Degeneration and regeneration in rabbit peripheral nerve with long-term nerve cuff electrode implant: A stereological study of myelinated and unmyelinated axons
Abstract
Selective and dynamically co-ordinated functional electrical stimulation (FES) of paralysed/paretic limbs in upper motor neuron lesioned people depends on optimal contact at the neural interface. Implanted nerve cuff electrodes may form a stable electrical neural interface, but may also inflict nerve damage. In this study the immediate and long-term effects of cuff implantation on the number and sizes of myelinated and unmyelinated axons have been evaluated with unbiased stereological techniques. Cuff electrodes were implanted in rabbit tibial nerves just below the knee joint, and the stereological analyses were carried out 2 weeks and 16 months after implantation. Myelinated axons were analysed at standardised levels proximal to, underneath, and distal to the cuff; unmyelinated axons underneath the cuff. A 27% loss of myelinated axons was found underneath and distal to the nerve cuff 2 weeks post surgery. All axonal sizes were equally lost except for the very smallest. At 16 months post surgery the number of myelinated axons was restored to control values at both levels. Except for the presence of regenerative sprouts at 2 weeks post surgery, no changes in the number or sizes of unmyelinated axons were revealed at either 2 weeks or 16 months post surgery. Our study demonstrates that implanted cuff electrodes may cause an initial loss of myelinated axons but with subsequent regeneration.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Chronic implantation of snug cuffs modifies nerve shape and produces loss of large nerve fibers, which are the most sensitive to compression.
“Although snug-fitting nerve cuffs have been advocated to reduce the stimulus charge injection or to obtain a high signal-to-noise ratio for neural recordings (Naples et al., 1990) , different studies have shown that chronic implantation of snug cuffs modifies the nerve shape and produces a loss of large nerve fibers, which are the most sensitive to compression (Krarup et al., 1989; Grill and Mort”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Larsen et al., 1998
Neural damage induced by implantation and presence of an extraneural electrode is reversible.
“Girsch et al. and Larsen et al. showed that neural damage induced by the implantation and presence of an extraneural electrode is reversible [53, 54].”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [54]
In rabbit tibial nerve, full regeneration occurred after an initial loss of 27% of myelinated axons following cuff electrode implantation.
“Larsen et al. evaluated axonal loss in the rabbit tibial nerve 2 weeks and 16 months after implantation of cuff electrodes, and demonstrated full regeneration after an initial loss of 27% of myelinated axons.”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [54]
These damages could be reversible if no further stimulation is applied.
“These damages could be reversible if no further stimulation is applied [54, 113].”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [54]
These drawbacks may induce irreversible damage to the nerves and cause serious biological issues, including inflammation, demyelination, axonal degradation, and blood vein compression.
“These drawbacks may induce irreversible damage to the nerves and cause serious biological issues, including inflammation, demyelination, axonal degradation, and blood vein compression (14–17).”
From Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording (2019) · cites this paper as [16]
Cites
GUNDERSEN, H. J. G., BENDTSEN, T. F., KORBO, L., MARCUSSEN, N., MØLLER, A., NIELSEN, K., NYENGAARD, J. R., PAKKENBERG, B., SØRENSEN, F. B., VESTERBY, A., et al. (1988) · APMIS · doi:10.1111/j.1699-0463.1988.tb05320.x · Cited by 2,929 · see citing works
Gundersen, H. J. G. (1986) · PubMed · Cited by 1,659 · see citing works
Notes on the estimation of the numerical density of arbitrary profiles: the edge effect
Gundersen, H. J. G. (1977) · Journal of Microscopy · doi:10.1111/j.1365-2818.1977.tb00062.x · Cited by 1,265 · see citing works
Selective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · see citing works
Stimulation of rat sciatic nerve regeneration with pulsed electromagnetic fields
Sisken, B. F., Kanje, M., Lundborg, G., Herbst, E., & Kurtz, W. H. (1989) · Brain Research · doi:10.1016/0006-8993(89)90575-1 · Cited by 164 · see citing works
Stereology of nerve cross sections
Larsen, J. O. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00129-0 · Cited by 134 · see citing works
Self-citationCuff electrodes for long-term recording of natural sensory information
Struijk, J. J., Thomsen, M., Larsen, J. O., & Sinkjaer, T. (1999) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.765194 · Cited by 97 · see citing works
Self-citationA flexible method for fabrication of nerve cuff electrodes
Haugland, M. K. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.656992 · Cited by 79 · see citing works
Self-citationNatural neural sensing and artificial muscle control in man
Sinkjær, T., Haugland, M. K., & Haase, J. (1994) · Experimental Brain Research · doi:10.1007/bf00233991 · Cited by 69 · see citing works
Self-citationTm, M. (1988) · PubMed · Cited by 61 · see citing works
Conditioning Effect Induced by Chronic Nerve Compression
Dahlin, L. B., & Kanje, M. (1992) · Scandinavian Journal of Plastic and Reconstructive Surgery and Hand Surgery · doi:10.3109/02844319209035181 · Cited by 42 · see citing works
Flagged first, then by citations
1 of the 12 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · see citing works
Soft and elastic hydrogel-based microelectronics for localized low-voltage neuromodulation
Liu, Y., Liu, J., Chen, S., Lei, T., Kim, Y., Niu, S., Wang, H., Wang, X., Foudeh, A. M., Tok, J. B.-H., et al. (2019) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0335-6 · Cited by 811 · see citing works
Neuromuscular electrical stimulation in neurorehabilitation
Sheffler, L. R., & Chae, J. (2007) · Muscle & Nerve · doi:10.1002/mus.20758 · Cited by 529 · see citing works
Implanted Neural Interfaces: Biochallenges and Engineered Solutions
Grill, W. M., Norman, S. E., & Bellamkonda, R. V. (2009) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-061008-124927 · Cited by 488 · see citing works
Functionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398 · see citing works
Zhang, Y., Zheng, N., Cao, Y., Wang, F., Wang, P., Ma, Y., Lu, B., Hou, G., Fang, Z., Liang, Z., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw1066 · Cited by 314 · see citing works
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · see citing works
Polyimide cuff electrodes for peripheral nerve stimulation
Rodri´guez, F. J., Ceballos, D., Schu¨ttler, M., Valero, A., Valderrama, E., Stieglitz, T., & Navarro, X. (2000) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(00)00192-8 · Cited by 207 · see citing works
Electroconductive, Adhesive, Non‐Swelling, and Viscoelastic Hydrogels for Bioelectronics
Han, I. K., Song, K., Jung, S., Jo, Y., Kwon, J., Chung, T., Yoo, S., Jang, J., Kim, Y., Hwang, D. S., et al. (2022) · Advanced Materials · doi:10.1002/adma.202203431 · Cited by 170 · see citing works
Chronic Response of the Rat Sciatic Nerve to the Flat Interface Nerve Electrode
