The DOI matches its Crossref record; 39 of the 39 references it cites were found in a registry or an index; 16 of the 42 sources its claims cite were checked for support.
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Open accessAbstract
Wearable sensor technologies are essential to the realization of personalized medicine through continuously monitoring an individual's state of health. Sampling human sweat, which is rich in physiological information, could enable non-invasive monitoring. Previously reported sweat-based and other non-invasive biosensors either can only monitor a single analyte at a time or lack on-site signal processing circuitry and sensor calibration mechanisms for accurate analysis of the physiological state. Given the complexity of sweat secretion, simultaneous and multiplexed screening of target biomarkers is critical and requires full system integration to ensure the accuracy of measurements. Here we present a mechanically flexible and fully integrated (that is, no external analysis is needed) sensor array for multiplexed in situ perspiration analysis, which simultaneously and selectively measures sweat metabolites (such as glucose and lactate) and electrolytes (such as sodium and potassium ions), as well as the skin temperature (to calibrate the response of the sensors). Our work bridges the technological gap between signal transduction, conditioning (amplification and filtering), processing and wireless transmission in wearable biosensors by merging plastic-based sensors that interface with the skin with silicon integrated circuits consolidated on a flexible circuit board for complex signal processing. This application could not have been realized using either of these technologies alone owing to their respective inherent limitations. The wearable system is used to measure the detailed sweat profile of human subjects engaged in prolonged indoor and outdoor physical activities, and to make a real-time assessment of the physiological state of the subjects. This platform enables a wide range of personalized diagnostic and physiological monitoring applications.
Overview
AI snapshot
Written from the paper’s full text.
TL;DR
A flexible wearable sensing array with full on-board integration was built to measure sweat glucose, lactate, sodium, potassium, and skin temperature at the same time, with signal conditioning, processing, and Bluetooth transmission. Human exercise trials showed it could track sweat profiles in real time and detect dehydration-related electrolyte changes.
Key findings
- The platform merged plastic-based sensors with silicon integrated circuits on a flexible printed circuit board, enabling multiplexed sweat analysis without external analysis; five sensors measured metabolites, electrolytes, and skin temperature.
- In vitro sensitivities were 2.35 nA μM−1 for glucose over 0–200 μM, 220 nA mM−1 for lactate over 0–30 mM, 64.2 mV per decade for Na+ over 10–160 mM, 61.3 mV per decade for K+ over 1–32 mM, and about 0.18% per degree Celsius for temperature over 20–40 °C.
- Sensors showed negligible interference from non-target species, and temperature compensation corrected enzymatic sensor overestimation; long-term tests indicated stable sensitivities for at least four weeks, with Na+/K+ at about 62.5 and 59.5 mV per decade in ambient conditions and glucose/lactate within 5% of original values when stored at 4 °C.
- On-body trials with 26 healthy subjects (4 females, 22 males, aged 20–40) tracked sweat during cycling and outdoor running; dehydration trials without water intake showed sweat [Na+] increases after about 80 min and roughly 2.5% body-weight loss, while euhydration with 150 ml water every 5 min kept [Na+] and [K+] stable after initial changes.
Methods
- Design: a five-sensor array on PET with 3-mm electrodes, including enzymatic amperometric glucose/lactate sensors, ion-selective Na+/K+ sensors, and a resistive temperature sensor, integrated with an FPCB containing analogue conditioning, an ATmega328P microcontroller, and Bluetooth.
- Data: in vitro characterization in analyte solutions and artificial sweat; on-body testing in constant-load cycling (14 volunteers: 4 females, 10 males), graded cycling (7 males), and outdoor running (12 males), plus ex situ sweat comparisons.
- Analysis: chronoamperometry for glucose/lactate, open-circuit potentiometry for Na+/K+, resistance for temperature, with on-board calibration/compensation and wireless data relay to a mobile app.
Limitations
- Stated: glucose and lactate sensor responses degraded after about two hours of use because of reduced enzyme activity, though the sensor array can be replaced via the sensor–FPCB interface.
- Stated: wearing the plastic-substrate device raised local temperature, which may slightly bias actual skin-temperature readings, but on-board calibration was reported to prevent impact on electrolyte and metabolite measurements; in vitro temperature-compensation tests also had noise from uneven solution heating.
- Evident: the human sample was small and heavily male-skewed (26 total; 4 females, 22 males; graded and outdoor running trials included only males), and the paper does not report long-term continuous use beyond the exercise sessions for the enzymatic sensors.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
36 Own result · 18 No basis given · 13 No full text of the sources · 10 Own method · 10 Own interpretation · 8 Partly supported · 2 Supported · 1 Supported so far · 1 Couldn’t confirm · 1 Not checked yet · accuracy not yet measured
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“skin temperature is clinically informative of a variety of diseases and skin injuries such as pressure ulcers23,24.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [23] Webb, R. C. et al. Ultrathin conformal devices for precise and continuous thermal characterization of human skin. Nature Mater. 12, 938–944 (2013) sourceExperimental — accuracy not yet measured
The Abstract says that skin thermometry can provide clinically relevant information about cardiovascular health, malignancy and other aspects of physiology. It does not link skin temperature to pressure ulcers or establish its clinical value for assessing skin injuries; its mention of wound healing concerns local heating as a possible intervention. (LLM, conf=0.86)
- Not checked — no full text of the sourcecites [24] Sprigle, S., Linden, M., McKenna, D., Davis, K. & Riordan, B. Clinical skin temperature measurement to predict incipient pressure ulcers. Adv. Skin Wound Care 14, 133–137 (2001) source
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The decreases in concentration of lactate and glucose in sweat are expected, owing to the dilution effect caused by an increase in sweat rate, which is visually observed as exercise continues13.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) sourceExperimental — accuracy not yet measured
The Introduction (page not available in our copy) explains that increasing sweat flow can dilute biomarkers whose transport into sweat is slower than the flow. It does not establish that lactate and glucose concentrations both decrease as exercise continues; instead, it says high-exertion tests can show a spike in sweat lactate associated with sweat rate. (LLM, conf=0.83)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The relatively stable behaviour of [K+] is explained by its passive ion partitioning mechanism13.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) sourceExperimental — accuracy not yet measured
The section headed “Passive ion partitioning: Potassium (mM)” reports that sweat K+ concentration is independent of sweat rate, supporting the description of relatively stable behavior. But it says the precise partitioning mechanism has not been fully confirmed, so it does not establish passive partitioning as the explanation. No page number is available in our copy. (LLM, conf=0.86)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Assuming a single-centred flow model13, the best-case sampling interval can be calculated to be less than 1 min, based on the sweat rate (~3–4 mg min−1 cm−2)39 and the pad size (1.5 cm × 2 cm × 50 μm).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
What the sources say
- Supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) source
The citation is attached to the single-centred flow model, not the sweat rate or pad dimensions. Section C, “Expectations for biomarker sampling intervals,” defines that model as the best-case limit and gives the refill-time calculation used by the citing authors. (LLM, conf=0.94)
- Partly supportedcites [39] Vimieiro-Gomes, A. C. Comparison of sweat rate during graded exercise and the local rate induced by pilocarpine. Braz. J. Med. Biol. Res. 38, 1133–1139 (2005) sourceExperimental — accuracy not yet measured
The Abstract (page unavailable) reports local exercise-induced sweat rates of 2.1 ± 1.4 and 2.7 ± 1.2 mg cm−2 min−1 in nine young men. That supports using a sweat rate on the order of 3 for an estimate, but not the cited value of approximately 3–4 as a reported rate. The citation is attached to the sweat-rate input, not to the flow model, pad dimensions, or calculated interval. (LLM, conf=0.84)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- The authors state this as an assumption: Sweat flow can be modeled as a single-centred flow.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“However, lactate concentration becomes relatively stable after 1,100 s, indicating the stabilization of physiological responses to continuous, sub-maximal constant-exercise power output22.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [22] Derbyshire, P. J., Barr, H., Davis, F. & Higson, S. P. Lactate in human sweat: a critical review of research to the present day. J. Physiol. Sci. 62, 429–440 (2012) sourceExperimental — accuracy not yet measured
