A Flexible and Wearable Human Stress Monitoring Patch
Open accessAbstract
A human stress monitoring patch integrates three sensors of skin temperature, skin conductance, and pulsewave in the size of stamp (25 mm × 15 mm × 72 μm) in order to enhance wearing comfort with small skin contact area and high flexibility. The skin contact area is minimized through the invention of an integrated multi-layer structure and the associated microfabrication process; thus being reduced to 1/125 of that of the conventional single-layer multiple sensors. The patch flexibility is increased mainly by the development of flexible pulsewave sensor, made of a flexible piezoelectric membrane supported by a perforated polyimide membrane. In the human physiological range, the fabricated stress patch measures skin temperature with the sensitivity of 0.31 Ω/°C, skin conductance with the sensitivity of 0.28 μV/0.02 μS, and pulse wave with the response time of 70 msec. The skin-attachable stress patch, capable to detect multimodal bio-signals, shows potential for application to wearable emotion monitoring.
AI snapshot
Written from the paper’s full text.
TL;DR
The paper presents a flexible, wearable multimodal stress-monitoring patch integrating skin temperature, skin conductance, and pulsewave sensors in a multilayer structure. It reports reduced skin contact area and characterizes each sensor in human physiological ranges, though validation is bench-top/artificial-artery rather than in vivo.
Key findings
- The fabricated patch is reported as 24.7 mm × 15.0 mm × 70 μm, weighing 49.5 mg, with a skin contact area of 3.71 cm²—about 1/125 of a conventional single-layer multimodal stress-monitoring device (451 cm²).
- Sensor performance: skin temperature sensitivity 0.31 Ω/°C with 0.021% nonlinearity; skin conductance sensitivity 0.28 μV/0.02 μS, resolution 0.02 μS, nonlinearity 0.447%; pulsewave time delay 70.00 ± 7.42 msec, with sensitivities of 0.35, 0.38, and 0.33 mV/mmHg at 50, 135, and 220 BPM.
- Adding windows to the polyimide support (Type A vs. Type B) theoretically increased patch flexibility by 6.60% (bending stiffness 89.21 vs. 95.51 μNm²) and pulsewave sensing sensitivity by 57% (0.373 vs. 0.237 mV/mmHg).
- Estimated patch lifetime is 9 days, limited by P(VDF-TrFE) retention (0.8 × 10⁶ seconds); aluminum corrosion gives an estimated 8–100 months for the temperature and conductance sensors.
Methods
- Design/fabrication: a three-layer patch with a skin-contact layer containing aluminum skin temperature and skin conductance sensors, a parylene-C insulation layer, and a pulsewave-sensing layer using a P(VDF-TrFE) piezoelectric membrane on a perforated polyimide support with a silver interlayer electrode. Fabrication included 50 μm Kapton, 0.5 μm Ag, 20 μm P(VDF-TrFE), 0.3 μm Al, 1 μm parylene-C, KOH-ethanol-DI wet etching, a 21-minute DI dip, and poling at 900 V, 90 °C for 30 minutes.
- Characterization: skin temperature sensor tested at 30–40 °C in 0.5 °C increments; skin conductance sensor tested at 2–20 μS; pulsewave sensor tested on an artificial artery actuator with a PDMS artificial radial artery at 40–120 mmHg and 50–220 BPM. Theoretical simulations used COMSOL Multiphysics and MATLAB to compare Type A (with windows) and Type B (without windows).
- No human subjects were used for the reported experimental characterization; testing was bench-top and artificial-artery based.
Limitations
- Stated: the study did not perform a subjective comfort test.
- Stated: the experimental pulsewave output voltage and sensitivity were lower than the simulation result; patch lifetime is estimated as only 9 days due to P(VDF-TrFE) retention.
- Evident: there is no in vivo human validation; the pulsewave sensor was tested on an artificial artery actuator rather than a human wrist, and the emotion-sensing claim (SVM categorization of surprise, anger, stress, and sadness) relies on a prior reference and is not experimentally demonstrated here.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 41 cited sources checked · 0 supported · 0 flagged
- Findingown data
The fabricated stress patch measures skin temperature with a sensitivity of 0.31 Ω/°C in the human physiological range.
“In the human physiological range, the fabricated stress patch measures skin temperature with the sensitivity of 0.31 Ω/°C, skin conductance with the sensitivity of 0.28 μV/0.02 μS, and pulse wave with the response time of 70 msec.”
- Also stated in: § Experimental Characterization · plain
“The skin temperature sensor nonlinearity and sensitivity (see Fig. 6a) are measured as 0.021% and 0.31 Ω/°C, respectively.”
Quote in the paperSubstantial claim - Also stated in: § Experimental Characterization · plain
- Findingown data
The fabricated stress patch measures skin conductance with a sensitivity of 0.28 μV/0.02 μS in the human physiological range.
“In the human physiological range, the fabricated stress patch measures skin temperature with the sensitivity of 0.31 Ω/°C, skin conductance with the sensitivity of 0.28 μV/0.02 μS, and pulse wave with the response time of 70 msec.”
- Also stated in: § Experimental Characterization · plain
“The skin conductance sensor has showed the nonlinearity, resolution, and sensitivity (see Fig. 6b) as 0.447%, 0.02 μS, and 0.28 μV/0.02 μS, respectively, within the human skin conductance range.”
Quote in the paperSubstantial claim - Also stated in: § Experimental Characterization · plain
- Findingown data
The skin contact area of the patch is reduced to 1/125 of that of the conventional single-layer multiple sensors.
“The skin contact area is minimized through the invention of an integrated multi-layer structure and the associated microfabrication process; thus being reduced to 1/125 of that of the conventional single-layer multiple sensors.”
- Also stated in: § Discussion · plain
“As a result, the skin contact area of the fabricated multi-layer patch (3.71 cm2) is as small as the 1/125 of that of the conventional single layer multimodal stress monitoring device (451 cm2)28.”
Quote in the paperSubstantial claim - Also stated in: § Discussion · plain
- Findingown data
The fabricated stress patch measures pulse wave with a response time of 70 msec.
“In the human physiological range, the fabricated stress patch measures skin temperature with the sensitivity of 0.31 Ω/°C, skin conductance with the sensitivity of 0.28 μV/0.02 μS, and pulse wave with the response time of 70 msec.”
- Also stated in: § Experimental Characterization · plain
“The time delay of the pulsewave sensor (see Fig. 7a) is maintained as 70.00 ± 7.42 msec.”
Quote in the paperSubstantial claim - Also stated in: § Experimental Characterization · plain
- Findingreasoning
The skin-attachable stress patch is capable of detecting multimodal bio-signals.
“The skin-attachable stress patch, capable to detect multimodal bio-signals, shows potential for application to wearable emotion monitoring.”
Quote in the paperSubstantial claim
- Findingnone given
A novel silver inter-layer electrode is prepared between the piezoelectric and support membranes.
“We develop a novel structure and process for the flexible multilayer stress patch, where a novel silver inter-layer electrode is prepared between the piezoelectric and support membranes.”
Quote in the paperSubstantial claim - Findingown data
The output signal of the pulsewave sensor is correlated well with the amplitude of 40 mmHg and the frequency of 50 BPM of input pressure pulse equivalent to the arterial pulsewave's pressure difference amplitude and frequency at human resting state.
“Figure 6c demonstrates the output signal of the pulsewave sensor is correlated well with the amplitude of 40 mmHg and the frequency of 50 BPM of input pressure pulse equivalent to the arterial pulsewave’s pressure difference amplitude and frequency at human resting state.”
Quote in the paperSubstantial claim - Findingown data
The time delay of the pulsewave sensor is maintained as 70.00 ± 7.42 msec.
“The time delay of the pulsewave sensor (see Fig. 7a) is maintained as 70.00 ± 7.42 msec.”
Quote in the paperSubstantial claim - Findingown data
For a free end load of 30 μN, the bending stiffness of Type A (with windows) and Type B (without windows) are 89.21 μNm2 and 95.51 μNm2, respectively.
