Approaches to a cortical vision prosthesis: implications of electrode size and placement
Open accessThe DOI matches its Crossref record; 68 of the 68 references it cites were found in a registry or an index.
Abstract
Objective. In order to move forward with the development of a cortical vision prosthesis, the critical issues in the field must be identified. Approach. To begin this process, we performed a brief review of several different cortical and retinal stimulation techniques that can be used to restore vision. Main results. Intracortical microelectrodes and epicortical macroelectrodes have been evaluated as the basis of a vision prosthesis. We concluded that an important knowledge gap necessitates an experimental in vivo performance evaluation of microelectrodes placed on the surface of the visual cortex. A comparison of the level of vision restored by intracortical versus epicortical microstimulation is necessary. Because foveal representation in the primary visual cortex involves more cortical columns per degree of visual field than does peripheral vision, restoration of foveal vision may require a large number of closely spaced microelectrodes. Based on previous studies of epicortical macrostimulation, it is possible that stimulation via surface microelectrodes could produce a lower spatial resolution, making them better suited for restoring peripheral vision. Significance. The validation of epicortical microstimulation in addition to the comparison of epicortical and intracortical approaches for vision restoration will fill an important knowledge gap and may have important implications for surgical strategies and device longevity. It is possible that the best approach to vision restoration will utilize both epicortical and intracortical microstimulation approaches, applying them appropriately to different visual representations in the primary visual cortex.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 ( A ) Photograph of human temporal neocortex taken after microelectrode array explantation. A bowing pattern on the cortical surface and minor hemorrhages were observed [ 31 ] . ( B ) Photograph of a feline brain implanted with a penetrating Utah microelectrode array before explantation. The array was implanted for a total of nine months. The back panel of the array is fl ush with the cortical surface. ( C ) The same specimen shown in fi gure 2 ( B ) , after explantation. The cortical surface exhibits an embedded footprint of the array after it has been removed.](https://figures.cochat.ai/paper-figures/dffe337f56534e6b93f5e150888468bb/approaches-to-a-cortical-vision-prosthesis-implications-of-electrode-size-and-figure-2.png)


Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“In addition, these large electrodes interacted with relatively large volumes of cortex (∼1 cm3), resulting in very low spatial resolution of the perceived phosphenes (Christie et al., 2016; Niketeghad et al., 2019).”
From Toward Long-Term Communication With the Brain in the Blind by Intracortical Stimulation: Challenges and Future Prospects (2020) · cites this paper as Christie et al., 2016
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Global estimates of visual impairment: 2010
Pascolini, D., & Mariotti, S. P. (2012) · British Journal of Ophthalmology · doi:10.1136/bjophthalmol-2011-300539 · Cited by 3,763
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · cited 1×
“meaning that over a short time interval there is zero net charge injected into the tissue [11].”
BackgroundElectrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · cited 1×
“The proliferation of astrocytes and glial scar formation can cause cell bodies to be pushed away from electrode recording sites [26], which may contribute to the decline in single unit activity that can be recorded over time [27].”
BackgroundCharge density and charge per phase as cofactors in neural injury induced by electrical stimulation
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.102812 · Cited by 708
Interim results from the international trial of Second Sight's visual prosthesis
Humayun, M. S., Dorn, J. D., da Cruz, L., Dagnelie, G., Sahel, J.-A., Stanga, P. E., Cideciyan, A. V., Duncan, J. L., Eliott, D., Filley, E., et al. (2012) · Ophthalmology · doi:10.1016/j.ophtha.2011.09.028 · Cited by 700 · cited 4×
“The Argus II allowed blind individuals to perform object localization, motion discrimination, and discrimination of oriented gratings [59].”
BackgroundMethodInfluentialSyringe-injectable electronics
Liu, J., Fu, T.-M., Cheng, Z., Hong, G., Zhou, T., Jin, L., Duvvuri, M., Jiang, Z., Kruskal, P., Xie, C., et al. (2015) · Nature Nanotechnology · doi:10.1038/nnano.2015.115 · Cited by 687
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672
Decoding motor imagery from the posterior parietal cortex of a tetraplegic human
Aflalo, T., Kellis, S., Klaes, C., Lee, B., Shi, Y., Pejsa, K., Shanfield, K., Hayes-Jackson, S., Aisen, M., Heck, C., et al. (2015) · Science · doi:10.1126/science.aaa5417 · Cited by 649
Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066014 · Cited by 600 · cited 1×
“stimulation, cracks occurred in the parylene-C insulation, which has been reported as a potential failure mode [39].”
