Comprehensive Review on Wearable Sweat-Glucose Sensors for Continuous Glucose Monitoring
Open accessAbstract
The incidence of diabetes is increasing at an alarming rate, and regular glucose monitoring is critical in order to manage diabetes. Currently, glucose in the body is measured by an invasive method of blood sugar testing. Blood glucose (BG) monitoring devices measure the amount of sugar in a small sample of blood, usually drawn from pricking the fingertip, and placed on a disposable test strip. Therefore, there is a need for non-invasive continuous glucose monitoring, which is possible using a sweat sensor-based approach. As sweat sensors have garnered much interest in recent years, this study attempts to summarize recent developments in non-invasive continuous glucose monitoring using sweat sensors based on different approaches with an emphasis on the devices that can potentially be integrated into a wearable platform. Numerous research entities have been developing wearable sensors for continuous blood glucose monitoring, however, there are no commercially viable, non-invasive glucose monitors on the market at the moment. This review article provides the state-of-the-art in sweat glucose monitoring, particularly keeping in sight the prospect of its commercialization. The challenges relating to sweat collection, sweat sample degradation, person to person sweat amount variation, various detection methods, and their glucose detection sensitivity, and also the commercial viability are thoroughly covered.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · cited 4×
“[39,107,118,119]”
BackgroundMethodInfluentialELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY
Clark, L. C., & Lyons, C. (1962) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1962.tb13623.x · Cited by 3,945
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Nathan, D. M., & for the DCCT/EDIC Research Group (2014) · Diabetes Care · doi:10.2337/dc13-2112 · Cited by 1,832
Lee, H., Choi, T. K., Lee, Y. B., Cho, H. R., Ghaffari, R., Wang, L., Choi, H. J., Chung, T. D., Lu, N., Hyeon, T., et al. (2016) · Nature Nanotechnology · doi:10.1038/nnano.2016.38 · Cited by 1,802
Nanozymes: From New Concepts, Mechanisms, and Standards to Applications
Liang, M., & Yan, X. (2019) · Accounts of Chemical Research · doi:10.1021/acs.accounts.9b00140 · Cited by 1,691
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486 · cited 1×
“The concentration of glucose in human sweat is from 0.06–0.2 mM and corresponds to 3.3–17.3 mM in BG [29,30].”
BackgroundInfluentialSelection in vitro of an RNA enzyme that specifically cleaves single-stranded DNA
Robertson, D. L., & Joyce, G. F. (1990) · Nature · doi:10.1038/344467a0 · Cited by 1,331
A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat
Koh, A., Kang, D., Xue, Y., Lee, S., Pielak, R. M., Kim, J., Hwang, T., Min, S., Banks, A., Bastien, P., et al. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf2593 · Cited by 1,308
Lee, H., Song, C., Hong, Y. S., Kim, M., Cho, H. R., Kang, T., Shin, K., Choi, S. H., Hyeon, T., & Kim, D.-H. (2017) · Science Advances · doi:10.1126/sciadv.1601314 · Cited by 1,118
Electrochemical Non-enzymatic Glucose Sensors: A Perspective and an Evaluation
Toghill, K. E., & Compton, R. G. (2010) · International Journal of Electrochemical Science · doi:10.1016/s1452-3981(23)15359-4 · Cited by 926
Electrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862
Emaminejad, S., Gao, W., Wu, E., Davies, Z. A., Yin Yin Nyein, H., Challa, S., Ryan, S. P., Fahad, H. M., Chen, K., Shahpar, Z., et al. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1701740114 · Cited by 839
Recent advances in electrochemical non-enzymatic glucose sensors – A review
Hwang, D.-W., Lee, S., Seo, M., & Chung, T. D. (2018) · Analytica Chimica Acta · doi:10.1016/j.aca.2018.05.051 · Cited by 788
Glucose Sensing for Diabetes Monitoring: Recent Developments
Bruen, D., Delaney, C., Florea, L., & Diamond, D. (2017) · Sensors · doi:10.3390/s17081866 · Cited by 777
Bandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav3294 · Cited by 761
Accessing analytes in biofluids for peripheral biochemical monitoring