The review’s “Sweat lactate variation with exercise intensity” section (page unavailable) reports that sweat lactate decayed and then reached a plateau during incremental cycling. It does not give 1,100 s as the time of stabilization or interpret the plateau as stabilization of physiological responses during continuous, sub-maximal constant-power exercise. (LLM, conf=0.84)
hedgeddirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“Owing to Faraday and capacitive currents38, the responses of both sensors showed drift initially but stabilized within 1 min of the data recording.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
What the sources say
- Could not verifycites [38] Grieshaber, D., MacKenzie, R., Vörös, J. & Reimhult, E. Electrochemical biosensors—sensor principles and architectures. Sensors 8, 1400–1458 (2008) sourceExperimental — accuracy not yet measured
The Electrochemical Detection Techniques section (page unavailable) describes time-dependent current in chronoamperometry, but attributes current changes to the diffusion layer. Searches for Faradaic and capacitive currents, sensor drift, and one-minute stabilization found no passage connecting those currents to initial drift in glucose and lactate sensor responses. (LLM, conf=0.73)
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The independent and selective operation of individual sensors is preserved during multiplexed measurements by employing highly specific surface chemistries and by electrically decoupling the operating points of each sensor's interface.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
In the paper
“By using poly(3,4-ethylenedioxythiophene) polystyrene sulfonate (PEDOT:PSS) as an ion-to-electron transducer in the ISEs and carbon nanotubes in the PVB reference membrane25, robust potentiometric sensors (with voltage output) can be obtained for long-term continuous measurements with negligible voltage drift.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [25] Zhu, J., Qin, Y. & Zhang, Y. Preparation of all solid-state potentiometric ion sensors with polymer-CNT composites. Electrochem. Commun. 11, 1684–1687 (2009) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“These enzymatic sensors autonomously generate current signals proportional to the abundance of the corresponding metabolites between the working electrode and the Ag/AgCl electrode.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The measurement of Na+ and K+ levels is facilitated through the use of ion-selective electrodes (ISEs), coupled with a polyvinyl butyral (PVB)-coated reference electrode to maintain a stable potential in solutions with different ionic strengths (Extended Data Fig. 2b–d).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Probably not a substantial claim. 22
- It may state two separate claims as one. 30
- The quote only partly states this claim. 1
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Wearable sensor technologies are essential to the realization of personalized medicine through continuously monitoring an individual's state of health1–12.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [1] Kim, D. H. et al. Epidermal electronics. Science 333, 838–843 (2011) sourceExperimental — accuracy not yet measured
The paper’s Abstract describes skin-integrated sensors for physiological-status monitoring, and its ECG, EMG and EEG demonstrations show that they can measure an individual’s signals. It does not establish that wearable sensors are essential to personalized medicine or demonstrate continuous health monitoring; it notes that long-term use requires further work (Abstract; page unavailable). (LLM, conf=0.84)
- Not checked — no full text of the sourcecites [2] Lipomi, D. J. et al. Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes. Nature Nanotechnol. 6, 788–792 (2011) source
- Not checked — no full text of the sourcecites [3] Takei, K. et al. Nanowire active-matrix circuitry for low-voltage macroscale artificial skin. Nature Mater. 9, 821–826 (2010) source
- Not checked — no full text of the sourcecites [4] Wu, W. et al. Piezoelectricity of single-atomic-layer MoS2 for energy conversion and piezotronics. Nature 514, 470–474 (2014) source
- Partly supportedcites [5] McAlpine, M. C., Ahmad, H., Wang, D. & Heath, J. R. Highly ordered nanowire arrays on plastic substrates for ultrasensitive flexible chemical sensors. Nature Mater. 6, 379–384 (2007) sourceExperimental — accuracy not yet measured
The Abstract (p. 1) says flexible chemical and biological sensors could enable wearable monitoring systems. It does not establish that wearable sensors are essential to personalized medicine or that they continuously monitor an individual's health; those claims are stronger and more specific than the paper's proposed application. (LLM, conf=0.86)
- Not checked — no full text of the sourcecites [6] Kaltenbrunner, M. et al. An ultra-lightweight design for imperceptible plastic electronics. Nature 499, 458–463 (2013) source
- Partly supportedcites [7] Xu, S. et al. Soft microfluidic assemblies of sensors, circuits, and radios for the skin. Science 344, 70–74 (2014) sourceExperimental — accuracy not yet measured
The Abstract says skin-mounted sensors can provide clinical-quality health monitoring for continuous use and that continuous physiological monitoring is needed to treat many medical conditions. It does not say wearable sensors are essential to the realization of personalized medicine; that part overstates what the paper supports. (LLM, conf=0.85)
- Not checked — no full text of the sourcecites [8] Wang, C. et al. User-interactive electronic skin for instantaneous pressure visualization. Nature Mater. 12, 899–904 (2013) source
- Not checked — no full text of the sourcecites [9] Liang, J., Li, L., Niu, X., Yu, Z. & Pei, Q. Elastomeric polymer light-emitting devices and displays. Nature Photon. 7, 817–824 (2013) source
- Not checked — no full text of the sourcecites [10] Kim, D. H., Ghaffari, R. & Rogers, J. A. Flexible and stretchable electronics for biointegrated devices. Annu. Rev. Biomed. Eng. 14, 113–128 (2012) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence.
- Partly supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“Previously reported sweat-based and other non-invasive biosensors either can only monitor a single analyte at a time or lack on-site signal processing circuitry and sensor calibration mechanisms for accurate analysis of the physiological state14–18.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Supportedcites [14] Schazmann, B. et al. A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration. Anal. Methods 2, 342–348 (2010) source
The Abstract describes this device as a sodium-selective sweat sensor, and the Conclusions discuss a sensor array for other ions as a future possibility. The paper therefore supports its use as an example of a previously reported single-analyte sweat sensor; the citation need not establish that this device also lacked on-site processing or calibration. (LLM, conf=0.90)
- Not checked — no full text of the sourcecites [15] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013) source
- Could not verifycites [16] Coyle, S. et al. BIOTEX—biosensing textiles for personalised healthcare management. IEEE Trans. Inf. Technol. Biomed. 14, 364–370 (2010) sourceExperimental — accuracy not yet measured
The Introduction describes limitations of earlier sweat-collection methods, but does not say that previously reported biosensors were restricted to one analyte or lacked on-site processing and calibration. It instead describes BIOTEX’s multiparametric patch; the sensor and control-electronics sections discuss calibration before trials and signal processing, without establishing the claimed either-or limitation of prior devices. (LLM, conf=0.76)
- Not checked — no full text of the sourcecites [17] Bandodkar, A. J. et al. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 54, 603–609 (2014) source
- Not checked — no full text of the sourcecites [18] Rose, D. P. et al. Adhesive RFID sensor patch for monitoring of sweat electrolytes. IEEE Trans. Biomed. Eng. 62, 1457–1465 (2015) source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“sweat lactate can potentially serve as a sensitive marker of pressure ischaemia22;”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Supportedcites [22] Derbyshire, P. J., Barr, H., Davis, F. & Higson, S. P. Lactate in human sweat: a critical review of research to the present day. J. Physiol. Sci. 62, 429–440 (2012) source
The “Potential applications for sweat lactate measurements” section (page not provided) calls sweat lactate “potentially a very useful early warning indicator of pressure ischemia.” This supports the claim’s tentative wording. (LLM, conf=0.93)
hedgedcorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Sampling human sweat, which is rich in physiological information13, could enable non-invasive monitoring.”