Tyler, D. J., & Durand, D. M. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1569263 · Cited by 144 · see citing works
Neural stimulation with optical radiation
Richter, C., Matic, A. I., Wells, J. D., Jansen, D., & Walsh, J. T. (2010) · Laser & Photonics Review · doi:10.1002/lpor.200900044 · Cited by 145 · see citing works
Ortiz-Catalan, M., Brånemark, R., Håkansson, B., & Delbeke, J. (2012) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-11-33 · Cited by 144 · see citing works
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
The foreign body response to the Utah Slant Electrode Array in the cat sciatic nerve
Christensen, M. B., Pearce, S. M., Ledbetter, N. M., Warren, D. J., Clark, G. A., & Tresco, P. A. (2014) · Acta Biomaterialia · doi:10.1016/j.actbio.2014.07.010 · Cited by 135 · see citing works
Stereology of nerve cross sections
Larsen, J. O. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00129-0 · Cited by 134 · see citing works
Self-citationXiang, Z., Yen, S., Sheshadri, S., Wang, J., Lee, S., Liu, Y., Liao, L., Thakor, N. V., & Lee, C. (2016) · Advanced Materials · doi:10.1002/adma.201503423 · Cited by 116 · see citing works
Cuff electrodes for long-term recording of natural sensory information
Struijk, J. J., Thomsen, M., Larsen, J. O., & Sinkjaer, T. (1999) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.765194 · Cited by 97 · see citing works
Self-citationMultiple-electrode nerve cuffs for low-velocity and velocity-selective neural recording
Taylor, J. T., Donaldson, N., & Winter, J. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02347545 · Cited by 79 · see citing works
Nimodipine promotes regeneration and functional recovery after intracranial facial nerve crush
Mattsson, P., Janson, A. M., Aldskogius, H., & Svensson, M. A. (2001) · The Journal of Comparative Neurology · doi:10.1002/cne.1273 · Cited by 79 · see citing works
Soft Electronics Based on Stretchable and Conductive Nanocomposites for Biomedical Applications
Zambrano, B. L., Renz, A. F., Ruff, T., Lienemann, S., Tybrandt, K., Vörös, J., & Lee, J. (2020) · Advanced Healthcare Materials · doi:10.1002/adhm.202001397 · Cited by 74 · see citing works
Frost, C. M., Ursu, D. C., Flattery, S. M., Nedic, A., Hassett, C. A., Moon, J. D., Buchanan, P. J., Brent Gillespie, R., Kung, T. A., Kemp, S. W. P., et al. (2018) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0452-1 · Cited by 68 · see citing works
Riso, R. R., Mosallaie, F., Jensen, W., & Sinkjær, T. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.847826 · Cited by 64 · see citing works
Self-citationPyridoxine-induced toxicity in rats: a stereological quantification of the sensory neuropathy
Perry, T., Weerasuriya, A., Mouton, P. R., Holloway, H. W., & Greig, N. H. (2004) · Experimental Neurology · doi:10.1016/j.expneurol.2004.07.013 · Cited by 60 · see citing works
PERIPHERAL NERVE AND MUSCLE STIMULATION
Mortimer, J. T., & Bhadra, N. (2004) · Series on bioengineering and biomedical engineering · doi:10.1142/9789812561763_0020 · Cited by 58 · see citing works
Cutrone, A., & Micera, S. (2019) · Advanced Healthcare Materials · doi:10.1002/adhm.201801345 · Cited by 57 · see citing works
A Parylene Self-Locking Cuff Electrode for Peripheral Nerve Stimulation and Recording
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2014.2333733 · Cited by 56 · see citing works
Long‐term biocompatibility of implanted polymer‐based intrafascicular electrodes
Lawrence, S. M., Larsen, J. O., Horch, K. W., Riso, R., & Sinkjær, T. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10303 · Cited by 55 · see citing works
Self-citationInfrared neural stimulation: Beam path in the guinea pig cochlea
Moreno, L. E., Rajguru, S. M., Matic, A. I., Yerram, N. K., Robinson, A. M., Hwang, M., Stock, S. R., & Richter, C. (2011) · Hearing Research · doi:10.1016/j.heares.2011.06.006 · Cited by 52 · see citing works
