Peripheral nerve bionic interface: a review of electrodes
Open accessThe DOI matches its Crossref record; 68 of the 68 references it cites were found in a registry or an index; 9 of the 74 sources its claims cite were checked for support.
Abstract
As the demand for sensory feedback to and from prosthetic limbs becomes increasingly desirable, implantable neural interfaces are becoming more attractive. Here, we briefly review the current landscape of extra-neural electrodes for interfacing the peripheral nervous system exploring both clinical and exploratory sciences.
Overview
CoChat Summary
TL;DR
This is a narrative review of extra-neural peripheral nerve electrodes, focused mainly on cuff-type interfaces for prosthetic and other neuroelectronic applications. It argues that extra-neural electrodes are less traumatic and easier to use long term, but that the field still lacks a single confidently reliable, clinically approved method for peripheral nervous system interfacing.
Key findings
- The review identifies three main surgically implanted peripheral nerve interface categories: regenerative electrodes, intra-fascicular electrodes, and extra-neural electrodes; extra-neural electrodes are described as the least invasive and the preferred choice in this work.
- Commercial options are limited: the paper notes only a small set of commercially available peripheral nerve electrodes, mostly split-cylinder or spiral cuff designs using silicone or polyimide substrates and platinum-based contacts.
- None of the commercial products listed in Table 1 are FDA approved, which the authors say prevents clinical application and slows development of clinically suited prosthetic technologies.
- The paper states that in the human arm, sensory axons outnumber motor axons by a ratio of 9:1, supporting the importance of sensory-feedback-oriented interface design.
Methods
- Narrative review of the “current landscape of extra-neural electrodes” for interfacing the peripheral nervous system, covering both clinical and exploratory sciences.
- Uses published literature and commercial-product comparison; Table 1 summarizes commercial peripheral nerve electrodes by company, substrate, interface material, thickness, channel count, and approval status.
- Main analysis is qualitative: the paper compares electrode types, electrode designs, materials, wireless/tethering considerations, sterilization, and regulatory constraints.
Limitations
- The paper does not report a systematic search strategy, inclusion/exclusion criteria, or formal quality assessment; this is evident from the text.
- The discussion states that there is “yet to be a single method” that can be used confidently for PNS interfacing and that the lifetime/stability of neural implants cannot yet be guaranteed; this is a limitation of the field, not a study-specific sample limitation.
- The text says the commercial devices listed are not FDA approved, limiting clinical application and generalizability to real-world prosthetic use.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
9 Valid · 1 Not valid · 48 Not checked · 33 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“In severe peripheral nerve injuries, these pathways can be blocked or lost, resulting in loss of function (Seddon 1942).”
What the sources say
- Supportedcites Seddon 1942 matched to “A Classification of Nerve Injuries” (1942) source
The Neurotmesis section describes complete nerve division with complete sensory and motor paralysis. The classification also describes a transient block that produces paralysis, supporting the paraphrase that interrupted or lost nerve function can result in loss of function. Page numbers are not provided in the supplied passages. (LLM, conf=0.88)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“These implants are constructed from a library of materials that are readily absorbed into the body such as silk and polyurethanes for insulation and magnesium for electrical conductors.”
What the sources say
- Partly supportedcites Bettinger 2018 matched to “Recent advances in materials and flexible electronics for peripheral nerve interfaces” (2018) sourceExperimental — accuracy not yet measured
The section “Bioabsorbable substrates and components for stretchable and bioresorbable electrodes” (page unavailable) identifies polyurethanes as bioabsorbable elastomers for transient implants and magnesium as a dissolvable metal for electronic components. It does not identify silk or polyurethanes specifically as insulation, or explicitly assign magnesium the role of electrical conductor; silk appears only in a cited reference title. (LLM, conf=0.86)
- Not matchedcites Koo etal. 2018
An author–year citation the paper’s reference list doesn’t settle. We don’t search these by name.
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“They are designed to dissolve post-implantation after a period has elapsed removing the need for surgery to remove the implant.”
What the sources say
- Supportedcites Bettinger 2018 matched to “Recent advances in materials and flexible electronics for peripheral nerve interfaces” (2018) source
The “Bioabsorbable substrates and components for stretchable and bioresorbable electrodes” section says temporary implants can be resorbed within the body and that bioabsorbable peripheral nerve interfaces do not have to be explanted. It also describes functional lifetimes of weeks or months after implantation. No page number is provided in our copy. (LLM, conf=0.93)
- Not matchedcites Koo etal. 2018
An author–year citation the paper’s reference list doesn’t settle. We don’t search these by name.