- Also stated in: § Abstract · plain
“Measurements of human sweat could enable such insight, because it contains physiologically and metabolically rich information that can be retrieved non-invasively13.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Supportedcites [13] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015) source
The Introduction describes sweat as a source of biomarkers and blood solutes accessible non-invasively. Section V explains that many blood solutes enter sweat quickly enough for continuous physiological monitoring, supporting the claim that sweat sampling could provide useful physiological information. (LLM, conf=0.92)
plainconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Also stated in: § Abstract · plain
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The two-electrode system is a common strategy for low-current electrochemical sensing34,35.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Design of electrochemical sensors
What the sources say
- Not checked — no full text of the sourcecites [34] Kudo, H. et al. A flexible and wearable glucose sensor based on functional polymers with Soft-MEMS techniques. Biosens. Bioelectron. 22, 558–562 (2006) source
- Partly supportedcites [35] Bandodkar, A. J. et al. Tattoo-based noninvasive glucose monitoring: a proof-of-concept study. Anal. Chem. 87, 394–398 (2015) sourceExperimental — accuracy not yet measured
The Results and Discussion section describes this tattoo glucose sensor using an Ag/AgCl electrode as a combined counter/reference electrode alongside a working electrode (page unavailable). It does not say that this two-electrode configuration is a common or standard strategy for low-current electrochemical sensing; its discussion of lower current density concerns glucose extraction by reverse iontophoresis, not the sensing configuration. (LLM, conf=0.85)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“These trends are probably caused by increased blood serum [Na+] and [K+] with dehydration and increased neural stimulation, a conclusion in agreement with previous ex situ sweat analyses33.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Partly supportedcites [33] Morgan, R. M., Patterson, M. J. & Nimmo, M. A. Acute effects of dehydration on sweat composition in men during prolonged exercise in the heat. Acta Physiol. Scand. 182, 37–43 (2004) sourceExperimental — accuracy not yet measured
The Results report higher serum Na+ and K+ during dehydration, and the Discussion proposes greater extracellular Na+ and sympathetic nervous activity as possible explanations for higher sweat Na+. The Discussion’s concluding mechanism emphasizes Na+ and Cl−, aldosterone, and sympathetic activity; it does not identify increased serum K+ as a specific cause. The Experimental procedures describe sweat collected from a forearm patch and analyzed after collection, consistent with ex situ analysis. Page numbers are unavailable in our copy. (LLM, conf=0.86)
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Partly supported
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Here we present a mechanically flexible and fully integrated (that is, no external analysis is needed) sensor array for multiplexed in situ perspiration analysis, which simultaneously and selectively measures sweat metabolites (such as glucose and lactate) and electrolytes (such as sodium and potassium ions), as well as the skin temperature (to calibrate the response of the sensors).”
- Also stated in: § Abstract · plain
“Here we present a wearable flexible integrated sensing array (FISA) for simultaneous and selective screening of a panel of biomarkers in sweat (Fig. 1a).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
- Also stated in: § Abstract · plain
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The wearable system is used to measure the detailed sweat profile of human subjects engaged in prolonged indoor and outdoor physical activities, and to make a real-time assessment of the physiological state of the subjects.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The use of Prussian blue dye as a mediator minimizes the reduction potentials to approximately 0 V (versus Ag/AgCl) (Extended Data Fig. 2a), and thus eliminates the need for an external power source to activate the sensors.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The pad could absorb about 10 μl of sweat, which was sufficient to provide stable sensor readings.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
hedgeddirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A claim is checked only against the cited paper’s full text, never its abstract, and none is available for its sources.
In the paper
“Sweat [Na+] increases and [K+] decreases in the beginning of perspiration, in line with the previous ex situ studies from the collected sweat samples28,29.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [28] Cage, G. W., Wolfe, S. M., Thompson, R. H. & Gordon, R. S. Jr. Effects of water intake on composition of thermal sweat in normal human volunteers. J. Appl. Physiol. 29, 687–690 (1970) source
- Not checked — no full text of the sourcecites [29] Patterson, M. J., Galloway, S. D. R. & Nimmo, M. A. Variations in regional sweat composition in normal human males. Exp. Physiol. 85, 869–875 (2000) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Importantly, the entire system is mechanically flexible, thus delivering a practical wearable sensor technology that can be used for prolonged indoor and outdoor physical activities.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“This conditioning process was important to minimize the potential drift further.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Preparation of Na+ and K+ selective sensors
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The quote only partly states this claim.
In the paper
“No apparent difference is observed for glucose concentration at different power output settings, a finding consistent with the response of blood glucose to graded, short-term exercise30.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [30] Messonnier, L. A. et al. Lactate kinetics at the lactate threshold in trained and untrained men. J. Appl. Physiol. 114, 1593–1602 (2013) source
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Extended Data Fig. 6a–d show that Na+ and K+ sensors had a relative standard deviation of ~1% in sensitivity while glucose and lactate sensors had a relative standard deviation of ~5% in sensitivity.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Although temperature has a minimal effect on the potentiometric sensors, it greatly influences the performance of the enzymatic sensors.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“After the power is raised, the sweat rate visibly increases, followed by a sharp increase in skin temperature and sweat [Na+] as well as a slight increase in [K+] (in three of the seven subjects, [K+] remained stable).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirectn = 3- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Parylene is used as an insulating layer to ensure reliable sensor reading by preventing electrical contact of the metal lines with skin and sweat.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Both [Na+] and [K+] stabilize as the cycling continues.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Figure 2h illustrates that with the increase of temperature, the uncompensated sensor readouts can lead to substantial overestimation of the actual concentration of the given glucose and lactate solutions; however, the temperature compensation allows for accurate and consistent readings.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Sweat analyte levels on the wrist follow similar trends but with concentrations different from those obtained at the forehead (Extended Data Fig. 9).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“During the exercise, the heart rate, oxygen consumption (VO2), and pulmonary minute ventilation were measured using external monitoring instruments, and were found to increase proportionally with increasing power output as shown in Fig. 3c.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“On the other hand, a substantial increase in sweat [Na+] and a smaller increase in sweat [K+] (no clear increase in [K+] was observed in two out of six subjects) were observed in dehydration trials (without water intake) after 80 min when subjects had lost a large amount of water (~2.5% of body weight) (Fig. 4d and e).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirectn = 2- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“With continued perspiration, the skin temperature rises at about 400 s because of muscle heat conductance to skin and then remains stable, while the concentration of both lactate and glucose in sweat decrease gradually.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“A linear relationship between current and analyte concentrations with sensitivities of 2.35 nA μM−1 for glucose sensors and 220 nA mM−1 for lactate sensors was observed.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaincorrelational- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The sensor–FPCB interface allows for convenient replacement of the fresh sensor arrays for subsequent use.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“With the cessation of exercise, these physiological responses decrease and then remain stable.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The sensors could be used without pre-conditioning (with a small drift of ~2–3 mV h−1).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Preparation of Na+ and K+ selective sensors
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The current sensing signal paths were capable of measuring current levels as low as 1 nA, which was much lower than the minimum signal in our measurements (tens of nanoamperes).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > Signal conditioning, processing and wireless transmission circuit design
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“An increase in temperature was observed when the ‘smart headband’ or ‘smart wristband’ was worn owing to the use of the plastic substrate on skin.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This platform enables a wide range of personalized diagnostic and physiological monitoring applications.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence.
In the paper
“For example, excessive loss of sodium and potassium in sweat could result in hyponatremia, hypokalemia, muscle cramps or dehydration20;”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [20] Speedy, D. B., Noakes, T. D. & Schneider, C. Exercise-associated hyponatremia: a review. Emerg. Med. 13, 17–27 (2001) source
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may state two separate claims as one.
In the paper
“For these applications, the sample collection and analysis are performed separately, failing to provide a real-time profile of sweat content secretion, while requiring extensive laboratory analysis using bulky instrumentation19.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [19] Boysen, T. C., Yanagawa, S., Sato, F. & Sato, K. A modified anaerobic method of sweat collection. J. Appl. Physiol. 56, 1302–1307 (1984) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Given the complexity of sweat secretion, simultaneous and multiplexed screening of target biomarkers is critical and requires full system integration to ensure the accuracy of measurements.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may state two separate claims as one.
- Rests on an assumption: Sweat secretion involves multivariate mechanisms and is complex.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Additionally, skin temperature measurements are needed to compensate for and eliminate the influence of temperature variation in the readings of the chemical sensors through a built-in signal processor.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: fails — The paper may give no reason why it holds.
Probably not a substantial claim. The paper may give no reason why it holds.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This approach decouples the stringent mechanical requirements at the sensor level and electrical requirements at the signal conditioning, processing and transmission levels, and at the same time exploits the strengths of the underlying technologies.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The convection and non-uniform distributions of solution temperature could result in noticeable noise in the signal measurements.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The characterization of the sensors
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This platform could be exploited or reconfigured for in situ analyses of other biomarkers within sweat and other human fluid samples to facilitate personalized and real-time physiological and clinical investigations.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“During on-body tests, the newly generated sweat would refill the pad and ‘rinse away’ the old sweat.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
hedgeddirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“On the other hand, the silicon integrated-circuit technology does not provide sufficiently large active areas nor the intimate skin contact required to achieve stable and sensitive on-body measurements.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim.
