Modelling the physics in the iterative reconstruction for transmission computed tomography
Open accessThe DOI matches its Crossref record; 129 of the 129 references it cites were found in a registry or an index.
Abstract
There is an increasing interest in iterative reconstruction (IR) as a key tool to improve quality and increase applicability of x-ray CT imaging. IR has the ability to significantly reduce patient dose; it provides the flexibility to reconstruct images from arbitrary x-ray system geometries and allows one to include detailed models of photon transport and detection physics to accurately correct for a wide variety of image degrading effects. This paper reviews discretization issues and modelling of finite spatial resolution, Compton scatter in the scanned object, data noise and the energy spectrum. The widespread implementation of IR with a highly accurate model-based correction, however, still requires significant effort. In addition, new hardware will provide new opportunities and challenges to improve CT with new modelling.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Radiation Detection and Measurement
(2013) · Physics for Radiation Protection · doi:10.1002/9783527667062.ch12 · Cited by 6,776
Bayesian-Based Iterative Method of Image Restoration*
Richardson, W. H. (1972) · Journal of the Optical Society of America · doi:10.1364/josa.62.000055 · Cited by 4,160
An iterative technique for the rectification of observed distributions
Lucy, L. B. (1974) · The Astronomical Journal · doi:10.1086/111605 · Cited by 3,866
Gordon, R., Bender, R., & Herman, G. T. (1970) · Journal of Theoretical Biology · doi:10.1016/0022-5193(70)90109-8 · Cited by 2,791 · cited 1×
“IR typically uses either no preferential weights (such as in ART-type algorithms (Gordon et al 1970)), treating all measurements equally, or statistical weights (such as in ML-type algorithms (Lange and Carson 1984, De Man et al 2000) or weighted least-squares approaches (Sauer and Bouman 1993,…”
MethodEnergy-selective reconstructions in X-ray computerised tomography
Alvarez, R. E., & Macovski, A. (1976) · Physics in Medicine & Biology · doi:10.1088/0031-9155/21/5/002 · Cited by 2,013
Sidky, E. Y., & Pan, X. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/17/021 · Cited by 1,995
Andersen, A. (1984) · Ultrasonic Imaging · doi:10.1016/0161-7346(84)90008-7 · Cited by 1,812
EM reconstruction algorithms for emission and transmission tomography.
Lange, K. L., & Carson, R. E. (1984) · PubMed · Cited by 1,769
Fast calculation of the exact radiological path for a three‐dimensional CT array
Siddon, R. L. (1985) · Medical Physics · doi:10.1118/1.595715 · Cited by 1,535
(2008) · Choice Reviews Online · doi:10.5860/choice.45-6220 · Cited by 1,093
Accurate image reconstruction from few-views and limited-angle data in divergent-beam CT
Sidky, E. Y., Kao, C.-M., & Pan, X. (2006) · Journal of X-Ray Science and Technology: Clinical Applications of Diagnosis and Therapeutics · doi:10.3233/xst-2006-00155 · Cited by 1,091
A three‐dimensional statistical approach to improved image quality for multislice helical CT
Thibault, J., Sauer, K. D., Bouman, C. A., & Hsieh, J. (2007) · Medical Physics · doi:10.1118/1.2789499 · Cited by 937
X-ray phase-contrast imaging: from pre-clinical applications towards clinics
Bravin, A., Coan, P., & Suortti, P. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/1/r1 · Cited by 846
Beam hardening in X-ray reconstructive tomography
Brooks, R. A., & Chiro, G. D. (1976) · Physics in Medicine and Biology · doi:10.1088/0031-9155/21/3/004 · Cited by 721 · cited 1×
“The polychromatic nature of standard x-ray sources not only complicates the noise statistics, if uncorrected it also leads to beam hardening and image artefacts (cupping and shadows) (McDavid et al 1975, Brooks and Chiro 1976).”
BackgroundStatistical image reconstruction for polyenergetic X-ray computed tomography
Elbakri, I. A., & Fessler, J. A. (2002) · IEEE Transactions on Medical Imaging · doi:10.1109/42.993128 · Cited by 691 · cited 1×
“…above (object discretization, detector resolution and measurement statistics) to develop model-based image reconstruction methods that ‘correct’ for the polyenergetic spectrum during the iterations (De Man et al 2001, Elbakri and Fessler 2002 and 2003a, O’Sullivan and Benac 2003, Lasio et al 2007).”
MethodSelf-citationKalender, W. A. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/13/r03 · Cited by 689
Jørgensen, B. (1987) · Journal of the Royal Statistical Society Series B: Statistical Methodology · doi:10.1111/j.2517-6161.1987.tb01685.x · Cited by 679 · cited 1×
“As a more tractable approach, Xu and Tsui (2007, 2009) proposed an exponential dispersion approximation (Jorgensen 1987) for which Var{Ki} = φ(Eμ[Ki]), where p > 1 and φ > 0 is the dispersion parameter.”
