Wearable Photoplethysmography for Cardiovascular Monitoring
Open accessAbstract
Smart wearables provide an opportunity to monitor health in daily life and are emerging as potential tools for detecting cardiovascular disease (CVD). Wearables such as fitness bands and smartwatches routinely monitor the photoplethysmogram signal, an optical measure of the arterial pulse wave that is strongly influenced by the heart and blood vessels. In this survey, we summarize the fundamentals of wearable photoplethysmography and its analysis, identify its potential clinical applications, and outline pressing directions for future research in order to realize its full potential for tackling CVD.
Overview
AI snapshot
Written from the paper’s full text.
TL;DR
Wearable photoplethysmography (PPG) is widely used for heart-rate monitoring and is being developed for broader cardiovascular monitoring, including blood pressure, respiratory rate, oxygen saturation, atrial fibrillation, and vascular risk. This narrative review summarizes PPG fundamentals, signal processing, clinical applications, and research directions, concluding that accuracy, validation, and clinical integration remain key challenges.
Key findings
- Wearable PPG can already estimate HR, but accuracy falls during exercise: summary MAE 2.15 bpm at rest vs 7.70 bpm during treadmill exercise and 10.64 bpm during cycling.
- In a community smartwatch study, possible AF notifications from 1-min PPG recordings had a PPV of 0.84; sensitivity and specificity were not assessed.
- Computational modeling of PPG in healthy subjects aged 25–75 found stiffness index increased with age and d/a decreased with age; both indices were predominantly influenced by arterial properties (large artery diameter, PWV, peripheral vascular resistance), with SI also strongly influenced by stroke volume and d/a by HR.
- The review identifies AF detection as perhaps nearest to clinical use, while BP, RR, SpO2, arterial stiffness, sleep, stress, sepsis, HF, and preeclampsia applications need further validation.
Methods
- Narrative review summarizing key literature on wearable PPG fundamentals, signal processing, physiological parameter estimation, clinical applications, future research, and resources; includes tables of further reading, PPG datasets, and wearable devices.
- No systematic search, meta-analysis, or formal risk-of-bias assessment is described.
- A computational model simulated PPG for healthy subjects aged 25–75 under varied cardiovascular conditions using the PWDB; SI and d/a were extracted and assessed via relative sensitivity to cardiovascular properties.
Limitations
- The paper states it is a narrative review; it notes that mechanisms underlying PPG pulse wave shape are not fully understood, it is not clear which pulse-wave detection approaches perform best, and many clinical applications require further validation.
- The paper notes that most wearable BP devices are not validated, and SpO2 measurements from conventional pulse oximeters have been found less accurate and less reliable in black patients than white patients.
- Evident from the text: the modeling findings use synthetic/simulated data from healthy adults aged 25–75, not daily-life wearable recordings, and the review does not provide a systematic synthesis of evidence quality.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 of 172 cited sources checked · 0 supported · 0 flagged
None of the sources cited for this claim have been checked yet.
In the paper
“Many datasets contain finger PPG signals, whereas only a few contain wrist PPG signals: pulse wave morphology may differ between these sites [58].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Resources > PPG Datasets
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesNone of the sources cited for this claim have been checked yet.
In the paper
“Furthermore, very few datasets have been acquired in daily life conditions (the PPG-DaLiA Dataset is a notable exception [295]), making it difficult to compare wearable PPG analysis techniques [28], [72].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Resources > PPG Datasets
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Both indices were predominantly influenced by arterial properties: large artery diameter, PWV, and peripheral vascular resistance.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Directions > Understanding the Determinants of the PPG Pulse Wave
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIn the paper
“The shape is influenced not only by the incident wave transmitted through the arteries from the heart but also by reflected waves from a range of arterial sites [6].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [6] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesThe paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The shape of a PPG pulse wave is determined primarily by the heart and blood vessels.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear? Based: unclearIn the paper
“The configuration of a PPG sensor has a large impact on the PPG signal quality.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
- Not checked yetcites [55] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIn the paper
“Breathing also influences the PPG signal at subcardiac frequencies in three ways (see the Supplementary Material): amplitude modulation, frequency modulation, and baseline wander [24].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [24] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIn the paper
“Key considerations include [55] the geometry of LED and photodiode positioning, since configurations in which LEDs surround the photodiode result in higher PPG signal amplitudes, the spacing between an LED and the photodiode, since a shorter spacing results in higher amplitudes, and the use of an optical barrier between the LED(s) and photodiode, which can increase the signal amplitude.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
- Not checked yetcites [55] source
Strength:plain· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIn the paper
“Key considerations include [55] the geometry of LED and photodiode positioning, since configurations in which LEDs surround the photodiode result in higher PPG signal amplitudes, the spacing between an LED and the photodiode, since a shorter spacing results in higher amplitudes, and the use of an optical barrier between the LED(s) and photodiode, which can increase the signal amplitude.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
- Not checked yetcites [55] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIn the paper
“Finally, the PPG can be affected by changes in the height of measurement relative to the heart [53].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [53] source
Strength:plain· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passes
- It may state two separate claims as one. 62
- Probably not a substantial claim. 44
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“several CV properties were found to have similar effects on the pulse wave: changes in arterial stiffness, mean BP, and peripheral compliance all impacted the portion of the pulse wave between systolic and diastolic peaks; HR and arterial diameter impacted the height of the diastolic peak.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Directions > Understanding the Determinants of the PPG Pulse Wave
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“The PPG signal is susceptible to motion artifact, as shown in Fig. 1(e), and good skin contact is required to obtain high-quality signals [7].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [7] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The pulsatile portion of the PPG is primarily determined by the volume of arterial blood in the tissue bed, which increases and decreases with each heartbeat.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“The mechanisms underlying the PPG pulse wave shape are not yet fully understood [6].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [6] source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“In this case, the pulsatile component of the PPG can be of the order of 1 mV in amplitude [55] although lower amplitudes can be caused by technical factors, such as inappropriate gain control, or physiological factors, such as vasoconstriction [9].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Consequently, signal processing techniques have been developed to extract additional physiological parameters from the PPG.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Introduction
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“If these techniques could be refined and validated for use in daily life, then potentially wearables could be used for CV monitoring and to inform clinical decisions.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Introduction
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
In the paper
“As shown in Fig. 1(d), it consists of a baseline component and a pulsatile portion.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [47] source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The choice of PPG measurement site influences both user acceptability and the usefulness of PPG signals.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Photoplethysmography has already had a profound impact on clinical care through its use in pulse oximeters, which are routinely used to assess blood oxygen saturation in a wide range of clinical settings.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Introduction
What the sources say
- Not checked yetcites [35] source
Strength:boosted· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The PPG also contains information on the cardiac, vascular, respiratory, and autonomic nervous systems.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Introduction
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds.
In the paper
“While there are similarities between PPG and arterial blood pressure (BP) pulse waves, their shapes differ [54] because the signals are produced by different physiological mechanisms.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
- Not checked yetcites [54] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This may be because the upper wrist PPG is dominated by microvascular blood, whereas the finger PPG is more strongly influenced by blood flow in the digital arteries [see Fig. 1(a) and (c)].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Other factors that influence the PPG include the temperature [48], [49], the use of reflected or transmitted light, and the anatomical measurement site [50].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
What the sources say
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Fluctuations in venous blood volume also affect the signal.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“There is no widely accepted measurement unit for the PPG signal.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > PPG Signal
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“Approaches to identify individuals at risk of CVD could prompt these interventions and help reduce CVD-associated mortality and morbidity.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Introduction
What the sources say
- Not checked yetcites [2] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
In the paper
“The photoplethysmogram (PPG) signal is a measure of arterial blood volume, which fluctuates with each heartbeat and is used by many wearables to monitor heart rate (HR).”
- Also stated in: § PPG Signal · plain
“The shape of a PPG pulse wave is determined primarily by the heart and blood vessels.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Introduction
What the sources say
- Not checked yetcites [6] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
- Also stated in: § PPG Signal · plain
None of the sources cited for this claim have been checked yet.