In the paper
“In this case, because the subject had a lower sweat rate at the wrist29, the sensors were activated at a later time.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [29] Patterson, M. J., Galloway, S. D. R. & Nimmo, M. A. Variations in regional sweat composition in normal human males. Exp. Physiol. 85, 869–875 (2000) source
plaindirectSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“Monitoring hydration status is of the utmost importance to athletes because fluid deficit impairs endurance performance and increases carbohydrate reliance32.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
What the sources say
- Not checked — no full text of the sourcecites [32] Barr, S. I. Effects of dehydration on exercise performance. Can. J. Appl. Physiol. 24, 164–172 (1999) source
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“We believe that this wearable platform may enable new fundamental physiology studies through further on-body evaluation.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: fails — It may be too vague to check against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“To ensure the fidelity of sensor readings, the data collection of each channel took place when a sufficient sweat sample was present, as shown by stabilization of sensor readings within the physiologically relevant range (see Methods).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > Abstract
plaindirect- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“The power output was calibrated and monitored through the ergometer.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“All subjects gave written, informed consent before participation in the study.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“The on-body evaluation of the FISA was performed in compliance with the protocol that was approved by the institutional review board at the University of California, Berkeley (CPHS 2014-08-6636).”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“Although this may result in a small error in measuring the actual skin temperature, it should be noted that this does not have an impact on the measurement of the electrolytes and metabolites, owing to the on-board temperature calibration.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > The setup of FISA for on-body testing
hedgedconditional- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
It may state two separate claims as one.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“The sensor arrays were calibrated, and the subjects’ foreheads and wrists were cleaned with alcohol swabs and gauze before sensors were worn on-body.”
§ Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis > METHODS > On-body sweat analysis
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: fails — The paper’s conclusions may not depend on it.
- Based: passes — The paper gives a reason why it holds.
Probably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper relies on this without support.
Commercially available wearable sensors only track physical activities and vital signs and fail to provide molecular-level health insight.
“At present, commercially available wearable sensors are only capable of tracking an individual's physical activities and vital signs (such as heart rate), and fail to provide insight into the user's health state at molecular levels.”
The paper relies on this without support.
Sweat secretion involves multivariate mechanisms and is complex.
“Given the complexity of sweat secretion, simultaneous and multiplexed screening of target biomarkers is critical and requires full system integration to ensure the accuracy of measurements.”
“Given the multivariate mechanisms that are involved in sweat secretion, an attractive strategy would be to devise a fully integrated multiplexed sensing system to extract the complex information available from sweat.”
The paper relies on this without support.
Amperometric sensors with a larger area produce a larger current signal.
“It is known that amperometric sensors with larger area provide larger current signal.”
The paper relies on this without support.
A thinner Prussian blue layer provides better sensitivity, which is essential for measuring low glucose levels in sweat.
“A thinner Prussian blue layer can provide better sensitivity, which is essential for low-glucose-level measurements in sweat.”
The paper relies on this without support.
A thicker Prussian blue layer yields a broader linear response range, which is important for measuring lactate in sweat.
“A thicker Prussian blue layer can provide a wider linear response range, which is crucial for lactate measurement in sweat.”
The authors state this as an assumption.
Sweat flow can be modeled as a single-centred flow.
“Assuming a single-centred flow model13, the best-case sampling interval can be calculated to be less than 1 min, based on the sweat rate (~3–4 mg min−1 cm−2)39 and the pad size (1.5 cm × 2 cm × 50 μm).”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The sensor remains stable throughout the measurement period with a sensitivity of 56.2 mV/decade, similar to sensitivities achieved by lithography-based Na+ sensors and close to the Nernstian ideal (23).”
From Regional and correlative sweat analysis using high-throughput microfluidic sensing patches toward decoding sweat (2019) · cites this paper as [23]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Recently, a fully integrated wearable sensor array for multiplexed in-situ perspiration analysis was reported, where the sodium, potassium, lactate, and glucose content of sweat can be measured simultaneously by electrochemical methods.”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [83]
The cited source says what the claim attributes to it.
In the citing paper
“Conventional commercially available integrated-circuit components (more than ten chips) can be applied on a FPCB to serve as data processing and transmission.[83]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [83]
The cited source says what the claim attributes to it.
In the citing paper
“Sweat [Na+] and [K+] were deemed stable throughout running in euhydration trials (with water intake of 150 ml per 5 min) after the initial [Na+] increase and [K+] decrease.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
The cited source says what the claim attributes to it.
In the citing paper
“Although such systems can support various types of biophysical measurements of physiological health (8–13), additional advances are needed to meet the challenging requirements of the NICU, where comprehensive, continuous sensing with wireless functionality, clinical-grade measurement fidelity, and mechanical form factors that eliminate risk of harm to exceptionally fragile neonatal skin are essent”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [10]
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Recently, tactile pressure sensors have been developed in the form of active‐matrix arrays consisting of pressure‐sensitive graphene transistors with air–dielectric layers.33”
From Wearables in Medicine (2018) · cites this paper as [33]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“These tactile sensors allowed detecting a broad pressure range from 250 Pa to 3 MPa.”
From Wearables in Medicine (2018) · cites this paper as [33]
Experimental — accuracy not yet measuredIn the citing paper
“Javey group, Gao group and Wang group also focused on electrochemical biosensing of sweat using various wearable sweat patches for sweat rate [30], L-Dopa [138,139], and C-reactive protein [140], cortisol [141,142], ketone [143], glucose [144], and therapeutic drugs [145].”
From The emergence of AI-based wearable sensors for digital health technology: a review (2023) · cites this paper as [30]
In the citing paper
“Wearable sensors are helping clinical and healthcare services with the possible shift from centralized clinical diagnostics to more individualized and personalized home care [29,30] (Figure 1a).”
From The emergence of AI-based wearable sensors for digital health technology: a review (2023) · cites this paper as [30]
In the citing paper
“In addition, since sweat is not sampled but sensed directly on the skin, the skin topography, skin temperature[ 10 ] and pH[ 11 ] may interfere considerably with the measurement results.”
From MicroSweat: A wearable microfluidic patch for noninvasive and reliable sweat collection enables human stress monitoring (2023) · cites this paper as [10]
In the citing paper
“In this work, the Berkeley team demonstrated the simultaneous multiplexed detection of sweat metabolites (glucose and lactate) and electrolytes (sodium and potassium ions), along with skin temperature by integrating a multi-sensing array.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [35]
In the citing paper
“The reported system proved advantageous for monitoring fitness parameters during exercise, but its utility would be limited in continuous monitoring applications due to its reliance on physical exertion for sweat generation.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [35]
In the citing paper
“Results of long-term stability studies of these electrochemical sensors indicate that the sensitivities of the biosensors are consistent over 2 wk with sensitivity variations of <5% (16).”
From Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose monitoring using a fully integrated wearable platform (2017) · cites this paper as [16]
In the citing paper
“We recently demonstrated a fully integrated wearable sensing system for real-time monitoring of multiple analytes in human perspiration during physical exercise which allows accurate measurement of sweat analytes through signal processing and calibration (16).”
From Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose monitoring using a fully integrated wearable platform (2017) · cites this paper as [16]
In the citing paper
“The system allowed for realtime monitoring of body temperature, electrolytes (sodium and potassium ions), and metabolites (glucose and lactate) during prolonged indoor and outdoor physical activities.”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“The multisensor array exhibited good mechanical resiliency with good RSD (1% for electrolytes and 5% for metabolites).”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“Javey and co-workers reported an integrated system consisting of a flexible multiplexed sensor array, a flexible printed circuitry board and a customized app for perspiration/sweat analysis (Figure 8b).[199]”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“Simultaneous and continuous monitoring of sweat profile during exercise can be conducted using wearable electrochemical sensors.[66,199]”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“As illustrated in Figure 10d, by wearing the device consisting of a multiplexed sensor array and a flexible printed circuitry board in the form of a wristband or headband, variations of skin temperature, electrolytes, and sweat metabolites in perspiration can be monitored wirelessly using a customized app on a smart phone.[199]”
From Nanomaterial-Enabled Wearable Sensors for Healthcare (2018) · cites this paper as [199]
In the citing paper
“An increase in sweat [Na+] and a smaller increase in sweat [K+] were observed in dehydration trials (without water intake) after 80 min when subjects had lost a large amount of water (~2.5% of body weight).”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“Gao et al. developed a flexible sweat sensor for the real-time detection of Na+, lactate, K+, glucose, and skin temperature.44”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“As previously mentioned, Gao et al.44 fabricated an epidermal sensor on a PET substrate for the concurrent and continuous detection of sodium, lactate, potassium, glucose, and skin temperature.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“Ex situ measurements of [Na+] and [K+] from collected sweat samples demonstrated similar phenomena.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [44]
In the citing paper
“Recently, Gao et al.41 demonstrated a wearable patch that can simultaneously track levels of metabolites and electrolytes in human sweat. However, electrophysiology sensors were not included, and such multimodal sensor fusion is crucial to obtain a more comprehensive knowledge about a wearer's well-being.”
