Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Open accessThe DOI matches its Crossref record, with a small difference in title or year; 65 of the 65 references it cites were found in a registry or an index.
Abstract
In general, patients with diabetes performing self-monitoring of blood glucose (SMBG) can strongly rely on the accuracy of measurement results. However, various factors such as application errors, extreme environmental conditions, extreme hematocrit values, or medication interferences may potentially falsify blood glucose readings. Incorrect blood glucose readings may lead to treatment errors, for example, incorrect insulin dosing. Therefore, the diabetes team as well as the patients should be well informed about limitations in blood glucose testing. The aim of this publication is to review the current knowledge on limitations and interferences in blood glucose testing with the perspective of their clinical relevance.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Newer blood glucose monitors tend to use a different reagent (glucose dehydrogenase, with or without a cofactor, rather than glucose oxidase). This reduces the susceptibility to interference from haematocrit, the partial pressure of oxygen, and most drugs (aside from icodextrin, a non–glucose-containing peritoneal dialysate).13”
From Blood glucose monitoring devices: current considerations (2023) · cites this paper as [13]
In the citing paper
“Numerous interfering factors have been reported to influence the results of blood glucose meters (135,136).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [135]
Citations
Cites
International Diabetes Federation 2017
Carracher, A. M., Marathe, P. H., & Close, K. L. (2018) · Journal of Diabetes · doi:10.1111/1753-0407.12644 · Cited by 13,418
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Standards of Medical Care in Diabetes
Power, D. (2006) · Diabetes Care · doi:10.2337/diacare.29.02.06.dc05-1593 · Cited by 8,673
Cosentino, F., Grant, P. J., Aboyans, V., Bailey, C. J., Ceriello, A., Delgado, V., Federici, M., Filippatos, G., Grobbee, D. E., Hansen, T. B., et al. (2020) · European Heart Journal · doi:10.1093/eurheartj/ehz486 · Cited by 4,763
(2013) · European Heart Journal · doi:10.1093/eurheartj/eht108 · Cited by 1,955
(2014) · Revista Española de Cardiología (English Edition) · doi:10.1016/j.rec.2013.10.013 · Cited by 609
Standards of Medical Care in Diabetes—2016: Summary of Revisions
(2016) · Diabetes Care · doi:10.2337/dc16-s003 · Cited by 448
Polonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Petersen, B., Schweitzer, M., & Wagner, R. S. (2011) · Diabetes Care · doi:10.2337/dc10-1732 · Cited by 429
Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406 · cited 5×
“Inappropriate handling of SMBG has been identified as the most common factor affecting BG results; more than 90% of overall inaccuracies result from incorrect use of BG meters.(4,18) Due to the minute blood samples utilized by mod-”
BackgroundInfluentialFactors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348 · cited 7×
“As a consequence, high levels of triglycerides may result in falsely low BG readings.(23) In BG systems utilizing GO-based test strips, the potential interference of pO 2 has been reported.”
BackgroundInfluentialTang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management
Dungan, K., Chapman, J., Braithwaite, S. S., & Buse, J. (2007) · Diabetes Care · doi:10.2337/dc06-1679 · Cited by 292
Montesinos, P., Lorenzo, I., Martin, G., Sanz, J., Perez-Sirvent, M. L., Martinez, D., Orti, G., Algarra, L., Martinez, J., Moscardo, F., et al. (2008) · Haematologica · doi:10.3324/haematol.11575 · Cited by 231 · cited 1×
“Uric acid is a DNA degradation product, and as such, very high uric acid concentrations, exceeding 20 mg/dL, can be observed in patients under chemotherapy, radiation therapy, or cancers with rapid cell turnover, such as certain lymphomas.(58) At very high levels, uric acid may be oxidized by the electrode, thereby po…”
BackgroundTang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · American Journal of Clinical Pathology · doi:10.1309/qaw1-x5xw-bvrq-5lkq · Cited by 207 · cited 1×
“A study on 30 substances—including, among others, acetaminophen (paracetamol), acetylsalicylic acid, various antibiotics, heparin, and warfarin—found glucose measurements of 6 handheld BG meters to be potentially affected by ascorbic acid, acetaminophen, dopamine, maltose, and mannitol.(50) Depending on the BG system…”
BackgroundTang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00038 · Cited by 188 · cited 3×
“With a GO-based measurement, however, an increase in pO 2 to >100 Torr (eg, in critically ill patients receiving oxygen treatment) may result in a remarkable underestimation of BG values.(22,54) An evaluation of 5 SMBG systems utilizing a GO enzyme reaction on test strips showed BG measurements to be affected by pO 2…”
BackgroundMethodGuideline for management of postmeal glucose in diabetes
(2014) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2012.08.002 · Cited by 183
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Bergenstal, R. M., Ahmann, A. J., Bailey, T., Beck, R. W., Bissen, J., Buckingham, B., Deeb, L., Dolin, R. H., Garg, S. K., Goland, R., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700234 · Cited by 166 · cited 1×
“In particular, patients with insulintreated diabetes perform SMBG as a substantial element of daily management of diabetes.(14,15) The term “BG system” denotes the combination of a BG meter and test strips, and both determine analytical performance.”