In the paper
“These alternative sites can have advantages over the wrist, such as being less prone to motion artifact [63] or vasoconstriction [62], and providing more accurate HR measurements [64].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“The attractor’s shape is determined by the shape of the pulse waves, and its density is determined by variability in pulse wave shape [117], [118].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Extracting Pulse Wave Features
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Approaches to identify individual pulse waves in the PPG seek to overcome two key challenges: 1) pulse waves can exhibit two peaks, particularly in young subjects [see Fig. 1(g)] and 2) artifacts can create spurious peaks [see Fig. 1(e)].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Identifying Individual Pulse Waves
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“A low-pass cutoff frequency as low as 2.25 Hz may be suitable for HR estimation [75], whereas higher cutoff frequencies are required for analyses of pulse wave timing (such as 5 Hz for interbeat interval (IBI) calculation [76]) and pulse wave shape (such as 20 Hz for feature extraction [77]) [11].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
What the sources say
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Similarly, novel techniques have been developed to determine whether derived RRs are accurate [89].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
What the sources say
- Not checked yetcites [89] source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Derivatives are typically derived from the PPG after eliminating high-frequency noise, which can otherwise greatly obscure the derivatives.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Extracting Pulse Wave Features
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“Reference motion signals that can be used include accelerometry [81], [82], the gyroscope signal [83], or a PPG signal of a different wavelength (such as using an infrared PPG to remove motion artifact from a green PPG) [84].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
What the sources say
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Second, the contact pressure resulting from the attachment can influence pulse wave shape and timing, meaning that it should ideally be kept constant or calibrated when performing analyses using these signal characteristics [70].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
- Not checked yetcites [70] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Some approaches, such as analyzing the local maxima scalogram, use minimal heuristic information, which may make them more suitable for use across settings, although heuristics may be useful for ensuring high performance in particular settings.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Identifying Individual Pulse Waves
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“It is important to only use high-quality signals for pulse wave analysis, as pulse wave features can be distorted in the presence of noise.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Extracting Pulse Wave Features
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it. The paper may give no reason why it holds. It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“It is not yet clear which approaches perform the best nor whether a single approach can be used for all patient groups and recording settings.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Identifying Individual Pulse Waves
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“It is important to select a threshold suitable for the intended application, as a very high-quality signal might be required for analysis of pulse wave shape, whereas a lower quality signal could be acceptable for HR monitoring.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
Strength:hedged· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it. It may state two separate claims as one.
In the paper
“For HR estimation, the high-pass cutoff should be less than the HR (e.g., 0.4 Hz [75], ensuring that even the lowest plausible HR of 30 bpm is preserved).”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
What the sources say
- Not checked yetcites [75] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“In contrast, green is often used in reflectance mode, resulting in higher quality PPG signals [67], greater robustness to temperature changes [48], and more accurate HR monitoring [68].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Longer wavelengths penetrate to deeper depths [47] and are, thus, influenced by different levels of the vasculature.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
What the sources say
- Not checked yetcites [47] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Typically, signal segments are deemed to be either high or low quality through the use of empirical thresholds.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“Several approaches have been proposed, including statistical analysis of pulse wave shape [56], assessing the level of perfusion through the perfusion index (the ratio of the amplitude of the pulsatile component of the PPG to its baseline) [56], assessing the similarity of successive pulse wave shapes using template matching [86] or dynamic time warping [87], and deep learning [88].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
What the sources say
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Individual pulse waves must be identified in order to obtain IBIs (the time between consecutive pulse waves) and extract pulse wave features from individual waves.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Identifying Individual Pulse Waves
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds.
In the paper
“Motion artifact removal is particularly helpful for facilitating HR monitoring during exercise [13] although, in many situations, it is preferable to only estimate parameters from high-quality signals to increase accuracy.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Preprocessing
What the sources say
- Not checked yetcites [13] source
Strength:plain· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“There are also benefits to acquiring other signals simultaneously with the PPG, such as the electrocardiogram (ECG) (for pulse arrival time (PAT) measurement), and an accelerometry, gyroscope, or second PPG signal (for motion artifact removal—see Section III-A2).”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Photoplethysmography > Wearable PPG Device Design
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“SpO2 can be measured from two PPG signals acquired using different wavelengths of light”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“There is a fundamental difference between HRV and PRV: HRV is obtained from the timings of electrical impulses causing ventricular contraction, whereas PRV is obtained from the timings of pulse waves arriving at the periphery.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“HRs have been found to be acceptably accurate during rest [e.g., summary mean absolute error (MAE) of 2.15 beats per minute (bpm)] although they are less accurate during exercise (e.g., MAE of 7.70 bpm during treadmill exercise and 10.64 bpm during cycling) [14].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [14] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Often machine learning is used to create a model relating pulse wave features to BP”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [23] source
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“HR is already widely measured by wearable PPG devices.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. The paper’s conclusions may not depend on it. The paper may give no reason why it holds.
None of the sources cited for this claim have been checked yet.
In the paper
“Several models relating BP to PTT or PAT have been proposed [156], including models that incorporate additional variables, such as HR [157].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
In the paper
“SpO2 measurement is now being incorporated into smart wearables [139], requiring two PPG signals at different wavelengths.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [139] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“BP has been estimated from the PPG using the following inputs [26]: a single PPG [19]; a proximal and a distal PPG to measure PPT [21], [22]; or a distal PPG signal and a signal indicating the time of ventricular contraction (e.g., ECG, phonocardiogram) to measure PAT [20], [21], [150].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“An alternative approach proposed in [138] has been found to provide greater accuracy when using low-quality PPG signals, although at greater computational cost”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Potential explanations for this include [17]: 1) motion artifact in the PPG rendering PPG-derived IBIs less accurate; 2) a lack of motion artifact removal in PRV studies; and 3) variability in preejection period and PPT causing additional variability in PPG-derived IBIs.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [17] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“The time difference between the electrical impulses and pulse wave arrival consists of the preejection period (the time between the impulses and ejection of blood into the aorta) and the pulse transit time (PPT) (the time taken for the pulse wave to propagate from the heart to the measurement site).”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Both of these components are variable, potentially causing differences between HRV and PRV.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“they have been found to be less accurate and less reliable in black patients than white patients”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [144]–[147]
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“However, they usually cannot be considered as a replacement for clinical tests often due to their inferior performance.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds.
None of the sources cited for this claim have been checked yet.
In the paper
“The volume-clamp method is an alternative PPG-based approach for measuring BP”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Deep learning approaches have the following advantages over traditional feature extraction approaches: 1) they are hypothesis-free, i.e., there is no need to hypothesize the pulse wave features of interest and 2) they do not require pulse wave analysis algorithms”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Developing Models to Analyze PPG Signals
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“The PPG has also been used to assess perfusion [179], [180], hemoglobin concentration [181], and haemorheology [182].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“it may be possible to identify changes in BP from a single PPG without calibration [159], which could have utility in detecting clinical deteriorations, such as sepsis.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [159] source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Since it is related to BP, it can be assessed using similar techniques to those used to estimate BP.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Therefore, the potential role of PPG-based approaches is likely to be in the early detection of CVD and ubiquitous measurement of risk factors, which could trigger a further clinical assessment.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications
Tests✓ Asserted: passes? Checkable: unclear✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“However, it can be difficult to identify AF as it may not produce symptoms and may occur only intermittently”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Detecting Atrial Fibrillation
What the sources say
- Not checked yetcites [190] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“The performance of techniques for detecting AF from the PPG has been assessed in several studies [76], [193]–[196] although further work is required to determine which approach is most suitable for clinical use”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Detecting Atrial Fibrillation
What the sources say
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Standards have been developed for the validation of wearable, cuffless BP devices [168], [169] although further work is required to refine validation standards to account for the issues presented by PPG-based devices [170].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
In the paper
“While frequent calibration may be necessary to accurately estimate absolute BP values from a single PPG, less frequent calibration may be required when estimating BP from certain PAT measurements”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [158] source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
In the paper
“Statistical modeling and machine learning can be used to develop models to analyze PPG signals”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Developing Models to Analyze PPG Signals
What the sources say
- Not checked yetcites [23] source
Strength:plain· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“Consequently, it is often not recognized in clinical practice [191], where gold-standard ECG assessments may not be indicated due to a lack of symptoms, and intermittent episodes of AF may be missed as 12-lead ECG assessments are usually conducted at a single time point in a clinic, rather than over several days during daily life.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Detecting Atrial Fibrillation
What the sources say
- Not checked yetcites [191] 2015
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Most devices are not validated [163] although some have recently been certified for medical use [164]–[166].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > PPG Signal Processing > Estimating Physiological Parameters
What the sources say
- Not checked yetcites [163] source
- Not checked yetcites [164]–[166]
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The potential utility of wearable PPG in these applications lies in its convenience: existing clinical tests can be invasive and more time-consuming, are conducted at a specific time rather than continuously in daily life, and often require a clinical operator and more expensive equipment.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“The rate of HR recovery (HRR) after exercise has been associated with CV events and mortality [244], and postoperative morbidity [245].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Response to Exercise
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“While it has been observed that stress tends to increase HR and increase PRV”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Mental Stress
What the sources say
- Not checked yetcites [230] source
Strength:hedged· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“HR, RR, and SpO2 have previously been derived from the PPG to identify deteriorations in health [239], including changes prior to cardiac arrests [46].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Identifying Clinical Deteriorations
What the sources say
- Not checked yetcites [239] source
- Not checked yetcites [46] Charlton, 2017
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“a desaturation of 4% is often used as an (imperfect) indicator of apnea”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Identifying Obstructive Sleep Apnea
What the sources say
- Not checked yetcites [207] source
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“The positive predictive value (PPV) of notifications was found to be 0.84, indicating that wearables may have utility for identifying possible AF.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Detecting Atrial Fibrillation
What the sources say
- Not checked yetcites [29] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“However, it is difficult to measure stress frequently in everyday life: salivary cortisol swab tests are often used as a reference marker of stress”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Mental Stress
What the sources say
- Not checked yetcites [229] source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“The utility of PRV for assessing stress has been investigated in several studies [230]–[232], including a study of PRV metrics derived from short recordings [233].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Mental Stress
What the sources say
- Not checked yetcites [230]–[232]
- Not checked yetcites [233] source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“It has been proposed that the PPG could be used to assess vascular age”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Vascular Age
Strength:hedged· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“Three broad approaches have been proposed to identify apnea events from the PPG.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Identifying Obstructive Sleep Apnea
What the sources say
- Not checked yetcites [206]–[211]
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Studies have demonstrated the potential utility of smartwatch data for improving detection of COVID-19 beyond that provided by symptom data alone [215] and for presymptomatic detection of COVID-19 [216].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Monitoring the Spread of Infectious Diseases
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“Some of these indices have been found to be independent predictors of CV risk and mortality [33], [238], demonstrating their potential utility for risk prediction.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Vascular Age
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
In the paper
“Vascular age may have particular utility for CV risk prediction since aortic PWV has been found to be predictive of CV events and all-cause mortality”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Vascular Age
What the sources say
- Not checked yetcites [171] source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This provides an incentive for using wearables to monitor stress and assist with stress management.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Assessing Mental Stress
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it.