From A wearable chemical–electrophysiological hybrid biosensing system for real-time health and fitness monitoring (2016) · cites this paper as [41]
In the citing paper
“Several research groups have recently developed mechanically flexible and fully integrated wearable sweat sensors (in wristband, temporary tattoo, eyeglasses, mouthguard) for analysis of biofluid metabolites and electrolytes (e.g., glucose, lactate, sodium, potassium) and pH sensing [156-160].”
From Wearables and the Medical Revolution (2018) · cites this paper as [156]
In the citing paper
“The wristband device also measures skin temperature for calibration, and that data is also useful in analyzing health as described earlier [126,156].”
From Wearables and the Medical Revolution (2018) · cites this paper as [156]
In the citing paper
“These devices have potential applications for measuring heat stress, dehydration and metabolism in athletes, soldiers, firefighters, and industrial laborers who exercise or work in hot environments.”
From A Critical Review of Consumer Wearables, Mobile Applications, and Equipment for Providing Biofeedback, Monitoring Stress, and Sleep in Physically Active Populations (2018) · cites this paper as Gao et al., 2016
Citations
Cites
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836 · cited 1×
“Cloud servers Transimpedance amplifier (1) Transimpedance amplifier (3)”
BackgroundSkin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes
Lipomi, D. J., Vosgueritchian, M., Tee, B. C.-K., Hellstrom, S. L., Lee, J. A., Fox, C. H., & Bao, Z. (2011) · Nature Nanotechnology · doi:10.1038/nnano.2011.184 · Cited by 3,188
An ultra-lightweight design for imperceptible plastic electronics
Kaltenbrunner, M., Sekitani, T., Reeder, J., Yokota, T., Kuribara, K., Tokuhara, T., Drack, M., Schwödiauer, R., Graz, I., Bauer-Gogonea, S., et al. (2013) · Nature · doi:10.1038/nature12314 · Cited by 2,485 · cited 2×
“Differential amplifier (6) Voltage buffer (8)”
BackgroundA review of wearable sensors and systems with application in rehabilitation
Patel, S., Park, H., Bonato, P., Chan, L., & Rodgers, M. (2012) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-9-21 · Cited by 2,396 · cited 1×
“Bluetooth transceiver (11) Iglucose Ilactate Rtemperature Vsodium Vref Vpotassium Ag/AgCl Gluc ose”
BackgroundElectrochemical Biosensors - Sensor Principles and Architectures
Grieshaber, D., MacKenzie, R., Vörös, J., & Reimhult, E. (2008) · Sensors · doi:10.3390/s80314000 · Cited by 2,329
Piezoelectricity of single-atomic-layer MoS2 for energy conversion and piezotronics
Wu, W., Wang, L., Li, Y., Zhang, F., Lin, L., Niu, S., Chenet, D., Zhang, X., Hao, Y., Heinz, T. F., et al. (2014) · Nature · doi:10.1038/nature13792 · Cited by 2,287 · cited 1×
“Low-pass filter (2) Low-pass filter (4) Voltage divider”
MethodNanowire active-matrix circuitry for low-voltage macroscale artificial skin
Takei, K., Takahashi, T., Ho, J. C., Ko, H., Gillies, A. G., Leu, P. W., Fearing, R. S., & Javey, A. (2010) · Nature Materials · doi:10.1038/nmat2835 · Cited by 1,274 · cited 1×
“Cloud servers Transimpedance amplifier (1) Transimpedance amplifier (3)”
BackgroundSelf-citationNon-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247
Ultrathin conformal devices for precise and continuous thermal characterization of human skin
Webb, R. C., Bonifas, A. P., Behnaz, A., Zhang, Y., Yu, K. J., Cheng, H., Shi, M., Bian, Z., Liu, Z., Kim, Y.-S., et al. (2013) · Nature Materials · doi:10.1038/nmat3755 · Cited by 1,229 · cited 1×
“…dehydration 20 ; sweat glucose is reported to be metabolically related to blood glucose 21 ; sweat lactate can potentially serve as a sensitive marker of pressure ischaemia 22 ; and skin temperature is clinically informative of a variety of diseases and skin injuries such as pressure ulcers 23,24 .”
BackgroundSoft Microfluidic Assemblies of Sensors, Circuits, and Radios for the Skin
Xu, S., Zhang, Y., Jia, L., Mathewson, K. E., Jang, K.-I., Kim, J., Fu, H., Huang, X., Chava, P., Wang, R., et al. (2014) · Science · doi:10.1126/science.1250169 · Cited by 1,199 · cited 1×
“Low-pass filter (7) Low-pass filter (9)”
Cited by
Sorted by citations
Items 801–900 of 5,047
Kirigami-inspired strain-insensitive sensors based on atomically-thin materials
Yong, K., De, S., Hsieh, E. Y., Leem, J., Aluru, N. R., & Nam, S. (2019) · Materials Today · doi:10.1016/j.mattod.2019.08.013 · Cited by 124
Reeder, J. T., Xue, Y., Franklin, D., Deng, Y., Choi, J., Prado, O., Kim, R., Liu, C. Y., Hanson, J., Ciraldo, J. P., et al. (2019) · Nature Communications · doi:10.1038/s41467-019-13431-8 · Cited by 125
Li, X., Koh, K. H., Farhan, M., & Lai, K. W. C. (2020) · Nanoscale · doi:10.1039/c9nr09306k · Cited by 126
Busolo, T., Szewczyk, P. K., Nair, M., Stachewicz, U., & Kar‐Narayan, S. (2021) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c00983 · Cited by 132
Khor, S. M., Choi, J., Won, P., & Ko, S. H. (2022) · Nanomaterials · doi:10.3390/nano12020221 · Cited by 133
“A self-contained wearable sweat glucose biosensor must be a fully integrated and autonomous system [24].”