BackgroundA model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526
Hassler, C., Boretius, T., & Stieglitz, T. (2011) · Journal of Polymer Science Part B: Polymer Physics · doi:10.1002/polb.22169 · Cited by 501 · cited 1×
“Although parylene-C is an established electrical insulator [40], its adhesion to metals in neural prosthesis devices is not sufficient [41].”
BackgroundSimeral, J. D., Kim, S.-P., Black, M. J., Donoghue, J. P., & Hochberg, L. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/2/025027 · Cited by 496
Factors influencing the biocompatibility of insertable silicon microshafts in cerebral cortex
Edell, D. J., Toi, V. V., McNeil, V. M., & Clark, L. D. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.141202 · Cited by 479 · cited 2×
“Just two weeks after implantation, a continuous sheath of cells will start to form around the insertion site [25, 26].”
BackgroundArtificial vision with wirelessly powered subretinal electronic implant alpha-IMS
Stingl, K., Bartz-Schmidt, K. U., Besch, D., Braun, A., Bruckmann, A., Gekeler, F., Greppmaier, U., Hipp, S., Hörtdörfer, G., Kernstock, C., et al. (2013) · Proceedings of the Royal Society B: Biological Sciences · doi:10.1098/rspb.2013.0077 · Cited by 473 · cited 1×
“A subretinal device, the Alpha IMS Wireless system, is on the market in Europe and can be used by blind individuals with untreatable forms of hereditary retinal degenerations [60].”
BackgroundDobelle, W. H., & Mladejovsky, M. G. (1974) · The Journal of Physiology · doi:10.1113/jphysiol.1974.sp010766 · Cited by 468 · cited 2×
“To validate Brindley and Lewin’s results, Dobelle conducted a series of acute experiments involving volunteers that were undergoing neurosurgical procedures in the occipital lobe [15].”
BackgroundBrain-Computer Interfaces Using Electrocorticographic Signals
Schalk, G., & Leuthardt, E. C. (2011) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2011.2172408 · Cited by 427 · cited 1×
“Macroelectrode diameters are 1 mm or greater, and typically are ∼4 mm in diameter in a clinical setting [5].”
BackgroundA neural interface for a cortical vision prosthesis
Normann, R. A., Maynard, E. M., Rousche, P. J., & Warren, D. J. (1999) · Vision Research · doi:10.1016/s0042-6989(99)00040-1 · Cited by 409
Williams, J. C., Rennaker, R. L., & Kipke, D. R. (1999) · Brain Research Protocols · doi:10.1016/s1385-299x(99)00034-3 · Cited by 395 · cited 1×
“However, microwire probes [28] and high-count penetrating microelectrode arrays [29] have been shown to record for over five years after implantation.”
BackgroundWeiland, J. D., Liu, W., & Humayun, M. S. (2005) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.7.060804.100435 · Cited by 371
A method for pneumatically inserting an array of penetrating electrodes into cortical tissue
Rousche, P. J., & Normann, R. A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368133 · Cited by 306 · cited 1×
“A pneumatic inserter is commonly used for the implantation of high-channel count penetrating electrode arrays [22], but it is also possible to place them manually at a low speed [23].”
BackgroundVisual sensations produced by intracortical microstimulation of the human occipital cortex
Bak, M., Girvin, J. P., Hambrecht, F. T., Kufta, C. V., Loeb, G. E., & Schmidt, E. M. (1990) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442682 · Cited by 299 · cited 1×
“To demonstrate the feasibility of intracortical microstimulation for use in a cortical vision prosthesis, Bak et al performed an acute, intra-operative study in awake humans in 1990 [32].”
MethodToward a comparison of microelectrodes for acute and chronic recordings
Ward, M. P., Rajdev, P., Ellison, C., & Irazoqui, P. P. (2009) · Brain Research · doi:10.1016/j.brainres.2009.05.052 · Cited by 296 · cited 1×
“The minimization of foreign body response can improve electrode longevity [14].”
