Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers
Open accessAbstract
In this review, we consider the ways in which vasopressin and oxytocin have been measured since their first discovery. Two different ways of measuring oxytocin in widespread use currently give values in human plasma that differ by two orders of magnitude, and the values measured by these two methods in the same samples show no correlation. The notion that we should accept this seems absurd. Either one (or both) methods is not measuring oxytocin, or , by ‘oxytocin’, the scientists that use these different methods mean something very different. If these communities are to talk to each other, it is important to validate one method and invalidate the other, or else to establish exactly what each community understands by ‘oxytocin’. A similar issue concerns vasopressin: again, different ways of measuring vasopressin give values in human plasma that differ by two orders of magnitude, and it appears that the same explanation for discrepant oxytocin measurements applies to discrepant vasopressin measurements. The first assays for oxytocin and vasopressin measured biological activity directly. When immunoassays were introduced, they encountered problems: high molecular weight factors in raw plasma interfered with the binding of antibodies to the hormones, leading to high and erroneous readings. When these interfering factors were removed by extraction of plasma samples, immunoassays gave measurements consistent with bioassays, with measures of turnover and with the sensitivity of target tissues to exogenous hormone. However, many recent papers use an enzyme‐linked immunoassay to measure plasma levels without extracting the samples. Like the first radioimmunassays of unextracted plasma, this generates impossibly high and wholly erroneous measurements.
Overview
AI snapshot
Written from the paper’s full text.
TL;DR
This review argues that widely used unextracted-plasma immunoassays for oxytocin and vasopressin generate impossibly high, erroneous values, while extracted bioassays and radioimmunoassays converge on much lower physiological concentrations. It calls for assay validation and warns that much recent oxytocin/vasopressin literature may be based on measurements no better than random numbers.
Key findings
- Two current methods for measuring oxytocin in human plasma differ by two orders of magnitude and show no correlation in the same samples; the same discrepancy applies to vasopressin.
- Many recent studies using unextracted plasma ELISA report oxytocin and vasopressin levels typically of 100–500 pg/ml, whereas bioassays and extracted radioimmunoassays indicate normal circulating levels are <10 pg/ml; normal plasma vasopressin in man is below 2 μU/ml (5 pg/ml), rising several-fold after 24 h dehydration.
- Comparisons of the same ELISA with and without extraction found no correlation: in seals, extracted samples gave approximately 8 pg/ml versus 300 pg/ml in unextracted heparin samples or 540 pg/ml in EDTA samples; in humans, extracted RIA gave 1.1 pg/ml while unextracted ELISA gave 97 pg/ml, and spiking unextracted samples exaggerated measurements 3.5-fold.
- The authors attribute the high values to plasma protein interference with antibody binding, producing a high erroneous baseline; they conclude such assays in unextracted plasma are “no more than a random number generator.”
Methods
- Narrative review of historical and current methods for measuring oxytocin and vasopressin, including bioassays, radioimmunoassays, ELISAs, LC-MS/MS, and copeptin as a vasopressin surrogate.
- Synthesizes published studies and historical standards, including comparisons of extracted versus unextracted plasma, assay validation studies, clearance and secretion estimates, and target-tissue sensitivity data.
- Main analysis is qualitative and quantitative plausibility checking rather than a systematic review or meta-analysis; it uses cited values to show that unextracted ELISA results are inconsistent with known production, storage, clearance, and bioactivity.
Limitations
- The paper states no formal limitations section.
- Evident from the text: it is a narrative review without a systematic search strategy or meta-analysis, and it presents no new primary experimental data.
- Evident from the text: its conclusions depend on the cited literature and on calculations from previously published values; the proposed validation recommendations are not prospectively tested here.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 of 122 cited sources checked · 0 supported · 0 flagged
None of the sources cited for this claim have been checked yet.
In the paper
“Indeed, none of the four studies cited in the Introduction 3, 4, 5, 6 as examples of the use of ELISAs to measure oxytocin and vasopressin in unextracted samples made any correction for multiple comparisons in their analyses.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Conclusions
What the sources say
- Not checked yetcites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source
- Not checked yetcites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source
- Not checked yetcites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source
- Not checked yetcites [6] Rubin LH, Carter CS, Bishop JR, Pournajafi‐Nazarloo H, Drogos LL, Hill SK, Ruocco AC, Keedy SK, Reilly JL, Keshavan MS, Pearlson GD, Tamminga CA, Gershon ES, Sweeney JA. Reduced levels of vasopressin and reduced behavioral modulation of oxytocin in psychotic disorders. Schizophr Bull 2014; 40: 1374– 1384. 24619535 PMC4193712 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesNone of the sources cited for this claim have been checked yet.
In the paper
“more than 100 papers have recently been published using assays that have been claimed to measure oxytocin and vasopressin in human plasma simply and reliably”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Introduction
What the sources say
- Not checked yetcites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source
- Not checked yetcites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source
- Not checked yetcites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIn the paper
“Measurements by radioimmunoassay reported that, during suckling in women, intermittent pulses were observed with a mean amplitude of just approximately 9 pg/ml from a basal level of approximately 2 pg/ml, as expected for bolus release of approximately 17 ng 64.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [64] Ueda T, Yokoyama Y, Irahara M, Aono T. Influence of psychological stress on suckling‐induced pulsatile oxytocin release. Obstet Gynecol 1994; 84: 259– 262. 8041543
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIn the paper
“high molecular weight factors in raw plasma interfered with the binding of antibodies to the hormones, leading to high and erroneous readings”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Abstract
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIn the paper
“Stimulation at 50 Hz releases approximately 7 pg of oxytocin per stimulus pulse 66, which is close to the release estimated from the intramammary pressure response to a similar electrical stimulation in vivo.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [66] Leng G, Bicknell R. Hormone release from the neural lobe In: Gross PM, ed. Circumventricular Organs and Body Fluids, vol. III Boca Raton, FL: CRC Press, 1987: 47– 62.
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesThe paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Two different ways of measuring oxytocin in widespread use currently give values in human plasma that differ by two orders of magnitude”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Abstract
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesNone of the sources cited for this claim have been checked yet.
In the paper
“The antidiuretic bioassay was capable of high sensitivity and precision”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioassays of vasopressin
What the sources say
- Not checked yetcites [29] Dettelbach HR. A method for assaying small amounts of antidiuretic substances with notes on some properties of vasopressin. Am J Physiol 1958; 192: 379– 386. 13508887 source
- Not checked yetcites [30] Heller J, Stulc J. [Significance of a new method of titration of antidiuretic hormone]. Cesk Fysiol 1959; 8: 194– 195. 13671532
- Not checked yetcites [33] Segar WE, Moore WW. The regulation of antidiuretic hormone release in man: I. Effects of change in position and ambient temperature on blood ADH levels. J Clin Invest 1968; 47: 2143– 2151. 16695953 PMC297375 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIn the paper
“Like the first radioimmunassays of unextracted plasma, this generates impossibly high and wholly erroneous measurements.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Abstract
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesNone of the sources cited for this claim have been checked yet.
