Integrated and flexible multichannel interface for electrotactile stimulation

Open access
Štrbac, Matija, Belić, Minja, Isaković, Milica, Kojić, Vladimir, Bijelić, Goran, Popović, Igor, Radotić, Milutin, Došen, Strahinja, Marković, Marko, Farina, Dario, Keller, Thierry · 2016 · Journal of Neural Engineering
Cited by 106

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: The aim of the present work was to develop and test a flexible electrotactile stimulation system to provide real-time feedback to the prosthesis user. The system requirements were to accommodate the capabilities of advanced multi-DOF myoelectric hand prostheses and transmit the feedback variables (proprioception and force) using intuitive coding, with high resolution and after minimal training. APPROACH: We developed a fully-programmable and integrated electrotactile interface supporting time and space distributed stimulation over custom designed flexible array electrodes. The system implements low-level access to individual stimulation channels as well as a set of high-level mapping functions translating the state of a multi-DoF prosthesis (aperture, grasping force, wrist rotation) into a set of predefined dynamic stimulation profiles. The system was evaluated using discrimination tests employing spatial and frequency coding (10 able-bodied subjects) and dynamic patterns (10 able-bodied and 6 amputee subjects). The outcome measure was the success rate (SR) in discrimination. MAIN RESULTS: The more practical electrode with the common anode configuration performed similarly to the more usual concentric arrangement. The subjects could discriminate six spatial and four frequency levels with SR >90% after a few minutes of training, whereas the performance significantly deteriorated for more levels. The dynamic patterns were intuitive for the subjects, although amputees showed lower SR than able-bodied individuals (86% ± 10% versus 99% ± 3%). SIGNIFICANCE: The tests demonstrated that the system was easy to setup and apply. The design and resolution of the multipad electrode was evaluated. Importantly, the novel dynamic patterns, which were successfully tested, can be superimposed to transmit multiple feedback variables intuitively and simultaneously. This is especially relevant for closing the loop in modern multifunction prostheses. Therefore, the proposed system is convenient for practical applications and can be used to implement sensory perception training and/or closed-loop control of myoelectric prostheses, providing grasping force and proprioceptive feedback.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/1741-2560/13/4/046014

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Research works printed but not linked: 43