BackgroundWorld Congress 2009 on Medical Physics and Biomedical Engineering
(2009) · Zeitschrift für Medizinische Physik · doi:10.1016/j.zemedi.2009.03.006 · Cited by 291
Retinal Remodeling in Inherited Photoreceptor Degenerations
Marc, R. E., & Jones, B. W. (2003) · Molecular Neurobiology · doi:10.1385/mn:28:2:139 · Cited by 258 · cited 1×
“ganglion cells is common in many retinal diseases [2].”
BackgroundDecoding spoken words using local field potentials recorded from the cortical surface
Kellis, S., Miller, K., Thomson, K., Brown, R., House, P., & Greger, B. (2010) · Journal of Neural Engineering · doi:10.1088/1741-2560/7/5/056007 · Cited by 235
Self-citationMapping of functional organization in human visual cortex: Electrical cortical stimulation
Lee, H. W., Hong, S. B., Seo, D. W., Tae, W. S., & Hong, S. C. (2000) · Neurology · doi:10.1212/wnl.54.4.849 · Cited by 184 · cited 1×
“However, a study with human subjects observed that the majority of evoked phosphenes were brighter than the background color [47].”
BackgroundWinslow, B. D., & Tresco, P. A. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2009.11.049 · Cited by 178
Hill, N. J., Gupta, D., Brunner, P., Gunduz, A., Adamo, M. A., Ritaccio, A. L., & Schalk, G. (2012) · Journal of Visualized Experiments · doi:10.3791/3993 · Cited by 178 · cited 1×
“Macroelectrode grids often contain a square of 8×8 electrodes with an inter-electrode spacing of 1 cm, while a strip contains a single line of 4–6 electrodes [6].”
BackgroundA model for intracortical visual prosthesis research
Troyk, P., Bak, M., Berg, J., Bradley, D., Cogan, S., Erickson, R., Kufta, C., McCreery, D., Schmidt, E., & Towle, V. (2003) · Artificial Organs · doi:10.1046/j.1525-1594.2003.07308.x · Cited by 166 · cited 2×
“microelectrodes were chronically implanted in the V1 operculum of a non-human primate [34].”
BackgroundMethodComparison of neural damage induced by electrical stimulation with faradaic and capacitor electrodes
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. A. (1988) · Annals of Biomedical Engineering · doi:10.1007/bf02368010 · Cited by 163
Visuotopic Mapping Through a Multichannel Stimulating Implant in Primate V1
Bradley, D. C., Troyk, P. R., Berg, J. A., Bak, M., Cogan, S., Erickson, R., Kufta, C., Mascaro, M., McCreery, D., Schmidt, E. M., et al. (2005) · Journal of Neurophysiology · doi:10.1152/jn.01213.2003 · Cited by 152 · cited 2×
“In a continuation of this study, 32 out of 152 electrodes were selected for stimulation [35].”
BackgroundToward the development of a cortically based visual neuroprosthesis
Normann, R. A., Greger, B. A., House, P., Romero, S. F., Pelayo, F., & Fernandez, E. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/3/035001 · Cited by 143 · cited 1×
“evoked using high levels of current, they tend to have unpredictable spatial properties and locations [17].”
BackgroundSelf-citationMcCreery, D., Pikov, V., & Troyk, P. R. (2010) · Journal of Neural Engineering · doi:10.1088/1741-2560/7/3/036005 · Cited by 137 · cited 3×
“prosthesis is time-variable, the majority of the neuronal loss could be attributed to the mechanical injury resulting from electrode insertion [48].”
