Clinical magnetocardiography: the unshielded bet—past, present, and future
Open accessThe DOI matches its Crossref record; 314 of the 339 references it cites were found in a registry or an index; 38 of the 131 sources its claims cite were checked for support.
Abstract
Magnetocardiography (MCG), which is nowadays 60 years old, has not yet been fully accepted as a clinical tool. Nevertheless, a large body of research and several clinical trials have demonstrated its reliability in providing additional diagnostic electrophysiological information if compared with conventional non-invasive electrocardiographic methods. Since the beginning, one major objective difficulty has been the need to clean the weak cardiac magnetic signals from the much higher environmental noise, especially that of urban and hospital environments. The obvious solution to record the magnetocardiogram in highly performant magnetically shielded rooms has provided the ideal setup for decades of research demonstrating the diagnostic potential of this technology. However, only a few clinical institutions have had the resources to install and run routinely such highly expensive and technically demanding systems. Therefore, increasing attempts have been made to develop cheaper alternatives to improve the magnetic signal-to-noise ratio allowing MCG in unshielded hospital environments. In this article, the most relevant milestones in the MCG's journey are reviewed, addressing the possible reasons beyond the currently long-lasting difficulty to reach a clinical breakthrough and leveraging the authors' personal experience since the early 1980s attempting to finally bring MCG to the patient's bedside for many years thus far. Their nearly four decades of foundational experimental and clinical research between shielded and unshielded solutions are summarized and referenced, following the original vision that MCG had to be intended as an unrivaled method for contactless assessment of the cardiac electrophysiology and as an advanced method for non-invasive electroanatomical imaging, through multimodal integration with other non-fluoroscopic imaging techniques. Whereas all the above accounts for the past, with the available innovative sensors and more affordable active shielding technologies, the present demonstrates that several novel systems have been developed and tested in multicenter clinical trials adopting both shielded and unshielded MCG built-in hospital environments. The future of MCG will mostly be dependent on the results from the ongoing progress in novel sensor technology, which is relatively soon foreseen to provide multiple alternatives for the construction of more compact, affordable, portable, and even wearable devices for unshielded MCG inside hospital environments and perhaps also for ambulatory patients.
Overview
CoChat Summary
TL;DR
This review traces 60 years of clinical magnetocardiography (MCG), contrasting shielded and unshielded systems and drawing on the authors’ four decades of experience. It argues that unshielded MCG is feasible and clinically informative, and that advances in sensors, active shielding, and denoising may finally bring it to the patient’s bedside.
Key findings
- Unshielded MCG has been demonstrated since 1980; the first 36-channel unshielded system became operational in January 2002, with raw peak-to-peak background noise of 7–30 pT, 1–2 pT after adaptive 50 Hz filtering, sensitivity of 20–40 fT/√Hz above 1 Hz, source-localization reproducibility ICC >0.7, and quasi-real-time mapping in 90 s with analysis in less than 2 min.
- Localization accuracy: single-channel unshielded MCG gave average 3D uncertainty of about 12 mm in patients and about 5 mm in phantom; shielded Neuromag MCG with amagnetic catheters gave 2 ± 0.7 mm SD in a realistic torso phantom and 4 ± 2.7 mm SD in patients, about twice as good as simultaneous body surface potential mapping.
- Clinical targets discussed include emergency triage of chest pain, ischemic and non-ischemic cardiomyopathies, heart transplant rejection, non-invasive 3D electroanatomical imaging of arrhythmogenic substrates, fetal MCG, and animal studies; newer systems include Avalon-H90, VitalScan/Corsens, and CardioFlux, while a multicenter ACS study concluded VitalScan did not yet meet accuracy to confidently rule out ACS in the ED in 2020.
- Emerging sensor/denoising results include scalar OPM gradiometers with <90 fT/√Hz equivalent to 18 fT/cm/√Hz at 5 cm baseline, another with 16 fT/cm/Hz1/2 at 3 cm baseline, and EEMD denoising improving SNR by approximately 18 dB with r = 0.9 between shielded and unshielded MCG after processing.
Methods
- Narrative review, not a systematic review or meta-analysis, synthesizing referenced experimental and clinical MCG literature with the authors’ personal experience since the early 1980s in shielded and unshielded settings.
- Sources include clinical studies, multicenter trials, conference proceedings, device-development reports, and validation work using systems such as CMI 3619a, Neuromag, amagnetic catheters, phantoms, and patient recordings.
- No formal primary data collection or quantitative synthesis is reported; the analysis is a qualitative historical and technical comparison of shielded versus unshielded MCG and of sensor, shielding, and denoising technologies.
Limitations
- Stated in the paper: lack of standardization of MCG acquisition, postprocessing, and analysis makes studies only partially comparable and less sharable; dedicated software for MCG 3D electroanatomical imaging is still lacking relative to ECGi; some new sensors still require electromagnetic shielding or have lower sensitivity; scalar OPMs are costly (about $15K per gradiometer unit, more than one order of magnitude higher than the present average cost of a SQUID); wearable MCG may suffer movement artifacts; and OPMs developed for MEG need evaluation for deeper cardiac source localization.
