”
From Wearables in Medicine (2018) · cites this paper as [57]
In the citing paper
“Not only do biomarker molecules in tears diffuse directly from the blood and exhibit close tear–blood concentration correlations, but also tear analysis presents opportunities for the diagnosis of ocular disease.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [94]
In the citing paper
“Not only do biomarker molecules in tears diffuse directly from the blood and exhibit close tear–blood concentration correlations, but also tear analysis presents opportunities for the diagnosis of ocular disease.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [94]
In the citing paper
“In particular, tear glucose concentrations have been demonstrated to exhibit good correlation with blood glucose levels, reflecting the diffusion from the lachrymal artery when the tears are sampled without any eye irritation or stimulation, which can compromise the relationship94, 95, 99-101.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [94]
In the citing paper
“Given this background, a smart contact lens is a promising example of a wearable, health monitoring device (13, 14).”
From Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays (2018) · cites this paper as [13]
In the citing paper
“By considering the correlation between the tear glucose level and blood glucose level (15), a glucose sensor fitted on a contact lens can provide the noninvasive monitoring of user’s glucose levels from tear fluids with a consideration of the lag time between tear glucose level and blood glucose level in the range of 10 to 20 min (13–16).”
From Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays (2018) · cites this paper as [13]
In the citing paper
“In addition, the user’s tears can be collected in the contact lens by completely natural means, such as normal secretion and blinking, and used to assess various biomarkers found in the blood, such as glucose, cholesterol, sodium ions, and potassium ions (13).”
From Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays (2018) · cites this paper as [13]
Citations
Cites
The present and future role of microfluidics in biomedical research
Sackmann, E. K., Fulton, A. L., & Beebe, D. J. (2014) · Nature · doi:10.1038/nature13118 · Cited by 2,995
Fluorescence Spectroscopy of Single Biomolecules
Weiss, S. (1999) · Science · doi:10.1126/science.283.5408.1676 · Cited by 2,060
Clinical Methods: The History, Physical, and Laboratory Examinations
Kruse, J. A. (1990) · JAMA · doi:10.1001/jama.1990.03450210108045 · Cited by 2,040
Paper-based microfluidic point-of-care diagnostic devices
Yetisen, A. K., Akram, M. S., & Lowe, C. R. (2013) · Lab on a Chip · doi:10.1039/c3lc50169h · Cited by 2,017
Self-citationClinical Methods: The History, Physical, and Laboratory Examinations
(1990) · Annals of Internal Medicine · doi:10.7326/0003-4819-113-7-563_2 · Cited by 1,408
Geigy, J. R. (1981) · Medical Entomology and Zoology · Cited by 1,221
The renaissance of fluorescence resonance energy transfer.
Selvin, P. R. (2000) · Nature Structural Biology · doi:10.1038/78948 · Cited by 998
Functional aspects of the tear film lipid layer
Bron, A. J., Tiffany, J. M., Gouveia, S. M., Yokoi, N., & Voon, L. W. (2003) · Experimental Eye Research · doi:10.1016/j.exer.2003.09.019 · Cited by 784
Solomon, A., Dursun, D., Liu, Z., Xie, Y., Macrì, A., & Pflugfelder, S. C. (2001) · PubMed · Cited by 721
Pflugfelder, S. C., Jones, D., Ji, Z., Afonso, A. A., & Monroy, D. C. (1999) · Current Eye Research · doi:10.1076/ceyr.19.3.201.5309 · Cited by 618
Fluorescence-based glucose sensors
Pickup, J. C., Hussain, F., Evans, N. D., Rolinski, O. J., & Birch, D. J. S. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.10.002 · Cited by 616
Kratz (2004) · doi:10.1056/nejmcpc049016 · Cited by 593
Human body fluid proteome analysis
Hu, S., Loo, J. A., & Wong, D. T. (2006) · PROTEOMICS · doi:10.1002/pmic.200600284 · Cited by 535
A contact lens with embedded sensor for monitoring tear glucose level
Yao, H., Shum, A. J., Cowan, M., Lähdesmäki, I., & Parviz, B. A. (2011) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.12.042 · Cited by 513
Definition of glaucoma: clinical and experimental concepts
Casson, R. J., Chidlow, G., Wood, J. P. M., Crowston, J. G., & Goldberg, I. (2012) · Clinical and Experimental Ophthalmology · doi:10.1111/j.1442-9071.2012.02773.x · Cited by 510
