Photoplethysmography in Wearable Devices: A Comprehensive Review of Technological Advances, Current Challenges, and Future Directions
Open accessThe DOI matches its Crossref record; 27 of the 27 references it cites were found in a registry or an index.
Abstract
Photoplethysmography (PPG) is an affordable and straightforward optical technique used to detect changes in blood volume within tissue microvascular beds. PPG technology has found widespread application in commercial medical devices, enabling measurements of oxygen saturation, blood pressure, and cardiac output; the assessment of autonomic nerve function; and the diagnosis of peripheral vascular disease. Recently, the growing demand for non-invasive, portable, cost-effective technology, along with advancements in small semiconductor components, has led to the integration of PPG into various wrist-worn wearable devices. Multiple sensor structures have been proposed and, through appropriate signal processing and algorithmic application, these wearable devices can measure a range of health indicators during daily life. This paper begins by addressing the market status of wrist-worn wearable devices, followed by an explanation of the fundamental principles underlying light operation and its interaction with living tissue for PPG measurements. Moving on to technological advancements, the paper addresses the analog front end for the measurement of the PPG signal, sensor configurations with multiple light emitters and receivers, the minimum sampling rate required for low-power systems, and the measurement of stress, sleep, blood pressure, blood glucose, and activity using PPG signals. Several challenges in the field are also identified, including selecting the appropriate wavelength for the PPG sensor’s light source, developing low-power interpolation methods to extract high-resolution inter-beat intervals at a low sampling rate, and exploring the measurement of physiological phenomena using multi-wavelength PPG signals simultaneously collected at the same location. Lastly, the paper presents future research directions, which encompass the development of new, reliable parameters specific to wearable PPG devices and conducting studies in real-world scenarios, such as 24-h long-term measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · cited 3×
“PPG is a non-invasive optical technology that measures changes in blood volume in tissue by detecting changes in the amount of light absorbed by the tissue [11].”
Bashkatov, A. N., Genina, E. A., Kochubey, V. I., & Tuchin, V. V. (2005) · Journal of Physics D: Applied Physics · doi:10.1088/0022-3727/38/15/004 · Cited by 1,720
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945 · cited 1×
“PPG sensors can be used to measure a variety of physiological parameters, including the heart rate, heart rate variability, blood pressure, blood oxygen saturation, and sleep quality [12,13].”
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · cited 2×
“In early 2010, as wrist-wearable devices with PPG sensors began to be commercialized, a review of wearable PPG systems was published in the journal Electronics [16].”
Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., et al. (2020) · JMIR mHealth and uHealth · doi:10.2196/18694 · Cited by 715 · cited 1×
“In a study conducted by Fuller et al. in 2020, the heart rate measurement accuracy of commercial wearable devices was investigated by analyzing data from 158 publications [50].”
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661 · cited 1×
“Pulse rate variability (PRV) measurement is a commonly used parameter in addition to pulse rate measurement [51,52].”
Optical properties of human skin
Lister, T., Wright, P. A., & Chappell, P. H. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.9.090901 · Cited by 467 · cited 2×
“Research suggests that around 4–7% of incident light is reflected from the skin surface, irrespective of the wavelength [69].”
DistancePPG: Robust non-contact vital signs monitoring using a camera
Kumar, M., Veeraraghavan, A., & Sabharwal, A. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.001565 · Cited by 417 · cited 2×
“Multi-wavelength PPG sensing technology is widely recognized as a robust method for the measurement of PPG signals [92,93].”
Hamaoka, T., McCully, K. K., Quaresima, V., Yamamoto, K., & Chance, B. (2007) · Journal of Biomedical Optics · doi:10.1117/1.2805437 · Cited by 329
Miller, D. J., Sargent, C., & Roach, G. D. (2022) · Sensors · doi:10.3390/s22166317 · Cited by 328 · cited 1×
“Wearable healthcare devices can help to detect and monitor a range of health conditions, such as heart disease, diabetes, and sleep disorders [4–6].”