BackgroundNoninvasive Body Fat Burn Monitoring from Exhaled Acetone with Si-doped WO3-sensing Nanoparticles
Güntner, A. T., Sievi, N. A., Theodore, S. J., Gulich, T., Kohler, M., & Pratsinis, S. E. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b02843 · Cited by 127
Two‐Layered and Stretchable e‐Textile Patches for Wearable Healthcare Electronics
La, T., Qiu, S., Scott, D. K., Bakhtiari, R., Kuziek, J. W. P., Mathewson, K. E., Rieger, J., & Chung, H. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201801033 · Cited by 126
Zhai, Q., Gong, S. P., Wang, Y., Lyu, Q., Liu, Y., Ling, Y., Wang, J., Simon, G. P., & Cheng, W. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b19383 · Cited by 124
Stimuli-responsive engineered living materials
Rivera‐Tarazona, L. K., Campbell, Z. T., & Ware, T. H. (2020) · Soft Matter · doi:10.1039/d0sm01905d · Cited by 132
Miniaturized surface engineered technologies for multiplex biosensing devices
Srivastava, V. R., Kumari, R., & Chandra, P. (2023) · Electroanalysis · doi:10.1002/elan.202200355 · Cited by 126
Active-reset protein sensors enable continuous in vivo monitoring of inflammation
Zargartalebi, H., Mirzaie, S., GhavamiNejad, A., Ahmed, S. U., Esmaeili, F., Geraili, A., Flynn, C. D., Chang, D., Das, J., Abdrabou, A. M., et al. (2024) · Science · doi:10.1126/science.adn2600 · Cited by 138
Sun, Y., He, W., Jiang, C., Li, J., Liu, J., & Liu, M. (2025) · Nano-Micro Letters · doi:10.1007/s40820-024-01597-w · Cited by 134
Novel Eco‐Friendly Starch Paper for Use in Flexible, Transparent, and Disposable Organic Electronics
Jeong, H., Baek, S., Han, S., Jang, H., Kim, S. H., & Lee, H. S. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201704433 · Cited by 123
Dervisevic, M., Dervisevic, E., Esser, L., Easton, C. D., Cadarso, V. J., & Voelcker, N. H. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114955 · Cited by 125
Flexible Electrochemical Biosensors for Health Monitoring
Yang, A., & Yan, F. (2021) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.0c00534 · Cited by 122
Mechanically Reinforced Skin‐Electronics with Networked Nanocomposite Elastomer
Han, S., Kim, M. K., Wang, B., Wie, D. S., Wang, S., & Lee, C. H. (2016) · Advanced Materials · doi:10.1002/adma.201603878 · Cited by 122
Polymer/carbon nanotube composite materials for flexible thermoelectric power generator
Song, H., Qiu, Y., Wang, Y., Cai, K., Li, D., Deng, Y., & He, J. (2017) · Composites Science and Technology · doi:10.1016/j.compscitech.2017.10.006 · Cited by 126
Mechano‐Based Transductive Sensing for Wearable Healthcare
Wang, T., Yang, H., Qi, D., Liu, Z., Cai, P., Zhang, H., & Chen, X. (2018) · Small · doi:10.1002/smll.201702933 · Cited by 121
Zhang, S., Zahed, M. A., Sharifuzzaman, M., Yoon, S., Hui, X., Barman, S. C., Sharma, S., Yoon, H. S., Park, C., & Park, J. Y. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112844 · Cited by 126
High-strength, highly conductive and woven organic hydrogel fibers for flexible electronics
Wang, X., Wang, X., Pi, M., & Ran, R. (2021) · Chemical Engineering Journal · doi:10.1016/j.cej.2021.131172 · Cited by 119
Wearable potentiometric biosensor for analysis of urea in sweat
Ibáñez-Redín, G., Cagnani, G. R., Gomes, N. O., Raymundo‐Pereira, P. A., Machado, S. A. S., Gutierrez, M. A., Krieger, J. E., & Oliveira, O. N. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114994 · Cited by 124
Recent progress of bio-based smart wearable sensors for healthcare applications
Banitaba, S. N., Khademolqorani, S., Jadhav, V. V., Chamanehpour, E., Mishra, Y. K., Mostafavi, E., & Kaushik, A. K. (2023) · Materials Today Electronics · doi:10.1016/j.mtelec.2023.100055 · Cited by 126
High performance flexible pH sensor based on polyaniline nanopillar array electrode
Yoon, J. H., Hong, S. B., Yun, S.-O., Lee, S. J., Lee, T. J., Lee, K. G., & Choi, B. G. (2016) · Journal of Colloid and Interface Science · doi:10.1016/j.jcis.2016.11.033 · Cited by 120
Hu, Y., Zhao, T., Zhu, P., Zhang, Y., Liang, X., Sun, R., & Wong, C. (2017) · Nano Research · doi:10.1007/s12274-017-1811-0 · Cited by 120
You, I., Choi, S., Hwang, H., Han, S. W., Kim, J. W., & Jeong, U. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201801858 · Cited by 120
Soft, skin-interfaced wearable systems for sports science and analytics
Ray, T. R., Choi, J., Reeder, J. T., Lee, S. P., Aranyosi, A. J., Ghaffari, R., & Rogers, J. A. (2019) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2019.01.003 · Cited by 126
Highly self-healable and flexible cable-type pH sensors for real-time monitoring of human fluids
Yoon, J. H., Kim, S., Park, H. J., Kim, Y. K., Oh, D. X., Cho, H.-W., Lee, K. G., Hwang, S. Y., Park, J., & Choi, B. G. (2019) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2019.111946 · Cited by 120
Tian, Z., Sun, Z., Shao, Y., et al. (2021) · Energy & Environmental Science · doi:10.1039/d0ee03623d · Cited by 121
A flexible transparent colorimetric wrist strap sensor
Wang, T., Guo, Y., Wan, P., Sun, X., Zhang, H., Yu, Z., & Chen, X. (2016) · Nanoscale · doi:10.1039/c6nr08265c · Cited by 120
Organic semiconductors for biological sensing
Borges‐González, J., Kousseff, C. J., & Nielsen, C. B. O. (2019) · Journal of Materials Chemistry C · doi:10.1039/c8tc05900d · Cited by 121
Design and applications of stretchable and self-healable conductors for soft electronics
Zhao, Y., Kim, A., Wan, G., & Tee, B. C. K. (2019) · Nano Convergence · doi:10.1186/s40580-019-0195-0 · Cited by 128
Multiscale Soft–Hard Interface Design for Flexible Hybrid Electronics
Gong, S. P., Yap, L. W., Zhu, B., & Cheng, W. (2019) · Advanced Materials · doi:10.1002/adma.201902278 · Cited by 122
Borophene pressure sensing for electronic skin and human-machine interface
Hou, C., Tai, G., Liu, Y., Liu, R., Liang, X., Wu, Z., & Wu, Z. (2022) · Nano Energy · doi:10.1016/j.nanoen.2022.107189 · Cited by 117
Chou, X., Zhu, J., Qian, S., et al. (2018) · Nano Energy · doi:10.1016/j.nanoen.2018.09.006 · Cited by 119
High‐Transconductance Stretchable Transistors Achieved by Controlled Gold Microcrack Morphology
Matsuhisa, N., Jiang, Y., Liu, Z., Chen, G., Wan, C., Kim, Y., Kang, J., Tran, H., Wu, H., You, I., et al. (2019) · Advanced Electronic Materials · doi:10.1002/aelm.201900347 · Cited by 122
Sweat as a Source of Next-Generation Digital Biomarkers
Brasier, N., & Eckstein, J. (2019) · Digital Biomarkers · doi:10.1159/000504387 · Cited by 120
A thread-based wearable sweat nanobiosensor
Zhao, C., Li, X., Wu, Q., & Liu, X. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113270 · Cited by 122
Stretchable and self-healable hydrogel artificial skin
Xue, B., Sheng, H. X., Li, Y., Li, L., Di, W., Xu, Z., Ma, L., Wang, X., Jiang, H., Qin, M., et al. (2021) · National Science Review · doi:10.1093/nsr/nwab147 · Cited by 121
“These characteristics have inspired the development of artificial skin-like materials for flexible electronics with various advanced applications in soft robotics [1,2], human-machine interfaces [3,4] andwearable electronic devices [5,6].”
BackgroundLi, T., Koh, J. Y. C., Moudgil, A., et al. (2022) · ACS Nano · doi:10.1021/acsnano.2c02191 · Cited by 119
Emerging 0D, 1D, 2D, and 3D nanostructures for efficient point-of-care biosensing
Byakodi, M., Shrikrishna, N. S., Sharma, R., Bhansali, S., Mishra, Y. K., Kaushik, A. K., & Gandhi, S. (2022) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2022.100284 · Cited by 127
“Epidermal sensors (Ciui et al., 2018) conjugated with wearable sweat platforms result in a POC device that should acquire, process, and transmit data from the electrochemical sweat (Bandodkar et al., 2019; Gao et al., 2016) to a user interface.”
