Abstract
Implantable electronics address therapeutical needs of patients with electrical signaling dysfunctions such as heart problems, neurological disorders or hearing impairments. While standard electronics are rigid, planar and made of hard materials, their surrounding biological tissues are soft, wet and constantly in motion. These intrinsic differences in mechanical and chemical properties cause physiological responses that constitute a fundamental challenge to create functional long-term interfaces. Using soft and stretchable materials for electronic implants decreases the mechanical mismatch between implant and biological tissues. As a result, tissue damage during and after implantation is reduced, leading not only to an attenuated foreign body response, but also enabling completely novel applications. However, but for a few exceptions, soft materials are not sufficient to create long-term stable functional implants. In this work, we review recent progress in interfacing both the central (CNS) and peripheral nervous system (PNS) for long-term functional devices. The basics of soft and stretchable devices are introduced by highlighting the importance of minimizing physical as well as mechanical mismatch between tissue and implant in the CNS and emphasizing the relevance of an appropriate surface chemistry for implants in the PNS. Finally, we report on the latest materials and techniques that provide further electronic enhancements while reducing the foreign body reaction. Thus, this review should serve as a guide for creating long-term functional implants to enable future healthcare technologies and as a discussion on current ideas and progress within the field.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Foreign body reaction to biomaterials
Anderson, J. M., Rodriguez, A., & Chang, D. T. (2008) · Seminars in Immunology · doi:10.1016/j.smim.2007.11.004 · Cited by 5,138 · cited 3×
“A more detailed discussion of the immunological response to non-CNS implants can be found elsewhere [15, 31].”
BackgroundInfluentialDAMPs, PAMPs and alarmins: all we need to know about danger
Bianchi, M. E. (2007) · Journal of Leukocyte Biology · doi:10.1189/jlb.0306164 · Cited by 2,794 · cited 1×
“Outside of the CNS, danger signals in the form of damage associated molecular patterns released from the tissue immediately activate a coagulation cascade and the complement system as schematically depicted in figure 1 [24, 25].”
BackgroundOn the mechanisms of biocompatibility
Williams, D. F. (2008) · Biomaterials · doi:10.1016/j.biomaterials.2008.04.023 · Cited by 2,772
Inflammation in Wound Repair: Molecular and Cellular Mechanisms
Eming, S. A., Krieg, T., & Davidson, J. M. (2007) · Journal of Investigative Dermatology · doi:10.1038/sj.jid.5700701 · Cited by 2,429 · cited 1×
“Resolution of the inflammation through the release of anti-inflammatory mediators, cleavage of chemokines, down-regulation of inflammatory mediators and receptors, as well as apoptosis of immune cells are prerequisites for successful tissue repair [30].”
BackgroundNeural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243
Engineered Nanoparticles for Drug Delivery in Cancer Therapy
Sun, T., Zhang, Y. S., Pang, B., Hyun, D. C., Yang, M., & Xia, Y. (2014) · Angewandte Chemie International Edition · doi:10.1002/anie.201403036 · Cited by 2,164 · cited 1×
“Drug delivery systems often contain either a reservoir from which a drug is released diffusively or consist of a degradable material that continuously releases the drug via material erosion [139].”
BackgroundResponse of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · cited 2×
“There are a number of excellent review papers discussing these processes in more detail [67, 69, 70].”
BackgroundThe rise of plastic bioelectronics
Someya, T., Bao, Z., & Malliaras, G. G. (2016) · Nature · doi:10.1038/nature21004 · Cited by 1,884
A Survey of Structure−Property Relationships of Surfaces that Resist the Adsorption of Protein
Ostuni, E., Chapman, R. G., Holmlin, R. E., Takayama, S., & Whitesides, G. M. (2001) · Langmuir · doi:10.1021/la010384m · Cited by 1,799 · cited 1×
“With few exceptions, typical properties of such materials are hydrophilicity, neutral charge, hydrogen bond acceptors and lack of hydrogen bond donors [128, 129].”
BackgroundGorbet, M. B., & Sefton, M. V. (2004) · Biomaterials · doi:10.1016/j.biomaterials.2004.01.023 · Cited by 1,347 · cited 1×
“Outside of the CNS, danger signals in the form of damage associated molecular patterns released from the tissue immediately activate a coagulation cascade and the complement system as schematically depicted in figure 1 [24, 25].”