  1. Biddiss, 2007, Upper limb prosthesis use and abandonment: a survey of the last 25 years, Prosthet. Orthot. Int., 31, 236, 10.1080/03093640600994581 · DOI · Paper page
  2. Atkins, 1996, Epidemiologic overview of individuals with upper-limb loss and their reported research priorities, J. Prosthet. Orthot., 8, 2, 10.1097/00008526-199600810-00003 · DOI · Paper page
  3. Dietrich, 2012, Sensory feedback prosthesis reduces phantom limb pain: proof of a principle, Neurosci. Lett., 507, 97, 10.1016/j.neulet.2011.10.068 · DOI · Paper page
  4. Dhillon, 2005, Direct neural sensory feedback and control of a prosthetic arm, IEEE Trans. Neural Syst. Rehabil. Eng., 13, 468, 10.1109/TNSRE.2005.856072 · DOI · Paper page
  5. Meek, 1989, Extended physiologic taction: design and evaluation of a proportional force feedback system, J. Rehabil. Res. Dev., 26, 53
  6. Lundborg, 1999, Hearing as substitution for sensation: a new principle for artificial sensibility, J. Hand Surg., 24, 219, 10.1053/jhsu.1999.0219 · DOI
  7. Antfolk, 2013, Transfer of tactile input from an artificial hand to the forearm: experiments in amputees and able-bodied volunteers, Disability Rehabil.: Assist. Technol., 8, 249, 10.3109/17483107.2012.713435 · DOI · Paper page
  8. Cipriani, 2009, A novel concept for a prosthetic hand with a bidirectional interface: a feasibility study, IEEE Trans. Biomed. Eng., 56, 2739, 10.1109/TBME.2009.2031242 · DOI
  9. Cipriani, 2012, A miniature vibrotactile sensory substitution device for multifingered hand prosthetics, IEEE Trans. Biomed. Eng., 59, 400, 10.1109/TBME.2011.2173342 · DOI · Paper page
  10. Ninu, 2014, 22, 1041, 10.1109/TNSRE.2014.2318431 · DOI
  11. Rombokas, 2013, Vibrotactile sensory substitution for electromyographic control of object manipulation, IEEE Trans. Biomed. Eng., 60, 2226, 10.1109/TBME.2013.2252174 · DOI · Paper page
  12. Saunders, 2011, The role of feed-forward and feedback processes for closed-loop prosthesis control, J. Neuroeng. Rehabil., 8, 1, 10.1186/1743-0003-8-60 · DOI · Paper page
  13. Stepp, 2012, Vibrotactile sensory substitution for object manipulation: amplitude versus pulse train frequency modulation, IEEE Trans. Neural Syst. Rehabil. Eng., 20, 31, 10.1109/TNSRE.2011.2170856 · DOI
  14. Antfolk, 2010, A tactile display system for hand prostheses to discriminate pressure and individual finger localization, J. Med. Biol. Eng., 30, 355, 10.5405/jmbe.766 · DOI
  15. Antfolk, 2013, Artificial redirection of sensation from prosthetic fingers to the phantom hand map on transradial amputees: vibrotactile versus mechanotactile sensory feedback, IEEE Trans. Neural Syst. Rehabil. Eng., 21, 112, 10.1109/TNSRE.2012.2217989 · DOI · Paper page
  16. Echenique, 2011, Electrical stimulation of mechanoreceptors, J. Phys.: Conf. Ser., 332, 10.1088/1742-6596/332/1/012044 · DOI
  17. Geng, 2014, Human ability in identification of location and pulse number for electrocutaneous stimulation applied on the forearm, J. Neuro Engineering Rehabil., 11, 97, 10.1186/1743-0003-11-97 · DOI
  18. Higashiyama, 1993, Localization of electrocutaneous stimuli on the fingers and forearm: effects of electrode configuration and body axis, Perception Psychophysics, 54, 108, 10.3758/BF03206942 · DOI
  19. Jorgovanovic, 2014, Virtual grasping: closed-loop force control using electrotactile feedback, Comput. Math. Methods Med., 2014, 120357, 10.1155/2014/120357 · DOI · Paper page
  20. Szeto, 1982, Electrocutaneous stimulation for sensory communication in rehabilitation engineering, Biomed. Eng., 29, 300, 10.1109/TBME.1982.324948 · DOI · Paper page
  21. Szeto, 1977, Comparison of codes for sensory feedback using electrocutaneous tracking, Ann. Biomed. Eng., 5, 367, 10.1007/BF02367316 · DOI
  22. Witteveen, 2012, Grasping force and slip feedback through vibrotactile stimulation to be used in myoelectric forearm prostheses, 2969, 10.1109/EMBC.2012.6346587 · DOI
  23. Antfolk, 2013, 10, 45, 10.1586/erd.12.68 · DOI · Paper page
  24. Peerdeman, 2011, 48, 719, 10.1682/JRRD.2010.08.0161 · DOI · Paper page