Khan, A., DeVoe, E., & Andreescu, S. (2023) · Sensors & Diagnostics · doi:10.1039/d2sd00226d · Cited by 131
Jeong, H., Lee, J. Y., Lee, K., Kang, Y. J., Kim, J.-T., Avila, R., Tzavelis, A., Kim, J., Ryu, H., Kwak, S. S., et al. (2021) · Science Advances · doi:10.1126/sciadv.abg3092 · Cited by 121
Wearable Ring-Based Sensing Platform for Detecting Chemical Threats
Sempionatto, J. R., Mishra, R. K., Martín, A., Tang, G., Nakagawa, T., Lu, X., Campbell, A. S., Lyu, K. M., & Wang, J. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00603 · Cited by 117
Zhang, H., Li, X., Yang, Y., Xiao, M., Han, J., Ding, L., Zhang, Z., Hu, Y., & Peng, L. (2018) · ACS Nano · doi:10.1021/acsnano.7b09145 · Cited by 118
Noninvasive wearable electroactive pharmaceutical monitoring for personalized therapeutics
Lin, S., Yu, W., Wang, B., Zhao, Y., En, K., Zhu, J., Cheng, X., Zhou, C., Lin, H., Wang, Z., et al. (2020) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2009979117 · Cited by 117
Self-citationMicrofluidics for flexible electronics
Guo, J., Yu, Y., Cai, L., Wang, Y., Shi, K., Shang, L., Pan, J., & Zhao, Y. (2021) · Materials Today · doi:10.1016/j.mattod.2020.08.017 · Cited by 120
Integrated water quality monitoring system with pH, free chlorine, and temperature sensors
Qin, Y., Alam, A. U., Pan, S. H., Howlader, M. M. R., Ghosh, R., Hu, N., Jin, H., Dong, S., Chen, C., & Deen, M. J. (2017) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2017.07.188 · Cited by 123
Chen, C.-H., Lee, P.-W., Tsao, Y.-H., & Lin, Z. (2017) · Nano Energy · doi:10.1016/j.nanoen.2017.10.064 · Cited by 117
Recent Advances in Smart Contact Lenses
Kim, J., Cha, E., & Park, J. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201900728 · Cited by 120
Material and structural considerations for high-performance electrodes for wearable skin devices
Lim, K., Seo, H., Chung, W. G., Song, H., Oh, M., Ryu, S. Y., Kim, Y., & Park, J. (2024) · Communications Materials · doi:10.1038/s43246-024-00490-8 · Cited by 142
Designing wearable microgrids: towards autonomous sustainable on-body energy management
Yin, L., Kim, K. N., Trifonov, A., Podhajny, T., & Wang, J. (2022) · Energy & Environmental Science · doi:10.1039/d1ee03113a · Cited by 111
Technological leap for sweat sensing
Heikenfeld, J. C. (2016) · Nature · doi:10.1038/529475a · Cited by 117
Jin, Z.-H., Liu, Y.-L., Chen, J.-J., Cai, S.-L., Xu, J.-Q., & Huang, W.-H. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.6b04616 · Cited by 116
Epidermal Inorganic Optoelectronics for Blood Oxygen Measurement
Li, H., Xu, Y. F., Li, X., Chen, Y., Jiang, Y., Zhang, C., Lu, B., Wang, J., Ma, Y., Chen, Y., et al. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201601013 · Cited by 118
Sun, C., Shi, Q., Hasan, D. M. N., Yazici, M. S., Zhu, M., Ma, Y., Dong, B., Liu, Y., & Lee, C. (2019) · Nano Energy · doi:10.1016/j.nanoen.2019.01.096 · Cited by 114
Li, G., & Wen, D. (2020) · Journal of Materials Chemistry B · doi:10.1039/c9tb02474c · Cited by 119
Gomes, N. O., Carrilho, E., Machado, S. A. S., & Sgobbi, L. F. (2020) · Electrochimica Acta · doi:10.1016/j.electacta.2020.136341 · Cited by 116
Self-powered and wearable biosensors for healthcare
Zeng, X., Peng, R., Fan, Z., & Lin, Y. (2021) · Materials Today Energy · doi:10.1016/j.mtener.2021.100900 · Cited by 123
“A variety of biosensors, such as glucose sensors, lactate sensors, uric acid (UA) sensors, has been successfully integrated into wearable platforms such as watches [1], glass [2], and mouthguards [3 e 6].”
Xiao, J., Fan, C. G., Xu, T., Su, L., & Zhang, X. (2022) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2022.131586 · Cited by 115
Chenani, H., Saeidi, M., Rastkhiz, M. A., Bolghanabadi, N., Aghaii, A. H., Orouji, M., Hatamie, A., & Simchi, A. (2024) · Analytical Chemistry · doi:10.1021/acs.analchem.3c03942 · Cited by 123
New Carbon Materials for Multifunctional Soft Electronics
Xue, J., Liu, D., Li, D., Hong, T., Li, C., Zhu, Z., Sun, Y., Gao, X., Guo, L., Shen, X., et al. (2024) · Advanced Materials · doi:10.1002/adma.202312596 · Cited by 123
Wearable Sensing Systems with Mechanically Soft Assemblies of Nanoscale Materials
Lee, Y., Kim, J., Joo, H., Raj, M. S., Ghaffari, R., & Kim, D. (2017) · Advanced Materials Technologies · doi:10.1002/admt.201700053 · Cited by 116
Flexible thermoelectric power generation system based on rigid inorganic bulk materials
Eom, Y., Wijethunge, D., Park, H., Park, S. H., & Kim, W. (2017) · Applied Energy · doi:10.1016/j.apenergy.2017.08.231 · Cited by 117
Superwettable Electrochemical Biosensor toward Detection of Cancer Biomarkers
Xu, T., Song, Y., Gao, W. F., Wu, T., Xu, L., Zhang, X., & Wang, S. (2018) · ACS Sensors · doi:10.1021/acssensors.7b00868 · Cited by 113
“ter electrode is chosen to simplify design and suit the small size of microdroplets, which is a common strategy for low-current and low-volume electrochemical sensing.(54) Two kinds of typical prostate cancer biomarkers including miRNAs (miRNA-141 and miRNA-375, (56)) and prostate-specific antigen (PSA),(57) were”
BackgroundSelf-citationKim, S. B., Lee, K., Raj, M. S., Lee, B.-R., Reeder, J. T., Koo, J., Hourlier‐Fargette, A., Bandodkar, A. J., Won, S. M., Sekine, Y., et al. (2018) · Small · doi:10.1002/smll.201802876 · Cited by 118
Kweon, O. Y., Samanta, S. K., Won, Y., Yoo, J. H., & Oh, J. H. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.9b04440 · Cited by 115
Yan, T., Zhang, G., Yu, K., Chai, H., Tian, M., Qu, L., Dong, H., & Zhang, X. (2022) · Chemical Engineering Journal · doi:10.1016/j.cej.2022.140779 · Cited by 120
Flexible Sweat Sensors: From Films to Textiles
Luo, D., Sun, H., Li, Q., Niu, X., He, Y., & Liu, H. (2023) · ACS Sensors · doi:10.1021/acssensors.2c02642 · Cited by 119
Wang, W., Chen, Y., Xiao, C., Xiao, S., Wang, C., Nie, Q., Xu, P., Chen, J., You, R., Zhang, G., et al. (2023) · Chemical Engineering Journal · doi:10.1016/j.cej.2023.145953 · Cited by 118
Bai, H., Zhao, T., & Cao, M. (2025) · Chemical Society Reviews · doi:10.1039/d4cs01073f · Cited by 121
Recent Advances in Biointegrated Optoelectronic Devices
Xu, H., Yin, L., Liu, C., Sheng, X., & Zhao, N. (2018) · Advanced Materials · doi:10.1002/adma.201800156 · Cited by 119
Multifunctional Fibers to Shape Future Biomedical Devices
Zhang, Y., Ding, J., Qi, B., Tao, W., Wang, J., Zhao, C., Peng, H., & Shi, J. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201902834 · Cited by 114
Ray, P. P., Dash, D., & Kumar, N. (2020) · Computer Communications · doi:10.1016/j.comcom.2020.05.029 · Cited by 121
Li, H., Wu, G., Weng, Z., Sun, H., Nistala, R., & Zhang, Y. (2021) · ACS Sensors · doi:10.1021/acssensors.0c02330 · Cited by 117
Recent progress in flexible micro-pressure sensors for wearable health monitoring
Hu, J., Dun, G., Geng, X., Chen, J., Wu, X., & Ren, T. (2023) · Nanoscale Advances · doi:10.1039/d2na00866a · Cited by 120
Lin, S., Wang, B., Zhao, Y., Shih, R., Cheng, X., Yu, W., Hojaiji, H., Lin, H., Hoffman, C., Ly, D., et al. (2020) · ACS Sensors · doi:10.1021/acssensors.9b01727 · Cited by 114
Self-citationBaker, L. B., Ungaro, C. T., Sopeña, B. C., Nuccio, R. P., Reimel, A. J., Carter, J. M., Stofan, J. R., & Barnes, K. A. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00867.2017 · Cited by 116
“Because of the variability in sweating among athletes and recommendations to personalize fluid replacement (27, 37), there has been recent interest in practical methods to measure sweat fluid and electrolyte losses during exercise (18, 23, 24, 32, 33, 42).”