BackgroundCone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
Siewerdsen, J. H., & Jaffray, D. A. (2001) · Medical Physics · doi:10.1118/1.1339879 · Cited by 612
Penalized weighted least-squares image reconstruction for positron emission tomography
Fessler, J. A. (1994) · IEEE Transactions on Medical Imaging · doi:10.1109/42.293921 · Cited by 611
Self-citationPan, X., Sidky, E. Y., & Vannier, M. (2009) · Inverse Problems · doi:10.1088/0266-5611/25/12/123009 · Cited by 572 · cited 2×
“The very fact that the object is discretized into a finite number of basis functions inherently leads to discrepancies between the measured projection data and the simulated re-projections estimated during the reconstruction (Zbijewski and Beekman 2004a, Pan et al 2009, Herman and Davidi 2008).”
BackgroundPenalized weighted least-squares image reconstruction for dual energy X-ray transmission tomography
Sukovic, P., & Clinthorne, N. H. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.896783 · Cited by 552
Statistical inverse problems: Discretization, model reduction and inverse crimes
Kaipio, J., & Somersalo, E. (2007) · Journal of Computational and Applied Mathematics · doi:10.1016/j.cam.2005.09.027 · Cited by 540 · cited 3×
“…are sufficient for investigating stability, upper performance bounds and theoretical aspects of a reconstruction algorithm (Bian et al 2013, Sidky and Pan 2008), they are likely to overestimate an algorithm’s performance compared to its behaviour with real data (Kaipio and Somersalo 2007).”
BackgroundOptimal short scan convolution reconstruction for fan beam CT
Parker, D. L. (1982) · Medical Physics · doi:10.1118/1.595078 · Cited by 525 · cited 1×
“In FBP, redundant data are handled by a projection domain or image domain weighting factors, such as Parker weighting (Parker 1982) and helical weighting (Wang et al 1993).”
MethodA local update strategy for iterative reconstruction from projections
Sauer, K., & Bouman, C. (1993) · IEEE Transactions on Signal Processing · doi:10.1109/78.193196 · Cited by 494 · cited 4×
“…treating all measurements equally, or statistical weights (such as in ML-type algorithms (Lange and Carson 1984, De Man et al 2000) or weighted least-squares approaches (Sauer and Bouman 1993, Sukovic and Clinthorne 2000)), emphasizing more reliable data and de-emphasizing less reliable data.”
BackgroundMethodInfluentialMonte Carlo Principles and Neutron Transport Problems
Coveyou, R. R. (1971) · Nuclear Science and Engineering · doi:10.13182/nse71-a22388 · Cited by 470
Distance-driven projection and backprojection in three dimensions
Man, B. D., & Basu, S. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/11/024 · Cited by 459
An iterative maximum-likelihood polychromatic algorithm for CT
De Man, B., Nuyts, J., Dupont, P., Marchal, G., & Suetens, P. (2001) · IEEE Transactions on Medical Imaging · doi:10.1109/42.959297 · Cited by 430 · cited 2×
“…above (object discretization, detector resolution and measurement statistics) to develop model-based image reconstruction methods that ‘correct’ for the polyenergetic spectrum during the iterations (De Man et al 2001, Elbakri and Fessler 2002 and 2003a, O’Sullivan and Benac 2003, Lasio et al 2007).”
BackgroundMethodSelf-citationNoise and Edge Artifacts in Maximum-Likelihood Reconstructions for Emission Tomography
Snyder, D. L., Miller, M. I., Thomas, L. J., & Politte, D. G. (1987) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.1987.4307831 · Cited by 421
An Improved Algorithm for Reprojecting Rays through Pixel Images
Joseph, P. M. (1982) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.1982.4307572 · Cited by 375
Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT
Bian, J., Siewerdsen, J. H., Han, X., Sidky, E. Y., Prince, J. L., Pelizzari, C. A., & Pan, X. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/22/001 · Cited by 373 · cited 1×
“In addition, since IR methods are inherently better suited to handling scanning geometries with missing data (Zbijewski and Beekman 2004b, Bian et al 2010), they could potentially be applied to projections obtained with beam blockers without the need for the interpolation in blocker shadows, such…”
MethodBadal, A., & Badano, A. (2009) · Medical Physics · doi:10.1118/1.3231824 · Cited by 358 · cited 1×
“One important recent development is the implementation of MC simulation of xray photon propagation on graphics processing units (Badal and Badano 2009).”
BackgroundDose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements
Kalender, W. A., Wolf, H., & Suess, C. (1999) · Medical Physics · doi:10.1118/1.598738 · Cited by 345 · cited 1×
“Often, x-ray CT systems include deliberate tube current modulation to reduce x-ray dose while maintaining image quality (Gies et al 1999, Kalender et al 1999, Greess et al 2000, Hurwitz et al 2009, Angel et al 2009) .”