BackgroundAstrocyte barriers to neurotoxic inflammation
Sofroniew, M. V. (2015) · Nature Reviews Neuroscience · doi:10.1038/nrn3898 · Cited by 1,202 · cited 1×
“There are a number of excellent review papers discussing these processes in more detail [67, 69, 70].”
BackgroundElectronic dura mater for long-term multimodal neural interfaces
Minev, I. R., Musienko, P., Hirsch, A., Barraud, Q., Wenger, N., Moraud, E. M., Gandar, J., Capogrosso, M., Milekovic, T., Asboth, L., et al. (2015) · Science · doi:10.1126/science.1260318 · Cited by 1,082
Self-citationZwitterionic hydrogels implanted in mice resist the foreign-body reaction
Zhang, L., Cao, Z., Bai, T., Carr, L., Ella-Menye, J.-R., Irvin, C., Ratner, B. D., & Jiang, S. (2013) · Nature Biotechnology · doi:10.1038/nbt.2580 · Cited by 1,035 · cited 1×
“In a recent breakthrough, Zhang et al could demonstrate that robust lowfouling zwitterionic hydrogels resisted capsule formation and promoted angiogenesis in mice [134].”
BackgroundNeuroGrid: recording action potentials from the surface of the brain
Khodagholy, D., Gelinas, J. N., Thesen, T., Doyle, W., Devinsky, O., Malliaras, G. G., & Buzsáki, G. (2015) · Nature Neuroscience · doi:10.1038/nn.3905 · Cited by 938
Brown, B. N., Ratner, B. D., Goodman, S. B., Amar, S., & Badylak, S. F. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.02.034 · Cited by 926
Veiseh, O., Doloff, J. C., Ma, M., Vegas, A. J., Tam, H. H., Bader, A. R., Li, J., Langan, E., Wyckoff, J., Loo, W. S., et al. (2015) · Nature Materials · doi:10.1038/nmat4290 · Cited by 897 · cited 1×
“5 mm significantly suppressed fibrosis compared to smaller spheres or to different geometries independent of the tested materials [125].”
BackgroundTranslational principles of deep brain stimulation
Kringelbach, M. L., Jenkinson, N., Owen, S. L. F., & Aziz, T. Z. (2007) · Nature Reviews Neuroscience · doi:10.1038/nrn2196 · Cited by 853 · cited 1×
“These insights have already led to the development of new devices especially for 1D wire electrodes, schematically shown in figure 2(C)(c), used to access deep brain regions for the treatment of various neurological disorders [91, 92].”
BackgroundMacrophage activation switching: an asset for the resolution of inflammation
Porcheray, F., Viaud, S., Rimaniol, A.-C., Léone, C., Samah, B., Dereuddre-Bosquet, N., Dormont, D., & Gras, G. (2005) · Clinical and Experimental Immunology · doi:10.1111/j.1365-2249.2005.02934.x · Cited by 840 · cited 1×
“Here, two different subtypes of macrophages play a pivotal role during inflammation, the mainly pro inflammatory M1 macrophages and the anti-inflammatory M2 macrophages [143, 144].”
BackgroundMesenchymal stem cell–educated macrophages: A novel type of alternatively activated macrophages
Kim, J., & Hematti, P. (2009) · Experimental Hematology · doi:10.1016/j.exphem.2009.09.004 · Cited by 785
Spiller, K. L., Nassiri, S., Witherel, C. E., Anfang, R. R., Ng, J., Nakazawa, K. R., Yu, T., & Vunjak-Novakovic, G. (2015) · Biomaterials · doi:10.1016/j.biomaterials.2014.10.017 · Cited by 773
What is immune privilege (not)?
Galea, I., Bechmann, I., & Perry, V. H. (2007) · Trends in Immunology · doi:10.1016/j.it.2006.11.004 · Cited by 768 · cited 1×
“In the CNS, an immune privileged site with restricted access of blood-borne antibodies and leukocytes, implants are primarily encapsulated with cells [61].”
BackgroundMaterials and technologies for soft implantable neuroprostheses
Lacour, S. P., Courtine, G., & Guck, J. (2016) · Nature Reviews Materials · doi:10.1038/natrevmats.2016.63 · Cited by 738
Kozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721 · cited 1×
“Similar devices also showed reduced inflammation and were capable of recording neural activity over an extended time window of more than seven weeks when compared to traditional electrodes, losing their function halfway [95].”