BackgroundMethodMinnikanti, S., Diao, G., Pancrazio, J. J., Xie, X., Rieth, L. W., Solzbacher, F., & Peixoto, N. (2013) · Acta Biomaterialia · doi:10.1016/j.actbio.2013.10.031 · Cited by 114
Perge, J. A., Zhang, S., Malik, W. Q., Homer, M. L., Cash, S., Friehs, G., Eskandar, E. N., Donoghue, J. P., & Hochberg, L. R. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/4/046007 · Cited by 100
Electrical Performance of Penetrating Microelectrodes Chronically Implanted in Cat Cortex
Kane, S. R., Cogan, S. F., Ehrlich, J., Plante, T. D., McCreery, D. B., & Troyk, P. R. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2248152 · Cited by 95
Davis, T. S., Parker, R. A., House, P. A., Bagley, E., Wendelken, S., Normann, R. A., & Greger, B. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/6/065003 · Cited by 91 · cited 1×
“tiguous electrodes resulted in an increased percentage of responsive regions across the array [9].”
BackgroundSelf-citationXie, X., Rieth, L. W., Caldwell, R., Negi, S., Bhandari, R., Sharma, R., Tathireddy, P., & Solzbacher, F. (2015) · Biomedical Microdevices · doi:10.1007/s10544-014-9904-y · Cited by 91
Torab, K., Davis, T. S., Warren, D. J., House, P. A., Normann, R. A., & Greger, B. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/3/035001 · Cited by 90 · cited 1×
“A later study set out to characterize the performance of intracortical microstimulation in non-human primates using silicon-based microelectrode arrays [8].”
BackgroundSelf-citationMulti-scale analysis of neural activity in humans: Implications for micro-scale electrocorticography
Kellis, S., Sorensen, L. B., Darvas, F., Sayres, C., O’Neill, K., Brown, R. B., House, P. A., Ojemann, J. G., & Greger, B. E. (2015) · Clinical Neurophysiology · doi:10.1016/j.clinph.2015.06.002 · Cited by 89
Self-citationHouse, P. A., MacDonald, J. D., Tresco, P. A., & Normann, R. A. (2006) · Neurosurgical Focus · doi:10.3171/foc.2006.20.5.5 · Cited by 89
Self-citationThe visual sensations produced by electrical stimulation of the medial occipital cortex.
Brindley, G. S., & Lewin, W. S. (1968) · PubMed · Cited by 89
Laminar Variation in Threshold for Detection of Electrical Excitation of Striate Cortex by Macaques
DeYoe, E. A., Lewine, J. D., & Doty, R. W. (2005) · Journal of Neurophysiology · doi:10.1152/jn.00407.2005 · Cited by 86 · cited 1×
“Fibers in layer 1 may make this more difficult, but microelectrodes acutely placed in layers 2/3 can evoke phosphenes with safe low levels of stimulation [20, 21].”
BackgroundCurrent steering in retinal stimulation via a quasimonopolar stimulation paradigm
Matteucci, P. B., Chen, S. C., Tsai, D., Dodds, C. W. D., Dokos, S., Morley, J. W., Lovell, N. H., & Suaning, G. J. (2013) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.13-11653 · Cited by 79
Xie, X., Rieth, L., Merugu, S., Tathireddy, P., & Solzbacher, F. (2012) · Applied Physics Letters · doi:10.1063/1.4748322 · Cited by 75
Seven years of recording from monkey cortex with a chronically implanted multiple microelectrode
Krüger, J. (2010) · Frontiers in Neuroengineering · doi:10.3389/fneng.2010.00006 · Cited by 73
Rennaker, R. L., Street, S. E., Ruyle, A. M., & Sloan, A. M. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.08.009 · Cited by 70 · cited 1×
“A pneumatic inserter is commonly used for the implantation of high-channel count penetrating electrode arrays [22], but it is also possible to place them manually at a low speed [23].”
Schiller, P. H., Slocum, W. M., Kwak, M. C., Kendall, G. L., & Tehovnik, E. J. (2011) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1108337108 · Cited by 54 · cited 1×
“A separate acute study set out to characterize the types of images that can be expected when V1 is stimulated [46].”
BackgroundDepth‐dependent detection of microampere currents delivered to monkey V1
Tehovnik, E. J., & Slocum, W. M. (2009) · European Journal of Neuroscience · doi:10.1111/j.1460-9568.2009.06695.x · Cited by 54 · cited 1×
“Fibers in layer 1 may make this more difficult, but microelectrodes acutely placed in layers 2/3 can evoke phosphenes with safe low levels of stimulation [20, 21].”