BackgroundA general cone-beam reconstruction algorithm
Wang, G., Lin, T.-H., Cheng, P., & Shinozaki, D. M. (1993) · IEEE Transactions on Medical Imaging · doi:10.1109/42.241876 · Cited by 333 · cited 3×
“An additional advantage of footprint-based (back)projectors is that they avoid the creation of Moiré patterns which are often produced by algorithms derived with straightforward discretization (De Man and Basu 2004). The footprint depends on the basis function assumed for the voxel. For traditional pixels, De Man and…”
BackgroundMethodX‐ray scatter correction algorithm for cone beam CT imaging
Ning, R., Tang, X., & Conover, D. (2004) · Medical Physics · doi:10.1118/1.1711475 · Cited by 322 · cited 1×
“…acquisition of only a small subset of projections with the blocker in place and the computation of the global scatter estimate by the interpolation (Ning et al 2004), by using blockers only in a first sequence of scans (prior scan, e.g. in the monitoring of radiation therapy treatment) (Niu et al…”
MethodImage recovery from data acquired with a charge-coupled-device camera
Snyder, D. L., White, R. L., & Hammoud, A. M. (1993) · Journal of the Optical Society of America A · doi:10.1364/josaa.10.001014 · Cited by 318
Practical considerations for 3-D image reconstruction using spherically symmetric volume elements
Matej, S., & Lewitt, R. M. (1996) · IEEE Transactions on Medical Imaging · doi:10.1109/42.481442 · Cited by 314
Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies
Gies, M., Kalender, W. A., Wolf, H., Suess, C., & Madsen, M. T. (1999) · Medical Physics · doi:10.1118/1.598779 · Cited by 306 · cited 1×
“Often, x-ray CT systems include deliberate tube current modulation to reduce x-ray dose while maintaining image quality (Gies et al 1999, Kalender et al 1999, Greess et al 2000, Hurwitz et al 2009, Angel et al 2009) .”
BackgroundFlat-detector computed tomography (FD-CT)
Kalender, W. A., & Kyriakou, Y. (2007) · European Radiology · doi:10.1007/s00330-007-0651-9 · Cited by 302 · cited 1×
“In the last 20 years, flat-panel detectors have been introduced for planar and tomographic 0031-9155/13/120063+34$33.00 © 2013 Institute of Physics and Engineering in Medicine Printed in the UK & the USA R63 x-ray imaging (Kalender and Kyriakou 2007).”
BackgroundInfluentialMultidimensional digital image representations using generalized Kaiser–Bessel window functions
Lewitt, R. M. (1990) · Journal of the Optical Society of America A · doi:10.1364/josaa.7.001834 · Cited by 292
Iterative reconstruction for helical CT: a simulation study
Nuyts, J., Man, B. D., Dupont, P., Defrise, M., Suetens, P., & Mortelmans, L. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/4/003 · Cited by 291 · cited 1×
“…than that assumed by the reconstructor, which usually involves a denser voxel grid, but often also denser detector sampling, depending on the chosen mechanism for forward projection (De Man et al 2000, De Man et al 2001, Nuyts et al 1998, Zbijewski and Beekman 2004a, Elbakri and Fessler 2003a).”
BackgroundSelf-citationSiewerdsen, J. H., Daly, M. J., Bakhtiar, B., Moseley, D. J., Richard, S., Keller, H., & Jaffray, D. A. (2006) · Medical Physics · doi:10.1118/1.2148916 · Cited by 283
A Method for Correcting Bone Induced Artifacts in Computed Tomography Scanners
Joseph, P. M., & Spital, R. D. (1978) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-197801000-00017 · Cited by 281 · cited 1×
“Based on these ideas, numerous correction methods for analytical reconstruction have been proposed (Joseph and Spital 1978, Herman 1979, Herman and Trivedi 1983, Joseph and Ruth 1997, Hsieh et al 2000, Yan et al 2000, Kyriakou et al 2010, Liu et al 2009).”
MethodAn iterative approach to the beam hardening correction in cone beam CT
Hsieh, J., Molthen, R. C., Dawson, C. A., & Johnson, R. H. (2000) · Medical Physics · doi:10.1118/1.598853 · Cited by 280 · cited 1×
“Based on these ideas, numerous correction methods for analytical reconstruction have been proposed (Joseph and Spital 1978, Herman 1979, Herman and Trivedi 1983, Joseph and Ruth 1997, Hsieh et al 2000, Yan et al 2000, Kyriakou et al 2010, Liu et al 2009).”
MethodThe Use of Sieves to Stabilize Images Produced with the EM Algorithm for Emission Tomography
Snyder, D. L., & Miller, M. I. (1985) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1985.4334521 · Cited by 278
Faigel, G., & Tegze, M. (1999) · Reports on Progress in Physics · doi:10.1088/0034-4885/62/3/002 · Cited by 266
Multi-Material Decomposition Using Statistical Image Reconstruction for Spectral CT
Long, Y. Q., & Fessler, J. A. (2014) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2014.2320284 · Cited by 252
Self-citationImproved scatter correction using adaptive scatter kernel superposition
Sun, M., & Star-Lack, J. M. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/22/007 · Cited by 233
Rührnschopf, E., & Klingenbeck, K. (2011) · Medical Physics · doi:10.1118/1.3599033 · Cited by 232
Greess, H., Wolf, H., Baum, U., Lell, M., Pirkl, M., Kalender, W., & Bautz, W. A. (2000) · European Radiology · doi:10.1007/s003300050062 · Cited by 230 · cited 1×
“Often, x-ray CT systems include deliberate tube current modulation to reduce x-ray dose while maintaining image quality (Gies et al 1999, Kalender et al 1999, Greess et al 2000, Hurwitz et al 2009, Angel et al 2009) .”
