A movable unshielded magnetocardiography system
Open accessThe DOI matches its Crossref record; 60 of the 62 references it cites were found in a registry or an index; 19 of the 55 sources its claims cite were checked for support.
Abstract
Magnetocardiography (MCG), which uses high-sensitivity magnetometers to record magnetic field signals generated by electrical activity in the heart, is a noninvasive method for evaluating heart diseases such as arrhythmia and ischemia. The MCG measurements usually require the participant keeping still in a magnetically shielded room due to the immovable sensor and noisy external environments. These requirements limit MCG applications, such as exercise MCG tests and long-term MCG observations, which are useful for early detections of heart diseases. Here, we introduce a movable MCG system that can clearly record MCG signals of freely behaving participants in an unshielded environment. On the basis of optically pumped magnetometers with a sensitivity of 140 fT/Hz 1/2 , we successfully demonstrated the resting MCG and the exercise MCG tests. Our method is promising to realize a practical movable multichannel unshielded MCG system that nearly sets no limits to participants and brings another kind of insight into the medical diagnosis of heart disease.
Overview
CoChat Summary
TL;DR
The paper presents a movable, unshielded magnetocardiography (MCG) system using self-oscillating optically pumped magnetometers (OPMs) that records cardiac magnetic signals from freely behaving participants in Earth’s field. It demonstrates resting MCG and exercise MCG (eMCG) measurements, including during high-knee exercise, with gradiometric noise suppression and signal processing.
Key findings
- Each OPM achieved a sensitivity of ∼140 fTrms/Hz1/2, and the two-sensor gradiometer achieved ∼200 fTrms/Hz1/2 with a 15-cm baseline; measured common-mode noise rejection ratio was about 80 at 1 Hz.
- In eMCG tests, motion artifacts were up to about ∼30 nT, but gradiometric operation suppressed most common-mode motion artifacts to less than 1 nT; breathing artifacts of about 20 pT with a period of ∼3 s were observed at rest.
- Resting MCG showed clear QRS and T-waves without signal averaging and matched ECG and PPG; pulse transit time was estimated at about 280 ms.
- In three healthy male participants aged 22–25, eMCG during exercise showed cardiac period decreasing from about 0.65 s at standing still to 0.52 s in stable exercise (115 beats per minute) and increasing to 0.69 s after rest; T-wave amplitude changed from ∼10 pT to ∼15 pT to 6 pT, and R-wave amplitude from ∼35 pT to ∼40 pT to 33 pT across S, E2, and R2 stages.
Methods
- Design: a movable unshielded MCG system using two robust self-oscillating cesium OPMs in gradiometric mode, with one sensor near the left chest and a reference sensor ∼15 cm away, mounted in a vest; the system operated in Earth’s field (about 54,500 nT in Beijing, China).
- Data/sample: resting MCG was recorded with participants sitting in a yard away from buildings; eMCG was measured in three healthy male individuals aged 22–25 during stand still (80 s), high-knee exercise (140 s), and standing rest (80 s). ECG and PPG were recorded synchronously for resting measurements.
- Analysis: raw signals were filtered with a fourth-order 0.6–45 Hz Butterworth band-pass filter and notch filters, then denoised using empirical mode decomposition with interval thresholding (EMD-IT); eMCG features including RR interval, QRS duration, RT interval, T-wave amplitude, and R-wave amplitude were extracted and compared across stages.
Limitations
- The paper states that technical issues remain for a wearable multichannel system: suppression of magnetic gradient noise, miniaturization of sensors for multichannel operation, and removal of motion artifacts.
- The paper states that the first-order gradiometer only suppresses common-mode magnetic noise, and that scalar OPM measurements depend on participant orientation relative to the background field, requiring an auxiliary three-axis fluxgate sensor for signal calibration.
- Evident from the text: the eMCG demonstration used only three healthy male participants aged 22–25, with no patient population; ECG and PPG failed during exercise due to vibration, so exercise-stage comparisons relied on MCG alone.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures






Verification
Claims the paper makes
9 Valid · 4 Not valid · 10 Not checked · 55 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“However, some heart abnormalities can be seen only during exercise (12), such as the common unexplained chest pain.”