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“However, for some of these coatings, in particular conducting polymers, there is an additional technological challenge of delamination of these films from the underlying conductors (Guo 2016).”
What the sources say
- Supportedcites Guo 2016 matched to “The Pursuit of Chronically Reliable Neural Interfaces: A Materials Perspective” (2016) source
The Physical integration section (page not provided) calls delamination a major drawback of conducting-polymer electrode coatings. That directly supports the cited claim about coating films separating from underlying conductors. (LLM, conf=0.94)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“These materials will also be assessed based on their chemical and mechanical similarity to the target tissue to prevent an undesirable immune response to these alien materials (Guo 2016).”
What the sources say
- Supportedcites Guo 2016 matched to “The Pursuit of Chronically Reliable Neural Interfaces: A Materials Perspective” (2016) source
In “Physical integration > Engineering principles” (page not supplied), Guo says an ideal neural implant should mimic host tissue physically and biochemically to minimize immune response. The biomimicry section describes mechanical and biochemical strategies for doing so. This supports the citing sentence’s rationale for assessing material similarity. (LLM, conf=0.91)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
39 more claims
Create a free account to see all 91 claims and every source verdict.
- Probably not a substantial claim. 26
- It may state two separate claims as one. 27
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Although they require the nerve to be freed from the surrounding tissues, extraneural electrodes are generally simple to implant [15] and have shown selective stimulation in multiple long-term clinical trials for motor and sensory restoration [16-18].”
From Directed stimulation with interfascicular interfaces for peripheral nerve stimulation (2021) · cites this paper as [15]
Citations
Cites
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · cited 1×
“Neural electrodes for electrical interfacing with the body have been used for a range of treatments of chronic conditions requiring electrical recording, stimulation to excite or inhibit the target organ (Cogan 2008).”
BackgroundBetter speech recognition with cochlear implants
Wilson, B. S., Finley, C. C., Lawson, D. T., Wolford, R. D., Eddington, D. K., & Rabinowitz, W. M. (1991) · Nature · doi:10.1038/352236a0 · Cited by 1,190
Targeted neurotechnology restores walking in humans with spinal cord injury
Wagner, F. B., Mignardot, J.-B., Le Goff-Mignardot, C. G., Demesmaeker, R., Komi, S., Capogrosso, M., Rowald, A., Seáñez, I., Caban, M., Pirondini, E., et al. (2018) · Nature · doi:10.1038/s41586-018-0649-2 · Cited by 1,087
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
Brain-Controlled Interfaces: Movement Restoration with Neural Prosthetics
Schwartz, A. B., Cui, X. T., Weber, D. J., & Moran, D. W. (2006) · Neuron · doi:10.1016/j.neuron.2006.09.019 · Cited by 792
A neural interface provides long-term stable natural touch perception
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008669 · Cited by 751
Recovery of Over-Ground Walking after Chronic Motor Complete Spinal Cord Injury
Angeli, C. A., Boakye, M., Morton, R. A., Vogt, J., Benton, K., Chen, Y., Ferreira, C. K., & Harkema, S. J. (2018) · New England Journal of Medicine · doi:10.1056/nejmoa1803588 · Cited by 670
Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066014 · Cited by 600
Gill, M. L., Grahn, P. J., Calvert, J. S., Linde, M. B., Lavrov, I. A., Strommen, J. A., Beck, L. A., Sayenko, D. G., Van Straaten, M. G., Drubach, D. I., et al. (2018) · Nature Medicine · doi:10.1038/s41591-018-0175-7 · Cited by 574
A Classification of Nerve Injuries
Seddon, H. J. (1942) · BMJ · doi:10.1136/bmj.2.4260.237 · Cited by 563 · cited 1×
“In severe peripheral nerve injuries, these pathways can be blocked or lost, resulting in loss of function (Seddon 1942).”
BackgroundVagus-nerve stimulation for the treatment of epilepsy
Ben-Menachem, E. (2002) · The Lancet Neurology · doi:10.1016/s1474-4422(02)00220-x · Cited by 560 · cited 1×
“1982), or treatment of epilepsy by vagal nerve stimulation (Ben-Menachem 2002).”