Background3D Forward and Back-Projection for X-Ray CT Using Separable Footprints
Long, Y., Fessler, J. A., & Balter, J. M. (2010) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2010.2050898 · Cited by 228 · cited 3×
“Long et al (2010) extended this to a trapezoidal shape, which enables more accurate modelling for projection lines obliquely intersecting the voxel grid.”
BackgroundMethodSelf-citationImage reconstruction from a small number of projections
Herman, G. T., & Davidi, R. (2008) · Inverse Problems · doi:10.1088/0266-5611/24/4/045011 · Cited by 228
Siewerdsen, J. H., Moseley, D. J., Bakhtiar, B., Richard, S., & Jaffray, D. A. (2004) · Medical Physics · doi:10.1118/1.1819789 · Cited by 226
Scatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results
Zhu, L., Bennett, N. R., & Fahrig, R. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.884636 · Cited by 220 · cited 1×
“…scatter and thus provides estimates of primary signal (Yang et al 2012, Sechopoulos 2012), or using an array of semi-transparent (instead of opaque) blockers that modulate the primary distribution so that Fourier techniques can be used to separate scatter and primary signals (Zhu et al 2006).”
MethodEffect of scattered radiation on image noise in cone beam CT
Endo, M., Tsunoo, T., Nakamori, N., & Yoshida, K. (2001) · Medical Physics · doi:10.1118/1.1357457 · Cited by 212 · cited 1×
“…include cupping and streak artefacts, decreased contrast and resolution and decreased image contrast-to-noise (Johns and Yaffe 1982, Joseph and Spital 1982, Glover 1982, Endo et al 2001, Siewerdsen and Jaffray 2001, Kyprianou et al 2005, Colijn et al 2004, Zbijewski and Beekman 2006b).”
BackgroundThe effects of scatter in x‐ray computed tomography
Joseph, P. M., & Spital, R. D. (1982) · Medical Physics · doi:10.1118/1.595111 · Cited by 211
Grids or air gaps for scatter reduction in digital radiography: A model calculation
Neitzel, U. (1992) · Medical Physics · doi:10.1118/1.596836 · Cited by 206 · cited 2×
“…(e.g. the increase in dose needed to compensate for primary attenuation in a grid while maintaining full spatial resolution (Siewerdsen et al 2004, Neitzel 1992, Kwan et al 2005)), design constraints and cost of devices (e.g. need of larger detectors and gantry when using gaps) and thus many…”
BackgroundWhiting, B. R., Massoumzadeh, P., Earl, O. A., O'Sullivan, J. A., Snyder, D. L., & Williamson, J. F. (2006) · Medical Physics · doi:10.1118/1.2230762 · Cited by 198
Variance analysis of x-ray CT sinograms in the presence of electronic noise background
Ma, J., Liang, Z., Fan, Y., Liu, Y., Huang, J., Chen, W., & Lu, H. (2012) · Medical Physics · doi:10.1118/1.4722751 · Cited by 177 · cited 1×
“…noise, and (ignoring A/D quantization) a reasonable measurement model would be the popular ‘Poisson+Gaussian’ model (Snyder et al 1993, 1995, Ma et al 2012): Y rawi ∼ K Poisson{Eμ[Mi]} + Nμεi , σ 2εi , (10) where μεi represents the dark current of the ith channel and σ 2 εi the electronic…”
MethodByrne, C. L. (2007) · A K Peters/CRC Press eBooks · doi:10.1201/9780429295492 · Cited by 174
Computer-Simulated Radiation Dose Reduction for Abdominal Multidetector CT of Pediatric Patients
Frush, D. P., Slack, C. C., Hollingsworth, C. L., Bisset, G. S., Donnelly, L. F., Hsieh, J., Lavin-Wensell, T., & Mayo, J. R. (2002) · American Journal of Roentgenology · doi:10.2214/ajr.179.5.1791107 · Cited by 174
Empirical beam hardening correction (EBHC) for CT
Kyriakou, Y., Meyer, E., Prell, D., & Kachelrieß, M. (2010) · Medical Physics · doi:10.1118/1.3477088 · Cited by 171 · cited 1×
“Based on these ideas, numerous correction methods for analytical reconstruction have been proposed (Joseph and Spital 1978, Herman 1979, Herman and Trivedi 1983, Joseph and Ruth 1997, Hsieh et al 2000, Yan et al 2000, Kyriakou et al 2010, Liu et al 2009).”
MethodCompton scatter effects in CT reconstructions
Glover, G. H. (1982) · Medical Physics · doi:10.1118/1.595197 · Cited by 167 · cited 3×
“…include cupping and streak artefacts, decreased contrast and resolution and decreased image contrast-to-noise (Johns and Yaffe 1982, Joseph and Spital 1982, Glover 1982, Endo et al 2001, Siewerdsen and Jaffray 2001, Kyprianou et al 2005, Colijn et al 2004, Zbijewski and Beekman 2006b).”