§ A movable unshielded magnetocardiography system > INTRODUCTION
What the sources say
- Partly supportedcites [12] G. Massoullié, B. Boyer, V. Sapin, F. Jean, M. Andronache, M. Peoc’h, G. Clerfond, R. Eschalier, Sudden cardiac death risk in contact sports increased by myocarditis: A case series. Eur. Heart J. Case Rep. 5, ytab054 (2021). 34113757 PMC8185879 sourceExperimental — accuracy not yet measured
Case 1 describes palpitations and loss of consciousness during rugby, exclusively after direct chest-wall impact. The Conclusion likewise links ventricular arrhythmia to blunt chest trauma during sport, not to exercise alone. Our copy does not support the claim that abnormalities appear only during exercise or identify common unexplained chest pain as an example. (LLM, conf=0.78)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Therefore, for a given biomagnetic signal, the measurement results of scalar OPMs are related to the orientation of participants with respect to the background field (52).”
§ A movable unshielded magnetocardiography system > MATERIALS AND METHODS > Measuring MCG signal with scalar magnetometers
What the sources say
- Supportedcites [52] R. J. Clancy, V. Gerginov, O. Alem, S. Becker, S. Knappe, A study of scalar optically-pumped magnetometers for use in magnetoencephalography without shielding. Phys. Med. Biol. 66, 175030 (2021). PMC9273178 34325403 source
The Forward model (p. 4) says that, when the background field is much larger than the biomagnetic signal, scalar sensor readings can be treated as the biomagnetic field projected onto the background field’s direction. Bias field angle dependence (p. 11) also examines how rotating the sensor array relative to that field affects localization accuracy. (LLM, conf=0.93)
plainstates an associationn = 2- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The MCG measurements usually require the participant keeping still in a magnetically shielded room due to the immovable sensor and noisy external environments.”
§ A movable unshielded magnetocardiography system > Abstract
What the sources say
- Partly supportedcites [35] R. Zhang, W. Xiao, Y. Ding, Y. Feng, X. Peng, L. Shen, C. Sun, T. Wu, Y. Wu, Y. Yang, Z. Zheng, X. Zhang, J. Chen, H. Guo, Recording brain activities in unshielded Earth’s field with optically pumped atomic magnetometers. Sci. Adv. 6, eaba8792 (2020). 32582858 PMC7292643 sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) says that MEG measurements commonly use magnetically shielded rooms to suppress ambient noise and that fixed commercial MEG sensors limit head movement. It does not establish the same requirements for MCG measurements, which are the subject of the claim. (LLM, conf=0.85)
- Partly supportedcites [36] Y. J. Kim, I. Savukov, S. Newman, Magnetocardiography with a 16-channel fiber-coupled single-cell Rb optically pumped magnetometer. Appl. Phys. Lett. 114, 143702 (2019). sourceExperimental — accuracy not yet measured
The paper states that its human MCG experiment took place in a magnetically shielded room to avoid magnetic noise (Abstract, p. 3). It does not say participants usually must keep still or that sensors are immovable; it describes its own sensor as portable. (LLM, conf=0.85)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“The emerging optically pumped magnetometer (OPM), which is based on the Zeeman effect and light-atom interactions, is one of the most sensitive magnetometers that can reach a sensitivity as high as the SQUID sensor (32–34) without the cryogenic operational condition.”
§ A movable unshielded magnetocardiography system > INTRODUCTION
What the sources say
- Not checked — no full text of the sourcecites [32] I. Kominis, T. Kornack, J. Allred, M. V. Romalis, A subfemtotesla multichannel atomic magnetometer. Nature 422, 596– 599 (2003). 12686995 source
- Supportedcites [33] H. Dang, A. C. Maloof, M. V. Romalis, Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer. Appl. Phys. Lett. 97, 151110 (2010). source
On p. 1, the paper reports 160 aT/Hz¹ᐟ² sensitivity for its optically pumped atomic magnetometer, compared with about 1 fT/Hz¹ᐟ² for cm-sized SQUID sensors at 4.2 K. Its conclusion (p. 3) explicitly notes the absence of cryogenics. (LLM, conf=0.94)
- Supportedcites [34] D. Sheng, S. Li, N. Dural, M. V. Romalis, Subfemtotesla scalar atomic magnetometry using multipass cells. Phys. Rev. Lett. 110, 160802 (2013). 23679590 source
The introduction (p. 1) says alkali-metal magnetometers can surpass SQUIDs in sensitivity, while noting that this requires operation near zero magnetic field. The experimental description (pp. 2–3) reports optical pumping in a vapor cell heated to 120 °C, not cryogenic operation; the conclusion (p. 4) reports a sensitivity of 0.54 fT/Hz¹ᐟ². (LLM, conf=0.91)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“It is clear to see the QRS and T-waves in the MCG signals without any signal averaging.”