BackgroundCanales, A., Jia, X., Froriep, U. P., Koppes, R. A., Tringides, C. M., Selvidge, J., Lu, C., Hou, C., Wei, L., Fink, Y., et al. (2015) · Nature Biotechnology · doi:10.1038/nbt.3093 · Cited by 668 · cited 1×
“Canales et al combined simultaneous neural recording, optogenetic stimulation and drug delivery integrated into a single, flexible polymer fiber that generated a decreased glial response to the implant [93].”
BackgroundGlial responses to implanted electrodes in the brain
Salatino, J. W., Ludwig, K. A., Kozai, T. D. Y., & Purcell, E. K. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0154-1 · Cited by 642
Brain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies
Kozai, T. D. Y., Jaquins-Gerstl, A. S., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn500256e · Cited by 622 · cited 1×
“Within the first 24 h post implantation, the device is already encapsulated by a thin sheath of microglial cells [64].”
BackgroundFailure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066014 · Cited by 600
Tissue–electronics interfaces: from implantable devices to engineered tissues
Feiner, R., & Dvir, T. (2017) · Nature Reviews Materials · doi:10.1038/natrevmats.2017.76 · Cited by 595 · cited 1×
“Such cell-implant hybrids as seen in figure 4(D), created by culturing patient-specific cells on a medical device in vitro, are an upcoming route to improve tissue formation and integration of implants [151].”
BackgroundHost response to tissue engineered devices
Babensee, J. E., Anderson, J. M., McIntire, L. V., & Mikos, A. G. (1998) · Advanced Drug Delivery Reviews · doi:10.1016/s0169-409x(98)00023-4 · Cited by 585
Vegas, A. J., Veiseh, O., Doloff, J. C., Ma, M., Tam, H. H., Bratlie, K., Li, J., Bader, A. R., Langan, E., Olejnik, K., et al. (2016) · Nature Biotechnology · doi:10.1038/nbt.3462 · Cited by 551
Implanted Neural Interfaces: Biochallenges and Engineered Solutions
Grill, W. M., Norman, S. E., & Bellamkonda, R. V. (2009) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-061008-124927 · Cited by 488
McConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480
A transverse intrafascicular multichannel electrode (TIME) to interface with the peripheral nerve
Boretius, T., Badia, J., Pascual-Font, A., Schuettler, M., Navarro, X., Yoshida, K., & Stieglitz, T. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.05.010 · Cited by 468 · cited 1×
“Due to the success of flexible intraneural electrodes, human clinical trials using TIMEs are underway mainly focusing on the human machine interface as well as phantom limb pain [50, 51].”
BackgroundNeural probe design for reduced tissue encapsulation in CNS
Seymour, J. P., & Kipke, D. R. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.024 · Cited by 462 · cited 1×
“10 devices resulted in a reduced tissue reaction compared to larger diameter devices [76, 77].”
BackgroundMolecular basis of biomaterial-mediated foreign body reactions
Hu, W.-J., Eaton, J. W., Ugarova, T. P., & Tang, L. (2001) · Blood · doi:10.1182/blood.v98.4.1231 · Cited by 431
Luk, Y.-Y., Kato, M., & Mrksich, M. (2000) · Langmuir · doi:10.1021/la0004653 · Cited by 414
Three-dimensional macroporous nanoelectronic networks as minimally invasive brain probes
Xie, C., Liu, J., Fu, T.-M., Dai, X., Zhou, W., & Lieber, C. M. (2015) · Nature Materials · doi:10.1038/nmat4427 · Cited by 404
Electrical brain stimulation for epilepsy
Fisher, R. S., & Velasco, A. L. (2014) · Nature Reviews Neurology · doi:10.1038/nrneurol.2014.59 · Cited by 402
A MEMS-based flexible multichannel ECoG-electrode array
Rubehn, B., Bosman, C., Oostenveld, R., Fries, P., & Stieglitz, T. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/3/036003 · Cited by 387
Soft Materials in Neuroengineering for Hard Problems in Neuroscience
Jeong, J.-W., Shin, G., Park, S. I., Yu, K. J., Xu, L., & Rogers, J. A. (2015) · Neuron · doi:10.1016/j.neuron.2014.12.035 · Cited by 333
Long-term stimulation and recording with a penetrating microelectrode array in cat sciatic nerve
Branner, A., Stein, R. B., Fernandez, E., Aoyagi, Y., & Normann, R. A. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.820321 · Cited by 301
High‐Density Stretchable Electrode Grids for Chronic Neural Recording
Tybrandt, K., Khodagholy, D., Dielacher, B., Stauffer, F., Renz, A. F., Buzsáki, G., & Vörös, J. (2018) · Advanced Materials · doi:10.1002/adma.201706520 · Cited by 286
Self-citationMechanically-compliant intracortical implants reduce the neuroinflammatory response
Nguyen, J. K., Park, D. J., Skousen, J. L., Hess-Dunning, A. E., Tyler, D. J., Rowan, S. J., Weder, C., & Capadona, J. R. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/5/056014 · Cited by 274
Engineering biomaterials to integrate and heal: The biocompatibility paradigm shifts
Bryers, J. D., Giachelli, C. M., & Ratner, B. D. (2012) · Biotechnology and Bioengineering · doi:10.1002/bit.24559 · Cited by 271 · cited 1×
“Additionally, microporous substrates seem to promote healing independently of material or implant site [124].”