In the paper
“However, many recent papers use an enzyme‐linked immunoassay to measure plasma levels without extracting the samples.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Abstract
What the sources say
- Not checked yetcites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source
- Not checked yetcites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source
- Not checked yetcites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesThe paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“different ways of measuring vasopressin give values in human plasma that differ by two orders of magnitude”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Abstract
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passes
- It may state two separate claims as one. 69
- Probably not a substantial claim. 15
- The quote only partly states this claim. 2
In the paper
“The most sensitive bioassays used mouse mammary strips and could detect 4 μU/ml (7 pg/ml) in 0.5 ml of sample 50, although even this was still not sufficiently sensitive to measure levels in raw plasma.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [50] Robinson IC, Walker JM. Bioassay of oxytocin on the superfused mammary gland of the mouse, using an automatic apparatus. Br J Pharmacol 1974; 50: 409– 417. 4368453 PMC1776677 source
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“the human posterior pituitary contains approximately 21 μg of oxytocin 72, 73, 74, whereas that of the rat contains 0.5–1 μg.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [72] Chard T, Boyd NR, Forsling ML, McNeilly AS, Landon J. The development of a radioimmunoassay for oxytocin: the extraction of oxytocin from plasma, and its measurement during parturition in human and goat blood. J Endocrinol 1970; 48: 223– 234. 5471863 source
- Not checked yetcites [73] Chard T, Forsling ML, James MA, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: sensitivity of the assay in aqueous buffer solution, specificity and the dissociation of immunological and biological activity. J Endocrinol 1970; 46: 533– 542. 5443195 source
- Not checked yetcites [74] Chard T, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: radioiodination, antibody production and separation techniques. J Endocrinol 1970; 46: 269– 278. 5492003 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“the normal plasma concentration of vasopressin in man is below 2 μU/ml (5 pg/ml) 39, with this increasing several fold after 24 h of dehydration.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioassays of vasopressin
What the sources say
- Not checked yetcites [39] Share L. Vasopressin, its bioassay and the physiological control of its release. Am J Med 1967; 42: 701– 712. 5337374 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Across mammalian species, blood volume is linearly proportional to body mass, although the mass of the pituitary is proportional only to the log of body mass 71.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [71] Stahl WR. Organ weights in primates and other mammals. Science 1965; 150: 1039– 1042. 5843618 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Each burst is followed by an increase in intramammary pressure that can be mimicked by i.v. injection of approximately 0.5 mU (0.8 ng) of oxytocin 16, 65.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [16] Leng G, Pineda R, Sabatier N, Ludwig M. 60 YEARS OF NEUROENDOCRINOLOGY: the posterior pituitary, from Geoffrey Harris to our present understanding. J Endocrinol 2015; 226: T173– T185. 25901040 source
- Not checked yetcites [65] Wakerley JB, Lincoln DW. Intermittent release of oxytocin during suckling in the rat. Nat New Biol 1971; 233: 180– 181. 5287087 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“the kidney responds to changes in the rate of vasopressin secretion of < 1 pg/min/kg, and also that the rate required to produce isosmolar urine during overhydration is approximately 3 pg/min/kg”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioassays of vasopressin
What the sources say
- Not checked yetcites [44] Andersen LJ, Andersen JL, Schutten HJ, Warberg J, Bie P. Antidiuretic effect of subnormal levels of arginine vasopressin in normal humans. Am J Physiol 1990; 259: R53– R60. 2375429
- Not checked yetcites [45] Hollander W Jr, Williams TF, Fordham CC 3rd, Welt LG. A study of the quantitative relationship between antidiuretic hormone (vasopressin) and the renal tubular reabsorption of water. J Clin Invest 1957; 36: 1059– 1071. 13449158 PMC1072692 source
- Not checked yetcites [46] Raisz LG, Mc NW, Saxon L. Studies on the renal concentrating mechanism. I. Role of vasopressin. J Lab Clin Med 1958; 52: 437– 445. 13575934
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“he calculated that the half‐life of oxytocin in pregnant women was approximately 3–4 min, and also that uterine contractions were initiated by infusions that raised plasma concentrations by 3 μU/ml (5 pg/ml).”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [52] Saameli K. An indirect method for the estimation of oxytocin blood concentration and half‐life in pregnant women near term. Am J Obstet Gynecol 1963; 85: 186– 192. 13975829 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“The plasma volume of a rat is approximately 4 ml/100 g 67, and so a burst that releases 0.5 mU (0.8 ng) oxytocin should raise circulating concentrations in a 300 g rat by approximately 70 pg/ml.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [67] Lee HB, Blaufox MD. Blood volume in the rat. J Nucl Med 1985; 26: 72– 76. 3965655
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Higuchi et al. 68 developed a radioimmunoassay that allowed oxytocin to be measured in unextracted rat plasma, and took samples at the time of suckling‐induced milk let‐down, as recognised by stretching movements of the suckling pups. Plasma concentrations rose from 16.5 pg/ml to approximately 65 pg/ml, and declined with a half‐life of approximately 1.5 min.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [68] Higuchi T, Tadokoro Y, Honda K, Negoro H. Detailed analysis of blood oxytocin levels during suckling and parturition in the rat. J Endocrinol 1986; 110: 251– 256. 3755747 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“In pigs (body weight 150–200 kg), suckling evokes increases in intramammary pressure equivalent to those induced by i.v. injection of 50 mU (83 ng) of oxytocin 69. For a plasma volume of 5 l, these would raise plasma concentrations by approximately 17 pg/ml, which is close to the peak concentration (approximately 18 pg/ml) measured by radioimmunoassay during suckling‐induced milk ejection.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Bioactivity of oxytocin
What the sources say
- Not checked yetcites [69] Ellendorff F, Forsling ML, Poulain DA. The milk ejection reflex in the pig. J Physiol 1982; 333: 577– 594. 6892113 PMC1197265 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Such interference has also been recognised in immunoassays for cholecystokinin 80, parathyroid hormone 81, thyrotropic hormone 82, glucagon 83, secretin 84, 85, angiotensin 86, growth hormone 87, vasoactive intestinal peptide 88, gonadotrophins 89 and gonadotrophin‐releasing hormone 90, all of which were resolved by sample extraction.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
What the sources say
- Not checked yetcites [80] Harvey RF, Dowsett L, Hartog M, Read AE. Radioimmunoassay of cholecystokinin‐pancreozymin. Gut 1974; 15: 690– 699. 4435583 PMC1413038 source
- Not checked yetcites [81] Christensen MS. A sensitive radioimmunoassay of parathyroid hormone in human serum using a specific extraction procedure. Scand J Clin Lab Invest 1976; 36: 313– 322. 959749 source
- Not checked yetcites [82] Spencer CA, Challand GS. Interference in a radioimmunoassay for human thyrotropin. Clin Chem 1977; 23: 584– 588. 556979 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“The first peak accounted for a 67% loss of binding of antibody to vasopressin, and the second peak accounted for a further 30% loss. Neither of these apparent ‘immunoreactivities’ were displaced by an excess of cold vasopressin, indicating that they did not involve competitive binding of vasopressin, but occlusion of antibody recognition of vasopressin.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“However, it is unlikely that an antibody will recognise a different ligand with higher affinity than it has for the ligand to which it was raised; it may recognise other ligands with lower affinity, although this presents a major problem only if that different ligand is present at higher concentrations.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. The paper’s conclusions may not depend on it. The paper may give no reason why it holds. It may state two separate claims as one.
In the paper
“Thus, close attention was given to extraction and concentration procedures 76.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
None of the sources cited for this claim have been checked yet.