BackgroundDURÁN, A., MARTÍN, P., RUNKLE, I., PÉREZ, N., ABAD, R., FERNÁNDEZ, M., DEL VALLE, L., SANZ, M. F., & CALLE‐PASCUAL, A. L. (2010) · Journal of Diabetes · doi:10.1111/j.1753-0407.2010.00081.x · Cited by 156
Franciosi, M., Lucisano, G., Pellegrini, F., Cantarello, A., Consoli, A., Cucco, L., Ghidelli, R., Sartore, G., Sciangula, L., Nicolucci, A., et al. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03268.x · Cited by 137
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
Bergenstal, R., Pearson, J., Cembrowski, G. S., Bina, D., Davidson, J., & List, S. (2000) · The Diabetes Educator · doi:10.1177/014572170002600610 · Cited by 108
Bosi, E., Scavini, M., Ceriello, A., Cucinotta, D., Tiengo, A., Marino, R., Bonizzoni, E., Giorgino, F., & on behalf of the PRISMA Study Group (2013) · Diabetes Care · doi:10.2337/dc13-0092 · Cited by 105
Breton, M. D., & Kovatchev, B. P. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400309 · Cited by 103 · cited 1×
“Simulations suggest an increasing likelihood of treatment errors in response to decreasing accuracy of BG systems.(26) Even when used by trained laboratory professionals, every BG meter entails a certain degree of imprecision and bias associated with it.”
BackgroundKlonoff, D. C., Blonde, L., Cembrowski, G., Chacra, A. R., Charpentier, G., Colagiuri, S., Dailey, G., Gabbay, R. A., Heinemann, L., Kerr, D., et al. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500630 · Cited by 97 · cited 5×
“Inappropriate handling of SMBG has been identified as the most common factor affecting BG results; more than 90% of overall inaccuracies result from incorrect use of BG meters.(4,18) Due to the minute blood samples utilized by mod-”
BackgroundInfluentialSelf-citationBaumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., & Freckmann, G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600511 · Cited by 90 · cited 2×
“A potential concern involving test strips is the variance between test strip lots (lot-to-lot variation).(31) In a study, 4 test strip lots for each of 5 different BG systems were evaluated, including measurement results ranging from <50 mg/dL to >400 mg/dL.”
BackgroundSelf-citationSelf-Monitoring of Blood Glucose in Type 2 Diabetes: Recent Studies
Schnell, O., Alawi, H., Battelino, T., Ceriello, A., Diem, P., Felton, A.-M., Grzeszczak, W., Harno, K., Kempler, P., Satman, I., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700225 · Cited by 88 · cited 1×
“SMBG is recommended to be performed in a structured approach.(2,5,8,9) It is reported to be only useful when blood glucose (BG) data are interpreted and utilized for immediate therapeutic actions.”
BackgroundSelf-citationMoore, K., Vizzard, N., Coleman, C., McMahon, J., Hayes, R., & Thompson, C. J. (2001) · Diabetic Medicine · doi:10.1046/j.0742-3071.2001.00568.x · Cited by 76
Self-Monitoring of Blood Glucose: The Use of the First or the Second Drop of Blood
Hortensius, J., Slingerland, R. J., Kleefstra, N., Logtenberg, S. J. J., Groenier, K. H., Houweling, S. T., & Bilo, H. J. G. (2011) · Diabetes Care · doi:10.2337/dc10-1694 · Cited by 76 · cited 1×
“considerable amounts of glucose on the skin, thereby causing falsely high SMBG results.(17,19,20) In daily practice, a substan-”
BackgroundHematocrit Interference of Blood Glucose Meters for Patient Self-Measurement
Ramljak, S., Lock, J. P., Schipper, C., Musholt, P. B., Forst, T., Lyon, M., & Pfützner, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700123 · Cited by 72 · cited 2×
“A study on 19 SMBG systems found that only a few meters were unaffected by hematocrit interference.(49) Nevertheless, modern BG systems correcting automatically for hematocrit are considered to be less susceptible to high or low hematocrit.”