“For P of 30 μN, the maximum deflections of Type A and Type B are estimated from COMSOL Multiphysics® and the bending stiffness of Type A and Type B are obtained by Eq. (1) as 89.21 μNm2 and 95.51 μNm2, respectively.”
Quote in the paperSubstantial claim - Findingown data
The pulsewave sensor sensitivities at 50 BPM, 135 BPM, and 220 BPM are 0.35 mV/mmHg, 0.38 mV/mmHg, and 0.33 mV/mmHg, respectively.
“The pulsewave sensor sensitivities (see Fig. 7b) at different frequencies of 50 BPM, 135 BPM, and 220 BPM are 0.35 mV/mmHg, 0.38 mV/mmHg, and 0.33 mV/mmHg, respectively.”
Quote in the paperNot a substantial claim - Findingown data
The skin temperature sensor of the patch has less than 0.5 msec of time constant, theoretically.
“The skin temperature sensor of the patch has less than 0.5 msec of time constant, theoretically.”
Quote in the paperSubstantial claim - Findingnone given
In the range of environment temperature (0-30 °C), the temperature of the skin temperature sensor on the patch reaches body temperature within maximum 52.5 msec.
“In the range of environment temperature (0–30 °C), the temperature of the skin temperature sensor on the patch reach to the body temperature within maximum 52.5 msec.”
Quote in the paperNot a substantial claim - Findingreasoning
Considering lifetimes of each sensor integrated in the patch, the patch lifetime is estimated as 9 days.
“Therefore, considering life times of each sensor integrated in the patch, the patch life time is estimated as 9 days.”
Quote in the paperSubstantial claim- Rests on an assumption: The overall lifetime of the patch is determined by the shortest-lived integrated sensor.
- Findingcitation
The skin contact area of the fabricated multi-layer patch (3.71 cm2) is as small as 1/125 of that of the conventional single layer multimodal stress monitoring device (451 cm2).
“As a result, the skin contact area of the fabricated multi-layer patch (3.71 cm2) is as small as the 1/125 of that of the conventional single layer multimodal stress monitoring device (451 cm2)28.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [28] Choi, J., Ahmed, B. & Gutierrez-Osuna, R. “Development and evaluation of an ambulatory stress monitor based on wearable sensors”, IEEE T. Inf. Technol. Biomed. 16(2), 279–286 (2012). source
Not checked in the free run.
- Findingreasoning
The skin temperature sensor and skin conductance sensor lifetime is estimated in the range of 8-100 months, considering the 0.5 μm-thick electrodes.
“Thus, the skin temperature sensor and the skin conductance sensor life time is estimated in the range of 8–100 months, considering the 0.5 μm-thick electrodes.”
Quote in the paperSubstantial claim - Findingnone given
The windows on the polyimide membrane reduce the patch's bending stiffness by 6.60%, theoretically.
“For additional flexibility enhancement, we design the windows on the polyimide membrane, thus reducing the patch’s bending stiffness in the amount of 6.60%, theoretically.”
Quote in the paperSubstantial claim - Findingreasoning
In the fabrication process of the stress monitoring patch increasing wearing comfort, windows formation by patterning polyimide is a key process to increase the piezoelectric sensitivity of the pulsewave sensor with the patch flexibility.
“In the fabrication process of the stress monitoring patch increasing wearing comfort, the windows formation by patterning polyimide is a key process to increase the piezoelectric sensitivity of the pulsewave sensor with the patch flexibility.”
Quote in the paperNot a substantial claim - Findingown data
The individual sensors of the patch have proper performances for the comprehensive analysis of human physiological stress.
“We experimentally show the individual sensors of the patch have proper performances for the comprehensive analysis of human physiological stress.”
Quote in the paperSubstantial claim - Findingreasoning
The windows fabrication on the polyimide support membrane enables the electric contact of the silver inter-layer electrode sandwiched between the polyimide and the P(VDF-TrFE) membranes.
“The windows fabrication on the polyimide support membrane, also, enables the electric contact of the silver inter-layer electrode sandwiched between the polyimide and the P(VDF-TrFE) membranes.”
Quote in the paperSubstantial claim - Findingown data
A 21-minute dipping into DI water is proper to remove the residue of the 50 um-thick polyimide support membrane after conventional wet etching.
“We experimentally find 21-minute dipping into DI water is proper to remove the residue of the 50 um-thick polyimide support membrane after the conventional wet etching.”
Quote in the paperSubstantial claim - Interpretationnone given
The patch flexibility is increased mainly by the development of a flexible pulsewave sensor made of a flexible piezoelectric membrane supported by a perforated polyimide membrane.
“The patch flexibility is increased mainly by the development of flexible pulsewave sensor, made of a flexible piezoelectric membrane supported by a perforated polyimide membrane.”
Quote in the paperSubstantial claim - Premisecitation
Physiological data for stress monitoring are responses of the human autonomous nervous system (ANS).
“Based on the previous studies, it is proved that the physiological data for stress monitoring are responses of human autonomous nervous system (ANS)10.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [10] Carlson, N. Foundation of physiological psychology 3rd edn Trans. Kim, H. et al. Ch. 11, 459–466 (Sigmapress, 1998).
Not checked in the free run.
- Interpretationreasoning
The skin-attachable stress patch shows potential for application to wearable emotion monitoring.
“The skin-attachable stress patch, capable to detect multimodal bio-signals, shows potential for application to wearable emotion monitoring.”
Quote in the paperSubstantial claim - Premisecitation
The self-report method is hard to monitor human stress consistently due to the lack of standards for stress status.
“The self-report method is hard to monitor human stress consistently due to the lack of standards for stress status7.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [7] Lee,J., Hwang, Y., Cheon, K. & Jung, H. “Emotion-on-a-chip (EOC): evolution of bio chip technology to measure human emotion using body fluids”, Med. Hypotheses 79, 827–832 (2012). source
Not checked in the free run.
- Interpretationreasoning
The multimodal physiological data analysis is the suitable method to monitor physiological stress in daily life because it measures psychological stress consistently and continuously.
“However, the multimodal physiological data analysis is the suitable method to monitor physiological stress in daily life because it measures psychological stress consistently and continuously.”
Quote in the paperSubstantial claim- Rests on an assumption: Body fluid analysis cannot measure stress continuously.
- Premisecitation
Among ANS responses, skin temperature, skin conductance, and arterial pulsewave signals are required for comprehensive multimodal physiological data analysis.
“Among ANS responses, the skin temperature1126, skin conductance112628, and arterial pulsewave11232628 signals are required for the comprehensive multimodal physiological data analysis.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [11] Kreibig, S., Wilhelm, F., Roth, W. & Gross, J. “Cardiovascular, electrodermal, and respiratory response patterns to fear- and sadness-inducing films”, Psychophysiology 44, 787–806 (2007). source
Not checked in the free run.
- cites [26] Quazi, M., Mukhopadhyay, S., Suryadevara, N. & Huang, Y. “Towards the smart sensors based human emotion recognition”, Instrumentation and Measurement Technology Conference (I2MTC), Graz. IEEE. (doi: 1109/12MTC.2012.6229646)(2012, May 13–16). source
Not checked in the free run.
- cites [28] Choi, J., Ahmed, B. & Gutierrez-Osuna, R. “Development and evaluation of an ambulatory stress monitor based on wearable sensors”, IEEE T. Inf. Technol. Biomed. 16(2), 279–286 (2012). source
Not checked in the free run.
- Interpretationreasoning
ANS signal analysis enables the various kinds of stress analysis.
“Therefore ANS signal analysis enables the various kinds of the stress analysis.”
Quote in the paperSubstantial claim - Premisecitation
The ANS is excited by various kinds of stressors such as audial, visual, thermal, working, and exercising stimulations.
“The ANS is excited by various kinds of stressors (origins of the stress), such as audial32, visual33, thermal3435, working836, and exercising3738 stimulations.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [32] Tanev, G., Saadi, D., Hoppe, K. & Sorensen, H. “Classification of acute stress using linear and non-linear heart rate variability ananlysis derived from sternal ecg”, Conf. Proc. IEEE Eng. Med. Biol. Soc. Chicago. IEEE. (doi: 10.1109/EMBC.2014.6944349)(2014, August, 26–30).