Heikenfeld, J., Jajack, A., Feldman, B., Granger, S. W., Gaitonde, S., Begtrup, G., & Katchman, B. A. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0040-3 · Cited by 758
Physiology of sweat gland function: The roles of sweating and sweat composition in human health
Baker, L. B. (2019) · Temperature · doi:10.1080/23328940.2019.1632145 · Cited by 732
Sonner, Z., Wilder, E., Heikenfeld, J., Kasting, G., Beyette, F., Swaile, D., Sherman, F., Joyce, J., Hagen, J., Kelley-Loughnane, N., et al. (2015) · Biomicrofluidics · doi:10.1063/1.4921039 · Cited by 724
Klonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Tattoo-Based Noninvasive Glucose Monitoring: A Proof-of-Concept Study
Bandodkar, A. J., Jia, W., Yardımcı, C., Wang, X., Ramirez, J., & Wang, J. (2015) · Analytical Chemistry · doi:10.1021/ac504300n · Cited by 700
Wireless battery-free wearable sweat sensor powered by human motion
Song, Y., Min, J., Yu, Y., Wang, H., Yang, Y., Zhang, H., & Gao, W. (2020) · Science Advances · doi:10.1126/sciadv.aay9842 · Cited by 652
Noninvasive Alcohol Monitoring Using a Wearable Tattoo-Based Iontophoretic-Biosensing System
Kim, J., Jeerapan, I., Imani, S., Cho, T. N., Bandodkar, A., Cinti, S., Mercier, P. P., & Wang, J. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00356 · Cited by 599
Wearable non-invasive epidermal glucose sensors: A review
Kim, J., Campbell, A. S., & Wang, J. (2018) · Talanta · doi:10.1016/j.talanta.2017.08.077 · Cited by 586
Non-invasive glucose monitoring technology in diabetes management: A review
Vashist, S. K. (2012) · Analytica Chimica Acta · doi:10.1016/j.aca.2012.03.043 · Cited by 575
Correlation Between Sweat Glucose and Blood Glucose in Subjects with Diabetes
Moyer, J., Wilson, D., Finkelshtein, I., Wong, B., & Potts, R. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0262 · Cited by 570
Integrated textile sensor patch for real-time and multiplex sweat analysis
He, W., Wang, C., Wang, H., Jian, M., Lu, W., Liang, X., Zhang, X., Yang, F., & Zhang, Y. (2019) · Science Advances · doi:10.1126/sciadv.aax0649 · Cited by 564
Biosensors in clinical chemistry
D'Orazio, P. (2003) · Clinica Chimica Acta · doi:10.1016/s0009-8981(03)00241-9 · Cited by 524
A contact lens with embedded sensor for monitoring tear glucose level
Yao, H., Shum, A. J., Cowan, M., Lähdesmäki, I., & Parviz, B. A. (2011) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.12.042 · Cited by 513
Bandodkar, A. J., Molinnus, D., Mirza, O., Guinovart, T., Windmiller, J. R., Valdés-Ramírez, G., Andrade, F. J., Schöning, M. J., & Wang, J. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.11.039 · Cited by 490
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475 · cited 1×
“of Copyright 2019, Elsevier), (b) Epidermal Microfluidic Electrochemical Detection System [55]”
BackgroundAdvances in non-enzymatic glucose sensors based on metal oxides
Zhu, H., Li, L., Zhou, W., Shao, Z., & Chen, X. (2016) · Journal of Materials Chemistry B · doi:10.1039/c6tb02037b · Cited by 473
Skin-like biosensor system via electrochemical channels for noninvasive blood glucose monitoring
Chen, Y., Lu, S., Zhang, S., Li, Y., Qu, Z., Chen, Y., Lu, B., Wang, X., & Feng, X. (2017) · Science Advances · doi:10.1126/sciadv.1701629 · Cited by 467
A wearable multisensing patch for continuous sweat monitoring
Anastasova, S., Crewther, B., Bembnowicz, P., Curto, V., Ip, H. M., Rosa, B., & Yang, G.-Z. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.09.038 · Cited by 443
Skin-Attachable, Stretchable Electrochemical Sweat Sensor for Glucose and pH Detection
Oh, S. Y., Hong, S. Y., Jeong, Y. R., Yun, J., Park, H., Jin, S. W., Lee, G., Oh, J. H., Lee, H., Lee, S.-S., et al. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b03342 · Cited by 440
Adhesive RFID Sensor Patch for Monitoring of Sweat Electrolytes
Rose, D. P., Ratterman, M. E., Griffin, D. K., Hou, L., Kelley-Loughnane, N., Naik, R. R., Hagen, J. A., Papautsky, I., & Heikenfeld, J. C. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2369991 · Cited by 416
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390 · cited 1×
“Included in the table are some studies for comparison’s sake that are minimally invasive [33,34], making use of ISF as biofluid for glucose detection.”