BackgroundMethodX‐ray scatter removal by deconvolution
Seibert, J. A., & Boone, J. M. (1988) · Medical Physics · doi:10.1118/1.596208 · Cited by 165
Compensation for readout noise in CCD images
Snyder, D. L., Helstrom, C. W., Lanterman, A. D., White, R. L., & Faisal, M. (1995) · Journal of the Optical Society of America A · doi:10.1364/josaa.12.000272 · Cited by 161
Rührnschopf and, E., & Klingenbeck, K. (2011) · Medical Physics · doi:10.1118/1.3589140 · Cited by 160
An efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT
Poludniowski, G. G., Evans, P., Hansen, V. N., & Webb, S. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/12/016 · Cited by 156 · cited 1×
“…the noise in the resulting scatter estimates (Colijn and Beekman 2004, Jarry et al 2006), or on (ii) forcing the photons from each interaction to a fixed, small number of nodes in the projection plane and interpolating between them to obtain the complete distribution (Poludniowski et al 2009).”
BackgroundCatSim: a new computer assisted tomography simulation environment
Man, B. D., Basu, S., Chandra, N. S., Dunham, B., Edic, P. M., Iatrou, M., Mcolash, S. M., Sainath, P., Shaughnessy, C., Tower, B., et al. (2007) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.710713 · Cited by 154 · cited 2×
“Ideally, each x-ray photon interaction in a detector element would produce the same number, say K, photoelectrons (De Man et al 2007, Iatrou et al 2007); these would then be recorded with additional electronic noise, and (ignoring A/D quantization) a reasonable measurement model would be the…”
BackgroundMethodEfficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT
Zbijewski, W. B., & Beekman, F. J. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.872328 · Cited by 146
Self-citationReconstruction algorithm for polychromatic CT imaging: application to beam hardening correction
Yan, C. H., Whalen, R. T., Beaupré, G. S., et al. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.832955 · Cited by 144
Scatter estimation for a digital radiographic system using convolution filtering
Love, L. A., & Kruger, R. A. (1987) · Medical Physics · doi:10.1118/1.596126 · Cited by 142 · cited 1×
“A potentially more accurate approach models the scatter as a PSF (kernel) applied to primary fluence and tries to estimate the scatter from projection data (scatter+primary) by deconvolution (Love and Kruger 1987, Seibert and Boone 1988).”
MethodElbakri, I. A., & Fessler, J. A. (2003) · Physics in Medicine and Biology · doi:10.1088/0031-9155/48/15/314 · Cited by 137 · cited 5×
“…than that assumed by the reconstructor, which usually involves a denser voxel grid, but often also denser detector sampling, depending on the chosen mechanism for forward projection (De Man et al 2000, De Man et al 2001, Nuyts et al 1998, Zbijewski and Beekman 2004a, Elbakri and Fessler 2003a).”
BackgroundMethodInfluentialSelf-citationKyriakou, Y., Riedel, T., & Kalender, W. A. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/18/008 · Cited by 136 · cited 1×
“Another recently proposed acceleration strategy combines an analytical model of first-order scatter with a coarse MC simulation of the typically homogeneous higher order scatter (Kyriakou et al 2006).”
MethodAccelerated Simulation of Cone Beam X-Ray Scatter Projections
Colijn, A. P., & Beekman, F. J. (2004) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2004.825600 · Cited by 131 · cited 1×
“…and rely on either (i) model-based fitting of smooth surfaces in the projection plane to reduce the noise in the resulting scatter estimates (Colijn and Beekman 2004, Jarry et al 2006), or on (ii) forcing the photons from each interaction to a fixed, small number of nodes in the projection…”
MethodSelf-citationScattered radiation in fan beam imaging systems
Johns, P. C., & Yaffe, M. J. (1982) · Medical Physics · doi:10.1118/1.595076 · Cited by 121
Shikhaliev, P. M. (2005) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/24/004 · Cited by 121
The exponential edge-gradient effect in X-ray computed tomography
Joseph, P. M., & Spital, R. D. (1981) · Physics in Medicine and Biology · doi:10.1088/0031-9155/26/3/010 · Cited by 113
Joseph, P. M., & Ruth, C. (1997) · Medical Physics · doi:10.1118/1.597970 · Cited by 111 · cited 1×
“Based on these ideas, numerous correction methods for analytical reconstruction have been proposed (Joseph and Spital 1978, Herman 1979, Herman and Trivedi 1983, Joseph and Ruth 1997, Hsieh et al 2000, Yan et al 2000, Kyriakou et al 2010, Liu et al 2009).”