In the paper
“PPG-based wearables can also facilitate remote physiological monitoring in the acute care setting, reducing the staff contact required to monitor patients with infectious diseases, such as COVID-19”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Identifying Clinical Deteriorations
What the sources say
- Not checked yetcites [240] Wehrly, 2019
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There are several approaches to integrate wearable PPG-based devices into clinical pathways, each of which has different performance requirements.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Conclusion
Strength:plain· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it. The paper may give no reason why it holds. It may state two separate claims as one.
In the paper
“pulse wave morphology may differ between these sites [58].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Resources > PPG Datasets
What the sources say
- Not checked yetcites [58] source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“In [29], an irregular pulse notification was generated if an irregular pulse was identified in at least five out of six PPG recordings in a 48-h period [29].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Directions > Acquiring PPG Data During Daily Living
What the sources say
- Not checked yetcites [29] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Even though photoplethysmography has been used clinically for 40 years in pulse oximeters, there is still much opportunity to exploit the technology for further benefit, aided by the widespread use of photoplethysmography in wearable devices.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Conclusion
Tests✓ Asserted: passes✗ Checkable: fails✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may be too vague to check against evidence. It may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Key signal processing challenges include the estimation of BP, RR, and arterial blood oxygen saturation (using reflectance photoplethysmography), each of which could be of great value clinically.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Conclusion
Load-bearingStrength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
In the paper
“The performance of these devices varies and can have a considerable impact on the success of a study, influencing the proportion of time for which data are captured, whether data capture is biased toward particular subjects, and the quality of data in different activities [72].”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Resources > Wearable Devices for Acquiring PPG Signals
What the sources say
- Not checked yetcites [72] source
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Further development of the technology will be aided by freely available datasets, particularly those acquired in daily living with reference labels, and open-source algorithms against which to compare new signal processing techniques.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Conclusion
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it. The paper may give no reason why it holds. It may state two separate claims as one.
In the paper
“it has been proposed that the PPG could be used to identify possible preeclampsia, particularly in resource-constrained settings”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Identifying Preeclampsia
What the sources say
- Not checked yetcites [264] source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The PPG signal contains a wealth of information on the CV system although there are several confounders that can either obscure this information (such as motion artifact) or alter it (such as contact pressure).”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Conclusion
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes? Based: unclearIt may state two separate claims as one.
In the paper
“Similarly, in [46], algorithms to detect clinical deteriorations in hospital patients generated an alert if at least 50% of the data acquired within a 30-min period were outside of normal ranges.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Research Directions > Acquiring PPG Data During Daily Living
What the sources say
- Not checked yetcites [46] Charlton, 2017
Strength:plain· not yet calibratedCausal:direct· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“the speed of HRR in response to standing is reduced in HF and is predictive of mortality”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Clinical Applications > Identifying Heart Failure
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The detection of an irregular pulse to prompt further assessment for AF is, perhaps, nearest to being ready for clinical use.”
§ Wearable Photoplethysmography for Cardiovascular Monitoring > Conclusion
Load-bearingStrength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds.
The paper relies on this without support.
A diagnosis of atrial fibrillation requires ECG verification.
“Currently, a diagnosis of AF requires ECG verification, so the potential role of PPG-based devices is limited to initial assessment.”
The paper relies on this without support.
Subsequent anticoagulant treatment for atrial fibrillation has the side effect of increased bleeding risk.
“This is particularly important in the case of AF because subsequent anticoagulant treatment has the side effect of increased bleeding risk.”
The paper relies on this without support.
A wide range of consumer devices is available with varying performance levels and no universal standards.
“Second, a wide range of consumer devices is available with varying performance levels and no universal standards.”
The paper relies on this without support.
At present, there is little evidence for the long-term health benefits of self-directed cardiovascular health monitoring.
“At present, there is little evidence for the long-term health benefits of this approach.”
The paper relies on this without support.
Even the best consumer devices may not perform sufficiently well to avoid excessive resource utilization.
“First, even the best devices may not perform sufficiently well to avoid excessive resource utilization.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Naturally, the signal quality of PPG data in general is often discussed in other work as it is highly susceptible to motion artifacts57, however, there is also a high variance in PPG signal quality across different devices67, such that a device-specific consideration seems appropriate.”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [57]
In the citing paper
“BVP data is dependent on measuring reflected light from a light-emitting diode (LED), and this signal can be heavily skewed by contamination from external light sources and motion artifacts, sometimes to the point of being irrecoverable through retrospective data processing40,57,58.”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [57]
In the citing paper
“underlining its high susceptibility to measurement artifacts57.”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [57]
In the citing paper
“For reflectance measurements used, for example, for heart rate monitoring, green light is usually used as this provides high quality PPG signals and is less affected by temperature changes and accuracy.132”
From Effect of skin color on optical properties and the implications for medical optical technologies: a review (2024) · cites this paper as [132]
Citations
Cites
2 retracted
Uçar, M. K., Bozkurt, M. R., Bilgin, C., & Polat, K. (2018) · Neural Computing and Applications · doi:10.1007/s00521-016-2365-x · Cited by 106