- Evident from the text: the review is authored by long-standing proponents of unshielded MCG and relies heavily on their own center’s experience, with no described systematic search or selection criteria, so selection and enthusiasm bias are possible.
- Evident from the text: many cited future or ongoing developments, including ongoing trials and not-yet-commercial sensors, are not yet clinically validated for widespread MCG use.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures
![Figure 1 (A) CNR's physicist Gian Luca Romani (left) and Riccardo Fenici (right) testing the prototype of the CNR unshielded MCG SQUID gradiometer at the Catholic University's Gemelli Hospital in Rome. (B) Example of real-time simultaneous MCG and ECG recordings and (C) example of 36-position real-time single-beat MCG recordings (Finnish grid) [modified from (98)].](https://figures.cochat.ai/paper-figures/a3abc1e771e54d61972be491c7091e59/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-1.jpg)

![Figure 3 (A) High-resolution MCG map of the “ramp-like” pattern during the PR interval with superimposed isomagnetic lines [modified from (120) data]. (B) Validation of the atrial repolarization nature of the PR “ramp-like” MCG pattern with simultaneous MCG and atrial monophasic action potential recording. The red asterisk indicates the His bundle MF (±0.2pT) disclosed during the last half of the PR interval after subtraction of the stronger (±1.8 pT) atrial repolarization MF (open red arrow). The pseudo-current reconstruction (multiple white arrows on the MF maps) indicates the opposite direction of atrial and His bundle depolarization current and of atrial repolarization current [standardized MF color code: in blue (−), MF outgoing from the chest; in red (+), MF entering in the chest].](https://figures.cochat.ai/paper-figures/b9947f432be9464fa9dec7893eb08049/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-3.jpg)


![Figure 6 Typical experimental setup used for the MCG study of small animals with the CMI 3649a system. (A) Non-invasive contactless MCG recording and source imaging. (B) Procedure for simultaneous minimally invasive electrophysiological study with a single AC for multiple monophasic action potential recording from the epicardial surface, localizable with the MCG mapping and EMD inverse solution [red-circled solid arrow in (C)].](https://figures.cochat.ai/paper-figures/9f5ff125893644d38b40045bd5019e53/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-6.jpg)

![Figure 8 First reported comparison between sequential 36-point MCG recording with a single-channel OPM developed at the University of Freiburg (Swiss) and 36-channel MCG mapping at the Catholic University Hospital in Rome. All data were converted to the same format and analyzed with the Neuromag MCG software. The identity of MCG waveforms, of the MF distribution, of the related source localization (green arrows onto the MF maps), and of the 3D current density imaging (CDI), calculated at the apex of the QRS and of the T waves, is immediately evident [modified from data in (275)].](https://figures.cochat.ai/paper-figures/f344048c25cb4f58b8bae070a9c93376/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-8.jpg)

Verification
Claims the paper makes
18 Valid · 19 Not valid · 34 Not checked · 24 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“More recently, another study carried out with a new 37-channel (19 active sensors and 18 for noise reduction) prototype by Leeds University suggests that, with appropriate modeling, 5 of 38 magnetic QRS parameters could be used to provide a MCG estimate of left ventricular ejection fraction (264).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A truly portable compact multichannel MCG device
What the sources say
- Partly supportedcites [264] Beadle, 2021, Assessing heart disease using a novel magnetocardiography device, Biomed Phys Eng Express sourceExperimental — accuracy not yet measured
The Introduction and Methods describe a 37-channel prototype with 19 active sensors and 18 noise-reduction channels. The Results and discussion report extracting 38 MCG parameters and fitting an LVEF model, but the curve-fit limitations state that the model used three predictors, not five of the 38 parameters. The paper’s references to five parameters concern differences between the post-MI and healthy-control groups, not the number used to estimate LVEF. (LLM, conf=0.89)
hedgedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A first pilot clinical study (protocol NCT02359773, on ClinicalTrials.gov) had shown that the device provided high sensitivity (95.4%) and a negative predictive value (NPV) (97.7%) for the rule-out of healthy subjects and of patients whose chest pain was non-ischemic from those with ischemic heart disease (262).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A truly portable compact multichannel MCG device
What the sources say
- Partly supportedcites [262] Ghasemi-Roudsari, 2018, A portable prototype magnetometer to differentiate ischemic and non-ischemic heart disease in patients with chest pain, PLoS One sourceExperimental — accuracy not yet measured
Results > Model performance (Table 3) reports 95.4% sensitivity and 97.7% NPV for the cross-validated model comparing IHD patients with controls and young healthy volunteers; the NPV assumes 15% disease prevalence. But Results > Combined data set shows that these figures come from pooled technical-performance and pilot-clinical-study data, not the pilot clinical study alone. (LLM, conf=0.88)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“concluded that, at least in 2020, the VitalScan did not yet meet the level of accuracy required to confidently rule out ACS in the ED clinical practice (263).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A truly portable compact multichannel MCG device
What the sources say