Not checked in the free run.
- cites [33] Fernandez, C. et al. “Physiological responses induced by emotion-eliciting films”, Appl. Psychophysiol. Biofeedback 37, 73–79 (2012), doi: 10.1007/s10484-012-90180-7. source
Not checked in the free run.
- cites [34] Atterhog, J., Eliasson, K. & Hjemdahl, P. “Sympathoadrenal andcardiovascular responses to mental stress, isometric handgrip, and cold pressor test in asymptomatic young men with primary t wave abnormalities in the electrocardiogram”, Br. Heart J. 46, 311–319 (1981). source
Not checked in the free run.
- Premisecitation
Peripheral skin temperature shows a negative relationship with the chronic stress level.
“Peripheral skin temperature shows negative relationship2340 with the chronic stress level.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [23] Lee, Y., Lee, B. & Lee, M. “Wearable sensor glove based on conducting fabric using electrodermal activity and pulse-wave sensors for e-health application”, Telemed. E-Health 16(2), 209–217 (2010). source
Not checked in the free run.
- cites [40] Msc, M., Yamasaki, M., Sasaki, T. & Nakayama, H. “Fall in skin temperature of exercising man”, Br. J. Sports Med. 26(1), 29–32 (1992). source
Not checked in the free run.
- Premisecitation
A watch-type daily time stress monitoring device, though having good wearing comfort, is not able to perform comprehensive stress analysis by measuring only skin conductance.
“On the other hand, though good wearing comfort, a watch type daily time stress monitoring device31 is not able to comprehensive stress analysis by measuring only skin conductance.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [31] Poh, M., Swenson, N. & Picard, R. “A wearable sensor for unobtrusive, long-term assessment of electrodermal activity”, IEEE Trans. Biomed. Eng. 57(5), 1243–1252 (2010). source
Not checked in the free run.
- Interpretationreasoning
The flexibility of the pulsewave sensor is enhanced by replacing the rigid piezoelectric material and support with the flexible piezoelectric material P(VDF-TrFE) on a flexible polyimide support.
“In this study, we enhance the flexibility of the pulsewave sensor by replacing the rigid piezoelectric material and support42 by the flexible flexible piezoelectric material of Poly(vinylidene fluoride-trifluoroethylene) [P(VDF-TrFE)]4344454647 on the flexible polyimide4849505152 support.”
Quote in the paperSubstantial claim - Premisecitation
A glove-type device uses multiple hard sensors assembled on a single layer, resulting in poor wearing comfort due to low flexibility and large skin contact area.
“For example, a glove-type device23 uses the multiple hard sensors assembled on a single layer, resulting in a poor wearing comfort due to low flexibility and large skin contact area.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Lee, Y., Lee, B. & Lee, M. “Wearable sensor glove based on conducting fabric using electrodermal activity and pulse-wave sensors for e-health application”, Telemed. E-Health 16(2), 209–217 (2010). source
Not checked in the free run.
- Interpretationnone given
The skin contact area minimization is achieved by the multi-layer integration of the skin temperature sensor and the skin conductance sensor on the patch-type pulsewave sensor.
“The skin contact area minimization is achieved by the multi-layer integration of the skin temperature sensor and the skin conductance sensor on the patch type pulsewave sensor.”
Quote in the paperSubstantial claim - Premisecitation
Conventional stress monitoring devices using commercial physiological sensors have bulky size to be worn or carried for daily time use.
“The conventional researches about stress monitoring devices are focused on designing completed systems using commercial physiological sensors; thereby having bulky size41 to be worn or carried for daily time use.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [41] Kim, D., Kim, J., Lee, E., Whang, M. & Cho, Y. “Interactive emotional content communications system using portable wireless bifeedback device”, IEEE Trans. Consum. Electron. 57(4), 1929–1935 (2011). source
Not checked in the free run.
- Interpretationreasoning
The windows on the polyimide support provide an additional effect of flexibility increase, desirable for conformal skin contact.
“The windows on the polyimide support provides the additional effect of flexibility increase, desirable for conformal skin contact.”
Quote in the paperNot a substantial claim - Interpretationnone given
Forming windows on the polyimide support membrane underneath the P(VDF-TrFE) membrane induces higher mechanical stress on the piezoelectric material for higher pulsewave sensing sensitivity.
“In the present patch, we form windows on the polyimide support membrane underneath the P(VDF-TrFE) membrane to induce the higher mechanical stress on the piezoelectric material for the higher pulsewave sensing sensitivity.”
Quote in the paperNot a substantial claim - Premisecitation
Human pulsewave systolic-diastolic pressure difference varies in the range of 40-120 mmHg at a frequency of 50-200 BPM between resting and exercising states.
“Human pulsewave is strongly depending on the Human states, such that the systolic-diastolic pressure difference is varying in the range of 40–120 mmHg at the frequency of 50–200 BPM (0.83–3.66 Hz)54 between Human resting state and exercising state.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [54] Platini, P. “Blood pressure behaviour during physical activity”, Sports Med. 5(6), 353–374 (1988). source
Not checked in the free run.
- Premisecitation
Polyimide is used as the support membrane due to high flexibility and high chemical resistance.
“We use the polyimide as the support membrane not only due to high flexibility but also due to high chemical resistance40.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [40] Msc, M., Yamasaki, M., Sasaki, T. & Nakayama, H. “Fall in skin temperature of exercising man”, Br. J. Sports Med. 26(1), 29–32 (1992). source
Not checked in the free run.
- Premisecitation
Wearing comfort of wearable devices is decided by the number of wearing devices, device size and weight, and flexibility of the device.
“The wearing comfort575859 of wearable devices is decided by; 1) the number of wearing devices; 2) device’s size and weight; 3) flexibility of the device.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [57] Knight, J. F. & Baber, C. “A tool to assess the comfort of wearable computers”, Hum. Factors 47(1), 77–91 (2005). source
Not checked in the free run.
- cites [58] Cho, G., Lee, S. & Cho, J. “Review and reappraisal of smart clothing”, Int. H. Hum.-Comput. Interact. 25(6), 582–617 (2009). source
Not checked in the free run.
- cites [59] Chuo, Y. et al. B. “Mechanically flexible wireless multisensor platform for human physical activity and vitals monitoring”, IEEE Trans. Biomed. Circuits Syst. 4(5), 281–294 (2010). source
Not checked in the free run.
- Interpretationown data
All experimentally verified performances of the patch demonstrate feasible performance for human stress monitoring applications.
“All the experimentally verified performances of the patch are listed in Table 1, demonstrating feasible performance for human stress monitoring applications.”
Quote in the paperSubstantial claim - Premisecitation
Heart rate variability can be estimated from the pulsewave signal by significantly reducing the influence of motion artifacts.
“HRV is able to be estimated from the pulsewave signal62 by significantly reducing the influence of motion artifacts.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [62] Lee, C. & Zhang, Y. “Reduction of motion artifacts from photoplethysmographic recordings using a wavelet demoising approach”, Biomedical Engineering, IEEE EMBS Asian-Pacific Conference on, Kyoto. IEEE. (doi: 10.1109/APBME.2003.1302650)(2003, October, 20–22).
Not checked in the free run.
- Premisecitation
Aluminum corrosion rate changes in the range of 0.06-0.7 μm/year according to surrounding environment.
“Aluminum corrosion rate60 changes in the range of 0.06–0.7 μm/year according to surrounding environment.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [60] Akhmedov, G., Shakov, V., Trifel, M. & Khanlarova, A. “Corrosion resistance of aluminum alloys used in hydraulic construction”, Hydrotechnical Construction 2(3), 215–218 (1968). source
Not checked in the free run.
- Premisecitation
Motion artifacts on the pulsewave signal can be removed by signal processing without using additional motion sensors.
“The motion artifacts on pulsewave signal is able to be removed by signal processing62636465 without using additional motion sensors.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites 62636465
Not checked in the free run.
- Premisecitation
The retention time of P(VDF-TrFE) fabricated in the same method as the present patch is known as 0.8 × 106 seconds.