Romero, E., Denef, J. F., Delbeke, J., Robert, A., & Veraart, C. (2001) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345271 · Cited by 49 · see citing works
Development and characterization of a chronic implant mouse model for vagus nerve stimulation
Mughrabi, I. T., Hickman, J. L., Jayaprakash, N., Thompson, D., Ahmed, U., Papadoyannis, E. S., Chang, Y.-C., Abbas, A., Datta, T., Chang, E. H., et al. (2021) · eLife · doi:10.7554/elife.61270 · Cited by 47 · see citing works
Interfacing the body’s own sensing receptors into neural prosthesis devices
Haugland, M. K., & Sinkjær, T. (1999) · Technology and Health Care · doi:10.3233/thc-1999-7603 · Cited by 46 · see citing works
Self-citationBiomechanical and functional variation in rat sciatic nerve following cuff electrode implantation
Restaino, S. M., Abliz, E., Wachrathit, K., Krauthamer, V., & Shah, S. B. (2014) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-11-73 · Cited by 44 · see citing works
Chronic histological effects of the flat interface nerve electrode
Leventhal, D. K., Cohen, M., & Durand, D. M. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/2/004 · Cited by 43 · see citing works
Cobo, A., Larson, C. E., Scholten, K., Miranda, J. A., Elyahoodayan, S., Song, D. B., Pikov, V., & Meng, E. F. (2018) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2018.2881908 · Cited by 43 · see citing works
Moldovan, M. (2006) · Brain · doi:10.1093/brain/awl184 · Cited by 43 · see citing works
Self-Closed Parylene Cuff Electrode for Peripheral Nerve Recording
Kang, X., Liu, J., Tian, H., Yang, B. T., NuLi, Y., & Yang, C. (2015) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2014.2381634 · Cited by 39 · see citing works
Fabrication of the Composite Regenerative Peripheral Nerve Interface (C-RPNI) in the Adult Rat
Svientek, S. R., Ursu, D. C., Cederna, P. S., & Kemp, S. W. P. (2020) · Journal of Visualized Experiments · doi:10.3791/60841 · Cited by 37 · see citing works
Vince, V., Thil, M.-A., Gérard, A.-C., Veraart, C., Delbeke, J., & Colin, I. M. (2005) · Journal of Neuroimmunology · doi:10.1016/j.jneuroim.2004.10.010 · Cited by 35 · see citing works
Thil, M., Duy, D. T., Colin, I. M., & Delbeke, J. (2007) · Journal of Neuroimmunology · doi:10.1016/j.jneuroim.2007.01.021 · Cited by 34 · see citing works
Vince, V., Brelen, M. E., Delbeke, J., & Colin, I. M. (2005) · Journal of Neuroimmunology · doi:10.1016/j.jneuroim.2005.04.019 · Cited by 30 · see citing works
Influence of nerve cuff channel count and implantation site on the separability of afferent ENG
Silveira, C. R. A., Brunton, E., Spendiff, S., & Nazarpour, K. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aabca0 · Cited by 29 · see citing works
Interfacing with the nervous system: a review of current bioelectric technologies
Sahyouni, R., Mahmoodi, A., Chen, J. W., Chang, D. T., Moshtaghi, O., Djalilian, H. R., & Lin, H. W. (2017) · Neurosurgical Review · doi:10.1007/s10143-017-0920-2 · Cited by 27 · see citing works
Gan, L. S., Ravid, E., Kowalczewski, J. A., Olson, J. L., Morhart, M. J., & Procházka, A. (2011) · Neurorehabilitation and neural repair · doi:10.1177/1545968311420443 · Cited by 27 · see citing works
Long-term Cuff Electrode Recordings from Peripheral Nerves in Animals and Humans
Sinkjær, T., Haugland, M. K., Struijk, J. J., & Riso, R. R. (1999) · Springer eBooks · doi:10.1007/978-3-642-58552-4_29 · Cited by 26 · see citing works
Self-citationSacral Neuromodulation in Lower Urinary Tract Dysfunction
Vignes, J., de Sèze, M., Dobremez, E., Joseph, P. A., & Guerin, J. B. (2005) · Advances and technical standards in neurosurgery · doi:10.1007/3-211-27208-9_5 · Cited by 22 · see citing works
Microscale Electrode Implantation during Nerve Repair