Berman, E. R. (1991) · Springer eBooks · doi:10.1007/978-1-4757-9441-0 · Cited by 482
Commercialization of microfluidic devices
Volpatti, L. R., & Yetisen, A. K. (2014) · Trends in biotechnology · doi:10.1016/j.tibtech.2014.04.010 · Cited by 437
Self-citationPhotonic Crystal Glucose-Sensing Material for Noninvasive Monitoring of Glucose in Tear Fluid
Alexeev, V. L., Das, S., Finegold, D. N., & Asher, S. A. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2004.039701 · Cited by 370
Russell, R. J., Pishko, M. V., Gefrides, C. C., McShane, M. J., & Coté, G. L. (1999) · Analytical Chemistry · doi:10.1021/ac990060r · Cited by 354
Fluid, Electrolyte, and Acid-Base Disorders in Small Animal Practice
DiBartola, S. P. (2012) · Elsevier eBooks · doi:10.1016/c2009-0-60133-1 · Cited by 351
Tear analysis in ocular surface diseases
Zhou, L., & Beuerman, R. W. (2012) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2012.06.002 · Cited by 318
Imaging of the Dynamics of Glucose Uptake in the Cytosol of COS-7 Cells by Fluorescent Nanosensors
Fehr, M., Lalonde, S., Lager, I., Wolff, M. W., & Frommer, W. B. (2003) · Journal of Biological Chemistry · doi:10.1074/jbc.m301333200 · Cited by 302
Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes
Leonardi, M., Pitchon, E. M., Bertsch, A., Renaud, P., & Mermoud, A. (2009) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.2008.01404.x · Cited by 300
Fast Responsive Crystalline Colloidal Array Photonic Crystal Glucose Sensors
Ben-Moshe, M., Alexeev, V. L., & Asher, S. A. (2006) · Analytical Chemistry · doi:10.1021/ac060643i · Cited by 292
A smartphone algorithm with inter-phone repeatability for the analysis of colorimetric tests
Yetisen, A. K., Martinez-Hurtado, J. L., Garcia-Melendrez, A., da Cruz Vasconcellos, F., & Lowe, C. R. (2014) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2014.01.077 · Cited by 285
Self-citationChu, M. X., Miyajima, K., Takahashi, D., Arakawa, T., Sano, K., Sawada, S.-I., Kudo, H., Iwasaki, Y., Akiyoshi, K., Mochizuki, M., et al. (2011) · Talanta · doi:10.1016/j.talanta.2010.10.055 · Cited by 279
Design and Analysis of Clinical Trials: Concepts and Methodologies
Chow, S., & Liu, J. (2013) · Cited by 272
Protein levels in nonstimulated and stimulated tears of normal human subjects.
Fullard, R. J., & Snyder, C. (1990) · PubMed · Cited by 250
24-Hour Monitoring of Intraocular Pressure in Glaucoma Management: A Retrospective Review
Hughes, E. H., Spry, P., & Diamond, J. P. (2003) · Journal of Glaucoma · doi:10.1097/00061198-200306000-00009 · Cited by 242
Tears as a source of biomarkers for ocular and systemic diseases
Hohenstein-Blaul, N. V. T. U., Funke, S., & Grus, F. H. (2013) · Experimental Eye Research · doi:10.1016/j.exer.2013.07.015 · Cited by 236
Laboratory findings in tear fluid analysis
Ohashi, Y., Doğru, M., & Tsubota, K. (2006) · Clinica Chimica Acta · doi:10.1016/j.cca.2005.12.035 · Cited by 236
Inflammatory Markers in the Tears of Patients with Ocular Surface Disease
Acera, A., Rocha, G., Vecino, E., Lema, I., & Durán, J. A. (2008) · Ophthalmic Research · doi:10.1159/000150445 · Cited by 229
Holographic Sensors: Three-Dimensional Analyte-Sensitive Nanostructures and Their Applications
Yetisen, A. K., Naydenova, I., da Cruz Vasconcellos, F., Blyth, J., & Lowe, C. R. (2014) · Chemical Reviews · doi:10.1021/cr500116a · Cited by 224
Self-citationA contact lens with an integrated lactate sensor
Thomas, N., Lähdesmäki, I., & Parviz, B. A. (2012) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2011.12.049 · Cited by 220
24-hour monitoring of intraocular pressure in glaucoma management: a retrospective review
Hughes, E. L., Spry, P., & Diamond, J. (2003) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2003.10.008 · Cited by 216
Biofuel cell as a power source for electronic contact lenses
Falk, M., Andoralov, V., Blum, Z., Sotres, J., Suyatin, D. B., Ruzgas, T. A., Arnebrant, T., & Shleev, S. (2012) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2012.04.030 · Cited by 215
Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus
Baca, J. T., Finegold, D. N., & Asher, S. A. (2007) · The Ocular Surface · doi:10.1016/s1542-0124(12)70094-0 · Cited by 187
Handbook of Optical Sensing of Glucose in Biological Fluids and Tissues
Tuchin, V. V. (2008) · CRC Press eBooks · doi:10.1201/9781584889755 · Cited by 186
Investigation of the human tear film proteome using multiple proteomic approaches.