§ A movable unshielded magnetocardiography system > RESULTS > Resting MCG signal
boostedSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
20 more claims
Create a free account to see all 78 claims and every source verdict.
- It may state two separate claims as one. 32
- Probably not a substantial claim. 24
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The most recent innovative application of scalar OPMs comes from China, where two cesium OPMs based on the self-oscillating frequency tracking technique and a sensitivity of 140 fT/Hz1/2 have been successfully used for dynamic MCG recordings during real-life activities, including postural changes and exercise, and could be suitable long-term (Holter-like) MCG recordings (285)”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [285]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“It represents the strongest endogenous magnetic signal in the human body, up to 100 times stronger than the brain’s [5,6]”
From The Heart’s Electromagnetic Field in Emotions, Empathy and Human Connection: Biosensor-Derived Insights into Heart–Brain Axis Mechanisms and a Basis for Novel BioMagnetoTherapies (2026) · cites this paper as [5]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“It represents the strongest endogenous magnetic signal in the human body, up to 100 times stronger than the brain’s [5,6]”
From The Heart’s Electromagnetic Field in Emotions, Empathy and Human Connection: Biosensor-Derived Insights into Heart–Brain Axis Mechanisms and a Basis for Novel BioMagnetoTherapies (2026) · cites this paper as [5]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ A movable unshielded magnetocardiography system > RESULTS > Exercise MCG signal
“we select six characteristic indicators to describe the eMCG signal periods and waveforms: the cardiac period (RR interval), QRS duration, RT interval, T-wave amplitude, and R-wave amplitude.”
§ A movable unshielded magnetocardiography system > RESULTS > Exercise MCG signal
“(B to F) The left diagram of each figure shows the variation of the characteristic indicators (cardiac period, QRS duration, RT interval, T-wave amplitude, and R-wave amplitude) with time and different states.”
Citations
Cites
BioelectromagnetismPrinciples and Applications of Bioelectric and Biomagnetic Fields
Malmivuo, J., & Plonsey, R. (1995) · doi:10.1093/acprof:oso/9780195058239.001.0001 · Cited by 1,834
A subfemtotesla multichannel atomic magnetometer
Kominis, I. K., Kornack, T. W., Allred, J. C., & Romalis, M. V. (2003) · Nature · doi:10.1038/nature01484 · Cited by 1,631
Moving magnetoencephalography towards real-world applications with a wearable system
Boto, E., Holmes, N., Leggett, J., Roberts, G., Shah, V., Meyer, S. S., Muñoz, L. D., Mullinger, K. J., Tierney, T. M., Bestmann, S., et al. (2018) · Nature · doi:10.1038/nature26147 · Cited by 1,377 · cited 1×
“Although, there have been some studies on recording the MEG signal of moving participants in a MSR ( 44 ) and the unshielded static MEG measurements ( 35 , 37 ), a MEG system that is both unshielded and movable remains to be developed.”
High-Sensitivity Atomic Magnetometer Unaffected by Spin-Exchange Relaxation
Allred, J. C., Lyman, R. N., Kornack, T. W., & Romalis, M. V. (2002) · Physical Review Letters · doi:10.1103/physrevlett.89.130801 · Cited by 1,049 · cited 1×
“The OPM used in the MCG system is based on a self-oscillating frequency tracking technique (see details in Materials and Methods), which is different from the conventionally used spin-exchange relaxation-free magne-tometer ( 38 , 39 ).”
Wavelet transforms and the ECG: a review
Addison, P. S. (2005) · Physiological Measurement · doi:10.1088/0967-3334/26/5/r01 · Cited by 908 · cited 1×
“Many denoising methods have been reported to suppress the noise of ECG signals, such as filter banks, principal components analysis ( 53 ), independent component analysis ( 54 , 55 ), empirical mode decomposition (EMD) ( 56 – 58 ), and wavelet transform ( 59 ).”
Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer
Dang, H. B., Maloof, A. C., & Romalis, M. V. (2010) · Applied Physics Letters · doi:10.1063/1.3491215 · Cited by 891
ECG beat classification using PCA, LDA, ICA and Discrete Wavelet Transform
Martis, R. J., Acharya, U. R., & Min, L. C. (2013) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2013.01.005 · Cited by 722
Denoising of ECG signals based on noise reduction algorithms in EMD and wavelet domains
Kabir, M. A., & Shahnaz, C. (2012) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2011.11.003 · Cited by 444
Subfemtotesla Scalar Atomic Magnetometry Using Multipass Cells
Sheng, D., Li, S., Dural, N., & Romalis, M. V. (2013) · Physical Review Letters · doi:10.1103/physrevlett.110.160802 · Cited by 364 · cited 1×
“The emerging optically pumped magnetometer (OPM), which is based on the Zeeman effect and light-atom interactions, is one of the most sensitive magnetometers that can reach a sensitivity as high as the SQUID sensor ( 32 – 34 ) without the cryogenic operational condition.”
Spin-Exchange Shift and Narrowing of Magnetic Resonance Lines in Optically Pumped Alkali Vapors
Happer, W., & Tang, H. (1973) · Physical Review Letters · doi:10.1103/physrevlett.31.273 · Cited by 323 · cited 1×
“The OPM used in the MCG system is based on a self-oscillating frequency tracking technique (see details in Materials and Methods), which is different from the conventionally used spin-exchange relaxation-free magne-tometer ( 38 , 39 ).”
(2013) · doi:10.1017/cbo9780511846380 · Cited by 286
Principles of Operation of the Rubidium Vapor Magnetometer
Bloom, A. L. (1962) · Applied Optics · doi:10.1364/ao.1.000061 · Cited by 282 · cited 1×
“The self-oscillating magnetometer could cover the Earth ’ s field ( 40 – 42 ) rather than can only be operated in near-zero magnetic fields.”
Portable Magnetometry for Detection of Biomagnetism in Ambient Environments
Limes, M. E., Foley, E. L., Kornack, T. W., Caliga, S., McBride, S., Braun, A., Lee, W., Lucivero, V. G., & Romalis, M. V. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.14.011002 · Cited by 257 · cited 2×
“Despite there being several works on un-shielded biomagnetic signal measurements ( 35 , 37 ), these shielded or unshielded measurements normally require the participant to remain still during the measurement.”
Circulating Spike Protein Detected in Post–COVID-19 mRNA Vaccine Myocarditis
Yonker, L. M., Swank, Z., Bartsch, Y. C., Burns, M. D., Kane, A., Boribong, B. P., Davis, J. P., Loiselle, M., Novak, T., Senussi, Y., et al. (2023) · Circulation · doi:10.1161/circulationaha.122.061025 · Cited by 224 · cited 1×
“The resting ECG test, which records the ECG signal of participants in rest, is typically taken as part of the routine checkups for people in clinic applications ( 10 – 11 ).”
Recording brain activities in unshielded Earth’s field with optically pumped atomic magnetometers
Zhang, R., Xiao, W., Ding, Y., Feng, Y., Peng, X., Shen, L., Sun, C., Wu, T., Wu, Y., Yang, Y., et al. (2020) · Science Advances · doi:10.1126/sciadv.aba8792 · Cited by 222 · cited 3×
“Previously, biomagnetic signal measurements based on OPMs are often performed in magnetically shielded environments ( 35 , 36 ) due to the magnetic field noise from surroundings.”
Self-citationEMD interval thresholding denoising based on similarity measure to select relevant modes
Yang, G., Liu, Y., Wang, Y., & Zhu, Z. (2015) · Signal Processing · doi:10.1016/j.sigpro.2014.10.038 · Cited by 149
Strand, S., Lutter, W., Strasburger, J. F., Shah, V., Baffa, O., & Wakai, R. T. (2019) · Journal of the American Heart Association · doi:10.1161/jaha.119.013436 · Cited by 136 · cited 1×
“Thus, MCG is expected to be more sensitive and can be especially helpful for early diagnosis of numerous heart diseases ( 2 – 6 ) and monitoring the fetal heart health ( 7 – 9 ).”
Magnetocardiography with a 16-channel fiber-coupled single-cell Rb optically pumped magnetometer
Kim, Y. J., Savukov, I., & Newman, S. (2019) · Applied Physics Letters · doi:10.1063/1.5094339 · Cited by 133 · cited 1×
“Previously, biomagnetic signal measurements based on OPMs are often performed in magnetically shielded environments ( 35 , 36 ) due to the magnetic field noise from surroundings.”
