Clinical magnetocardiography: the unshielded bet—past, present, and future
Open accessThe DOI matches its Crossref record; 314 of the 339 references it cites were found in a registry or an index; 38 of the 131 sources its claims cite were checked for support.
Abstract
Magnetocardiography (MCG), which is nowadays 60 years old, has not yet been fully accepted as a clinical tool. Nevertheless, a large body of research and several clinical trials have demonstrated its reliability in providing additional diagnostic electrophysiological information if compared with conventional non-invasive electrocardiographic methods. Since the beginning, one major objective difficulty has been the need to clean the weak cardiac magnetic signals from the much higher environmental noise, especially that of urban and hospital environments. The obvious solution to record the magnetocardiogram in highly performant magnetically shielded rooms has provided the ideal setup for decades of research demonstrating the diagnostic potential of this technology. However, only a few clinical institutions have had the resources to install and run routinely such highly expensive and technically demanding systems. Therefore, increasing attempts have been made to develop cheaper alternatives to improve the magnetic signal-to-noise ratio allowing MCG in unshielded hospital environments. In this article, the most relevant milestones in the MCG's journey are reviewed, addressing the possible reasons beyond the currently long-lasting difficulty to reach a clinical breakthrough and leveraging the authors' personal experience since the early 1980s attempting to finally bring MCG to the patient's bedside for many years thus far. Their nearly four decades of foundational experimental and clinical research between shielded and unshielded solutions are summarized and referenced, following the original vision that MCG had to be intended as an unrivaled method for contactless assessment of the cardiac electrophysiology and as an advanced method for non-invasive electroanatomical imaging, through multimodal integration with other non-fluoroscopic imaging techniques. Whereas all the above accounts for the past, with the available innovative sensors and more affordable active shielding technologies, the present demonstrates that several novel systems have been developed and tested in multicenter clinical trials adopting both shielded and unshielded MCG built-in hospital environments. The future of MCG will mostly be dependent on the results from the ongoing progress in novel sensor technology, which is relatively soon foreseen to provide multiple alternatives for the construction of more compact, affordable, portable, and even wearable devices for unshielded MCG inside hospital environments and perhaps also for ambulatory patients.
Overview
CoChat Summary
TL;DR
This review traces 60 years of clinical magnetocardiography (MCG), contrasting shielded and unshielded systems and drawing on the authors’ four decades of experience. It argues that unshielded MCG is feasible and clinically informative, and that advances in sensors, active shielding, and denoising may finally bring it to the patient’s bedside.
Key findings
- Unshielded MCG has been demonstrated since 1980; the first 36-channel unshielded system became operational in January 2002, with raw peak-to-peak background noise of 7–30 pT, 1–2 pT after adaptive 50 Hz filtering, sensitivity of 20–40 fT/√Hz above 1 Hz, source-localization reproducibility ICC >0.7, and quasi-real-time mapping in 90 s with analysis in less than 2 min.
- Localization accuracy: single-channel unshielded MCG gave average 3D uncertainty of about 12 mm in patients and about 5 mm in phantom; shielded Neuromag MCG with amagnetic catheters gave 2 ± 0.7 mm SD in a realistic torso phantom and 4 ± 2.7 mm SD in patients, about twice as good as simultaneous body surface potential mapping.
- Clinical targets discussed include emergency triage of chest pain, ischemic and non-ischemic cardiomyopathies, heart transplant rejection, non-invasive 3D electroanatomical imaging of arrhythmogenic substrates, fetal MCG, and animal studies; newer systems include Avalon-H90, VitalScan/Corsens, and CardioFlux, while a multicenter ACS study concluded VitalScan did not yet meet accuracy to confidently rule out ACS in the ED in 2020.
- Emerging sensor/denoising results include scalar OPM gradiometers with <90 fT/√Hz equivalent to 18 fT/cm/√Hz at 5 cm baseline, another with 16 fT/cm/Hz1/2 at 3 cm baseline, and EEMD denoising improving SNR by approximately 18 dB with r = 0.9 between shielded and unshielded MCG after processing.
Methods
- Narrative review, not a systematic review or meta-analysis, synthesizing referenced experimental and clinical MCG literature with the authors’ personal experience since the early 1980s in shielded and unshielded settings.
- Sources include clinical studies, multicenter trials, conference proceedings, device-development reports, and validation work using systems such as CMI 3619a, Neuromag, amagnetic catheters, phantoms, and patient recordings.
- No formal primary data collection or quantitative synthesis is reported; the analysis is a qualitative historical and technical comparison of shielded versus unshielded MCG and of sensor, shielding, and denoising technologies.
Limitations
- Stated in the paper: lack of standardization of MCG acquisition, postprocessing, and analysis makes studies only partially comparable and less sharable; dedicated software for MCG 3D electroanatomical imaging is still lacking relative to ECGi; some new sensors still require electromagnetic shielding or have lower sensitivity; scalar OPMs are costly (about $15K per gradiometer unit, more than one order of magnitude higher than the present average cost of a SQUID); wearable MCG may suffer movement artifacts; and OPMs developed for MEG need evaluation for deeper cardiac source localization.
- Evident from the text: the review is authored by long-standing proponents of unshielded MCG and relies heavily on their own center’s experience, with no described systematic search or selection criteria, so selection and enthusiasm bias are possible.
- Evident from the text: many cited future or ongoing developments, including ongoing trials and not-yet-commercial sensors, are not yet clinically validated for widespread MCG use.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures
![Figure 1 (A) CNR's physicist Gian Luca Romani (left) and Riccardo Fenici (right) testing the prototype of the CNR unshielded MCG SQUID gradiometer at the Catholic University's Gemelli Hospital in Rome. (B) Example of real-time simultaneous MCG and ECG recordings and (C) example of 36-position real-time single-beat MCG recordings (Finnish grid) [modified from (98)].](https://figures.cochat.ai/paper-figures/a3abc1e771e54d61972be491c7091e59/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-1.jpg)

![Figure 3 (A) High-resolution MCG map of the “ramp-like” pattern during the PR interval with superimposed isomagnetic lines [modified from (120) data]. (B) Validation of the atrial repolarization nature of the PR “ramp-like” MCG pattern with simultaneous MCG and atrial monophasic action potential recording. The red asterisk indicates the His bundle MF (±0.2pT) disclosed during the last half of the PR interval after subtraction of the stronger (±1.8 pT) atrial repolarization MF (open red arrow). The pseudo-current reconstruction (multiple white arrows on the MF maps) indicates the opposite direction of atrial and His bundle depolarization current and of atrial repolarization current [standardized MF color code: in blue (−), MF outgoing from the chest; in red (+), MF entering in the chest].](https://figures.cochat.ai/paper-figures/b9947f432be9464fa9dec7893eb08049/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-3.jpg)


![Figure 6 Typical experimental setup used for the MCG study of small animals with the CMI 3649a system. (A) Non-invasive contactless MCG recording and source imaging. (B) Procedure for simultaneous minimally invasive electrophysiological study with a single AC for multiple monophasic action potential recording from the epicardial surface, localizable with the MCG mapping and EMD inverse solution [red-circled solid arrow in (C)].](https://figures.cochat.ai/paper-figures/9f5ff125893644d38b40045bd5019e53/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-6.jpg)

![Figure 8 First reported comparison between sequential 36-point MCG recording with a single-channel OPM developed at the University of Freiburg (Swiss) and 36-channel MCG mapping at the Catholic University Hospital in Rome. All data were converted to the same format and analyzed with the Neuromag MCG software. The identity of MCG waveforms, of the MF distribution, of the related source localization (green arrows onto the MF maps), and of the 3D current density imaging (CDI), calculated at the apex of the QRS and of the T waves, is immediately evident [modified from data in (275)].](https://figures.cochat.ai/paper-figures/f344048c25cb4f58b8bae070a9c93376/clinical-magnetocardiography-the-unshielded-betpast-present-and-future-figure-8.jpg)

Verification
Claims the paper makes
18 Valid · 19 Not valid · 34 Not checked · 24 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“More recently, another study carried out with a new 37-channel (19 active sensors and 18 for noise reduction) prototype by Leeds University suggests that, with appropriate modeling, 5 of 38 magnetic QRS parameters could be used to provide a MCG estimate of left ventricular ejection fraction (264).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A truly portable compact multichannel MCG device
What the sources say
- Partly supportedcites [264] Beadle, 2021, Assessing heart disease using a novel magnetocardiography device, Biomed Phys Eng Express sourceExperimental — accuracy not yet measured