Green‐Church, K. B., Nichols, K. K., Kleinholz, N. M., Zhang, L., & Nichols, J. J. (2008) · PubMed Central · Cited by 183
Kabilan, S., Marshall, A. J., Sartain, F. K., Lee, M., Hussain, A., Yang, X., Blyth, J., Karangu, N., James, K., Zeng, J., et al. (2004) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.07.005 · Cited by 168
Self-citationGaras, A., Vargha, P., & Holló, G. (2010) · Eye · doi:10.1038/eye.2010.139 · Cited by 158
初, 明. (2011) · Medical Entomology and Zoology · Cited by 157
The effect of collection technique on tear composition.
Stuchell, R. N., Feldman, J. J., Farris, R. L., & Mandel, I. D. (1984) · PubMed · Cited by 155
Tensile properties of soft contact lens materials
Tranoudis, I., & Efron, N. (2004) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2004.08.002 · Cited by 150
Proteomic analysis of human meibomian gland secretions
Tsai, P.-I. (2006) · British Journal of Ophthalmology · doi:10.1136/bjo.2005.080846 · Cited by 141
Contact lens interactions with the tear film
Mann, A., & Tighe, B. J. (2013) · Experimental Eye Research · doi:10.1016/j.exer.2013.07.013 · Cited by 138
Capacitive contact lens sensor for continuous non-invasive intraocular pressure monitoring
Chen, G.-Z., Chan, I.-S., & Lam, D. C. C. (2013) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2013.08.029 · Cited by 133
Koo, B.-S., Lee, D., Ha, H.-S., Kim, J., & Kim, C. (2005) · Journal of Proteome Research · doi:10.1021/pr0498133 · Cited by 132
Levels of lactoferrin, secretory IgA and serum albumin in the tear film of people with keratoconus
Balasubramanian, S. A., Pye, D. C., & Willcox, M. D. P. (2011) · Experimental Eye Research · doi:10.1016/j.exer.2011.12.010 · Cited by 132
Water properties of soft contact lens materials
Tranoudis, I., & Efron, N. (2004) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2004.08.003 · Cited by 129
Reusable, Robust, and Accurate Laser-Generated Photonic Nanosensor
Yetisen, A. K., Montelongo, Y., da Cruz Vasconcellos, F., Martinez-Hurtado, J. L., Neupane, S., Butt, H., Qasim, M. M., Blyth, J., Burling, K., Carmody, J. B., et al. (2014) · Nano Letters · doi:10.1021/nl5012504 · Cited by 123
Self-citationLight‐Directed Writing of Chemically Tunable Narrow‐Band Holographic Sensors
Yetisen, A. K., Butt, H., da Cruz Vasconcellos, F., Montelongo, Y., Davidson, C. A. B., Blyth, J., Chan, L., Carmody, J. B., Vignolini, S., Steiner, U., et al. (2014) · Advanced Optical Materials · doi:10.1002/adom.201300375 · Cited by 123
Self-citationMalvitte, L., Montange, T., Véjux, A., Baudouin, C., Bron, A. M., Creuzot-Garcher, C. P., & Lizard, G. (2006) · British Journal of Ophthalmology · doi:10.1136/bjo.2006.101485 · Cited by 122
The regulation of mobile medical applications
Yetisen, A. K., Martinez-Hurtado, J. L., Vasconcellos, F. D. C., Simsekler, M. C. E., Akram, M. S., & Lowe, C. R. (2014) · Lab on a Chip · doi:10.1039/c3lc51235e · Cited by 119
Self-citationLab-on-a-Chip Devices and Micro-Total Analysis Systems
Castillo-León, J., & Svendsen, W. E. (2014) · Springer eBooks · doi:10.1007/978-3-319-08687-3 · Cited by 111
Stanková, L., Riddle, M. C., Larned, J., Burry, K., Menashe, D., Hart, J., & Bigley, R. H. (1984) · Metabolism · doi:10.1016/0026-0495(84)90197-5 · Cited by 109
Electromagnetic fields (EMF): Do they play a role in children's environmental health (CEH)?
