Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults
Open accessThe DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Abstract
Introduction With the increasing marketing of automated and semi–automated devices for the measurement of blood pressure, potential purchasers need to be able to satisfy themselves that such devices have been evaluated according to agreed criteria [1]. With this in mind, the Association for the Advancement of Medical Instrumentation (AAMI) published a standard for electronic or aneroid sphygmomanometers in 1987 [2] that included a protocol for evaluating the accuracy of devices, this being followed in 1990 by the protocol of the British Hypertension Society (BHS) [3]. Both of these were revised in 1993 [4,5]. These protocols, which differed in detail, had a common objective, namely the standardization of validation procedures to establish minimum standards of accuracy and performance, and to facilitate the comparison of one device with another [6]. Since their introduction, a large number of blood pressure measuring devices have been evaluated according to one or both protocols. Experience has, however, demonstrated that the conditions demanded by the protocols are extremely difficult to fulfil. This is especially so because of the large number of subjects who have to be recruited and the ranges of blood pressure required. The time required to complete a validation study is such that it is difficult to recruit trained staff for the duration of an investigation. These factors have made validation studies difficult to perform and very costly, with the result that fewer centers are prepared to undertake them. This is particularly unfortunate as more devices than ever before are in need of independent validation. When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood Pressure Monitoring of the European Society of Hypertension (ESH) undertook to produce an updated protocol, which it has named the International Protocol. The ESH Working Group on Blood Pressure Monitoring is composed of experts in blood pressure measurement, many of whom have considerable experience in validating blood pressure measuring devices. In setting about its objective, the ESH Working Group recognized the urgent imperative to provide a simplified protocol that does not sacrifice the integrity of the earlier protocols. When the AAMI and BHS protocols [2–5] were published, the relevant committees did not have evidence from previous studies on which to base their recommendations. The ESH Working Group has had the advantage of being able to examine and analyze the data from 19 validation studies performed according to the AAMI and BHS protocols at the Blood Pressure Unit in Dublin [7–23]. A critical assessment of this database of evidence has permitted a rationalization and simplification of validation procedures without losing the merits of the much more complicated earlier protocols. The basic recommendations of the simplified International Protocol have been presented at meetings of the ESH Working Group, and the proceedings of these meetings have been published in order to invite comment and discussion [24–27]. The International Protocol has been drafted in such a way as to be applicable to the majority of blood pressure measuring devices on the market. The validation procedure is therefore confined to adults over the age of 30 years (as these will constitute the majority of subjects with hypertension), and does not make recommendations for special groups, such as children, pregnant women and the elderly, or for special circumstances, for example exercise. It is anticipated that the relative ease of performance of the International Protocol will encourage manufacturers to submit blood pressure measuring devices for validation in order to obtain the minimum approval necessary for a device to be used in clinical medicine, and that, in time, most devices on the market will be assessed according to the protocol for basic accuracy. It does not preclude manufacturers of devices from subjecting their products to more rigorous assessment and validation. Validation procedure Summary The validation team should consist of four persons experienced in blood pressure measurement: two observers and a supervisor (generally nurses), and an ‘expert’ (a doctor overseeing the entire procedure). If the doctor can be present throughout the entire validation procedure, he/she can take over the role of supervisor, thereby reducing the number of personnel to three. The validation procedure consists of the following steps Observer training and assessment. Two observers are trained in accurate blood pressure measurement. Familiarization session. The validation team becomes familiar with the workings of the device and the accompanying software. Validation measurements. Observer and device measurements are recorded on subjects in two phases. In the first phase, 15 subjects are recruited; devices passing this primary phase proceed to the secondary phase, for which a further 18 subjects are recruited. Analysis. An analysis of