BackgroundÖberg, D., & Östenson, C.-G. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1261 · Cited by 71
Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815580160 · Cited by 70
Self-citationKempf, K., Kruse, J., & Martin, S. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0008 · Cited by 66
Heinemann, L. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0076 · Cited by 62 · cited 10×
“ably when stored in closed vials than in open vials, which is of special importance when used under extreme environmental conditions.(17) The impact of inappropriate storage of test strips on BG readings is demonstrated by the case of a 72-year-old Japanese patient with type 2 diabetes.”
BackgroundInfluentialSelf-citationEffect of High Altitude on Blood Glucose Meter Performance
Fink, K. S., Christensen, D. B., & Ellsworth, A. J. (2002) · Diabetes Technology & Therapeutics · doi:10.1089/152091502320798259 · Cited by 57
Effect of Ambient Temperature on Analytical Performance of Self-Monitoring Blood Glucose Systems
Nerhus, K., Rustad, P., & Sandberg, S. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0255 · Cited by 51 · cited 2×
“Conversely, after a rapid change from 30°C to room temperature, 4 of the 9 BG systems presented downward discrepancies >5%.(35) A period of acclimatization (up to 15-30 minutes) seemed to lessen these effects.”
BackgroundHirose, T., Mita, T., Fujitani, Y., Kawamori, R., & Watada, H. (2011) · Diabetes Care · doi:10.2337/dc10-1705 · Cited by 50 · cited 1×
“considerable amounts of glucose on the skin, thereby causing falsely high SMBG results.(17,19,20) In daily practice, a substan-”
BackgroundAdverse Impact of Temperature and Humidity on Blood Glucose Monitoring Reliability: A Pilot Study
Haller, M. J., Shuster, J. J., Schatz, D., & Melker, R. J. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0051 · Cited by 49 · cited 4×
“SMBG.(39) High temperature and humidity, but mostly temperature were found to affect the reliability of many BG meters.”
BackgroundMethodInfluentialChallenges to Glycemic Measurement in the Perioperative and Critically Ill Patient: A Review
Pitkin, A. D., & Rice, M. J. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300606 · Cited by 47 · cited 3×
“Particularly in critically ill patients, an incorrect diagnosis of hypoglycemia due to poor peripheral perfusion (eg, circulatory shock) may be harmful.(34) Peripheral hypoperfusion may result in an increased tissue glucose extraction and a lower glucose value in capillary in comparison to venous blood.”
BackgroundFreckmann, G., Link, M., Schmid, C., Pleus, S., Baumstark, A., & Haug, C. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0085 · Cited by 46
Self-citationBaum, J. M., Monhaut, N. M., Parker, D. R., & Price, C. P. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.347 · Cited by 42 · cited 1×
“Incorrect coding may lead to measurement errors of ±30% or more.(21) However, most modern BG systems no longer require a coding step.”
BackgroundInfluence of Simulated Altitude on the Performance of Five Blood Glucose Meters
Gautier, J., Bigard, X. A., Douce, P., Duvallet, A. L., & Cathelineau, G. (1996) · Diabetes Care · doi:10.2337/diacare.19.12.1430 · Cited by 42
Louie, R. F., Sumner, S. L., Belcher, S., Mathew, R., Tran, N. K., & Kost, G. J. (2009) · Disaster Medicine and Public Health Preparedness · doi:10.1097/dmp.0b013e3181979a06 · Cited by 39 · cited 3×
“In addition, there is a need for complete, comprehensible, and clear information in the labeling, for example, about an oxygen dependency, operational limits or storage under extreme conditions, such as high and low temperature and humidity.(22,37) It is pertinent to note, however, that many patients (and members of t…”
BackgroundResultExtreme Altitude Mountaineering and Type 1 Diabetes
Pavan, P., Sarto, P., Merlo, L., Casara, D., Ponchia, A., Biasin, R., Noventa, D., & Avogaro, A. (2003) · Diabetes Care · doi:10.2337/diacare.26.11.3196 · Cited by 35
Blevins, T. (2013) · Postgraduate Medicine · doi:10.3810/pgm.2013.05.2668 · Cited by 34 · cited 1×
“In particular, patients with insulintreated diabetes perform SMBG as a substantial element of daily management of diabetes.(14,15) The term “BG system” denotes the combination of a BG meter and test strips, and both determine analytical performance.”