In the paper
“In unextracted plasma from healthy subjects, they found a mean vasopressin concentration of 2 pg/ml in the supine position and 6.2 pg/ml in the upright position, which is in agreement with the levels measured in bioassays by Segar and Moore 33.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
What the sources say
- Not checked yetcites Fyhrquist et al. 92
- Not checked yetcites Segar and Moore 33
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“By filtering plasma samples on G‐25 Sephadex, they resolved three peaks of ‘immunoreactivity’, only the third and smallest of which was consistent with true vasopressin, being recovered in the same location as pure vasopressin 78.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Fortunately, enzymatic degradation of oxytocin is a problem only when measuring in plasma from pregnant women 54, and extraction procedures can remove such enzymes efficiently.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays for oxytocin
What the sources say
- Not checked yetcites [54] Leake RD, Weitzman RE, Glatz TH, Fisher DA. Plasma oxytocin concentrations in men, nonpregnant women, and pregnant women before and during spontaneous labor. J Clin Endocrinol Metab 1981; 53: 730– 733. 7287862 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“In 1973, Chard 76 noted that best immunoassays for vasopressin were no more sensitive than the best bioassays; the same is still true today for most radioimmunoassays and most enzyme‐linked immunoassays (ELISAs).”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“The immunoreactivity in the third peak reproduced the expected relationship with plasma osmotic pressure in normal adults, the expected high levels in nephrogenic diabetes insipidus”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one. The quote only partly states this claim.
In the paper
“After accounting for the plasma matrix interference, Robertson et al. 78 calculated that normal plasma concentrations of vasopressin in man were approximately 1 pg/ml, increasing to approximately 10 pg/ml with dehydration.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“There were considerable difficulties in developing immunoassays for vasopressin and oxytocin. Because these have a low molecular weight, they are poor immunogens, making the preparation of high‐affinity antisera difficult, and circulating levels were clearly very low.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“By the end of the 20th century, bioassays were still ‘of critical importance’ in the discovery of novel antagonists and antagonists of vasopressin and oxytocin because biological activity cannot be inferred from structure alone 75.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [75] Chan WY, Wo NC, Stoev ST, Cheng LL, Manning M. Discovery and design of novel and selective vasopressin and oxytocin agonists and antagonists: the role of bioassays. Exp Physiol 2000; 85S: 7S– 18S. 10795902 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Overestimation can also arise when enzymes are present that degrade oxytocin. The plasma of late pregnant women contains abundant oxytocinase, which inactivates oxytocin by splitting the peptide linkage between the cysteine and tyrosine residues 102.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays for oxytocin
What the sources say
- Not checked yetcites [102] Nomura S, Ito T, Yamamoto E, Sumigama S, Iwase A, Okada M, Shibata K, Ando H, Ino K, Kikkawa F, Mizutani S. Gene regulation and physiological function of placental leucine aminopeptidase/oxytocinase during pregnancy. Biochim Biophys Acta 2005; 1751: 19– 25. 15894523 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“By the 1980s, many laboratories had generated radioimmunoassays for oxytocin, using independently produced antibodies 58, 72, 73, 74, 94, 95, 96, 97, 98] and most, but not all 99, 100, required extraction.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays for oxytocin
What the sources say
- Not checked yetcites [58] Dawood MY. Evolving concepts of oxytocin for induction of labor. Am J Perinatol 1989; 6: 167– 172. 2653335 source
- Not checked yetcites [72] Chard T, Boyd NR, Forsling ML, McNeilly AS, Landon J. The development of a radioimmunoassay for oxytocin: the extraction of oxytocin from plasma, and its measurement during parturition in human and goat blood. J Endocrinol 1970; 48: 223– 234. 5471863 source
- Not checked yetcites [73] Chard T, Forsling ML, James MA, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: sensitivity of the assay in aqueous buffer solution, specificity and the dissociation of immunological and biological activity. J Endocrinol 1970; 46: 533– 542. 5443195 source
- Not checked yetcites [74] Chard T, Kitau MJ, Landon J. The development of a radioimmunoassay for oxytocin: radioiodination, antibody production and separation techniques. J Endocrinol 1970; 46: 269– 278. 5492003 source
- Not checked yetcites [94] Dawood MY, Raghavan KS, Pociask C. Radioimmunoassay of oxytocin. J Endocrinol 1978; 76: 261– 270. 627820 source
- Not checked yetcites [95] Schams D. Oxytocin determination by radioimmunoassay. III. Improvement to subpicogram sensitivity and application to blood levels in cyclic cattle. Acta Endocrinol (Copenh) 1983; 103: 180– 183. 6683055
- Not checked yetcites [96] Chiodera P, Louis F, Legros JJ. Simultaneous radioimmunoassay for plasma arginine‐vasopressin and oxytocin using DEAE Sephadex A 25 extraction. J Endocrinol Invest 1984; 7: 287– 293. 6501802 source
- Not checked yetcites [97] Sheldrick EL, Flint AP. Circulating concentrations of oxytocin during the estrous cycle and early pregnancy in sheep. Prostaglandins 1981; 22: 631– 636. 7323312 source
- Not checked yetcites [98] Landgraf R. Plasma oxytocin concentrations in man after different routes of administration of synthetic oxytocin. Exp Clin Endocrinol 1985; 85: 245– 248. 4018162 source
- Not checked yetcites [99] Burd JM, Weightman DR, Baylis PH. Solid phase radioimmunoassay for direct measurement of human plasma oxytocin. J Immunoassay 1985; 6: 227– 243. 3900138 source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“When oxytocin is incubated in this plasma, there is a rapid loss of biological activity but, after 2 h, oxytocin ‘immunoreactivity’ exceeds bioactivity by three‐fold. The obvious explanation is that oxytocinase in the sample also degrades the labelled oxytocin. The consequence will be that less label will bind to antibody, leading to overestimation of the amount of oxytocin 76.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays for oxytocin
What the sources say
- Not checked yetcites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
- Rests on an assumption: When oxytocin or vasopressin are secreted, the entire contents of the vesicle are discharged, including elements of incompletely processed precursor that may not be biologically active.