MethodWearable Sensors for Biochemical Sweat Analysis
Bandodkar, A. J., Jeang, W. J., Ghaffari, R., & Rogers, J. A. (2019) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061318-114910 · Cited by 387 · cited 1×
“…of the current produced (using enzyme-based sensing) from the redox reactions of the targeted analyte activated by constantly applied potentials [56,121] by using mediators such as ferric material and Prussian blue in between difficult-to-access redox centers in the enzyme to the electrode as…”
BackgroundProspects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy
Yadav, J., Rani, A., Singh, V., & Murari, B. M. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.01.005 · Cited by 374
A Glucose‐Selective Molecular Fluorescence Sensor
James, T. D., Sandanayake, K. R. A. S., & Shinkai, S. (1994) · Angewandte Chemie International Edition in English · doi:10.1002/anie.199422071 · Cited by 368
Choi, J., Bandodkar, A. J., Reeder, J. T., Ray, T. R., Turnquist, A., Kim, S. B., Nyberg, N., Hourlier-Fargette, A., Model, J. B., Aranyosi, A. J., et al. (2019) · ACS Sensors · doi:10.1021/acssensors.8b01218 · Cited by 365
Abellán-Llobregat, A., Jeerapan, I., Bandodkar, A., Vidal, L., Canals, A., Wang, J., & Morallón, E. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2017.01.058 · Cited by 348
A potentiometric tattoo sensor for monitoring ammonium in sweat
Guinovart, T., Bandodkar, A. J., Windmiller, J. R., Andrade, F. J., & Wang, J. (2013) · The Analyst · doi:10.1039/c3an01672b · Cited by 343 · cited 1×
“This technique has been further employed by De Guzman et al. for measuring the impedance and changes to skin barrier function [113] using sweat containing lactate [114], pH [110], and ammonium [115] that could aid in the management of diseases such as atopic dermatitis and psoriasis.”
MethodA short history of sweat gland biology
Wilke, K., Martin, A., Terstegen, L., & Biel, S. S. (2007) · International Journal of Cosmetic Science · doi:10.1111/j.1467-2494.2007.00387.x · Cited by 334
Yao, S., Ren, P., Song, R., Liu, Y., Huang, Q., Dong, J., O'Connor, B. T., & Zhu, Y. (2020) · Advanced Materials · doi:10.1002/adma.201902343 · Cited by 324
Eyeglasses based wireless electrolyte and metabolite sensor platform
Sempionatto, J. R., Nakagawa, T., Pavinatto, A., Mensah, S. T., Imani, S., Mercier, P., & Wang, J. (2017) · Lab on a Chip · doi:10.1039/c7lc00192d · Cited by 311
Bandage‐Based Wearable Potentiometric Sensor for Monitoring Wound pH
Guinovart, T., Valdés‐Ramírez, G., Windmiller, J. R., Andrade, F. J., & Wang, J. (2014) · Electroanalysis · doi:10.1002/elan.201300558 · Cited by 308
Noninvasive Blood Glucose Monitoring With Optical Coherence Tomography
Larin, K. V., Eledrisi, M. S., Motamedi, M., & Esenaliev, R. O. (2002) · Diabetes Care · doi:10.2337/diacare.25.12.2263 · Cited by 293
Noninvasive Monitoring of Blood Glucose with Raman Spectroscopy
Pandey, R., Paidi, S. K., Valdez, T. A., Zhang, C., Spegazzini, N., Dasari, R. R., & Barman, I. (2017) · Accounts of Chemical Research · doi:10.1021/acs.accounts.6b00472 · Cited by 290
Functionalized carbon nanotubes and nanofibers for biosensing applications
Wang, J., & Lin, Y. (2008) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2008.05.009 · Cited by 289
Smart Sensor Systems for Wearable Electronic Devices
An, B. W., Shin, J. H., Kim, S.-Y., Kim, J., Ji, S., Park, J., Lee, Y., Jang, J., Park, Y.-G., Cho, E., et al. (2017) · Polymers · doi:10.3390/polym9080303 · Cited by 289
Injectable hydrogel microbeads for fluorescence-based in vivo continuous glucose monitoring
Shibata, H., Heo, Y. J., Okitsu, T., Matsunaga, Y., Kawanishi, T., & Takeuchi, S. (2010) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1006911107 · Cited by 287