The Introduction and Methods describe a 37-channel prototype with 19 active sensors and 18 noise-reduction channels. The Results and discussion report extracting 38 MCG parameters and fitting an LVEF model, but the curve-fit limitations state that the model used three predictors, not five of the 38 parameters. The paper’s references to five parameters concern differences between the post-MI and healthy-control groups, not the number used to estimate LVEF. (LLM, conf=0.89)
hedgedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A first pilot clinical study (protocol NCT02359773, on ClinicalTrials.gov) had shown that the device provided high sensitivity (95.4%) and a negative predictive value (NPV) (97.7%) for the rule-out of healthy subjects and of patients whose chest pain was non-ischemic from those with ischemic heart disease (262).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A truly portable compact multichannel MCG device
What the sources say
- Partly supportedcites [262] Ghasemi-Roudsari, 2018, A portable prototype magnetometer to differentiate ischemic and non-ischemic heart disease in patients with chest pain, PLoS One sourceExperimental — accuracy not yet measured
Results > Model performance (Table 3) reports 95.4% sensitivity and 97.7% NPV for the cross-validated model comparing IHD patients with controls and young healthy volunteers; the NPV assumes 15% disease prevalence. But Results > Combined data set shows that these figures come from pooled technical-performance and pilot-clinical-study data, not the pilot clinical study alone. (LLM, conf=0.88)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“concluded that, at least in 2020, the VitalScan did not yet meet the level of accuracy required to confidently rule out ACS in the ED clinical practice (263).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A truly portable compact multichannel MCG device
What the sources say
- Supportedcites [263] Goodacre, 2021, Diagnostic accuracy of the magnetocardiograph for patients with suspected acute coronary syndrome, Emerg Med J source
The Methods (p. 6) describe a prospective, multicentre cohort study of patients presenting to the ED with suspected ACS. The Discussion (p. 13) concludes that the VitalScan MCG could not yet meet the accuracy required to rule out ACS for clinical use; the Conclusion (p. 3) likewise says it was not ready for clinical practice. (LLM, conf=0.95)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“although theory and experimental research suggest and confirm that it can provide additional diagnostic information not achievable with electrocardiography alone (1, 2, 4–6, 9, 16, 107).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Discussion
What the sources say
- Not checked — no full text of the sourcecites [1] Plonsey, 1972, Comparative capabilities of electrocardiography and magnetocardiography, Am J Cardiol source
- Not matchedcites [2] Malmivuo
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Not checked — no full text of the sourcecites [4] Wikswo, 1982, Possible sources of new information in the magnetocardiogram, J Theor Biol source
- Not checked — no full text of the sourcecites [5] Roth, 1986, Electrically silent magnetic fields, Biophys J source
- Not checked — no full text of the sourcecites [6] Lant, 1990, Complementary nature of electrocardiographic and magnetocardiographic data in patients with ischemic heart disease, J Electrocardiol source
- Not checked — no full text of the sourcecites [9] Baule, 1963, Detection of the magnetic field of the heart, Am Heart J source
- Not checked — no full text of the sourcecites [16] Cohen, 1975, Magnetic determination of the relationship between the S-T segment shift and the injury current produced by coronary artery occlusion, Circ Res source
- Supportedcites [107] Savard, 1983, Magnetic measurement of S-T and T-Q segment shifts in humans. Part I: early repolarization and left bundle branch block, Circ Res source
The paper’s introductory passage says the ECG cannot measure T-Q shifts or distinguish two causes of S-T shifts, whereas dcMCG can. The Discussion qualifies its practical use: extraneous signals prevented its use as a clinical diagnostic tool at the time, though it could clarify S-T shifts in selected subjects as a research tool. (LLM, conf=0.88)
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Every source the paper cites for this claim was checked and supports it.
In the paper
“Another non-cryogenic alternative to SQUID-based MCG systems is the use of OPMs (269).”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the present > A still shielded but more compact and non-cryogenic (OPM-based) multichannel MCG system for emergency departments
What the sources say
- Supportedcites [269] Lin, 2013, A compact, high performance atomic magnetometer for biomedical applications, Phys Med Biol source
The Introduction describes the atomic magnetometer as a viable alternative to SQUID magnetometers used in MCG systems. The Discussion states that atomic magnetometers do not require cryogenic cooling, and the Results report comparable MCG recordings from the atomic and SQUID magnetometers. (LLM, conf=0.94)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
36 more claims · 4 with their cited sources checked
Create a free account to see all 95 claims and to verify the 8 references still open.
- It may state two separate claims as one. 40
- Probably not a substantial claim. 17
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the past > From single-channel MCG mapping to multichannel MCG imaging in unshielded hospital settings
“the same unshielded CMI 3619a installation currently used for clinical MCG, mapping the cardiac MF of small animals (195) has been also performed”
§ Clinical magnetocardiography: the unshielded bet—past, present, and future > Unshielded MCG: the past > From single-channel MCG mapping to multichannel MCG imaging in unshielded hospital settings
“Figure: Figure 6 Typical experimental setup used for the MCG study of small animals with the CMI 3649a system.”
Citations
Cites
Yang (2021) · doi:10.1038/s41598-020-79139-8 · Cited by 5,035
Hämäläinen, M., Hari, R., Ilmoniemi, R. J., Knuutila, J., & Lounasmaa, O. V. (1993) · Reviews of Modern Physics · doi:10.1103/revmodphys.65.413 · Cited by 4,684
A subfemtotesla multichannel atomic magnetometer
Kominis, I. K., Kornack, T. W., Allred, J. C., & Romalis, M. V. (2003) · Nature · doi:10.1038/nature01484 · Cited by 1,631
Cohen, D. (1972) · Science · doi:10.1126/science.175.4022.664 · Cited by 785
Gepstein, L., Hayam, G., & Ben-Haim, S. A. (1997) · Circulation · doi:10.1161/01.cir.95.6.1611 · Cited by 759
Noninvasive electrocardiographic imaging for cardiac electrophysiology and arrhythmia
Ramanathan, C., Ghanem, R. N., Jia, P., Ryu, K., & Rudy, Y. (2004) · Nature Medicine · doi:10.1038/nm1011 · Cited by 646
Optical magnetic detection of single-neuron action potentials using quantum defects in diamond
Barry, J. F., Turner, M. J., Schloss, J. M., Glenn, D. R., Song, Y., Lukin, M. D., Park, H., & Walsworth, R. L. (2016) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1601513113 · Cited by 637
Potential clinical relevance of the “little brain” on the mammalian heart
Armour, J. A. (2008) · Experimental Physiology · doi:10.1113/expphysiol.2007.041178 · Cited by 417
Optically pumped magnetometers: from quantum origins to multi-channel magnetoencephalography
Tierney, T. M., Holmes, N., Mellor, S., López, J. D., Roberts, G., Hill, R. M., Boto, E., Leggett, J., Shah, V., Brookes, M. J., et al. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.05.063 · Cited by 400
Zimmerman, J. E., Thiene, P., & Harding, J. T. (1970) · Journal of Applied Physics · doi:10.1063/1.1659074 · Cited by 394 · cited 20×
“The concept was clearly demonstrated by the excellent work by the Indira Gandhi Centre for Atomic Research’s (IGCAR) researchers who, after EEMD processing, obtained a similar MCG signal quality with their DC-SQUID-based first gradiometer inside and outside the MSR (314); thus, we could anticipate that a similar resul…”
BackgroundInfluentialDetection of the magnetic field of the heart
Baule, G., & McFee, R. (1963) · American Heart Journal · doi:10.1016/0002-8703(63)90075-9 · Cited by 392 · cited 2×
“These attempts were finally achieved in the early 1960s, experimentally by recording the cardiac MF from an isolated rabbit heart preparation with a toroidal solenoid (8) and in humans by Baule and McFee (9), using two large magnetic sensor coils, although at the very low spatial resolution, which obviously made it st…”
MethodMagnetocardiograms taken inside a shielded room with a superconducting point-contact magnetometer
Cohen, D., Edelsack, E. A., & Zimmerman, J. E. (1970) · Applied Physics Letters · doi:10.1063/1.1653195 · Cited by 361 · cited 1×
“Providing a significant improvement in sensitivity and spatial resolution of MF measurements, a major milestone in MCG was the invention of the first superconducting quantum interference device (SQUID) (10) and the installation of SQUID within the magnetically shielded room (MSR) of the Massachusetts Institute of Tech…”
BackgroundDistribution of heart potentials on the thoracic surface of normal human subjects
TACCARDI, B. (1963) · Circulation Research · doi:10.1161/01.res.12.4.341 · Cited by 337
A compact, high performance atomic magnetometer for biomedical applications
Shah, V. K., & Wakai, R. T. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/22/8153 · Cited by 322
Magnetic sensors-A review and recent technologies
Khan, M. A., Sun, J., Li, B., Przybysz, A., & Kosel, J. (2021) · Engineering Research Express · doi:10.1088/2631-8695/ac0838 · Cited by 315
Ultrasensitive Magnetic Field Sensors for Biomedical Applications
Murzin, D., Mapps, D. J., Levada, K., Belyaev, V., Omelyanchik, A., Panina, L., & Rodionova, V. (2020) · Sensors · doi:10.3390/s20061569 · Cited by 277
Portable Magnetometry for Detection of Biomagnetism in Ambient Environments
Limes, M. E., Foley, E. L., Kornack, T. W., Caliga, S., McBride, S., Braun, A., Lee, W., Lucivero, V. G., & Romalis, M. V. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.14.011002 · Cited by 257
The inverse problem in electrocardiography: solutions in terms of epicardial potentials.