In the paper
“This assay was sensitive to 1 pg/ml (0.3 mU/ml) but, when whole plasma was assayed, apparent concentrations were ‘several hundredfold higher than anticipated, failed to fluctuate appropriately in response to physiological manipulation or in disease states, and were not reduced substantially by dialysis of the plasma under conditions that completely removed vasopressin’.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“In some cases, an antibody might recognise elements unrelated to the antigen against which it was raised. Antibodies are sensitive to differences in amino acid sequence, although they recognise the structure of a molecule, not its amino acid sequence per se, and so unrelated molecules might cross‐react 76, 79.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
What the sources say
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“At that time, seven other vasopressin radioimmunoassays had been described, using different antibodies and different extraction procedures. All had reported plasma concentrations of 1–10 pg/ml, and all showed that concentrations changed with physiological conditions in the manner expected from bioassays.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“When describing their vasopressin assay, Baylis and Heath 91 emphasised that ‘even small increases in protein concentration lower binding and therefore careful control of extract protein concentration is needed to avoid the misinterpretation that a fall in binding is due to immunoreactive vasopressin’.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
What the sources say
- Not checked yetcites [91] Baylis PH, Heath DA. The development of a radioimmunoassay for the measurement of human plasma arginine vasopressin. Clin Endocrinol (Oxf) 1977; 7: 91– 102. 891002 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
In the paper
“Antibodies do not recognise the whole molecule, and so an immunoassay may overestimate hormone concentrations by measuring fragments of no physiological significance 76.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays for oxytocin
What the sources say
- Not checked yetcites [76] Chard T. The radioimmunoassay of oxytocin and vasopressin. J Endocrinol 1973; 58: 143– 160. 4578963 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“With few exceptions, assays in extracted plasma reported basal levels of < 5 pg/ml in man: of 17 studies reviewed by Szeto et al. 101, only two reported higher levels; the highest (71 pg/ml) were in samples taken from the corpus cavernosum of the penis, a circumstance that might be expected to affect hormone secretion.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays for oxytocin
What the sources say
- Not checked yetcites Szeto et al. 101
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Henneberry et al. 93 took a different approach: they assayed unextracted samples using standard curves generated in charcoal‐filtered plasma, and reported basal levels of 1.8 pg/ml, rising to 6.3 pg/ml after a saline infusion that raised plasma osmotic pressure from 280 to 293 mOsm/l.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
What the sources say
- Not checked yetcites Henneberry et al. 93
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“High salt concentrations have a nonlinear effect, and serial dilutions of samples with a high salt content produce measurements that do not parallel the standard curve.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Immunoassays > Problems with immunoassays
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. The paper’s conclusions may not depend on it. The paper may give no reason why it holds. It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Thus, to sustain a plasma concentration of 350 pg/ml in man, oxytocin must be produced at 40 000 times the rate measured in the rat. Given a difference in pituitary mass and oxytocin content of just 40‐fold, this seems beyond plausibility.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“In rats, Ginsburg and Heller 104 estimated that the initial half‐life of exogenously administered vasopressin was less than 1 min. From experiments in acutely nephrectomised rats, and by ligating arteries to reduce blood flow through the liver, it was concluded that most of the vasopressin was cleared by the kidneys and liver.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of vasopressin
What the sources say
- Not checked yetcites Ginsburg and Heller 104
Strength:hedged· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Claims that the Enzo ELISA is a valid way of measuring oxytocin in unextracted plasma do not rest on manufacturer's advice, which, as we have noted, strongly recommends extraction, but, instead, converge on two papers, neither of which actually investigated human plasma at all.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Ginsburg and Smith 49, using a rat uterus bioassay, found that, when oxytocin was added to rat plasma in vitro at 37 °C, it was relatively stable, losing just 6–13% of biological activity after 40 min of incubation. By contrast, when oxytocin was injected into rats, the biological activity disappeared with a half‐life of 1.65 min.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
What the sources say
- Not checked yetcites Ginsburg and Smith 49
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“In dogs, rather than measure vasopressin by either radioimmunoassay or bioassay, Ang and Jenkins 106 injected radiolabelled vasopressin and used thin‐layer chromatography and autoradiography to quantify labelled vasopressin in plasma samples: it disappeared with an initial half‐life of 3 min and a later half‐life of 13 min.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of vasopressin
What the sources say
- Not checked yetcites Ang and Jenkins 106
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“In patients with chronic renal failure, the clearance is much slower: MCR is reduced by approximately half 108, which is consistent with the conclusion that vasopressin is mainly cleared from the circulation through the kidneys.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of vasopressin
What the sources say
- Not checked yetcites [108] Argent NB, Wilkinson R, Baylis PH. Metabolic clearance rate of arginine vasopressin in severe chronic renal failure. Clin Sci (Lond) 1992; 83: 583– 587. 1335395 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Ultrafiltration of the plasma indicated that, whereas approximately 30% of similarly infused vasopressin was bound to large molecules, there was no apparent binding of oxytocin. In vitro, oxytocin bioactivity was stable, as found in rats 49; only approximately 20% was lost after 24 h at 37 °C.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
What the sources say
- Not checked yetcites [49] Ginsburg M, Smith MW. The fate of oxytocin in male and female rats. Br J Pharmacol Chemother 1959; 14: 327– 333. 13850241 PMC1481739 source
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“These values are inconsistent, by two orders of magnitude, with both bioassays of oxytocin and vasopressin and with radioimmunoassay measurements on extracted samples. They are also inconsistent with the known sensitivity of peripheral targets to oxytocin and vasopressin, and with what is known of the production, storage and clearance of these hormones.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“In nephrectomised rats, bioactivity declined with a half‐life of approximately 3 min in the first 7 min, and much more slowly thereafter. When the splanchnic vessels were also ligated, the initial decline was not further changed but, after 7 min, oxytocin concentrations ‘stuck’ at a high, steady level.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
What the sources say
- Not checked yetcites Ginsburg and Smith 49
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Ginsburg and Smith deduced that, when oxytocin enters the circulation, it is rapidly removed by the kidneys and liver. Without these organs, injected oxytocin diffuses within approximately 7 min throughout the extracellular fluid, where it is degraded only very slowly.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“As Zhang et al. 125 recognised, ‘One common challenge in biomarker analysis is to select an appropriate surrogate matrix to prepare the standard curve, which is free of target analytes and is identical or similar to unknown sample matrix with regard to ionisation effects and extraction recovery’.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Other methods of measurement
What the sources say
- Not checked yetcites Zhang et al. 125
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“In man, Fabian et al. 43 measured oxytocin with a mammary gland bioassay and studied its clearance after the end of an i.v. infusion. Bioactivity disappeared with a half‐life of 4.8 min with a distribution volume of 10.5 ml/100 g, which is close to the extracellular fluid volume, whereas an acute large injection disappeared with a half‐life of 3.2 min and a distribution volume of 7.4 ml/100 g.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
What the sources say
- Not checked yetcites Fabian et al. 43
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Three studies have compared measurements of oxytocin in the same samples by the same ELISA, with and without extraction 101, 116, 117. Robinson et al. 116 measured oxytocin levels in seals: in extracted samples, they found plasma levels of approximately 8 pg/ml, compared to 300 pg/ml in unextracted samples collected in heparin vacutainers or 540 pg/ml in unextracted samples collected in ethylenedi”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
What the sources say
- Not checked yetcites [101] Szeto A, McCabe PM, Nation DA, Tabak BA, Rossetti MA, McCullough ME, Schneiderman N, Mendez AJ. Evaluation of enzyme immunoassay and radioimmunoassay methods for the measurement of plasma oxytocin. Psychosom Med 2011; 73: 393– 400102. 21636661 PMC3118424 source
- Not checked yetcites [116] Robinson KJ, Hazon N, Lonergan M, Pomeroy PP. Validation of an enzyme‐linked immunoassay (ELISA) for plasma oxytocin in a novel mammal species reveals potential errors induced by sampling procedure. J Neurosci Meth 2014; 226: 73– 79. 24485867 source
- Not checked yetcites [117] Christensen JC, Shiyanov PA, Estepp JR, Schlager JJ. Lack of association between human plasma oxytocin and interpersonal trust in a Prisoner's Dilemma paradigm. PLoS ONE 2014; 9: e116172. 25549255 PMC4280178 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one. The quote only partly states this claim.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As a proportion of body weight, the rat pituitary contains much more oxytocin and vasopressin than the human pituitary and, in rats, as in other small mammals, circulating levels are higher. In man and other large animals, lower circulating concentrations are compensated for by higher sensitivity of the peripheral targets.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear? Based: unclearIt may state two separate claims as one.