“The retention time of P(VDF-TrFE) fabricated in a same method of the present patch is known as 0.8 × 106 seconds61.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [61] Choi, H., Hong, S., Sung, T. & No, K. “Effects of surface morphology on retention loss of ferroelectric domains in poly(vinylidenefluoride-co-trifluoroethylene) thin films”, Appl. Phys. Lett., 99, 092905, doi: 10.1063/1.3632042 (2011). source
Not checked in the free run.
- Interpretationreasoning
The patch enhances wearing comfort by the multi-layer integration of the three sensors and the windows fabrication on its support membrane.
“In conclusion, the patch enhances wearing comfort by the multi-layer integration of the three sensors and the windows fabrication on its support membrane.”
Quote in the paperSubstantial claim- Rests on an assumption: The objective criteria of wearing comfort (fewer devices, smaller/lighter size, higher flexibility) are sufficient to establish enhanced wearing comfort without subjective comfort testing.
- Premisecitation
The pulsewave sensor lifetime is decided by retention time of the piezoelectric P(VDF-TrFE).
“The pulsewave sensor life time is decided by retention time of the piezoelectric P(VDF-TrFE).”
Quote in the paperSubstantial claim - Interpretationreasoning
The present patch is able to measure pulsewave for human stress monitoring with additional signal processing.
“Therefore, the present patch is able to measure pulsewave for human stress monitoring with additional signal processing.”
Quote in the paperSubstantial claim- Rests on an assumption: Muscle contractions can interfere with the pulsewave sensor, producing motion artifacts.
- Interpretationown data
The present patch is weakly affected by external environment changes such as temperature and muscle contractions.
“The present patch is weakly affected by external environment changes such as temperature and muscle contractions.”
Quote in the paperSubstantial claim - Interpretationreasoning
The present patch is weakly affected by environmental temperature change with a period over 52.5 msec.
“Therefore the present patch is weekly affected by the environmental temperature change period over 52.5 msec.”
Quote in the paperSubstantial claim - Premisenone given
The skin temperature sensor and skin conductance sensor lifetime is related to aluminum corrosion time.
“The skin temperature sensor and the skin conductance sensor life time is related with aluminum corrosion time.”
Quote in the paperSubstantial claim - Interpretationreasoning
The patch with enhanced wearing comfort has potential to be a human emotion sensor for daily use.
“Therefore, the patch enhanced wearing comfort has potentials to human emotion sensor for daily use.”
Quote in the paperSubstantial claim - Interpretationreasoning
The small size (24.7 mm × 15.0 mm × 70 μm) and light weight (49.5 mg) of the stress monitoring patch with improved wearing comfort have potential for multimodal bio-signal monitoring applications and emotion sensing applications using mobile electronics and wearable devices.
“The small size (24.7 mm × 15.0 mm × 70 μm) and light weight (49.5 mg) of the stress monitoring patch with improved wearing comfort has potentials for multimodal bio-signal monitoring applications and emotion sensing applications using mobile electronics and wearable devices.”
Quote in the paperSubstantial claim - Interpretationreasoning
The integrated patch enables continuous measurement and comprehensive analysis of human psychological stress by quantitative and continuous measurement of skin temperature, skin conductance, and pulsewave in the human physiological range.
“The integrated patch enables the continuous measurement and the comprehensive analysis of human psychological stress by quantitative and continuous measurement of skin temperature, skin conductance, and pulsewave in the human physiological range.”
Quote in the paperSubstantial claim - Premisecitation
The patch categorizes the four kinds of human emotions (surprise, anger, stress, and sadness) based on the singular vector machine (SVM) algorithm.
“The patch categorizes the 4 kinds of human emotions (surprise, anger, stress, and sadness) based on singular vector machine (SVM) algorithm17.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17] Kim, K., Bang, S. & Kim, S. “Emotion recognition system using short-term monitoring of physiological signals”, Med. Biol. Comput. 42, 419–427 (2004). source
Not checked in the free run.
- May combine two claims. 25
- Classified as probably not a substantial claim. 12
May combine two claims.
FindingreasoningA human stress monitoring patch integrates three sensors of skin temperature, skin conductance, and pulsewave in the size of a stamp (25 mm × 15 mm × 72 μm) to enhance wearing comfort with small skin contact area and high flexibility.
“A human stress monitoring patch integrates three sensors of skin temperature, skin conductance, and pulsewave in the size of stamp (25 mm × 15 mm × 72 μm) in order to enhance wearing comfort with small skin contact area and high flexibility.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationThe silver electrode layer is designed to perform triple functions: an etch stop for protecting the piezoelectric membrane during polyimide window etching, an electrode layer for piezoelectric poling, and an electrode layer for sensing pulsewave signals.
“The silver electrode layer is specially designed to perform triple functions: 1) an etch stop for protecting the piezoelectric membrane during the polyimide window etching process using KOH, Ethanol, and DI water (1:8:2) mixture52, 2) an electrode layer for piezoelectric poling process; 3) an electrode layer for sensing the pulsewave signals generated by piezoelectric material.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [52] Choi, W. “Polymer micromachined flexible tactile sensor for three-axial loads detection”, Trans. Electr. Electron. Mater. 11(3), 130–133 (2010). source
Not checked in the free run.
May combine two claims.
FindingcitationThe window of the pulsewave sensor is designed as 6 mm × 6 mm, capable of covering the radial artery sufficiently by having twice larger width than the radial artery diameter of 2.19-3.01 mm.
“The window of the pulsewave sensor is designed as 6 mm × 6 mm, capable to cover the radial artery sufficiently by having twice larger width than the radial artery diameter of 2.19–3.01 mm55.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [55] Yoo, B. et al. “Anatomical consideration of the radial artery for transradial coronary procedures: arterial diameter, branching anomaly and vessel tortuosity”, Int. J. Cardiol. 101, 421–427 (2005). source
Not checked in the free run.
May combine two claims.
FindingreasoningThe overall size of the flexible human stress monitoring patch is 25 mm × 15 mm × 72 μm, with a skin contact area of 3.75 cm2.
“Therefore, the overall size of the flexible human stress monitoring patch (see Fig. 4) is decided as 25 mm × 15 mm × 72 μm (see Fig. 3), having the skin contact area of 3.75 cm2.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe skin conductance sensor has a nonlinearity of 0.447%, a resolution of 0.02 μS, and a sensitivity of 0.28 μV/0.02 μS within the human skin conductance range.
“The skin conductance sensor has showed the nonlinearity, resolution, and sensitivity (see Fig. 6b) as 0.447%, 0.02 μS, and 0.28 μV/0.02 μS, respectively, within the human skin conductance range.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe skin temperature sensor has a nonlinearity of 0.021% and a sensitivity of 0.31 Ω/°C.
“The skin temperature sensor nonlinearity and sensitivity (see Fig. 6a) are measured as 0.021% and 0.31 Ω/°C, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataType A with windows shows a 6.60% increase in patch flexibility and a 57% increase in pulsewave sensing sensitivity compared to Type B without windows.
“The theoretical results indicate that Type A with window shows the 6.60% increase of the patch flexibility and 57% increase of the pulsewave sensing sensitivity compared to Type B without window.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe pulsewave sensing sensitivities of Type A and Type B are 0.373 mV/mmHg and 0.237 mV/mmHg, respectively.
“The pulsewave sensing sensitivities of Type A and Type B are shown as 0.373 mV/mmHg and 0.237 mV/mmHg, respectively.”
Quote in the paperSubstantial claimMay combine two claims.
FindingreasoningEach skin conductance electrode is designed as 7.5 mm × 8.5 mm to have a skin contact area of 63.75 mm2, larger than the single electrode area (6.36 mm2) of a commercial skin conductance sensor (HK-228s, iWorx).
“Thus each electrode is designed as 7.5 mm × 8.5 mm to have skin contact area of 63.75 mm2, larger than the single electrode area (6.36 mm2) of a commercial skin conductance sensor (HK-228s, iWorx).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn all cases, the temperature of the skin temperature sensor is saturated at the skin temperature, therefore being dominantly affected not by external temperature but by skin temperature.