Otto, M., & von Mühlendahl, K. E. (2007) · International Journal of Hygiene and Environmental Health · doi:10.1016/j.ijheh.2007.07.007 · Cited by 109
Yang, X., Pan, X., Blyth, J., & Lowe, C. R. (2008) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2007.09.016 · Cited by 103
Self-citationHypomagnesemia and Inflammation: Clinical and Basic Aspects
Weglicki, W. B. (2012) · Annual Review of Nutrition · doi:10.1146/annurev-nutr-071811-150656 · Cited by 103
Liu, R., Meng, F., Zhang, L., Liu, A., Qin, H., Lan, X., Li, L., & Du, G. (2011) · Molecules · doi:10.3390/molecules16032084 · Cited by 103
Gehrmann, C., Persson, J., & Smeets, B. (2004) · Cited by 102
Analysis of tear glucose concentration with electrospray ionization mass spectrometry
Taormina, C. R., Baca, J. T., Asher, S. A., Grabowski, J. J., & Finegold, D. N. (2006) · Journal of the American Society for Mass Spectrometry · doi:10.1016/j.jasms.2006.10.002 · Cited by 99
Tear fluid levels of MMP-8 are elevated in ocular rosacea—treatment effect of oral doxycycline
Määttã, M., Kari, O., Tervahartiala, T. I., Peltonen, S., Kari, M., Saari, M., & Sorsa, T. A. (2006) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-005-0212-3 · Cited by 99
Mottet, B., Aptel, F., Romanet, J.-P., Hubanova, R., Pépin, J.-L., & Chiquet, C. (2013) · JAMA Ophthalmology · doi:10.1001/jamaophthalmol.2013.5297 · Cited by 94
Excretion of Urea, Sodium, Potassium and Chloride in Human Tears
Thaysen, J. H., & Thorn, N. A. (1954) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1954.178.1.160 · Cited by 93
Glucose‐sensitive holographic sensors
Kabilan, S., Blyth, J., Lee, M. C., Marshall, A. J., Hussain, A., Yang, X., & Lowe, C. R. (2004) · Journal of Molecular Recognition · doi:10.1002/jmr.663 · Cited by 91
Self-citationLacryglobin in human tears, a potential marker for cancer
Evans, V., Vockler, C., Friedlander, M. L., Walsh, B. J., & Willcox, M. D. P. (2001) · Clinical and Experimental Ophthalmology · doi:10.1046/j.1442-9071.2001.00408.x · Cited by 90
Grus, F. H., Joachim, S. C., & Pfeiffer, N. (2007) · PROTEOMICS - CLINICAL APPLICATIONS · doi:10.1002/prca.200700105 · Cited by 83
Intra- and inter-day variation of cytokines and chemokines in tears of healthy subjects
Benito, M. J., González-García, M. J., Tesón, M., García, N., Fernández, I., Calonge, M., & Enríquez‐de‐Salamanca, A. (2014) · Experimental Eye Research · doi:10.1016/j.exer.2013.12.017 · Cited by 81
Initial Clinical Testing of a Holographic Non-Invasive Contact Lens Glucose Sensor
Domschke, A., March, W. F., Kabilan, S., & Lowe, C. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.89 · Cited by 81
Self-citationTherapeutic uses of contact lenses
McDermott, M. L., & Chandler, J. (1989) · Survey of Ophthalmology · doi:10.1016/0039-6257(89)90015-5 · Cited by 80
Alterations in the Tear Proteome of Dry Eye Patients—A Matter of the Clinical Phenotype
Boehm, N., Funke, S., Wiegand, M., Wehrwein, N., Pfeiffer, N., & Grus, F. H. (2013) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.11-8751 · Cited by 80
Haeringen, N. J. V., & Glasius, E. (1977) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/bf00496763 · Cited by 75
Clinical Trial of a Noninvasive Contact Lens Glucose Sensor
March, W. F., Mueller, A., & Herbrechtsmeier, P. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.782 · Cited by 74
1968 International Solid-State Circuits Conference
(1968) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1968.1049838 · Cited by 74
Printable Surface Holograms via Laser Ablation