RetractedQawqzeh, Y. K., Bajahzar, A. S., Jemmali, M., Otoom, M. M., & Thaljaoui, A. (2020) · BioMed Research International · doi:10.1155/2020/3764653 · Cited by 71
RetractedMIMIC-III, a freely accessible critical care database
Johnson, A. E. W., Pollard, T. J., Shen, L., Lehman, L.-W. H., Feng, M., Ghassemi, M., Moody, B., Szolovits, P., Anthony Celi, L., & Mark, R. G. (2016) · Scientific Data · doi:10.1038/sdata.2016.35 · Cited by 8,141
Atrial fibrillation as an independent risk factor for stroke: the Framingham study
Wolf, P. A., Abbott, R. D., & Kannel, W. B. (1991) · Stroke · doi:10.1161/01.str.22.8.983 · Cited by 7,388
An overview of heart rate variability metrics and norms
Shaffer, F., & Ginsberg, J. P. (2017) · Frontiers in Public Health · doi:10.3389/fpubh.2017.00258 · Cited by 7,185
DEAP: A Database for Emotion Analysis ;Using Physiological Signals
Koelstra, S., Muhl, C., Soleymani, M., Lee, J.-S., Yazdani, A., Ebrahimi, T., Pun, T., Nijholt, A., & Patras, I. (2012) · IEEE Transactions on Affective Computing · doi:10.1109/t-affc.2011.15 · Cited by 5,093
Benjafield, A. V., Ayas, N. T., Eastwood, P. R., Heinzer, R., Ip, M. S. M., Morrell, M. J., Nunez, C. M., Patel, S. R., Penzel, T., Pépin, J.-L., et al. (2019) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(19)30198-5 · Cited by 4,326
Prediction of Cardiovascular Events and All-Cause Mortality With Arterial Stiffness
Vlachopoulos, C., Aznaouridis, K., & Stefanadis, C. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.10.061 · Cited by 4,165
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
Seymour, C. W., Liu, V. X., Iwashyna, T. J., Brunkhorst, F. M., Rea, T. D., Scherag, A., Rubenfeld, G., Kahn, J. M., Shankar-Hari, M., Singer, M., et al. (2016) · JAMA · doi:10.1001/jama.2016.0288 · Cited by 3,939
Global, Regional, and National Burden of Cardiovascular Diseases for 10 Causes, 1990 to 2015
Roth, G. A., Johnson, C., Abajobir, A., Abd-Allah, F., Abera, S. F., Abyu, G., Ahmed, M., Aksut, B., Alam, T., Alam, K., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.04.052 · Cited by 3,791
Yancy, C. W., Jessup, M., Bozkurt, B., Butler, J., Casey, D. E., Drazner, M. H., Fonarow, G. C., Geraci, S. A., Horwich, T., Januzzi, J. L., et al. (2013) · Circulation · doi:10.1161/cir.0b013e31829e8807 · Cited by 3,304
Global Public Health Burden of Heart Failure
Savarese, G., & Lund, L. H. (2017) · Cardiac failure review · doi:10.15420/cfr.2016:25:2 · Cited by 2,818
Anderson, R. R., & Parrish, J. A. (1981) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12479191 · Cited by 2,456
Stress and Heart Rate Variability: A Meta-Analysis and Review of the Literature
Kim, H.-G., Cheon, E.-J., Bai, D.-S., Lee, Y. H., & Koo, B.-H. (2018) · Psychiatry Investigation · doi:10.30773/pi.2017.08.17 · Cited by 2,308
O'Donnell, M. J., Chin, S. L., Rangarajan, S., Xavier, D., Liu, L., Zhang, H., Rao-Melacini, P., Zhang, X., Pais, P., Agapay, S., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)30506-2 · Cited by 2,296
Salivary cortisol as a biomarker in stress research
Hellhammer, D. H., Wüst, S., & Kudielka, B. M. (2009) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2008.10.026 · Cited by 1,925
Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation
Perez, M. V., Mahaffey, K. W., Hedlin, H., Rumsfeld, J. S., Garcia, A., Ferris, T., Balasubramanian, V., Russo, A. M., Rajmane, A., Cheung, L., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1901183 · Cited by 1,917
Bashkatov, A. N., Genina, E. A., Kochubey, V. I., & Tuchin, V. V. (2005) · Journal of Physics D: Applied Physics · doi:10.1088/0022-3727/38/15/004 · Cited by 1,720
Mol, B. W., Roberts, C. T., Thangaratinam, S., Magee, L. A., de Groot, C. J. M., & Hofmeyr, G. J. (2015) · The Lancet · doi:10.1016/s0140-6736(15)00070-7 · Cited by 1,667
Introducing WESAD, a Multimodal Dataset for Wearable Stress and Affect Detection
Schmidt, P., Reiss, A., Dürichen, R., Marberger, C., & Laerhoven, K. V. (2018) · INTERNATIONAL CONFERENCE ON MULTIMODAL INTERACTION · doi:10.1145/3242969.3242985 · Cited by 1,508
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206
Psychological Stress and Cardiovascular Disease
Dimsdale, J. E. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.12.024 · Cited by 1,117
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114
Marini, C., De Santis, F., Sacco, S., Russo, T., Olivieri, L., Totaro, R., & Carolei, A. (2005) · Stroke · doi:10.1161/01.str.0000166053.83476.4a · Cited by 949
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
Racial/Ethnic Differences in Sleep Disturbances: The Multi-Ethnic Study of Atherosclerosis (MESA)
Chen, X., Wang, R., Zee, P., Lutsey, P. L., Javaheri, S., Alcántara, C., Jackson, C. L., Williams, M. A., & Redline, S. (2015) · SLEEP · doi:10.5665/sleep.4732 · Cited by 922
Stress and Cardiovascular Disease: An Update on Current Knowledge
Steptoe, A., & Kivimäki, M. (2013) · Annual Review of Public Health · doi:10.1146/annurev-publhealth-031912-114452 · Cited by 879
Gray, H., Gray, H., & Lewis, W. H. (1918) · Lea & Febiger eBooks · doi:10.5962/bhl.title.20311 · Cited by 839
Zhang, Z., Pi, Z., & Liu, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2359372 · Cited by 759
Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., et al. (2020) · JMIR mHealth and uHealth · doi:10.2196/18694 · Cited by 715
Covid-19: a remote assessment in primary care
Greenhalgh, T., Koh, G. C. H., & Car, J. (2020) · BMJ · doi:10.1136/bmj.m1182 · Cited by 705
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · cited 1×
“Furthermore, different devices have different levels of accuracy [122].”
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661
The use of photoplethysmography for assessing hypertension
Elgendi, M., Fletcher, R., Liang, Y., Howard, N., Lovell, N. H., Abbott, D., Lim, K., & Ward, R. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0136-7 · Cited by 625
Emergency Duties and Deaths from Heart Disease among Firefighters in the United States
Kales, S. N., Soteriades, E. S., Christophi, C. A., & Christiani, D. C. (2007) · New England Journal of Medicine · doi:10.1056/nejmoa060357 · Cited by 622
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589
Photoplethysmography Revisited: From Contact to Noncontact, From Point to Imaging
Sun, Y., & Thakor, N. V. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2476337 · Cited by 589
Passive Detection of Atrial Fibrillation Using a Commercially Available Smartwatch
Tison, G. H., Sanchez, J. M., Ballinger, B., Singh, A., Olgin, J. E., Pletcher, M. J., Vittinghoff, E., Lee, E. S., Fan, S. M., Gladstone, R. A., et al. (2018) · JAMA Cardiology · doi:10.1001/jamacardio.2018.0136 · Cited by 566
Primary prevention of cardiovascular disease: A review of contemporary guidance and literature
Stewart, J., Manmathan, G. P. R., & Wilkinson, P. (2017) · JRSM Cardiovascular Disease · doi:10.1177/2048004016687211 · Cited by 516
Mobile Photoplethysmographic Technology to Detect Atrial Fibrillation
Guo, Y., Wang, H., Zhang, H., Liu, T., Liang, Z., Xia, Y., Yan, L., Xing, Y., Shi, H., Li, S., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2019.08.019 · Cited by 489
Multiparameter Respiratory Rate Estimation From the Photoplethysmogram
Karlen, W., Raman, S., Ansermino, J. M., & Dumont, G. A. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2246160 · Cited by 486
Pre-symptomatic detection of COVID-19 from smartwatch data