“In all the cases, the temperature of the skin temperature sensor is saturated at the skin temperature, therefore being dominantly affected not by external temperature but by skin temperature.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe fabricated stress monitoring patch is as small as 24.7 mm × 15.0 mm × 70 μm with a skin contact area of 3.71 cm2 and a weight of 49.5 mg.
“The fabricated stress monitoring patch is as small as 24.7 mm × 15.0 mm × 70 μm with the skin contact area of 3.71 cm2 and the weight of 49.5 mg.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Findingnone givenThe wearing comfort of the patch is enhanced by the windows fabrication and the multi-layer integration using the silver inter-layer electrode.
“The wearing comfort of the patch is enhanced by the windows fabrication and the multi-layer integration using the silver inter-layer electrode.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationStress leads to many kinds of diseases such as shock, cardiovascular diseases, neuropsychiatric diseases, gastro intestinal diseases, and even tumors.
“Also, stress leads many kinds of diseases4 such as shock, cardiovascular diseases, neuropsychiatric diseases, gastro intestinal diseases, and even tumors.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [4] Hans, S. Stress in health and disease Ch. 4, 725–895 (Butterworth-Heinemann, 1976).
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationStress affects human society by increasing suicide and reducing work efficiency.
“Stress affects human society by increasing suicide1 and reducing work efficiency2.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [1] Mann, J. & Currier, D. “Stress, genetics and epigenetic effects on the neurobiology of suicidal behavior and depression”, Eur. Psychiat. 25(5), 268–271 (2010). source
Not checked in the free run.
- cites [2] Macan, T., Shahani, C., Dipboye, R. & Philips, A. “College students’ time management: correlations with academic performance and stress”, J. Educ. Psychol. 82(4), 760–768 (1990). source
Not checked in the free run.
May combine two claims.
PremisecitationThe stress signals of skin temperature, skin conductance, and pulsewave are dependent on the stimulation time and categorized into acute stress (response under 3 seconds stimulation) and chronic stress (response over 3 seconds stimulation).
“The stress signals of skin temperature, skin conductance, and pulsewave are dependent on the stimulation time and categorized in two types: acute stress (response under 3 seconds stimulation)11 and chronic stress (response over 3 seconds stimulation)39.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [11] Kreibig, S., Wilhelm, F., Roth, W. & Gross, J. “Cardiovascular, electrodermal, and respiratory response patterns to fear- and sadness-inducing films”, Psychophysiology 44, 787–806 (2007). source
Not checked in the free run.
- cites [39] Gil, E. et al. “Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions”, Physiol. Meas. 31, 1271–1290 (2010). source
Not checked in the free run.
May combine two claims.
PremisecitationSkin conductance on palm and volar part of the wrist has a positive relationship with both chronic and acute stress levels.
“Skin conductance on palm and volar part of the wrist has positive relationship11 with both of the chronic and the acute stress levels.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [11] Kreibig, S., Wilhelm, F., Roth, W. & Gross, J. “Cardiovascular, electrodermal, and respiratory response patterns to fear- and sadness-inducing films”, Psychophysiology 44, 787–806 (2007). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
Premisenone givenThe low flexibility of conventional sensors is mainly due to the hardness of the pulsewave sensor.
“The low flexibility of the conventional sensors is mainly due to the hardness of pulsewave sensor.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
PremisecitationFor stress characterization, arterial pulsewave signals are required to be transformed into human heart rate variability (HRV).
“For stress characterization, the arterial pulsewave signals is required to be transformed into the human heart rate variability (HRV)39.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [39] Gil, E. et al. “Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions”, Physiol. Meas. 31, 1271–1290 (2010). source
Not checked in the free run.
May combine two claims.
PremisecitationHRV represents the chronic stress level and also individual's stress vulnerability.
“The HRV represents the chronic stress level10 and also individual’s stress vulnerability12.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [10] Carlson, N. Foundation of physiological psychology 3rd edn Trans. Kim, H. et al. Ch. 11, 459–466 (Sigmapress, 1998).
Not checked in the free run.
- cites [12] Porges, S. “Vagal tone: a physiologic marker of stress vulnerability”, Pediatrics 90(3), 498–504 (1992). source
Not checked in the free run.
May combine two claims.
PremisecitationThe human peripheral skin temperature and skin conductance change in the range of 30-40 °C and 2-20 μS, respectively.
“The human peripheral skin temperature and skin conductance are changed in the range of 30–40 °C53 and 2–20 μS9, respectively.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [53] Sim, J., Youn, S. & Cho, Y. “A thermal peripheral blood flowmeter with contact force compensation”, J. Micromech. Microeng. 22, 125014, doi: 10.1088/0960-1317/22/12/125014 (2012). source
Not checked in the free run.
- cites [9] Cacioppo, J. Handbook of psychophysiology 3rd edn (eds Tassinary, L. et al.) Ch. 7, 159–181 (Cambridge, New York, 2007).
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe flexible multi-layer patch integrated with the skin temperature sensor and skin conductance sensor on the pulsewave sensor results in skin contact area reduction and patch flexibility increase, thereby enhancing wearing comfort.
“The flexible multi-layer patch (see Fig. 1), integrated with the skin temperature sensor, the skin conductance sensor on the pulsewave sensor, results in skin contact area reduction and patch flexibility increase; thereby enhancing the wearing comfort, required for wearable or skin-attachable stress monitoring by human ANS monitoring.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationP(VDF-TrFE) is used as a flexible piezoelectric material with high mechanical strength and biocompatibility.
“We use P(VDF-TrFE) as flexible piezoelectric materials with high mechanical strength and biocompatibility43.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [43] Li, C. et al. “Flexible dome and bump shape piezoelectric tactile sensors using PVDF-TrFE copolymer”, J. Microelectromech. Syst. 17(2), 334–341 (2008). source
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe fabricated pulsewave sensor integrated in the patch is suitable for human stress monitoring with HRV, though its experimental output voltage and sensitivity are lower than the simulation result.
“Therefore the fabricated pulsewave sensor integrated in the patch is suitable for human stress monitoring with HRV, though the experimental output voltage and sensitivity of the pulsewave sensor are lower than that of the simulation result.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe present patch satisfies the standards of small size, low number of devices, and high flexibility, even though the present study did not perform a subjective comfort test.
“The present patch satisfies not only small but also other two standards even though the present study didn’t perform subjective comfortable test.”
Quote in the paperSubstantial claim- Rests on an assumption: The objective criteria of wearing comfort (fewer devices, smaller/lighter size, higher flexibility) are sufficient to establish enhanced wearing comfort without subjective comfort testing.
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
InterpretationreasoningPeople feel wearable devices are comfortable when they wear fewer devices, small and light devices, and highly flexible devices.
“In other words, people feel the wearable devices are comfortable when they wearing the less number of devices, small and light devices, and highly flexible devices.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningIn stress measurement, the time delay is a more important factor than output voltage amplitude or sensitivity because HRV is directly related to the time interval between peaks of the pulsewave sensor output signal, not amplitude or sensitivity.
“In the stress measurement application, the time delay is more important factor than the output voltage amplitude or sensitivity because HRV is directly related with time interval between peaks of the output voltage signal of the pulsewave sensor, not amplitude or sensitivity.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe present patch composed of three sensors integrated in a single multi-layer structure reduces the number of devices and the skin contact area of the patch.
“The present patch composed of the three sensors are integrated in a single multi-layer structure to reduce the number of devices and the skin contact area of the patch.”
Quote in the paperSubstantial claimMay combine two claims.
Premisenone givenStress monitoring ability in daily time is decided by not only wearing comfort but also lifetime of the patch.
“Stress monitoring ability in daily time is decided by not only a wearing comfort but also a life time of the patch.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
InterpretationreasoningThe present patch has applications for not only patients but also non-patients by monitoring the autonomic nervous system.
“The present patch has applications for not only patients but also non-patients by monitoring ANS.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
InterpretationreasoningFor patients, the present patch can alert or notify doctors of potential risks related to the autonomic nervous system for medical patient monitoring, using communication modules of wearable devices.