A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration
Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D. (2010) · Analytical Methods · doi:10.1039/b9ay00184k · Cited by 286
Stretchable, Wireless Sensors and Functional Substrates for Epidermal Characterization of Sweat
Huang, X., Liu, Y., Chen, K., Shin, W., Lu, C., Kong, G., Patnaik, D., Lee, S., Cortes, J. F., & Rogers, J. A. (2014) · Small · doi:10.1002/smll.201400483 · Cited by 282
Promphet, N., Rattanawaleedirojn, P., Siralertmukul, K., Soatthiyanon, N., Potiyaraj, P., Thanawattano, C., Hinestroza, J. P., & Rodthongkum, N. (2019) · Talanta · doi:10.1016/j.talanta.2018.09.086 · Cited by 254
Transdermal monitoring of glucose and other analytes using ultrasound
Kost, J., Mitragotri, S., Gabbay, R. A., Pishko, M., & Langer, R. (2000) · Nature Medicine · doi:10.1038/73213 · Cited by 253
Shirreffs, S. M., & Maughan, R. J. (1997) · Journal of Applied Physiology · doi:10.1152/jappl.1997.82.1.336 · Cited by 251
Mini Review on Flexible and Wearable Electronics for Monitoring Human Health Information
Gu, Y., Zhang, T., Chen, H., Wang, F., Pu, Y., Gao, C., & Li, S. (2019) · Nanoscale Research Letters · doi:10.1186/s11671-019-3084-x · Cited by 247
Wearable temporary tattoo sensor for real-time trace metal monitoring in human sweat
Kim, J., de Araujo, W. R., Samek, I. A., Bandodkar, A. J., Jia, W., Brunetti, B., Paixão, T. R. L. C., & Wang, J. (2015) · Electrochemistry Communications · doi:10.1016/j.elecom.2014.11.024 · Cited by 245
Real-Time Electrochemical Monitoring: Toward Green Analytical Chemistry
Wang, J. (2002) · Accounts of Chemical Research · doi:10.1021/ar010066e · Cited by 234
Wearable and Implantable Electronics: Moving toward Precision Therapy
Song, Y., Min, J., & Gao, W. (2019) · ACS Nano · doi:10.1021/acsnano.9b08323 · Cited by 232
Evolution of Insulin Delivery Devices: From Syringes, Pens, and Pumps to DIY Artificial Pancreas
Kesavadev, J., Saboo, B., Krishna, M. B., & Krishnan, G. (2020) · Diabetes Therapy · doi:10.1007/s13300-020-00831-z · Cited by 221
Shu, Y., Su, T., Lu, Q., Shang, Z., Xu, Q., & Hu, X. (2021) · Analytical Chemistry · doi:10.1021/acs.analchem.1c04106 · Cited by 221
Bae, C. W., Toi, P. T., Kim, B. Y., Lee, W. I., Lee, H. B., Hanif, A., Lee, E. H., & Lee, N.-E. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.9b00848 · Cited by 215
Eccrine sweat gland development and sweat secretion
Cui, C., & Schlessinger, D. (2015) · Experimental Dermatology · doi:10.1111/exd.12773 · Cited by 213
Potentiometric sensors using cotton yarns, carbon nanotubes and polymeric membranes
Guinovart, T., Parrilla, M., Crespo, G. A., Rius, F. X., & Andrade, F. J. (2013) · The Analyst · doi:10.1039/c3an00710c · Cited by 208
Trends in Nanomaterial-Based Non-Invasive Diabetes Sensing Technologies
Makaram, P., Owens, D., & Aceros, J. (2014) · Diagnostics · doi:10.3390/diagnostics4020027 · Cited by 200
Low-cost, μm-thick, tape-free electronic tattoo sensors with minimized motion and sweat artifacts
Wang, Y., Qiu, Y., Ameri, S. K., Jang, H., Dai, Z., Huang, Y., & Lu, N. (2018) · npj Flexible Electronics · doi:10.1038/s41528-017-0019-4 · Cited by 192
Radio-Frequency and Microwave Techniques for Non-Invasive Measurement of Blood Glucose Levels
Yilmaz, T., Foster, R., & Hao, Y. (2019) · Diagnostics · doi:10.3390/diagnostics9010006 · Cited by 192
Pu, Z., Zhang, X., Yu, H., Tu, J., Chen, H., Liu, Y., Su, X., Wang, R., Zhang, L., & Li, D. (2021) · Science Advances · doi:10.1126/sciadv.abd0199 · Cited by 186
Naikoo, G. A., Salim, H., Hassan, I. U., Awan, T., Arshad, F., Pedram, M. Z., Ahmed, W., & Qurashi, A. (2021) · Frontiers in Chemistry · doi:10.3389/fchem.2021.748957 · Cited by 181