Rudy, Y., & Messinger‐Rapport, B. J. (1988) · PubMed · Cited by 254
LocaLisa: New Technique for Real-Time 3-Dimensional Localization of Regular Intracardiac Electrodes
Wittkampf, F. H. M., Wever, E. F. D., Derksen, R., Wilde, A. A. M., Ramanna, H., Hauer, R. N. W., & Robles de Medina, E. O. (1999) · Circulation · doi:10.1161/01.cir.99.10.1312 · Cited by 238
Romani, G. L., Williamson, S. J., & Kaufman, L. (1982) · Review of Scientific Instruments · doi:10.1063/1.1136907 · Cited by 235
Model Studies of the Magnetocardiogram
Grynszpan, F., & Geselowitz, D. B. (1973) · Biophysical Journal · doi:10.1016/s0006-3495(73)86034-5 · Cited by 226
Recording brain activities in unshielded Earth’s field with optically pumped atomic magnetometers
Zhang, R., Xiao, W., Ding, Y., Feng, Y., Peng, X., Shen, L., Sun, C., Wu, T., Wu, Y., Yang, Y., et al. (2020) · Science Advances · doi:10.1126/sciadv.aba8792 · Cited by 222
Biomagnetism using SQUIDs: status and perspectives
Sternickel, K., & Braginski, A. I. (2006) · Superconductor Science and Technology · doi:10.1088/0953-2048/19/3/024 · Cited by 180
A microfabricated optically-pumped magnetic gradiometer
Sheng, D., Perry, A. R., Krzyzewski, S. P., Geller, S., Kitching, J., & Knappe, S. (2017) · Applied Physics Letters · doi:10.1063/1.4974349 · Cited by 179
Clinical application of magnetocardiography
Fenici, R., Brisinda, D., & Meloni, A. M. (2005) · Expert Review of Molecular Diagnostics · doi:10.1586/14737159.5.3.291 · Cited by 174 · cited 3×
“Ambulatory MCG study became a routine ambulatory procedure to study cardiac patients, including those with arrhythmias, to improve the non-invasive mechanistic diagnostic accuracy provided by electrocardiographic methods (67, 102).”
BackgroundMethodBiomagnetism: An Interdisciplinary Approach
Kaufman, L., & Williamson, S. J. (1983) · Medical Entomology and Zoology · Cited by 168
Zimmerman, J. E., & Frederick, N. V. (1971) · Applied Physics Letters · doi:10.1063/1.1653725 · Cited by 152
Yoshida, N., Inden, Y., Uchikawa, T., Kamiya, H., Kitamura, K., Shimano, M., Tsuji, Y., Hirai, M., & Murohara, T. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.11.023 · Cited by 146
Performance and limitations of noninvasive cardiac activation mapping
Duchateau, J., Sacher, F., Pambrun, T., Derval, N., Chamorro-Servent, J., Denis, A., Ploux, S., Hocini, M., Jaïs, P., Bernus, O., et al. (2019) · Heart Rhythm · doi:10.1016/j.hrthm.2018.10.010 · Cited by 146
Strand, S., Lutter, W., Strasburger, J. F., Shah, V., Baffa, O., & Wakai, R. T. (2019) · Journal of the American Heart Association · doi:10.1161/jaha.119.013436 · Cited by 136
Magnetocardiography with a 16-channel fiber-coupled single-cell Rb optically pumped magnetometer
Kim, Y. J., Savukov, I., & Newman, S. (2019) · Applied Physics Letters · doi:10.1063/1.5094339 · Cited by 133
Giant Magnetoresistance Biosensors in Biomedical Applications
Wu, K., Tonini, D., Liang, S. A., Saha, R., Chugh, V. K., & Wang, J.-P. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c20141 · Cited by 133
Cohen, D., & Kaufman, L. A. (1975) · Circulation Research · doi:10.1161/01.res.36.3.414 · Cited by 131
A new wearable multichannel magnetocardiogram system with a SERF atomic magnetometer array
Yang, Y., Xu, M., Liang, A., Yin, Y., Ma, X., Gao, Y., & Ning, X. (2021) · Scientific Reports · doi:10.1038/s41598-021-84971-7 · Cited by 125
Magnetic fields around the torso: production by electrical activity of the human heart
Cohen, D. (1967) · Science · doi:10.1126/science.156.3775.652 · Cited by 116 · cited 1×
“Providing a significant improvement in sensitivity and spatial resolution of MF measurements, a major milestone in MCG was the invention of the first superconducting quantum interference device (SQUID) (10) and the installation of SQUID within the magnetically shielded room (MSR) of the Massachusetts Institute of Tech…”
BackgroundA movable unshielded magnetocardiography system
Xiao, W., Sun, C., Shen, L., Feng, Y., Liu, M., Wu, Y., Liu, X., Wu, T., Peng, X., & Guo, H. (2023) · Science Advances · doi:10.1126/sciadv.adg1746 · Cited by 116
(1974) · Journal of Perinatal Medicine · doi:10.1515/jpme.1974.2.3.214 · Cited by 112
GEPSTEIN, L., & EVANS, S. J. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb00187.x · Cited by 112
Magnetocardiography Predicts Coronary Artery Disease in Patients with Acute Chest Pain
Park, J., Hill, P. M., Chung, N., Hugenholtz, P. G., & Jung, F. (2005) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2005.00634.x · Cited by 109
Single-beat noninvasive imaging of cardiac electrophysiology of ventricular pre-excitation
Berger, T., Fischer, G., Pfeifer, B., Modre, R., Hanser, F., Trieb, T., Roithinger, F. X., Stuehlinger, M., Pachinger, O., Tilg, B., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.08.019 · Cited by 107
Kwong, J. S. W., Leithäuser, B., Park, J.-W., & Yu, C.-M. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.12.056 · Cited by 103
Miniature cavity-enhanced diamond magnetometer
Chatzidrosos, G., Wickenbrock, A., Bougas, L., Leefer, N., Wu, T., Jensen, K., Dumeige, Y., & Budker, D. (2017) · Physical Review Applied · doi:10.1103/physrevapplied.8.044019 · Cited by 103
BROCKMEIER, K., SCHMITZ, L., CHAVEZ, J. D. J. B., BURGHOFF, M., KOCH, H., ZIMMERMANN, R., & TRAHMS, L. (1997) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.1997.tb01824.x · Cited by 99 · cited 1×
“Frontiers in Cardiovascular Medicine 13 capability in numerous clinical fields, such as but not limited to the earlier detection of (sometimes electrically silent) myocardial ischemia (59, 88, 90, 323, 319), the preventive arrhythmogenic risk assessment in adults (53, 225, 325–322) as well as in fetuses (83, 323, 324)…”
BackgroundMagnetocardiographic functional localization using a current dipole in a realistic torso
Nenonen, J., Purcell, C. J., Horacek, B. M., Stroink, G., & Katila, T. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83565 · Cited by 97
Localized structural alterations underlying a subset of unexplained sudden cardiac death
Haïssaguerre, M., Hocini, M., Cheniti, G., Duchateau, J., Sacher, F., Puyo, S., Cochet, H., Takigawa, M., Denis, A., Martin, R., et al. (2018) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.117.006120 · Cited by 96
Revishvili, A. S., Wissner, E., Lebedev, D. S., Lemes, C., Deiss, S., Metzner, A., Kalinin, V. V., Sopov, O. V., Labartkava, E. Z., Kalinin, A. V., et al. (2015) · Europace · doi:10.1093/europace/euu339 · Cited by 89
Millimetre-scale magnetocardiography of living rats with thoracotomy
Arai, K., Kuwahata, A., Nishitani, D., Fujisaki, I., Matsuki, R., Nishio, Y., Xin, Z., Cao, X., Hatano, Y., Onoda, S., et al. (2022) · Communications Physics · doi:10.1038/s42005-022-00978-0 · Cited by 85
High Resolution Electrocardiography
Peralta, M. R., González-Hermosillo, A., & Alfaro, J. K. (1992) · Annals of Internal Medicine · doi:10.7326/0003-4819-117-2-176_6 · Cited by 84
Nademanee, K., Chung, F.-P., Sacher, F., Nogami, A., Nakagawa, H., Jiang, C., Hocini, M., Behr, E., Veerakul, G., Jan Smit, J., et al. (2023) · Circulation · doi:10.1161/circulationaha.122.063367 · Cited by 83
Jensen, K., Skarsfeldt, M. A., Stærkind, H., Arnbak, J., Balabas, M. V., Olesen, S.-P., Bentzen, B. H., & Polzik, E. S. (2018) · Scientific Reports · doi:10.1038/s41598-018-34535-z · Cited by 79 · cited 2×
“The interest for MCG in animals’ experimental models is progressively increasing, and its application for the electrophysiological study of transgenic models of cardiomyopathy, experimental myocardial injury, and regulatory pharmacological preclinical evaluation has been confirmed by other authors, working in MSR with…”
