Integrated and flexible multichannel interface for electrotactile stimulation
Open accessThe DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The aim of the present work was to develop and test a flexible electrotactile stimulation system to provide real-time feedback to the prosthesis user. The system requirements were to accommodate the capabilities of advanced multi-DOF myoelectric hand prostheses and transmit the feedback variables (proprioception and force) using intuitive coding, with high resolution and after minimal training. APPROACH: We developed a fully-programmable and integrated electrotactile interface supporting time and space distributed stimulation over custom designed flexible array electrodes. The system implements low-level access to individual stimulation channels as well as a set of high-level mapping functions translating the state of a multi-DoF prosthesis (aperture, grasping force, wrist rotation) into a set of predefined dynamic stimulation profiles. The system was evaluated using discrimination tests employing spatial and frequency coding (10 able-bodied subjects) and dynamic patterns (10 able-bodied and 6 amputee subjects). The outcome measure was the success rate (SR) in discrimination. MAIN RESULTS: The more practical electrode with the common anode configuration performed similarly to the more usual concentric arrangement. The subjects could discriminate six spatial and four frequency levels with SR >90% after a few minutes of training, whereas the performance significantly deteriorated for more levels. The dynamic patterns were intuitive for the subjects, although amputees showed lower SR than able-bodied individuals (86% ± 10% versus 99% ± 3%). SIGNIFICANCE: The tests demonstrated that the system was easy to setup and apply. The design and resolution of the multipad electrode was evaluated. Importantly, the novel dynamic patterns, which were successfully tested, can be superimposed to transmit multiple feedback variables intuitively and simultaneously. This is especially relevant for closing the loop in modern multifunction prostheses. Therefore, the proposed system is convenient for practical applications and can be used to implement sensory perception training and/or closed-loop control of myoelectric prostheses, providing grasping force and proprioceptive feedback.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Myoelectric control systems—A survey
Asghari Oskoei, M., & Hu, H. (2007) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2007.07.009 · Cited by 1,244
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
Sensory substitution and the human-machine interface
Bach-y-Rita, P., & W. Kercel, S. (2003) · Trends in Cognitive Sciences · doi:10.1016/j.tics.2003.10.013 · Cited by 858
Control of limb dynamics in normal subjects and patients without proprioception
Sainburg, R. L., Ghilardi, M. F., Poizner, H., & Ghez, C. (1995) · Journal of Neurophysiology · doi:10.1152/jn.1995.73.2.820 · Cited by 662
Effect of sensory discrimination training on cortical reorganisation and phantom limb pain
Flor, H., Denke, C., Schaefer, M., & Grüsser, S. (2001) · The Lancet · doi:10.1016/s0140-6736(00)04890-x · Cited by 560
Closed-loop control in prosthetic systems: Historical perspective
Childress, D. S. (1980) · Annals of Biomedical Engineering · doi:10.1007/bf02363433 · Cited by 158
Design and Evaluation of a Low-Cost Force Feedback System for Myoelectric Prosthetic Hands
Pylatiuk, C., Kargov, A., & Schulz, S. (2006) · JPO Journal of Prosthetics and Orthotics · doi:10.1097/00008526-200604000-00007 · Cited by 156
Vibro- and Electrotactile User Feedback on Hand Opening for Myoelectric Forearm Prostheses
Witteveen, H. J. B., Droog, E. A., Rietman, J. S., & Veltink, P. H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2200678 · Cited by 150
Anani, A. B., Ikeda, K., & Körner, L. M. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457988 · Cited by 74
Perovic, M., Stevanovic, M., Jevtic, T., Strbac, M., Bijelic, G., Vucetic, C., Popovic-Maneski, L., & Popovic, D. (2013) · Journal of Automatic Control · doi:10.2298/jac1301013p · Cited by 28
Self-citation
Research works printed but not linked: 43
- Biddiss, 2007, Upper limb prosthesis use and abandonment: a survey of the last 25 years, Prosthet. Orthot. Int., 31, 236, 10.1080/03093640600994581 · DOI · Paper page