- Supportedcites [263] Goodacre, 2021, Diagnostic accuracy of the magnetocardiograph for patients with suspected acute coronary syndrome, Emerg Med J source
The Methods (p. 6) describe a prospective, multicentre cohort study of patients presenting to the ED with suspected ACS. The Discussion (p. 13) concludes that the VitalScan MCG could not yet meet the accuracy required to rule out ACS for clinical use; the Conclusion (p. 3) likewise says it was not ready for clinical practice. (LLM, conf=0.95)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“although theory and experimental research suggest and confirm that it can provide additional diagnostic information not achievable with electrocardiography alone (1, 2, 4–6, 9, 16, 107).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Discussion
What the sources say
- Not checked — no full text of the sourcecites [1] Plonsey, 1972, Comparative capabilities of electrocardiography and magnetocardiography, Am J Cardiol source
- Not matchedcites [2] Malmivuo
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Not checked — no full text of the sourcecites [4] Wikswo, 1982, Possible sources of new information in the magnetocardiogram, J Theor Biol source
- Not checked — no full text of the sourcecites [5] Roth, 1986, Electrically silent magnetic fields, Biophys J source
- Not checked — no full text of the sourcecites [6] Lant, 1990, Complementary nature of electrocardiographic and magnetocardiographic data in patients with ischemic heart disease, J Electrocardiol source
- Not checked — no full text of the sourcecites [9] Baule, 1963, Detection of the magnetic field of the heart, Am Heart J source
- Not checked — no full text of the sourcecites [16] Cohen, 1975, Magnetic determination of the relationship between the S-T segment shift and the injury current produced by coronary artery occlusion, Circ Res source
- Supportedcites [107] Savard, 1983, Magnetic measurement of S-T and T-Q segment shifts in humans. Part I: early repolarization and left bundle branch block, Circ Res source
The paper’s introductory passage says the ECG cannot measure T-Q shifts or distinguish two causes of S-T shifts, whereas dcMCG can. The Discussion qualifies its practical use: extraneous signals prevented its use as a clinical diagnostic tool at the time, though it could clarify S-T shifts in selected subjects as a research tool. (LLM, conf=0.88)
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Every source the paper cites for this claim was checked and supports it.
In the paper
“Another non-cryogenic alternative to SQUID-based MCG systems is the use of OPMs (269).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A still shielded but more compact and non-cryogenic (OPM-based) multichannel MCG system for emergency departments
What the sources say
- Supportedcites [269] Lin, 2013, A compact, high performance atomic magnetometer for biomedical applications, Phys Med Biol source
The Introduction describes the atomic magnetometer as a viable alternative to SQUID magnetometers used in MCG systems. The Discussion states that atomic magnetometers do not require cryogenic cooling, and the Results report comparable MCG recordings from the atomic and SQUID magnetometers. (LLM, conf=0.94)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
36 more claims · 4 with their cited sources checked
Create a free account to see all 95 claims and to verify the 8 references still open.
- It may state two separate claims as one. 40
- Probably not a substantial claim. 17
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the past > From single-channel MCG mapping to multichannel MCG imaging in unshielded hospital settings
“the same unshielded CMI 3619a installation currently used for clinical MCG, mapping the cardiac MF of small animals (195) has been also performed”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the past > From single-channel MCG mapping to multichannel MCG imaging in unshielded hospital settings
“Figure: Figure 6 Typical experimental setup used for the MCG study of small animals with the CMI 3649a system.”
Citations
Cites
Items 201–315 of 315
Fenici, R., Brisinda, D., & Meloni, A. M. (2004) · PubMed · Cited by 13
Self-citationLehto, M., Jurkko, R., Parikka, H. J., Mäntynen, V., Väänänen, H., Montonen, J., Voipio‐Pulkki, L., Toivonen, L. K., & Laine, M. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.02205.x · Cited by 13
Fenici, R., & Brisinda, D. (2006) · International journal of cardiac imaging · doi:10.1007/s10554-006-9076-9 · Cited by 13
Self-citationPESOLA, K., Lojtjonen, J., Nenonen, J., Magnin, I. E., Lauerma, K., Fenici, R., & Katila, T. E. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.867958 · Cited by 13
Self-citationAnalysis of Fetal Heart Rate Variability Obtained by Magnetocardiography
Verklan, M. T., Padhye, N. S., & Brazdeikis, A. (2006) · The Journal of Perinatal & Neonatal Nursing · doi:10.1097/00005237-200610000-00014 · Cited by 13
Rat Magnetocardiography Using a Flux-Gate Sensor Based on Iron Garnet Films
Vetoshko, P. M., Gusev, N. A., Chepurnova, D. A., Самойлова, Е. В., Zvezdin, A. K., Korotaeva, A. A., & Belotelov, V. I. (2016) · Biomedical Engineering · doi:10.1007/s10527-016-9628-9 · Cited by 13
Diagnostic accuracy of the magnetocardiograph for patients with suspected acute coronary syndrome
Goodacre, S., Walters, S. J., Qayyum, H., Coffey, F., Carlton, E., Coats, T., Glazebrook, W., & Unitt, L. (2021) · Emergency Medicine Journal · doi:10.1136/emermed-2020-210396 · Cited by 13
Udovychenko, Y. Y., Popov, A. O., & Chaikovsky, I. A. (2021) · Frontiers in Medical Technology · doi:10.3389/fmedt.2021.779800 · Cited by 13 · cited 2×
“Interesting data were also obtained in patients with non-ischemic arrhythmogenic cardiomyopathy (96, 138, 218, 219), confirmed also by more recent multichannel MCG studies in MSRs (220, 221).”