Mishra, T., Wang, M., Metwally, A. A., Bogu, G. K., Brooks, A., Bahmani, A. H., Alavi, A., Celli, A., Higgs, E., Dagan‐Rosenfeld, O., et al. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-00640-6 · Cited by 484
An Efficient Algorithm for Automatic Peak Detection in Noisy Periodic and Quasi-Periodic Signals
Scholkmann, F., Boss, J. M., & Wolf, M. (2012) · Algorithms · doi:10.3390/a5040588 · Cited by 483
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461
Wearable sensor data and self-reported symptoms for COVID-19 detection
Quer, G., Radin, J. M., Gadaleta, M., Baca-Motes, K., Ariniello, L., Ramos, E., Kheterpal, V., Topol, E. J., & Steinhubl, S. R. (2021) · Nature Medicine · doi:10.1038/s41591-020-1123-x · Cited by 447
Schnall, P. L. (1990) · JAMA · doi:10.1001/jama.263.14.1929 · Cited by 440
Toward a Robust Estimation of Respiratory Rate From Pulse Oximeters
Pimentel, M. A. F., Johnson, A. E. W., Charlton, P. H., Birrenkott, D. A., Watkinson, P., Tarassenko, L., & Clifton, D. A. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2613124 · Cited by 429
Self-citationBlood Pressure Estimation from Photoplethysmogram Using a Spectro-Temporal Deep Neural Network
Slapničar, G., Mlakar, N., & Luštrek, M. (2019) · Sensors · doi:10.3390/s19153420 · Cited by 387
Accuracy of wrist-worn heart rate monitors
Wang, R., Blackburn, G., Desai, M., Phelan, D., Gillinov, L., Houghtaling, P., & Gillinov, M. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2016.3340 · Cited by 386
Deep PPG: Large-Scale Heart Rate Estimation with Convolutional Neural Networks
Reiss, A., Indlekofer, I., Schmidt, P., & Laerhoven, K. V. (2019) · Sensors · doi:10.3390/s19143079 · Cited by 383
Breathing Rate Estimation From the Electrocardiogram and Photoplethysmogram: A Review
Charlton, P. H., Birrenkott, D. A., Bonnici, T., Pimentel, M. A. F., Johnson, A. E. W., Alastruey, J., Tarassenko, L., Watkinson, P. J., Beale, R., & Clifton, D. A. (2018) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2017.2763681 · Cited by 379
Self-citationOrphanidou, C., Bonnici, T., Charlton, P., Clifton, D., Vallance, D., & Tarassenko, L. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2338351 · Cited by 368
Self-citationCharlton, P. H., Bonnici, T., Tarassenko, L., Clifton, D. A., Beale, R., & Watkinson, P. J. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/4/610 · Cited by 366
Self-citationContour analysis of the photoplethysmographic pulse measured at the finger
Millasseau, S., Ritter, J. M., Takazawa, K., & Chowienczyk, P. (2006) · Journal of Hypertension · doi:10.1097/01.hjh.0000239277.05068.87 · Cited by 351
Self-citationCassini, A., Colzani, E., Pini, A., Mangen, M.-J. J., Plaß, D., McDonald, S., Maringhini, G., van Lier, A., Haagsma, J. A., Havelaar, A. H., et al. (2018) · Eurosurveillance · doi:10.2807/1560-7917.es.2018.23.16.17-00454 · Cited by 350
Boriani, G., Laroche, C., Diemberger, I., Fantecchi, E., Popescu, M. I., Rasmussen, L. H., Sinagra, G., Petrescu, L., Tavazzi, L., Maggioni, A. P., et al. (2014) · The American Journal of Medicine · doi:10.1016/j.amjmed.2014.11.026 · Cited by 344
Optimal Signal Quality Index for Photoplethysmogram Signals
Elgendi, M. (2016) · Bioengineering · doi:10.3390/bioengineering3040021 · Cited by 338
Sleep and Cardio-Metabolic Disease
Cappuccio, F. P., & Miller, M. A. (2017) · Current Cardiology Reports · doi:10.1007/s11886-017-0916-0 · Cited by 323
High Prevalence of Atrial Fibrillation Among Patients With Ischemic Stroke
Friberg, L., Rosenqvist, M., Lindgren, A. G., Terént, A., Norrving, B. G., & Asplund, K. (2014) · Stroke · doi:10.1161/strokeaha.114.006070 · Cited by 319
Clinical consequences and economic costs of untreated obstructive sleep apnea syndrome
Knauert, M. P., Naik, S., Gillespie, M. B., & Kryger, M. H. (2015) · World Journal of Otorhinolaryngology - Head and Neck Surgery · doi:10.1016/j.wjorl.2015.08.001 · Cited by 311
Monte-Moreno, E. (2011) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2011.05.001 · Cited by 311
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617328 · Cited by 305
Radin, J. M., Wineinger, N. E., Topol, E. J., & Steinhubl, S. R. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30222-5 · Cited by 300
Photoplethysmography based atrial fibrillation detection: a review
Pereira, T., Tran, N., Gadhoumi, K., Pelter, M. M., Do, D. H., Lee, R. J., Colorado, R., Meisel, K., & Hu, X. (2020) · npj Digital Medicine · doi:10.1038/s41746-019-0207-9 · Cited by 300
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101870 · Cited by 300
Self-citationSeshadri, D. R., Davies, E. V., Harlow, E. R., Hsu, J. J., Knighton, S. C., Walker, T. A., Voos, J. E., & Drummond, C. K. (2020) · Frontiers in Digital Health · doi:10.3389/fdgth.2020.00008 · Cited by 300
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296
Continuous Multi-Parameter Heart Rate Variability Analysis Heralds Onset of Sepsis in Adults
Ahmad, S. S., Ramsay, T., Huebsch, L. B., Flanagan, S., McDiarmid, S. A., Batkin, I., McIntyre, L., Sundaresan, S. R., Maziak, D. E., Shamji, F. M., et al. (2009) · PLoS ONE · doi:10.1371/journal.pone.0006642 · Cited by 292
Noninvasive Assessment of the Digital Volume Pulse
Millasseau, S., Guigui, F. G., Kelly, R. P., Prasad, K., Cockcroft, J., Ritter, J. M., & Chowienczyk, P. (2000) · Hypertension · doi:10.1161/01.hyp.36.6.952 · Cited by 292
Self-citationAnalysis of heart rate variability in the presence of ectopic beats using the heart timing signal
Mateo, J., & Laguna, P. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.808831 · Cited by 287
Masimo Signal Extraction Pulse Oximetry
Goldman, J. M., Petterson, M. T., Kopotic, R. J., & Barker, S. J. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011493521730 · Cited by 283
Elgendi, M., Norton, I., Brearley, M. B., Abbott, D., & Schuurmans, D. (2013) · PLoS ONE · doi:10.1371/journal.pone.0076585 · Cited by 282
Dynamic time warping and machine learning for signal quality assessment of pulsatile signals
Li, Q., & Clifford, G. D. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/9/1491 · Cited by 274
An open source benchmarked toolbox for cardiovascular waveform and interval analysis
Vest, A. N., Poian, G. D., Li, Q., Liu, C., Nemati, S., Shah, A. J., & Clifford, G. D. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aae021 · Cited by 273
Heart Rate Estimation From Wrist-Worn Photoplethysmography: A Review
Biswas, D., Simões‐Capela, N., Hoof, C. A. V., & Helleputte, N. V. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2914166 · Cited by 273
Charlton, P. H., Harana, J. M., Vennin, S., Li, Y., Chowienczyk, P., & Alastruey, J. (2019) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00218.2019 · Cited by 269
Self-citationWong, A.-K. I., Charpignon, M., Kim, H., Josef, C., de Hond, A. A. H., Fojas, J. J., Tabaie, A., Liu, X., Mireles-Cabodevila, E., Carvalho, L., et al. (2021) · JAMA Network Open · doi:10.1001/jamanetworkopen.2021.31674 · Cited by 268
Pulse oximetry: its invention, theory, and future
Aoyagi, T. (2003) · Journal of Anesthesia · doi:10.1007/s00540-003-0192-6 · Cited by 263
Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography
Mendelson, Y., & Ochs, B. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7286 · Cited by 251
The WATCH AF trial: SmartWATCHes for detection of atrial fibrillation
Dörr, M., Nohturfft, V., Brasier, N., Bosshard, E., Djurdjevic, A., Gross, S., Raichle, C. J., Rhinisperger, M., Stöckli, R., & Eckstein, J. (2019) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2018.10.006 · Cited by 245 · cited 1×
“The potential utility of wrist-worn PPG devices for detecting possible AF has been assessed in recent studies [197], [198].”