BackgroundZhou, T., Hong, G., Fu, T.-M., Yang, X., Schuhmann, T. G., Viveros, R. D., & Lieber, C. M. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1705509114 · Cited by 227
Lago, N., Yoshida, K., Koch, K. P., & Navarro, X. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.886617 · Cited by 227
McNally, A. K., & Anderson, J. M. (2002) · The American Journal of Pathology · doi:10.1016/s0002-9440(10)64882-1 · Cited by 225
Kozai, T. D. Y., Vazquez, A. L., Weaver, C. L., Kim, S.-G., & Cui, X. T. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/6/066001 · Cited by 215
Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants
Prasad, A., Xue, Q.-S., Dieme, R., Sankar, V., Mayrand, R. C., Nishida, T., Streit, W. J., & Sanchez, J. C. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00002 · Cited by 178
Borton, D., Micera, S., Millán, J. D. R., & Courtine, G. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3005968 · Cited by 171
Sensory nerve recording for closed‐loop control to restore motor functions
Popovic, D. B., Stein, R. B., Jovanovic, K. L., Dai, R., Kostov, A., & Armstrong, W. W. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.247801 · Cited by 159
Biomaterial–microvasculature interactions
Sieminski, A. L., & Gooch, K. J. (2000) · Biomaterials · doi:10.1016/s0142-9612(00)00149-6 · Cited by 155
Ware, T., Simon, D., Hearon, K., Liu, C., Shah, S., Reeder, J., Khodaparast, N., Kilgard, M. P., Maitland, D. J., Rennaker, R. L., et al. (2012) · Macromolecular Materials and Engineering · doi:10.1002/mame.201200241 · Cited by 138
The foreign body response to the Utah Slant Electrode Array in the cat sciatic nerve
Christensen, M. B., Pearce, S. M., Ledbetter, N. M., Warren, D. J., Clark, G. A., & Tresco, P. A. (2014) · Acta Biomaterialia · doi:10.1016/j.actbio.2014.07.010 · Cited by 135
A Microchannel Neuroprosthesis for Bladder Control After Spinal Cord Injury in Rat
Chew, D. J., Zhu, L., Delivopoulos, E., Minev, I. R., Musick, K. M., Mosse, C. A., Craggs, M., Donaldson, N., Lacour, S. P., McMahon, S. B., et al. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3007186 · Cited by 123
Xiang, Z., Yen, S., Sheshadri, S., Wang, J., Lee, S., Liu, Y., Liao, L., Thakor, N. V., & Lee, C. (2016) · Advanced Materials · doi:10.1002/adma.201503423 · Cited by 116
Badia, J., Boretius, T., Pascual-Font, A., Udina, E., Stieglitz, T., & Navarro, X. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2153850 · Cited by 103
Flexible and highly biocompatible nanofiber-based electrodes for neural surface interfacing
Heo, D. N., Kim, H.-J., Lee, Y. J., Heo, M., Lee, S. J., Lee, D., Do, S. H., Lee, S. H., & Kwon, I. K. (2017) · ACS Nano · doi:10.1021/acsnano.6b08390 · Cited by 86
A three-dimensional self-opening intraneural peripheral interface (SELINE)
Cutrone, A., Valle, J. D., Santos, D., Badia, J., Filippeschi, C., Micera, S., Navarro, X., & Bossi, S. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016016 · Cited by 82 · cited 1×
“These use a three-dimensional device geometry that enhances mechanical anchorage to nerves, reducing shifts in electrode position caused by tissue movement [49].”