BackgroundIntrasulcal Electrocorticography in Macaque Monkeys with Minimally Invasive Neurosurgical Protocols
Matsuo, T., Kawasaki, K., Osada, T., Sawahata, H., Suzuki, T., Shibata, M., Miyakawa, N., Nakahara, K., Iijima, A., Sato, N., et al. (2011) · Frontiers in Systems Neuroscience · doi:10.3389/fnsys.2011.00034 · Cited by 54 · cited 1×
“With the development of ultra-thin flexible probes using micro-electro-mechanical systems technology [67], it may be possible to place surface microelectrode arrays within the cortical sulci, thereby allowing stimulation of the entire parafoveal and peripheral visual field representation.”
MethodThe functional consequences of chronic, physiologically effective intracortical microstimulation
Parker, R. A., Davis, T. S., House, P. A., Normann, R. A., & Greger, B. (2011) · Progress in Brain Research · doi:10.1016/b978-0-444-53815-4.00010-8 · Cited by 47
Self-citationTranslating image sequences into spike patterns for cortical neuro-stimulation
Pelayo, F. J., Romero, S. F., Morillas, C. A., Martínez-Álvarez, A., Ros, E., & Fernández, E. (2004) · Neurocomputing · doi:10.1016/j.neucom.2004.01.142 · Cited by 39
Lane, F. J., HUYCK, M. H., & Troyk, P. R. (2010) · Disability and Rehabilitation Assistive Technology · doi:10.3109/17483107.2010.514381 · Cited by 30 · cited 1×
“In addition, they have pursued guidance from individuals who would use these devices [55, 56].”
BackgroundYamagishi, F. G. (1991) · Thin Solid Films · doi:10.1016/0040-6090(91)90539-a · Cited by 30 · cited 1×
“Although parylene-C is an established electrical insulator [40], its adhesion to metals in neural prosthesis devices is not sufficient [41].”
BackgroundIntraoperative Neurophysiologic Monitoring
Galloway, G. M., Nuwer, M. R., López, J. R., & Zamel, K. M. (2010) · Cambridge University Press eBooks · doi:10.1017/cbo9780511777950 · Cited by 28
Lane, F. J., HUYCK, M. H., Troyk, P. R., & Schug, K. (2012) · Disability and Rehabilitation Assistive Technology · doi:10.3109/17483107.2011.629326 · Cited by 27 · cited 1×
“In addition, they have pursued guidance from individuals who would use these devices [55, 56].”
BackgroundTransformative Reality: Improving bionic vision with robotic sensing
Lui, W. L. D., Browne, D., Kleeman, L., Drummond, T., & Li, W. H. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6345929 · Cited by 23
Forsling, M. L., & Widdas, W. F. (1968) · The Journal of Physiology · doi:10.1113/jphysiol.1968.sp008423 · Cited by 19
Rizzo III, J. F., & Ayton, L. N. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/2/020301 · Cited by 18 · cited 1×
“Progress is also being made in establishing a uniform testing procedure for vision prostheses [57].”
BackgroundToward the development of a cortically based visual neuroprosthesis
Normann, R. A., Greger, B. A., House, P., Romero, S. F., Pelayo, F., & Fernandez, E. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2552/6/4/049802 · Cited by 18
Self-citationHenle, C., Meier, W. P., Schüettler, M., Boretius, T., & Stieglitz, T. (2009) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-03889-1_27 · Cited by 12
Parker, R. A., Negi, S., Davis, T. S., Keefer, E. W., Wiggins, H., House, P. A., & Greger, B. E. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346050 · Cited by 11
Self-citationXie, X., Rieth, L. W., Tathireddy, P., & Solzbacher, F. (2013) · International Conference on Solid-State Sensors, Actuators and Microsystems · doi:10.1109/transducers.2013.6626949 · Cited by 11
Sensing millimeter-scale dynamics in cortical surface potentials for neural prosthetics
Kellis, S., Greger, B. E., Hanrahan, S. J., House, P. A., & Brown, R. B. (2011) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2011.6127165 · Cited by 7 · cited 1×
“cerebrospinal fluid between the electrode and the cortical surface [64].”