Ratiometric Fluorescent Nanohybrid for Noninvasive and Visual Monitoring of Sweat Glucose
Cui, Y., Duan, W., Jin, Y., Wo, F., Xi, F., & Wu, J. (2020) · ACS Sensors · doi:10.1021/acssensors.0c00718 · Cited by 177
Sweat composition in exercise and in heat
Verde, T., Shephard, R. J., Corey, P., & Moore, R. (1982) · Journal of Applied Physiology · doi:10.1152/jappl.1982.53.6.1540 · Cited by 174
Electrochemical sensing based on printable temporary transfer tattoos
Windmiller, J. R., Bandodkar, A. J., Valdés-Ramírez, G., Parkhomovsky, S., Martinez, A. G., & Wang, J. (2012) · Chemical Communications · doi:10.1039/c2cc32839a · Cited by 169
Kim, S. B., Zhang, Y., Won, S. M., Bandodkar, A. J., Sekine, Y., Xue, Y., Koo, J., Harshman, S. W., Martin, J. A., Park, J. M., et al. (2018) · Small · doi:10.1002/smll.201703334 · Cited by 167
Wang, Y., Wang, X., Lu, W., Yuan, Q., Zheng, Y., & Yao, B. (2019) · Talanta · doi:10.1016/j.talanta.2019.01.104 · Cited by 166
‘SWEATCH’: A Wearable Platform for Harvesting and Analysing Sweat Sodium Content
Glennon, T., O’Quigley, C., McCaul, M. V., Matzeu, G., Beirne, S., Wallace, G. G., Stroiescu, F., O'Mahoney, N., White, P., & Diamond, D. (2016) · Electroanalysis · doi:10.1002/elan.201600106 · Cited by 164
Toi, P. T., Trung, T. Q., Dang, T. M. L., Bae, C. W., & Lee, N. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b20583 · Cited by 162
Materials for Diabetes Therapeutics
Bratlie, K. M., York, R. L., Invernale, M. A., Langer, R., & Anderson, D. G. (2012) · Advanced Healthcare Materials · doi:10.1002/adhm.201200037 · Cited by 161
Sim, J. Y., Ahn, C.-G., Jeong, E.-J., & Kim, B. K. (2018) · Scientific Reports · doi:10.1038/s41598-018-19340-y · Cited by 161
Tehrani, F., & Bavarian, B. B. (2016) · Scientific Reports · doi:10.1038/srep27975 · Cited by 158
Multivariate calibration of NIR spectroscopic sensors for continuous glucose monitoring
Goodarzi, M., Sharma, S., Ramón, H., & Saeys, W. (2015) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2014.12.005 · Cited by 140
Human sweat metabolomics for lung cancer screening
Calderón-Santiago, M., Priego‐Capote, F., Turck, N., Robin, X., Gámez, B. J., Sanchez, J., & de Castro, M. D. L. (2015) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-015-8700-8 · Cited by 138
Fang, L., Liu, B., Liu, L., Li, Y., Huang, K., & Zhang, Q. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.08.032 · Cited by 135
Sweat testing for cocaine, codeine and metabolites by gas chromatography–mass spectrometry
Huestis, M. A., Oyler, J. M., Cone, E. J., Wstadik, A. T., Schoendorfer, D., & Joseph, R. E. (1999) · Journal of Chromatography B Biomedical Sciences and Applications · doi:10.1016/s0378-4347(99)00246-7 · Cited by 125
Exercise-related hypoglycemia in diabetes mellitus
Younk, L. M., Mikeladze, M., Tate, D., & Davis, S. N. (2011) · Expert Review of Endocrinology & Metabolism · doi:10.1586/eem.10.78 · Cited by 114
Recent progress in the fabrication of flexible materials for wearable sensors
Liu, H., Wang, L., Lin, G., & Feng, Y. (2021) · Biomaterials Science · doi:10.1039/d1bm01136g · Cited by 113
Review of non-invasive continuous glucose monitoring based on impedance spectroscopy
Huang, J., Zhang, Y., & Wu, J. J. (2020) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2020.112103 · Cited by 109
A Lab-on-a-Chip-Based Non-Invasive Optical Sensor for Measuring Glucose in Saliva
Jung, D., Jung, D., & Kong, S. (2017) · Sensors · doi:10.3390/s17112607 · Cited by 107
Delbeck, S., Vahlsing, T., Leonhardt, S. K., Steiner, G. R., & Heise, H. M. (2018) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1395-x · Cited by 99
AUTOMATION CONTROL OF BLOOD SUGAR. I. A SERVOMECHANISM FOR GLUCOSE MONITORING AND CONTROL.