BackgroundResultStroink, G. (1989) · Advances in Biomagnetism · doi:10.1007/978-1-4613-0581-1_4 · Cited by 12
Lim, H. K., Chung, N., Kim, K., Ko, Y., Kwon, H., Lee, Y., Kim, J., Cho, J. R., Kim, J., Kim, I.-S., et al. (2006) · Annals of Biomedical Engineering · doi:10.1007/s10439-006-9210-9 · Cited by 12
Three-dimensional spectral map of atrial fibrillation by a 64-channel magnetocardiogram
Nakai, K., Oka, T., Okabayashi, H., Tsuboi, J., Fukuhiro, Y., Fukushima, A., Suwabe, A., Itoh, M., & Yoshizawa, M. (2008) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2007.06.008 · Cited by 12
Noninvasive Determination of HV Interval Using Magnetocardiography
Senthilnathan, S., Selvaraj, R. J., Patel, R., Satheesh, S., Katholil, G., Janawadkar, M. P., & Radhakrishnan, T. S. (2017) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13067 · Cited by 12
Lachlan, T., He, H., Sharma, K., Khan, J. N., Rajappan, K., Morley‐Davies, A., Patwala, A., Randeva, H. S., & Osman, F. (2020) · BMJ Open · doi:10.1136/bmjopen-2020-038804 · Cited by 12
Fenici, R., & Melillo, G. (1991) · International journal of cardiac imaging · doi:10.1007/bf01797753 · Cited by 12 · cited 3×
“This pioneering vision of MCG mapping as a unique novel method for non-invasive 3D electroanatomical imaging (EAI) and localization of cardiac electrophysiological mechanisms with its potential to guide “aimed” myocardial biopsy and interventional transcatheter treatment of arrhythmogenic substrates prompted the devel…”
BackgroundMethodInfluentialSelf-citationQuantifying Cognitive Workload Using a Non-Contact Magnetocardiography (MCG) Wearable Sensor
Wang, Z., Zhu, K., Kaur, A., Recker, R. S., Yang, J. G., & Kiourti, A. (2022) · Sensors · doi:10.3390/s22239115 · Cited by 12
Fenici, R., Melillo, G., Cappelli, A., Luca, C. D., & Masselli, M. (1989) · Springer eBooks · doi:10.1007/978-1-4613-0581-1_94 · Cited by 11
Self-citationMagnetocardiographic Localization of a Pacing Catheter
Fenici, R., Melillo, G., Cappelli, A., Luca, C. D., & Masselli, M. (1989) · Springer eBooks · doi:10.1007/978-1-4613-0581-1_76 · Cited by 11
Self-citationHigh-resolution magnetic measurements of human cardiac electrophysiological events
Fenici, R., Romani, G. L., & Erne, S. N. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455927 · Cited by 11 · cited 1×
“(A) High-resolution MCG map of the “ramp-like” pattern during the PR interval with superimposed isomagnetic lines [modified from (120) data].”
MethodSelf-citationComani, S., Mantini, D., Alleva, G., Luzio, S. D., & Romani, G. L. (2005) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/23/006 · Cited by 11
The P wave in the magnetocardiogram
Takeuchi, A., Watanabe, K., Nomura, M., Ishihara, S., Sumi, M., Murakami, M., Saito, K., Nakaya, Y., & Mori, H. (1988) · Journal of Electrocardiology · doi:10.1016/s0022-0736(88)80012-8 · Cited by 11
Göudde, P., Agrawal, R., Müuller, H. ‐., Czerski, K., Endt, P., Steinhoff, U., Oeff, M., Schultheiss, H. ‐., & Behrens, S. (2001) · Clinical Cardiology · doi:10.1002/clc.4960241009 · Cited by 11
BRISINDA, D., & FENICI, R. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00627.x · Cited by 11
Self-citationKomamura, K., Adachi, Y., Miyamoto, M., et al. (2008) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.2003979 · Cited by 11
Desai, L., Wakai, R. T., Tsao, S. S. L., Strasburger, J. F., Gotteiner, N. L., & Patel, A. (2020) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13900 · Cited by 11
(2004) · Cited by 10
Self-citationA Novel Filter Scheme of Data Processing for SQUID-Based Magnetocardiogram
Dang-Ting, L., Ye, T., Ren, Y.-F., et al. (2008) · Chinese Physics Letters · doi:10.1088/0256-307x/25/7/105 · Cited by 10
Malmivuo, J. A., Wikswo, J. P., Barry, W. H., Harrison, D. C., & Fairbank, W. M. (1977) · Medical and Biological Engineering and Computing · doi:10.1007/bf02457995 · Cited by 10
From 3D to 4D imaging: is that useful for interventional cardiac electrophysiology?