Smartwatch Performance for the Detection and Quantification of Atrial Fibrillation
Wasserlauf, J., You, C., Patel, R., Valys, A., Albert, D., & Passman, R. (2019) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.118.006834 · Cited by 245
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244
An optimal filter for short photoplethysmogram signals
Liang, Y., Elgendi, M., Chen, Z., & Ward, R. (2018) · Scientific Data · doi:10.1038/sdata.2018.76 · Cited by 242
Said, M. A., Eppinga, R. N., Lipšic, E., Verweij, N., & van der Harst, P. (2018) · Journal of the American Heart Association · doi:10.1161/jaha.117.007621 · Cited by 241
Qiu, S., Cai, X., Sun, Z., Li, L., Zügel, M., Steinacker, J. M., & Schümann, U. (2017) · Journal of the American Heart Association · doi:10.1161/jaha.117.005505 · Cited by 240
Walch, O., Huang, Y., Forger, D., & Goldstein, C. (2019) · Sleep · doi:10.1093/sleep/zsz180 · Cited by 240
Lee, H., & Jung, C.-W. (2018) · Scientific Reports · doi:10.1038/s41598-018-20062-4 · Cited by 239
Validation of the Empatica E4 wristband
McCarthy, C., Pradhan, N., Redpath, C., & Adler, A. (2016) · 2016 IEEE EMBS International Student Conference (ISC) · doi:10.1109/embsisc.2016.7508621 · Cited by 235
HeartPy: A novel heart rate algorithm for the analysis of noisy signals
van Gent, P., Farah, H., van Nes, N., & van Arem, B. (2019) · Transportation Research Part F: Traffic Psychology and Behaviour · doi:10.1016/j.trf.2019.09.015 · Cited by 234
Hannon, N., Sheehan, O. C., Kelly, L., Marnane, M., Merwick, Á., Moore, A. R., Kyne, L., Duggan, J., Moroney, J. T., McCormack, P. M. E., et al. (2009) · Cerebrovascular Diseases · doi:10.1159/000255973 · Cited by 231
Obstructive sleep apnea, hypertension and cardiovascular diseases
Gonzaga, C., Bertolami, A., Bertolami, M., Amodeo, C., & Calhoun, D. (2015) · Journal of Human Hypertension · doi:10.1038/jhh.2015.15 · Cited by 230
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226
An Automatic Beat Detection Algorithm for Pressure Signals
Aboy, M., McNames, J., Thong, T., Tsunami, D., Ellenby, M. S., & Goldstein, B. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.855725 · Cited by 216
A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China
Liang, Y., Chen, Z., Liu, G., & Elgendi, M. (2018) · Scientific Data · doi:10.1038/sdata.2018.20 · Cited by 215
Evaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals
Mukkamala, R., Yavarimanesh, M., Natarajan, K., Hahn, J.-O., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S. (2021) · Hypertension · doi:10.1161/hypertensionaha.121.17747 · Cited by 211
The advantages of wearable green reflected photoplethysmography
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9506-z · Cited by 209
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203
Beattie, Z., Oyang, Y., Statan, A., Ghoreyshi, A., Pantelopoulos, A., Russell, A., & Heneghan, C. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa9047 · Cited by 198
Quer, G., Gouda, P., Galarnyk, M., Topol, E. J., & Steinhubl, S. R. (2020) · PLOS ONE · doi:10.1371/journal.pone.0227709 · Cited by 185
Kario, K., Matsuo, T., Kobayashi, H., Yamamoto, K., & Shimada, K. (1997) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(97)00002-8 · Cited by 184
Ding, X., Zhao, N., Yang, G., Pettigrew, R. I., Lo, B. P. L., Miao, F., Li, Y., Liu, J., & Zhang, Y. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2620995 · Cited by 184
Screening for atrial fibrillation: a call for evidence
Jones, N. R., Taylor, C. J., Hobbs, R. F., Bowman, L. J., & Casadei, B. (2019) · European Heart Journal · doi:10.1093/eurheartj/ehz834 · Cited by 182
Shah, S. A., Velardo, C., Farmer, A. A., & Tarassenko, L. (2017) · Journal of Medical Internet Research · doi:10.2196/jmir.7207 · Cited by 177
Hartmann, V., Liu, H., Chen, F., Qiu, Q., Hughes, S., & Zheng, D. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00198 · Cited by 173
Cattivelli, F. S., & Garudadri, H. (2009) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.35 · Cited by 172
Motion Artifact Removal Techniques for Wearable EEG and PPG Sensor Systems
Seok, D., Lee, S.-H., Kim, M., Cho, J., & Kim, C. (2021) · Frontiers in Electronics · doi:10.3389/felec.2021.685513 · Cited by 168
Charlton, P. H., Paliakaitė, B., Pilt, K., Bachler, M., Zanelli, S., Kulin, D., Allen, J., Hallab, M., Bianchini, E., Mayer, C. C., et al. (2021) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00392.2021 · Cited by 166
Self-citationInfluence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162
Mejía‐Mejía, E., May, J. M., Torres, R., & Kyriacou, P. A. (2020) · Physiological Measurement · doi:10.1088/1361-6579/ab998c · Cited by 160
Self-citationRemote management of covid-19 using home pulse oximetry and virtual ward support
Greenhalgh, T., Knight, M., Inada-Kim, M., Fulop, N., Leach, J., & Vindrola‐Padros, C. (2021) · BMJ · doi:10.1136/bmj.n677 · Cited by 157
Esmaili, A., Kachuee, M., & Shabany, M. (2017) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2017.2745081 · Cited by 155
University of Queensland Vital Signs Dataset
Liu, D., Görges, M., & Jenkins, S. A. (2011) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e318241f7c0 · Cited by 152
Charlton, P. H., Bonnici, T., Tarassenko, L., Alastruey, J., Clifton, D. A., Beale, R. J., & Watkinson, P. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa670e · Cited by 149
Self-citationArza, A., Garzón-Rey, J. M., Lázaro, J., Gil, E., Lopez-Anton, R., de la Camara, C., Laguna, P., Bailon, R., & Aguiló, J. (2019) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-018-1879-z · Cited by 148
Infectious Disease Surveillance in the Big Data Era: Towards Faster and Locally Relevant Systems
Simonsen, L., Gog, J. R., Olson, D., & Viboud, C. G. (2016) · The Journal of Infectious Diseases · doi:10.1093/infdis/jiw376 · Cited by 146
Mashhadi, M. B., Asadi, E., Eskandari, M., Kiani, S., & Marvasti, F. (2015) · IEEE Signal Processing Letters · doi:10.1109/lsp.2015.2509868 · Cited by 144
Taj-Eldin, M., Ryan, C., O’Flynn, B., & Galvin, P. (2018) · Sensors · doi:10.3390/s18124271 · Cited by 142
Motion Artifact Reduction for Wrist-Worn Photoplethysmograph Sensors Based on Different Wavelengths
Zhang, Y., Song, S., Vullings, R., Biswas, D., Simões-Capela, N., van Helleputte, N., van Hoof, C., & Groenendaal, W. (2019) · Sensors · doi:10.3390/s19030673 · Cited by 140
Yamakoshi, K.-I., Shimazu, H., & Togawa, T. (1980) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1980.326616 · Cited by 138
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138
Self-citationCuffless Single-Site Photoplethysmography for Blood Pressure Monitoring
Hosanee, M., Chan, G., Welykholowa, K., Cooper, R., Kyriacou, P. A., Zheng, D., Allen, J., Abbott, D., Menon, C., Lovell, N. H., et al. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9030723 · Cited by 136
Self-citationvon Wowern, E., Östling, G., Nilsson, P. M., & Olofsson, P. (2015) · PLOS ONE · doi:10.1371/journal.pone.0135659 · Cited by 134
Maximising benefit and minimising harm of screening
Gray, J. A. M., Patnick, J., & Blanks, R. G. (2008) · BMJ · doi:10.1136/bmj.39470.643218.94 · Cited by 134
Zhang, Q., Zhou, D., & Zeng, X. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-017-0317-z · Cited by 132
Assessing mental stress from the photoplethysmogram: a numerical study
Charlton, P. H., Celka, P., Farukh, B., Chowienczyk, P., & Alastruey, J. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aabe6a · Cited by 131
Self-citationMühlen, J. M., Stang, J. S., Skovgaard, E. L., Júdice, P. B., Molina‐García, P., Johnston, W., Sardinha, L. B., Ortega, F. B., Caulfield, B., Bloch, W., et al. (2021) · British Journal of Sports Medicine · doi:10.1136/bjsports-2020-103148 · Cited by 126
Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
Karlen, W., Kobayashi, K., Ansermino, J. M., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1617 · Cited by 123
PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement
Chandrasekhar, A., Yavarimanesh, M., Natarajan, K., Hahn, J., & Mukkamala, R. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.2976989 · Cited by 123
Development of a Screening Tool for Sleep Disordered Breathing in Children Using the Phone Oximeter™
Garde, A., Dehkordi, P., Karlen, W., Wensley, D., Ansermino, J. M., & Dumont, G. A. (2014) · PLoS ONE · doi:10.1371/journal.pone.0112959 · Cited by 122