BackgroundMethodCharacterizing atomic magnetic gradiometers for fetal magnetocardiography
Sulai, I. A., DeLand, Z. J., Bulatowicz, M. D., Wahl, C. P., Wakai, R. T., & Walker, T. G. (2019) · Review of Scientific Instruments · doi:10.1063/1.5091007 · Cited by 79
Tolstrup, K., Madsen, B. E., Ruiz, J. A., Greenwood, S. D., Camacho, J., Siegel, R. J., Gertzen, H. C., Park, J.-W., & Smars, P. A. (2006) · Cardiology · doi:10.1159/000093490 · Cited by 78
Distribution of heart potentials on dog's thoracic surface
Taccardi, B. (1962) · Circulation Research · doi:10.1161/01.res.11.5.862 · Cited by 76
Multichannel biomagnetic system for study of electrical activity in the brain and heart
Schneider, S., Hoenig, E., Reichenberger, H., Abraham-Fuchs, K., Moshage, W., Oppelt, A., Stefan, H., Weikl, A., & Wirth, A. (1990) · Radiology · doi:10.1148/radiology.176.3.2389043 · Cited by 74
Angiotensin II-induced sudden arrhythmic death and electrical remodeling
Fischer, R., Dechend, R., Gapelyuk, A., Shagdarsuren, E., Gruner, K., Gruner, A., Gratze, P., Qadri, F., Wellner, M., Fiebeler, A., et al. (2007) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.01400.2006 · Cited by 71
Tsukada, K., Miyashita, T., Kandori, A., Mitsui, T., Terada, Y., Sato, M., Shiono, J., Horigome, H., Yamada, S., & Yamaguchi, I. (2000) · International journal of cardiac imaging · doi:10.1023/a:1006376326755 · Cited by 70
YAMADA, S., & YAMAGUCHI, I. (2005) · Internal Medicine · doi:10.2169/internalmedicine.44.1 · Cited by 70
Pena, M. E., Pearson, C. L., Goulet, M. P., Kazan, V. M., DeRita, A. L., Szpunar, S. M., & Dunne, R. B. (2020) · IJC Heart & Vasculature · doi:10.1016/j.ijcha.2019.100466 · Cited by 70
Baule, G. M., & McFee, R. (1970) · American Heart Journal · doi:10.1016/0002-8703(70)90312-1 · Cited by 69
Possible sources of new information in the magnetocardiogram
Wikswo, J. P., & Barach, J. P. (1982) · Journal of Theoretical Biology · doi:10.1016/0022-5193(82)90350-2 · Cited by 68
Wearable OPM-MEG: a changing landscape for epilepsy
Pedersen, M., Abbott, D. F., & Jackson, G. D. (2022) · Epilepsia · doi:10.1111/epi.17368 · Cited by 68
Fiber-Coupled Diamond Magnetometry with an Unshielded Sensitivity of 30 pT / Hz
Graham, S. M., Rahman, A., Munn, L., Patel, R. L., Newman, A. J., Stephen, C. J., Colston, G., Nikitin, A. P., Edmonds, A. M., Twitchen, D. J., et al. (2023) · Physical Review Applied · doi:10.1103/physrevapplied.19.044042 · Cited by 66
Biomagnetism: The First Sixty Years
Roth, B. J. (2023) · Sensors · doi:10.3390/s23094218 · Cited by 66
Lant, J., Stroink, G., ten Voorde, B., Horacek, B. M., & Montague, T. J. (1990) · Journal of Electrocardiology · doi:10.1016/0022-0736(90)90121-h · Cited by 63
Petrini, G., Moreva, E., Bernardi, E., Traina, P., Tomagra, G., Carabelli, V., Degiovanni, I. P., & Genovese, M. (2020) · Advanced Quantum Technologies · doi:10.1002/qute.202000066 · Cited by 63
100 years of superconductivity
(2012) · Choice Reviews Online · doi:10.5860/choice.49-6336 · Cited by 63
A 3-D model-based registration approach for the PET, MR and MCG cardiac data fusion
Mäkelä, T. J., Pham, Q. C., Clarysse, P., Nenonen, J., Lötjönen, J. M. P., Sipilä, O., Hänninen, H., Lauerma, K., Knuuti, J. M., Katila, T. E., et al. (2003) · Medical Image Analysis · doi:10.1016/s1361-8415(03)00012-4 · Cited by 62
The magnetocardiogram in cardiac disorders
SAARINEN, M., KARP, P. J., KATILA, T. E., & SILTANEN, P. (1974) · Cardiovascular Research · doi:10.1093/cvr/8.6.820 · Cited by 61
The inverse problem in electrocardiography: solutions in terms of equivalent sources.
Gulrajani, R. M., Savard, P., & ROBERGE, F. A. (1988) · PubMed · Cited by 59
Solving the inverse problem in magnetocardiography
Nenonen, J. T. (1994) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.310989 · Cited by 59
Wacker‐Gussmann, A., Strasburger, J. F., & Wakai, R. T. (2022) · Journal of the American Heart Association · doi:10.1161/jaha.121.025224 · Cited by 58 · cited 2×
“Frontiers in Cardiovascular Medicine 13 capability in numerous clinical fields, such as but not limited to the earlier detection of (sometimes electrically silent) myocardial ischemia (59, 88, 90, 323, 319), the preventive arrhythmogenic risk assessment in adults (53, 225, 325–322) as well as in fetuses (83, 323, 324)…”
BackgroundElectrically silent magnetic fields
Roth, B. J., & Wikswo, J. P. (1986) · Biophysical Journal · doi:10.1016/s0006-3495(86)83513-5 · Cited by 57
Kurashima, K., Kataoka, M., Nakano, T., Fujiwara, K., Kato, S., Nakamura, T., Yuzawa, M., Masuda, M., Ichimura, K., Okatake, S., et al. (2023) · Sensors · doi:10.3390/s23020646 · Cited by 56
HAILER, B., CHAIKOVSKY, I., AUTH‐EISERNITZ, S., SCHÄFER, H., & VAN LEEUWEN, P. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.09318.x · Cited by 54
Brala, D., Thevathasan, T., Grahl, S., Barrow, S., Violano, M., Bergs, H., Golpour, A., Suwalski, P., Poller, W., Skurk, C., et al. (2023) · Journal of the American Heart Association · doi:10.1161/jaha.122.027619 · Cited by 54
Cohen, D., Norman, J. C., Molokhia, F., & Hood, W. (1971) · Science · doi:10.1126/science.172.3990.1329 · Cited by 53
Magnetocardiography: current status and perspectives. Part II: Clinical applications.
Tavarozzi, I., Comani, S., Gratta, C. D., Luzio, S. D., Romani, G. L., Gallina, S., Zimarino, M., Brisinda, D., Fenici, R., & Caterina, R. D. (2002) · PubMed · Cited by 53
Self-citationLim, H. K., Kwon, H., Chung, N., Ko, Y.-G., Kim, J.-M., Kim, I.-S., & Park, Y.-K. (2009) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2008.10.013 · Cited by 52
Noninvasive epicardial and endocardial mapping of premature ventricular contractions
Wissner, E., Revishvili, A., Metzner, A., Tsyganov, A., Kalinin, V., Lemes, C., Saguner, A. M., Maurer, T., Deiss, S., Sopov, O., et al. (2016) · Europace · doi:10.1093/europace/euw103 · Cited by 52
Cohen, D., Savard, P., Rifkin, R. D., Lepeschkin, E., & Strauss, W. E. (1983) · Circulation Research · doi:10.1161/01.res.53.2.274 · Cited by 51
Neuromagnetic functional localization: principles, state of the art, and perspectives
Romani, G. L., & Rossini, P. (1988) · Brain Topography · doi:10.1007/bf01129335 · Cited by 51
Moshage, W. E. L., Achenbach, S. S., Göhl, K., & Bachmann, K. F. (1996) · International journal of cardiac imaging · doi:10.1007/bf01798116 · Cited by 51
KORHONEN, P., HUSA, T., TIERALA, I., VÄÄNÄNEN, H., MÄKIJÄRVI, M., KATILA, T., & TOIVONEN, L. (2006) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2005.00332.x · Cited by 51
Conversion of magnetocardiographic recordings between two different multichannel SQUID devices
Burghoff, M., Nenonen, J., Trahms, L., & Katila, T. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.846680 · Cited by 51
Complex and novel arrhythmias precede stillbirth in fetuses with de novo long QT syndrome
Strand, S., Strasburger, J. F., Cuneo, B. F., & Wakai, R. T. (2020) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.119.008082 · Cited by 50
Hänninen, H., Takala, P., Korhonen, P., Oikarinen, L., Mäkijärvi, M., Nenonen, J., Katila, T., & Toivonen, L. (2002) · Annals of Medicine · doi:10.1080/07853890252953518 · Cited by 49
Non-invasive imaging of arrhythmogenic left-ventricular myocardium after infarction
Leder, U., Haueisen, J., Huck, M., & Nowak, H. (1998) · The Lancet · doi:10.1016/s0140-6736(98)00082-8 · Cited by 48
Zhang, R., Mhaskar, R., Smith, K., & Prouty, M. (2020) · Applied Physics Letters · doi:10.1063/5.0004746 · Cited by 48
Investigations of sensitivity and resolution of ECG and MCG in a realistically shaped thorax model
Mäntynen, V., Konttila, T., & Stenroos, M. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/23/7141 · Cited by 45
The normal human magnetocardiogram. II. A multipole analysis.