BackgroundEmergence of microfluidic wearable technologies
Yeo, J. C., Kenry, K., & Lim, C. T. (2016) · Lab on a Chip · doi:10.1039/c6lc00926c · Cited by 115
Recent progress in flexible and wearable bio-electronics based on nanomaterials
Yang, Y., Xiang-Dong, Y., Tan, Y., & Yuan, Q. (2017) · Nano Research · doi:10.1007/s12274-017-1476-8 · Cited by 114
All flexible electrospun papers based self-charging power system
Sun, N., Wen, Z., Zhao, F., Yang, Y., Shao, H., Zhou, C., Shen, Q., Feng, K., Peng, M., Li, Y., et al. (2017) · Nano Energy · doi:10.1016/j.nanoen.2017.05.048 · Cited by 114
Fu, Y., He, H., Zhao, T., Dai, Y., Han, W., Ma, J., Xing, L., Zhang, Y., & Xue, X. (2018) · Nano-Micro Letters · doi:10.1007/s40820-018-0228-y · Cited by 116
Wearable and Implantable Cortisol‐Sensing Electronics for Stress Monitoring
Ok, J., Park, S., Jung, Y. H., & Kim, T. (2023) · Advanced Materials · doi:10.1002/adma.202211595 · Cited by 131
Soft Materials and Devices Enabling Sensorimotor Functions in Soft Robots
Su, J., He, K., Li, Y., Tu, J., & Chen, X. (2025) · Chemical Reviews · doi:10.1021/acs.chemrev.4c00906 · Cited by 121
Kutbee, A. T., Bahabry, R. R., Alamoudi, K. O., Ghoneim, M. T., Cordero, M. D., Almuslem, A. S., Gümüş, A., Diallo, E. M., Nassar, J. M., Hussain, A. M., et al. (2017) · npj Flexible Electronics · doi:10.1038/s41528-017-0008-7 · Cited by 113
Interactive Skin Display with Epidermal Stimuli Electrode
Kim, E. H., Han, H., Yu, S., Park, C., Kim, G., Jeong, B., Lee, S. W., Kim, J. S., Lee, S., Kim, J., et al. (2019) · Advanced Science · doi:10.1002/advs.201802351 · Cited by 112
Modular and Reconfigurable Wireless E‐Tattoos for Personalized Sensing
Jeong, H., Wang, L., Ha, T., Mitbander, R., Yang, X., Dai, Z., Qiao, S., Shen, L., Sun, N., & Lu, N. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201900117 · Cited by 111
Zhang, D., Xu, S., Zhao, X. Q., Qian, W., Bowen, C., & Yang, Y. F. (2020) · Advanced Functional Materials · doi:10.1002/adfm.201910809 · Cited by 112
Directed Assembly of Liquid Metal–Elastomer Conductors for Stretchable and Self‐Healing Electronics
Krisnadi, F., Nguyen, L. L., Ankit, A., Ma, J., Kulkarni, M. R., Mathews, N., & Dickey, M. D. (2020) · Advanced Materials · doi:10.1002/adma.202001642 · Cited by 115
Recent status and future perspectives of ultracompact and customizable micro-supercapacitors
Wang, S., Ma, J., Shi, X., et al. (2022) · Nano Research Energy · doi:10.26599/nre.2022.9120018 · Cited by 112
Inhomogeneity in pore size appreciably lowering thermal conductivity for porous thermal insulators
Qiu, L., Zou, H., Tang, D., Wen, D., Feng, Y., & Zhang, X. (2017) · Applied Thermal Engineering · doi:10.1016/j.applthermaleng.2017.11.066 · Cited by 111
“AITS are devices that can be mated with to closely monitor a material’s thermophysical performance, without destroying or limiting the material’s integrity, which could enable in situ analyzing heat transfer process [17-19] .”
BackgroundContinuous biomarker monitoring by particle mobility sensing with single molecule resolution
Visser, E. W. A., Yan, J., van IJzendoorn, L. J., et al. (2018) · Nature Communications · doi:10.1038/s41467-018-04802-8 · Cited by 111
Wireless powered wearable micro light-emitting diodes
Lee, H. E., Lee, D., Lee, T., Shin, J. H., Choi, G., Kim, C., Lee, S. H., Lee, J. H., Kim, Y. H., Kang, S., et al. (2018) · Nano Energy · doi:10.1016/j.nanoen.2018.11.017 · Cited by 113
Progress on wearable triboelectric nanogenerators in shapes of fiber, yarn, and textile
Xiong, J., & Lee, P. S. (2019) · Science and Technology of Advanced Materials · doi:10.1080/14686996.2019.1650396 · Cited by 112
Jeerapan, I., & Poorahong, S. (2020) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ab7117 · Cited by 116
Recent advances in bioelectronics chemistry
Fang, Y., Meng, L., Promiński, A., Schaumann, E. N., Seebald, M., & Tian, B. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00333f · Cited by 121
Huang, L., Wang, H., Wu, P., Huang, W., Gao, W., Fang, F., Cai, N., Chen, R., & Zhu, Z. (2020) · Sensors · doi:10.3390/s20154266 · Cited by 111
Self-citationEpidermal self-powered sweat sensors for glucose and lactate monitoring
Huang, X., Li, J., Liu, Y., Wong, T., Su, J., Yao, K., Zhou, J., Huang, Y., Li, H., Li, D., et al. (2021) · Bio-Design and Manufacturing · doi:10.1007/s42242-021-00156-1 · Cited by 115
Electrochemical biosensors for point-of-care testing
Kim, J., Jeong, J., & Ko, S. H. (2024) · Bio-Design and Manufacturing · doi:10.1007/s42242-024-00301-6 · Cited by 126
Hybrid Integration of Wearable Devices for Physiological Monitoring
Zhang, Y., Zheng, X. T., Zhang, X., Pan, J., & Thean, A. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00471 · Cited by 126
Highly Sensitive, Durable, and Multifunctional Sensor Inspired by a Spider
Luo, C., Jia, J., Gong, Y., Wang, Z., Fu, Q., & Pan, C. (2017) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.7b02988 · Cited by 109
Metadata
- Authors
- Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., Lien, D.-H., Brooks, G. A., Davis, R. W., & Javey, A.
- Affiliations
- Lawrence Berkeley National Laboratory (US)
- Stanford Medicine (US)
- University of California, Berkeley (US)
- Corresponding author
- Javey, Ali
- ORCID
- Venue
- Nature
- Published
- 2016-01-27
- Volume
- 529
- Issue
- 7587
- Pages
- 509-514
- ISSN
- 0028-0836 (ISSN-L)
- 1476-4687
- PMC
- PMC4996079
- OpenAlex
- W2266175888
- Semantic Scholar
- 52709f85aaaaea3dbceb33a9666e3cb1dfb932a3
- Cited by
- 5,087
- References
- 39
- Citations by year
- 2026: 462 · 2025: 581 · 2024: 528 · 2023: 543 · 2022: 531 · 2021: 585 · 2020: 538 · 2019: 502 · 2018: 415 · 2017: 282 · 2016: 117
- FWCI
- 211.10
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Nature (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) (submittedVersion)
- CaltechAUTHORS (California Institute of Technology) (submittedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- eScholarship (California Digital Library) (submittedVersion)
- Rare & Special e-Zone (The Hong Kong University of Science and Technology) (submittedVersion)
- Funders
- National Science Foundation (DEAC02-05CH11231)
- U.S. Department of Energy (-AC02-05CH11231, 05CH11231, DE AC02 05CH11231, DE-AC02-)
- National Institutes of Health (DE-AC02–05CH11231, HG000205)
- Office of Science (DEAC02–05CH11231, AC02-05CH11231, DE-AC02)
- Japan Society for the Promotion of Science
- National Human Genome Research Institute (P01HG000205)
- Basic Energy Sciences (DE-AC02, DE-AC02-05CH11231., AC02–05CH11231)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Keywords
- wearable sensors, sweat analysis, sweat electrolytes, glucose, lactate, sodium, potassium, skin temperature, flexible electronics, silicon integrated circuits, multiplexed biosensing, wireless transmission, real-time physiological monitoring, non-invasive monitoring, wearable biosensors, physical activity monitoring, personalized medicine, drug abuse detection, athletic performance
- MeSH terms
- Adult, Bicycling, Body Water, Calibration, Electrolytes, Female, Glucose, Humans, Male, Monitoring, Physiologic, Running, Skin, Skin Temperature, Sweat, Reproducibility of Results, Lactic Acid, Young Adult, Precision Medicine, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-09-28, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-09-28 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review