Adaptive reduction of motion artifact in the electrocardiogram
Tong, D. A., Bartels, K. A., & Honeyager, K. S. (2002) · Proceedings of the Second Joint 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society] [Engineering in Medicine and Biology · doi:10.1109/iembs.2002.1106451 · Cited by 122 · cited 1×
“…sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters ( 46 ) and the stationary wavelet transform ( 47 , 48 ).”
Willigenburg, N. W., Daffertshofer, A., Kingma, I., & van Dieën, J. H. (2012) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2012.01.001 · Cited by 120 · cited 1×
“Many denoising methods have been reported to suppress the noise of ECG signals, such as filter banks, principal components analysis ( 53 ), independent component analysis ( 54 , 55 ), empirical mode decomposition (EMD) ( 56 – 58 ), and wavelet transform ( 59 ).”
(1974) · Journal of Perinatal Medicine · doi:10.1515/jpme.1974.2.3.214 · Cited by 112 · cited 1×
“Thus, MCG is expected to be more sensitive and can be especially helpful for early diagnosis of numerous heart diseases ( 2 – 6 ) and monitoring the fetal heart health ( 7 – 9 ).”
Kwong, J. S. W., Leithäuser, B., Park, J.-W., & Yu, C.-M. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.12.056 · Cited by 103
Fractal and EMD based removal of baseline wander and powerline interference from ECG signals
Agrawal, S., & Gupta, A. (2013) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2013.07.030 · Cited by 102 · cited 1×
“Many denoising methods have been reported to suppress the noise of ECG signals, such as filter banks, principal components analysis ( 53 ), independent component analysis ( 54 , 55 ), empirical mode decomposition (EMD) ( 56 – 58 ), and wavelet transform ( 59 ).”
BROCKMEIER, K., SCHMITZ, L., CHAVEZ, J. D. J. B., BURGHOFF, M., KOCH, H., ZIMMERMANN, R., & TRAHMS, L. (1997) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.1997.tb01824.x · Cited by 99
Hänninen, H., Takala, P., Mäkijärvi, M., Montonen, J., Korhonen, P., Oikarinen, L., Nenonen, J., Katila, T., & Toivonen, L. (2000) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2000.tb00380.x · Cited by 74 · cited 1×
“…currents and the ECG is more sensitive to the radial currents in principle ( 1 , 17 , 18 ), it is believed that the exercise MCG (eMCG) contains information complementary to the eECG ( 19 , 20 ) and shows a better sensitivity to the deviations from the normal activation direction ( 21 , 22 ).”
1-pT Noise Fluxgate Magnetometer for Geomagnetic Measurements and Unshielded Magnetocardiography
Janosek, M., Butta, M., Dressler, M., Saunderson, E., Novotny, D., & Fourie, C. (2020) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2019.2949205 · Cited by 70 · cited 1×
“Many efforts have been made to develop such kinds of mag-netometers, such as the induction coil magnetometer ( 28 ), fluxgate magnetometers ( 29 ), and magnetoresistive magnetometers ( 30 , 31 ).”
Lant, J., Stroink, G., ten Voorde, B., Horacek, B. M., & Montague, T. J. (1990) · Journal of Electrocardiology · doi:10.1016/0022-0736(90)90121-h · Cited by 63
On the fetal magnetocardiogram
Peters, M. J., Stinstra, J. G., van den Broek, S. P., Huirne, J. A. F., Quartero, H. W. F., ter Brake, H. J. M., & Rogalla, H. (1998) · Bioelectrochemistry and Bioenergetics · doi:10.1016/s0302-4598(98)00199-8 · Cited by 60
Electrically silent magnetic fields
Roth, B. J., & Wikswo, J. P. (1986) · Biophysical Journal · doi:10.1016/s0006-3495(86)83513-5 · Cited by 57
PCA and ICA applied to noise reduction in multi-lead ECG
Romero, I. (2011) · Computing in cardiology · Cited by 55
Cohen, D., Savard, P., Rifkin, R. D., Lepeschkin, E., & Strauss, W. E. (1983) · Circulation Research · doi:10.1161/01.res.53.2.274 · Cited by 51
Motion artefact removals for wearable ECG using stationary wavelet transform
Nagai, S., Anzai, D., & Wang, J. (2017) · Healthcare Technology Letters · doi:10.1049/htl.2016.0100 · Cited by 48 · cited 1×
“…sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters ( 46 ) and the stationary wavelet transform ( 47 , 48 ).”