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Blood pressure measuring devices: recommendations of the European Society of Hypertension
O'Brien, E., Waeber, B., Parati, G., Staessen, J., & Myers, M. G. (2001) · BMJ · doi:10.1136/bmj.322.7285.531 · Cited by 903
Self-citationParati, G., Casadei, R., Groppelli, A., Di Rienzo, M., & Mancia, G. (1989) · Hypertension · doi:10.1161/01.hyp.13.6.647 · Cited by 899
Self-citationO??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772
Self-citationAccuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol
O??Brien, E., Mee, F., Atkins, N., & O??Malley, K. (1991) · Journal of Hypertension · doi:10.1097/00004872-199106000-00016 · Cited by 451
Self-citationO'Brien, E. T., Mee, F. I., Atkins, N., & Thomas (1996) · PubMed · Cited by 409
Self-citationAsmar, R., & Zanchetti, A. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018050-00001 · Cited by 213
Self-citationTwenty-four-hour ambulatory blood pressure measurement in a primigravid population
Halligan, A., O'Brien, E. T., O??Malley, K., Mee, F. I., Atkins, N., Conroy, R., Walshe, J. J., & Darling, M. (1993) · Journal of Hypertension · doi:10.1097/00004872-199308000-00014 · Cited by 147
Self-citationInaccuracy of the Hawksley random zero sphygmomanometer
O'Brien, E., Mee, F., Atkins, N., & O'MALLEY, K. (1990) · The Lancet · doi:10.1016/0140-6736(90)93177-q · Cited by 107
Self-citationBlood pressure measurement : recommendations of the British Hypertension Society
O'Brien, E. T., Petrie, J. C., Littler, W. A., de Swiet, M., Padfield, P., Dillon, M. J., Coats, A. J. S., & Mee, F. (1997) · OpenGrey (Institut de l'Information Scientifique et Technique) · Cited by 87
Self-citation
Cited by
National High Blood Pressure Education Program Working Group on High Blood Pressure in Children and Adolescents (2004) · PEDIATRICS · doi:10.1542/peds.114.2.s2.555 · Cited by 6,938
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Self-citationPickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
Self-citationThe Japanese Society of Hypertension Guidelines for the Management of Hypertension (JSH 2019)
Umemura, S., Arima, H., Arima, S., Asayama, K., Dohi, Y., Hirooka, Y., Horio, T., Hoshide, S., Ikeda, S., Ishimitsu, T., et al. (2019) · Hypertension Research · doi:10.1038/s41440-019-0284-9 · Cited by 1,880
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationEuropean Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
Self-citationLurbe, E., Agabiti-Rosei, E., Cruickshank, J. K., Dominiczak, A., Erdine, S., Hirth, A., Invitti, C., Litwin, M., Mancia, G., Pall, D., et al. (2016) · Journal of Hypertension · doi:10.1097/hjh.0000000000001039 · Cited by 1,379
“If an oscillometric method is applied, the monitor should have passed the validation procedure recommended by the British Hypertension Society [53], the American Association for the Advancement of Medical Instrumentation [54] or the ESH International Protocol [55].”
Williams, B., Poulter, N. R., Brown, M. J., Davis, M., McInnes, G. T., Potter, J. F., Sever, P. S., & McG Thom, S. (2004) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001683 · Cited by 1,272
Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837
Self-citationAmbulatory Blood-Pressure Monitoring
Pickering, T. G., Shimbo, D., & Haas, D. (2006) · New England Journal of Medicine · doi:10.1056/nejmra060433 · Cited by 778
Self-citationLurbe, E., Cifkova, R., Cruickshank, J. K., Dillon, M. J., Ferreira, I., Invitti, C., Kuznetsova, T., Laurent, S., Mancia, G., Morales-Olivas, F., et al. (2009) · Journal of Hypertension · doi:10.1097/hjh.0b013e32832f4f6b · Cited by 771
Pickering, T. G., Miller, N. H., Ogedegbe, G., Krakoff, L. R., Artinian, N. T., & Goff, D. (2008) · Hypertension · doi:10.1161/hypertensionaha.107.189010 · Cited by 668
Self-citationO'Brien, E., Atkins, N., Stergiou, G., Karpettas, N., Parati, G., Asmar, R., Imai, Y., Wang, J., Mengden, T., & Shennan, A. (2010) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283360e98 · Cited by 610
Self-citationVI Diretrizes Brasileiras de Hipertensão
(2010) · Arquivos Brasileiros de Cardiologia · doi:10.1590/s0066-782x2010001700001 · Cited by 596
European Society of Hypertension Practice Guidelines for home blood pressure monitoring
Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2010) · Journal of Human Hypertension · doi:10.1038/jhh.2010.54 · Cited by 482
Self-citationStergiou, G. S., Alpert, B., Mieke, S., Asmar, R., Atkins, N., Eckert, S., Frick, G., Friedman, B., Graßl, T., Ichikawa, T., et al. (2018) · Hypertension · doi:10.1161/hypertensionaha.117.10237 · Cited by 469
Self-citationStaessen, J. A., Wang, J., Bianchi, G., & Birkenhäger, W. H. (2003) · The Lancet · doi:10.1016/s0140-6736(03)13302-8 · Cited by 446
Self-citationPickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J. W., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · The Journal of Clinical Hypertension · doi:10.1111/j.1524-6175.2005.04377.x · Cited by 436
Self-citationA new oscillometric method for pulse wave analysis: comparison with a common tonometric method
Wassertheurer, S., Kropf, J., Weber, T., van der Giet, M., Baulmann, J., Ammer, M., Hametner, B., Mayer, C. C., Eber, B., & Magometschnigg, D. (2010) · Journal of Human Hypertension · doi:10.1038/jhh.2010.27 · Cited by 363
Rezk, C. C., Marrache, R. C. B., Tinucci, T., Mion, D., & Forjaz, C. L. M. (2006) · European Journal of Applied Physiology · doi:10.1007/s00421-006-0257-y · Cited by 349
Avolio, A. P., Butlin, M., & Walsh, A. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/1/r01 · Cited by 310
Jananese Society of Hypertension (JSH) Guidelines for Self-Monitoring of Blood Pressure at Home
IMAI, Y., OTSUKA, K., KAWANO, Y., SHIMADA, K., HAYASHI, H., TOCHIKUBO, O., MIYAKAWA, M., & FUKIYAMA, K. (2003) · Hypertension Research · doi:10.1291/hypres.26.771 · Cited by 309
Self-citationBlood Pressure Percentiles by Age and Height From Nonoverweight Children and Adolescents in Germany
Neuhauser, H. K., Thamm, M., Ellert, U., Hense, H. W., & Rosario, A. S. (2011) · Pediatrics · doi:10.1542/peds.2010-1290 · Cited by 281
Stergiou, G. S., Alpert, B., Mieke, S., Asmar, R., Atkins, N., Eckert, S., Frick, G., Friedman, B., Graßl, T., Ichikawa, T., et al. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001634 · Cited by 278
Self-citationAccuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277
Measuring Blood Pressure Accurately: New and Persistent Challenges
Jones, D. W., Appel, L. J., Sheps, S. G., Roccella, E. J., & Lenfant, C. (2003) · JAMA · doi:10.1001/jama.289.8.1027 · Cited by 264
Validation of the mobil-O-Graph: 24 h-blood pressure measurement device
Wei, W., Tölle, M., Zidek, W., & van der Giet, M. (2010) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e328338892f · Cited by 260
Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A. (2008) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3282cb57b6 · Cited by 259
Sharman, J. E., Avolio, A. P., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C. L., Boutouyrie, P., Chen, C.-H., Chowienczyk, P., Cockcroft, J. R., et al. (2017) · European Heart Journal · doi:10.1093/eurheartj/ehw632 · Cited by 231
Self-citationCuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226
El Assaad, M. A., Topouchian, J. A., & Asmar, R. G. (2003) · Blood Pressure Monitoring · doi:10.1097/00126097-200306000-00006 · Cited by 224
Self-citationNexfin Noninvasive Continuous Blood Pressure Validated Against Riva-Rocci/Korotkoff
Eeftinck Schattenkerk, D. W., van Lieshout, J. J., van den Meiracker, A. H., Wesseling, K. R., Blanc, S., Wieling, W., van Montfrans, G. A., Settels, J. J., Wesseling, K. H., & Westerhof, B. E. (2009) · American Journal of Hypertension · doi:10.1038/ajh.2008.368 · Cited by 222
Home Blood Pressure Measurement
Verberk, W. J., Kroon, A. A., Kessels, A. G. H., & de Leeuw, P. W. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.05.058 · Cited by 212
as members of the Japanese Society of Hypertension Committee for Guidelines for Self-monitoring of Blood Pressure at Home, Imai, Y., Kario, K., Shimada, K., Kawano, Y., Hasebe, N., Matsuura, H., Tsuchihashi, T., Ohkubo, T., Kuwajima, I., et al. (2012) · Hypertension Research · doi:10.1038/hr.2012.56 · Cited by 203
Self-citationWilkinson, I. B., McEniery, C. M., Schillaci, G., Boutouyrie, P., Segers, P., Donald, A., & Chowienczyk, P. J. (2010) · Artery Research · doi:10.1016/j.artres.2010.03.001 · Cited by 202
Oscillometric Blood Pressure Estimation: Past, Present, and Future
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolic, M., Rajan, S., & Batkin, I. (2015) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2015.2434215 · Cited by 196
Stergiou, G. S., Yiannes, N. G., & Rarra, V. C. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000209074.38331.16 · Cited by 192
Ambulatory Blood Pressure Measurement: What Is the International Consensus?