Fenici, R., & Brisinda, D. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4353714 · Cited by 10
Self-citationLombardi, G., Sorbo, A. R., Guida, G., La Brocca, L., Fenici, R., & Brisinda, D. (2018) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2018.02.004 · Cited by 10
Magnetocardiography detects left atrial dysfunction in paroxysmal atrial fibrillation
Her, A.-Y., Shin, E.-S., Zhou, Q., Wierzbinski, J., Vidal-Lopez, S., Saleh, A., Kim, Y. H., Garg, S., Jung, F., & Brachmann, J. (2019) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-180528 · Cited by 10
Huang, X., Hua, N., Tang, F., & Zhang, S. (2020) · Cardiovascular Diagnosis and Therapy · doi:10.21037/cdt-20-121 · Cited by 10
Zhang, J., Liu, J., Zhang, Q., Филиппов, Д. А., Li, K., Wu, J., Tao, J., Jiang, L., Cao, L., & Srinivasan, G. (2021) · Review of Scientific Instruments · doi:10.1063/5.0035059 · Cited by 10
Biomagnetic Imaging in the Cardiac Catheterization Laboratory
Fenici, R., & Melillo, G. (1989) · Springer eBooks · doi:10.1007/978-1-4613-0581-1_87 · Cited by 9
Self-citationClinical multichannel MCG in unshielded hospital environment.
Brisinda, D., Meloni, A. M., & Fenici, R. (2004) · PubMed · Cited by 9
Self-citationBrisinda, D., Caristo, M. E., & Fenici, R. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/8/002 · Cited by 9
Self-citationMagnetocardiographically‐Guided Catheter Ablation
FENICI, R. R., COVINO, M., CELLERINO, C., DI LILLO, M., DE FILIPPO, M. C., & MELILLO, G. (1995) · Journal of Interventional Cardiology · doi:10.1111/j.1540-8183.1995.tb00936.x · Cited by 9
Self-citationChang, Y., Wu, C., Lin, C., Wu, Y., Yang, Y., Chang, T., Jiang, Y., & Chuang, L. (2015) · PLoS ONE · doi:10.1371/journal.pone.0133192 · Cited by 9
Ultra-sensitive vector magnetometer for magnetocardiographic mapping
Gusev, N. A., Vetoshko, P. M., Kuzmichev, A. N., Chepurnova, D. A., Samoilova, E. V., Zvezdin, A. K., Korotaeva, A. A., & Belotelov, V. I. (2017) · Biomedical Engineering · doi:10.1007/s10527-017-9705-8 · Cited by 9
Guida, G., Sorbo, A. R., Fenici, R., & Brisinda, D. (2018) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12569 · Cited by 9
Self-citationMäkijärvi, M. A., Korhonen, P., Jurkko, R., Väänänen, H., SILTANEN, P., & Hänninen, H. (2010) · Springer eBooks · doi:10.1007/978-1-84882-046-3_44 · Cited by 9
Medical applications of diamond magnetometry: commercial viability
Dale, M. W., & Morley, G. W. (2017) · arXiv · doi:10.48550/arxiv.1705.01994 · Cited by 9
Farrell, D. E., Tripp, J., & Norgren, R. (1979) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.956997 · Cited by 9
Senthilnathan, S., Parasakthi, C., Mariyappa, N., Patel, R., Katholil, G., Janawadkar, M. P., Radhakrishnan, T. S., & Thoddi, M. R. (2012) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2012.00523.x · Cited by 8
Current Status and Future of Cardiac Mapping in Atrial Fibrillation
Ds, K., & Ah, H. (2012) · InTech eBooks · doi:10.5772/25905 · Cited by 8
Magnetic measurements and modelling for the investigation of the human-heart conduction system
Fenici, R. R., Romani, G. L., & Leoni, R. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455931 · Cited by 8 · cited 1×
“The first attempts for MCG localization of the human His bundle (34), of the accessory pathways in the Wolff–Parkinson–White syndrome (35), and of supraventricular (137) and ventricular arrhythmias (138, 139) were initially validated by off-line comparison with the fluoroscopic position of intracardiac catheters recor…”
ResultSelf-citationFenici, R., Brisinda, D., Nenonen, J., Morana, G., & Fenici, P. (2001) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmte.2001.46.s2.219 · Cited by 8
Self-citationBrisinda, D., Venuti, A., Sorbo, A. R., & Fenici, R. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2013.07.215 · Cited by 8
Self-citationIndependent component analysis of normal and abnormal rhythm in twin pregnancies
Mensah-Brown, N. A., Lutter, W. J., Comani, S., Strasburger, J. F., & Wakai, R. T. (2010) · Physiological Measurement · doi:10.1088/0967-3334/32/1/004 · Cited by 8
Optical Magnetometer Employing Adaptive Noise Cancellation for Unshielded Magnetocardiography
Tiporlini, V., & Alameh, K. (2013) · Universal Journal of Biomedical Engineering · doi:10.13189/ujbe.2013.010104 · Cited by 8
Fenici, R., & Brisinda, D. (2007) · PubMed · Cited by 8
Self-citation[Magnetocardiography: a comparison with electrocardiography].