Karp, P. J., Katila, T. E., Saarinen, M., Siltanen, P., & Varpula, T. T. (1980) · Circulation Research · doi:10.1161/01.res.47.1.117 · Cited by 43
Savard, P., Cohen, D., Lepeschkin, E., Cuffin, B. N., & Madias, J. E. (1983) · Circulation Research · doi:10.1161/01.res.53.2.264 · Cited by 43
Integrated Giant Magnetoresistance Technology for Approachable Weak Biomagnetic Signal Detections
Shen, H., Hu, L., & Fu, X. H. (2018) · Sensors · doi:10.3390/s18010148 · Cited by 43
Magnetocardiogram recordings in a nonshielded environment—reproducibility and ischemia detection
Steinberg, B. A., Roguin, A., Watkins, S. P., Hill, P., Fernando, D., & Resar, J. R. (2005) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2005.05611.x · Cited by 42
Kirchhof, P. F., Loh, P. K. K., Eckardt, L., Ribbing, M., Rolf, S., EICK, O. J., Wittkampf, F. H. M., Borggrefe, M., Breithardt, G. Ü., & Haverkamp, W. (2002) · The American Journal of Cardiology · doi:10.1016/s0002-9149(02)02481-5 · Cited by 42
Clinical utility of magnetocardiography in cardiology for the detection of myocardial ischemia
Camm, A. J., Henderson, R., Brisinda, D., Body, R., Charles, R. G., Varcoe, B., & Fenici, R. (2019) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2019.07.009 · Cited by 42 · cited 4×
“Frontiers in Cardiovascular Medicine 08 myocardial ischemia (71, 72, 74, 76–78, 211–216) also during effort test (217) and the emergency triage of patients with chest pain (87, 88).”
BackgroundMethodInfluentialSelf-citationMagnetic Study of the His-Purkinje Conduction System in Man
Farrell, D. E., Tripp, J. H., & Norgren, R. (1980) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1980.326646 · Cited by 41 · cited 1×
“Although a preliminary magnetic measurement of the PR interval’s MF had been reported in 1978 (116), high-resolution (HR) MCG recordings of the PR segment were carried out independently both with unshielded MCG and in MSR since the early 1980s (27, 33, 117, 118).”
BackgroundA consensus statement on relative merits of EEG and MEG
Anogianakis, G., Badier, J., Barrett, G., Erne, S. N., Fenici, R., Fenwick, P. B. C., Grandori, F., Hari, R., Ilmoniemi, R. J., Mauguìere, F., et al. (1992) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(92)90001-x · Cited by 40
Self-citationHailer, B., Leeuwen, P. V., Lange, S., & Wehr, M. (1999) · Journal of Electrocardiology · doi:10.1016/s0022-0736(99)90103-6 · Cited by 39
Magnetocardiography: Ventricular arrhythmias
Fenici, R., & Melillo, G. (1993) · European Heart Journal · doi:10.1093/eurheartj/14.suppl_e.53 · Cited by 39
Self-citationDutz, S., Bellemann, M. E., Leder, U., & Haueisen, J. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/1/011 · Cited by 39
A portable diagnostic device for cardiac magnetic field mapping
Mooney, J. W., Ghasemi-Roudsari, S., Banham, E. R., Symonds, C., Pawlowski, N., & Varcoe, B. T. H. (2017) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/3/1/015008 · Cited by 39
Calibration of Room Temperature Magnetic Sensor Array for Biomagnetic Measurement
Adachi, Y., Oyama, D., Terazono, Y., Hayashi, T., Shibuya, T., & Kawabata, S. (2019) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2019.2895355 · Cited by 39
Perry, A. R., Bulatowicz, M. D., Larsen, M., Walker, T. G., & Wyllie, R. (2020) · Optics Express · doi:10.1364/oe.408486 · Cited by 39
COMANI, S., LIBERATI, M., MANTINI, D., GABRIELE, E., BRISINDA, D., DI LUZIO, S., FENICI, R., & LUCA ROMANI, G. (2004) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2004.00699.x · Cited by 38 · cited 1×
“The parallel research efforts carried out with shielded and unshielded MCG over the last three decades have enlarged the knowledge about the pros and cons of these two approaches and provided evidence of well-defined fields of clinical application, such as the emergency triage of patients with chest pain, the diagnosi…”
BackgroundSelf-citationA Shielded Facility for Low-Level Magnetic Measurements
Cohen, D. (1967) · Journal of Applied Physics · doi:10.1063/1.1709590 · Cited by 38 · cited 1×
“Providing a significant improvement in sensitivity and spatial resolution of MF measurements, a major milestone in MCG was the invention of the first superconducting quantum interference device (SQUID) (10) and the installation of SQUID within the magnetically shielded room (MSR) of the Massachusetts Institute of Tech…”
BackgroundMariyappa, N., Sengottuvel, S., Parasakthi, C., Gireesan, K., Janawadkar, M. P., Radhakrishnan, T. S., & Sundar, C. S. (2014) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2014.06.023 · Cited by 38
Magnetometry with nitrogen-vacancy centers in diamond
Jensen, K., Kehayias, P., & Budker, D. (2017) · Smart Sensors, Measurement and Instrumentation · doi:10.1007/978-3-319-34070-8_18 · Cited by 38
Spatial Distribution of Repolarization Times in Patients with Coronary Artery Disease
LEEUWEN, P. V., HAILER, B., LANGE, S., & GRÖNEMEYER, D. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.t01-1-00256.x · Cited by 37
Body surface electrocardiographic mapping for non-invasive identification of arrhythmic sources
Shah, A. J., Hocini, M., Pascale, P., Roten, L., Komatsu, Y., Daly, M., Ramoul, K., Denis, A., Derval, N., Sacher, F., et al. (2013) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2013.2.1.16 · Cited by 37
A study of scalar optically-pumped magnetometers for use in magnetoencephalography without shielding
Clancy, R. J., Gerginov, V., Alem, O., Becker, S., & Knappe, S. (2021) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ac18fb · Cited by 37
Magnetocardiography: Supraventricular arrhythmias and preexcitation syndromes
Makijarvi, M., Nenonen, J., Toivonen, L., Montonen, J., Katila, T., & Siltanen, P. (1993) · European Heart Journal · doi:10.1093/eurheartj/14.suppl_e.46 · Cited by 36
Wang, S., Zhou, X., Huang, B., Wang, Z., Zhou, L., Wang, M., Yu, L., & Jiang, H. (2016) · Scientific Reports · doi:10.1038/srep30783 · Cited by 36
Fujino, K., Sumi, M., Saito, K., Murakami, M., Higuchi, T., Nakaya, Y., & Mori, H. (1984) · Journal of Electrocardiology · doi:10.1016/s0022-0736(84)80058-8 · Cited by 35
Recording the heart beat of cattle using a gradiometer system of optically pumped magnetometers
Sutter, J. U., Lewis, O., Robinson, C., McMahon, A., Boyce, R., Bragg, R., Macrae, A., Orton, J., Shah, V., Ingleby, S. J., et al. (2020) · Computers and Electronics in Agriculture · doi:10.1016/j.compag.2020.105651 · Cited by 35
Nakai, K., Izumoto, H., Kawazoe, K., Tsuboi, J., Fukuhiro, Y., Oka, T., Yoshioka, K., Shozushima, M., Itoh, M., Suwabe, A., et al. (2005) · International journal of cardiac imaging · doi:10.1007/s10554-005-9019-x · Cited by 34
Bioelectromagnetic localization of a pacing catheter in the heart
Pesola, K., Nenonen, J., Fenici, R., Lötjönen, J., Mäkijärvi, M., Fenici, P., Korhonen, P., Lauerma, K., Valkonen, M., Toivonen, L., et al. (1999) · Physics in Medicine and Biology · doi:10.1088/0031-9155/44/10/314 · Cited by 34
Self-citationThe normal magnetocardiogram: I Morphology.