Kadish, A. H. (1964) · PubMed · Cited by 94
Wearable, robust, non-enzymatic continuous glucose monitoring system and its in vivo investigation
Yoon, H., Xuan, X., Jeong, S., & Park, J. Y. (2018) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2018.06.008 · Cited by 92
A modified anaerobic method of sweat collection
Boysen, T. C., Yanagawa, S., Sato, F., & Sato, K. (1984) · Journal of Applied Physiology · doi:10.1152/jappl.1984.56.5.1302 · Cited by 91
Wearable piezoelectric mass sensor based on pH sensitive hydrogels for sweat pH monitoring
Scarpa, E., Mastronardi, V. M., Guido, F., Algieri, L., Qualtieri, A., Fiammengo, R., Rizzi, F., & De Vittorio, M. (2020) · Scientific Reports · doi:10.1038/s41598-020-67706-y · Cited by 90
Eccrine Sweat as a Biofluid for Profiling Immune Biomarkers
Katchman, B. A., Zhu, M., Blain Christen, J., & Anderson, K. S. (2018) · PROTEOMICS – Clinical Applications · doi:10.1002/prca.201800010 · Cited by 86
Badugu, R., Lakowicz, J. R., & Geddes, C. D. (2004) · Talanta · doi:10.1016/j.talanta.2004.08.003 · Cited by 86
Wearable sweat sensing for prolonged, semicontinuous, and nonobtrusive health monitoring
Moonen, E. J. M., Haakma, J. R., Peri, E., Pelssers, E., Mischi, M., & den Toonder, J. M. J. (2020) · VIEW · doi:10.1002/viw.20200077 · Cited by 86
Mechanisms underlying the age-related decrement in the human sweating response
Inoue, Y., Shibasaki, M., Ueda, H., & Ishizashi, H. (1999) · European Journal of Applied Physiology · doi:10.1007/s004210050485 · Cited by 85
Iron, copper and zinc concentrations in human sweat and plasma; the effect of exercise
Aruoma, O. I., Reilly, T., MacLaren, D., & Halliwell, B. (1988) · Clinica Chimica Acta · doi:10.1016/0009-8981(88)90310-5 · Cited by 83
Wearable Fluid Capture Devices for Electrochemical Sensing of Sweat
Li, G., Mo, X., Law, W., & Chan, K. C. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b17419 · Cited by 83
Thomson, M. L. (1954) · The Journal of Physiology · doi:10.1113/jphysiol.1954.sp005045 · Cited by 80
Yu, M., Li, Y., Hu, Y., Tang, L., Yang, F., Lv, W.-L., Zhang, Z., & Zhang, G. (2021) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2021.115029 · Cited by 80
Stamp transfer electrodes for electrochemical sensing on non-planar and oversized surfaces
Windmiller, J. R., Bandodkar, A. J., Parkhomovsky, S., & Wang, J. (2012) · The Analyst · doi:10.1039/c2an35041f · Cited by 76
Regional and Total Body Sweat Composition of Men Fed Controlled Diets
Costa, F., GALLOWAY, D. H., & Margen, S. (1969) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/22.1.52 · Cited by 75
Optical sensors for continuous glucose monitoring
Yu, Z., Jiang, N., Kazarian, S. G., Taşoğlu, S., & Yetisen, A. K. (2021) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/abe6f8 · Cited by 73
Role of Continuous Glucose Monitoring in Diabetes Treatment
Hirsch, I., Battelino, T., Peters, A., Chamberlain, J., Aleppo, G., & Bergenstal, R. (2018) · American Diabetes Association eBooks · doi:10.2337/db20181 · Cited by 73
Wang, C., Shim, E., Chang, H., Lee, N., Kim, H. R., & Park, J. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112652 · Cited by 69
Baker, L. B., Ungaro, C. T., Barnes, K. A., Nuccio, R. P., Reimel, A. J., & Stofan, J. R. (2014) · Physiological Reports · doi:10.14814/phy2.12007 · Cited by 67
Poongodi, T., Rathee, A., Indrakumari, R., & Suresh, P. (2019) · Intelligent systems reference library · doi:10.1007/978-3-030-33596-0_5 · Cited by 67
A Wearable Device for Monitoring Sweat Rates via Image Analysis