MethodEvaluation of x‐ray scatter properties in a dedicated cone‐beam breast CT scanner
Kwan, A. L. C., Boone, J. M., & Shah, N. (2005) · Medical Physics · doi:10.1118/1.1954908 · Cited by 110
Image Reconstruction for Hybrid True-Color Micro-CT
Xu, Q., Yu, H., Bennett, J., He, P., Zainon, R., Doesburg, R., Opie, A. M. T., Walsh, M., Shen, H., Butler, A. P. H., et al. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2192119 · Cited by 108
Li, H., Mohan, R., & Zhu, X. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/23/006 · Cited by 103 · cited 2×
“…of the scatter kernel is essential for this deconvolution; the kernels are usually assumed to depend on object thickness (estimated locally based on water equivalent projections) and are either measured (Li et al 2008) or presimulated with MC methods (Maltz et al 2008, Sun and Star-Lack 2010).”
MethodTomographic Image Reconstruction Based on a Content-Adaptive Mesh Model
Brankov, J. G., Yang, Y., & Wernick, M. N. (2004) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2003.822822 · Cited by 103
Maximum-likelihood dual-energy tomographic image reconstruction
Fessler, J. A., Elbakri, I. A., Sukovic, P., et al. (2002) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.467189 · Cited by 99
Self-citationMaltz, J. S., Gangadharan, B., Bose, S., Hristov, D. H., Faddegon, B., Paidi, A., & Bani‐Hashemi, A. R. (2008) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2008.928922 · Cited by 93
Tomographic Reconstruction in the 21st Century
Clackdoyle, R., & Defrise, M. (2010) · IEEE Signal Processing Magazine · doi:10.1109/msp.2010.936743 · Cited by 92 · cited 1×
“In addition, as a side effect of these developments, new insight has been gained in the reconstruction from truncated projection (Clackdoyle and Defrise 2010).”
BackgroundZhou, J., & Qi, J. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/20/015 · Cited by 90 · cited 1×
“Zhou and Qi (2011) proposed to accurately measure the projection matrix and then model it as a combination of sinogram blurring, ideal projection and image blurring.”
BackgroundKarmazyn, B. W., Frush, D. P., Applegate, K. E., Maxfield, C. M., Cohen, M. D., & Jones, R. P. (2008) · American Journal of Roentgenology · doi:10.2214/ajr.08.1391 · Cited by 89 · cited 1×
“…a variety of simplified noise models, numerous methods have been proposed for adding ‘synthetic’ noise to x-ray CT sinograms to simulate lower dose scans (Mayo et al 1997, Frush et al 2002, Karmazyn et al 2009, Massoumzadeh et al 2009, Benson and De Man 2010, Wang et al 2012, Zabic et al 2012).”
MethodJin, J., Ren, L., Liu, Q., Kim, J., Wen, N. W., Guan, H., Movsas, B., & Chetty, I. J. (2010) · Medical Physics · doi:10.1118/1.3497272 · Cited by 89
Badano, A. G., & Sempau, J. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/6/013 · Cited by 87 · cited 1×
“Accurate modelling usually involves MC simulations (Badano and Sempau 2006).”
Kyprianou, I. S., Rudin, S., Bednarek, D. R., & Hoffmann, K. R. (2005) · Medical Physics · doi:10.1118/1.1844151 · Cited by 85 · cited 1×
“…include cupping and streak artefacts, decreased contrast and resolution and decreased image contrast-to-noise (Johns and Yaffe 1982, Joseph and Spital 1982, Glover 1982, Endo et al 2001, Siewerdsen and Jaffray 2001, Kyprianou et al 2005, Colijn et al 2004, Zbijewski and Beekman 2006b).”
BackgroundA Simple Regularizer for B-spline Nonrigid Image Registration That Encourages Local Invertibility
Chun, S. Y., & Fessler, J. A. (2009) · IEEE Journal of Selected Topics in Signal Processing · doi:10.1109/jstsp.2008.2011116 · Cited by 82
Self-citationOptimization-based image reconstruction from sparse-view data in offset-detector CBCT
Bian, J., Wang, J., Han, X., Sidky, E. Y., Shao, L., & Pan, X. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/2/205 · Cited by 81 · cited 3×
“Having such a perfect match is sometimes referred to as the ‘inverse crime’ (Herman and Davidi 2008, Kaipio and Somersalo 2007, Bian et al 2013).”
BackgroundMethodNiu, T., Al-Basheer, A. K., & Zhu, L. (2012) · Medical Physics · doi:10.1118/1.3693050 · Cited by 80 · cited 1×
“…(Ning et al 2004), by using blockers only in a first sequence of scans (prior scan, e.g. in the monitoring of radiation therapy treatment) (Niu et al 2012), by employing moving blockers (Liu et al 2005, Zhu et al 2005, Jin et al 2010, Wang et al 2010), or by exploiting data redundancy in…”
MethodVariance reduction techniques for fast Monte Carlo CBCT scatter correction calculations
Mainegra‐Hing, E., & Kawrakow, I. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/16/s05 · Cited by 76 · cited 1×
“…variance reduction techniques (e.g. forced detection, Woodcock tracking, interaction splitting), which, when optimized for simulation of x-ray scatter, can potentially result in 10–100× reduction of computation time necessary to reach a given level of noise (Mainegra-Hing and Kawrakow 2010).”