Urbanchek, M. G., Wei, B. P. C., Egeland, B. M., Abidian, M. R., Kipke, D. R., & Cederna, P. S. (2011) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0b013e3182268ac8 · Cited by 19 · see citing works
The foreign body response and morphometric changes associated with mesh-style peripheral nerve cuffs
Christensen, M. B., & Tresco, P. A. (2018) · Acta Biomaterialia · doi:10.1016/j.actbio.2017.11.059 · Cited by 19 · see citing works
A Preclinical Study of Laryngeal Motor-Evoked Potentials as a Marker Vagus Nerve Activation
Grimonprez, A., Raedt, R., Taeye, L. D., Larsen, L. E., Delbeke, J., Boon, P., & Vonck, K. E. J. (2015) · International Journal of Neural Systems · doi:10.1142/s0129065715500343 · Cited by 18 · see citing works
Multichannel neural cuff electrodes with integrated multiplexer circuit
Schüettler, M., Koch, K. P., Stieglitz, T., Scholz, O., Haberer, W., Keller, R., & Meyer, J.-U. (2000) · International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine and Biology (MMB) · doi:10.1109/mmb.2000.893861 · Cited by 15 · see citing works
PERIPHERAL NERVE RECORDING ELECTRODES AND TECHNIQUES
Yoshida, K., & Riso, R. (2004) · Series on bioengineering and biomedical engineering · doi:10.1142/9789812561763_0021 · Cited by 15 · see citing works
Stereology of Neural Connections: An Overview
Avendaño, C. (2006) · Springer eBooks · doi:10.1007/0-387-28942-9_16 · Cited by 14 · see citing works
Integrating Sensory Nerve Signals Into Neural Prosthesis Devices
Sinkjær, T. (2000) · Neuromodulation Technology at the Neural Interface · doi:10.1046/j.1525-1403.2000.00035.x · Cited by 14 · see citing works
Self-citationStimulation of Neurons with Infrared Radiation
Zhao, K., Tan, X., Young, H. K., & Richter, C. (2016) · Springer eBooks · doi:10.1007/978-1-4939-1758-7_17 · Cited by 13 · see citing works
Two-way communication for programming and measurement in a miniature implantable stimulator
Thil, M., Gérard, B., Jarvis, J. C., & Delbeke, J. (2005) · Medical & Biological Engineering & Computing · doi:10.1007/bf02344736 · Cited by 12 · see citing works
A fully implanted drug delivery system for peripheral nerve blocks in behaving animals
Pohlmeyer, E. A., Jordon, L. R., Kim, P., & Miller, L. E. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.06.006 · Cited by 12 · see citing works
A Nerve Cuff Electrode for Controlled Reshaping of Nerve Geometry
Caparso, A., Durand, D. M., & Mansour, J. M. (2008) · Journal of Biomaterials Applications · doi:10.1177/0885328208097426 · Cited by 12 · see citing works
Zhang, W., Li, C., Yang, S., Xu, C., Wang, W., Nyengaard, J. R., & Tang, Y. C. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.04.005 · Cited by 11 · see citing works
Electrical and Physical Sensors for Biomedical Implants
Kassanos, P., Anastasova, S., & Yang, G. (2018) · Springer eBooks · doi:10.1007/978-3-319-69748-2_3 · Cited by 11 · see citing works
Du, J., Morales, A., Paknahad, J., Kosta, P., Bouteiller, J. C., Fernández, E., & Lazzi, G. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9631068 · Cited by 11 · see citing works
Schiønning, J. D. (2000) · Apmis · doi:10.1111/j.1600-0463.2000.tb05696.x · Cited by 11 · see citing works
Liu, J., Grayden, D. B., Keast, J. R., & John, S. E. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/aca69e · Cited by 10 · see citing works
Long-term implantable flexible neural interfaces for electrophysiological monitoring
Chen, B., Lu, J., Chen, Z., Han, X., Sun, Y., Lin, X., Tang, Z., Jia, S., Xie, G., Huo, F., et al. (2024) · Journal of Materials Chemistry C · doi:10.1039/d4tc04289a · Cited by 10 · see citing works
Lahiri, A., Delgado-Martínez, I., Sheshadri, S., Ng, K., Nag, S., Yen, S., & Thakor, N. V. (2015) · Muscle & Nerve · doi:10.1002/mus.24928 · Cited by 9 · see citing works
Histopathologic features of the vagus nerve after electrical stimulation in swine.