  25. Prior, 1975, Electrocutaneous feedback for artificial limbs. Summary progress report, February 1, 1974, through July 31, 1975, Bull. Prosthet. Res., 3 (10–24), 3
  26. Mann, 1970, Kinesthetic sensing for the EMG controlled Boston arm, IEEE Trans. Man-Mach. Syst., 11, 110, 10.1109/TMMS.1970.299971 · DOI · Paper page
  27. Chatterjee, 2008, Testing a prosthetic haptic feedback simulator with an interactive force matching task, J. Prosthet. Orthot., 20, 27, 10.1097/01.JPO.0000311041.61628.be · DOI
  28. Cipriani, 2008, On the shared control of an EMG-controlled prosthetic hand: analysis of user–prosthesis interaction, IEEE Trans. Robot., 24, 170, 10.1109/TRO.2007.910708 · DOI · Paper page
  29. Shannon, 1979, A myoelectrically-controlled prosthesis with sensory feedback, Med. Biol. Eng. Comput., 17, 73, 10.1007/BF02440956 · DOI
  30. Belter, 2013, Mechanical design and performance specifications of anthropomorphic prosthetic hands: a review, J. Rehabil. Res. Dev., 50, 599, 10.1682/JRRD.2011.10.0188 · DOI · Paper page
  31. Solomonow, 1977, Electrotactile two-point discrimination as a function of frequency, body site, laterality, and stimulation codes, Ann. Biomed. Eng., 5, 47, 10.1007/BF02409338 · DOI
  32. Cody, 2008, Tactile spatial acuity varies with site and axis in the human upper limb, Neurosci. Lett., 433, 103, 10.1016/j.neulet.2007.12.054 · DOI
  33. Geng, 2011, Impacts of selected stimulation patterns on the perception threshold in electrocutaneous stimulation, J. Neuroeng. Rehabil., 8, 9, 10.1186/1743-0003-8-9 · DOI
  34. Gescheider, 2013, The classical psychophysical methods, 10.4324/9780203774458 · DOI · Paper page
  35. Saunders, 1977, Recommended procedures for electrocutaneous displays, Functional Electrical Stimulation: Applications in Neural Prostheses, 303
  36. Szeto, 1979, Electrocutaneous pulse rate and pulse width psychometric functions for sensory communications, Hum. Factors: J. Hum. Factors Ergon. Soc., 21, 241, 10.1177/001872087902100212 · DOI · Paper page
  37. Goff, 1967, Differential discrimination of frequency of cutaneous mechanical vibration, J. Exp. Psychol., 74, 294, 10.1037/h0024561 · DOI · Paper page
  38. Kaczmarek, 1991, Electrotactile and vibrotactile displays for sensory substitution systems, IEEE Trans. Biomed. Eng., 38, 1, 10.1109/10.68204 · DOI · Paper page
  39. Damian, 2012, Slip speed feedback for grip force control, IEEE Trans. Biomed. Eng., 59, 2200, 10.1109/TBME.2012.2199491 · DOI
  40. Dosen, 2016, 10.1109/TNSRE.2016.2550864 · DOI · Paper page
  41. D’Alonzo, 2014, HyVE—hybrid vibro-electrotactile stimulation—is an efficient approach to multi-channel sensory feedback, IEEE Trans. Haptics, 7, 181, 10.1109/TOH.2013.52 · DOI
  42. Riso, 1989, Electrocutaneous sensations elicited using subdermally located electrodes, Automedica, 11, 25
  43. Jiang, 2012, Myoelectric control of artificial limbs—Is there a need to change focus, IEEE Signal Process. Mag., 29, 152, 10.1109/MSP.2012.2203480 · DOI · Paper page

Cited by

See all 106

Metadata

Authors
Štrbac, M., Belić, M., Isaković, M., Kojić, V., Bijelić, G., Popović, I., Radotić, M., Došen, S., Marković, M., Farina, D., & Keller, T.
Venue
Journal of Neural Engineering
Published
2016-06-14
Volume
13
Issue
4
Pages
046014
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W2441345851
Cited by
106
References
10
FWCI
5.48
Type
journal article
Language
en
Open access
bronze
Licence
Funders
  • Tecnalia Research & Innovation, FIK, European Commission (FP7-PEOPLE-2011-IAPP-286208)
  • Ministry of Education, Science and Technological Development of Serbia (Project no. 175016)

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Integrated and flexible multichannel interface for electrotactile stimulation. https://app.cochat.ai/papers/10.1088/1741-2560/13/4/046014
The paper
Štrbac, M., Belić, M., Isaković, M., Kojić, V., Bijelić, G., Popović, I., Radotić, M., Došen, S., Marković, M., Farina, D., & Keller, T. (2016). Integrated and flexible multichannel interface for electrotactile stimulation. Journal of Neural Engineering. https://doi.org/10.1088/1741-2560/13/4/046014

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.