Saarinen, M., SILTANEN, P., Karp, P., & Katila, T. E. (1978) · PubMed · Cited by 34
Advantages and pitfalls of noninvasive electrocardiographic imaging
Bear, L. R., Bouhamama, O., Cluitmans, M. J. M., Duchâteau, J., Walton, R. D., Abell, E., Belterman, C. N. W., Haı̈ssaguerre, M., Bernus, O., Coronel, R., et al. (2019) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2019.08.007 · Cited by 34
Optimizing NV magnetometry for magnetoneurography and magnetomyography applications
Zhang, C., Zhang, J., Widmann, M., Benke, M., Kübler, M., Dasari, D., Klotz, T., Gizzi, L., Röhrle, O., Brenner, P., et al. (2023) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.1034391 · Cited by 34
Body Surface Potential Mapping: Historical Background, Present Possibilities, Diagnostic Challenges
Medvegy, M., Duray, G. Z., Pintér, A., & Préda, I. (2002) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2002.tb00155.x · Cited by 33
Biomagnetometry: imaging the heart's magnetic field.
Hart, G. (1991) · Heart · doi:10.1136/hrt.65.2.61 · Cited by 33
Part I Abnormal patterns and physiological variations in magnetocardiograms
Cohen, D., Lepeschkin, E., Hosaka, H., Massell, B. F., & Myers, G. S. (1976) · Journal of Electrocardiology · doi:10.1016/s0022-0736(76)80040-4 · Cited by 32
Tsyganov, A., Wißner, E., Metzner, A., Mironovich, S., Chaykovskaya, M., Kalinin, V., Chmelevsky, M. P., Lemeš, C., & Kück, K. (2017) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2017.07.018 · Cited by 32
Inaba, T., Nakazawa, Y., Yoshida, K., Kato, Y., Hattori, A., Kimura, T., Hoshi, T., Ishizu, T., Seo, Y., Sato, A., et al. (2017) · Superconductor Science and Technology · doi:10.1088/1361-6668/aa8c26 · Cited by 32
Detection of patients with coronary artery disease using cardiac magnetic field mapping at rest
Gapelyuk, A., Wessel, N., Fischer, R., Zacharzowsky, U. B., Koch, L., Selbig, D., Schütt, H., Sawitzki, B., Luft, F. C., Dietz, R., et al. (2007) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2007.03.013 · Cited by 31
The use of low-level electromagnetic fields to suppress atrial fibrillation
Yu, L., Dyer, J. W., Scherlag, B. J., Stavrakis, S., Sha, Y., Sheng, X., Garabelli, P., Jacobson, J., & Po, S. S. (2015) · Heart Rhythm · doi:10.1016/j.hrthm.2014.12.022 · Cited by 31
Magnetocardiography in coronary artery disease with a new system in an unshielded setting
Hailer, B., Chaikovsky, I., Auth‐Eisernitz, S., Schäfer, H., Steinberg, F., & Grönemeyer, D. H. W. (2003) · Clinical Cardiology · doi:10.1002/clc.4960261007 · Cited by 30
Kim, D., Kim, K., Lee, Y.-H., & Ahn, H. (2007) · Interactive CardioVascular and Thoracic Surgery · doi:10.1510/icvts.2006.142869 · Cited by 29
Use of a superconducting instrumentation for biomagnetic measurements performed in a hospital
Barbanera, S., Carelli, P., Fenici, R., Leoni, R., Modena, I., & Romani, G. (1981) · IEEE Transactions on Magnetics · doi:10.1109/tmag.1981.1061052 · Cited by 29 · cited 2×
“Although a preliminary magnetic measurement of the PR interval’s MF had been reported in 1978 (116), high-resolution (HR) MCG recordings of the PR segment were carried out independently both with unshielded MCG and in MSR since the early 1980s (27, 33, 117, 118).”
BackgroundMethodSelf-citationPark, J.-W., Shin, E.-S., Ann, S. H., Gödde, M., Park, L. S.-I., Brachmann, J., Vidal-Lopez, S., Wierzbinski, J., Lam, Y.-Y., & Jung, F. (2015) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-141912 · Cited by 29
FENICI, R., NENONEN, J., PESOLA, K., KORHONEN, P., LÖTJÖNEN, J., MÄKIJÄRVI, M., TOIVONEN, L., POUTANEN, V., KETO, P., & KATILA, T. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00602.x · Cited by 29
Self-citationMagnetocardiography Using a Magnetoresistive Sensor Array
Shirai, Y., Hirao, K., Shibuya, T., Okawa, S., Hasegawa, Y., Adachi, Y., Sekihara, K., & Kawabata, S. (2019) · International Heart Journal · doi:10.1536/ihj.18-002 · Cited by 29
Aita, S., Ogata, K., Yoshida, K., Inaba, T., Kosuge, H., Machino, T., Tsumagari, Y., Hattori, A., Ito, Y., Komatsu, Y., et al. (2019) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2019.06.010 · Cited by 29
Park, J.-W., Leithäuser, B., Vršansky, M., & Jung, F. (2008) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-2008-1064 · Cited by 27
Ability of magnetocardiography to detect regional dominant frequencies of atrial fibrillation
Yoshida, K., Ogata, K., Inaba, T., Nakazawa, Y., Ito, Y., Yamaguchi, I., Kandori, A., & Aonuma, K. (2015) · Journal of Arrhythmia · doi:10.1016/j.joa.2015.05.003 · Cited by 27
Sensor‐Based Electromagnetic Navigation (Mediguide®): How Accurate Is It? A Phantom Model Study
Bourier, F. J., Reents, T., Ammar‐Busch, S., et al. (2015) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.12741 · Cited by 27
Brisinda, D., Meloni, A. M., & Fenici, R. (2003) · Lecture notes in computer science · doi:10.1007/3-540-44883-7_13 · Cited by 27
Self-citationNenonen, J., Pesola, K., H[auml ]nninen, H., Lauerma, K., Takala, P., M[auml ]kel[auml ], T., M[auml ]kij[auml ]rvi, M., Knuuti, J., Toivonen, L., & Katila, T. (2001) · Journal of Electrocardiology · doi:10.1054/jelc.2001.28824 · Cited by 27
Kawakami, S., Takaki, H., Hashimoto, S., Kimura, Y., Nakashima, T., Aiba, T., Kusano, K. F., Kamakura, S., Yasuda, S., & Sugimachi, M. (2017) · Circulation Journal · doi:10.1253/circj.cj-16-0683 · Cited by 27
(2003) · Cited by 26
Self-citationMagnetism and cardiac arrhythmias
Scherlag, B. J., Yamanashi, W. S., Hou, Y., Jacobson, J. I., Jackman, W. M., & Lazzara, R. (2004) · Cardiology in Review · doi:10.1097/01.crd.0000094029.10223.2f · Cited by 26
Magnetocardiography: An Overview
Geselowitz, D. B. (1979) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1979.326430 · Cited by 26
Escalona-Vargas, D., Bolin, E. H., Lowery, C. L., Siegel, E. R., & Eswaran, H. (2020) · Physiological Measurement · doi:10.1088/1361-6579/abc353 · Cited by 26
Erne, S. N., Fenici, R., Hahlbohm, H.-D., Jaszczuk, W., Lehmann, H. P., & Masselli, M. (1983) · Il Nuovo Cimento D · doi:10.1007/bf02455936 · Cited by 25
Self-citationFragmentation of bandpass-filtered QRS-complex of patients prone to malignant arrhythmia
ENDT, P., Hahlbohm, H.-D., Kreiseler, D., Oeff, M., Steinhoff, U., & Trahms, L. (1998) · Medical & Biological Engineering & Computing · doi:10.1007/bf02518875 · Cited by 24
Comparative capabilities of electrocardiography and magnetocardiography
Plonsey, R. (1972) · The American Journal of Cardiology · doi:10.1016/0002-9149(72)90179-8 · Cited by 24 · cited 1×
“Trying to identify the possible reasons for such translational difficulty, already in 2005, we had suggested two major causes that (1) MCG technology was at that time still too sophisticated, expensive, and not available to the patient’s bedside and (2) a sort of vicious circle, where the unavailability of MCG devices…”
BackgroundClinical magnetocardiography: 10 Years experience at the catholic university
Fenici, R. R., Melillo, G., & Masselli, M. (1991) · The International Journal of Cardiac Imaging · doi:10.1007/bf01797748 · Cited by 23 · cited 1×
“The parallel research efforts carried out with shielded and unshielded MCG over the last three decades have enlarged the knowledge about the pros and cons of these two approaches and provided evidence of well-defined fields of clinical application, such as the emergency triage of patients with chest pain, the diagnosi…”
BackgroundSelf-citationBrisinda, D., Caristo, M. E., & Fenici, R. (2006) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.01048.2005 · Cited by 23
Self-citationNon-invasive mapping of cardiac arrhythmias