Martín-Portugués, I. D.-G., Sanchez-Margallo, F. M., Sánchez-Gijón, S. P., Ayala, V. C., & Gargallo, J. U. (2005) · PubMed · doi:10.14670/hh-20.851 · Cited by 9 · see citing works
Nerve cuffs for nerve repair and regeneration
Hoffer, J. A., & Kallesøe, K. (2000) · Progress in brain research · doi:10.1016/s0079-6123(00)28012-6 · Cited by 9 · see citing works
Sammut, S., Koh, R. G. L., & Zariffa, J. (2021) · Sensors · doi:10.3390/s21020506 · Cited by 8 · see citing works
Thil, M., Vince, V., Veraart, C., Delbeke, J., & Colin, I. M. (2005) · Journal of Neuroimmunology · doi:10.1016/j.jneuroim.2005.06.011 · Cited by 7 · see citing works
Kumar, L., Gupta, R., Puri, M. M., Jaiswal, A., Srinath, Murar, A., & Behera, D. (2011) · PubMed · Cited by 6 · see citing works
Flexible Electrode for Implantable Neural Devices
Tsang, W. M., & Je, M. (2014) · Springer eBooks · doi:10.1007/978-1-4614-8151-5_6 · Cited by 6 · see citing works
An implant for long-term cervical vagus nerve stimulation in micePreprint
Mughrabi, I. T., Hickman, J. L., Jayaprakash, N., Papadoyannis, E. S., Abbas, A., Chang, Y.-C., Lee, S., Datta, T., Chang, E. H., Zanos, T. P., et al. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.06.20.160473 · Cited by 6 · see citing works
Feasibility of endovascular stimulation of the femoral nerve using a stent-mounted electrode array
Liu, J., Grayden, D. B., Keast, J. R., Booth, L. C., May, C. N., & John, S. E. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad4f16 · Cited by 6 · see citing works
Effects of chronic nerve cuff and intramuscular electrodes on rat triceps surae motor units
Carp, J. S., Chen, X.-Y., Sheikh, H., & Wolpaw, J. R. (2001) · Neuroscience Letters · doi:10.1016/s0304-3940(01)02183-8 · Cited by 5 · see citing works
Endovascular stimulation of the pudendal nerve using a stent-mounted electrode array
Liu, J., Grayden, D. B., Keast, J. R., & John, S. E. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad6590 · Cited by 4 · see citing works
Koppaka, S. (2013) · OhioLink ETD Center (Ohio Library and Information Network) · Cited by 4 · see citing works
Xiang, Z., Yen, S.-C., Sheshadri, S., Xue, N., Lee, S.-H., Wang, J., Thakor, N. V., & Lee, C. (2015) · 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319062 · Cited by 4 · see citing works
Wu, Y., & Guo, L. (2020) · Neural Interface Engineering · doi:10.1007/978-3-030-41854-0_5 · Cited by 3 · see citing works
Schiefer, M. A. (2009) · OhioLink ETD Center (Ohio Library and Information Network) · Cited by 3 · see citing works
Development of regenerative peripheral nerve interfaces for motor control of neuroprosthetic devices
Kemp, S. W. P., Urbanchek, M. G., Irwin, Z. T., Chestek, C. A., & Cederna, P. S. (2017) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2264130 · Cited by 2 · see citing works
Neural Interfacing with the Peripheral Nervous System: A FINE Approach
Durand, D. M., Wodlinger, B., & Park, H. (2012) · Springer eBooks · doi:10.1007/978-1-4614-5227-0_16 · Cited by 2 · see citing works
Bioelectric source localization in peripheral nerves.