Nelson, B. W., & Allen, N. B. (2019) · JMIR mHealth and uHealth · doi:10.2196/10828 · Cited by 310
Steven Eyobu, O., & Han, D. S. (2018) · Sensors · doi:10.3390/s18092892 · Cited by 259 · cited 1×
“Together, these sensors provide valuable information about the wearer’s physical activity and movement patterns, which can be used for healthcare applications, such as steps taken, distance traveled, calories burned, and fall detection [8–10].”
The advantages of wearable green reflected photoplethysmography
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9506-z · Cited by 209
Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., & Yamakoshi, T. (2013) · 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609852 · Cited by 202 · cited 1×
“Green light, which has shallower tissue penetration compared to wavelengths such as red or infrared [38,82], is less susceptible to motion artifacts.”
Wearable-band type visible-near infrared optical biosensor for non-invasive blood glucose monitoring
Rachim, V. P., & Chung, W.-Y. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.01.121 · Cited by 200 · cited 1×
“The results demonstrated a mean correlation coefficient (R) of 0.86 between the estimated and baseline blood glucose concentrations, with a standard error of prediction (SPE) of 6.16 mg/dL [100].”
InfluentialCuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180
The effect of contacting force on photoplethysmographic signals
Teng, X. F., & Zhang, Y. T. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/5/020 · Cited by 171
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155
Lee, J., Kim, M., Park, H.-K., & Kim, I. Y. (2020) · Sensors · doi:10.3390/s20051493 · Cited by 153
Motion Artifact Reduction for Wrist-Worn Photoplethysmograph Sensors Based on Different Wavelengths
Zhang, Y., Song, S., Vullings, R., Biswas, D., Simões-Capela, N., van Helleputte, N., van Hoof, C., & Groenendaal, W. (2019) · Sensors · doi:10.3390/s19030673 · Cited by 140
Choi, A., & Shin, H. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa5efa · Cited by 116
Multi-wavelength photoplethysmography method for skin arterial pulse extraction
Liu, J., Yan, B. P.-Y., Dai, W.-X., Ding, X.-R., Zhang, Y.-T., & Zhao, N. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.004313 · Cited by 107
Gorny, A. W., Liew, S. J., Tan, C. S., & Müller-Riemenschneider, F. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.8233 · Cited by 94
Overview of Artificial Intelligence–Driven Wearable Devices for Diabetes: Scoping Review
Ahmed, A., Aziz, S., Abd-alrazaq, A., Farooq, F., & Sheikh, J. (2022) · Journal of Medical Internet Research · doi:10.2196/36010 · Cited by 57
Yan, L., Hu, S., Alzahrani, A., Alharbi, S., & Blanos, P. (2017) · Biosensors · doi:10.3390/bios7020022 · Cited by 40
Current Status and Prospects of Health-Related Sensing Technology in Wearable Devices
Cho, J. (2019) · Journal of Healthcare Engineering · doi:10.1155/2019/3924508 · Cited by 33
Effects of skin surface temperature on photoplethysmograph
Jeong, I. C., Yoon, H., Kang, H., & Yeom, H. (2014) · Journal of Healthcare Engineering · doi:10.1260/2040-2295.5.4.429 · Cited by 33 · cited 2×
“Furthermore, studies have indicated that the use of an optimal wavelength can enhance the signal-to-noise ratio (SNR), even in scenarios where reduced blood perfusion at low temperatures leads to a decrease in the SNR [95,96].”
Cited by
Kostrna, J., Oparina, E., Palacios, C., Rodriguez, A. J., Pei, J., Ajmal, A., & Ramella-Roman, J. C. (2026) · Sensors · doi:10.3390/s26102922 · Cited by 5
Metadata
- Authors
- Kim, K. B., & Baek, H. J.
- Venue
- Electronics
- Publisher
- MDPI AG
- Published
- 2023-07-03
- Volume
- 12
- Issue
- 13
- Pages
- 2923
- ISSN
- 2079-9292 (ISSN-L)
- OpenAlex
- W4383068587
- Cited by
- 212
- References
- 27
- FWCI
- 15.55
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.