Shah, A., Hocini, M., Haissaguerre, M., & Jaïs, P. (2015) · Current Cardiology Reports · doi:10.1007/s11886-015-0616-6 · Cited by 22
Fenici, R., Brisinda, D., Venuti, A., & Sorbo, A. R. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2013.07.238 · Cited by 22
Self-citationFENICI, R., PESOLA, K., KORHONEN, P., MÄKIJÄRVI, M., NENONEN, J., TOIVONEN, L., FENICI, P., & KATILA, T. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb01207.x · Cited by 22
Self-citationApplication of superconducting magnetometers to the measurement of the vector magnetocardiogram
Wikswo, J., & Fairbank, W. (1977) · IEEE Transactions on Magnetics · doi:10.1109/tmag.1977.1059333 · Cited by 22
Ito, Y., Shiga, K., Yoshida, K., Ogata, K., Kandori, A., Inaba, T., Nakazawa, Y., Sekiguchi, Y., Tada, H., Sekihara, K., et al. (2014) · Heart Rhythm · doi:10.1016/j.hrthm.2014.05.032 · Cited by 22
Wu, Y., Lin, L. C., Tseng, W., Liu, Y., Kao, H.-L., Lin, M. S., Huang, H. C., Wang, S. Y., Horng, H. E., Yang, H. C., et al. (2014) · PubMed · Cited by 22
Miniature Coil Array for Passive Magnetocardiography in Non-Shielded Environments
Zhu, K., Shah, A. M., Berkow, J., & Kiourti, A. (2021) · IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology · doi:10.1109/jerm.2020.3019891 · Cited by 22
Park, J., Leithäuser, B., Hill, P., & Jung, F. (2008) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2008.00217.x · Cited by 21
Measurements of transmembrane potential and magnetic field at the apex of the heart
McBride, K. K., Roth, B. J., Sidorov, V. Y., Wikswo, J. P., & Baudenbacher, F. J. (2010) · Biophysical Journal · doi:10.1016/j.bpj.2010.08.040 · Cited by 21 · cited 2×
“In fact, when the ECG and cardiac enzyme patterns cannot yet be diagnostic, the higher sensitivity of MCG in detecting myocardial ischemia at an early stage is only one of the advantages arising by the more comprehensive biomagnetic electrophysiological assessment, which has the potential to non-invasively detect elec…”
BackgroundMethodMeloni, A. M., Fenici, R., & Brisinda, D. (2004) · Basic Research in Cardiology · doi:10.1007/s00395-003-0453-4 · Cited by 21
Self-citationMagnetocardiographic Localization of the Origin of Ventricular Ectopic Beats
OEFF, M., & BURGHOFF, M. (1994) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1994.tb01420.x · Cited by 21
Forty Years of Magnetocardiology
Stroink, G. (2010) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-12197-5_1 · Cited by 21
Swain, P. P., Sengottuvel, S., Patel, R., Mani, A., & Gireesan, K. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101664 · Cited by 21
Clinical Validation of Machine Learning for Automatic Analysis of Multichannel Magnetocardiography
Fenici, R., Brisinda, D., Meloni, A. M., Sternickel, K., & Fenici, P. (2005) · Lecture Notes in Computer Science · doi:10.1007/11494621_15 · Cited by 20
Self-citationSteinhoff, U., Knappe-Grüneberg, S., Schnabel, A., Trahms, L., Smith, F. E., Langley, P., Murray, A., & Koch, H. C. (2004) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2004.08.055 · Cited by 20
Magnetocardiography: clinical investigations with a biomagnetic multichannel system
Achenbach, S., & Moshage, W. (1993) · Physiological Measurement · doi:10.1088/0967-3334/14/4a/011 · Cited by 20
Shin, E.-S., Chung, J.-H., Park, S. G., Saleh, A., Lam, Y.-Y., Bhak, J., Jung, F., Morita, S., & Brachmann, J. (2019) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-180485 · Cited by 20
Compact SQUID realized in a double-layer graphene heterostructure
Indolese, D. I., Karnatak, P., Kononov, A., Delagrange, R., Haller, R., Wang, L., Makk, P., Watanabe, K., Taniguchi, T., & Schönenberger, C. (2020) · Nano Letters · doi:10.1021/acs.nanolett.0c02412 · Cited by 20
Sato, Y., Yoshida, K., Ogata, K., Inaba, T., Tada, H., Sekiguchi, Y., Ito, Y., Ishizu, T., Seo, Y., Yamaguchi, I., et al. (2012) · Circulation Journal · doi:10.1253/circj.cj-11-1419 · Cited by 19
FENICI, R., PESOLA, K., MÄKIJÄRVI, M., NENONEN, J., TEENER, U., FENICI, P., & KATILA, T. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb01206.x · Cited by 19
Self-citationChaikovsky, I., Hailer, B., Sosnytskyy, V., Lutay, M., Mjasnikov, G., Kazmirchuk, A., Bydnyk, M., Lomakovskyy, A., & Sosnytskaja, T. (2014) · Coronary Artery Disease · doi:10.1097/mca.0000000000000107 · Cited by 19
Ghasemi-Roudsari, S., Al-Shimary, A., Varcoe, B., Byrom, R., Kearney, L., & Kearney, M. (2018) · PLoS ONE · doi:10.1371/journal.pone.0191241 · Cited by 19
A high performance active noise control system for magnetic fields
Pyragius, T., & Jensen, K. (2021) · Review of Scientific Instruments · doi:10.1063/5.0062650 · Cited by 19
Heidecker, B. (2023) · European Heart Journal · doi:10.1093/eurheartj/ehad213 · Cited by 19 · cited 2×
“Frontiers in Cardiovascular Medicine 13 capability in numerous clinical fields, such as but not limited to the earlier detection of (sometimes electrically silent) myocardial ischemia (59, 88, 90, 323, 319), the preventive arrhythmogenic risk assessment in adults (53, 225, 325–322) as well as in fetuses (83, 323, 324)…”
BackgroundBody surface mapping and cardiac electric sources
TACCARDI, B. (1990) · Journal of Electrocardiology · doi:10.1016/0022-0736(90)90091-f · Cited by 18
Wu, Y.-W., Lee, C.-M., Liu, Y.-B., Wang, S.-S., Huang, H.-C., Tseng, W.-K., Jui, H.-Y., Wang, S.-Y., Horng, H.-E., Yang, H.-C., et al. (2013) · Circulation Journal · doi:10.1253/circj.cj-12-1170 · Cited by 18 · cited 1×
“The parallel research efforts carried out with shielded and unshielded MCG over the last three decades have enlarged the knowledge about the pros and cons of these two approaches and provided evidence of well-defined fields of clinical application, such as the emergency triage of patients with chest pain, the diagnosi…”
BackgroundBiomagnetic Noninvasive Localization of Accessory Pathways in Wolff‐Parkinson‐White Syndrome
WEISMÜLLER, P., Abraham‐Fuchs, K., SCHNEIDER, S., RICHTER, P., Kochs, M., EDRICH, J., & Hombach, V. (1991) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1991.tb02798.x · Cited by 18
The Value of Magnetocardiography in the Course of Coronary Intervention
Hailer, B., Van Leeuwen, P., Chaikovsky, I., Auth‐Eisernitz, S., Schäfer, H., & Grönemeyer, D. (2005) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2005.05625.x · Cited by 18
Multichannel mapping of fetal magnetocardiogram in an unshielded hospital setting
Brisinda, D., Comani, S., Meloni, A. M., Alleva, G., Mantini, D., & Fenici, R. (2005) · Prenatal Diagnosis · doi:10.1002/pd.1160 · Cited by 18
Self-citationNoninvasive Study of Ventricular Preexcitation Using Multichannel Magnetocardiography
Fenici, R., Brisinda, D., Nenonen, J., & Fenici, P. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.00064.x · Cited by 18
Self-citationThe magnetocardiogram—a new approach to the fields surrounding the heart
Stratbucker, R. A., Hyde, C. M., & Wixson, S. E. (1963) · IRE Transactions on Bio-Medical Electronics · doi:10.1109/tbmel.1963.4322823 · Cited by 18 · cited 1×
“These attempts were finally achieved in the early 1960s, experimentally by recording the cardiac MF from an isolated rabbit heart preparation with a toroidal solenoid (8) and in humans by Baule and McFee (9), using two large magnetic sensor coils, although at the very low spatial resolution, which obviously made it st…”
MethodSorbo, A. R., Lombardi, G., La Brocca, L., Guida, G., Fenici, R., & Brisinda, D. (2018) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12526 · Cited by 18
Self-citationAssessing heart disease using a novel magnetocardiography device