O’Brien, E., Parati, G., & Stergiou, G. (2013) · Hypertension · doi:10.1161/hypertensionaha.113.02148 · Cited by 183
Self-citationColeman, A., Freeman, P., Steel, S., & Shennan, A. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000189788.05736.5f · Cited by 179
Maitland, M. L., Kasza, K. E., Karrison, T., Moshier, K., Sit, L., Black, H. R., Undevia, S. D., Stadler, W. M., Elliott, W. J., & Ratain, M. J. (2009) · Clinical Cancer Research · doi:10.1158/1078-0432.ccr-09-0058 · Cited by 174
Khalid, S. G., Zhang, J., Chen, F., & Zheng, D. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/1548647 · Cited by 171
“To prevent the selection of an overfitted algorithm, the 10-fold cross-validation was used to test the overall measurement accuracy of the algorithms. results showed that the regression tree achieved better overall accuracy in terms of mean and SD of BP difference as required by the ISO [29].”
MethodWong, S.-N., Tz Sung, R. Y., & Leung, L. C.-K. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000209082.09623.b4 · Cited by 166
Asmar, R. (2011) · Vascular Health and Risk Management · doi:10.2147/vhrm.s27193 · Cited by 163
Self-citationHow to assess mean blood pressure properly at the brachial artery level
Bos, W. J., Verrij, E., Vincent, H. H., Westerhof, B. E., Parati, G., & van Montfrans, G. A. (2007) · Journal of Hypertension · doi:10.1097/hjh.0b013e32803fb621 · Cited by 161
Self-citationTopouchian, J. A., El Assaad, M. A., Orobinskaia, L. V., El Feghali, R. N., & Asmar, R. G. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000209078.17246.34 · Cited by 158
Self-citationSharman, J. E., O’Brien, E., Alpert, B., Schutte, A. E., Delles, C., Hecht Olsen, M., Asmar, R., Atkins, N., Barbosa, E., Calhoun, D., et al. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002246 · Cited by 139
“In 1999, the German Hypertension League developed another protocol [32] and in 2002 the European Society of Hypertension (ESH) Working Group on BP Monitoring developed their International Protocol, which they revised in 2010 [33,34].”
BackgroundSelf-citationPicone, D. S., Deshpande, R. A., Schultz, M. G., Fonseca, R., Campbell, N. R. C., Delles, C., Hecht Olsen, M., Schutte, A. E., Stergiou, G., Padwal, R., et al. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14719 · Cited by 118
Validation protocols for blood pressure measuring devices in the 21st century
Stergiou, G. S., Alpert, B. S., Mieke, S., Wang, J., & O'Brien, E. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13294 · Cited by 92
Self-citationImprovement of albuminuria after renal denervation
Ott, C., Mahfoud, F., Schmid, A., Ditting, T., Veelken, R., Ewen, S., Ukena, C., Uder, M., Böhm, M., & Schmieder, R. E. (2014) · International Journal of Cardiology · doi:10.1016/j.ijcard.2014.03.017 · Cited by 75
Cameron, J. D., Stevenson, I., Reed, E., McGrath, B. P., Dart, A. M., & Kingwell, B. A. (2004) · Blood Pressure Monitoring · doi:10.1097/00126097-200410000-00007 · Cited by 75
Stergiou, G. S., Lourida, P., Tzamouranis, D., & Baibas, N. M. (2009) · Journal of Human Hypertension · doi:10.1038/jhh.2009.20 · Cited by 74
Accuracy of the pressure scale of sphygmomanometers in clinical use within primary care
Coleman, A. J., Steel, S. D., Ashworth, M., Vowler, S. L., & Shennan, A. (2005) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000168398.87167.c2 · Cited by 64
Over-the-counter mouthwash use, nitric oxide and hypertension risk
Joshipura, K., Muñoz-Torres, F., Fernández-Santiago, J., Patel, R. P., & Lopez-Candales, A. (2020) · Blood Pressure · doi:10.1080/08037051.2019.1680270 · Cited by 58
“premature death and disability worldwide [1].”