“For patients, the present patch can alert or notice the potential risks related with ANS to doctors for medical patients monitoring, using communication modules of the wearable devices.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThe novelty of the present patch is compact size and high flexibility, so it can be easily built into commercial wearable and mobile devices.
“The novelty of the present patch is compact size and high flexibility; thereby, easily built into the commercial wearable and mobile devices.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisecitationConventional polyimide patterning processes use reactive ion etching or chemical etching with the polyimide etchant of KOH-ethanol-DI water mixture.
“The conventional polyimide patterning processes use reactive ion etching (RIE)66 or chemical etching process5152 with the polyimide etchant of KOH-ethanol-DI water mixture.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [66] J. Engel, J. Chen, Z. Fan & C. Liu “Polymer micromachined multimodal tactile sensors”, Sens. Actuator A-Phys. 117(1), 50–61 (2004). source
Not checked in the free run.
- cites [51] Kwon, H., Kim, J. & Choi, W. “Development of a flexible three-axial tactile sensor aray for a roboric finger”, Microsyst. Technol. 17, 1721–1726 (2011). source
Not checked in the free run.
- cites [52] Choi, W. “Polymer micromachined flexible tactile sensor for three-axial loads detection”, Trans. Electr. Electron. Mater. 11(3), 130–133 (2010). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
InterpretationreasoningThe present patch alerts emergency situations such as heart attack to a rescue team.
“Also, the present patch alert the emergency situations like heart attack to rescue team.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based).
InterpretationreasoningFor non-patients, the present patch can provide wellness information required by individual users.
“For non-patients, the present patch can provide wellness information required to individual users.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
InterpretationreasoningThe application of the patch can be expandable to sense certain dimensions of human emotion.
“Moreover, the application of the patch can be expandable to sense certain dimensions of human emotion.”
Quote in the paperSubstantial claim
The paper relies on this without support.
Body fluid analysis cannot measure stress continuously.
“The body fluid analysis cannot measure stress continuously.”
The paper relies on this without support.
The objective criteria of wearing comfort (fewer devices, smaller/lighter size, higher flexibility) are sufficient to establish enhanced wearing comfort without subjective comfort testing.
“The present patch satisfies not only small but also other two standards even though the present study didn’t perform subjective comfortable test.”
The paper relies on this without support.
The overall lifetime of the patch is determined by the shortest-lived integrated sensor.
“Therefore, considering life times of each sensor integrated in the patch, the patch life time is estimated as 9 days.”
The paper relies on this without support.
Muscle contractions can interfere with the pulsewave sensor, producing motion artifacts.
“Also, muscle-contractions can interfere to pulsewave sensor, occurring motion artifacts.”
The paper relies on this without support.
The conventional reactive ion etching polyimide etching process causes thermal damages.
“The conventional RIE polyimide etching process causes thermal damages.”
The paper relies on this without support.
The conventional chemical etching process leaves highly viscous polyimide residue that hinders electrical contacts to the silver electrode through the patterned window.
“The conventional chemical etching process leaves highly viscous polyimide residue which hinders electrical contacts to the silver electrode through the patterned window.”
The paper relies on this without support.
Removing the viscous polyimide residue requires an additional boiling or rinsing process in a DI water stream, which causes layer delamination.
“The viscous polyimide residue removal requires an additional boiling or rinsing process in DI water stream, which causes layer delamination.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
There are three primary methods to monitor stress levels for athletes: self-reporting, multimodal physiological analysis, and body-fluid analysis.
“There are three primary methods to monitor stress levels for athletes: (1) self-reporting (current method), (2) multimodal physiological analysis, and (3) body-fluid analysis74 (Table 6).”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [74]
An epidermal sensor for multimodal physiological data analysis of stress measured skin temperature with a sensitivity of 0.31 Ω/°C, skin conductance sensitivity of 0.28 μV/0.02 μS, and a pulse wave response time of 70 ms.
“The stress patch measured skin temperature with a sensitivity of 0.31 Ω/°C, skin conductance sensitivity of 0.28 μV/0.02 μS, and a pulse wave response time of 70 ms.74”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [74]
The stress patch categorized four types of human emotions (surprise, anger, stress, and sadness) based on a singular vector machine algorithm.
“The patch categorized the four types of human emotions (surprise, anger, stress, and sadness) based on a singular vector machine algorithm.”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [74]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Upper pulse-rate conversion is incorrect
MinorA pulse rate of 200 BPM is 3.33 Hz, not 3.66 Hz. The stated upper frequency corresponds to about 220 BPM.
Reported 3.66 · recomputed 3.333333
“at the frequency of 50–200 BPM (0.83–3.66 Hz)”
Simulation pressures exceed the stated physiological range
MinorThe paper describes all five simulated pressures as within the human systolic-diastolic pressure-difference range, but two of them, 160 and 200 mmHg, exceed its stated upper limit of 120 mmHg.
“For pulsewave sensing sensitivity comparison, the piezoelectric output voltages of Type A and Type B are estimated for the five different pressures of 40, 80, 120, 160 and 200 mmHg within the range of pressure difference between human systolic and diastolic arterial pressure.”
“the systolic-diastolic pressure difference is varying in the range of 40–120 mmHg”
Claimed skin-contact-area ratio does not match the areas
MinorDividing the conventional device's stated 451 cm2 by the patch's 3.71 cm2 gives about 121.6, not 125.
Reported 125 · recomputed 121.563342
“the skin contact area of the fabricated multi-layer patch (3.71 cm2) is as small as the 1/125 of that of the conventional single layer multimodal stress monitoring device (451 cm2)28.”
Cites
Impact of Event Scale: A Measure of Subjective Stress
Horowitz, M., Wilner, N., & Alvarez, W. (1979) · Psychosomatic Medicine · doi:10.1097/00006842-197905000-00004 · Cited by 7,813 · see citing works
(2001) · doi:10.1017/cbo9780511546396 · Cited by 4,482 · see citing works
Detecting Stress During Real-World Driving Tasks Using Physiological Sensors
Healey, J. A., & Picard, R. W. (2005) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2005.848368 · Cited by 2,207 · see citing works
Emotion recognition system using short-term monitoring of physiological signals
Kim, K. H., Bang, S. W., & Kim, S. R. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02344719 · Cited by 944 · see citing works
College students' time management: Correlations with academic performance and stress.