Kwon, H., Kim, K., Lee, Y.-H., Kim, J.-M., Yu, K. K., Chung, N., & Ko, Y.-G. (2010) · Circulation Journal · doi:10.1253/circj.cj-09-0975 · Cited by 47
Kandori, A., Kanzaki, H., Miyatake, K., Hashimoto, S., Itoh, S., Tanaka, N., Miyashita, T., & Tsukada, K. (2001) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345263 · Cited by 45 · cited 1×
“However, despite the great clinical significance of movable MCG systems, few studies have been reported ( 14 , 15 , 23 – 27 ).”
Magnetocardiographic and Electrocardiographic Exercise Mapping in Healthy Subjects
Takala, P., Hänninen, H., Montonen, J., Mäkijärvi, M., Nenonen, J., Oikarinen, L., Simelius, K., Toivonen, L., & Katila, T. (2001) · Annals of Biomedical Engineering · doi:10.1114/1.1376388 · Cited by 42
Lanata, A., Guidi, A., Baragli, P., Valenza, G., & Scilingo, E. P. (2015) · PLOS ONE · doi:10.1371/journal.pone.0140783 · Cited by 42 · cited 1×
“…sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters ( 46 ) and the stationary wavelet transform ( 47 , 48 ).”
Clinical utility of magnetocardiography in cardiology for the detection of myocardial ischemia
Camm, A. J., Henderson, R., Brisinda, D., Body, R., Charles, R. G., Varcoe, B., & Fenici, R. (2019) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2019.07.009 · Cited by 42
Kariniemi, V., Ahopelto, J., Karp, P. J., & Katila, T. E. (1974) · edoc Publication server (Humboldt University of Berlin) · doi:10.18452/12635 · Cited by 41
Calibration of Room Temperature Magnetic Sensor Array for Biomagnetic Measurement
Adachi, Y., Oyama, D., Terazono, Y., Hayashi, T., Shibuya, T., & Kawabata, S. (2019) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2019.2895355 · Cited by 39 · cited 1×
“Many efforts have been made to develop such kinds of mag-netometers, such as the induction coil magnetometer ( 28 ), fluxgate magnetometers ( 29 ), and magnetoresistive magnetometers ( 30 , 31 ).”
A study of scalar optically-pumped magnetometers for use in magnetoencephalography without shielding
Clancy, R. J., Gerginov, V., Alem, O., Becker, S., & Knappe, S. (2021) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ac18fb · Cited by 37 · cited 1×
“…as B tot ¼ ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi q Therefore, for a given biomagnetic signal, the measurement results of scalar OPMs are related to the orientation of participants with respect to the background field ( 52 ).”
Active Magnetic-Field Stabilization with Atomic Magnetometer
Zhang, R., Ding, Y., Yang, Y., Zheng, Z., Chen, J., Peng, X., Wu, T., & Guo, H. (2020) · Sensors · doi:10.3390/s20154241 · Cited by 32 · cited 1×
“If we want to realize a high CMRR in a low-frequency band for MCG signals, then the OPMs should have wide bandwidths, which usually benefit the consistency of OPMs ( 51 ).”
Self-citationMagnetocardiography Using a Magnetoresistive Sensor Array
Shirai, Y., Hirao, K., Shibuya, T., Okawa, S., Hasegawa, Y., Adachi, Y., Sekihara, K., & Kawabata, S. (2019) · International Heart Journal · doi:10.1536/ihj.18-002 · Cited by 29 · cited 1×
“Many efforts have been made to develop such kinds of mag-netometers, such as the induction coil magnetometer ( 28 ), fluxgate magnetometers ( 29 ), and magnetoresistive magnetometers ( 30 , 31 ).”
Park, J.-W., Leithäuser, B., Vršansky, M., & Jung, F. (2008) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-2008-1064 · Cited by 27
Magnetocardiography and exercise testing
Brockmeier, K., Comani, S., Erné, S. N., Di Luzio, S., Pasquarelli, A., & Romani, G. L. (1994) · Journal of Electrocardiology · doi:10.1016/s0022-0736(05)80096-2 · Cited by 24 · cited 1×
“However, despite the great clinical significance of movable MCG systems, few studies have been reported ( 14 , 15 , 23 – 27 ).”