Beadle, R., McDonnell, D., Ghasemi-Roudsari, S., Unitt, L., Parker, S. J., & Varcoe, B. T. H. (2021) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/abe5c5 · Cited by 18
Fenici, R., Romani, G. L., Barbanera, S., Zeppilli, P., Carelli, P., & Modena, I. (1980) · PubMed · Cited by 17
Self-citationQuantification of cardiac magnetic field orientation during ventricular de- and repolarization
Leeuwen, P. V., Hailer, B., Lange, S., Klein, A., Geue, D., Seybold, K., Poplutz, C., & Grönemeyer, D. H. W. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/9/006 · Cited by 17
Clinical magnetocardiography: experience with a biomagnetic multichannel system
Moshage, W., Achenbach, S., Weikl, A., Göhl, K., Bachmann, K., Abraham-Fuchs, K., Härer, W., & Schneider, S. (1991) · The International Journal of Cardiac Imaging · doi:10.1007/bf01797754 · Cited by 17
Bridging noninvasive and interventional electroanatomical imaging: role of magnetocardiography
Fenici, R., & Brisinda, D. (2007) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2006.10.032 · Cited by 17
Lindseth, B. J., Schwindt, P. D. D., Kitching, J. E., Fischer, D., & Shusterman, V. (2007) · Computers in cardiology · doi:10.1109/cic.2007.4745517 · Cited by 17
Optimal magnetic sensor vests for cardiac source imaging
Lau, S., Petković, B., & Haueisen, J. (2016) · Sensors · doi:10.3390/s16060754 · Cited by 17
Bang, W.-D., Kim, K., Lee, Y.-H., Kwon, H., Park, Y., Pak, H.-N., Ko, Y.-G., Lee, M., & Joung, B. (2016) · Yonsei Medical Journal · doi:10.3349/ymj.2016.57.6.1339 · Cited by 17
Gapelyuk, A., Schirdewan, A., Fischer, R., & Wessel, N. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/10/004 · Cited by 16
Interatrial conduction in patients with paroxysmal atrial fibrillation and in healthy subjects
Jurkko, R., Mäntynen, V., Lehto, M., Tapanainen, J. M., Montonen, J., Parikka, H. J., & Toivonen, L. K. (2009) · International Journal of Cardiology · doi:10.1016/j.ijcard.2009.05.064 · Cited by 16
Thompson, M. D., Ben Shalom, M., Geim, A. K., Matthews, A. J., White, J., Melhem, Z., Pashkin, Y. A., Haley, R. P., & Prance, J. R. (2017) · Applied Physics Letters · doi:10.1063/1.4981904 · Cited by 16
Roth (2023) · doi:10.3390/s23094215 · Cited by 16
Noninvasive Recording of His Potential Using Magnetocardiograms
Yamada, S., Kuga, K., On, K., & Yamaguchi, I. (2003) · Circulation Journal · doi:10.1253/circj.67.622 · Cited by 15
Uchida, S., Iramina, K., Goto, K., & Ueno, S. (2000) · Journal of Applied Physics · doi:10.1063/1.372655 · Cited by 15
Unshielded fetal magnetocardiography system using two-dimensional gradiometers
Seki, Y., Kandori, A., Kumagai, Y., Ohnuma, M., Ishiyama, A., Ishii, T., Nakamura, Y., Horigome, H., & Chiba, T. (2008) · Review of Scientific Instruments · doi:10.1063/1.2897588 · Cited by 15
Liao, Y., He, C., & Guo, Q. (2018) · Symmetry · doi:10.3390/sym10070269 · Cited by 15
Schirdewan, A., Gapelyuk, A., Fischer, R., Koch, L., Schütt, H., Zacharzowsky, U. B., Dietz, R., Thierfelder, L., & Wessel, N. (2007) · Chaos An Interdisciplinary Journal of Nonlinear Science · doi:10.1063/1.2432059 · Cited by 14
Averaging algorithm based on data statistics in magnetocardiography.
Kim, K., Yh, L., Kwon, H., Jm, K., Is, K., & Yk, P. (2004) · PubMed · Cited by 14
Development of Active Shielding-Type MI Gradiometer and Application for Magnetocardiography
Takiya, T., & Uchiyama, T. (2017) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2017.2726111 · Cited by 14
Research works printed but not linked: 12
- Cohen, 1970, Magnetocardiograms taken inside a shielded room with a superconducting point-contact magnetometer, Appl Phys Lett, 16, 80, 10.1063/1.1653195 · DOI · Paper page
- Stroink, 1989, Principles of cardiomagnetism, 47, 10.1007/978-1-4613-0581-1_4 · DOI · Paper page
- Stroink, 1998, Magnetism in medicine: a handbook, chapter cardiomagnetism
- Siltanen, 1989, Magnetocardiography, 1405
- Erne, 1985, Modelling of the His-Purkinje heart conduction system, 126
- Fenici, 1985, First simultaneous magnetocardiographic and invasive recordings of the PR interval electrophysiological phenomena in man, Med Biol Eng Comput, 23, 1483
- Dössel, 2000, Inverse problem of electro- and magnetocardiography: review and recent progress, Int J Bioelectromagn, 2, 262
- Fenici, 1989, Biomagnetic imaging in the cardiac catheterization laboratory, 409, 10.1007/978-1-4613-0581-1_87 · DOI
- Pesola, 1997, Comparison of regularization methods when applied to epicardial minimum norm estimates, Biomed Tech, 42, 273
- Fenici, 2001, First MCG multichannel instrumentation operative in an unshielded hospital laboratory for multimodal cardiac electrophysiology. Preliminary experience, Biomed Tech, 2019
- Park, 2005, Magnetocardiography predicts coronary artery disease in patients with acute chest pain, Ann Noninvasive Electrocardiol, 10, 312, 10.1111/j.1542-474X.2005.00634.x · DOI · Paper page
- Petrenko, 2022, Towards a practical implementation of a single-beam all-optical non-zero-field magnetic sensor for magnetoencephalographic complexes, Sensors, 22, 9862, 10.3390/s22249862 · DOI · Paper page
Cited by
Spintronic devices for biomedical applications
Mostufa, S., Liang, S. A., Chugh, V. K., Wang, J., & Wu, K. (2024) · npj Spintronics · doi:10.1038/s44306-024-00031-6 · Cited by 68
“It necessitates operation within a magnetically shielded room and at low temperatures 95 .”
Design and Analysis of a High-Sensitivity Atomic Magnetic Gradiometer With Adjustable Baseline
Zhang, Y., Tang, J., Xu, X., Wang, Z., & Zhai, Y. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3544375 · Cited by 27
“Conversely, in magnetocardiography (MCG), the magnetic field of the heart is typically modeled as a current dipole with p values ranging from 1.8 to 2.2 [34], [35].”
InfluentialAtrial Cardiomyopathy in Atrial Fibrillation: A Multimodal Diagnostic Framework
Karakasis, P., Vlachakis, P. K., Theofilis, P., Ktenopoulos, N., Patoulias, D., Fyntanidou, B., Antoniadis, A. P., & Fragakis, N. (2025) · Diagnostics · doi:10.3390/diagnostics15101207 · Cited by 33
“In addition to catheter-based electroanatomic mapping, emerging non-invasive current-based techniques such as magnetocardiography (MCG) offer novel avenues for atrial substrate assessment [77].”
Roth, B. J. (2024) · Biophysics Reviews · doi:10.1063/5.0201950 · Cited by 24
Harnessing the Heart’s Magnetic Field for Advanced Diagnostic Techniques
Elfouly, T., & Alouani, A. (2024) · Sensors · doi:10.3390/s24186017 · Cited by 19
Metadata
- Authors
- Brisinda, D., Fenici, P., & Fenici, R.
- Affiliations
- Università Cattolica del Sacro Cuore (IT)
- Corresponding author
- Fenici, R.
- Venue
- Frontiers in Cardiovascular Medicine
- Publisher
- Frontiers Media SA
- Published
- 2023-08-10
- Volume
- 10
- Pages
- 1232882-1232882
- ISSN
- 2297-055X (ISSN-L)
- PMC
- PMC10448194
- OpenAlex
- W4385724211
- Semantic Scholar
- 7dc43883741b2cd8ebeae3d2cd7cac34a798ade9
- Cited by
- 80
- References
- 339
- Citations by year
- 2026: 30 · 2025: 30 · 2024: 18 · 2023: 2
- FWCI
- 3.91
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Frontiers in Cardiovascular Medicine (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- PubliCatt (Università Cattolica del Sacro Cuore) (submittedVersion, cc-by)
- Funders
- Università degli Studi di Parma
- Università Cattolica del Sacro Cuore
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Keywords
- magnetocardiography, magnetic shielding, signal-to-noise ratio, cardiac electrophysiology, wearable sensors, clinical adoption
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.