In the paper
“In similar studies on rabbits, vasopressin was cleared with a half‐life of approximately 4.5 min, again mainly by the kidneys and liver 105.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of vasopressin
What the sources say
- Not checked yetcites [105] Hasan SH. The clearance of vasopressin and deaminovasopressin from the circulation of normal and hypercalcaemic rabbits. J Endocrinol 1970; 47: 135– 136. 5428912 source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Plasma oxytocin was measured using human plasma diluted with water (1 : 6) as a surrogate matrix, and basal levels of 2.4 pg/ml for human but 298 pg/ml for rat were reported; these high levels in the rat may indicate that human plasma is not a suitable surrogate matrix for measurements in rat.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Other methods of measurement
What the sources say
- Not checked yetcites [125] Zhang G, Zhang Y, Fast DM, Lin Z, Steenwyk R. Ultra sensitive quantitation of endogenous oxytocin in rat and human plasma using a two‐dimensional liquid chromatography‐tandem mass spectrometry assay. Anal Biochem 2011; 416: 45– 52. 21609710 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“The ELISAs in most widespread use are marketed by Enzo, who draw attention to problems of plasma matrix interference, illustrate the benefits of plasma extraction, and ‘strongly recommend’ that users extract plasma samples. They state that it might be possible to dilute plasma samples enough to minimise interference (suggesting at least an eight‐fold dilution) but warn that the user ‘must verify’”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
What the sources say
- Not checked yetcites [113] Designs A. Oxytocin Enzyme Immunoassay (EIA) kit: Manufacturer's instructions. Available at: http://www.enzolifesciences.com/fileadmin/redacteur/pdf/adi/ADI-900-153.pdf.
Load-bearingStrength:boosted· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Carter et al. 119 further reported that, in voles, the ELISA measured spiked levels of oxytocin with a gain of 0.79. This is consistent with a reduction in antibody affinity, although not sufficient to account for the discrepancy between expected and measured oxytocin in Kramer's measurements. In fractions separated by high‐performance liquid chromatography (HPLC), apparent immunoreactivity”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
What the sources say
- Not checked yetcites Carter et al. 119
- Not checked yetcites Wismer Fries et al. 120
- Not checked yetcites Anderson 121
Strength:hedged· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Lauson and Bacanegra 103 concluded that, at physiological levels (< 25 pg/ml), most of the vasopressin in the plasma is bound to protein and is cleared by the tubular cells of the kidney by enzymatic inactivation and secretion. At higher concentrations, more of the vasopressin is unbound, and some of this is filtered at the glomerulus and excreted into urine.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of vasopressin
What the sources say
- Not checked yetcites Lauson and Bacanegra 103
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“The binding of plasma proteins to an immunoglobulin can hinder its ability to bind an antigen, reducing its affinity, and/or may completely prevent some of the antibody from binding to antigen.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
What the sources say
- Not checked yetcites Bjerner et al. 111
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“A plasma concentration of 350 pg/ml oxytocin in man implies a total blood content of approximately 1 μg (given a plasma volume of 3 l) and, for a half‐life of 10 min, this implies a secretion (and production) rate of 500 ng/min, or 720 IU/day.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds. It may state two separate claims as one.
In the paper
“In the baboon, oxytocin clearance was also consistent with a two‐compartmental system, corresponding to the plasma volume and the extravascular fluid volume. The initial half‐life in the first phase of distribution was approximately 1 min, followed by a terminal elimination half‐life of approximately 10 min 109.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
What the sources say
- Not checked yetcites [109] Kowalski WB, Diveky L, Mehendale R, Parsons M, Wilson L Jr. Effect of pregnancy on the metabolic clearance rate and the volume of distribution of oxytocin in the baboon. Am J Physiol 1998; 274: E791– E795. 9612235 source
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“This advice has been largely ignored. Many studies have used this assay on unextracted human plasma, reporting mean concentrations of oxytocin and vasopressin of typically 100–500 pg/ml 3, 4, 5, 6, 114. Those that have diluted samples have done so by less than the manufacturer's recommendations, without confirming that their dilution is adequate.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
What the sources say
- Not checked yetcites [3] Jobst A, Dehning S, Ruf S, Notz T, Buchheim A, Henning‐Fast K, Meissner D, Meyer S, Bondy B, Muller N, Zill P. Oxytocin and vasopressin levels are decreased in the plasma of male schizophrenia patients. Acta Neuropsychiatr 2014; 26: 347– 355. 25288094 source
- Not checked yetcites [4] Apter‐Levi Y, Zagoory‐Sharon O, Feldman R. Oxytocin and vasopressin support distinct configurations of social synchrony. Brain Res 2014; 1580: 124– 132. 24184577 source
- Not checked yetcites [5] Gouin JP, Pournajafi‐Nazarloo H, Carter CS. Changes in social functioning and circulating oxytocin and vasopressin following the migration to a new country. Physiol Behav 2015; 139: 67– 72. 25446216 source
- Not checked yetcites [6] Rubin LH, Carter CS, Bishop JR, Pournajafi‐Nazarloo H, Drogos LL, Hill SK, Ruocco AC, Keedy SK, Reilly JL, Keshavan MS, Pearlson GD, Tamminga CA, Gershon ES, Sweeney JA. Reduced levels of vasopressin and reduced behavioral modulation of oxytocin in psychotic disorders. Schizophr Bull 2014; 40: 1374– 1384. 24619535 PMC4193712 source
- Not checked yetcites [114] McCullough ME, Churchland PS, Mendez AJ. Problems with measuring peripheral oxytocin: can the data on oxytocin and human behavior be trusted? Neurosci Biobehav Rev 2013; 37: 1485– 1492. 23665533 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“With a reduction in maximum binding, an assay will report erroneously high values even in the absence of any antigen, and will exaggerate measurements of added antigen. The effect of a reduction in affinity is different; in this case, the assay will under‐report added antigen. In both cases, there will be a linear relationship between the amount of oxytocin added and the amount measured, although”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Bolus injections disappear rapidly, with an initial half‐life that varies between species and with physiological state: in the rat, both hormones disappear with an initial half‐life of approximately 2 min; in man, oxytocin has a half‐life of 4–10 min, except in pregnancy, when oxytocinase shortens this to 2–3 min.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Later studies using radioimmunoassay indicated that vasopressin disappeared from the circulation with a bi‐exponential decay 107. The fast phase had a half‐life of 5.4 min and a distribution volume close to the plasma volume; the slower phase had a half‐life of 78 min with a distribution volume close to the extracellular fluid volume.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of vasopressin
What the sources say
- Not checked yetcites [107] Benmansour M, Rainfray M, Paillard F, Ardaillou R. Metabolic clearance rate of immunoreactive vasopressin in man. Eur J Clin Invest 1982; 12: 475– 480. 6818035 source
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Finally, Carter et al. 119 claimed a biological validation for the ELISA in reporting that breastfeeding in women lowered salivary concentrations of oxytocin. Because suckling‐induced oxytocin release is indispensable for milk let‐down in mammals 123, it is difficult to understand how such an anomalous observation qualifies as a validation.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Clearance of oxytocin > ELISAs
What the sources say
- Not checked yetcites Carter et al. 119
- Not checked yetcites [123] Russell JA, Leng G, Douglas AJ. The magnocellular oxytocin system, the fount of maternity: adaptations in pregnancy. Front Neuroendocrinol 2003; 24: 27– 61. 12609499 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Plasma matrix proteins interact differently with different antibodies: there is no single ‘radioimmunoassay’ and no single ‘ELISA’, and their susceptibility to plasma interference will vary with the antibody.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds. It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In mice or rats, validation could be supported by measurements in gene knockout animals or hypophysectomised animals, where vasopressin and/or oxytocin are completely absent.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. The paper’s conclusions may not depend on it. The paper may give no reason why it holds.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Measuring oxytocin or vasopressin in cerebrospinal fluid, urine or saliva should have fewer problems because of the relative absence of proteins but, in urine, high salt concentrations may cause problems.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Plasma proteins differ from individual to individual and differ substantially from species to species, and so any assay in unextracted plasma should be validated for the species in which it is measured and, ideally, in the individual in which it is measured.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
- Rests on an assumption: Plasma proteins differ from individual to individual and differ substantially from species to species.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“If so, these assays are, for unextracted plasma, no more than a random number generator, returning arbitrary values with a high variance and a skewed distribution.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Conclusions
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes? Based: unclearIt may state two separate claims as one.