BackgroundWireless Power Transfer Strategies for Implantable Bioelectronics
Agarwal, K., Jegadeesan, R., Guo, Y.-X., & Thakor, N. V. (2017) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2017.2683520 · Cited by 540
Koo, J., MacEwan, M. R., Kang, S.-K., Won, S. M., Stephen, M., Gamble, P., Xie, Z., Yan, Y., Chen, Y.-Y., Shin, J., et al. (2018) · Nature Medicine · doi:10.1038/s41591-018-0196-2 · Cited by 496
A transverse intrafascicular multichannel electrode (TIME) to interface with the peripheral nerve
Boretius, T., Badia, J., Pascual-Font, A., Schuettler, M., Navarro, X., Yoshida, K., & Stieglitz, T. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.05.010 · Cited by 468
Myoelectric Control of Artificial Limbs—Is There a Need to Change Focus? [In the Spotlight]
Jiang, N., Dosen, S., Muller, K.-R., & Farina, D. (2012) · IEEE Signal Processing Magazine · doi:10.1109/msp.2012.2203480 · Cited by 434
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415
Eldabe, S., Buchser, E., & Duarte, R. V. (2016) · Pain Medicine · doi:10.1093/pm/pnv025 · Cited by 403
Functionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398
Double nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396
Development of Implantable Medical Devices: From an Engineering Perspective
Joung, Y.-H. (2013) · International Neurourology Journal · doi:10.5213/inj.2013.17.3.98 · Cited by 367 · cited 2×
“…such as dry heat, electron beam- and gammairradiation, ethylene oxide, steam (autoclave), hydrogen peroxide, formaldehyde sterilization and ozone, as well as new methods such as vaporized peracetic acid, high intensity or pulsed light and microwave radiation (Joung 2013; Tipnis and Burgess 2018).”
BackgroundMethodDhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354
Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces
Stieglitz, T., Beutel, H., Schuettler, M., & Meyer, J.-U. (2000) · Biomedical Microdevices · doi:10.1023/a:1009955222114 · Cited by 343
The first 500 patients with sacral anterior root stimulator implants: general description
Brindley, G. S. (1994) · Spinal Cord · doi:10.1038/sc.1994.126 · Cited by 318 · cited 1×
“2002), bladder control in paraplegia (Brindley 1994; Brindley et al.”
BackgroundSacral anterior root stimulators for bladder control in paraplegia
Brindley, G. S., Polkey, C. E., & Rushton, D. N. (1982) · Spinal Cord · doi:10.1038/sc.1982.65 · Cited by 311
A review of portable FES-based neural orthoses for the correction of drop foot
Lyons, G. M., Sinkjaer, T., Burridge, J. H., & Wilcox, D. J. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806832 · Cited by 277
Selective Electrical Interfaces with the Nervous System
Rutten, W. L. C. (2002) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.4.020702.153427 · Cited by 276 · cited 1×
“As the understanding of peripheral nerves has increased and concurrent technologies have advanced, neural electrodes are beginning to enable the development of clinically applicable neuroelectronic interfaces (Rutten 2002).”
BackgroundProsthetic Myoelectric Control Strategies: A Clinical Perspective
Roche, A. D., Rehbaum, H., Farina, D., & Aszmann, O. C. (2014) · Current Surgery Reports · doi:10.1007/s40137-013-0044-8 · Cited by 261
Self-citationSchiefer, M., Tan, D., Sidek, S. M., & Tyler, D. J. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/1/016001 · Cited by 255
Criteria for the Selection of Materials for Implanted Electrodes
Geddes, L. A., & Roeder, R. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1581292 · Cited by 240
Sterilization of implantable polymer-based medical devices: A review
Tipnis, N. P., & Burgess, D. J. (2018) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2017.12.003 · Cited by 233
Graczyk, E. L., Resnik, L., Schiefer, M. A., Schmitt, M. S., & Tyler, D. J. (2018) · Scientific Reports · doi:10.1038/s41598-018-26952-x · Cited by 216
Principles Underlying New Methods for Chronic Neural Recording
Stein, R. B., Charles, D., Davis, L., Jhamandas, J., Mannard, A., & Nichols, T. R. (1975) · Canadian Journal of Neurological Sciences / Journal Canadien des Sciences Neurologiques · doi:10.1017/s0317167100020333 · Cited by 210
Implantable microscale neural interfaces
Cheung, K. C. (2007) · Biomedical Microdevices · doi:10.1007/s10544-006-9045-z · Cited by 202 · cited 1×
“An additional method of reducing the foreign body response is to incorporate eluting chemicals such as anti-inflammatory agents which are released in the surrounding tissue post-implantation (Cheung 2007).”