Macan, T. H., Shahani, C., Dipboye, R. L., & Phillips, A. P. (1990) · Journal of Educational Psychology · doi:10.1037/0022-0663.82.4.760 · Cited by 933 · see citing works
The effect of mental stress on heart rate variability and blood pressure during computer work
Hjortskov, N., Rissén, D., Blangsted, A., Fallentin, N., Lundberg, U., & Søgaard, K. (2004) · European Journal of Applied Physiology · doi:10.1007/s00421-004-1055-z · Cited by 861 · see citing works
Effects of Work Stress on Ambulatory Blood Pressure, Heart Rate, and Heart Rate Variability
Vrijkotte, T. G. M., van Doornen, L. J. P., & de Geus, E. J. C. (2000) · Hypertension · doi:10.1161/01.hyp.35.4.880 · Cited by 756 · see citing works
A Wearable Sensor for Unobtrusive, Long-Term Assessment of Electrodermal Activity
Poh, M.-Z., Swenson, N. C., & Picard, R. W. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2038487 · Cited by 693 · see citing works
Vagal Tone: A Physiologic Marker of Stress Vulnerability
Porges, S. W. (1992) · Pediatrics · doi:10.1542/peds.90.3.498 · Cited by 541 · see citing works
Cardiovascular, electrodermal, and respiratory response patterns to fear‐ and sadness‐inducing films
Kreibig, S. D., Wilhelm, F. H., Roth, W. T., & Gross, J. J. (2007) · Psychophysiology · doi:10.1111/j.1469-8986.2007.00550.x · Cited by 495 · see citing works
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461 · see citing works
Emotional sweating across the body: Comparing 16 different skin conductance measurement locations
van Dooren, M., de Vries, G.-J., & Janssen, J. H. (2012) · Physiology & Behavior · doi:10.1016/j.physbeh.2012.01.020 · Cited by 413 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
Patterning piezoelectric thin film PVDF–TrFE based pressure sensor for catheter application
Sharma, T., Je, S.-S., Gill, B. S., & Zhang, J. X. J. (2011) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2011.08.019 · Cited by 317 · see citing works
Flexible Dome and Bump Shape Piezoelectric Tactile Sensors Using PVDF-TrFE Copolymer
Li, C., Wu, P.-M., Lee, S., Gorton, A., Schulz, M. J., & Ahn, C. H. (2008) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2007.911375 · Cited by 245 · see citing works
Yoo, B.-S., Yoon, J., Ko, J.-Y., Kim, J.-Y., Lee, S.-H., Hwang, S.-O., & Choe, K.-H. (2005) · International Journal of Cardiology · doi:10.1016/j.ijcard.2004.03.061 · Cited by 237 · see citing works
Mental stress as a causal factor in the development of hypertension and cardiovascular disease
Pickering, T. G. (2001) · Current Hypertension Reports · doi:10.1007/s11906-001-0047-1 · Cited by 235 · see citing works
Polymer micromachined multimodal tactile sensors
Engel, J., Chen, J., Fan, Z., & Liu, C. (2005) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2004.05.037 · Cited by 224 · see citing works
Development and Evaluation of an Ambulatory Stress Monitor Based on Wearable Sensors
Choi, J., Ahmed, B., & Gutierrez-Osuna, R. (2012) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2169804 · Cited by 221 · see citing works
Stress, genetics and epigenetic effects on the neurobiology of suicidal behavior and depression
Mann, J. J., & Currier, D. M. (2010) · European Psychiatry · doi:10.1016/j.eurpsy.2010.01.009 · Cited by 169 · see citing works
Instantaneous changes in heart rate regulation due to mental load in simulated office work
Taelman, J., Vandeput, S., Vlemincx, E., Spaepen, A. J., & Huffel, S. V. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1776-0 · Cited by 164 · see citing works
A tool to assess the comfort of wearable computers
Knight, J. F., & Baber, C. (2005) · Human Factors The Journal of the Human Factors and Ergonomics Society · doi:10.1518/0018720053653875 · Cited by 164 · see citing works
Fletcher, R., Dobson, K., Goodwin, M. S., Eydgahi, H., Wilder-Smith, O., Fernholz, D., Kuboyama, Y., Hedman, E. B., Poh, M., & Picard, R. W. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2038692 · Cited by 155 · see citing works
Vaillancourt, J., Zhang, H., Vasinajindakaw, P., et al. (2008) · Applied Physics Letters · doi:10.1063/1.3043682 · Cited by 154 · see citing works
Fall in skin temperature of exercising man.
Torii, M., Yamasaki, M., Sasaki, T., & Nakayama, H. (1992) · British Journal of Sports Medicine · doi:10.1136/bjsm.26.1.29 · Cited by 145 · see citing works
A novel fabrication process of MEMS devices on polyimide flexible substrates
Xiao, S., Che, L., Li, X., et al. (2007) · Microelectronic Engineering · doi:10.1016/j.mee.2007.08.004 · Cited by 144 · see citing works
Physiological Responses Induced by Emotion-Eliciting Films
Fernández, C., Pascual, J. C., Soler, J., Elices, M., Portella, M. J., & Fernández-Abascal, E. G. (2012) · Applied Psychophysiology and Biofeedback · doi:10.1007/s10484-012-9180-7 · Cited by 140 · see citing works
Ohigashi, H., Koga, K., Suzuki, M., et al. (1984) · Ferroelectrics · doi:10.1080/00150198408017527 · Cited by 137 · see citing works
Psychological issues of human-computer interaction in the work place
Frese, M. H., Ulich, E., & Dzida, W. (1988) · Medical Entomology and Zoology · Cited by 126 · see citing works
Lee, C. M., & Zhang, Y. (2004) · doi:10.1109/apbme.2003.1302650 · Cited by 114 · see citing works
Ark, W., Dryer, D. C., & Lu, D. J. (1999) · International Conference on Human-Computer Interaction · Cited by 114 · see citing works
Adams, P., Rabbi, M., Rahman, T., Matthews, M. L., Voida, A., Gay, G. K., Choudhury, T., & Voida, S. (2014) · International Conference on Pervasive Computing Technologies for Healthcare · doi:10.4108/icst.pervasivehealth.2014.254959 · Cited by 113 · see citing works
Flexible polyimide-based force sensor
Dobrzynska, J., & Gijs, M. A. M. (2011) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2011.11.006 · Cited by 110 · see citing works
Chuo, Y., Marzencki, M., Hung, B., Jaggernauth, C., Tavakolian, K., Lin, P., & Kaminska, B. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2052616 · Cited by 102 · see citing works
Review and Reappraisal of Smart Clothing
Cho, G., Lee, S., & Cho, J. (2009) · International Journal of Human-Computer Interaction · doi:10.1080/10447310902997744 · Cited by 93 · see citing works
Flexible Pressure Sensor Based on a Poly(VDF‐TrFE) Nanofiber Web
Ren, G., Cai, F., Li, B., Zheng, J., & Xu, C. (2012) · Macromolecular Materials and Engineering · doi:10.1002/mame.201200218 · Cited by 90 · see citing works
A review on stress inducement stimuli for assessing human stress using physiological signals
Karthikeyan, P., Murugappan, M., & Yaacob, S. (2011) · IEEE International Colloquium on Signal Processing and its Applications (CSPA) · doi:10.1109/cspa.2011.5759914 · Cited by 86 · see citing works
Motion Artifact Reduction in Photoplethysmographic Signals using Singular Value Decomposition
Reddy, K. A., & Kumar, V. J. (2007) · Conference proceedings - IEEE Instrumentation/Measurement Technology Conference · doi:10.1109/imtc.2007.379467 · Cited by 83 · see citing works
Blood Pressure Behaviour During Physical Activity
Palatini, P. (1988) · Sports Medicine · doi:10.2165/00007256-198805060-00002 · Cited by 77 · see citing works
Pervasive and Unobtrusive Emotion Sensing for Human Mental Health
Guo, R., Li, S., He, L., Gao, W., Qi, H., & Owens, G. P. (2013) · Proceedings of the ICTs for improving Patients Rehabilitation Research Techniques · doi:10.4108/icst.pervasivehealth.2013.252133 · Cited by 76 · see citing works
Flexible PZT Thin Film Tactile Sensor for Biomedical Monitoring
Tseng, H.-J., Tian, W., & Wu, W.-J. (2013) · Sensors · doi:10.3390/s130505478 · Cited by 70 · see citing works
Kushki, A., Fairley, J. A., Merja, S., King, G. A., & Chau, T. T. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/10/002 · Cited by 63 · see citing works
Atterhog, J. H., Eliasson, K., & Hjemdahl, P. (1981) · Heart · doi:10.1136/hrt.46.3.311 · Cited by 53 · see citing works