BackgroundSchmid, C., Haug, C., Heinemann, L., & Freckmann, G. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0047 · Cited by 34 · cited 10×
“For this reason, patients should be instructed to use BG systems within the specified operating temperature range only.(22) Nowadays, many modern BG systems have a built-in temperature sensor that utilizes the measured temperature to correct the glucose measurement result.”
BackgroundMethodInfluentialSelf-citationAccuracy of Handheld Blood Glucose Meters at High Altitude
de Mol, P., Krabbe, H. G., de Vries, S. T., Fokkert, M. J., Dikkeschei, B. D., Rienks, R., Bilo, K. M., & Bilo, H. J. G. (2010) · PLoS ONE · doi:10.1371/journal.pone.0015485 · Cited by 34 · cited 1×
“Comparable results have been obtained in another study using a hypobaric chamber.(41,45)”
ResultBudiman, E. S., Samant, N., & Resch, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700213 · Cited by 30 · cited 1×
“Erroneous BG readings may result in contraindicated treatment decisions.(61,62) Small measurement errors might not impact insulin dosing.”
BackgroundBilen, H., Kiliçaslan, A., Akcay, G., & Albayrak, F. (2007) · Journal of Medical Engineering & Technology · doi:10.1080/03091900600861590 · Cited by 28 · cited 1×
“In a study conducted in Tanzania, 3 BG meters taken to Mt Kilimanjaro (>5800 m) showed BG readings of 50, 214, and 367 mg/dL on the same sample.(40) Comparable results have been obtained in another study using a hypobaric chamber.”
ResultLyon, M. E., & Lyon, A. W. (2011) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2011.01.005 · Cited by 28 · cited 1×
“Hematocrit levels outside the reference interval are reported to be more prevalent than expected, even in Western countries.(49,52) In general, low hematocrit values (< 35%) frequently result in too high readings, while an increase in hematocrit is associated with a decrease in BG readings.”
BackgroundBaumstark, A., Schmid, C., Pleus, S., Haug, C., & Freckmann, G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700611 · Cited by 26 · cited 2×
“On the other hand, low oxygen levels (eg, in patients with severe chronic obstructive pulmonary disease) might help deliver falsely high BG values.(23,56) On the other hand, a study using GD test strips showed them not to be significantly affected by different oxygen pressure.”
BackgroundMethodSelf-citationPartial Pressure of Oxygen in Capillary Blood Samples from the Fingertip
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700627 · Cited by 23 · cited 1×
“A broad range of capillary pO 2 values was demonstrated to occur in daily clinical practice.(57) Uric acid is a DNA degradation product, and as such, very high uric acid concentrations, exceeding 20 mg/dL, can be observed in patients under chemotherapy, radiation therapy, or cancers with rapid cell turnover, such as c…”
BackgroundSelf-citationLam, M., Louie, R. F., Curtis, C. M., Ferguson, W., Vy, J. H., Truong, A.-T., Sumner, S. L., & Kost, G. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813514325 · Cited by 23
Solnica, B., SKUPIEN, J. K., Kuśnierz‐Cabala, B., Słowińska-Solnica, K., Witek, P., Cempa, A., & Malecki, M. T. (2011) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2011.770 · Cited by 21 · cited 1×
“In general, low hematocrit values (< 35%) frequently result in too high readings, while an increase in hematocrit is associated with a decrease in BG readings.(33,53) A study on 19 SMBG systems found that only a few meters were unaffected by hematocrit interference.”
BackgroundSchnell, O., Hinzmann, R., Kulzer, B., Freckmann, G., Erbach, M., Lodwig, V., & Heinemann, L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700619 · Cited by 21 · cited 3×
“For instance, the need for adequate dosing of insulin heavily depends on reliable glucose information.(8) In particular, patients with insulintreated diabetes perform SMBG as a substantial element of daily management of diabetes.”
BackgroundSelf-citationIn This Issue of Diabetes Care
(2016) · Diabetes Care · doi:10.2337/dc16-ti01 · Cited by 18
Schnell, O., Alawi, H., Battelino, T., Ceriello, A., Diem, P., Felton, A., Grzeszczak, W., Harno, K., Kempler, P., Satman, I., et al. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600323 · Cited by 14
Self-citationCook and Chill: Effect of Temperature on the Performance of Nonequilibrated Blood Glucose Meters
Deakin, S., Steele, D., Clarke, S., Gribben, C., Bexley, A.-M., Laan, R., & Kerr, D. E. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815598775 · Cited by 12
Mortazavi, S., Gholampour, M., Haghani, M., Mortazavi, G., & Mortazavi, A. (2014) · PubMed Central · Cited by 12
Parameters Affecting Postprandial Blood Glucose: Effects of Blood Glucose Measurement Errors
Koschinsky, T., Heckermann, S., & Heinemann, L. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200109 · Cited by 10 · cited 2×
“However, if they are large, clinically relevant insulin doses might be administered.(61) Even relatively small measurement errors may add up to a substancial total deviation (Figure 1).”