Nakaya, Y. (1984) · PubMed · Cited by 8
Current density imaging of focal cardiac sources
Nenonen, J., Pesola, K., Feneici, R., Lauerma, K., Mäkijärvi, M., & Katila, T. (2001) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmte.2001.46.s2.50 · Cited by 7
Ma, J., Yang, Z., & Uchiyama, T. (2019) · AIP Advances · doi:10.1063/1.5129842 · Cited by 7
Fenici, R., Mashkar, R., & Brisinda, D. (2020) · European Heart Journal · doi:10.1093/ehjci/ehaa946.0386 · Cited by 7
Self-citationFenici, R., Masselli, M., & Sabetta, F. (1986) · Cited by 6
Self-citationFenici, R., & Brisinda, D. (2007) · Lecture Notes in Computer Science · doi:10.1007/978-3-540-72907-5_24 · Cited by 6
Self-citationMagnetocardiography capabilities in myocardium injuries diagnosis
Sosnytskyy, V., Chaikovsky, I. A., Stadnyuk, L., Miasnykov, G., Kazmirchyk, A., Sosnytska, T., & Gurjeva, O. (2013) · World Journal of Cardiovascular Diseases · doi:10.4236/wjcd.2013.35059 · Cited by 6
Brisinda, D., Bottelli, G., Napolitano, C., Priori, S. G., & Fenici, R. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.079 · Cited by 6
Self-citationBrisinda, D., Sorbo, A. R., Venuti, A., & Fenici, R. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/3/521 · Cited by 6
Self-citationScalable, tunable Josephson junctions and DC SQUIDs based on CVD graphene
Li, T., Gallop, J. C., Hao, L., & Romans, E. J. (2019) · IEEE Transactions on Applied Superconductivity · doi:10.1109/tasc.2019.2897999 · Cited by 6
High Resolution Cardiomagnetism
Farrell, D. E., Tripp, J. H., & VanDoren, C. L. (1981) · De Gruyter eBooks · doi:10.1515/9783110863529-019 · Cited by 6
Brisinda, D., Fenici, R., & Smars, P. (2022) · Contemporary Cardiology · doi:10.1007/978-3-031-05520-1_17 · Cited by 6
Self-citationPetrenko, M., & Vershovskii, A. (2022) · Sensors · doi:10.3390/s22249862 · Cited by 6
Mäntynen, V., Vitikainen, A.-M., Koskinen, R., Mäkijärvi, M. A., Toivonen, L. K., & Montonen, J. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.081 · Cited by 5
Biomagnetic study of cardiac arrhythmias
Fenici, R., & Melillo, G. (1991) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/12/a/001 · Cited by 5
Self-citationBrazdeikis, A., Chu, C. W., Cherukuri, P., et al. (2004) · PubMed · Cited by 5
Monitoring fetal development with magnetocardiography
Padhye, N. S., Brazdeikis, A., & Verklan, M. T. (2005) · doi:10.1109/iembs.2004.1404014 · Cited by 5
Lombardi, G., Sorbo, A. R., Fenici, R., & Brisinda, D. (2017) · Annals of Heart · doi:10.36959/652/388 · Cited by 5
Self-citationSmall Animal Biomagnetism Applications
Jensen, K., Bentzen, B. H., & Polzik, E. S. (2022) · Springer eBooks · doi:10.1007/978-3-031-05363-4_3 · Cited by 5
Magnetocardiographic recording of the his-purkinje system activity in man
Fenici, R., Romani, G. L., & Leoni, R. (1982) · Japanese Heart Journal · Cited by 4
Self-citationBrisinda, D., Meloni, A. M., Fenici, P., & Fenici, R. (2004) · Cited by 4
Self-citationChaikovsky, I. A., Primin, M., Владимирович, Н. И., & Будник, М. М. (2012) · InTech eBooks · doi:10.5772/30122 · Cited by 4
(2004) · PubMed · Cited by 4
Self-citationKleemann, T., Kouraki, K., Strauss, M., Skarlos, A., Zeymer, U., & Zahn, R. (2012) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-012-9754-6 · Cited by 4
Time course of changes in cardiac magnetic field mapping after myocardial infarction in rats
Fischer, R., Gapelyuk, A., Wessel, N., et al. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.018 · Cited by 4
The T wave abnormality in the magnetocardiogram.
Nakaya, Y., Nomura, M., Fujino, K., et al. (1989) · PubMed · Cited by 4
Theory Of The Pr Sequent Of The Human Magnetocardiogram
Tripp, J. H., & Farrell, D. E. (1981) · Biomagnetism · doi:10.1515/9783110863529-018 · Cited by 4
Lachlan, T., He, H., Miller, A., Chandan, N., Siddiqui, S., Beadle, R. M., Wilson, D. G., Petkar, S., Randeva, H. S., & Osman, F. (2022) · Heart Rhythm · doi:10.1016/j.hrthm.2022.12.020 · Cited by 4
First simultaneous magnetocardiographic and invasive Kent bundle localization in man
Fenici, R., Masselli, M., López, L., et al. (1985) · Cited by 3
Self-citationMagnetocardiography and cardiac risk
Mäkijärvi, M. (1997) · Herzschrittmachertherapie und Elektrophysiologie · doi:10.1007/bf03042400 · Cited by 3
Komamura, K., Kawai, J., Miyamoto, M., et al. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.062 · Cited by 3
Clinical application of MCG in ischemic heart disease
Hailer, B., & Leeuwen, P. V. (2007) · International Congress Series · doi:10.1016/j.ics.2007.01.045 · Cited by 3
Kashyap, R., Smars, P. A., Gartzen, C., Bellolio, M. F., & Reeder, G. (2008) · Annals of Emergency Medicine · doi:10.1016/j.annemergmed.2008.01.133 · Cited by 3
The magnetic field produced by the conduction system of the human heart
Patrick, J. L., Hess, D. W., Tripp, J. H., et al. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455929 · Cited by 3
Atrial activity during thePR segment of the MCG
Leifer, M. C., Capos, N., Griffin, J. C., & Wikswo, J. P. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455930 · Cited by 3
Brisinda, D., Caristo, M. E., & Fenici, R. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.047 · Cited by 3 · cited 1×
“Moreover, MCG has proven useful for the non-invasive study of experimental intact animals (84) and contactless high-resolution investigation of experimental electrophysiology of isolated heart (85) and cardiac tissue models (86).”