BackgroundSelf-citationCORTICAL VISUAL NEUROPROSTHESES FOR THE BLIND
Fernández, E., Pelayo, F. J., Ahnelt, P. K., Ammermüller, J., & Normann, R. A. (2004) · Cited by 7
Recent Results from Second Sight’s Argus® II Retinal Prosthesis Study
Humayun, M. S., da Cruz, L., Dagnelie, G., Stanga, P. E., Ho, A. C., Greenberg, R. J., Birch, D. G., Duncan, J. L., & Sahel, J. (2013) · Investigative Ophthalmology & Visual Science · Cited by 3
Intraoperative neurophysiologic monitoring of vascular disorders
López, J. R. (2010) · Cambridge University Press eBooks · doi:10.1017/cbo9780511777950.015 · Cited by 2
Research works printed but not linked: 2
- 20. DeYoe E A 1983 An Investigation in the Awake Macaque of the Threshold for Detection of Electrical Currents Applied to Striate Cortex: Psychophysical Properties and Laminar Differences Doctoral Thesis University of Rochester
- 62. Kellis S, Hanrahan S, Davis T, House P A, Brown R and Greger B 2012 Decoding hand trajectories from micro- electrocorticography in human patients Annual Int. Conf. IEEE Engineering in Medicine and Biology Society (EMBC) pp 4091-4
Cited by
Dynamic Stimulation of Visual Cortex Produces Form Vision in Sighted and Blind Humans
Beauchamp, M. S., Oswalt, D., Sun, P., Foster, B. L., Magnotti, J. F., Niketeghad, S., Pouratian, N., Bosking, W. H., & Yoshor, D. (2020) · Cell · doi:10.1016/j.cell.2020.04.033 · Cited by 257
“This has inspired hope for the development of a visual cortical prosthetic (VCP), a device that would bypass the eyes and optic nerve, transmitting visual information from a camera directly into the visual cortex (Bosking et al., 2017a; Brindley and Lewin, 1968; Christie et al., 2016; Lewis et al., 2015, 2016; Najarpo…”
MethodCaldwell, R., Street, M. G., Sharma, R., Takmakov, P., Baker, B., & Rieth, L. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2019.119731 · Cited by 100
Welle, E. J., Patel, P. R., Woods, J. E., Petrossians, A., della Valle, E., Vega-Medina, A., Richie, J. M., Cai, D., Weiland, J. D., & Chestek, C. A. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab8343 · Cited by 72
Fernández, E., Alfaro, A., & González-López, P. (2020) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.00681 · Cited by 35
“In addition, these large electrodes interacted with relatively large volumes of cortex (∼1 cm3), resulting in very low spatial resolution of the perceived phosphenes (Christie et al., 2016; Niketeghad et al., 2019).”
BackgroundOswalt, D., Bosking, W., Sun, P., Sheth, S. A., Niketeghad, S., Salas, M. A., Patel, U., Greenberg, R., Dorn, J., Pouratian, N., et al. (2021) · Brain Stimulation · doi:10.1016/j.brs.2021.08.024 · Cited by 40
Metadata
- Authors
- Christie, B. P., Ashmont, K. R., House, P. A., & Greger, B.
- Affiliations
- Arizona State University (US)
- University of Utah (US)
- Corresponding author
- Greger, Bradley
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2016-02-23
- Volume
- 13
- Issue
- 2
- Pages
- 025003
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2288591390
- Cited by
- 26
- References
- 67
- Citations by year
- 2026: 2 · 2025: 2 · 2024: 2 · 2023: 2 · 2022: 2 · 2021: 3 · 2020: 6 · 2019: 4 · 2018: 1 · 2017: 1 · 2016: 1
- FWCI
- 0.26
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- publishingsupport.iopscience.iop.org/iop-standard/v1 (vor)
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health
- Research to Prevent Blindness
- Telemedicine and Advanced Technology Research Center
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- intracortical microelectrodes, microstimulation, macrostimulation, primary visual cortex, foveal vision, peripheral vision, spatial resolution, electrode size, electrode placement, retinal stimulation, vision restoration
- MeSH terms
- Animals, Electric Stimulation, Electrodes, Implanted, Evoked Potentials, Visual, Humans, Microelectrodes, Vision Disorders, Visual Cortex, Visual Prosthesis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.