In the paper
“Levels from an individual could be verified by constructing a standard curve in hormone‐stripped plasma from that individual 93, although it would be impractical to do this for every individual sampled.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
What the sources say
- Not checked yetcites [93] Henneberry HP, Slater JD, Eisen V, Fuhr S. Arginine vasopressin response to hypertonicity in hypertension studied by arginine vasopressin assay in unextracted plasma. J Hypertens 1992; 10: 221– 228. 1315818 source
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The current impetus for measuring vasopressin and oxytocin in plasma derives from the assumption that such measurements are informative of the release of these peptides in the brain.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Conclusions
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There are no known conditions in which oxytocin is absent or grossly elevated in man. However, given the sensitivity of peripheral targets in man to very low concentrations, levels exceeding 10 pg/ml should be treated with scepticism.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
- Rests on an assumption: There are no known conditions in which oxytocin is absent or grossly elevated in man.
In the paper
“Analytical verification by HPLC with chromatographic separation 78 can confirm that measurements reflect the actual peptide, although care is needed to ensure that all high molecular weight proteins are eluted from the column and are checked for assay interference. There should be quantitative agreement between the amount of hormone in the appropriate fraction, and the level reported by the assay.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
What the sources say
- Not checked yetcites [78] Robertson GL, Klein LA, Roth J, Gorden P. Immunoassay of plasma vasopressin in man. Proc Natl Acad Sci USA 1970; 66: 1298– 1305. 5273455 PMC335819 source
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Such interference will be apparent if, when samples are spiked with known amounts of hormone, the assay recovers amounts different to the expected amount. However, the absence of a difference is not conclusive because interference that reduces antibody affinity will lead to underestimation of added hormone, whereas interference that blocks antibody recognition will lead to overestimation, and both”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Samples with high levels will not dilute in parallel to the standard curve if the levels arise from factors such as high salt levels or pH that interfere nonlinearly with antibody binding, although this test may not recognise protein interference that blocks antibody recognition of the antigen.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“The fallacies in this supposition have been discussed elsewhere 2.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Conclusions
What the sources say
- Not checked yetcites [2] Leng G, Ludwig M. Intranasal oxytocin: myths and delusions. Biol Psychiatry 2016; 79: 243– 250. 26049207 source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
None of the sources cited for this claim have been checked yet.
In the paper
“In a steady‐state, unbound oxytocin and vasopressin should distribute in extravascular fluids, and measurements in saliva might be useful for estimating steady‐state secretion 119, 130, 131.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Recommendations for validation of assays
What the sources say
- Not checked yetcites [119] Carter CS, Pournajafi‐Nazarloo H, Kramer KM, Ziegler TE, White‐Traut R, Bello D, Schwertz D. Oxytocin: behavioral associations and potential as a salivary biomarker. Ann N Y Acad Sci 2007; 1098: 312– 322. 17435137 source
- Not checked yetcites [130] Jong TR, Menon R, Bludau A, Grund T, Biermeier V, Klampfl SM, Jurek B, Bosch OJ, Hellhammer J, Neumann ID. Salivary oxytocin concentrations in response to running, sexual self‐stimulation, breastfeeding and the TSST: the Regensburg Oxytocin Challenge (ROC) study. Psychoneuroendocrinology 2015; 62: 381– 388. 26385109 source
- Not checked yetcites [131] Frijling JL, van Zuiden M, Nawijn L, Koch SB, Neumann ID, Veltman DJ, Olff M. Salivary oxytocin and vasopressin levels in police officers with and without post‐traumatic stress disorder. J Neuroendocrinol 2015; 27: 743– 751. 26184739 source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Tsukazaki et al. 127 developed an LC‐MS/MS assay for vasopressin, and showed that, correcting for plasma matrix effects, this assay gave close quantitative agreement with radioimmunoassay results from extracted human plasma samples, and had higher sensitivity and needed smaller sample volumes.”
§ Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers > Other methods of measurement
What the sources say
- Not checked yetcites Tsukazaki et al. 127
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper relies on this without support.
Antibodies are generally raised to the tail end of oxytocin, which confers its biological activity.
“Generally, antibodies are raised to the tail end of oxytocin, which confers its biological activity.”
The paper relies on this without support.
When oxytocin or vasopressin are secreted, the entire contents of the vesicle are discharged, including elements of incompletely processed precursor that may not be biologically active.
“When oxytocin or vasopressin are secreted, the entire contents of the vesicle are discharged, including elements of incompletely processed precursor that may not be biologically active.”
The paper relies on this without support.
There are no known conditions in which oxytocin is absent or grossly elevated in man.
“There are no known conditions in which oxytocin is absent or grossly elevated in man.”
The paper relies on this without support.
Plasma proteins differ from individual to individual and differ substantially from species to species.
“Plasma proteins differ from individual to individual and differ substantially from species to species”
The paper relies on this without support.
Oxytocin injected subcutaneously is expected to distribute throughout the extracellular fluid volume.
“Oxytocin injected s.c. is expected to distribute throughout the extracellular fluid volume”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“there is as yet no consensus on the best methodology to measure OXT22,23”
From Effect of age and autism spectrum disorder on oxytocin receptor density in the human basal forebrain and midbrain (2018) · cites this paper as [22]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Daily oxytocin production is inconsistent with the stated infusion rate
MajorAt the paper's stated potency of 600 IU/mg, 500 ng/min amounts to 0.72 mg/day, or 432 IU/day-not 720 IU/day. The erroneous daily value underlies the subsequent comparison with production in rats.
“500 ng/min, or 720 IU/day.”
“1 mg of oxytocin a potency of 600 IU”
Vasopressin assay sensitivity uses the wrong unit scale
MinorThe paper elsewhere equates 4 μU with 10 pg of vasopressin, so 1 pg corresponds to about 0.4 μU-not 0.3 mU (300 μU). The stated sensitivity is therefore hundreds of times larger than its stated equivalent.