MethodFitzGerald, J. J., Lago, N., Benmerah, S., Serra, J., Watling, C. P., Cameron, R. E., Tarte, E., Lacour, S. P., McMahon, S. B., & Fawcett, J. W. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/1/016010 · Cited by 73
Chronic multichannel neural recordings from soft regenerative microchannel electrodes during gait
Musick, K. M., Rigosa, J., Narasimhan, S., Wurth, S., Capogrosso, M., Chew, D. J., Fawcett, J. W., Micera, S., & Lacour, S. P. (2015) · Scientific Reports · doi:10.1038/srep14363 · Cited by 70
Microchannel-based regenerative scaffold for chronic peripheral nerve interfacing in amputees
Srinivasan, A., Tahilramani, M., Bentley, J. T., Gore, R. K., Millard, D. C., Mukhatyar, V. J., Joseph, A., Haque, A. S., Stanley, G. B., English, A. W., et al. (2015) · Biomaterials · doi:10.1016/j.biomaterials.2014.11.035 · Cited by 67
Modification of Biomaterials with a Self‐Protein Inhibits the Macrophage Response
Kim, Y. K., Que, R., Wang, S., & Liu, W. F. (2014) · Advanced Healthcare Materials · doi:10.1002/adhm.201300532 · Cited by 56
A Parylene Self-Locking Cuff Electrode for Peripheral Nerve Stimulation and Recording
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2014.2333733 · Cited by 56
Wark, H. A. C., Mathews, K. S., Normann, R. A., & Fernandez, E. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/4/046027 · Cited by 54
The design of and chronic tissue response to a composite nerve electrode with patterned stiffness
Freeberg, M. J., Stone, M. A., Triolo, R. J., & Tyler, D. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6632 · Cited by 54
Regenerated sciatic nerve axons stimulated through a chronically implanted macro-sieve electrode
MacEwan, M. R., Zellmer, E. R., Wheeler, J. J., Burton, H., & Moran, D. W. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00557 · Cited by 48
Biomechanical and functional variation in rat sciatic nerve following cuff electrode implantation
Restaino, S. M., Abliz, E., Wachrathit, K., Krauthamer, V., & Shah, S. B. (2014) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-11-73 · Cited by 44
Chronic histological effects of the flat interface nerve electrode
Leventhal, D. K., Cohen, M., & Durand, D. M. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/2/004 · Cited by 43
Fascicular nerve stimulation and recording using a novel double-aisle regenerative electrode
Delgado-Martínez, I., Righi, M., Santos, D., Cutrone, A., Bossi, S., D’Amico, S., Del Valle, J., Micera, S., & Navarro, X. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6bac · Cited by 35
Suppression of scarring in peripheral nerve implants by drug elution
FitzGerald, J. J. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/026006 · Cited by 27
Cited by
Stretchable Electronics Based on PDMS Substrates
Qi, D., Zhang, K., Tian, G., Jiang, B., & Huang, Y. (2021) · Advanced Materials · doi:10.1002/adma.202003155 · Cited by 663
Compliant peripheral nerve interfaces.
Paggi, V., Akouissi, O., Micera, S., & Lacour, S. P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abcdbe · Cited by 60
“[86] Renz A F, Reichmuth A M, Stauffer F, Thompson-Steckel G”
BackgroundStretchable gold nanowire-based cuff electrodes for low-voltage peripheral nerve stimulation
Lienemann, S., Zötterman, J., Farnebo, S., & Tybrandt, K. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abfebb · Cited by 48
Organic electronics for neuroprosthetics
Airaghi Leccardi, M. J. I., & Ghezzi, D. (2020) · Healthcare Technology Letters · doi:10.1049/htl.2019.0108 · Cited by 12
Metadata
- Authors
- Renz, A. F., Reichmuth, A. M., Stauffer, F., Thompson-Steckel, G., & Vörös, J.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2018-10-16
- Volume
- 15
- Issue
- 6
- Pages
- 061001
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2897988811
- Cited by
- 50
- References
- 71
- FWCI
- 2.40
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.