Knight, J. F., Schwirtz, A., Psomadelis, F., et al. (2004) · Personal and Ubiquitous Computing · doi:10.1007/s00779-004-0269-8 · Cited by 51 · see citing works
Towards the smart sensors based human emotion recognition
Quazi, M. T., Mukhopadhyay, S. C., Suryadevara, N. K., & Huang, Y. M. (2012) · 2012 IEEE International Instrumentation and Measurement Technology Conference Proceedings · doi:10.1109/i2mtc.2012.6229646 · Cited by 49 · see citing works
Polymer Micromachined Flexible Tactile Sensor for Three-Axial Loads Detection
Choi, W.-C. (2010) · Transactions on Electrical and Electronic Materials · doi:10.4313/teem.2010.11.3.130 · Cited by 44 · see citing works
Heart rate variability during high-intensity exercise
Sarmiento, S., Manso, J. M. G., González, J. M. M., et al. (2013) · Journal of Systems Science and Complexity · doi:10.1007/s11424-013-2287-y · Cited by 43 · see citing works
Body area network for monitoring autonomic nervous system responses
Brown, L., Grundlehner, B., van de Molengraft, J., et al. (2009) · doi:10.4108/icst.pervasivehealth2009.5973 · Cited by 40 · see citing works
ECG-based Mental Stress Assessment Using Fuzzy Computing and Associative Petri Net
Chiang, H. (2015) · Journal of Medical and Biological Engineering · doi:10.1007/s40846-015-0095-7 · Cited by 40 · see citing works
Massot, B., Baltenneck, N., Géhin, C., et al. (2011) · IEEE Sensors Journal · doi:10.1109/jsen.2011.2132703 · Cited by 37 · see citing works
Luminescence investigations on LiAl5O8:Tb3+ nanocrystalline phosphors
Pitale, S. S., Kumar, V., Nagpure, I. M., et al. (2010) · Current Applied Physics · doi:10.1016/j.cap.2010.08.002 · Cited by 36 · see citing works
Tanev, G., Saadi, D. B., Hoppe, K., & Sørensen, H. B. D. (2014) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2014.6944349 · Cited by 36 · see citing works
Stress Monitoring Based on Stochastic Fuzzy Analysis of Heartbeat Intervals
Kumar, M., Neubert, S., Behrendt, S., Rieger, A., Weippert, M., Stoll, N., Thurow, K., & Stoll, R. (2012) · IEEE Transactions on Fuzzy Systems · doi:10.1109/tfuzz.2012.2183602 · Cited by 35 · see citing works
Development of a flexible three-axial tactile sensor array for a robotic finger
Kwon, H.-J., Kim, J.-H., & Choi, W.-C. (2011) · Microsystem Technologies · doi:10.1007/s00542-011-1368-x · Cited by 24 · see citing works
Interactive emotional content communications system using portable wireless biofeedback device
Kim, D., Kim, J., Lee, E., Whang, M., & Cho, Y. (2011) · IEEE Transactions on Consumer Electronics · doi:10.1109/tce.2011.6131173 · Cited by 22 · see citing works
Self-citationEmotion-on-a-chip (EOC): Evolution of biochip technology to measure human emotion using body fluids
Lee, J.-H., Hwang, Y., Cheon, K.-A., & Jung, H.-I. (2012) · Medical Hypotheses · doi:10.1016/j.mehy.2012.09.002 · Cited by 20 · see citing works
Lee, Y., Lee, B., & Lee, M. (2010) · Telemedicine and e-Health · doi:10.1089/tmj.2009.0039 · Cited by 19 · see citing works
Selective current collecting design for spring-type energy harvesters
Kim, D.-J., Roh, H. S., Kim, Y., et al. (2015) · RSC Advances · doi:10.1039/c4ra16443a · Cited by 18 · see citing works
Reddy, K. A., George, B., & Kumar, V. J. (2008) · Conference proceedings - IEEE Instrumentation/Measurement Technology Conference · doi:10.1109/imtc.2008.4547026 · Cited by 16 · see citing works
Choi, H., Hong, S., Sung, T.-H., & No, K. (2011) · Applied Physics Letters · doi:10.1063/1.3632042 · Cited by 10 · see citing works
A thermal peripheral blood flowmeter with contact force compensation
Sim, J. K., Youn, S., & Cho, Y.-H. (2012) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/22/12/125014 · Cited by 10 · see citing works
Self-citationKim, W. Y., Ka, D. Y., Kwon, I. W., et al. (2011) · Current Applied Physics · doi:10.1016/j.cap.2011.03.021 · Cited by 8 · see citing works
HEART RATE VARIABILITY DURING HIGH-INTENSITY EXERCISE
Sarmiento, Samuel, S., Garcia-Manso, et al. (2013) · Acta Scientiarum Naturalium Universitatis Sunyatseni · Cited by 2 · see citing works
Corrosion resistance of aluminum alloys used in hydraulic construction
Akhmedov, G. M., Shakov, V. I., Trifel', M. S., & Khanlarova, A. G. (1968) · Hydrotechnical Construction · doi:10.1007/bf02377400 · Cited by 1 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- Carlson, N. Foundation of physiological psychology 3rd edn Trans. Kim, H. et al. Ch. 11, 459–466 (Sigmapress, 1998).
- Sun, F. Mobile computing, applications, and services (eds Kuo, C. et al.) 211–230 (Springer Berlin Heidelberg, 2012).
- Hibbleler, R. Mechanics of Materials 8th edn Ch. 12, 569–579 (Prentice Hall, 2011).
Cited by
Sorted by citations
Parlak, O., Keene, S. T., Marais, A., Curto, V. F., & Salleo, A. (2018) · Science Advances · doi:10.1126/sciadv.aar2904 · Cited by 603 · see citing works
Wearable sensors for monitoring the physiological and biochemical profile of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0150-9 · Cited by 537 · see citing works
A survey on wearable sensor modality centred human activity recognition in health care
Wang, Y., Cang, S., & Yu, H. (2019) · Expert Systems with Applications · doi:10.1016/j.eswa.2019.04.057 · Cited by 409 · see citing works
Wearable Devices for Ambulatory Cardiac Monitoring
Sana, F., Isselbacher, E. M., Singh, J. P., Heist, E. K., Pathik, B., & Armoundas, A. A. (2020) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2020.01.046 · Cited by 362 · see citing works
Mental health monitoring with multimodal sensing and machine learning: A survey
Garcia-Ceja, E., Riegler, M., Nordgreen, T., Jakobsen, P., Oedegaard, K. J., & Tørresen, J. (2018) · Pervasive and Mobile Computing · doi:10.1016/j.pmcj.2018.09.003 · Cited by 356 · see citing works
Su, Y., Ma, C., Chen, J., Wu, H., Luo, W., Peng, Y., Luo, Z., Li, L., Tan, Y., Omisore, O. M., et al. (2020) · Nanoscale Research Letters · doi:10.1186/s11671-020-03428-4 · Cited by 287 · see citing works
Taj-Eldin, M., Ryan, C., O’Flynn, B., & Galvin, P. (2018) · Sensors · doi:10.3390/s18124271 · Cited by 142 · see citing works
Jang, H., Sel, K., Kim, E., Kim, S., Yang, X., Kang, S., Ha, K.-H., Wang, R., Rao, Y., Jafari, R., et al. (2022) · Nature Communications · doi:10.1038/s41467-022-34406-2 · Cited by 127 · see citing works
Portable and wearable real-time stress monitoring: A critical review
Parlak, O. (2021) · Sensors and Actuators Reports · doi:10.1016/j.snr.2021.100036 · Cited by 60 · see citing works
Cohen, A. B., & Mathews, S. C. (2018) · Digital Biomarkers · doi:10.1159/000492396 · Cited by 24 · see citing works
- Authors
- Yoon, S., Sim, J. K., & Cho, Y.-H.
- Affiliations
- Korea Advanced Institute of Science and Technology (KR)
- Corresponding author
- Cho, Young-Ho
- Venue
- Scientific Reports
- Published
- 2016-03-23
- Volume
- 6
- Issue
- 1
- Pages
- 23468-23468
- ISSN
- 2045-2322 (ISSN-L)
- PMC
- PMC4804278
- OpenAlex
- W2314545921
- Semantic Scholar
- 774182daabc1319dc10ec7269b3e1b2bb4502867
- Cited by
- 151
- References
- 65
- Citations by year
- 2026: 8 · 2025: 8 · 2024: 15 · 2023: 15 · 2022: 14 · 2021: 28 · 2020: 21 · 2019: 20 · 2018: 15 · 2017: 7
- FWCI
- 6.74
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Scientific Reports (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- American Psychological Association
- Ministry of Science, ICT and Future Planning (2015054201)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Social Sciences › Psychology › Experimental and Cognitive Psychology › Emotion and Mood Recognition
- Keywords
- wearable patch, stress monitoring, skin temperature sensing, skin conductance, pulse wave detection, microfabrication, multimodal biosensors, emotion monitoring
- MeSH terms
- Equipment Design, Heart Rate, Humans, Skin Physiological Phenomena, Skin Temperature, Stress, Psychological, Monitoring, Ambulatory, Remote Sensing Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review