“This assay was sensitive to 1 pg/ml (0.3 mU/ml)”
“antidiuretic assays could measure vasopressin at doses as low as 4 μU (10 pg)”
Citations
Cites
Why Most Published Research Findings Are False
Ioannidis, J. P. A. (2005) · PLoS Medicine · doi:10.1371/journal.pmed.0020124 · Cited by 10,505
Simmons, J. P., Nelson, L. D., & Simonsohn, U. (2011) · Psychological Science · doi:10.1177/0956797611417632 · Cited by 6,567
Why Most Published Research Findings Are False
Ioannidis, J. P. A. (2005) · CHANCE · doi:10.1080/09332480.2005.10722754 · Cited by 4,702
Oxytocin: The Great Facilitator of Life
Lee, H.-J., Macbeth, A. H., Pagani, J., & Young, W. S. (2009) · Progress in Neurobiology · doi:10.1016/j.pneurobio.2009.04.001 · Cited by 1,037
On some physiological actions of ergot
Dale, H. H. (1906) · The Journal of Physiology · doi:10.1113/jphysiol.1906.sp001148 · Cited by 1,011 · cited 1×
“In 1906, as an aside in a study of the actions of ergot, Dale (7) reported that extracts of the posterior pituitary gland could stimulate uterine contractions.”
BackgroundDendritic peptide release and peptide-dependent behaviours
Ludwig, M., & Leng, G. (2006) · Nature Reviews Neuroscience · doi:10.1038/nrn1845 · Cited by 945
Self-citationIntranasal Oxytocin: Myths and Delusions
Leng, G., & Ludwig, M. (2016) · Biological Psychiatry · doi:10.1016/j.biopsych.2015.05.003 · Cited by 616 · cited 2×
“The sense behind this has been questioned (2).”
BackgroundSelf-citationCOMPUTER ANALYSIS OF RADIOLIGAND ASSAY AND RADIOIMMUNOASSAY DATA
Rodbard, D., & Lewald, J. E. (1970) · European Journal of Endocrinology · doi:10.1530/acta.0.065s079 · Cited by 608 · cited 1×
“This is a hyperbolic curve, which, after allowing for nonspecific binding, can be reduced by a logit-log transformation to a straight line (77).”
BackgroundFries, A. B. W., Ziegler, T. E., Kurian, J. R., Jacoris, S., & Pollak, S. D. (2005) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0504767102 · Cited by 570
Lee, H. B., & Blaufox, M. D. (1985) · PubMed · Cited by 565
Flagged first, then by citations
5 of the 139 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- Heller J, 1959, [Significance of a new method of titration of antidiuretic hormone], Cesk Fysiol, 8, 194
Cited by
Sorted by citations
Items 201–242 of 242
Oxytocin: much more than childbirth and milk letdown
Hollenberg, M. D. (2021) · Clinical Science · doi:10.1042/cs20210180 · Cited by 1
Oxytocin in Lactation and Parturition
Leng, G. (2024) · Masterclass in neuroendocrinology · doi:10.1007/978-3-031-51138-7_6 · Cited by 2
Self-citationFecteau, K. M., Shnitko, T. A., Grant, K. A., & Erikson, D. W. (2024) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.9839 · Cited by 1
Azar, M. E., & Vardast, M. R. (2024) · Studies in Medical Sciences · doi:10.61186/umj.35.12.1035 · Cited by 1
Role of oxytocin in social cognition in psychosis spectrum disorders
Benavides, C., Fred-Torres, S., Rutter, S., Larsen, E. M., Fastman, J., Bulbena‐Cabré, A., Lepow, L., Mahon, K., Russo, M., Ungar, A. K., et al. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-815315-4.00006-9 · Cited by 1
Circulating copeptin level and the clinical prognosis of patients with chronic liver disease
Tan, H.-Q., Zhao, M. D., Huang, Z., et al. (2023) · World Journal of Gastroenterology · doi:10.3748/wjg.v29.i31.4797 · Cited by 1
Moscovice, L. R., Sobczak, B., Gimsa, U., & Kanitz, E. (2021) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2021.06.11.448025 · Cited by 1
“for measuring peripheral OXT, since blood and plasma contain proteins that may 455 interfere with detection, and urine has a high salt concentration that may interfere 456 with measurement [35,58].”
BackgroundValidation of a Commercial Radioimmunoassay for Measurement of Human Peripheral OxytocinPreprint
Gerred, K., & Kapoor, A. (2023) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2023.03.22.533823 · Cited by 1
“Traditionally, OT was measured by radioimmunoassays (RIA) for which full in-house method validations were conducted 10 .”
BackgroundMethodThe effects of oxytocin on social behaviour: the influence of context and individual differences
Daughters, K. (2016) · ORCA Online Research @Cardiff (Cardiff University)
Tracking effects of oxytocin crossing the blood brain barrier in vivo and in vitro
Chakraborty, T. (2024) · Library, Museums and Press - UDSpace (University of Delaware) · doi:10.58088/6tnt-6n23
On oxytocin and social behavior
Hovey, D. (2018) · Gothenburg University Publications Electronic Archive (Gothenburg University)
The role of oxytocin, testosterone and cortisol in affiliation and bonding in male Barbary macaqaues
Rincon, A. V. (2020) · eDiss (Georg-August-Universität Göttingen) · doi:10.53846/goediss-7818
Gajewska‐Knapik, K., Kumar, S., Sutton-Cole, A., et al. (2021) · SSRN Electronic Journal · doi:10.2139/ssrn.3798543
Evaluating Methylation of the Oxytocin Receptor Gene and the Oxytocin Intergenic Region
King, L., Robins, S., Chen, G., Turecki, G., & Zelkowitz, P. (2021) · Methods in molecular biology · doi:10.1007/978-1-0716-1759-5_6
Diagnostic Utility of Serum Copeptin Level in Children with Febrile Seizures
Abdulhafith, A. A. A., Siam, A. G., Mohammed, A. A. S., et al. (2023) · The Egyptian Journal of Hospital Medicine · doi:10.21608/ejhm.2023.288377
Normal concentrations of steroids and their regulators in blood
Honour, J. W. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-818124-9.00003-6
Regulation of steroid production
Honour, J. W. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-818124-9.00017-6
Reciprocal interactions between the oxytocin and somatosensory systems
Küppers, S., Lefèvre, A., & Grinevich, V. (2024) · Elsevier eBooks · doi:10.1016/b978-0-12-820480-1.00085-1
Gnanadesikan, G. E., Bray, E. E., Cook, E. N., et al. (2024) · SSRN Electronic Journal · doi:10.2139/ssrn.4828294
Changes in peripheral oxytocin and vasopressin during a silent month-long Insight meditation retreat
Conklin, Q. A., Zanesco, A. P., King, B. G., Epel, E. S., & Saron, C. D. (2024) · Frontiers in Endocrinology · doi:10.3389/fendo.2024.1345527 · Cited by 4
“There are several compelling arguments against the assessment of oxytocin in unextracted samples, which suggest that resultant oxytocin values may be grossly overestimated, contaminated by other molecular species, or even completely arbitrary (59, 60, 109).”