Matzeu, G., Fay, C., Vaillant, A., Coyle, S. M., & Diamond, D. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2477676 · Cited by 66
Monitoring of glucose permeability in monkey skin in vivo using Optical Coherence Tomography
Ghosn, M. G., Sudheendran, N., Wendt, M., Glasser, A., Tuchin, V. V., & Larin, K. V. (2009) · Journal of Biophotonics · doi:10.1002/jbio.200910075 · Cited by 63
Comparison of Tear Lipid Profile among Basal, Reflex, and Flush Tear Samples
Rohit, A., Stapleton, F. J., Brown, S. H. J., et al. (2014) · Optometry and Vision Science · doi:10.1097/opx.0000000000000411 · Cited by 62
Deiss, D., Szadkowska, A., Gordon, D., Mallipedhi, A., Schütz-Fuhrmann, I., Aguilera, E., Ringsell, C., Block, C. E. M. D., & Irace, C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0397 · Cited by 62
Glucose detection in human sweat using an electronic nose
Olarte, O., Chilo, J., Pelegrí-Sebastiá, J., Barbé, K., & Moer, W. V. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609787 · Cited by 60
Modeling Glucose Transport From Systemic Circulation to Sweat
La Count, T. D., Jajack, A., Heikenfeld, J., & Kasting, G. B. (2019) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2018.09.026 · Cited by 60
Yuan, Y., Wang, Y., Wang, H., & Hou, S. (2019) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2019.113495 · Cited by 60
Portable and wearable real-time stress monitoring: A critical review
Parlak, O. (2021) · Sensors and Actuators Reports · doi:10.1016/j.snr.2021.100036 · Cited by 60 · cited 1×
“For example, cortisol sensors [42] are now used particularly for a few days to help measure human stress levels [43].”
BackgroundPeripheral modification of thermoregulatory function during heat acclimation.
Chen, W. Y., & Elizondo, R. S. (1974) · Journal of Applied Physiology · doi:10.1152/jappl.1974.37.3.367 · Cited by 59
Weschler, L. B. (2008) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00924.2007 · Cited by 57
The Composition of Human Perspiration (Samuel Hyde Memorial Lecture)
McSwiney, B. A. (1934) · Proceedings of the Royal Society of Medicine · doi:10.1177/003591573402700710 · Cited by 56
Nanoduct® sweat testing for rapid diagnosis in newborns, infants and children with cystic fibrosis
Desax, M.-C., Ammann, R. A., Hammer, J., Schoeni, M. H., & Barben, J. (2007) · European Journal of Pediatrics · doi:10.1007/s00431-007-0485-0 · Cited by 56
Pfützner, A., Caduff, A., Larbig, M., Schrepfer, T., & Forst, T. M. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/1520915041705839 · Cited by 55
Waterhouse, J. M., Aizawa, S., Nevill, A. M., Edwards, B., Weinert, D., Atkinson, G., & Reilly, T. P. (2007) · Chronobiology International · doi:10.1080/07420520601142551 · Cited by 53
Metal oxide-based composite for non-enzymatic glucose sensors
Liu, S., Zeng, W., Guo, Q. K., & Li, Y. (2020) · Journal of Materials Science Materials in Electronics · doi:10.1007/s10854-020-04239-0 · Cited by 52
Construction of highly accessible single Co site catalyst for glucose detection
Xiong, C., Tian, L., Xiao, C., Xue, Z., Zhou, F., Zhou, H., Zhao, Y., Chen, M., Wang, Q., Qu, Y., et al. (2020) · 中国科学通报:英文版 · doi:10.1016/j.scib.2020.07.031 · Cited by 50
The Evolution of Non-invasive Blood Glucose Monitoring System for Personal Application
Salam, N. A. A., Saad, W. H. M., Manap, Z., & Salehuddin, F. (2016) · Journal of Telecommunication Electronic and Computer Engineering (JTEC) · Cited by 41
A review of wearable technology: Moving beyond the hype: From need through sensor implementation