Vasconcellos, F. D. C., Yetisen, A. K., Montelongo, Y., Butt, H., Grigore, A., Davidson, C. A. B., Blyth, J., Monteiro, M. J., Wilkinson, T. D., & Lowe, C. R. (2014) · ACS Photonics · doi:10.1021/ph400149m · Cited by 72
Self-citationUltrasensitive Piezoresistive All‐Organic Flexible Thin Films
Laukhina, E. E., Pfattner, R., Ferreras, L. R., Galli, S., Mas‐Torrent, M., Masciocchi, N., Laukhin, V. N., Rovira, C., & Veciana, J. (2009) · Advanced Materials · doi:10.1002/adma.200902639 · Cited by 71
Pulsed laser writing of holographic nanosensors
Yetisen, A. K., Qasim, M. M., Nosheen, S., Wilkinson, T. D., & Lowe, C. R. (2014) · Journal of Materials Chemistry C · doi:10.1039/c3tc32507e · Cited by 67
Self-citationComplex Antibody Profile Changes in an Experimental Autoimmune Glaucoma Animal Model
Joachim, S. C., Grus, F. H., Kraft, D., White-Farrar, K., Barnes, G., Barbeck, M., Ghanaati, S., Cao, S., Li, B., & Wax, M. B. (2009) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.08-3144 · Cited by 67
The Fall and Rise of Tear Albumin Levels: A Multifactorial Phenomenon
Eden, G. R., Mann, A., & Tighe, B. J. (2013) · The Ocular Surface · doi:10.1016/j.jtos.2013.03.001 · Cited by 65
A microsystem-based assay for studying pollen tube guidance in plant reproduction
Yetisen, A. K., Jiang, L., Cooper, J., Qin, Y., Palanivelu, R., & Zohar, Y. (2011) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/21/5/054018 · Cited by 63
Self-citationZhou, L., Beuerman, R. W., Ang, L. P. K., Chan, C. M., Li, S. F. Y., Chew, F. T., & Tan, D. T. H. (2009) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.08-2604 · Cited by 63
The effect of an artificially elevated intraocular pressure on the central corneal curvature
Lam, A. K. C. (1997) · Ophthalmic and Physiological Optics · doi:10.1016/s0275-5408(96)00033-6 · Cited by 63
Tolerability of 24-Hour Intraocular Pressure Monitoring of a Pressure-sensitive Contact Lens
Lorenz, K., Korb, C. A., Herzog, N., Vetter, J. M., Elflein, H. M., Keilani, M. M., & Pfeiffer, N. (2013) · Journal of Glaucoma · doi:10.1097/ijg.0b013e318241b874 · Cited by 58
Fluorescent Measurement in the Non-Invasive Contact Lens Glucose Sensor
March, W., Lazzaro, D., & Rastogi, S. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.312 · Cited by 58
Laukhin, V., Sánchez, I., Moya, A., Laukhina, E., Martin, R., Ussa, F., Rovira, C., Guimera, A., Villa, R., Aguiló, J., et al. (2011) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2011.05.021 · Cited by 57
DAUM, K. M., & Hill, R. M. (1982) · PubMed · Cited by 56
Vitamin C Levels in Human Tears
Paterson, C. A., & O'Rourke, M. C. (1987) · Archives of Ophthalmology · doi:10.1001/archopht.1987.01060030096034 · Cited by 56
Computational modelling and characterisation of nanoparticle-based tuneable photonic crystal sensors
Tsangarides, C. P., Yetisen, A. K., Vasconcellos, F. D. C., Montelongo, Y., Qasim, M. M., Wilkinson, T. D., Lowe, C. R., & Butt, H. (2014) · RSC Advances · doi:10.1039/c3ra47984f · Cited by 55
Self-citationGlaucoma: diagnosis and management
Infeld, D. A., & O'shea, J. G. (1998) · Postgraduate Medical Journal · doi:10.1136/pgmj.74.878.709 · Cited by 54
Lack of Compliance With Home Blood Glucose Monitoring Predicts Hospitalization in Diabetes
Burge, M. R. (2001) · Diabetes Care · doi:10.2337/diacare.24.8.1502 · Cited by 54
Changes in Tear Protein Profile in Patients With Conjunctivochalasis
Acera, A., Suárez, T., Rodríguez-Agirretxe, I., Vecino, E., & Durán, J. A. (2010) · Cornea · doi:10.1097/ico.0b013e3181dea7d7 · Cited by 53