- Atkins, 1996, Epidemiologic overview of individuals with upper-limb loss and their reported research priorities, J. Prosthet. Orthot., 8, 2, 10.1097/00008526-199600810-00003 · DOI · Paper page
- Dietrich, 2012, Sensory feedback prosthesis reduces phantom limb pain: proof of a principle, Neurosci. Lett., 507, 97, 10.1016/j.neulet.2011.10.068 · DOI · Paper page
- Dhillon, 2005, Direct neural sensory feedback and control of a prosthetic arm, IEEE Trans. Neural Syst. Rehabil. Eng., 13, 468, 10.1109/TNSRE.2005.856072 · DOI · Paper page
- Meek, 1989, Extended physiologic taction: design and evaluation of a proportional force feedback system, J. Rehabil. Res. Dev., 26, 53
- Lundborg, 1999, Hearing as substitution for sensation: a new principle for artificial sensibility, J. Hand Surg., 24, 219, 10.1053/jhsu.1999.0219 · DOI
- Antfolk, 2013, Transfer of tactile input from an artificial hand to the forearm: experiments in amputees and able-bodied volunteers, Disability Rehabil.: Assist. Technol., 8, 249, 10.3109/17483107.2012.713435 · DOI · Paper page
- Cipriani, 2009, A novel concept for a prosthetic hand with a bidirectional interface: a feasibility study, IEEE Trans. Biomed. Eng., 56, 2739, 10.1109/TBME.2009.2031242 · DOI
- Cipriani, 2012, A miniature vibrotactile sensory substitution device for multifingered hand prosthetics, IEEE Trans. Biomed. Eng., 59, 400, 10.1109/TBME.2011.2173342 · DOI · Paper page
- Ninu, 2014, 22, 1041, 10.1109/TNSRE.2014.2318431 · DOI
- Rombokas, 2013, Vibrotactile sensory substitution for electromyographic control of object manipulation, IEEE Trans. Biomed. Eng., 60, 2226, 10.1109/TBME.2013.2252174 · DOI · Paper page
- Saunders, 2011, The role of feed-forward and feedback processes for closed-loop prosthesis control, J. Neuroeng. Rehabil., 8, 1, 10.1186/1743-0003-8-60 · DOI · Paper page
- Stepp, 2012, Vibrotactile sensory substitution for object manipulation: amplitude versus pulse train frequency modulation, IEEE Trans. Neural Syst. Rehabil. Eng., 20, 31, 10.1109/TNSRE.2011.2170856 · DOI
- Antfolk, 2010, A tactile display system for hand prostheses to discriminate pressure and individual finger localization, J. Med. Biol. Eng., 30, 355, 10.5405/jmbe.766 · DOI
- Antfolk, 2013, Artificial redirection of sensation from prosthetic fingers to the phantom hand map on transradial amputees: vibrotactile versus mechanotactile sensory feedback, IEEE Trans. Neural Syst. Rehabil. Eng., 21, 112, 10.1109/TNSRE.2012.2217989 · DOI · Paper page
- Echenique, 2011, Electrical stimulation of mechanoreceptors, J. Phys.: Conf. Ser., 332, 10.1088/1742-6596/332/1/012044 · DOI
- Geng, 2014, Human ability in identification of location and pulse number for electrocutaneous stimulation applied on the forearm, J. Neuro Engineering Rehabil., 11, 97, 10.1186/1743-0003-11-97 · DOI
- Higashiyama, 1993, Localization of electrocutaneous stimuli on the fingers and forearm: effects of electrode configuration and body axis, Perception Psychophysics, 54, 108, 10.3758/BF03206942 · DOI
- Jorgovanovic, 2014, Virtual grasping: closed-loop force control using electrotactile feedback, Comput. Math. Methods Med., 2014, 120357, 10.1155/2014/120357 · DOI · Paper page
- Szeto, 1982, Electrocutaneous stimulation for sensory communication in rehabilitation engineering, Biomed. Eng., 29, 300, 10.1109/TBME.1982.324948 · DOI · Paper page
- Szeto, 1977, Comparison of codes for sensory feedback using electrocutaneous tracking, Ann. Biomed. Eng., 5, 367, 10.1007/BF02367316 · DOI
- Witteveen, 2012, Grasping force and slip feedback through vibrotactile stimulation to be used in myoelectric forearm prostheses, 2969, 10.1109/EMBC.2012.6346587 · DOI
- Antfolk, 2013, 10, 45, 10.1586/erd.12.68 · DOI · Paper page
- Peerdeman, 2011, 48, 719, 10.1682/JRRD.2010.08.0161 · DOI · Paper page
- Prior, 1975, Electrocutaneous feedback for artificial limbs. Summary progress report, February 1, 1974, through July 31, 1975, Bull. Prosthet. Res., 3 (10–24), 3
- Mann, 1970, Kinesthetic sensing for the EMG controlled Boston arm, IEEE Trans. Man-Mach. Syst., 11, 110, 10.1109/TMMS.1970.299971 · DOI · Paper page
- Chatterjee, 2008, Testing a prosthetic haptic feedback simulator with an interactive force matching task, J. Prosthet. Orthot., 20, 27, 10.1097/01.JPO.0000311041.61628.be · DOI
- Cipriani, 2008, On the shared control of an EMG-controlled prosthetic hand: analysis of user–prosthesis interaction, IEEE Trans. Robot., 24, 170, 10.1109/TRO.2007.910708 · DOI · Paper page