Seshadri, D. R., Rowbottom, J. R., Drummond, C. K., Voos, J. E., & Craker, J. (2016) · Cairo International Biomedical Engineering Conference · doi:10.1109/cibec.2016.7836118 · Cited by 37
Chen, X., Gąsecka, P., Formanek, F., Galey, J.-B., & Rigneault, H. (2015) · British Journal of Dermatology · doi:10.1111/bjd.14292 · Cited by 35
Amano, T., Gerrett, N., Inoue, Y., Nishiyasu, T., Havenith, G., & Kondo, N. (2015) · European Journal of Applied Physiology · doi:10.1007/s00421-015-3275-9 · Cited by 34
Morzel, M., Truntzer, C., Neyraud, E., Brignot, H., Ducoroy, P., Lucchi, G., Canlet, C., Gaillard, S., Nicod, F., Nicklaus, S., et al. (2017) · Physiology & Behavior · doi:10.1016/j.physbeh.2017.02.005 · Cited by 34
Screen‐printed Tattoo Sensor towards the Non‐invasive Assessment of the Skin Barrier
Guzman, K. D., & Morrin, A. (2016) · Electroanalysis · doi:10.1002/elan.201600572 · Cited by 33
Intra‐operative blood glucose measurements: The effect of haematocrit on glucose test strips
Smith, E. A., & Kilpatrick, E. S. (1994) · Anaesthesia · doi:10.1111/j.1365-2044.1994.tb03369.x · Cited by 28
Heikenfeld, J. C. (2014) · IEEE Spectrum · doi:10.1109/mspec.2014.6934933 · Cited by 28
In vivo calibration of the subcutaneous amperometric glucose sensors using a non-enzyme electrode
Jeong, R.-A., Hwang, J. Y., Joo, S., Chung, T. D., Park, S., Kang, S. K., Lee, W., & Kim, H. C. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00219-7 · Cited by 28
Mugweru, A. M., Clark, B. L., & Pishko, M. V. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100308 · Cited by 26
Zheng, X. T., Choi, Y., Phua, G. G. D., & Tan, Y. N. (2020) · Bioconjugate Chemistry · doi:10.1021/acs.bioconjchem.9b00856 · Cited by 21
CHEMISTRY OF PALMAR SWEAT: III. Reducing Substances (Glucose)
Lobitz, W. C., & OSTERBERG, A. E. (1947) · Archives of Dermatology and Syphilology · doi:10.1001/archderm.1947.01520120107012 · Cited by 19
Data Reliability and Quality in Body Area Networks for Diabetes Monitoring
Huzooree, G., Khedo, K. K., & Joonas, N. (2018) · EAI/Springer Innovations in Communication and Computing · doi:10.1007/978-3-030-00865-9_4 · Cited by 15
RET nanobiosensors using affinity of an apo-enzyme toward its substrate
Chinnayelka, S., & McShane, M. J. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1403747 · Cited by 7
Tiller, B. P. (2016) · Enlighten: Theses (The University of Glasgow) · Cited by 2
Cited by
Ramesh, M., Janani, R., Deepa, C., & Rajeshkumar, L. (2022) · Biosensors · doi:10.3390/bios13010040 · Cited by 443
Wearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442
Nanomaterials-based biosensor and their applications: A review
Malik, S., Singh, J., Goyat, R., Saharan, Y., Chaudhry, V., Umar, A., Ibrahim, A. A., Akbar, S., Ameen, S., & Baskoutas, S. (2023) · Heliyon · doi:10.1016/j.heliyon.2023.e19929 · Cited by 441
Theyagarajan, K., & Kim, Y.-J. (2024) · Biosensors · doi:10.3390/bios14100492 · Cited by 31
Continuous glucose monitoring for children with hypoglycaemia: Evidence in 2023
Worth, C., Hoskyns, L., Salomon-Estebanez, M., Nutter, P. W., Harper, S., Derks, T. G. J., Beardsall, K., & Banerjee, I. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1116864 · Cited by 29
Metadata
- Authors
- Zafar, H., Channa, A., Jeoti, V., & Stojanović, G. M.
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2022-01-14
- Volume
- 22
- Issue
- 2
- Pages
- 638
- ISSN
- 1424-8220 (ISSN-L)
- OpenAlex
- W4205830475
- Cited by
- 329
- References
- 141
- FWCI
- 18.15
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.