Zariffa, J. (2010) · Library and Archives Canada (Government of Canada) · Cited by 4 · see citing works
Bolat, D., Bahar, S., KÜRÜM, A., & Gültiken, M. E. (2013) · Kafkas Universitesi Veteriner Fakultesi Dergisi · doi:10.9775/kvfd.2012.7919 · Cited by 1 · see citing works
Silveira, C. R. A., Brunton, E., Spendiff, S., & Nazarpour, K. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512751 · Cited by 1 · see citing works
Wang, R., Yu, H., & Li, Z. (2017) · Toxinology · doi:10.1007/978-981-10-2798-7_41-1 · Cited by 1 · see citing works
Sabetian, P. (2015) · TSpace (University of Toronto) · Cited by 1 · see citing works
Cox, T. G. H. (2017) · Purdue e-Pubs (Purdue University System) · doi:10.7912/c2gd32 · Cited by 1 · see citing works
Kameswaran, N. (2010) · ScholarlyCommons (University of Pennsylvania)
Historical Background from 1986 – 2005 for the Development of the ActiGait System
Haase, J., Haugland, M. K., & Sinkjær, T. (2014) · Biosystems & biorobotics · doi:10.1007/978-3-319-08072-7_61
Self-citationApplication of sensory nerve signals in neural prosthesis devices
Sinkjær, T. (1999) · VBN Forskningsportal (Aalborg Universitet)
Self-citationPeripheral Nerve Signal Processing, Multipole Cuff Methods
Yoo, P. (2015) · Encyclopedia of Computational Neuroscience · doi:10.1007/978-1-4614-6675-8_216
Wang, R., Yu, H., & Li, Z. (2018) · Micro/Nano technologies · doi:10.1007/978-981-10-5945-2_41
Peripheral Nerve Signal Processing, Multipole Cuff Methods
Yoo, P. (2022) · Encyclopedia of Computational Neuroscience · doi:10.1007/978-1-0716-1006-0_216
Peripheral Nerve Signal Processing, Multipole Cuff Methods
Yoo, P. (2014) · Springer eBooks · doi:10.1007/978-1-4614-7320-6_216-1
Advancing Peripheral Nerve Interfaces in a Large Animal Model
Andreis, F. R. (2022) · doi:10.54337/aau478976327
Nikam, P. L., Gayke, A. U., Avhad, P. S., Bhabad, R. B., & Malviya, R. (2023) · Wiley eBooks · doi:10.1002/9781394200344.ch10
Zhao, Q., Liu, Q., Li, B., Yang, X., Yan, J., Zou, J., Zhang, R., Tian, G., Cui, M., Wang, Q., et al. (2026) · Advanced Materials · doi:10.1002/adma.73685
- Authors
- Larsen, J. O., Thomsen, M., Haugland, M., & Sinkjær, T.
- Affiliations
- Aalborg University Hospital (DK)
- Aalborg University (DK)
- Venue
- Acta Neuropathologica
- Published
- 1998-10-09
- Volume
- 96
- Issue
- 4
- Pages
- 365-378
- ISSN
- 0001-6322 (ISSN-L)
- 1432-0533
- OpenAlex
- W2024281494
- Cited by
- 96
- References
- 12
- Citations by year
- 2026: 1 · 2024: 3 · 2023: 1 · 2022: 4 · 2021: 3 · 2020: 4 · 2019: 5 · 2018: 6 · 2017: 6 · 2016: 1 · 2015: 7 · 2014: 6 · 2013: 2 · 2012: 2
- FWCI
- 1.84
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Acta Neuropathologica (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Nerve injury and regeneration
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- cuff electrodes, peripheral nerve injury, axonal loss, axon regeneration, myelinated axons, unmyelinated axons, axon diameter, functional electrical stimulation, neural interfaces, tibial nerve, rabbit model
- MeSH terms
- Anatomy, Artistic, Animals, Axons, Electric Stimulation, Electrodes, Implanted, Female, Nerve Degeneration, Nerve Fibers, Nerve Fibers, Myelinated, Nerve Regeneration, Peripheral Nerves, Rabbits, Time Factors
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review