MethodSelf-citation[Magnetocardiography: principles and potential uses in cardiology].
Carré, F., Painvin, I., Poiseau, É., Mabo, P., Lessard, Y., Daubert, C., & Toulouse, P. (2002) · PubMed · Cited by 3
[Certain aspects of the clinical application of magnetocardiography].
Mn, T., Safonov, I. D., & Vm, P. (1967) · PubMed · Cited by 3
Brisinda, D., Caristo, M. E., & Fenici, R. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.060 · Cited by 3
Self-citationCurrent density imaging on the epicardial surface of the heart
PESOLA, K., & Nenonen, J. (2001) · Cited by 3
Vector Fetal Magnetocardiography
Rassi, D., Zhuravlev, Y. E., Lewis, M. J., & Emery, S. J. (2000) · Springer eBooks · doi:10.1007/978-1-4612-1260-7_131 · Cited by 2
Brisinda, D., Meloni, A. M., Nenonen, J., & Fenici, R. (2004) · Cited by 2
Self-citationMagnetocardiographic study of patients with right and left bundle branch blocks
Brisinda, D., & Fenici, R. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.061 · Cited by 2
Self-citationDenis, B., Machecourt, J., Favier, C., & Martin-Noël, P. (1978) · PubMed · Cited by 2
Clinical application of magnetic mapping.
Sosnytska, T. (2011) · Likarska sprava · Cited by 2
Chaikovsky, I. A., Primin, M., Владимирович, Н. И., Verba, A., Mjasnikov, G., Kazmirchyk, A., Ji, W., Fan, X., Tan, X., Lei, M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.07.309 · Cited by 2
The Beginning of Biomagnetism and Magnetoencephalography Research in Finland in the 1970s
Katila, T. E. (2020) · Oxford University Press eBooks · doi:10.1093/oso/9780190935689.003.0002 · Cited by 2
Romani, G. L., & Williamson, S. J. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455916 · Cited by 2
Wang (2022) · doi:10.1109/jsen.2022.3222553 · Cited by 2
Fenici, R., Brisinda, D., Mäkijärvi, M. A., Toivonen, L., PESOLA, K., Nenonen, J. T., & Fenici, P. (2001) · Cited by 2
Self-citationClinical assessment of the magnetocardiogram
Fenici, R. (1983) · Cited by 2
Self-citationKotani, M., Mori, H., Kuriki, S., Uchikawa, Y., Chiyotani, K., & Nemoto, I. (1987) · Springer eBooks · doi:10.1007/978-94-009-3361-3_20 · Cited by 1
Brisinda, D., Giordano, V., Calvani, M., & Fenici, R. (2007) · International Congress Series · doi:10.1016/j.ics.2006.12.039 · Cited by 1
Self-citationA portable biomagnetic measuring system
Vvedensky, V. L., Naurzakov, S. P., Ozhogin, V. I., et al. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455926 · Cited by 1
Study of P-R Segment Using Magnetocardiographic Measurement
Uchikawa, Y., Kotani, M., & Erne, S. N. (1987) · IEEE Translation Journal on Magnetics in Japan · doi:10.1109/tjmj.1987.4549630 · Cited by 1
Brisinda, D., Meloni, A. M., Nenonen, J., et al. (2003) · Cited by 1
Self-citationNakai, K., Kazui, T., Okabayashi, H., Hayashi, R., Fukushima, A., & Suwabe, A. (2008) · PubMed · Cited by 1
Fenici (1991) · doi:10.1007/bf01797746 · Cited by 1
Fenici, R., Sorbo, A. R., & Brisinda, D. (2018) · European Heart Journal · doi:10.1093/eurheartj/ehy565.p2269 · Cited by 1
Self-citationHigh resolution Electrocardiography and Magnetocardiography
Fenici, R., Masselli, M., Lopez, L., & Melillo, G. (1992) · Cited by 1
Self-citationUntitled work
Makijarvi (1993) · doi:10.1093/eurheartj/14.suppl-7b_-7de.46
Fiber-coupled Diamond Magnetometry with an Unshielded 30 pT/$\sqrt{\textrm{Hz}}$ Sensitivity
Graham, S. M., Rahman, A. T. M. A., Munn, L., Patel, R. L., Newman, A. J., Stephen, C. J., Colston, G., Nikitin, A., Edmonds, A. M., Twitchen, D. J., et al. (2022) · arXiv · doi:10.48550/arxiv.2211.09170
Williamson, S. J., Romani, G.-L., Kaufman, L., & Modena, I. (1983) · Biomagnetism · doi:10.1007/978-1-4757-1785-3_19
(1985)
Research works printed but not linked: 12
- Cohen, 1970, Magnetocardiograms taken inside a shielded room with a superconducting point-contact magnetometer, Appl Phys Lett, 16, 80, 10.1063/1.1653195 · DOI · Paper page