MethodSchultz, A. E., & Kuiken, T. A. (2011) · PM&R · doi:10.1016/j.pmrj.2010.06.016 · Cited by 186
Lago, N., Ceballos, D., J Rodrı́guez, F., Stieglitz, T., & Navarro, X. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.06.025 · Cited by 178
Axonal components of nerves innervating the human arm
Gesslbauer, B., Hruby, L. A., Roche, A. D., Farina, D., Blumer, R., & Aszmann, O. C. (2017) · Annals of Neurology · doi:10.1002/ana.25018 · Cited by 170
Self-citationCui, X., & Martin, D. C. (2003) · Sensors and Actuators A: Physical · doi:10.1016/s0924-4247(02)00427-2 · Cited by 160
A Review of Implant Communication Technology in WBAN: Progress and Challenges
Teshome, A. K., Kibret, B., & Lai, D. T. H. (2018) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2018.2848228 · Cited by 147
Energy harvesting for human wearable and implantable bio-sensors
Mitcheson, P. D. (2010) · 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology · doi:10.1109/iembs.2010.5627952 · Cited by 132
Charkhkar, H., Shell, C. E., Marasco, P. D., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aac964 · Cited by 131
Fins, J. J., Mayberg, H. S., Nuttin, B., Kubu, C. S., Galert, T., Sturm, V., Stoppenbrink, K., Merkel, R., & Schlaepfer, T. E. (2011) · Health Affairs · doi:10.1377/hlthaff.2010.0157 · Cited by 117
Xiang, Z., Yen, S., Sheshadri, S., Wang, J., Lee, S., Liu, Y., Liao, L., Thakor, N. V., & Lee, C. (2016) · Advanced Materials · doi:10.1002/adma.201503423 · Cited by 116
Context-Dependent Upper Limb Prosthesis Control for Natural and Robust Use
Amsuess, S., Vujaklija, I., Goebel, P., Roche, A. D., Graimann, B., Aszmann, O. C., & Farina, D. (2016) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2015.2454240 · Cited by 109
Self-citationA slowly penetrating interfascicular nerve electrode for selective activation of peripheral nerves
Tyler, D. J., & Durand, D. M. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559349 · Cited by 108
McDonnall, D., Clark, G. A., & Normann, R. A. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.828425 · Cited by 98
Toward Bioelectronic Medicine—Neuromodulation of Small Peripheral Nerves Using Flexible Neural Clip
Lee, S., Peh, W. Y. X., Wang, J., Yang, F., Ho, J. S., Thakor, N. V., Yen, S., & Lee, C. (2017) · Advanced Science · doi:10.1002/advs.201700149 · Cited by 96
The Evolution of Neuroprosthetic Interfaces
Adewole, D. O., Serruya, M. D., Harris, J. P., Burrell, J. C., Petrov, D., Chen, H.-C. I., Wolf, J. A., & Cullen, D. K. (2016) · Critical Reviews in Biomedical Engineering · doi:10.1615/critrevbiomedeng.2016017198 · Cited by 91
Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88
A flexible method for fabrication of nerve cuff electrodes
Haugland, M. (2002) · Proceedings of 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.656992 · Cited by 79
Recent advances in materials and flexible electronics for peripheral nerve interfaces
Bettinger, C. J. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0007-6 · Cited by 77
Selective stimulation and neural recording on peripheral nerves using flexible split ring electrodes
Lee, S., Sheshadri, S., Xiang, Z., Delgado-Martinez, I., Xue, N., Sun, T., Thakor, N. V., Yen, S.-C., & Lee, C. (2017) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2016.09.127 · Cited by 73
A Parylene Self-Locking Cuff Electrode for Peripheral Nerve Stimulation and Recording
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2014.2333733 · Cited by 56
A micro-scale printable nanoclip for electrical stimulation and recording in small nerves
Lissandrello, C. A., Gillis, W. F., Shen, J., Pearre, B. W., Vitale, F., Pasquali, M., Holinski, B. J., Chew, D. J., White, A. E., & Gardner, T. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa5a5b · Cited by 55
The design of and chronic tissue response to a composite nerve electrode with patterned stiffness
Freeberg, M. J., Stone, M. A., Triolo, R. J., & Tyler, D. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6632 · Cited by 54
Nanopatterning effects on astrocyte reactivity
Ereifej, E. S., Matthew, H. W., Newaz, G., Mukhopadhyay, A., Auner, G., Salakhutdinov, I., & VandeVord, P. J. (2013) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.34480 · Cited by 45
Electroceuticals in medicine – The brave new future
Mishra, S. (2017) · Indian Heart Journal · doi:10.1016/j.ihj.2017.10.001 · Cited by 37 · cited 1×
“Further, this electroceutical treatment is a welcomed alternative to pharmaceutical options for the same conditions to avoid unwanted side effects as well as added specificity of the delivery (Mishra 2017).”