PCA and ICA applied to noise reduction in multi-lead ECG
Romero (2011) · Computing in cardiology · Cited by 24
Changes of ECG parameters after BNT162b2 vaccine in the senior high school students
Chiu, S., Chen, Y., Hsu, C.-C., Hua, Y., Tseng, W., Lu, C., Lin, M., Chen, C., Wu, M., Chen, Y., et al. (2023) · European Journal of Pediatrics · doi:10.1007/s00431-022-04786-0 · Cited by 23 · cited 1×
“The resting ECG test, which records the ECG signal of participants in rest, is typically taken as part of the routine checkups for people in clinic applications ( 10 – 11 ).”
Frequency shifts of self-oscillating magnetometer with cesium vapor
Yabuzaki, T., & Ogawa, T. (1974) · Journal of Applied Physics · doi:10.1063/1.1663412 · Cited by 22 · cited 1×
“Furthermore, the OPMs feature a low heading error ( 43 ) of less than ±0.5 nT, which is an important noise for magnetic measurement in a moving platform where the relative direction of the sensor to the magnetic field is variable.”
InfluentialMiniature Coil Array for Passive Magnetocardiography in Non-Shielded Environments
Zhu, K., Shah, A. M., Berkow, J., & Kiourti, A. (2021) · IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology · doi:10.1109/jerm.2020.3019891 · Cited by 22 · cited 1×
“Many efforts have been made to develop such kinds of mag-netometers, such as the induction coil magnetometer ( 28 ), fluxgate magnetometers ( 29 ), and magnetoresistive magnetometers ( 30 , 31 ).”
Shin, E.-S., Chung, J.-H., Park, S. G., Saleh, A., Lam, Y.-Y., Bhak, J., Jung, F., Morita, S., & Brachmann, J. (2019) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-180485 · Cited by 20
Measurement of MCG in Unshielded Environment Using a Second-Order SQUID Gradiometer
Kang, C. S., Lee, Y. H., Yu, K. K., Kwon, H., Kim, J. M., Kim, K., Lim, H. K., Park, Y. K., & Lee, S. G. (2009) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2009.2018871 · Cited by 18 · cited 1×
“The second-order gradiometer, made up of two first-order gradiometers, can greatly reduce the magnetic field gradient noise from surroundings ( 45 ).”
Bang, W.-D., Kim, K., Lee, Y.-H., Kwon, H., Park, Y., Pak, H.-N., Ko, Y.-G., Lee, M., & Joung, B. (2016) · Yonsei Medical Journal · doi:10.3349/ymj.2016.57.6.1339 · Cited by 17
Suppression of current source noise with an atomic magnetometer
Shen, L., Zhang, R., Wu, T., Peng, X., Yu, S., Chen, J., & Guo, H. (2020) · Review of Scientific Instruments · doi:10.1063/5.0009987 · Cited by 16
Self-citationRubidium Self-oscillating Magnetometer for Earth's Field Measurements
STUART, W. F., USHER, M. J., & HALL, S. H. (1964) · Nature · doi:10.1038/202076a0 · Cited by 12
Magnetocardiography in the diagnosis of fetal arrhythmias
Van Hare, G. F. (2013) · Heart Rhythm · doi:10.1016/j.hrthm.2013.06.005 · Cited by 11
Magnetocardiography detects left atrial dysfunction in paroxysmal atrial fibrillation
Her, A.-Y., Shin, E.-S., Zhou, Q., Wierzbinski, J., Vidal-Lopez, S., Saleh, A., Kim, Y. H., Garg, S., Jung, F., & Brachmann, J. (2019) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-180528 · Cited by 10 · cited 1×
“Thus, MCG is expected to be more sensitive and can be especially helpful for early diagnosis of numerous heart diseases ( 2 – 6 ) and monitoring the fetal heart health ( 7 – 9 ).”
All-optical self-oscillating 4He atomic mangnetometer with optical phase shift
Wang, H., Wu, T., Wang, H., Liu, Y., Mao, X., Peng, X., & Guo, H. (2020) · Optics Express · doi:10.1364/oe.390375 · Cited by 7 · cited 1×
“The self-oscillating magnetometer could cover the Earth ’ s field ( 40 – 42 ) rather than can only be operated in near-zero magnetic fields.”