Jarchi, D., & Casson, A. J. (2016) · Data · doi:10.3390/data2010001 · Cited by 120
Picone, D. S., Deshpande, R. A., Schultz, M. G., Fonseca, R., Campbell, N. R. C., Delles, C., Hecht Olsen, M., Schutte, A. E., Stergiou, G., Padwal, R., et al. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14719 · Cited by 118
Choi, A., & Shin, H. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa5efa · Cited by 116
Left Ventricular Systolic Time Intervals as Indices of Postural Circulatory Stress in Man
STAFFORD, R. W., HARRIS, W. S., & WEISSLER, A. M. (1970) · Circulation · doi:10.1161/01.cir.41.3.485 · Cited by 114
The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals
Lee, H.-W., Lee, J., Jung, W.-G., & Lee, G.-K. (2007) · International Journal of Control Automation and Systems · Cited by 114
Organic Multi-Channel Optoelectronic Sensors for Wearable Health Monitoring
Khan, Y., Han, D., Ting, J., Ahmed, M., Nagisetty, R., & Arias, A. C. (2019) · IEEE Access · doi:10.1109/access.2019.2939798 · Cited by 111
Wang, L., Zhou, W., Xing, Y., & Zhou, X. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/7804243 · Cited by 109
AffectiveROAD system and database to assess driver's attention
Elhaouij, N., Poggi, J., Sevestre-Ghalila, S., Ghozi, R., & Jaïdane-Saïdane, M. (2018) · ACM Symposium on Applied Computing · doi:10.1145/3167132.3167395 · Cited by 107
Multilevel mental stress detection using ultra-short pulse rate variability series
Zubair, M., & Yoon, C. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101736 · Cited by 106
Population-Based Screening for Atrial Fibrillation
Khurshid, S. S., Healey, J. S., McIntyre, W. F., & Lubitz, S. A. (2020) · Circulation Research · doi:10.1161/circresaha.120.316341 · Cited by 103
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102
Prevalence and treatment of atrial fibrillation in UK general practice from 2000 to 2016
Adderley, N. J., Ryan, R. P., Nirantharakumar, K., & Marshall, T. P. (2018) · Heart · doi:10.1136/heartjnl-2018-312977 · Cited by 102
Detection of a and b waves in the acceleration photoplethysmogram
Elgendi, M., Norton, I., Brearley, M. B., Abbott, D., & Schuurmans, D. (2014) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-13-139 · Cited by 100
A Noninvasive Glucose Monitoring SoC Based on Single Wavelength Photoplethysmography
Hina, A., & Saadeh, W. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.2979514 · Cited by 99
The Various Oximetric Techniques Used for the Evaluation of Blood Oxygenation
Nitzan, M., Nitzan, I., & Arieli, Y. S. (2020) · Sensors · doi:10.3390/s20174844 · Cited by 99
An Ultra Low Power Pulse Oximeter Sensor Based on Compressed Sensing
Baheti, P. K., & Garudadri, H. (2009) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.32 · Cited by 97
The vascular impact of aging andvasoactive drugs: comparison of twodigital volume pulse measurements
Millasseau, S. (2003) · American Journal of Hypertension · doi:10.1016/s0895-7061(03)00569-7 · Cited by 96
Reflectance Photoplethysmography as Noninvasive Monitoring of Tissue Blood Perfusion
Abay, T. Y., & Kyriacou, P. A. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2417863 · Cited by 96
Self-citationAn Algorithm for Real-Time Pulse Waveform Segmentation and Artifact Detection in Photoplethysmograms
Fischer, C., Dömer, B., Wibmer, T., & Penzel, T. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2518202 · Cited by 95
Detection of atrial fibrillation episodes using a wristband device
Corino, V., Laureanti, R., Ferranti, L., Scarpini, G., Lombardi, F., & Mainardi, L. T. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa5dd7 · Cited by 95
Adaptive pulse segmentation and artifact detection in photoplethysmography for mobile applications
Karlen, W., Ansermino, J. M., & Dumont, G. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346628 · Cited by 94
Ye, Y., Cheng, Y., He, W., Hou, M., & Zhang, Z. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2597265 · Cited by 94
Ebmeier, S. J., Barker, M., Bacon, M., Beasley, R. C., Bellomo, R., Chong, C. K., Eastwood, G. M., Gilchrist, J., Kagaya, H., Pilcher, J., et al. (2018) · Anaesthesia and Intensive Care · doi:10.1177/0310057x1804600307 · Cited by 92
An Ultra Low Power Personalizable Wrist Worn ECG Monitor Integrated With IoT Infrastructure
Beach, C., Krachunov, S., Pope, J., Fafoutis, X., Piechocki, R. J., Craddock, I., & Casson, A. J. (2018) · IEEE Access · doi:10.1109/access.2018.2864675 · Cited by 91
Walker, A. M., Drozd, M., Hall, M., Patel, P. A., Paton, M. F., Lowry, J. E., Gierula, J., Byrom, R. J., Kearney, L. C., Sapsford, R. J., et al. (2018) · Journal of the American Heart Association · doi:10.1161/jaha.118.009684 · Cited by 91
Estimating blood pressure trends and the nocturnal dip from photoplethysmography
Radha, M., de Groot, K., Rajani, N. B., Wong, C. C. P., Kobold, N., Vos, V., Fonseca, P. M., Mastellos, N., Wark, P. A., Velthoven, N., et al. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab030e · Cited by 90
Signal quality and patient experience with wearable devices for epilepsy management
Nasseri, M., Nurse, E., Glasstetter, M., Böttcher, S., Gregg, N. M., Laks Nandakumar, A., Joseph, B., Pal Attia, T., Viana, P. F., Bruno, E., et al. (2020) · Epilepsia · doi:10.1111/epi.16527 · Cited by 90
Papini, G. B., Fonseca, P., van Gilst, M. M., Bergmans, J. W. M., Vullings, R., & Overeem, S. (2020) · Scientific Reports · doi:10.1038/s41598-020-69935-7 · Cited by 90
Bed-Based Ballistocardiography: Dataset and Ability to Track Cardiovascular Parameters
Carlson, C., Turpin, V. G., Suliman, A., Ade, C. J., Warren, S., & Thompson, D. E. (2020) · Sensors · doi:10.3390/s21010156 · Cited by 90
Chan, G., Cooper, R. E., Hosanee, M., Welykholowa, K., Kyriacou, P. A., Zheng, D., Allen, J., Abbott, D., Lovell, N. H., Fletcher, R., et al. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8111827 · Cited by 89
Self-citationGil, E., Vergara, J. M., & Laguna, P. (2008) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2007.12.002 · Cited by 88
Zhang, Y., Weaver, R. G., Armstrong, B., Burkart, S., Zhang, S., & Beets, M. W. (2020) · Journal of Sports Sciences · doi:10.1080/02640414.2020.1767348 · Cited by 88
Vybornova, A., Polychronopoulou, E., Wurzner-Ghajarzadeh, A., Fallet, S., Solà, J., & Wuerzner, G. (2021) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000531 · Cited by 87
Bonomi, A., Schipper, F., Eerikäinen, L. M., Margarito, J., van Dinther, R., Muesch, G., de Morree, H. M., Aarts, R. M., Babaeizadeh, S., McManus, D. D., et al. (2018) · Journal of the American Heart Association · doi:10.1161/jaha.118.009351 · Cited by 86
Pankaj, Kumar, A., Komaragiri, R., & Kumar, M. (2021) · Archives of Computational Methods in Engineering · doi:10.1007/s11831-021-09597-4 · Cited by 84
Multi-Scale Peak and Trough Detection Optimised for Periodic and Quasi-Periodic Neuroscience Data
Bishop, S. M., & Ercole, A. (2018) · Acta neurochirurgica. Supplementum · doi:10.1007/978-3-319-65798-1_39 · Cited by 80
Motion artifact cancellation in wearable photoplethysmography using gyroscope
Lee, H., Chung, H., & Lee, J. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2879970 · Cited by 79
The form of the volume pulse in the finger pad in health, arteriosclerosis, and hypertension
Dillon, J. B., & Hertzman, A. B. (1941) · American Heart Journal · doi:10.1016/s0002-8703(41)90966-3 · Cited by 77
Biering-Sørensen, T., Querejeta Roca, G., Hegde, S. M., Shah, A. M., Claggett, B., Mosley, T. H., Butler, K. R., & Solomon, S. D. (2018) · European Journal of Heart Failure · doi:10.1002/ejhf.928 · Cited by 76
Sun, X., Yang, P., Li, Y., Gao, Z., & Zhang, Y. (2012) · IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI ...) · doi:10.1109/bhi.2012.6211698 · Cited by 75
Detection of c, d, and e waves in the acceleration photoplethysmogram
Elgendi, M. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2014.08.001 · Cited by 73
Market and Patent Analyses of Wearables in Medicine
Mück, J. E., Ünal, B., Butt, H., & Yetisen, A. K. (2019) · Trends in biotechnology · doi:10.1016/j.tibtech.2019.02.001 · Cited by 73
Scardulla, F., D’Acquisto, L., Colombarini, R., Hu, S., Pasta, S., & Bellavia, D. (2020) · Sensors · doi:10.3390/s20185052 · Cited by 72
Unobtrusive and Multimodal Wearable Sensing to Quantify Anxiety
Zheng, Y., Wong, T. C. H., Leung, B. H. K., & Poon, C. C. Y. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2539383 · Cited by 71