BackgroundThe oxytocin system in patients with craniopharyngioma: A systematic reviewPreprint
Mann, A., Kalitsi, J., Jani, K., et al. (2024) · medRxiv · doi:10.1101/2024.07.31.24311260
“In patients with craniopharyngioma (n=70), mean/median baseline salivary oxytocin concentrations ranged between 0.32 pg/mL and 1.90 pg/mL for samples quantified using RIA (Brandi et al., 2020; Gebert et al., 2018; Hoffmann et al., 2017), and between 3.3 pg/mL and 3.6 pg/mL for samples quantified using EIA (Daubenbüche…”
InfluentialA Validated Method for the Simultaneous Determination of Oxytocin and Cortisol in Human Saliva
Polledri, E., Mercadante, R., Campo, L., & Fustinoni, S. (2024) · Separations · doi:10.3390/separations11080240 · Cited by 1
Rajamannar, P., Raz, O., & Levkowitz, G. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.08.07.606980
Snowden, A. W., Schwartz, S., Smith, A. L., et al. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.08.23.609421
Хрущова, А. М., Vasilieva, N. Y., & Rogovin, K. A. (2024) · Журнал общей биологии · doi:10.31857/s0044459624050014
Establishment of the oxytocin measurement assay ELISA and applications
Takeyama, K.-I. (2024) · Journal for the Integrated Study of Dietary Habits · doi:10.2740/jisdh.35.3_121
Circadian Variation of Salivary Oxytocin in young adult womenPreprint
de Almeida, M. F. R., Quattrocchi, L., Perroud, N. A., et al. (2025) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2025.03.26.645423
Goulter, N., Kaouar, S., & Kimonis, E. R. (2024) · OSF Preprints (OSF Preprints) · doi:10.31219/osf.io/yu9wv_v1
Goulter, N., Kaouar, S., & Kimonis, E. R. (2025) · OSF Preprints (OSF Preprints) · doi:10.31219/osf.io/yu9wv_v2
Wei, S., Tseng, H., Chang, H. H., & Chen, P. S. (2025) · International Journal of Bipolar Disorders · doi:10.1186/s40345-025-00393-8
“OXT immunoreactivity levels cannot accurately infer true values of OXT, and therefore cannot be compared between studies (Leng and Sabatier 2016; McCullough et al. 2013; Robinson et al. 2014).”
Biotinidase as a novel biomarker for pain assessment in dairy cattle
Cohen-Zinder, M., Shabtay, A., Shor-Shimoni, E., Honig, H. H., Asher, A., Friedman, S., Ross, M., Yitzhak, N., Shilo‐Benjamini, Y., & Salzer, Y. (2025) · PAIN Reports · doi:10.1097/pr9.0000000000001338
Copeptin (Not Only) in Water and Sodium Disorders
Drugda, J., Gabalec, F., & Horáček, J. (2025) · Acta Medica (Hradec Kralove Czech Republic) · doi:10.14712/18059694.2025.24
“However, direct measurement of AVP levels is significantly limited by its short half-life of approximately 24 minutes (4), secretory pulsatility, and rapid in vitro degradation, which requires complicated preanalytical steps and makes interpretation rather difficult.”
InfluentialZhang, X., Fralick, L., & Ethun, K. F. (2026) · Biomedical Chromatography · doi:10.1002/bmc.70349
“…the use of radioactive material, cross-reactivity issues with commercial kits, potential underestimation of circulating OT levels due to peptide aggregation, and insufficient sensitivity among nonpregnant and nonlactating individuals (Tabak et al. 2023; Leng and Sabatier 2016; Kastin et al. 1984).”
Tully, J., & Hind, E. (2026) · Aggression and Violent Behavior · doi:10.1016/j.avb.2026.102144
Tanaka, K., Takahashi, K., Hidaka, C., Osako, Y., & Nakane, H. (2026) · Neuroscience · doi:10.1016/j.neuroscience.2026.03.047
Mendonça, T., Ribeiro, G., Casqueiro, T., Prazeres, J., & Mills, D. S. (2026) · BMC Veterinary Research · doi:10.1186/s12917-026-05509-5
“Like-wise, the preferred laboratory method is based on a pragmatic consideration of both the accessibility and reduced costs of the ELISA technique compared to HPLC. Cross-reactions within ELISA and other immunoassays have been described [58], presenting a potential limitation for these approaches.”
Ouyang, J., León, G. A., Rinne, G. R., & Saxbe, D. (2026) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2026.107901
Gao, Z., Sun, M., Xun, G., Li, Z., Pan, Z., Li, Y., Kendrick, K. M., Li, N., Xu, X., Becker, B., et al. (2026) · Science China Chemistry · doi:10.1007/s11426-025-3338-6
Vasotocin influences mother–offspring associations in a facultatively family-living lizard
Lake, J., While, G. M., Ophir, A. G., Rubenstein, D. R., Witczak, L. R., & Whiting, M. J. (2026) · Behavioral Ecology · doi:10.1093/beheco/arag082
“There is disagreement in the literature regarding whether to extract plasma samples or not, with unex-tracted values typically leading to higher concentrations of hormone (Leng and Sabatier 2016).”
Grifnée, E., Cavalier, E., & Goff, C. L. (2026) · Clinica Chimica Acta · doi:10.1016/j.cca.2026.121247
Altered mechanisms of social rank recognition in schizophrenia spectrum disorders
Koçyiğit, D., Mutlu, E., Ecemiş, D., Özerhan, Ş., Korkmaz, H., Pehli̇vanoğlu, B., Yağcioğlu, A. E. A., & Goto, Y. (2026) · Journal of Psychiatric Research · doi:10.1016/j.jpsychires.2026.09.020
A multiscale PBPK‐PD model of oxytocin‐induced uterine excitation and contraction
Yang, Y., Bradley, C. P., Zhang, X., Clark, A. R., Gao, M., Li, G., Yang, L., Ortiz, I. R. M., Meng, Y., Hao, D., et al. (2026) · The Journal of Physiology · doi:10.1113/jp291462
“…released much less than 20% of the pituitary pool during labour, whereas the human posterior pituitary contains approximately 21 μg of oxytocin (Leng & Sabatier, 2016), 1.05% of release (21 μg × 0.0105 = 0.2205 μg) into 2.5 L (venous plasma) (Gaohua et al., 2012), which will give rise to 88…”
Influential
Metadata
- Authors
- Leng, G., & Sabatier, N.
- Affiliations
- University of Edinburgh (GB)
- Corresponding author
- Leng, G.
- ORCID
- Leng, G.
- Venue
- Journal of Neuroendocrinology
- Publisher
- Wiley
- Published
- 2016-09-23
- Volume
- 28
- Issue
- 10
- ISSN
- 0953-8194 (ISSN-L)
- 1365-2826
- PMC
- PMC5096068
- OpenAlex
- W2498209759
- Semantic Scholar
- 98dc3b3b12cdb3a15b162393aa170b70b7862e72
- Cited by
- 237
- References
- 139
- Citations by year
- 2026: 14 · 2025: 22 · 2024: 29 · 2023: 21 · 2022: 16 · 2021: 36 · 2020: 22 · 2019: 24 · 2018: 29 · 2017: 21 · 2016: 3
- FWCI
- 4.04
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Journal of Neuroendocrinology (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Edinburgh Research Explorer (University of Edinburgh) (submittedVersion, cc-by-nc-nd)
- Funders
- Seventh Framework Programme (607310)
- Topics
- Life Sciences › Neuroscience › Endocrine and Autonomic Systems › Neuroendocrine regulation and behavior
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuropeptides and Animal Physiology
- Keywords
- oxytocin, vasopressin, plasma hormone measurement, immunoassay interference, bioassay, enzyme-linked immunoassay, radioimmunoassay, plasma extraction, method validation
- MeSH terms
- Animals, Biological Assay, Humans, Immunoassay, Oxytocin, Vasopressins, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-08, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-08 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review