BackgroundStatistical reconstruction for x-ray computed tomography using energy-integrating detectors
Lasio, G., Whiting, B. R., & Williamson, J. F. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/8/014 · Cited by 76
Electronic Noise Modeling in Statistical Iterative Reconstruction
Xu, J., & Tsui, B. M. W. (2009) · IEEE Transactions on Image Processing · doi:10.1109/tip.2009.2017139 · Cited by 76
Angel, E., Yaghmai, N., Jude, C. M., DeMarco, J. J., Cagnon, C. H., Goldin, J. G., Primak, A. N., Stevens, D. M., CODY, D. D., McCollough, C. H., et al. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/3/003 · Cited by 76
Temporal resolution improvement in cardiac CT using PICCS (TRI‐PICCS): Performance studies
Tang, J., Hsieh, J., & Chen, G. (2010) · Medical Physics · doi:10.1118/1.3460318 · Cited by 75 · cited 1×
“In Tang et al (2010), IR is performed based on a 90◦ view interval.”
MethodX-ray scatter correction for cone-beam CT using moving blocker array
Zhu, L., Strobel, N. K., & Fahrig, R. (2005) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.594699 · Cited by 70 · cited 1×
“…of scans (prior scan, e.g. in the monitoring of radiation therapy treatment) (Niu et al 2012), by employing moving blockers (Liu et al 2005, Zhu et al 2005, Jin et al 2010, Wang et al 2010), or by exploiting data redundancy in the design of blocker pattern (Niu and Zhu 2011, Lee et al…”
MethodA Comparative Study of Two Postreconstruction Beam Hardening Correction Methods
Herman, G. T., & Trivedi, S. S. (1983) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.1983.4307626 · Cited by 70
Isola, A. A., Graß, M., & Niessen, W. J. (2010) · Medical Physics · doi:10.1118/1.3301600 · Cited by 70
A recursive filter for noise reduction in statistical iterative tomographic imaging
Thibault, J., Bouman, C. A., Sauer, K. D., et al. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.660281 · Cited by 67
Slambrouck, K. V., & Nuyts, J. L. (2012) · Medical Physics · doi:10.1118/1.4762567 · Cited by 67
Self-citationMaximum-likelihood transmission image reconstruction for overlapping transmission beams
Dong, Y., Fessler, J. A., & Ficaro, E. P. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.896785 · Cited by 66
Self-citationWang, J., Mao, W., & Solberg, T. D. (2010) · Medical Physics · doi:10.1118/1.3495819 · Cited by 64
Bootsma, G. J., Verhaegen, F., & Jaffray, D. A. (2015) · Medical Physics · doi:10.1118/1.4903260 · Cited by 62
Characterization and suppression of edge and aliasing artefacts in iterative x-ray CT reconstruction
Zbijewski, W. B., & Beekman, F. J. (2003) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/1/010 · Cited by 56
Self-citationLee, H., Xing, L., Lee, R., & Fahimian, B. (2012) · Medical Physics · doi:10.1118/1.3691901 · Cited by 54 · cited 1×
“…of scans (prior scan, e.g. in the monitoring of radiation therapy treatment) (Niu et al 2012), by employing moving blockers (Liu et al 2005, Zhu et al 2005, Jin et al 2010, Wang et al 2010), or by exploiting data redundancy in the design of blocker pattern (Niu and Zhu 2011, Lee et al 2012).”
MethodMeyer, M., Kalender, W. A., & Kyriakou, Y. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/1/007 · Cited by 52 · cited 1×
“…or estimated at relatively low cost using MC simulations (Bertram et al 2006), a hybrid method combining this approach with scatter kernels (Meyer et al 2010) and algorithms utilizing calibration scans to establish relationships between scatter properties of typical objects (e.g. spatial…”
MethodExperimental validation of a rapid Monte Carlo based micro-CT simulator
Colijn, A. P., Zbijewski, W. B., Sasov, A., & Beekman, F. J. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/18/009 · Cited by 43
Self-citationLiu, X., Shaw, C. C., Wang, T., Chen, L., Altunbas, M. C., & Kappadath, S. C. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.656655 · Cited by 41
Potential of software-based scatter corrections in cone-beam volume CT
Bertram, M., Wiegert, J., & Rose, G. (2005) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.595032 · Cited by 28 · cited 1×
“In their simplest form, such computational models represent the scatter fluence as a constant across the projection plane, assuming the same value for entire scan (Glover 1982) or different values for individual views (Bertram et al 2005).”
BackgroundSemerci, O., & Miller, E. L. (2012) · IEEE Transactions on Image Processing · doi:10.1109/tip.2012.2186308 · Cited by 27
Yao, W., & Leszczynski, K. W. (2009) · Medical Physics · doi:10.1118/1.3152115 · Cited by 26
Semerci, O., Miller, E. L., & Member, S. (2012) · Cited by 26
Evans, J. D., Whiting, B. R., Politte, D. G., O’Sullivan, J. A., Klahr, P., & Williamson, J. F. (2013) · Physica Medica · doi:10.1016/j.ejmp.2012.12.005 · Cited by 26
Zbijewski, W. B., & Beekman, F. J. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/7/017 · Cited by 25
Self-citationTemporal and spectral imaging with micro‐CT
Johnston, S. M., Johnson, G. A., & Badea, C. T. (2012) · Medical Physics · doi:10.1118/1.4736809 · Cited by 24
de Carvalho, B. M., & Herman, G. T. (2006) · Image and Vision Computing · doi:10.1016/j.imavis.2006.03.003 · Cited by 21 · cited 1×
“An alternative, but closely related expansion replaces the voxels with a set of spherically symmetric Kaiser–Bessel functions, known as ‘blobs’ (Lewitt 1990, Matej and Lewitt 1996, Ziegler et al 2006, Carvalho and Herman 2007).”