Birrenkott, D. A., Pimentel, M. A. F., Watkinson, P. J., & Clifton, D. A. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2778265 · Cited by 71
Heart Rate Monitoring in Newborn Babies: A Systematic Review
Anton, O., Alvarez, J. R. F., Rendón-Morales, E., Avilés‐Espinosa, R., Jordan, H., & Rabe, H. (2019) · Neonatology · doi:10.1159/000499675 · Cited by 71
Diagnosis of Obstructive Sleep Apnea Using Pulse Oximeter Derived Photoplethysmographic Signals
Romem, A., Romem, A., Koldobskiy, D. H., & Scharf, S. M. (2014) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.3530 · Cited by 70
Global Survey of the Frequency of Atrial Fibrillation–Associated Stroke
Perera, K. S., Vanassche, T., Bosch, J., Swaminathan, B., Mundl, H., Giruparajah, M., Barboza, M. A., O'Donnell, M. J., Schneider, M. G., Hankey, G. J., et al. (2016) · Stroke · doi:10.1161/strokeaha.116.013378 · Cited by 70
Speed of Heart Rate Recovery in Response to Orthostatic Challenge
McCrory, C., Berkman, L. F., Nolan, H., O’Leary, N., Foley, M. M., & Kenny, R. A. (2016) · Circulation Research · doi:10.1161/circresaha.116.308577 · Cited by 70
Pamula, V. R., Valero-Sarmiento, J. M., Yan, L., Bozkurt, A. Y., Hoof, C. A. V., Helleputte, N. V., Yazıcıoğlu, R. F., & Verhelst, M. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2661701 · Cited by 70
Rajala, S., Lindholm, H., & Taipalus, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac7ac · Cited by 70
Second Derivative of Photoplethysmography for Estimating Vascular Aging
Baek, H. J., Kim, J. S., Kim, Y. S., Lee, H. B., & Park, K. S. (2007) · IEEE International Conference on Information Technology and Applications in Biomedicine · doi:10.1109/itab.2007.4407346 · Cited by 69
Gunarathne, A., Patel, J. V., Hughes, E., & Lip, G. Y. H. (2008) · American Journal of Hypertension · doi:10.1038/ajh.2008.207 · Cited by 69
Modeling of the photoplethysmogram during atrial fibrillation
Sološenko, A., Petrėnas, A., Marozas, V., & Sörnmo, L. (2017) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2016.12.016 · Cited by 69
Self-citationSevere infections and subsequent delayed cardiovascular disease
Bergh, C., Fall, K., Udumyan, R., Sjöqvist, H., Fröbert, O., & Montgomery, S. (2017) · European Journal of Preventive Cardiology · doi:10.1177/2047487317724009 · Cited by 68
Hemon, M. C., & Phillips, J. P. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9695-6 · Cited by 67
Budidha, K., & Kyriacou, P. A. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/2/111 · Cited by 67
Self-citationOjha, V., Griego, D., Kuliga, S., Bielik, M., Búš, P., Schaeben, C., Treyer, L., Standfest, M., Schneider, S., Koenig, R., et al. (2018) · Information Sciences · doi:10.1016/j.ins.2018.09.061 · Cited by 67
Influence of mental stress on the pulse wave features of photoplethysmograms
Celka, P., Charlton, P., Farukh, B., Chowienczyk, P., & Alastruey, J. (2019) · Healthcare technology letters · doi:10.1049/htl.2019.0001 · Cited by 66
Self-citationvan Gilst, M. M., van Dijk, J. P., Krijn, R., Hoondert, B., Fonseca, P. M., van Sloun, R. J. G., Arsenali, B., Vandenbussche, N., Pillen, S., Maass, H., et al. (2019) · BMJ Open · doi:10.1136/bmjopen-2019-030996 · Cited by 66
Couceiro, R., Carvalho, P., Paiva, R. P., Henriques, J. M. O., & Muehlsteff, J. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/12/2369 · Cited by 64
Identification of Atrial Fibrillation by Quantitative Analyses of Fingertip Photoplethysmogram
Tang, S.-C., Huang, P.-W., Hung, C.-S., Shan, S.-M., Lin, Y.-H., Shieh, J., Lai, D., Wu, A.-Y. (., & Jeng, J. (2017) · Scientific Reports · doi:10.1038/srep45644 · Cited by 64
Detection of Respiratory Arousals Using Photoplethysmography (PPG) Signal in Sleep Apnea Patients
Karmakar, C. K., Khandoker, A. H., Penzel, T., Schöbel, C., & Palaniswami, M. S. (2013) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2282338 · Cited by 62
Aston, P. J., Christie, M. I., Huang, Y. H., & Nandi, M. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aaa93d · Cited by 62
Lee, H., Chung, H., Ko, H., & Lee, J. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2801385 · Cited by 61
Eerikäinen, L. M., Bonomi, A. G., Schipper, F., Dekker, L. R. C., Vullings, R., de Morree, H. M., & Aarts, R. M. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aad2c0 · Cited by 61
In vivo investigation of ear canal pulse oximetry during hypothermia
Budidha, K., & Kyriacou, P. A. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-9975-4 · Cited by 60
Self-citationIn-Ear Pulse Rate Measurement: A Valid Alternative to Heart Rate Derived from Electrocardiography?
Passler, S., Müller, N., & Senner, V. (2019) · Sensors · doi:10.3390/s19173641 · Cited by 60
Flagged first, then by citations
Research works printed but not linked: 9
- 2019, Cisco Visual Networking Index (VNI) Global Mobile Data Traffic Forecast Update, 2017–2022 White Paper
- 2021, Gartner Forecasts Global Spending on Wearable Devices to Total $81.5 Billion in 2021
- 10.7717/peerj.6947/table-1 · DOI
- Charlton, 2017, Continuous Respiratory Rate Monitoring to Detect Clinical Deteriorations Using Wearable Sensors
- 2015, Atrial fibrillation prevalence estimates in England: Application of recent population estimates of AF in Sweden
- 2020, What Should I Know About the Fitbit Heart Study?
- Charlton, 2021, A Study of the Acceptability and Performance of Wearables for Atrial Fibrillation Screening in Older Adults
- Petersen, 2002, Coronary heart disease statistics: Heart failure supplement
- Healey, 2000, Wearable and Automotive Systems for the Recognition of Affect From Physiology
Cited by
Sorted by citations
de Zambotti, M., Goldstein, C. A., Cook, J. D., Menghini, L., Altini, M., Cheng, P., & Robillard, R. (2023) · SLEEP · doi:10.1093/sleep/zsad325 · Cited by 276
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 233
Self-citationWang, W., Mohseni, P., Kilgore, K. L., & Najafizadeh, L. (2023) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.1090854 · Cited by 119
Zhang, Y., Yu, Y., Liu, X., Miao, J., Han, Y., Liu, J., & Wang, L. (2023) · Advanced Materials · doi:10.1002/adma.202211714 · Cited by 98
Detecting beats in the photoplethysmogram: benchmarking open-source algorithms
Charlton, P. H., Kotzen, K., Mejía‐Mejía, E., Aston, P. J., Budidha, K., Mant, J., Pettit, C., Behar, J. A., & Kyriacou, P. A. (2022) · Physiological Measurement · doi:10.1088/1361-6579/ac826d · Cited by 116
Self-citation
Metadata
- Authors
- Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J.
- Affiliations
- University of London (GB)
- University of Cambridge (GB)
- Kaunas University of Technology (LT)
- King's College London (GB)
- Corresponding author
- Charlton, Peter H.
- Venue
- Proceedings of the IEEE
- Published
- 2022-03
- Volume
- 110
- Issue
- 3
- Pages
- 355-381
- ISSN
- 0018-9219 (ISSN-L)
- 1558-2256
- PMC
- PMC7612541
- OpenAlex
- W4220840763
- Semantic Scholar
- e2397a66794ee73ff264acb1b40c4a1ceffd871b
- Cited by
- 250
- References
- 296
- Citations by year
- 2026: 108 · 2025: 62 · 2024: 37 · 2023: 33 · 2022: 10
- FWCI
- 10.32
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Proceedings of the IEEE (publishedVersion, cc-by-sa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- City Research Online (City University London) (acceptedVersion)
- Research Portal (King's College London) (submittedVersion, cc-by)
- Funders
- Wellcome Trust
- National Institute for Health and Care Research (203148/Z/16/Z)
- British Heart Foundation (WT203148/Z/16/Z, PG/15/104/31913, FS/20/20/34626, SP/14/8/31352)
- University of Cambridge
- King's College London (WT203148/Z/16/Z)
- European Cooperation in Science and Technology (CA18216)
- Lietuvos Mokslo Taryba (01.2.2-LMT-K-718, 01.2.2-LMT-K-718-01-0030)
- Engineering and Physical Sciences Research Council ([WT203148/Z/16/Z], WT 203148/Z/16/, 203148/ Z/16/Z)
- European Regional Development Fund (01.2.2-LMT-K-718, 01.2.2-LMT-K-718-01-0030)
- Centre For Medical Engineering, King’s College London (203148/Z/16/Z, WT 203148/Z/16/Z)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- cardiovascular monitoring, cardiovascular disease detection, smartwatches, fitness bands, arterial pulse wave, optical pulse measurement
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-09, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-09 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review