BackgroundTopography and Coating of Platinum Improve the Electrochemical Properties and Neuronal Guidance
Schlie-Wolter, S., Deiwick, A., Fadeeva, E., Paasche, G., Lenarz, T., & Chichkov, B. N. (2013) · ACS Applied Materials & Interfaces · doi:10.1021/am3028487 · Cited by 34
Hoffmann, K. ‐., Koch, K. P., Doerge, T., & Micera, S. (2006) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2006.1639123 · Cited by 33
Cuff electrodes for very small diameter nerves — Prototyping and first recordings in vivo
Ordonez, J. S., Pikov, V., Wiggins, H., Patten, C., Stieglitz, T., Rickert, J., & Schuettler, M. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6945201 · Cited by 30
Bacterial contamination of test stimulation leads during percutaneous nerve stimulation
Pannek, J., Grigoleit, U., & Hinkel, A. (2005) · Urology · doi:10.1016/j.urology.2005.01.004 · Cited by 24
Flexible, polyimide-based neural interfaces
Stieglitz, T., Beutel, H., Keller, R., Schüettler, M., & Meyer, J.-U. (1999) · International Conference on Microelectronics for Neural, Fuzzy and Bio-Inspired Systems (MicroNeuro) · doi:10.1109/mn.1999.758853 · Cited by 24
Ereifej, E. S., Cheng, M. M.-C., Mao, G., & VandeVord, P. J. (2013) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2013.04.023 · Cited by 19
The Pursuit of Chronically Reliable Neural Interfaces: A Materials Perspective
Guo, L. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00599 · Cited by 19 · cited 2×
“These materials will also be assessed based on their chemical and mechanical similarity to the target tissue to prevent an undesirable immune response to these alien materials (Guo 2016).”
BackgroundA parylene cuff electrode for peripheral nerve recording and drug delivery
Cobo, A., Boyajian, B., Larson, C. E., Kee, S., Pikov, V., & Meng, E. F. (2017) · IEEE International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2017.7863454 · Cited by 12
Russell, C., Kissane, R. W. P., Steenson, D. P., & Chakrabarty, S. (2018) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2018.10.036 · Cited by 6
Self-citationGill, M. L., Grahn, P. J., Calvert, J. S., et al. (2018) · Nature Medicine · doi:10.1038/s41591-018-0248-7
Classification of Nerve Injuries
Seddon, H. J. (1942) · BMJ · doi:10.1136/bmj.2.4270.560-b
Research works printed but not linked: 1
- Yoshida, K., Pellinen, D., Pivin, D., Rousche, P., Kipke, D.: Development of the thin-film longitudinal intra-fascicular electrode. In: Center for Sensory-Motor Interaction (SMI), Department of Health Science and Technology, Aalborg University, pp. 279–281 (2000)
Cited by
Bioelectronics for electrical stimulation: materials, devices and biomedical applications
Huang, Y., Yao, K., Zhang, Q., Huang, X., Chen, Z., Zhou, Y., & Yu, X. (2024) · Chemical Society Reviews · doi:10.1039/d4cs00413b · Cited by 183
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113
Advances in Soft Bioelectronics for Brain Research and Clinical Neuroengineering
Sunwoo, S., Han, S. I., Joo, H., Cha, G. D., Kim, D., Choi, S. H., Hyeon, T., & Kim, D. (2020) · Matter · doi:10.1016/j.matt.2020.10.020 · Cited by 114
Nizamis, K., Athanasiou, A., Almpani, S., Dimitrousis, C., & Astaras, A. (2021) · Sensors · doi:10.3390/s21062084 · Cited by 147
Metadata
- Authors
- Russell, C., Roche, A. D., & Chakrabarty, S.
- Affiliations
- University of Leeds (GB)
- Sheffield Teaching Hospitals NHS Foundation Trust (GB)
- Corresponding author
- Chakrabarty, Samit
- Venue
- International Journal of Intelligent Robotics and Applications
- Published
- 2019-02-22
- Volume
- 3
- Issue
- 1
- Pages
- 11-18
- ISSN
- 2366-5971 (ISSN-L)
- 2366-598X
- OpenAlex
- W2916220960
- Cited by
- 67
- References
- 65
- Citations by year
- 2026: 5 · 2025: 9 · 2024: 14 · 2023: 6 · 2022: 8 · 2021: 15 · 2020: 7 · 2019: 3
- FWCI
- 3.95
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- Locations
- International Journal of Intelligent Robotics and Applications (publishedVersion, cc-by)
- White Rose Research Online (University of Leeds, The University of Sheffield, University of York) (acceptedVersion)
- Funders
- Engineering and Physical Sciences Research Council (CYCLOPS)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- peripheral nerve interfaces, neural electrodes, implantable neural interfaces, artificial limbs, sensory feedback, peripheral nervous system
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.