BackgroundJarry, G., Graham, S. A., Jaffray, D. A., Moseley, D. J., & Verhaegen, F. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.653803 · Cited by 20 · cited 1×
“…(i) model-based fitting of smooth surfaces in the projection plane to reduce the noise in the resulting scatter estimates (Colijn and Beekman 2004, Jarry et al 2006), or on (ii) forcing the photons from each interaction to a fixed, small number of nodes in the projection plane and interpolating…”
BackgroundThree-dimensional focus of attention for iterative cone-beam micro-CT reconstruction
Benson, T. M., & Gregor, J. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/18/006 · Cited by 18
Maximum-likelihood x-ray computed-tomography finite-beamwidth considerations
Browne, J. A., Boone, J. M., & Holmes, T. J. (1995) · Applied Optics · doi:10.1364/ao.34.005199 · Cited by 13 · cited 2×
“Browne et al (1995) did not use a footprint approach, but represented the detectors by subsampling those as well, using ray tracing between all detector points and source points.”
MethodMaximum likelihood transmission image reconstruction for overlapping transmission beams
Fessler, J. A., Dong, Y., & Ficaro, E. P. (2003) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.1999.845797 · Cited by 13
Self-citationLiu, X., Shaw, C., Altunbas, M. C., & Wang, T. (2005) · Medical Physics · doi:10.1118/1.1998386 · Cited by 11
An iterative dual energy CT reconstruction method for a K-edge contrast material
Depypere, M., Nuyts, J. L., van Gastel, N., Carmeliet, G., Maes, F., & Suetens, P. (2011) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.878162 · Cited by 11 · cited 1×
“By using constraints, one can estimate more materials than measured energy bins (Mendonca et al 2010, Long et al 2011, Depypere et al 2011, Long and Fessler 2012).”
MethodSelf-citationMotion and Positional Error Correction for Cone Beam 3D-Reconstruction with Mobile C-Arms
Bodensteiner, C., Darolti, C., Schumacher, H., Matthäus, L., & Schweikard, A. (2007) · Lecture notes in computer science · doi:10.1007/978-3-540-75757-3_22 · Cited by 11 · cited 1×
“This was demonstrated for rigid head motion in Bodensteiner et al (2007) and Nuyts et al (2011).”
MethodYang, K., Burkett, G. W., & Boone, J. M. (2012) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.911093 · Cited by 8
Dose and scatter characteristics of a novel cone beam CT system for musculoskeletal extremities
Zbijewski, W. B., Sisniega, A., Vaquero, J. J., Muhit, A. A., Packard, N. J., Senn, R., Yang, D., Yorkston, J. I., Carrino, J. A., & Siewerdsen, J. H. (2012) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.911843 · Cited by 7 · cited 1×
“…objects in the patient), significant heterogeneity may be introduced into the scatter distribution and thus more accurate scatter estimates may be needed to achieve complete artefact correction and maximize improvement of quantitative accuracy (MainegraHing and Kawrakow 2010, Zbijewski et al 2012).”
BackgroundSelf-citationBootsma, G. J., Verhaegen, F., & Jaffray, D. A. (2012) · Medical Physics · doi:10.1118/1.4735742 · Cited by 2
Fast iterative image reconstruction using sparse matrix factorization with GPU acceleration
Zhou, J., & Qi, J. (2011) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.878799 · Cited by 2 · cited 1×
“Zhou and Qi (2011) proposed to accurately measure the projection matrix and then model it as a combination of sinogram blurring, ideal projection and image blurring.”
Background
Cited by
X-ray phase-contrast imaging: from pre-clinical applications towards clinics
Bravin, A., Coan, P., & Suortti, P. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/1/r1 · Cited by 846
Deep learning for tomographic image reconstruction
Wang, G., Ye, J. C., & De Man, B. (2020) · Nature Machine Intelligence · doi:10.1038/s42256-020-00273-z · Cited by 531
Self-citationCompressive sensing in medical imaging
Graff, C. G., & Sidky, E. Y. (2015) · Applied Optics · doi:10.1364/ao.54.000c23 · Cited by 184
“x-ray CT has shifted focus from analytic methods to iterative image reconstruction [23], which aims to solve the imaging model implicitly.”
Background
Metadata
- Authors
- Nuyts, J., De Man, B., Fessler, J. A., Zbijewski, W., & Beekman, F. J.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2013-06-05
- Volume
- 58
- Issue
- 12
- Pages
- R63-R96
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2118916039
- Cited by
- 199
- References
- 129
- FWCI
- 17.35
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.