- Shannon, 1979, A myoelectrically-controlled prosthesis with sensory feedback, Med. Biol. Eng. Comput., 17, 73, 10.1007/BF02440956 · DOI
- Belter, 2013, Mechanical design and performance specifications of anthropomorphic prosthetic hands: a review, J. Rehabil. Res. Dev., 50, 599, 10.1682/JRRD.2011.10.0188 · DOI · Paper page
- Solomonow, 1977, Electrotactile two-point discrimination as a function of frequency, body site, laterality, and stimulation codes, Ann. Biomed. Eng., 5, 47, 10.1007/BF02409338 · DOI
- Cody, 2008, Tactile spatial acuity varies with site and axis in the human upper limb, Neurosci. Lett., 433, 103, 10.1016/j.neulet.2007.12.054 · DOI
- Geng, 2011, Impacts of selected stimulation patterns on the perception threshold in electrocutaneous stimulation, J. Neuroeng. Rehabil., 8, 9, 10.1186/1743-0003-8-9 · DOI
- Gescheider, 2013, The classical psychophysical methods, 10.4324/9780203774458 · DOI · Paper page
- Saunders, 1977, Recommended procedures for electrocutaneous displays, Functional Electrical Stimulation: Applications in Neural Prostheses, 303
- Szeto, 1979, Electrocutaneous pulse rate and pulse width psychometric functions for sensory communications, Hum. Factors: J. Hum. Factors Ergon. Soc., 21, 241, 10.1177/001872087902100212 · DOI · Paper page
- Goff, 1967, Differential discrimination of frequency of cutaneous mechanical vibration, J. Exp. Psychol., 74, 294, 10.1037/h0024561 · DOI · Paper page
- Kaczmarek, 1991, Electrotactile and vibrotactile displays for sensory substitution systems, IEEE Trans. Biomed. Eng., 38, 1, 10.1109/10.68204 · DOI · Paper page
- Damian, 2012, Slip speed feedback for grip force control, IEEE Trans. Biomed. Eng., 59, 2200, 10.1109/TBME.2012.2199491 · DOI
- Dosen, 2016, 10.1109/TNSRE.2016.2550864 · DOI · Paper page
- D’Alonzo, 2014, HyVE—hybrid vibro-electrotactile stimulation—is an efficient approach to multi-channel sensory feedback, IEEE Trans. Haptics, 7, 181, 10.1109/TOH.2013.52 · DOI
- Riso, 1989, Electrocutaneous sensations elicited using subdermally located electrodes, Automedica, 11, 25
- Jiang, 2012, Myoelectric control of artificial limbs—Is there a need to change focus, IEEE Signal Process. Mag., 29, 152, 10.1109/MSP.2012.2203480 · DOI · Paper page
Cited by
Toward higher-performance bionic limbs for wider clinical use
Farina, D., Vujaklija, I., Brånemark, R., Bull, A. M. J., Dietl, H., Graimann, B., Hargrove, L. J., Hoffmann, K.-P., Huang, H., Ingvarsson, T., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00732-x · Cited by 263
Self-citationA Review of Non-Invasive Sensory Feedback Methods for Transradial Prosthetic Hands
Stephens-Fripp, B., Alici, G., & Mutlu, R. (2018) · IEEE Access · doi:10.1109/access.2018.2791583 · Cited by 210
Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88
Stephens-Fripp, B., Sencadas, V., Mutlu, R., & Alici, G. (2018) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2018.00179 · Cited by 29
Muceli, S., Poppendieck, W., Hoffmann, K.-P., Dosen, S., Benito-León, J., Barroso, F. O., Pons, J. L., & Farina, D. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab047a · Cited by 33
Self-citation
Metadata
- Authors
- Štrbac, M., Belić, M., Isaković, M., Kojić, V., Bijelić, G., Popović, I., Radotić, M., Došen, S., Marković, M., Farina, D., & Keller, T.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2016-06-14
- Volume
- 13
- Issue
- 4
- Pages
- 046014
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2441345851
- Cited by
- 106
- References
- 10
- FWCI
- 5.48
- Type
- journal article
- Language
- en
- Open access
- bronze
- Licence
- publishingsupport.iopscience.iop.org/iop-standard/v1 (vor)
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- Funders
- Tecnalia Research & Innovation, FIK, European Commission (FP7-PEOPLE-2011-IAPP-286208)
- Ministry of Education, Science and Technological Development of Serbia (Project no. 175016)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Integrated and flexible multichannel interface for electrotactile stimulation. https://app.cochat.ai/papers/10.1088/1741-2560/13/4/046014
- The paper
- Štrbac, M., Belić, M., Isaković, M., Kojić, V., Bijelić, G., Popović, I., Radotić, M., Došen, S., Marković, M., Farina, D., & Keller, T. (2016). Integrated and flexible multichannel interface for electrotactile stimulation. Journal of Neural Engineering. https://doi.org/10.1088/1741-2560/13/4/046014
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.