Proteomic analysis of dog tears for potential cancer markers
Campos, C. D. F., Cole, N., Dyk, D. D. V., Walsh, B. J., Diakos, P., de Almeida, D. E., Torrecilhas, A., Laus, J. L., & Willcox, M. D. P. (2008) · Research in Veterinary Science · doi:10.1016/j.rvsc.2007.11.006 · Cited by 52
Hubanova, R., Aptel, F., Chiquet, C., Mottet, B., & Romanet, J. (2014) · Acta Ophthalmologica · doi:10.1111/aos.12241 · Cited by 51
Mathew, M. E., & Sandhyarani, N. (2013) · Electrochimica Acta · doi:10.1016/j.electacta.2013.07.010 · Cited by 50
Eye Blink Rate as a biological marker of Mild Cognitive Impairment
Ladas, A. I., Frantzidιs, C. A., Bamidis, P. D., & Vivas, A. B. (2013) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2013.07.010 · Cited by 49
In-eye performance of soft contact lenses made from different materials
Tranoudis, I., & Efron, N. (2004) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2004.02.004 · Cited by 48
Naroo, S. A. (2016) · Cited by 48
Frauscher, B., Ehrmann, L., & Högl, B. (2012) · Sleep Medicine · doi:10.1016/j.sleep.2012.09.028 · Cited by 46
Proteomics in Autoimmune Thyroid Eye Disease
Okrojek, R., Grus, F. H., Matheis, N., & Kahaly, G. J. (2009) · Hormone and Metabolic Research · doi:10.1055/s-0029-1214413 · Cited by 45
Immunoglobulin concentrations in human tears in ocular diseases.
Sen, D., & Gs, S. (1979) · British Journal of Ophthalmology · doi:10.1136/bjo.63.5.297 · Cited by 45
Patent protection and licensing in microfluidics
Yetisen, A. K., & Volpatti, L. R. (2014) · Lab on a Chip · doi:10.1039/c4lc00399c · Cited by 40
Self-citationCompositional Profiling and Biomarker Identification of the Tear Film
Jacob, J. T., & Ham, B. M. (2008) · The Ocular Surface · doi:10.1016/s1542-0124(12)70178-7 · Cited by 37
A Wireless Health Monitoring System based on Android Operating System
Lou, D., Chen, X., Zhao, Z., Xuan, Y., Xu, Z., Jin, H., Guo, X., & Fang, Z. (2013) · IERI Procedia · doi:10.1016/j.ieri.2013.11.030 · Cited by 35
Malik, B. H., & Coté, G. L. (2010) · Journal of Biomedical Optics · doi:10.1117/1.3447923 · Cited by 35
Surface enhanced Raman scattering of herpes simplex virus in tear film
Reyes-Goddard, J., Barr, H., & Stone, N. (2008) · Photodiagnosis and Photodynamic Therapy · doi:10.1016/j.pdpdt.2008.01.002 · Cited by 34
Computational modelling of a graphene Fresnel lens on different substrates
Deng, S., Yetisen, A. K., Jiang, K. S., & Butt, H. (2014) · RSC Advances · doi:10.1039/c4ra03991b · Cited by 34
Self-citationAndoralov, V., Shleev, S., Arnebrant, T., & Ruzgas, T. A. (2013) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-013-6756-x · Cited by 34
Corneal thickness after overnight wear of an intraocular pressure fluctuation contact lens sensor
Freiberg, F. J., Lindell, J., Thederan, L., Leippi, S., Shen, Y., & Klink, T. (2012) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.2012.02495.x · Cited by 33
Powering electronic contact lenses: current achievements, challenges, and perspectives
Blum, Z., Pankratov, D., & Shleev, S. (2014) · Expert Review of Ophthalmology · doi:10.1586/17469899.2014.922873 · Cited by 33
Cerebrospinal fluid lactate and pyruvate concentrations and their ratio
Zhang, W., & Natowicz, M. R. (2012) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2012.11.008 · Cited by 31
Fibre optics sensors in tear electrolyte analysis: Towards a novel point of care potassium sensor
Harvey, D., Hayes, N. W., & Tighe, B. J. (2012) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2012.02.004 · Cited by 28