- Stroink, 1989, Principles of cardiomagnetism, 47, 10.1007/978-1-4613-0581-1_4 · DOI · Paper page
- Stroink, 1998, Magnetism in medicine: a handbook, chapter cardiomagnetism
- Siltanen, 1989, Magnetocardiography, 1405
- Erne, 1985, Modelling of the His-Purkinje heart conduction system, 126
- Fenici, 1985, First simultaneous magnetocardiographic and invasive recordings of the PR interval electrophysiological phenomena in man, Med Biol Eng Comput, 23, 1483
- Dössel, 2000, Inverse problem of electro- and magnetocardiography: review and recent progress, Int J Bioelectromagn, 2, 262
- Fenici, 1989, Biomagnetic imaging in the cardiac catheterization laboratory, 409, 10.1007/978-1-4613-0581-1_87 · DOI
- Pesola, 1997, Comparison of regularization methods when applied to epicardial minimum norm estimates, Biomed Tech, 42, 273
- Fenici, 2001, First MCG multichannel instrumentation operative in an unshielded hospital laboratory for multimodal cardiac electrophysiology. Preliminary experience, Biomed Tech, 2019
- Park, 2005, Magnetocardiography predicts coronary artery disease in patients with acute chest pain, Ann Noninvasive Electrocardiol, 10, 312, 10.1111/j.1542-474X.2005.00634.x · DOI · Paper page
- Petrenko, 2022, Towards a practical implementation of a single-beam all-optical non-zero-field magnetic sensor for magnetoencephalographic complexes, Sensors, 22, 9862, 10.3390/s22249862 · DOI · Paper page
Cited by
Spintronic devices for biomedical applications
Mostufa, S., Liang, S. A., Chugh, V. K., Wang, J., & Wu, K. (2024) · npj Spintronics · doi:10.1038/s44306-024-00031-6 · Cited by 68
“It necessitates operation within a magnetically shielded room and at low temperatures 95 .”
Design and Analysis of a High-Sensitivity Atomic Magnetic Gradiometer With Adjustable Baseline
Zhang, Y., Tang, J., Xu, X., Wang, Z., & Zhai, Y. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3544375 · Cited by 27
“Conversely, in magnetocardiography (MCG), the magnetic field of the heart is typically modeled as a current dipole with p values ranging from 1.8 to 2.2 [34], [35].”
InfluentialAtrial Cardiomyopathy in Atrial Fibrillation: A Multimodal Diagnostic Framework
Karakasis, P., Vlachakis, P. K., Theofilis, P., Ktenopoulos, N., Patoulias, D., Fyntanidou, B., Antoniadis, A. P., & Fragakis, N. (2025) · Diagnostics · doi:10.3390/diagnostics15101207 · Cited by 33
“In addition to catheter-based electroanatomic mapping, emerging non-invasive current-based techniques such as magnetocardiography (MCG) offer novel avenues for atrial substrate assessment [77].”
Roth, B. J. (2024) · Biophysics Reviews · doi:10.1063/5.0201950 · Cited by 24
Harnessing the Heart’s Magnetic Field for Advanced Diagnostic Techniques
Elfouly, T., & Alouani, A. (2024) · Sensors · doi:10.3390/s24186017 · Cited by 19
Metadata
- Authors
- Brisinda, D., Fenici, P., & Fenici, R.
- Affiliations
- Università Cattolica del Sacro Cuore (IT)
- Corresponding author
- Fenici, R.
- Venue
- Frontiers in Cardiovascular Medicine
- Publisher
- Frontiers Media SA
- Published
- 2023-08-10
- Volume
- 10
- Pages
- 1232882-1232882
- ISSN
- 2297-055X (ISSN-L)
- PMC
- PMC10448194
- OpenAlex
- W4385724211
- Semantic Scholar
- 7dc43883741b2cd8ebeae3d2cd7cac34a798ade9
- Cited by
- 80
- References
- 339
- Citations by year
- 2026: 30 · 2025: 30 · 2024: 18 · 2023: 2
- FWCI
- 3.91
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Frontiers in Cardiovascular Medicine (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- PubliCatt (Università Cattolica del Sacro Cuore) (submittedVersion, cc-by)
- Funders
- Università degli Studi di Parma
- Università Cattolica del Sacro Cuore
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Keywords
- magnetocardiography, magnetic shielding, signal-to-noise ratio, cardiac electrophysiology, wearable sensors, clinical adoption
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.