BackgroundSelf-citationControl Solutions for Blood Glucose Meters
Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815587602 · Cited by 8 · cited 1×
“In addition, the implementation of inmeter safety features (ie, validity of test strips check) has further increased the safety of these devices.(16) Consequently, patients with appropriate training and a good performance of BG testing can typically rely on the precision of BG measurement results.”
BackgroundSelf-citationOverestimation by a hand-held glucometer of blood glucose level due to icodextrin.
King, D., Ericson, R. P., & Todd, N. W. (2010) · PubMed · Cited by 7
Jadhav, P. P., & Jadhav, M. (2013) · International Journal of Case Reports and Images · doi:10.5348/ijcri-2013-09-362-cr-6 · Cited by 5 · cited 2×
“High uric acid and low hematocrit values have been suggested to cause falsely high BG readings, thereby resulting in inappropriate therapeutic decisions.(59)”
BackgroundA case of false hypoglycemia by SMBG due to improper storage of glucometer test strips
Tamaki, M., Kanazawa, A., Shirakami, A., Yamaguchi, H., Tamura, Y., Mita, T., Fujitani, Y., & Watada, H. (2013) · Diabetology International · doi:10.1007/s13340-013-0147-6 · Cited by 4 · cited 2×
“Furthermore, BG measurement may be compromised by usage of deteriorated test strips, which may result from inappropriate storage, mechanical stress, or usage after the expiry date.(18,23,24) BG test strips were found to perform more reli-”
BackgroundGlucose Information for Tight Glycemic Control: Different Methods with Different Challenges
Weber, C., & Neeser, K. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400529 · Cited by 4 · cited 2×
“Accuracy of SMBG measurement may also be affected by high or low hematocrit values.(18,23,33,34,51) Hematocrit levels outside the reference interval are reported to be more prevalent than expected, even in Western countries.”
Background
Research works printed but not linked: 3
- Downie P., 2013, Diabetes Primary Care, 15, 149
- Mortazavi S, 2014, J Biomed Phys Eng, 4, 111
- King DA, 2010, Isr Med Assoc J, 12, 314
Cited by
From pre-diabetes to diabetes: diagnosis, treatments and translational research
Khan, R., Chua, Z., Tan, J., Yang, Y., Liao, Z., & Zhao, Y. (2019) · Medicina · doi:10.3390/medicina55090546 · Cited by 484
“[117] However, what cannot be avoided during the patient’s self-monitoring of blood glucose (SMBC) is the need for finger pricking.”
BackgroundSacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Glycemic Monitoring and Management in Advanced Chronic Kidney Disease
Galindo, R. J., Beck, R. W., Scioscia, M. F., Umpierrez, G. E., & Tuttle, K. R. (2020) · Endocrine Reviews · doi:10.1210/endrev/bnaa017 · Cited by 202
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 206
Lorestani, F., Zhang, X., Abdullah, A. M., Xin, X., Liu, Y., Rahman, M. M., Biswas, M. A. S., Li, B., Dutta, A., Niu, Z., et al. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202306117 · Cited by 132
Metadata
- Authors
- Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O.
- Affiliations
- Sciarc Institute, Baierbrunn, Germany
- Institut für Diabetes-Technologie (Germany) (DE)
- Roche Diabetes Care GmbH (Germany) (DE)
- Forschungsinstitut der Diabetes Akademie Mergentheim (DE)
- Diabetes Clinic for Children and Adolescents, Muenster, Germany
- Science & Co GmbH, Düsseldorf, Germany
- Forschergruppe Diabetes e.V., Munich-Neuherberg, Germany
- Corresponding author
- Schnell, Oliver
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2016-07-09
- Volume
- 10
- Issue
- 5
- Pages
- 1161-1168
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC5032951
- OpenAlex
- W2340313604
- Cited by
- 115
- References
- 65
- Citations by year
- 2026: 12 · 2025: 18 · 2024: 16 · 2023: 18 · 2022: 9 · 2021: 3 · 2020: 13 · 2019: 10 · 2018: 7 · 2017: 8 · 2016: 1
- FWCI
- 1.28
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- self-monitoring of blood glucose, blood glucose monitoring, measurement interference, diabetes mellitus, insulin dosing errors, hematocrit, environmental factors, patient education, treatment errors
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Female, Humans, Male, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.