Torgrip, R. J. O., Lindberg, J., Linder, M., Karlberg, B., Jacobsson, S. P., Kolmert, J., Gustafsson, I., & Schuppe‐Koistinen, I. (2006) · Metabolomics · doi:10.1007/s11306-005-0013-z · Cited by 27
Singh, D., & Saxena, R. (2012) · Expert Review of Ophthalmology · doi:10.1586/eop.12.35 · Cited by 26
Changes in the Tear Protein Patterns of Diabetic Patients Using Two-Dimensional Electrophoresis
Herber, S., Grus, F. H., Sabuncuo, P., & Augustin, A. J. (2002) · Advances in experimental medicine and biology · doi:10.1007/978-1-4615-0717-8_88 · Cited by 25
Contact Lens Optics and Lens Design
(2006) · Elsevier eBooks · doi:10.1016/b978-0-7506-8879-6.x5001-5 · Cited by 24
Enhanced reflection from inverse tapered nanocone arrays
Kong, X., Butt, H., Yetisen, A. K., Kangwanwatana, C., Montelongo, Y., Deng, S., Vasconcellos, F. D. C., Qasim, M. M., Wilkinson, T. D., & Dai, Q. (2014) · Applied Physics Letters · doi:10.1063/1.4892580 · Cited by 24
Self-citationFarkas, Á., Vámos, R., Bajor, T., Müllner, N., Lazar, A. C., & Hrabá, A. (2003) · Experimental Eye Research · doi:10.1016/s0014-4835(02)00276-2 · Cited by 21
Nanomedical Device and Systems Design
Boehm, F. (2013) · CRC Press eBooks · doi:10.1201/b15626 · Cited by 19
Antibodies in human tears during and after infection
Friedman, M. G. (1990) · Survey of Ophthalmology · doi:10.1016/0039-6257(90)90070-c · Cited by 18
Patel, J. N., Gray, B. L., Kaminska, B., & Gates, B. D. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500503 · Cited by 18
van Delft, J. L., MEIJER, F., van Best, J. A., & Haeringen, N. J. V. (1997) · Mediators of Inflammation · doi:10.1080/09629359791532 · Cited by 18
Auvray, P., Rousseau, L., Lissorgues, G., Soulier, F., Potin, O., Bernard, S., de Dieuleveult, F., Scorsone, E., Bergonzo, P., Chicaud, L., et al. (2012) · IRBM · doi:10.1016/j.irbm.2012.01.002 · Cited by 16
HIV-specific IgA antibodies in tears of children with aids or at risk of AIDS
Renom, G., Bouquety, J. C., Lanckriet, C., Georges, A. J., Siopathis, M. R., & Martin, P. M. V. (1990) · Research in Virology · doi:10.1016/0923-2516(90)90007-6 · Cited by 16
Time series analysis of age related cataract hospitalizations and phacoemulsification
Leong, A. M., Crighton, E., Moineddin, R., Mamdani, M., & Upshur, R. E. G. (2006) · BMC Ophthalmology · doi:10.1186/1471-2415-6-2 · Cited by 15
Quantitative Analyse der Tränenproteinmuster bei weichen Kontaktlinsen - Klinische Studie1
Grus, F. H., Sabuncuo, P., & Augustin, A. J. (2001) · Klinische Monatsblätter für Augenheilkunde · doi:10.1055/s-2001-14920 · Cited by 14
Contact lens optics and lens design
Whittle, J. P. (2007) · Eye · doi:10.1038/sj.eye.6702607 · Cited by 13
Candiano, G., Musante, L., Zennaro, C., Bruschi, M., Carraro, M., Artero, M. L., Gusmano, R., Ginevri, F., Perfumo, F., & Ghiggeri, G. M. (2001) · Electrophoresis · doi:10.1002/1522-2683(200105)22:9 · Cited by 11
Detection of Antichlamydial Antibodies in Tears
Haller, E.-M., Auer‐Grumbach, P., Stünzner, D., Kessler, H. H., Pierer, K., Müllner, K., & Zenz, H. (1997) · Ophthalmology · doi:10.1016/s0161-6420(97)30350-9 · Cited by 10
A personal perspective on the history of contact lenses
GOODLAW, E. I. (2000) · International Contact Lens Clinic · doi:10.1016/s0892-8967(02)00085-8 · Cited by 10
March, W. F. (2002) · Diabetes Technology & Therapeutics · doi:10.1089/15209150252924085 · Cited by 6
Fink, B. A., Mitchell, G. L., & Hill, R. M. (2006) · Optometry and Vision Science · doi:10.1097/01.opx.0000232806.06677.97 · Cited by 5
[Do you wear contact lenses?].