Blood Pressure Response to Submaximal Exercise Test in Adults
Wielemborek-Musial, K., Szmigielska, K., Leszczynska, J., & Jegier, A. (2016) · BioMed Research International · doi:10.1155/2016/5607507 · Cited by 54
Stergiou, G. S., O'Brien, E., Myers, M., Palatini, P., Parati, G., Kollias, A., Birmpas, D., Kyriakoulis, K., Bountzona, I., Stambolliu, E., et al. (2019) · The Journal of Clinical Hypertension · doi:10.1111/jch.13710 · Cited by 53
Self-citationParati, G., Omboni, S., Compare, A., Grossi, E., Callus, E., Venco, A., Destro, M., Villa, G., Palatini, P., Rosei, E. A., et al. (2013) · Trials · doi:10.1186/1745-6215-14-22 · Cited by 49
Self-citationClark, C. E., McDonagh, S. T. J., & McManus, R. J. (2019) · Journal of Human Hypertension · doi:10.1038/s41371-018-0153-z · Cited by 49
“They reviewed eight studies of 11 devices, and observed that overall study methodology was variable and sample sizes were usually lower than those dictated for validation studies [11, 46].”
BackgroundThe pursuit of accurate blood pressure measurement: A 35‐year travail
O'Brien, E., & Stergiou, G. S. (2017) · The Journal of Clinical Hypertension · doi:10.1111/jch.13005 · Cited by 39
Self-citationDevelopment of an accurate oscillometric blood pressure device for low resource settings
de Greeff, A., Nathan, H., Stafford, N., Liu, B., & Shennan, A. H. (2008) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e32830fd07c · Cited by 38
Sánchez, R. A., Boggia, J., Peñaherrera, E., Barroso, W. S., Barbosa, E., Villar, R., Cobos, L., Hernández Hernández, R., Lopez, J., Octavio, J. A., et al. (2020) · The Journal of Clinical Hypertension · doi:10.1111/jch.13816 · Cited by 38
Self-citationBoonyasai, R. T., Carson, K. A., Marsteller, J. A., Dietz, K. B., Noronha, G. J., Hsu, Y., Flynn, S. J., Charleston, J. M., Prokopowicz, G. P., Miller, E. R., et al. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13166 · Cited by 34
Indirect measurement of systemic blood pressure in conscious dogs in a clinical setting.
HSIANG, T.-Y., LIEN, Y.-H., & HUANG, H.-P. (2008) · Journal of Veterinary Medical Science · doi:10.1292/jvms.70.449 · Cited by 29
Validation of the Stabil-O-Graph blood pressure self-measurement device
Westhoff, T. H., Schmidt, S., Zidek, W., & van der Giet, M. (2008) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1002287 · Cited by 26
Akpolat, T., Erdem, E., & Aydogdu, T. (2012) · Kidney and Blood Pressure Research · doi:10.1159/000330719 · Cited by 25
O’Brien, E., Stergiou, G., Palatini, P., Asmar, R., Ioannidis, J. P., Kollias, A., Lacy, P., McManus, R. J., Myers, M. G., Shennan, A., et al. (2019) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000391 · Cited by 23
“studies of blood pressure (BP) measuring devices using the European Society of Hypertension International Protocol (ESH-IP) [1,2] should no longer be accepted for publication in scientific journals [3].”
MethodSelf-citationAutomated blood pressure measuring devices: how are they clinically validated for accuracy?
Ringrose, J. S., & Padwal, R. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00761-2 · Cited by 9
“The ESH-IP protocol was designed to be a compromise between the need for clinical validation and feasibility (time, resources, and cost).”
MethodInfluential
Metadata
- Authors
- OʼBrien, E., Pickering, T., Asmar, R., Myers, M., Parati, G., Staessen, J., Mengden, T., Imai, Y., Waeber, B., Palatini, P., & Gerin, W.
- Venue
- Blood Pressure Monitoring
- Published
- 2002-02
- Volume
- 7
- Issue
- 1
- Pages
- 3-17
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2131288011
- Cited by
- 744
- References
- 26
- FWCI
- 19.72
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.