Self-citationSudden cardiac death risk in contact sports increased by myocarditis: a case series
Massoullié, G., Boyer, B., Sapin, V., Jean, F., Andronache, M., Peoc’h, M., Clerfond, G., & Eschalier, R. (2021) · European Heart Journal - Case Reports · doi:10.1093/ehjcr/ytab054 · Cited by 7 · cited 1×
“However, some heart abnormalities can be seen only during exercise ( 12 ), such as the common unexplained chest pain.”
Changes of ECG parameters after BNT162b2 vaccine in the senior high school studentsPreprint
Chiu, S., Chen, Y., Hsu, C.-C., et al. (2022) · Research Square · doi:10.21203/rs.3.rs-2198233/v1 · Cited by 4
P6432Predictive value of rest-MCG and stress-MCG in patients with angina pectoris
Brala, D., Morita, S., Landmesser, U., & Park, J.-W. (2019) · European Heart Journal · doi:10.1093/eurheartj/ehz746.1026 · Cited by 1
Laddu, A. A., Alderson, L., & Chaitman, B. R. (2015) · Cardiovascular Medicine · doi:10.1007/978-1-4471-2828-1_8 · Cited by 1
Research works printed but not linked: 3
- J. P. Wikswo Jr Theoretical aspects of the ECG-MCG relationship in Biomagnetism An Interdisciplinary Approach S. J. Williamson G.-L. Romani L. Kaufman I. Modena Eds. (Plenum 1983) pp. 311–326.
- P. Siltanen, Magnetocardiography. Compr. Electrocardiol 2, 1405–1438 (1989).
- M. Blanco-Velasco B. Weng K. E. Barner A new ECG enhancement algorithm for stress ECG tests in 2006 Computers in Cardiology Valencia Spain (IEEE 2006) pp. 917–920.
Cited by
Clinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
All-optical dual-axis zero-field atomic magnetometer using light-shift modulation
Li, X., Han, B., Zhang, K., Liu, Z., Wang, S., Yan, Y., & Lu, J. (2024) · Physical Review Applied · doi:10.1103/physrevapplied.21.014023 · Cited by 75
Zhao, Z., Xu, X., Liu, W., et al. (2025) · Measurement · doi:10.1016/j.measurement.2025.117216 · Cited by 45
Picotesla-sensitivity microcavity optomechanical magnetometry
Hu, Z., Gao, Y., Liu, J., Yang, H., Wang, M., Lei, Y., Zhou, X. G., Li, J., Cao, X., Liang, J., et al. (2024) · Light Science & Applications · doi:10.1038/s41377-024-01643-7 · Cited by 45
“…including spin physics in condensed matter systems [1], dark matter detection [2], nuclear magnetic resonance [3], magnetocardiography [4], magnetoencephalography [5], mineral exploration [6], nondestructive inspection [7], magnetic induction tomography [8, 9], and current detection…”
Optical Parameter Decoupling and Optimization for Elliptically Polarized Atomic Magnetometer
Yan, Y., Zhang, S., Liu, Z., Zhan, D., Liu, Z., Li, X., Zhou, Y., & Lu, J. (2023) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2023.3343783 · Cited by 35
“In particular, AMs have been employed to biomagnetism research, significantly enhancing the established field of medical diagnostics that demand precise measurements of magnetic fields generated within living organisms [15], [16], [17].”
Method
Metadata
- Authors
- Xiao, W., Sun, C., Shen, L., Feng, Y., Liu, M., Wu, Y., Liu, X., Wu, T., Peng, X., & Guo, H.
- Affiliations
- Peking University (CN)
- Corresponding author
- Xiao, Wei
- Sun, Chenxi
- Wu, Teng
- Guo, Hong
- ORCID
- Venue
- Science Advances
- Published
- 2023-03-31
- Volume
- 9
- Issue
- 13
- Pages
- eadg1746-eadg1746
- ISSN
- 2375-2548 (ISSN-L)
- PMC
- PMC10058232
- OpenAlex
- W4361302002
- Cited by
- 116
- References
- 62
- Citations by year
- 2026: 37 · 2025: 41 · 2024: 27 · 2023: 11
- FWCI
- 13.44
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by-nc
- Locations
- Science Advances (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- National Natural Science Foundation of China (61531003, 91436210, 62071012, 61571018)
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Keywords
- magnetocardiography, optically pumped magnetometers, heart disease diagnosis, cardiac arrhythmias, ischemia
- MeSH terms
- Arrhythmias, Cardiac, Exercise Test, Heart, Heart Diseases, Humans, Magnetocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.