Dm, W., & Pratt-Johnson, J. A. (1965) · PubMed · Cited by 1
Flagged first, then by citations
Research works printed but not linked: 22
- Novartis International AG Press Release “Novartis to license Google “smart lens” technology” 2014 1.
- Google Press Release “What is Google doing with a smart contact lens?”2014.
- Yan J., 2011, Conf. Proc. IEEE Eng. Med. Biol. Soc
- Brusini P., 2014, Atlas of Glaucoma
- Mochizuki M., 2004, Yamagata Med. J., 22, 9
- Daum K. M., 1982, Invest. Ophthalmol. Vis. Sci., 22, 509
- Green‐Church K. B., 2008, Mol. Vis., 14, 456
- Solomon A., 2001, Invest. Ophthalmol. Vis. Sci., 42, 2283
- Grus F. H., 2002, BMC Ophthalmol., 6, 2
- Tsuji F., 2012, Metabolomics, 2, 1
- McEwen W. K., 1969, Eye, 3, 34
- Chen R., 1996, J. Capillary Electrophor., 3, 243
- Sfuchell R. N., 1984, Invest. Ophthalmol. Vis. Sci., 25, 374
- Fullard R. J., 1990, Invest. Ophthalmol. Vis. Sci., 31, 1119
- Geddes C. D., 2006, Topics in Fluoresence Spectroscopy: Glucose Sensing
- Bennett E. S., 2005, Clinical Contact Lens Practice
- Tseng C.‐K., 2012, 2012 IEEE Sensors
- Domschke A., 2004, Proc. IEEE Sensors 2004
- Newlove D., 1981, Journal of the British Contact Lens Association
- Moezzi A. M., 2014, J. Optom, 1
- Barr J. T., 2005, Clinical Contact Lens Practice
- Schifrin L. G., 1984, The Contact Lens Industry: Structure, Competition and Public Policy (Health Technology Case Study 31)
Cited by
Sorted by citations
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373
Advanced Carbon for Flexible and Wearable Electronics
Wang, C., Xia, K., Wang, H., Liang, X., Yin, Z., & Zhang, Y. (2018) · Advanced Materials · doi:10.1002/adma.201801072 · Cited by 1,312
Wearable smart sensor systems integrated on soft contact lenses for wireless ocular diagnostics
Kim, J., Kim, M., Lee, M.-S., Kim, K., Ji, S., Kim, Y.-T., Park, J., Na, K., Bae, K.-H., Kyun Kim, H., et al. (2017) · Nature Communications · doi:10.1038/ncomms14997 · Cited by 951
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 915
Enzyme‐Based Glucose Sensor: From Invasive to Wearable Device
Lee, H., Hong, Y. J., Baik, S., Hyeon, T., & Kim, D. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201701150 · Cited by 778
Metadata
- Authors
- Farandos, N. M., Yetisen, A. K., Monteiro, M. J., Lowe, C. R., & Yun, S. H.
- Affiliations
- Imperial College London (GB)
- University of Cambridge (GB)
- The University of Queensland (AU)
- Harvard University (US)
- Corresponding author
- Yetisen, Ali K.
- Venue
- Advanced Healthcare Materials
- Publisher
- Wiley
- Published
- 2014-11-17
- Volume
- 4
- Issue
- 6
- Pages
- 792-810
- ISSN
- 2192-2640 (ISSN-L)
- 2192-2659
- OpenAlex
- W2055793978
- Cited by
- 466
- References
- 133
- Citations by year
- 2026: 16 · 2025: 29 · 2024: 42 · 2023: 48 · 2022: 47 · 2021: 51 · 2020: 52 · 2019: 50 · 2018: 44 · 2017: 37 · 2016: 24 · 2015: 18 · 2014: 8
- FWCI
- 2.48
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Advanced Healthcare Materials (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Ophthalmology › Ocular Surface and Contact Lens
- Keywords
- contact lens sensor, ocular diagnostics, glucose monitoring, tear glucose, glaucoma diagnosis, intraocular pressure, minimally invasive diagnosis, point-of-care monitoring, wireless power transfer, smartphone integration, contact lens manufacturing, FDA regulation, ocular drug delivery
- MeSH terms
- Contact Lenses, Electronics, Glaucoma, Humans, Intraocular Pressure, Biosensing Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review